Drager X-am 5000 Technical Manual

04723999_01.eps
Dräger X-am® 5000
approved as type MQG 0010
Multi-Gas Monitor
Technical Manual
Content
Content
For Your Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Tests and Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Intended operating area and operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
What is What . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Front panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Rear panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Special symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Standard gas configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Standard device configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Preparations for operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Switching on the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Switching off the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Before entering the workplace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
During operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Calling the Info Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Calling the Info-Off Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Calling the Quick Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Possible functions of the quick menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Quick menu "Delete peak values" . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Calling the Calibration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Calibration menu functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Identifying Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Concentration pre-alarm A1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Concentration main alarm A2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
STEL / TWA exposure alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Battery pre-alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Battery main alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Device alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Operation with pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Configuring the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Device settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Read Database and Display Graphically . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Faults, Cause and Remedy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Warning messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Content
Fault messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
ToxicTwins . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
CO H
2
Carry out manual bump test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Manual implementation without the documentation of results in the device
memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Menu implementation with the documentation of results in the device memory . 41
Automatic implementation with the Bump Test Station . . . . . . . . . . . . . . . . . . . . 43
Calibrating the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Carrying out the fresh air calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Carrying out 1-button calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Calibrating the sensitivity for an individual measuring channel . . . . . . . . . . . . . . 51
Span calibration for CatEx . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Replacing the batteries / rechargeable batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Charging the rechargeable batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Charging with the multiple charging station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Charging with charging module and plug-in power pack or vehicle charging
adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Replacing the Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Sensor warm-up acceleration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
X-am 5000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Order List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

For Your Safety

For Your Safety
General safety statements
Before using this product, carefully read the associated Instructions for Use. This
document does not replace the Instructions for Use.
Definitions of alert icons
The following alert icons are used in this document to provide and highlight areas of the associated text that require a greater awareness by the user. A definition of the meaning of each icon is as follows:
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in physical injury, or damage to the product or environment. It may also be used to alert against unsafe practices.
NOTICE
Indicates additional information on how to use the product.

Intended Use

Portable gas detection instrument for the continuous monitoring of the concentration of several gases in the ambient air within the working area and in explosion-hazard areas.
X-am 5000, depending on the device type and configuration of DrägerSensors: independent measurement of one up to five gases.

Tests and Approvals

Tests and Approvals
Copies of the name plate and the declaration of conformity are provided in the enclosed supplementary documentation (order no. 90 33 890). The BVS 10 ATEX E 080 X technical suitability test is based on the adjustment with the target gas. Do not stick anything on the name plate on the gas detector. The technical suitability tests are valid for the X-am 5000 gas detector and the calibration cradle. The explosion-protection approvals are only valid for the X-am 5000 gas detector; the calibration cradle must not be used in the Ex zone.
Tests and Approvals

Intended operating area and operating conditions

Areas subject to explosion hazards, classified by zones
The instrument is intended for the use in explosion-hazard areas of Zone 0, Zone 1 or Zone 2 or in mines at risk due to fire damp. It is intended for use within a temperature range of -20 °C to +50 °C, and for areas in which gases of explosion groups IIA, IIB or IIC and temperature class T3 or T4 (depending on the batteries and rechargeable battery) may be present. If used in mines, the instrument is only to be used in areas known to have a low risk of mechanical impact.
Areas subject to explosion hazards, classified by divisions.
The instrument is intended for use in explosion-hazard areas according to Class I&II, Div. 1 or Div. 2 within a temperature range of -20 °C to +50 °C, and for areas where gases or dusts of groups A, B, C, D, E, F, G and temperature class T3 or T4 may be present (depending on the rechargeable battery and batteries).
WARNING:
CSA requirement: The sensitivity must be tested on a daily basis before first use with a known concentration of the gas to be measured in accordance with 25 to 50 % of the concentration limit value. The accuracy must be 0 to +20 % of the actual value. The accuracy can be corrected via calibration.

Safety instructions

WARNING
To reduce the danger of explosion, do not mix new batteries with old batteries and do not mix batteries made by different manufacturers.
WARNING
Always disconnect the device from the power pack before carrying out any maintenance operations.
WARNING
Substitution of components may impair intrinsic safety.
CAUTION
Not tested in an oxygen-enriched atmosphere (>21 % O
WARNING
High off-scale readings may indicate an explosive concentration.
).
2
Tests and Approvals
CAUTION
Use only supply units ABT 0100 (83 22 237), HBT 0000 (83 18 704) or HBT 0100 (83 22 244). Check the supply unit for approved batteries and applicable temperature class.
Note the following for CSA (Canadian Standards Association) applications:
For the CSA approval only the functions of the device component that is used to measure flammable gases are tested. The device is not approved by CSA for use in mining.
WARNING:
Before daily use, test the sensitivity with a known concentration of the applicable gas corresponding to 25 to 50% of the maximum concentration. The accuracy must be within a range of 0 to +20% of the actual value. Perform a calibration to correct the accuracy if necessary.
Only applies for class II certification. CSA standard C22.2 no. 152 does not contain any requirements for class II hazard areas. Therefore this device was not tested for class II by the CSA. The sensor may become blocked and not measure the gas correctly or not warn the user that gas measurement is not possible.
WARNING
CSA requirement: Measured values over the full scale value are displayed as an ex­plosive concentration.
NOTICE
CSA requirement: a measurement suitability test must be conducted for flammable gases.

What is What

00223999_02.eps
0
1
2
6
5
4
3
2
X-am 5000
7
OK
00323999_02.eps
2
1
4
6
3
5
456
CH4
O
2
CO
%LEL
Vol %
ppm
123
CH4 %LEL O
2
Vol % CO
ppm H
2
S
ppm
NH
3
ppm
What is What

Front panel

1 Gas entry 2 Alarm LED 3 Buzzer 4 key 5 key 6 Display 7 Tool for replacing
the sensor

Rear panel

1 IR interface 2 Fastening clip 3 Type plate 4 Charging contacts 5 Power pack 6 Serial no.

Display

for 5 measuring channels only:
1 Measured gas
display with unit
2 Measured value
display
3 Special symbols
other: 4 Measured gas
display
5 Measured gas
display with unit
6 Special symbols
The following only shows the device version with 5 measuring channels.

Special symbols

Fault message, refer to page 17 Warning message, refer to page 17 The peak value display for all measuring gases,
refer to page 17 The exposure evaluation display (TWA) for measuring gases, e.g., H
CO, refer to page 17
The exposure evaluation display (STEL) for measuring gases, e.g,. H CO, refer to page 17
The device is set to function test with gas (bump test), refer to page 40
The device is set to the fresh air calibration function, refer to page 46
The device is set to the 1-button calibration function, refer to page 47
The device is set to the single gas calibration function, refer to page 50
The function for password entry is active, refer to page 20
Battery / rechargeable battery 100 % full Battery / rechargeable battery 2/3 full Battery / rechargeable battery 1/3 full Battery / rechargeable battery empty
What is What
S and
2
S and
2
Marking offset channels:
Feature Reading on the display
ToxicTwins HCN+
compensation CO+
CO H
2

Configuration

Configuration

Standard gas configuration

DrägerSensor Measuring
range
1)
Alarm A1
1)
Alarm A2
1)
CatEx 125 PR
0 to 100 20 Yes No 40 No Yes
setpoint
can be
acknowledged
self-latching
setpoint
can be
acknowledged
[%LEL] CatEx 125 PR Gas
[%LEL] XXS O2 [vol. %] 0 to 25 19 XXS O2 100 [vol. %] 0 to 100 18.5 XXS O2 LS / CO-LC
[vol. %], [ppm] XXS O2 / H2S-LC
[Vol%], [ppm]
0 to 100 20 Yes No 40 No Yes
2)
No Ye s 23 No Yes
2)
No Ye s 24 No Yes
0 to 25 O 0 to 2000 CO
0 to 25 O 0 to 100 H2S
2
2
19 O 30 CO
19 O 5H2SNoYes
No
2
Yes
2
YesNo23 O
60 CO
YesNo23 O
10 H2SNoNo
2
2
NoNoYes
Yes Yes
Yes
XXS CO [ppm] 0 to 2,000 30 Yes No 60 No Yes XXS CO LC [ppm] 0 to 2,000 30 Ye s No 60 No Yes XXS CO-LC/H2S-LC
[ppm] XXS CO-LC/O
[ppm], [vol.-%]
2
0 to 2,000 0 to 100
0 to 2,000 CO 0 to 25 O
2
30 5
30 CO 19 O
Yes YesNoNo6010
YesNoNo
2
Yes
60 CO 23 O
2
NoNoYes
Yes
NoNoYes
Yes
XXS CO HC [ppm] 0 to 10,000 600 Ye s No 1,200 No Yes XXS CO H2-CP [ppm] 0 to 2.000 30 Yes No 60 No Yes XXS H2 [ppm] 0 to 2.000 200 Ye s No 400 No Yes XXS H2S [ppm] 0 to 200 5 Yes No 10 No Yes XXS H2S LC [ppm] 0 to 100 5 Yes No 10 No Yes XXS H2S HC [ppm] 0 to 1,000 10 Yes No 20 No Yes XXS H2S/CO [ppm] 0 to 200 H2S
XXS H2S-LC / CO-LC [ppm]
0 to 2,000 CO 0 to 100 H2S
0 to 2000 CO
5 H2S 30 CO Yes No
5 H2S 30 CO
Yes YesNoNo
10 H2S 60 CO No Yes
10 H2S 60 CONoNo
Yes Yes
XXS NO [ppm] 0 to 200 25 Yes No 50 No Yes XXS NO2 [ppm] 0 to 50 5 Ye s No 10 No Yes XXS SO2 [ppm] 0 to 100 0,5 Ye s No 1 No Yes XXS PH3 [ppm] 0 to 20 0.1 Yes No 0.2 No Yes XXS PH3 HC [ppm] 0 to 2,000 5 Ye s No 10 No Yes XXS HCN [ppm] 0 to 50 1.9 Ye s No 3.8 No Yes XXS HCN PC [ppm] 0 to 50 5 Ye s No 10 No Yes XXS NH3 [ppm] 0 to 300 20 Yes No 40 No Yes XXS CO2 [vol. %] 0 to 5 0.5 Ye s No 1 No Yes XXS Cl2 [ppm] 0 to 20 0.5 Ye s No 1 No Ye s XXS H2 HC [vol. %] 0 to 4 0.8 Ye s No 1.6 No Ye s
self-latching
10
Configuration
DrägerSensor Measuring
range
1)
Alarm A1
setpoint
can be
acknowledged
1)
self-latching
Alarm A2
setpoint
can be
acknowledged
1)
self-latching
XXS OV [ppm] 0 to 200 10 Ye s No 20 No Ye s XXS OV A [ppm] 0 to 200 10 Ye s No 20 No Ye s XXS Odorant [ppm] 0 to 40 10 Yes No 20 No Ye s XXS Amine [ppm] 0 to 100 10 Ye s No 20 No Ye s XXS COCI2 [ppm] 0 to 10 0.1 Ye s No 0.2 No Yes XXS O3 [ppm] 0 to 10 0.1 Ye s No 0.2 No Yes XXS NO2 LC [ppm] 0 to 50 0,5 Yes No 1,0 No Yes
1) Different settings can be selected to meet customer requirements on delivery. The current setting can be checked and changed with the Dräger CC Vision software. A version of Dräger CC-Vision suitable for the Dräger X-am 5000/2000 can be downloaded on the product page of the X-am 5000: www.draeger.com.
2) In the case of O
A1 is the lower alarm setpoint: an alarm is issued if the value is too low.
2
11
Configuration

Standard device configuration

NOTICE
Only trained personnel are permitted to make changes to the device configuration.
Dräger X-am® 5000
Bump test mode Fresh air calibration Operating signal Capture range ON Switch off LEL factor
2)
2)
(ch4)
2) 4) 5)
STEL (short-term average)
2) 5) 6)
TWA (shift average)
Alarm A1
7)
Alarm A1 at O
Alarm A2
1) X-am® is a registered trademark of Dräger.
2) Different settings can be selected to meet customer requirements on delivery. The current setting can
3) A periodic short signal indicates the operating capacity of the instrument. If there is no operating signal,
4) STEL: average value of an exposure over a short period, generally 15 minutes.
5) Interpretation only if the sensor is designed for this.
6) TWA: shift averages are workplace limit values for generally eight hours per day of exposure for five
7) Latching and acknowledgement of alarms A1 and A2 can be configured with the Dräger CC Vision PC
7)
be checked and changed with the Dräger CC Vision software.
correct operation cannot be guaranteed.
days a week during a working life.
software.
Changing the standard configuration: See “Configuring the Device” on page 26.
2)
2) 3)
sensor
2
1)
Extended bump test
2)
ON ON
allowed
4.4 (vol. %) (4.4 vol. % corresponds to 100 %LEL)
STEL function - disabled Average value duration = 15 minutes
TWA function - disabled Average value duration = 8 hours
can be acknowledged, non-latching, pre-alarm, rising flank
7)
cannot be acknowledged, latching, like main alarm, falling flank
cannot be acknowledged, latching, main alarm, rising flank
WARNING
After a basic initialization has been carried out with the PC software Dräger CC Vision, individual alarm settings may have been changed.
Selecting or disabling the capture ranges (only applies for the measuring mode):
The capture range is selected in the measuring mode (factory setting) and permanently disabled in calibration mode. The CC-Vision PC software can be used to select or disable the capture ranges for the measuring mode.
12

Operation

OK
Operation

Preparations for operation

Before using the instrument for the first time, insert a charged NiMH T4 power pack
or batteries approved by Dräger, see "Changing the batteries" on page 55.
The instrument is now ready for operation.
WARNING
To reduce the risk of ignition of a flammable or explosive atmosphere, strictly adhere to the following warning statements:
Only use power pack types ABT 01xx, HBT 00xx or HBT 01xx. See the marking on the rechargeable battery for permitted rechargeable batteries and the corresponding temperature class.
Substitution of components may impair intrinsic safety.

Switching on the device

Press and hold the key for approx. 3 seconds until the countdown »3.2.1«
shown in the display has elapsed.
All the display segments, including the visual, audible and vibration alarms, are
activated for a short time.
The software version is displayed.The device performs a self test.The next sensor which is next due for calibration is displayed with the days
remaining until the next calibration, e.g., » Ex %LEL CAL 20 «.
 The time until the bump test interval elapses is displayed in days, e.g., » bt 123 «.  All A1 and A2 alarm thresholds and »« (TWA)
gases (e. g. H
During the sensor warm-up phase:
The display for the measured value flashes The special symbol »« is displayed.No alarms are issued during the warm-up phase.The red LEDs flash.The gas detector is ready to measure when the measured values no longer flash
and the red LEDs are no longer illuminated. The special symbol »« may con­tinue to be displayed if corresponding warnings (e.g. not yet ready for calibration) are active (to view the warnings, see the technical manual).
Press the key to cancel the display of the activation sequence.
OK
S or CO) are displayed consecutively.
2
1)
and »« (STEL)1) for all toxic
1) Only when activated in the instrument configuration. Delivery condition: not activated.
13
Operation

Switching off the device

Press and hold the key and key at the same time until the countdown
»3.2.1« shown in the display has elapsed.
Before the device is switched off, the visual, audible and vibration alarms are activated for a short time.
OK

Before entering the workplace

WARNING
Before any measurements relevant to safety are made, check the adjustment with a bump test, adjust if necessary and check all alarm elements. If national regulations apply, a bump test must be performed according to the national regulations. Faulty adjustment may result in incorrect measuring results, with possible serious consequences.
WARNING
In an oxygen enriched atmosphere (>21 vol. % O be guaranteed; remove instrument from the explosion-hazard area.
CAUTION
The CatEx sensor is intended for measurements of flammable gases and vapours mixed with air (i.e. O
content 21 vol.%). Incorrect measured values may be displayed
2
in the case of oxygen deficient or oxygen enriched environments.
NOTICE
If the gas detector is used for offshore applications, a distance of 5 m to a compass must be complied with.
), the explosion protection cannot
2
Switch on the device. The current measured values are shown in the display. Observe any warning » « or fault messages » «.
The device can be operated normally. If the warning message does not go out automatically during operation, the device must be maintained after the end of use.
The device is not ready to measure and requires maintenance.
If one of these special symbols is displayed, appropriate measures, refer to page 30
to page 33, must be taken.
14
Operation
Check that the gas inlet opening on the device is not covered.
WARNING
Explosion hazard! To reduce the risk of flammable or explosive atmospheres igniting, it is essential that the warning notices below are observed: Fractions of catalytic poisons in the measuring gas (e.g. volatile silicon, sulphur,
heavy metal compounds or halogenated hydrocarbon) can damage the Cat Ex sensor. If the CatEx sensor can no longer be calibrated to the target concentration, the sensor must be replaced.
In case of measurements in oxygen-deficient atmosphere (<12 vol. % O
CatEx sensor may show incorrect displays; in this case, a reliable measurement
) the
2
with a CatEx sensor is not possible.
In an oxygen enriched atmosphere (>21 vol. % O
cannot be guaranteed; remove instrument from the Ex area.
), the explosion protection
2
High values outside the display area indicate an explosive concentration where
applicable.
15
Operation

During operation

During operation, the measured values for every measured gas are displayed.In the event of an alarm, the corresponding displays, the visual, audible and vibration
alarms are activated – see chapter “Identifying Alarms”.
If a measuring range is exceeded or a negative drift occurs, the following displays
are shown instead of the measured value display:
»«(concentration too high) or »«(measuring range not reached) or
»«
If an O2 sensor is fitted and this sensor measures an O2 concentration of below
After the measuring range of the TOX measuring channels has been exceeded
In the event of an alarm, the corresponding displays, including the visual, audible and vibration alarms, are activated – see “Identifying Alarms” on page 21.
Special states in which there is no measuring operation (quick menu, calibration menu, warm-up of sensors, password input) are indicated by a visual signal (slow flashing of the alarm LED ).
(blocking alarm).
12 vol. %, an error is indicated with » « on the ex-channel instead of a measured value if the measured value is below the pre-alarm threshold (not with the setting CH4 with measuring range >100 %LEL).
temporarily (up to one hour), checking the measuring channels is not necessary.
NOTICE
If the measuring range is exceeded significantly on the CatEx channel (very high concentration of flammable materials), a blocking alarm is triggered. This CatEx blocking alarm can be acknowledged manually by switching the device off and back on again in fresh air. In configuration setting CH triggered as the heat conductance measurement principle is used.
The measuring range 0 to 100 vol. % CH4 is not suitable for monitoring explosive mixtures in the measuring range from 0 to 100%LEL.
If the DrägerSensor CatEx 125 PR is used in the gas detector, a zero point and sensitivity calibration must be carried out after experiencing an impact load that results in a non-zero display when exposed to fresh air. This warning does not apply if the DrägerSensor CatEx 125 PR Gas is used.
16
with measuring range 100 vol. %, no blocking alarm is
4
CAUTION
WARNING
Operation
OK
OK
OK
OK
OK
OK
OK

Calling the Info Mode

In measuring mode, press the key for approx. 3 seconds.
If any warning or fault messages exist, the corresponding information or error codes are displayed (page 30 to page 37). Press the key successively for the next display. The peak values and the exposition values TWA
Warning messages are displayed. Numerical codes of warning messages: see page 30.
key
Fault messages are displayed. Numerical codes of fault messages: see page 33.
key
The peak values = the maximum measured values in the case of, e.g., CO, H
S, ... or the minimum measured values in the case of O2 within the storage
2
interval are displayed
key
The average values of the exposures based on a shift of, e.g., 8 hours (TWA) of all the active sensors for the exposure evaluation are displayed
key
The short-term values (STEL) = average values of the concentrations over the average value duration of all the active sensors for the exposure evaluation are displayed
key
The device is in measuring mode again
OK
1)
and STEL1) are displayed.
If no key is pressed for 10 seconds, the device returns automatically to measuring
mode.

Calling the Info-Off Mode

When the device is in a deactivated state, press the key. The name of the gas, measuring unit and measuring range limit value are displayed for all channels.
Pressing the key again exits the Info Off mode (or via timeout).
1) Only when activated in the device configuration. Delivery status: not activated.
17
Operation

Calling the Quick Menu

Only the fresh air calibration is activated in the quick menu on delivery. The PC
software Dräger CC Vision can be used to activate the bump test for the quick menu and/or the function for displaying and deleting peak values.
In measuring mode, press the key three times.
If no functions have been activated in the quick menu, the device remains in measuring mode.
You can select the activated functions of the quick menu by pressing the key.
Press the key to call the selected function.
OK

Possible functions of the quick menu

Bump test, refer to page 40
Fresh air calibration, refer to page 46
Delete peak values, refer to page 19
Display pump information, refer to page 23
Activate or deactivate pump, refer to page 23
Press the key to cancel the active function and to switch to measuring mode.
If no key is pressed for 60 seconds, the device returns automatically to measuring
mode.
18
Quick menu "Delete peak val-
04823999_01_en.eps
ch4 %LEL O
2
Vol% CO
ppm H
2
S
ppm
NH
3
ppm
04923999_01_en.eps
ch4 %LEL O
2
Vol% CO
ppm H
2
S
ppm
NH
3
ppm
OK
OK
ues"
After the function has been selected, the current peak values are displayed; the peak values special symbol appears in the display at the same time.
The peak values can be deleted by
pressing the key for 5 sec. The adjacent display appears.
Operation
Press the key to end the function.
19
Operation
02323999_01.eps
OK

Calling the Calibration Menu

The calibration menu can only be accessed by entering a password.
Password on delivery: »001«
The default password on delivery can be changed using the PC software
Dräger CC Vision.
In measuring mode, press the key for at least 4 seconds.The function for entering the password is selected.The » « special symbol (for the "enter password" function) is displayed.
The display shows »000«, with the
first digit flashing.
Use the key to set the flashing digit.
Press the key, the second digit
starts flashing.
Use the key to set the flashing digit.
Press the key, the third digit starts
flashing.
Use the key to set the flashing digit.
Press the key to confirm the
password once it has been set completely.
The calibration menu functions can
now be selected by pressing the key.
Press the key to call the selected
function.
OK
OK
OK

Calibration menu functions

Fresh air calibration, refer to page 46
1-button calibration, refer to page 47
Single gas calibration, refer to page 50
Press the key to cancel the active function.
If no key is pressed for 10 minutes, the device automatically returns to measuring
mode.
20

Identifying Alarms

OK
OK
Identifying Alarms
An alarm is displayed visually, audibly and through vibration in a specific pattern.
NOTICE
At low temperatures the legibility of the display can be improved by switching on the backlight.

Concentration pre-alarm A1

The alarm is indicated by an intermittent alarm message: Display »A1« and measured value alternating: not for O
The pre-alarm A1 is not self-latching and stops when the concentration has dropped
below the alarm setpoint A1.
In the case of A1 a single tone is audible and the alarm LED flashes. Acknowledging the pre-alarm: Press the key. Only the audible alarm and the vibration alarm are switched off.

Concentration main alarm A2

The alarm is indicated by an intermittent alarm message: Display »A2« and measured value alternating:
In the case of A2 a double tone is audible and the alarm LED flashes twice. For O
:»A1« and measured value alternating = oxygen deficiency
2
»A2« and measured value alternating = oxygen surplus
!
2
DANGER
Leave the area immediately. Danger to life! A main alarm is self-latching and cannot be acknowledged or cancelled.
After leaving the area, if the concentration is less than the alarm setpoint A2: Press the key. The alarm messages are switched off. If the measuring range is exceeded significantly on the CatEx channel (very high concentration of flammable materials), a blocking alarm is triggered. This CatEx blocking alarm can be acknowledged manually by switching the device off and back on again in fresh air. In configuration setting CH triggered as the heat conductance measurement principle is used.
with measuring range 100 vol. %, no blocking alarm is
4
CAUTION
The measuring range 0 to 100 vol. % CH mixtures in the measuring range from 0 to 100 %LEL.
is not suitable for monitoring explosive
4
21
Identifying Alarms

STEL / TWA exposure alarm

The alarm is indicated by an intermittent alarm message: Display »A2« and »« (TWA) or » « (STEL) and measured
value alternating:
CAUTION
Health hazard! personnel is subject to the relevant national regulations.
STEL and TWA alarms cannot be acknowledged or canceled. Switch off the device. The values for the exposure evaluation are deleted after the
device is switched on again.
The STEL alarm can be triggered with a maximum delay of one minute.
Leave the area immediately. After this alarm, the deployment of
NOTICE

Battery pre-alarm

The alarm is indicated by an intermittent alarm message: Flashing special symbol »« on the right side of the display:
Acknowledging the pre-alarm:
Press the key. Only the audible alarm and the vibration alarm are switched off.The battery still lasts approx. 20 minutes after the first battery pre-alarm.
OK

Battery main alarm

The alarm is indicated by an intermittent alarm message: Flashing special symbol »« on the right side of the display:
The battery main alarm cannot be acknowledged or canceled:
The device is automatically switched off again after 10 seconds.Before the device is switched off, the visual, audible and vibration alarms are
activated for a short time.

Device alarm

The alarm is indicated by an intermittent alarm message: Special symbol » « on the right side of the display:
The device or one or several sensor channels are not ready for operation.For remedies, refer to page 30 to page 37. If necessary, commission the Dräger Service Center to eliminate the error.
22
Loading...
+ 50 hidden pages