Dräger X-am 2500 Technical Manual

Dräger X-am 2500 (MQG 0011)
Technical Manual
Content
Content
1.1 General safety statements . . . . . . . . . . . . . . . . . . . .4
1.2 Definitions of alert icons . . . . . . . . . . . . . . . . . . . . . .4
2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1 Product overview . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1.1 Front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1.2 Rear side . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1.3 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1.4 Special symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.3 Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.3.1 Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . .5
3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
3.1 Preparations for operation . . . . . . . . . . . . . . . . . . . .6
3.1.1 Charging the batteries . . . . . . . . . . . . . . . . . . . . . . . .6
3.1.2 Replacing the batteries / rechargeable batteries . . .7
3.1.3 Switching on the instrument . . . . . . . . . . . . . . . . . . .8
3.1.4 Switching off the instrument . . . . . . . . . . . . . . . . . . .8
3.2 Before entering the workplace . . . . . . . . . . . . . . . . .8
3.3 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
3.3.1 Standard gas configuration . . . . . . . . . . . . . . . . . . . .9
3.3.2 Standard instrument configuration . . . . . . . . . . . . .10
3.3.3 Configuring the device . . . . . . . . . . . . . . . . . . . . . .10
3.3.4 Export data memory and display graphically . . . . .11
3.4 Running the bump test . . . . . . . . . . . . . . . . . . . . . .11
3.4.1 Manual implementation without documentation
of the results in the instrument memory . . . . . . . . .11
3.4.2 Menu implementation with the documentation of
results in the instrument memory . . . . . . . . . . . . . .12
3.4.3 Automatic implementation with the
Bump Test Station . . . . . . . . . . . . . . . . . . . . . . . . .13
3.5 During operation . . . . . . . . . . . . . . . . . . . . . . . . . . .13
3.6 Identifying alarms . . . . . . . . . . . . . . . . . . . . . . . . . .14
3.6.1 Concentration pre-alarm A1 . . . . . . . . . . . . . . . . . .14
3.6.2 Concentration main alarm A2 . . . . . . . . . . . . . . . . .14
3.6.3 STEL/TWA exposure alarm . . . . . . . . . . . . . . . . . .14
3.6.4 Battery pre-alarm . . . . . . . . . . . . . . . . . . . . . . . . . .14
3.6.5 Battery main alarm . . . . . . . . . . . . . . . . . . . . . . . . .14
3.6.6 Instrument alarm . . . . . . . . . . . . . . . . . . . . . . . . . . .14
4 Menu functions . . . . . . . . . . . . . . . . . . . . . . . . . . .15
4.1 Activating the Info mode . . . . . . . . . . . . . . . . . . . . .15
4.2 Opening Info-Off Mode . . . . . . . . . . . . . . . . . . . . . .15
4.3 Quick Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
4.3.1 Quick menu functions . . . . . . . . . . . . . . . . . . . . . . .15
4.3.2 Opening the Quick Menu . . . . . . . . . . . . . . . . . . . .15
4.3.3 Quick menu "Delete peak values" . . . . . . . . . . . . . .15
4.4 Calibration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . .15
4.4.1 Calibration menu functions . . . . . . . . . . . . . . . . . . .15
4.4.2 Open the Calibration Menu . . . . . . . . . . . . . . . . . . .15
5 Calibrate instrument . . . . . . . . . . . . . . . . . . . . . . .16
5.1 Adjustment interval . . . . . . . . . . . . . . . . . . . . . . . . .16
5.2 Run fresh air calibration . . . . . . . . . . . . . . . . . . . . .16
5.3 Automatic fresh air calibration of the CatEx
sensor in the charging cradle . . . . . . . . . . . . . . . . .17
5.4 1-button calibration . . . . . . . . . . . . . . . . . . . . . . . . 17
5.4.1 Calibrating the sensitivity for an individual
measuring channel . . . . . . . . . . . . . . . . . . . . . . . . 18
5.4.2 Sensitivity calibration for CatEx . . . . . . . . . . . . . . . 18
5.5 Calibration with the X-dock maintenance
station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6 Operation with pump . . . . . . . . . . . . . . . . . . . . . 19
6.1 Performing a measurement with the Dräger
X-am Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.2 Performing a measurement with a manual pump
adapter and rubber ball pump . . . . . . . . . . . . . . . . 20
7 Replacing the sensors . . . . . . . . . . . . . . . . . . . . 20
8 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . 21
8.1 Warning messages . . . . . . . . . . . . . . . . . . . . . . . . 21
8.2 Fault message . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
9 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
9.1 Maintenance table . . . . . . . . . . . . . . . . . . . . . . . . . 26
10 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
11 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
12 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
13 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
13.1 X-am 2500 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
14 Order list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Dräger X-am 2500 3
For your safety
!
!
i
i
00133366.eps
0
1
2
6
5
4
3
2
X-am 2500
7
OK
00233366.eps
2
1
4
6
3
5
00333366_en.eps
123
ch4
O
2
CO
123
ch4 %LEL
0.0
O2 Vol%
20.9
CO
ppm
0.0
H2S
ppm
0.0
%LEL
0.0
Vol%
20.9
ppm
0.0

1 For your safety

1.1 General safety statements

Before using this product, carefully read the associated
Instructions for Use. This document does not replace the Instructions for Use.

1.2 Definitions of alert icons

The following alert icons are used in this document to provide and highlight areas of the associated text that require a greater awareness by the user. A definition of the meaning of each icon is as follows:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in physical injury, or damage to the product or environment. It may also be used to alert against unsafe practices.
NOTICE
Indicates additional information on how to use the product.

2.1.2 Rear side

1
IR interface
2
Fastening clip
3
Nameplate
4
Charging contacts
5
Power pack
6
Serial no.

2.1.3 Display

2 Description

2.1 Product overview

2.1.1 Front

Left: 4 measuring channels, right: 3 measuring channels
1
Measured gas display
2
Measuring value display with unit
3
Special symbols
The following only shows the instrument version with 4 measuring channels.
1
Gas entry
2
Alarm LED
3
Horn
4
key
5
key
6
Display
7
Tool for changing sensor
4 Dräger X-am 2500

2.1.4 Special symbols

Fault message, see section 4.1 on page 15 Warning message, see section 4.1 on page 15 Display of peak values for all measured gases,
see section 4.1 on page 15 The exposure evaluation display (TWA) for measured
gases, e.g. H The exposure evaluation display (STEL) for measured
gases, e.g. H
S and CO, see section 4.1 on page 15
2
S und CO, see section 4.1 on page 15
2
Description
!
!
The instrument is set to the bump test function, see section 3.4 on page 11
The instrument is set to the fresh air calibration function, see section 5.2 on page 16
The instrument is set to the 1-button calibration/ adjustment function, see section 5.4 on page 17
The instrument is set to the single gas calibration function, see section 5.4.1 on page 18
Function for password input is active, see section 4.4 on page 15
Battery / rechargeable battery 100 % full Battery / rechargeable battery 2/3 full Battery / rechargeable battery 1/3 full Battery / rechargeable battery empty

2.2 Intended use

Portable gas detection instrument for the continuous monitoring of the concentration of several gases in the ambient air within the working area and in explosion-hazard areas. Independent measurement of up to 4 gases, in accordance with the installed Dräger sensors.
Areas subject to explosion hazards, classified by zones
The instrument is intended for the use in explosion-hazard areas of Zone 0, Zone 1 or Zone 2 or in mines at risk due to fire damp. It is intended for use within a temperature range of ­20 °C to +50 °C, and for areas in which gases of explosion groups IIA, IIB or IIC and temperature class T3 or T4 (depending on the batteries and rechargeable battery) may be present. If used in mines, the instrument is only to be used in areas known to have a low risk of mechanical impact.
Areas subject to explosion hazards, classified by divisions.
The instrument is intended for use in explosion-hazard areas according to Class I&II, Div. 1 or Div. 2 within a temperature range of -20 °C to +50 °C, and for areas where gases or dusts of groups A, B, C, D, E, F, G and temperature class T3 or T4 may be present (depending on the rechargeable battery and batteries).
NOTICE
i
i
CSA requirement: Only the combustible gas detection portion of this instrument has been assessed for performance. The instrument has not been classified by the CSA for use in mines.

2.3 Approvals

Copies of the name plate and the declaration of conformity are provided in the enclosed supplementary documentation (order no. 90 33 890). Do not stick anything on the name plate on the gas detector. The technical approvals are valid for the X-am 2500 gas detection instrument and the calibration cradle. The explosion­protection approvals are only valid for the X-am 2500 gas detection instrument; the calibration cradle must not be used in the Ex zone. The BVS 10 ATEX E 080 X technical suitability test is based on the calibration with the target gas.

2.3.1 Safety Instructions

WARNING
!
Do not replace or charge batteries in potentially explosive areas. Explosion hazard!
Charge the NiMH power pack T4 (type HBT 0000) or T4 HC (type HBT 0100) with the associated Dräger charger. Charge NiMH single cells for ABT 0100 battery holder in accordance with the manufacturer's specifications. Ambient temperature during the charging process: 0 to +40 °C.
To reduce the danger of explosion, do not mix new batteries with old batteries and do not mix batteries made by different manufacturers.
Always disconnect the instrument from the power pack before carrying out any maintenance operations.
WARNING
CSA requirement: Measured values over the full scale value may indicate an explosive atmosphere.
Only applicable to Class II certification. CSA Std. C22.2 No 152 does not have any requirement for Class II hazardous locations and therefore this device has not been performance tested by CSA for Class II. The sensor may become clogged and not detect gas properly or warn the user of its inability to detect gas.
WARNING
CSA requirement: The sensitivity must be tested on a daily basis before first use with a known concentration of the gas to be measured in accordance with 25 to 50 % of the concentration limit value. The accuracy must be 0 to +20 % of the actual value. The accuracy can be corrected via calibration.
Substitution of components may impair intrinsic safety.
Only use power packs ABT 0100 (order no. 83 22 237), HBT 0000 (order no. 83 18 704) or HBT 0100 (order no. 83 22 244). See marking on power pack for approved batteries and related temperature classes.
Not tested in an oxygen-enriched atmosphere (>21 % O
High off-scale readings may indicate an explosive concentration.
).
2
Dräger X-am 2500 5
Operation
!
!
!
i
i
02733366.eps
0 0 0
Ex %UEG CO
2
ppm
O
2
Vol% CO ppm H
2S
ppm
Ex %UEG CO
2
ppm
O
2
Vol% CO ppm H
2S
ppm
Ex %UEG CO
2
ppm
O
2
Vol% CO ppm H
2S
ppm
X-am 2500
X-am 2500 X-am 2500
83 25 736 83 16 639
83 21 849 / 83 21 850
2
1
2
1
2
1
1
3
3
3
Note the following for CSA (Canadian Standards Association) applications:
For the CSA approval only the functions of the device component that is used to measure flammable gases are tested. The device is not approved by CSA for use in mining.
WARNING
Before daily use, test the sensitivity with a known concentration of the applicable gas corresponding to 25 to 50 % of the maximum concentration. The accuracy must be within a range of 0 to +20 % of the actual value. Perform a calibration to correct the accuracy if necessary.

3 Operation

3.1 Preparations for operation

WARNING
To reduce the risk of ignition of a flammable or explosive atmosphere, strictly adhere to the following warning statements: Only use power pack types ABT 01xx, HBT 00xx or HBT 01xx. See the marking on the rechargeable battery for permitted rechargeable batteries and the corresponding temperature class.
The charging time is typically 4 hours.A new NiMH power pack reaches its full capacity after three
complete charging/discharging cycles.
Never store the instrument for extended periods without
being connected to a power source (maximum of 2 months) because the internal buffer battery will drain.
Charging with the charging module adapter and the power supply unit 83 21 849 or 83 21 850
A maximum of 5 instruments in charging modules (order
no. 83 18 639) can be charged at the same time on the charging module adapter (order no. 83 25 736) in connection with the power supply unit (order no. 83 21 849). Up to 20 instruments can be charged at the same time with the power supply unit 83 21 850.
Before attaching the charging modules to the charging
module adapter, disconnect the power supply unit from the mains supply!
CAUTION
!
Always connect or disconnect the charging modules individually and never in groups in order to prevent the charging module adapter from becoming damaged. Even during transportation, always handle the power supply unit and the charging modules individually and without instruments inserted.
Substitution of components may impair intrinsic safety.
Before using the instrument for the first time, insert a
charged NiMH T4 power pack or batteries approved by Dräger see section 3.1.2 on page 7.
The instrument is now ready for operation.

3.1.1 Charging the batteries

WARNING
Explosion hazard! To reduce the risk of flammable or explosive atmospheres igniting, it is essential that the warning notices below are observed: Do not charge underground or in explosion hazard areas!
The chargers are not designed in accordance with the regulations for fire damp and explosion protection.
Charge the NiMH power pack T4 (type HBT 0000) or T4 HC (type HBT 0100) with the associated Dräger charger. Charge NiMH single cells for ABT 0100 battery holder in accordance with the manufacturer's specifications. Ambient temperature during the charging process: 0 to +40 °C.
NOTICE
Even if the instrument is not used, Dräger recommends storing the instrument in the charging cradle (chargingmodule X-am 1/2/5000, order no. 83 18 639).
To maintain the lifetime of the batteries, charging is
temperature controlled and only performed in a temperature range of 5 to 35 °C. When outside this temperature range, the charging automatically interrupted and automatically recommenced after the temperature range has been reached again.
Position the instrument on an even and level surface.
1. Turn the slots of the interlock into a horizontal position by using a screwdriver or coin.
2. Insert the fastening lug (2) of the charger module (simultaneous power feed) until it engages.
3. Close the lock (1) with a quarter turn (slot is positioned vertically).
4. Attach additional charging modules in the same way.
5. Connect the power pack to the mains.
The green LED (3) lights up.
6. Insert the switched off instrument into the charger module.
Display LED (3) on the charger module:
Charging
Fault
Full
If a fault occurs:
Remove the instrument from the charging module and
insert it again.
If the fault still occurs, have the charging module repaired.
It takes approx. 4 hours to fully charge an empty
rechargeable battery.
6 Dräger X-am 2500
Operation
!
i
i
02833366.eps
83 16 994 (100 ... 240 V) 83 16 997 (100 ... 240 V)
45 30 057
0
Ex %UEG CO
2
ppm
O
2
Vol% CO ppm H
2S
ppm
X-am 2500
CAUTION
A short circuit of the charging contacts in the charging modules, e. g., by metallic objects that have fallen in, does not result in damage to the charging station. It should, however, be avoided due to possible heating hazards and incorrect displays on the charger module.
NOTICE
If combining different charging modules, follow the instructions in the manual supplied with the charging module adapter.
Charge using charger module and plug-in power pack or vehicle charging adapter
When using the power supply unit (order no. 83 16 994),
up to 5 instruments can be charged at the same time, with the power supply unit (order no. 83 16 997) one instrument individually.
When using the vehicle charging adapter (order no.
45 30 057) it is recommended that you supply every charging module separately.
The charging process is carried out analogous to charging
with the multiple charging station.

3.1.2 Replacing the batteries / rechargeable batteries

WARNING
!
Explosion hazard! To reduce the risk of flammable or explosive atmospheres igniting, it is essential that the warning notices below are observed:
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in areas at risk of an explosion hazard.
Do not mix new batteries with used batteries, and do not mix batteries from different manufacturers or of different types.
Remove batteries before maintenance work.
Batteries / rechargeable batteries are part of the Ex approval. Only the following types may be used:
Alkaline batteries – T3 – (non rechargeable!)Panasonic LR6 PowerlineVarta Type 4106 Varta Type 4006 Alkaline batteries – T4 – (non rechargeable!)Duracell Procell MN1500
MN1500
NiMHy rechargeable batteries – T3 – (rechargeable)GP 180AAHC
1)
1)
(power one) or
1
(industrial)
1)
, Duracell Plus Power
1
(1800 mAh) max. 40 °C ambient
temperature.
Charge the NiMH power pack T4 (type HBT 0000) or T4 HC (type HBT 0100) with the associated Dräger charger. Charge NiMH rechargeable batteries for battery holder ABT 0100 in accordance with the manufacturer's specifications. Ambient temperature during the charging process: 0 to +40 °C.
1) Not part of the measurement performance tests BVS10 ATEX E 080X and PFG 10 G 001X.
Dräger X-am 2500 7
Operation
02633366.eps
1
2
3
+
+
OK
1. Switching off the instrument: key and key are held down simultaneously.
2. Loosen the screw (2.0 mm hexagon socket) on the power pack and remove the power pack.
With battery holder (order no. 83 22 237): Replace alkaline
batteries or NiMHy rechargeable batteries. Ensure correct polarity.
With the T4 NiMH power pack (type HBT 0000) / T4 HC
(type HBT 0100): Completely replace the power pack.
3. Insert the power pack into the instrument and tighten the screw, the instrument switches on automatically.
After replacing the T4 NiMH power pack (type HBT 0000)/ T4 HC (type HBT 0100), a full charge is recommended.
After the batteries have been replaced:
The settings and data are stored when the battery is replaced.
The sensors warm up again.

3.1.3 Switching on the instrument

1. Hold down the [OK] button for approx. 3 seconds until the »3.2.1« countdown shown on the display has elapsed.
All the display segments, including the visual, audible
and vibration alarms, are activated for a short time.
The software version is displayed.The instrument performs a self-test.The sensor that is up next for calibration/adjustment is
displayed with the remaining days until the next calibration/adjustment e. g. »Ex%LEL CAL20«.
The time until the bump test interval elapses is
displayed in days, e.g. »bt123«.
All A1 and A2 alarm thresholds and »« (TWA)
»« (STEL)
1
for all toxic gases (e. g. H2S or CO) are
1)
and
displayed consecutively.
During the sensor warm-up phase:
The display for the measured value flashes The special symbol » « is displayed.No alarms are issued during the warm-up phase.The red LEDs flash.The gas detector is ready to measure when the
measured values no longer flash and the red LEDs are no longer illuminated. The special symbol » « may continue to be displayed if corresponding warnings (e.g. not yet ready for calibration) are active.
2. Press the [OK] key to cancel the display of the activation sequence.
1) Only when activated in the instrument configuration. Delivery condition: not activated.

3.1.4 Switching off the instrument

Press and hold the [OK] key and [+] key simultaneously
until the countdown »3.2.1« shown on the display has elapsed. When the instrument is switched off, the visual, audible and vibration alarms are activated for a short time in order to protect against inadvertent switch off.

3.2 Before entering the workplace

WARNING
!
Before any measurements relevant to safety are made, check the adjustment with a bump test, adjust if necessary and check all alarm elements. If national regulations apply, a bump test must be performed according to the national regulations. Faulty adjustment may result in incorrect measuring results, with possible serious consequences.
CAUTION
!
The CatEx sensor is intended for measurements of flammable gases and vapours mixed with air (i.e. O
content 21 vol.%). Incorrect measured values may be displayed in the case of oxygen deficient or oxygen enriched environments.
NOTICE
i
i
If the gas detector is used for offshore applications, a distance of 5 m to a compass must be complied with.
1. Switch on the instrument. The current measured values are shown in the display.
2. Observe any warning »« or fault messages »«.
The instrument can be operated normally. If the warning message does not disappear automatically during operation, the instrument must be serviced after the end of use.
The instrument is not ready to measure and requires maintenance.
3. Check that the gas inlet opening on the instrument is not covered.
WARNING
!
Explosion hazard! To reduce the risk of flammable or explosive atmospheres igniting, it is essential that the warning notices below are observed:
Fractions of catalytic poisons in the measuring gas
(e.g. volatile silicone, sulphur, heavy metal compounds or halogenated hydrocarbon) can damage the CatEx sensor. If the CatEx sensor can no longer be calibrated to the target concentration, the sensor must be replaced.
In case of measurements in an oxygen-deficient
atmosphere (<12 Vol.-% O may show incorrect displays; in this case, a reliable
) the CatEx sensor
2
measurement with a CatEx sensor is not possible.
In an oxygen enriched atmosphere (>21 vol. %
O
), the explosion protection cannot be
2
guaranteed; remove instrument from the Ex area.
High values outside the display area indicate an
explosive concentration where applicable.
2
8 Dräger X-am 2500

3.3 Configuration

3.3.1 Standard gas configuration

Operation
DrägerSensor Measuring range
1)
threshold
Alarm A1
can be acknowledged
1)
self-latching
threshold
Alarm A2
can be acknowledged
1)
self-latching
CatEx 125 PR [%LEL] 0 to 100 20 yes no 40 no yes CatEx 125 PR Gas [%LEL] 0 to 100 20 yes no 40 no yes XXS O2 [Vol.-%] 0 to 25
19
2)
no yes 23 no yes
XXS CO LC [ppm] 0 to 2000 30 yes no 60 no yes XXS H2S LC [ppm] 0 to 100 5 yes no 10 no yes
XXS NO2 [ppm] 0 to 50 5 yes no 10 no yes
XXS SO2 [ppm] 0 to 100 0.5 yes no 1 no yes
1) Different settings can be selected to meet customer requirements on delivery. The current setting can be checked and changed with the Dräger CC Vision software. A version of the CC-Vision software that can be used for Dräger X-am 2500 is available for download from the product page for the X-am 2500 at the following web address: www.draeger.com
2) With O
, A1 is the lower alarm threshold: an alarm is triggered if the value is too low.
2
Dräger X-am 2500 9
Operation
i
i
!
IR
Calibration cradle (order no. 83 18 752) with inserted USB DIRA with USB cable (order no. 83 17 409)
USB DIRA with USB cable (order no. 83 17 409)
USB 2.0
USB 2.0
0
0
X-am 2500

3.3.2 Standard instrument configuration

NOTICE
Only trained persons are permitted to carry out modifications to the instrument configuration.
2)
2) 3)
1)
Extended bump test
2)
ON
ON
Dräger X-am® 2500
Bump test mode
Fresh air calibration
Operating signal Capture range ON
Switch off
LEL factor
2)
2)
(ch4)
2) 4) 5)
STEL (short-term average)
2) 5) 6)
TWA (shift average)
Alarm A1
7)
(4.4 vol. % corresponds to
STEL function - disabled
Average value duration =
TWA function - disabled
Average value duration = 8 hours
can be acknowledged, non-latching,
allowed
4.4 (vol. %)
100 %LEL)
15 minutes
pre-alarm,
rising flank
Alarm A1 at O
Alarm A2
sensor cannot be acknowledged, latching,
2
7)
like main alarm, falling flank
cannot be acknowledged, latching,
main alarm,
rising flank
1) X-am® is a registered trademark of Dräger.
2) Different settings can be selected to meet customer requirements on delivery. The current setting can be checked and changed with the Dräger CC Vision software.
3) A periodic short flashing indicates the operating capacity of the instrument. If there is no operating signal, correct operation cannot be guaranteed.
4) STEL: average value of an exposure over a short period, generally 15 minutes.
5) Interpretation only if the sensor is designed for this.
6) TWA: shift averages are workplace limit values for generally eight hours per day of exposure for five days a week during a working life.
7) Latching and acknowledgement of alarms A1 and A2 can be configured with the Dräger CC Vision PC software.
Changing the configuration: see “Replacing the sensors” on page 20.
WARNING
After a basic initialization has been carried out with the PC software Dräger CC Vision, individual alarm settings may have been changed.
Selecting or disabling the capture ranges (only applies for the measuring mode):
The capture range is selected in the measuring mode (factory setting) and permanently disabled in calibration mode. The CC-Vision PC software can be used to select or disable the capture ranges for the measuring mode.

3.3.3 Configuring the device

To individually configure a instrument with standard configuration, the instrument must be connected with a PC. The installed PC software Dräger CC Vision is used for configuration. The PC software Dräger CC Vision can be downloaded from the following web address free of charge: www.draeger.com/software.
Observe the documentation and online help for the
software.
Device settings
The following changes can be made to the device parameters for a device:
Designation Field
Password Numeric field (3-figure)
Operating signal LED
Operating signal horn
1)
1)
Yes/ No
Yes/ No
Switch-off mode “Switch off permitted” or
“Switch off prohibited” or “Switch off prohibited at A2”
Shift length (TWA) 2) (in minutes) Short-term exposure limit
3) 4)
(STEL)
(in minutes)
60 - 1440 (setting for exposure alarm)
0 - 15 (setting for exposure alarm)
User ID(12 characters) Alphanumeric field Switch database on or off On/Off Overwrite database Yes/No Database mode Peak/Average Database interval
1 s / 10 s / 30 s / 1 min / 2 min /
5 min / 10 min / 30 min Date (date on the PC) Time (time on the PC) Warning after expiry of
Yes/ No calibration interval Error after expiry of calibration
Yes/ No interval Delay until error after expiry of
0 - 10 calibration interval (days) Automatic detection of Bump
Yes/ No Test Station Activate sensitivity calibration following negative bump test
Yes/No (relates only to a
device connected to the
Dräger Bump Test Station) Bump test mode “extended bump test” or “quick
bump test” or “bump test
deactivated” Warning after expiry of bump
Yes/ No test interval
10 Dräger X-am 2500
Loading...
+ 22 hidden pages