Dräger X-am 2500 Technical Manual

Dräger X-am 2500 (MQG 0011)
Technical Manual
Content
Content
1.1 General safety statements . . . . . . . . . . . . . . . . . . . .4
1.2 Definitions of alert icons . . . . . . . . . . . . . . . . . . . . . .4
2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1 Product overview . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1.1 Front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1.2 Rear side . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1.3 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1.4 Special symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.3 Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.3.1 Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . .5
3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
3.1 Preparations for operation . . . . . . . . . . . . . . . . . . . .6
3.1.1 Charging the batteries . . . . . . . . . . . . . . . . . . . . . . . .6
3.1.2 Replacing the batteries / rechargeable batteries . . .7
3.1.3 Switching on the instrument . . . . . . . . . . . . . . . . . . .8
3.1.4 Switching off the instrument . . . . . . . . . . . . . . . . . . .8
3.2 Before entering the workplace . . . . . . . . . . . . . . . . .8
3.3 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
3.3.1 Standard gas configuration . . . . . . . . . . . . . . . . . . . .9
3.3.2 Standard instrument configuration . . . . . . . . . . . . .10
3.3.3 Configuring the device . . . . . . . . . . . . . . . . . . . . . .10
3.3.4 Export data memory and display graphically . . . . .11
3.4 Running the bump test . . . . . . . . . . . . . . . . . . . . . .11
3.4.1 Manual implementation without documentation
of the results in the instrument memory . . . . . . . . .11
3.4.2 Menu implementation with the documentation of
results in the instrument memory . . . . . . . . . . . . . .12
3.4.3 Automatic implementation with the
Bump Test Station . . . . . . . . . . . . . . . . . . . . . . . . .13
3.5 During operation . . . . . . . . . . . . . . . . . . . . . . . . . . .13
3.6 Identifying alarms . . . . . . . . . . . . . . . . . . . . . . . . . .14
3.6.1 Concentration pre-alarm A1 . . . . . . . . . . . . . . . . . .14
3.6.2 Concentration main alarm A2 . . . . . . . . . . . . . . . . .14
3.6.3 STEL/TWA exposure alarm . . . . . . . . . . . . . . . . . .14
3.6.4 Battery pre-alarm . . . . . . . . . . . . . . . . . . . . . . . . . .14
3.6.5 Battery main alarm . . . . . . . . . . . . . . . . . . . . . . . . .14
3.6.6 Instrument alarm . . . . . . . . . . . . . . . . . . . . . . . . . . .14
4 Menu functions . . . . . . . . . . . . . . . . . . . . . . . . . . .15
4.1 Activating the Info mode . . . . . . . . . . . . . . . . . . . . .15
4.2 Opening Info-Off Mode . . . . . . . . . . . . . . . . . . . . . .15
4.3 Quick Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
4.3.1 Quick menu functions . . . . . . . . . . . . . . . . . . . . . . .15
4.3.2 Opening the Quick Menu . . . . . . . . . . . . . . . . . . . .15
4.3.3 Quick menu "Delete peak values" . . . . . . . . . . . . . .15
4.4 Calibration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . .15
4.4.1 Calibration menu functions . . . . . . . . . . . . . . . . . . .15
4.4.2 Open the Calibration Menu . . . . . . . . . . . . . . . . . . .15
5 Calibrate instrument . . . . . . . . . . . . . . . . . . . . . . .16
5.1 Adjustment interval . . . . . . . . . . . . . . . . . . . . . . . . .16
5.2 Run fresh air calibration . . . . . . . . . . . . . . . . . . . . .16
5.3 Automatic fresh air calibration of the CatEx
sensor in the charging cradle . . . . . . . . . . . . . . . . .17
5.4 1-button calibration . . . . . . . . . . . . . . . . . . . . . . . . 17
5.4.1 Calibrating the sensitivity for an individual
measuring channel . . . . . . . . . . . . . . . . . . . . . . . . 18
5.4.2 Sensitivity calibration for CatEx . . . . . . . . . . . . . . . 18
5.5 Calibration with the X-dock maintenance
station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6 Operation with pump . . . . . . . . . . . . . . . . . . . . . 19
6.1 Performing a measurement with the Dräger
X-am Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.2 Performing a measurement with a manual pump
adapter and rubber ball pump . . . . . . . . . . . . . . . . 20
7 Replacing the sensors . . . . . . . . . . . . . . . . . . . . 20
8 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . 21
8.1 Warning messages . . . . . . . . . . . . . . . . . . . . . . . . 21
8.2 Fault message . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
9 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
9.1 Maintenance table . . . . . . . . . . . . . . . . . . . . . . . . . 26
10 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
11 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
12 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
13 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
13.1 X-am 2500 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
14 Order list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Dräger X-am 2500 3
For your safety
!
!
i
i
00133366.eps
0
1
2
6
5
4
3
2
X-am 2500
7
OK
00233366.eps
2
1
4
6
3
5
00333366_en.eps
123
ch4
O
2
CO
123
ch4 %LEL
0.0
O2 Vol%
20.9
CO
ppm
0.0
H2S
ppm
0.0
%LEL
0.0
Vol%
20.9
ppm
0.0

1 For your safety

1.1 General safety statements

Before using this product, carefully read the associated
Instructions for Use. This document does not replace the Instructions for Use.

1.2 Definitions of alert icons

The following alert icons are used in this document to provide and highlight areas of the associated text that require a greater awareness by the user. A definition of the meaning of each icon is as follows:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in physical injury, or damage to the product or environment. It may also be used to alert against unsafe practices.
NOTICE
Indicates additional information on how to use the product.

2.1.2 Rear side

1
IR interface
2
Fastening clip
3
Nameplate
4
Charging contacts
5
Power pack
6
Serial no.

2.1.3 Display

2 Description

2.1 Product overview

2.1.1 Front

Left: 4 measuring channels, right: 3 measuring channels
1
Measured gas display
2
Measuring value display with unit
3
Special symbols
The following only shows the instrument version with 4 measuring channels.
1
Gas entry
2
Alarm LED
3
Horn
4
key
5
key
6
Display
7
Tool for changing sensor
4 Dräger X-am 2500

2.1.4 Special symbols

Fault message, see section 4.1 on page 15 Warning message, see section 4.1 on page 15 Display of peak values for all measured gases,
see section 4.1 on page 15 The exposure evaluation display (TWA) for measured
gases, e.g. H The exposure evaluation display (STEL) for measured
gases, e.g. H
S and CO, see section 4.1 on page 15
2
S und CO, see section 4.1 on page 15
2
Description
!
!
The instrument is set to the bump test function, see section 3.4 on page 11
The instrument is set to the fresh air calibration function, see section 5.2 on page 16
The instrument is set to the 1-button calibration/ adjustment function, see section 5.4 on page 17
The instrument is set to the single gas calibration function, see section 5.4.1 on page 18
Function for password input is active, see section 4.4 on page 15
Battery / rechargeable battery 100 % full Battery / rechargeable battery 2/3 full Battery / rechargeable battery 1/3 full Battery / rechargeable battery empty

2.2 Intended use

Portable gas detection instrument for the continuous monitoring of the concentration of several gases in the ambient air within the working area and in explosion-hazard areas. Independent measurement of up to 4 gases, in accordance with the installed Dräger sensors.
Areas subject to explosion hazards, classified by zones
The instrument is intended for the use in explosion-hazard areas of Zone 0, Zone 1 or Zone 2 or in mines at risk due to fire damp. It is intended for use within a temperature range of ­20 °C to +50 °C, and for areas in which gases of explosion groups IIA, IIB or IIC and temperature class T3 or T4 (depending on the batteries and rechargeable battery) may be present. If used in mines, the instrument is only to be used in areas known to have a low risk of mechanical impact.
Areas subject to explosion hazards, classified by divisions.
The instrument is intended for use in explosion-hazard areas according to Class I&II, Div. 1 or Div. 2 within a temperature range of -20 °C to +50 °C, and for areas where gases or dusts of groups A, B, C, D, E, F, G and temperature class T3 or T4 may be present (depending on the rechargeable battery and batteries).
NOTICE
i
i
CSA requirement: Only the combustible gas detection portion of this instrument has been assessed for performance. The instrument has not been classified by the CSA for use in mines.

2.3 Approvals

Copies of the name plate and the declaration of conformity are provided in the enclosed supplementary documentation (order no. 90 33 890). Do not stick anything on the name plate on the gas detector. The technical approvals are valid for the X-am 2500 gas detection instrument and the calibration cradle. The explosion­protection approvals are only valid for the X-am 2500 gas detection instrument; the calibration cradle must not be used in the Ex zone. The BVS 10 ATEX E 080 X technical suitability test is based on the calibration with the target gas.

2.3.1 Safety Instructions

WARNING
!
Do not replace or charge batteries in potentially explosive areas. Explosion hazard!
Charge the NiMH power pack T4 (type HBT 0000) or T4 HC (type HBT 0100) with the associated Dräger charger. Charge NiMH single cells for ABT 0100 battery holder in accordance with the manufacturer's specifications. Ambient temperature during the charging process: 0 to +40 °C.
To reduce the danger of explosion, do not mix new batteries with old batteries and do not mix batteries made by different manufacturers.
Always disconnect the instrument from the power pack before carrying out any maintenance operations.
WARNING
CSA requirement: Measured values over the full scale value may indicate an explosive atmosphere.
Only applicable to Class II certification. CSA Std. C22.2 No 152 does not have any requirement for Class II hazardous locations and therefore this device has not been performance tested by CSA for Class II. The sensor may become clogged and not detect gas properly or warn the user of its inability to detect gas.
WARNING
CSA requirement: The sensitivity must be tested on a daily basis before first use with a known concentration of the gas to be measured in accordance with 25 to 50 % of the concentration limit value. The accuracy must be 0 to +20 % of the actual value. The accuracy can be corrected via calibration.
Substitution of components may impair intrinsic safety.
Only use power packs ABT 0100 (order no. 83 22 237), HBT 0000 (order no. 83 18 704) or HBT 0100 (order no. 83 22 244). See marking on power pack for approved batteries and related temperature classes.
Not tested in an oxygen-enriched atmosphere (>21 % O
High off-scale readings may indicate an explosive concentration.
).
2
Dräger X-am 2500 5
Operation
!
!
!
i
i
02733366.eps
0 0 0
Ex %UEG CO
2
ppm
O
2
Vol% CO ppm H
2S
ppm
Ex %UEG CO
2
ppm
O
2
Vol% CO ppm H
2S
ppm
Ex %UEG CO
2
ppm
O
2
Vol% CO ppm H
2S
ppm
X-am 2500
X-am 2500 X-am 2500
83 25 736 83 16 639
83 21 849 / 83 21 850
2
1
2
1
2
1
1
3
3
3
Note the following for CSA (Canadian Standards Association) applications:
For the CSA approval only the functions of the device component that is used to measure flammable gases are tested. The device is not approved by CSA for use in mining.
WARNING
Before daily use, test the sensitivity with a known concentration of the applicable gas corresponding to 25 to 50 % of the maximum concentration. The accuracy must be within a range of 0 to +20 % of the actual value. Perform a calibration to correct the accuracy if necessary.

3 Operation

3.1 Preparations for operation

WARNING
To reduce the risk of ignition of a flammable or explosive atmosphere, strictly adhere to the following warning statements: Only use power pack types ABT 01xx, HBT 00xx or HBT 01xx. See the marking on the rechargeable battery for permitted rechargeable batteries and the corresponding temperature class.
The charging time is typically 4 hours.A new NiMH power pack reaches its full capacity after three
complete charging/discharging cycles.
Never store the instrument for extended periods without
being connected to a power source (maximum of 2 months) because the internal buffer battery will drain.
Charging with the charging module adapter and the power supply unit 83 21 849 or 83 21 850
A maximum of 5 instruments in charging modules (order
no. 83 18 639) can be charged at the same time on the charging module adapter (order no. 83 25 736) in connection with the power supply unit (order no. 83 21 849). Up to 20 instruments can be charged at the same time with the power supply unit 83 21 850.
Before attaching the charging modules to the charging
module adapter, disconnect the power supply unit from the mains supply!
CAUTION
!
Always connect or disconnect the charging modules individually and never in groups in order to prevent the charging module adapter from becoming damaged. Even during transportation, always handle the power supply unit and the charging modules individually and without instruments inserted.
Substitution of components may impair intrinsic safety.
Before using the instrument for the first time, insert a
charged NiMH T4 power pack or batteries approved by Dräger see section 3.1.2 on page 7.
The instrument is now ready for operation.

3.1.1 Charging the batteries

WARNING
Explosion hazard! To reduce the risk of flammable or explosive atmospheres igniting, it is essential that the warning notices below are observed: Do not charge underground or in explosion hazard areas!
The chargers are not designed in accordance with the regulations for fire damp and explosion protection.
Charge the NiMH power pack T4 (type HBT 0000) or T4 HC (type HBT 0100) with the associated Dräger charger. Charge NiMH single cells for ABT 0100 battery holder in accordance with the manufacturer's specifications. Ambient temperature during the charging process: 0 to +40 °C.
NOTICE
Even if the instrument is not used, Dräger recommends storing the instrument in the charging cradle (chargingmodule X-am 1/2/5000, order no. 83 18 639).
To maintain the lifetime of the batteries, charging is
temperature controlled and only performed in a temperature range of 5 to 35 °C. When outside this temperature range, the charging automatically interrupted and automatically recommenced after the temperature range has been reached again.
Position the instrument on an even and level surface.
1. Turn the slots of the interlock into a horizontal position by using a screwdriver or coin.
2. Insert the fastening lug (2) of the charger module (simultaneous power feed) until it engages.
3. Close the lock (1) with a quarter turn (slot is positioned vertically).
4. Attach additional charging modules in the same way.
5. Connect the power pack to the mains.
The green LED (3) lights up.
6. Insert the switched off instrument into the charger module.
Display LED (3) on the charger module:
Charging
Fault
Full
If a fault occurs:
Remove the instrument from the charging module and
insert it again.
If the fault still occurs, have the charging module repaired.
It takes approx. 4 hours to fully charge an empty
rechargeable battery.
6 Dräger X-am 2500
Operation
!
i
i
02833366.eps
83 16 994 (100 ... 240 V) 83 16 997 (100 ... 240 V)
45 30 057
0
Ex %UEG CO
2
ppm
O
2
Vol% CO ppm H
2S
ppm
X-am 2500
CAUTION
A short circuit of the charging contacts in the charging modules, e. g., by metallic objects that have fallen in, does not result in damage to the charging station. It should, however, be avoided due to possible heating hazards and incorrect displays on the charger module.
NOTICE
If combining different charging modules, follow the instructions in the manual supplied with the charging module adapter.
Charge using charger module and plug-in power pack or vehicle charging adapter
When using the power supply unit (order no. 83 16 994),
up to 5 instruments can be charged at the same time, with the power supply unit (order no. 83 16 997) one instrument individually.
When using the vehicle charging adapter (order no.
45 30 057) it is recommended that you supply every charging module separately.
The charging process is carried out analogous to charging
with the multiple charging station.

3.1.2 Replacing the batteries / rechargeable batteries

WARNING
!
Explosion hazard! To reduce the risk of flammable or explosive atmospheres igniting, it is essential that the warning notices below are observed:
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in areas at risk of an explosion hazard.
Do not mix new batteries with used batteries, and do not mix batteries from different manufacturers or of different types.
Remove batteries before maintenance work.
Batteries / rechargeable batteries are part of the Ex approval. Only the following types may be used:
Alkaline batteries – T3 – (non rechargeable!)Panasonic LR6 PowerlineVarta Type 4106 Varta Type 4006 Alkaline batteries – T4 – (non rechargeable!)Duracell Procell MN1500
MN1500
NiMHy rechargeable batteries – T3 – (rechargeable)GP 180AAHC
1)
1)
(power one) or
1
(industrial)
1)
, Duracell Plus Power
1
(1800 mAh) max. 40 °C ambient
temperature.
Charge the NiMH power pack T4 (type HBT 0000) or T4 HC (type HBT 0100) with the associated Dräger charger. Charge NiMH rechargeable batteries for battery holder ABT 0100 in accordance with the manufacturer's specifications. Ambient temperature during the charging process: 0 to +40 °C.
1) Not part of the measurement performance tests BVS10 ATEX E 080X and PFG 10 G 001X.
Dräger X-am 2500 7
Operation
02633366.eps
1
2
3
+
+
OK
1. Switching off the instrument: key and key are held down simultaneously.
2. Loosen the screw (2.0 mm hexagon socket) on the power pack and remove the power pack.
With battery holder (order no. 83 22 237): Replace alkaline
batteries or NiMHy rechargeable batteries. Ensure correct polarity.
With the T4 NiMH power pack (type HBT 0000) / T4 HC
(type HBT 0100): Completely replace the power pack.
3. Insert the power pack into the instrument and tighten the screw, the instrument switches on automatically.
After replacing the T4 NiMH power pack (type HBT 0000)/ T4 HC (type HBT 0100), a full charge is recommended.
After the batteries have been replaced:
The settings and data are stored when the battery is replaced.
The sensors warm up again.

3.1.3 Switching on the instrument

1. Hold down the [OK] button for approx. 3 seconds until the »3.2.1« countdown shown on the display has elapsed.
All the display segments, including the visual, audible
and vibration alarms, are activated for a short time.
The software version is displayed.The instrument performs a self-test.The sensor that is up next for calibration/adjustment is
displayed with the remaining days until the next calibration/adjustment e. g. »Ex%LEL CAL20«.
The time until the bump test interval elapses is
displayed in days, e.g. »bt123«.
All A1 and A2 alarm thresholds and »« (TWA)
»« (STEL)
1
for all toxic gases (e. g. H2S or CO) are
1)
and
displayed consecutively.
During the sensor warm-up phase:
The display for the measured value flashes The special symbol » « is displayed.No alarms are issued during the warm-up phase.The red LEDs flash.The gas detector is ready to measure when the
measured values no longer flash and the red LEDs are no longer illuminated. The special symbol » « may continue to be displayed if corresponding warnings (e.g. not yet ready for calibration) are active.
2. Press the [OK] key to cancel the display of the activation sequence.
1) Only when activated in the instrument configuration. Delivery condition: not activated.

3.1.4 Switching off the instrument

Press and hold the [OK] key and [+] key simultaneously
until the countdown »3.2.1« shown on the display has elapsed. When the instrument is switched off, the visual, audible and vibration alarms are activated for a short time in order to protect against inadvertent switch off.

3.2 Before entering the workplace

WARNING
!
Before any measurements relevant to safety are made, check the adjustment with a bump test, adjust if necessary and check all alarm elements. If national regulations apply, a bump test must be performed according to the national regulations. Faulty adjustment may result in incorrect measuring results, with possible serious consequences.
CAUTION
!
The CatEx sensor is intended for measurements of flammable gases and vapours mixed with air (i.e. O
content 21 vol.%). Incorrect measured values may be displayed in the case of oxygen deficient or oxygen enriched environments.
NOTICE
i
i
If the gas detector is used for offshore applications, a distance of 5 m to a compass must be complied with.
1. Switch on the instrument. The current measured values are shown in the display.
2. Observe any warning »« or fault messages »«.
The instrument can be operated normally. If the warning message does not disappear automatically during operation, the instrument must be serviced after the end of use.
The instrument is not ready to measure and requires maintenance.
3. Check that the gas inlet opening on the instrument is not covered.
WARNING
!
Explosion hazard! To reduce the risk of flammable or explosive atmospheres igniting, it is essential that the warning notices below are observed:
Fractions of catalytic poisons in the measuring gas
(e.g. volatile silicone, sulphur, heavy metal compounds or halogenated hydrocarbon) can damage the CatEx sensor. If the CatEx sensor can no longer be calibrated to the target concentration, the sensor must be replaced.
In case of measurements in an oxygen-deficient
atmosphere (<12 Vol.-% O may show incorrect displays; in this case, a reliable
) the CatEx sensor
2
measurement with a CatEx sensor is not possible.
In an oxygen enriched atmosphere (>21 vol. %
O
), the explosion protection cannot be
2
guaranteed; remove instrument from the Ex area.
High values outside the display area indicate an
explosive concentration where applicable.
2
8 Dräger X-am 2500

3.3 Configuration

3.3.1 Standard gas configuration

Operation
DrägerSensor Measuring range
1)
threshold
Alarm A1
can be acknowledged
1)
self-latching
threshold
Alarm A2
can be acknowledged
1)
self-latching
CatEx 125 PR [%LEL] 0 to 100 20 yes no 40 no yes CatEx 125 PR Gas [%LEL] 0 to 100 20 yes no 40 no yes XXS O2 [Vol.-%] 0 to 25
19
2)
no yes 23 no yes
XXS CO LC [ppm] 0 to 2000 30 yes no 60 no yes XXS H2S LC [ppm] 0 to 100 5 yes no 10 no yes
XXS NO2 [ppm] 0 to 50 5 yes no 10 no yes
XXS SO2 [ppm] 0 to 100 0.5 yes no 1 no yes
1) Different settings can be selected to meet customer requirements on delivery. The current setting can be checked and changed with the Dräger CC Vision software. A version of the CC-Vision software that can be used for Dräger X-am 2500 is available for download from the product page for the X-am 2500 at the following web address: www.draeger.com
2) With O
, A1 is the lower alarm threshold: an alarm is triggered if the value is too low.
2
Dräger X-am 2500 9
Operation
i
i
!
IR
Calibration cradle (order no. 83 18 752) with inserted USB DIRA with USB cable (order no. 83 17 409)
USB DIRA with USB cable (order no. 83 17 409)
USB 2.0
USB 2.0
0
0
X-am 2500

3.3.2 Standard instrument configuration

NOTICE
Only trained persons are permitted to carry out modifications to the instrument configuration.
2)
2) 3)
1)
Extended bump test
2)
ON
ON
Dräger X-am® 2500
Bump test mode
Fresh air calibration
Operating signal Capture range ON
Switch off
LEL factor
2)
2)
(ch4)
2) 4) 5)
STEL (short-term average)
2) 5) 6)
TWA (shift average)
Alarm A1
7)
(4.4 vol. % corresponds to
STEL function - disabled
Average value duration =
TWA function - disabled
Average value duration = 8 hours
can be acknowledged, non-latching,
allowed
4.4 (vol. %)
100 %LEL)
15 minutes
pre-alarm,
rising flank
Alarm A1 at O
Alarm A2
sensor cannot be acknowledged, latching,
2
7)
like main alarm, falling flank
cannot be acknowledged, latching,
main alarm,
rising flank
1) X-am® is a registered trademark of Dräger.
2) Different settings can be selected to meet customer requirements on delivery. The current setting can be checked and changed with the Dräger CC Vision software.
3) A periodic short flashing indicates the operating capacity of the instrument. If there is no operating signal, correct operation cannot be guaranteed.
4) STEL: average value of an exposure over a short period, generally 15 minutes.
5) Interpretation only if the sensor is designed for this.
6) TWA: shift averages are workplace limit values for generally eight hours per day of exposure for five days a week during a working life.
7) Latching and acknowledgement of alarms A1 and A2 can be configured with the Dräger CC Vision PC software.
Changing the configuration: see “Replacing the sensors” on page 20.
WARNING
After a basic initialization has been carried out with the PC software Dräger CC Vision, individual alarm settings may have been changed.
Selecting or disabling the capture ranges (only applies for the measuring mode):
The capture range is selected in the measuring mode (factory setting) and permanently disabled in calibration mode. The CC-Vision PC software can be used to select or disable the capture ranges for the measuring mode.

3.3.3 Configuring the device

To individually configure a instrument with standard configuration, the instrument must be connected with a PC. The installed PC software Dräger CC Vision is used for configuration. The PC software Dräger CC Vision can be downloaded from the following web address free of charge: www.draeger.com/software.
Observe the documentation and online help for the
software.
Device settings
The following changes can be made to the device parameters for a device:
Designation Field
Password Numeric field (3-figure)
Operating signal LED
Operating signal horn
1)
1)
Yes/ No
Yes/ No
Switch-off mode “Switch off permitted” or
“Switch off prohibited” or “Switch off prohibited at A2”
Shift length (TWA) 2) (in minutes) Short-term exposure limit
3) 4)
(STEL)
(in minutes)
60 - 1440 (setting for exposure alarm)
0 - 15 (setting for exposure alarm)
User ID(12 characters) Alphanumeric field Switch database on or off On/Off Overwrite database Yes/No Database mode Peak/Average Database interval
1 s / 10 s / 30 s / 1 min / 2 min /
5 min / 10 min / 30 min Date (date on the PC) Time (time on the PC) Warning after expiry of
Yes/ No calibration interval Error after expiry of calibration
Yes/ No interval Delay until error after expiry of
0 - 10 calibration interval (days) Automatic detection of Bump
Yes/ No Test Station Activate sensitivity calibration following negative bump test
Yes/No (relates only to a
device connected to the
Dräger Bump Test Station) Bump test mode “extended bump test” or “quick
bump test” or “bump test
deactivated” Warning after expiry of bump
Yes/ No test interval
10 Dräger X-am 2500
Operation
00833366.eps
0,5 L/min
0
Designation Field
Error after expiry of bump test
Yes/ No interval (if warning activated) Capture range Yes / No
Remote configuration Yes / No Bump test interval (days) 1 - 732 Delay until error after expiry of
0-10 cal. interval (days) Activate user service life Yes/No
User service life (days)
0 - 999 (if activated)
Running in Yes/No LEL category “---” or “PTB” or “IEC” or
“NIOSH” (if this is changed,
the LEL factor will be altered to
match)
1) At least one of the two operating signals must be switched on.
2) Corresponds to the averaging time and is used to calculate the exposure value TWA.
3) Only evaluated if the sensor is provided for the purpose.
4) Corresponds to the averaging time and is used to calculate the exposure value STEL.
Sensor settings
The following changes can be made to the sensor parameters for the sensors:
Designation Field
Alarm threshold A1
0 - A2
(in measurement unit) Alarm threshold A2
(in measurement unit)
Type of evaluation
1)
Alarm threshold STEL (in measurement unit) Alarm threshold TWA (in measurement unit)
1)
1)
A1 – Measuring range limit value
Inactive, TWA, STEL, TWA+STEL
0 – Measuring range limit value
0 – Measuring range limit value
Calibration interval (days) 0 - 180 (sensor-dependent) Unit (sensor-dependent) Vol%, %UEG, %LEL, %LIE,
3
Gas name: “Ex”
ppm, mbar, ppb, mg/m Yes/ No
(CatEx sensor only) Automatic fresh air calibration
Yes/ No in the charging cradle (CatEx sensor only)
1) Only evaluated if the sensor is provided for the purpose.
Testing the parameters
In order to ensure that the values have been correctly transferred to the gas measuring device:
1. Press the touch button Data from X-am 1/2/5x00
2. Check parameters.

3.3.4 Export data memory and display graphically

To read the database of the instrument and display it graphically, the instrument must be connected with a PC.
The installed Dräger GasVision PC software is used for exporting and displaying the database.
Observe the documentation and online help of the software.

3.4 Running the bump test

3.4.1 Manual implementation without documentation of the results in the instrument memory

1. Prepare a test gas cylinder,
the volume flow must be
0.5 l/min and the gas concentration must be higher than the alarm threshold concentration that is to be tested. Example test gas cylinder 68 11 130 = mixed gas with 50 ppm CO, 15 ppm H
2.5 vol. % CH
, 18 vol. % O
4
2. Connect the test gas
cylinder with the calibration cradle (order no. 83 18 752).
3. Vent the test gas into
a fume cupboard or into the open air (with a hose connected to the second connector of the calibration cradle).
WARNING
!
CSA requirement: carry out a bump test before use. It should be carried out in the measuring range 25­50 % of the full scale value, whereby the displayed measured value may deviate from the actual measured value by 0-20 %. Accuracy may be corrected via calibration.
CAUTION
!
Never inhale the test gas. Health hazard! Observe the hazard warnings of the relevant Safety Data Sheets.
4. Switch on the instrument and insert it into the calibration
cradle – press downwards until it engages.
5. Open the test gas cylinder valve to let test gas flow over
the sensors. Recommendation: Wait until the instrument displays the test gas concentration with sufficient tolerance –
Ex: ±20 % of the test gas concentration O2: ±0.6 vol. %
1
TOX: ±20 % of the test gas concentration Wait until at least alarm threshold A1 or A2 has been exceeded, however.
If the alarm thresholds are exceeded, the instrument displays the gas concentration in alternation with »A1« or »A2« depending on the test gas concentration.
6. Close the test gas cylinder valve and remove the
instrument from the calibration cradle.
NOTICE
i
i
To check the measured value response times, apply t90 test gas to the X-am via the calibration cradle. Check the results in accordance with the information in the table in the enclosed supplementary documentation (order no. 90 33 890) until 90 % of the end display is reached.
S,
2
2
1)
1
1) Upon application of the Dräger mixed gas (order no. 68 11 130) the displays
should be within this range.
Dräger X-am 2500 11
Operation
i
i
00833366.eps
0,5 L/min
0
!
00933366_en.eps
CH4 %LEL O
2
Vol% CO
ppm H
2S
ppm
OK
01033366_en.eps
CH4 %LEL O
2
Vol% CO
ppm H
2S
ppm
01133366_en.eps
CH
4
%LEL O
2
Vol% CO
ppm H
2
S
ppm
NOTICE
After the bump test (menu), the display shows a printer icon even if there is no printer connected to the bump test station.
If the concentration has now fallen under the A1 alarm threshold:
Acknowledge the alarm.
If the displays are outside of the above-mentioned ranges:
Calibrating/adjusting the instrument, see section 5 on page 16.

3.4.2 Menu implementation with the documentation of results in the instrument memory

The setting to "Quick bump test" or "Extended bump test" is made using the PC software Dräger CC Vision. In the "Quick bump test" a check is carried out as to whether or not the gas concentration has exceeded alarm threshold 1 (with oxygen, the test checks that alarm threshold 1 has not been reached). In the case of the “Extended bump test”, a check is made as to whether the gas concentration has reached the set bump test concentration within a tolerance window.
Setting on delivery: Extended bump test.
1. Prepare a test gas cylinder, the volume flow must be 0.5 l/min
and the gas concentration must be higher than the alarm threshold concentration that is to be tested. Example test gas cylinder 68 11 130 = mixed gas with 50 ppm CO, 15 ppm H
S, 2.5 vol. % CH4, 18 vol. % O
2
2
2. Connect the test gas cylinder
with the calibration cradle (order no. 83 18 752)).
3. Vent the test gas into a fume
cupboard or into the open air (with a hose connected to the second connector of the calibration cradle).
CAUTION
Never inhale the test gas. Health hazard! Observe the hazard warnings of the relevant Safety Data Sheets.
5. Open the Quick menu and select the bump test, page 15. The current gas concentration values and the special symbol » « (for bump test) flash.
6. Press the key to start the bump test.
7. Open the test gas cylinder valve to let test gas flow over the sensor.
If gas concentration exceeds the alarm thresholds A 1 or A 2 the corresponding alarm will occur.
Ending the bump test:
After the set bump test concentration has been reached or a gas alarm has been triggered (with "Quick bump test"):
The display containing the
current gas concentration changes with the display »OK«.
The bump test that was
carried out is documented with the result and date in the instrument memory.
8. Close the test gas cylinder valve and remove the instrument from the calibration cradle.
If the concentration values have now fallen under the
A1 alarm thresholds, the instrument returns to the measuring mode.
If the set bump test concentration is not reached within the
set time, an error is issued.
The fault message
» « appears and » « is displayed instead of the measured value on the faulty measuring channel.
In this case, repeat the
bump test or calibrate the instrument, page 20.
4. Switch on the instrument and insert it into the calibration cradle – press downwards until it engages.
The bump test can also be run automatically. The "Bump Test Station" is required for this function, see section 3.4.3 on page 13.
12 Dräger X-am 2500
Operation
01333366.eps
0
Ex %UEG O
2
Vol% CO ppm H
2S
ppm
01133366_en.eps
CH
4
%LEL O
2
Vol% CO
ppm H
2
S
ppm
01133366_en.eps
CH
4
%LEL O
2
Vol% CO
ppm H
2
S
ppm

3.4.3 Automatic implementation with the Bump Test Station

The fault message
» « appears and » « is displayed instead of the measured value on the faulty measuring channel.
In this case, repeat the
bump test or calibrate the instrument, page 20.
The bump test can also be run manually, see section 3.4.1 on page 11.
Prerequisite: The instrument first needs to be configured for the automatic
The Dräger CC Vision PC software can be used to enable the "Automatic calibration after incorrect bump test" option.
bump test using the Dräger CC-Vision PC sofware.
Activating the instrument for the automatic bump test.Adjust the test gas concentration (mixed gas) with the
Dräger CC-Vision PC software if it deviates from the following default values – standard on delivery: 50 ppm CO, 15 ppm H
Define which measuring channels should participate in the
S, 2.5 vol. % CH4, 18 vol. % O
2
2
automatic bump test. All measuring channels participate in the bump test by default.
1. Prepare the Bump Test Station according to the instructions.
2. Switch on the instrument
and insert it into the receptacle of the Bump Test Station until it engages. The bump test will be started automatically. The special symbol » «

3.5 During operation

During operation, the measured values for every measured
In the event of an alarm, the corresponding displays,
If a measuring range is exceeded or not reached,
NOTICE
i
i
After the bump test, the display shows a printer icon even if there is no printer connected to the bump gas station.
gas are displayed.
including the visual, audible and vibration alarms, are activated see section 6 on page 19.
the following displays are shown instead of the measured value display:
(for bump test) flashes.
»«
»«
»
(measuring range exceeded) or
(measuring range not reached) or
(blocking alarm).
«
In the event of an alarm, the corresponding displays,
including the visual, audible and vibration alarms,
If a gas alarm (quick bump test) is initiated and the set bump test concentration (Accelerated bump test) is reached within the set time, the current gas concentration will be displayed alternately with »OK«.
3. Remove the instrument from the Bump Test Station.
If the concentration values have now fallen under the
A1 alarm thresholds, the instrument returns to the measuring mode.
If there is no alarm during the bump test and the current
measurements do not reach the set target concentration ("Accelerated bump test" only), an error is issued.
are activated, see section 6 on page 19.
If an O
concentration of below 12 vol. %, an error is indicated with
sensor is fitted and this sensor measures an O
2
» « on the ex-channel instead of a measured value if the measured value is below the pre-alarm threshold.
After the measuring range of the TOX measuring channels
has been exceeded temporarily (up to one hour), checking the measuring channels is not necessary.
NOTICE
i
i
Special states in which there is no measuring operation (quick menu, calibration menu, warm-up of sensors, password input) are indicated by a visual signal (slow flashing of the alarm LED
).
Dräger X-am 2500 13
2
Operation
!
i
i
OK
!
OK
If the measuring range is exceeded significantly at the CatEx channel (very high concentration of flammable materials), a blocking alarm is triggered. This CatEx blocking alarm can be acknowledged manually by switching the instrument off and back on again in fresh air.
WARNING
If the DrägerSensor CatEx 125 PR is used in the gas detector, a zero point and sensitivity calibration must be carried out after experiencing an impact load that results in a non-zero display when exposed to fresh air. This warning does not apply if the DrägerSensor CatEx 125 PR Gas is used.

3.6 Identifying alarms

An alarm is displayed visually, audibly and through vibration in a specific pattern.
NOTICE
At low temperatures the legibility of the display can be improved by switching on the backlight.

3.6.1 Concentration pre-alarm A1

Intermittent alarm: Display »A1« and measured value alternating:
The pre-alarm A1 is not latching and stops when the concentration has dropped below the alarm threshold A1.
In case of A1, a single tone is audible and the alarm
LED flashes.
Acknowledging the pre-alarm:
Press the key. Only the audible alarm and the vibration
alarm are switched off.

3.6.2 Concentration main alarm A2

WARNING
Risk of fatal injury! Leave the area immediately. A main alarm is self-latching and cannot be acknowledged or cancelled.
Intermittent alarm: Display » A2 and measured value alternating:
In case of A2, a double tone is audible and the alarm
LED flashes twice.
After leaving the area, when the concentration has dropped below the alarm threshold:
Press the key. The alarm messages are switched off.
If the measuring range is exceeded significantly on the CatEx channel (very high concentration of flammable materials), a blocking alarm is triggered. This CatEx blocking alarm can be acknowledged manually by switching the instrument off and back on again in fresh air.
OK

3.6.3 STEL/TWA exposure alarm

CAUTION
!
Health hazard! Leave the area immediately. After this alarm, the deployment of personnel is subject to the relevant national regulations.
NOTICE
i
i
The STEL alarm can be triggered with a maximum delay of one minute.
Intermittent alarm: Display »A2« and »« (STEL) or »« (TWA) and
measured value alternating:
The STEL and TWA alarm cannot be acknowledged
or cancelled.
Switch off the instrument. The values for the exposure
evaluation are deleted after the instrument is switched on again.

3.6.4 Battery pre-alarm

Intermittent alarm: Flashing special symbol »« on the right
side of the display:
Acknowledging the pre-alarm:
Press the key. Only the audible alarm and the vibration
alarm are switched off.
The battery still lasts min. 20 minutes after the first battery
pre-alarm.

3.6.5 Battery main alarm

Intermittent alarm: Flashing special symbol »« on the right side
of the display:
The battery main alarm cannot be acknowledged or cancelled:
The device automatically switches off after 10 seconds.Before the instrument is switched off, the visual, audible
and vibration alarms are activated for a short time.

3.6.6 Instrument alarm

Intermittent alarm: Special symbol »« displayed on the right
side of the display:
The instrument is not ready for operation.For corrective measures, see “Replacing the sensors” on
page 20 to page 23.
Contact maintenance or DrägerService to rectify the problem.
14 Dräger X-am 2500
Menu functions
OK
OK
OK
OK
OK
OK
OK
00433366_en.eps
CH4 %LEL O
2
Vol% CO
ppm H
2S
ppm
00533366_en.eps
CH4 %LEL O
2
Vol% CO
ppm H
2S
ppm
OK
OK

4 Menu functions

4.1 Activating the Info mode

In measuring mode, press the key for approx. 3 seconds.If any warning or fault messages exist, the corresponding
information or error codes will be displayed (see section 8 on page 21).
Press the key successively for the next display.The peak values and the exposition values TWA
STEL
1)
are displayed.
Warning messages are displayed. Numerical codes of warning messages: see section 8.1 on page 21.
key
Fault messages are displayed. Numerical codes of fault messages: see section 8.2 on page 23.
key
The peak values = the maximum measured values in the case of, e.g., CO, H measured values in the case of O interval are displayed
S, ... or the minimum
2
within the storage
2
key
The average values of the exposures based on a shift of, e.g., 8 hours (TWA) of all the active sensors for the exposure evaluation are displayed
key
The short-term values (STEL) = average values of the concentrations over the average value duration of all the active sensors for the exposure evaluation are displayed
key
The instrument is in measuring mode again
If no key is pressed for 10 seconds, the instrument returns
automatically to measuring mode.
1)
and

4.3.2 Opening the Quick Menu

On delivery, only the fresh air calibration is activated in the Quick Menu. The PC software Dräger CC Vision can be used to activate the bump test for the quick menu and/or the function for displaying and deleting peak values.
1. In measuring mode, press the key three times. If no functions have been activated in the quick menu, the instrument remains in measuring mode.
2. You can select the activated functions of the quick menu by pressing the key.
Press the key to call the selected function.Press the key to cancel the active function and to
OK
switch to measuring mode.
If no key is pressed for 60 seconds, the instrument
returns automatically to measuring mode.

4.3.3 Quick menu "Delete peak values"

After the function has been selected, the current peak values are displayed; the peak values special symbol appears in the display at the same time.
1. The peak values can be deleted by pressing the
key for 5 sec. and the adjacent display appears, for example.
2. Press the key to end the function.

4.2 Opening Info-Off Mode

When the instrument is in a deactivated state, press the key. The name of the gas, measuring unit, and measuring range limit value are displayed for all channels.
Pressing the key again exits the Info Off mode
OK
(or via timeout).

4.3 Quick Menu

4.3.1 Quick menu functions

Bump test see section 3.4 on page 11
Fresh air calibration, see section 5.2 on page 16
Delete peak values, see section 4.3.3 on page 15
Display pump information, see page 19
Activate or deactivate pump, see page 19
1) Only when activated in the instrument configuration. Delivery status: not activated.

4.4 Calibration Menu

4.4.1 Calibration menu functions

Fresh air calibration, see section 5.2 on page 16
1-button calibration, see section 5.4 on page 17
Single gas calibration, see section 5.4.1 on page 18

4.4.2 Open the Calibration Menu

The calibration menu can only be accessed by entering
a password. Password on delivery: »001«
The default password on delivery can be changed using
the PC software Dräger CCVision.
Dräger X-am 2500 15
Calibrate instrument
00633366.eps
OK
OK
OK
OK
!
i
i
!
01333366_en.eps
CH
4
%LEL O
2
Vol% CO
ppm H
2
S
ppm
OK
01433366_en.eps
CH4 %LEL
O
2
Vol% CO
ppm H
2S
ppm
OK
1. In measuring mode, press the key for at least 4 seconds. The function for entering the password is selected. The special symbol » « (for the "enter password" function) is displayed.
The display shows »000«, with the first digit flashing.
2. Use the key to set the flashing digit.
3. Press the key, the second digit starts flashing.
4. Use the key to set the flashing digit.
5. Press the key, the third digit starts flashing.
6. Use the key to set the flashing digit.
7. Press the key to confirm the password once it has been set completely.
8. The calibration menu functions can now be selected by pressing the key.
Press the key to call the selected function.Press the key to cancel the active function.If no key is pressed for 10 minutes, the instrument
automatically returns to measuring mode.

5 Calibrate instrument

WARNING
Always calibrate the zero-point before span. Otherwise, the calibration will contain errors!
NOTICE
Dräger recommends using the extended bump test for cross calibrations.
Adjustment may not be possible due to instrument and
channel errors.
Allow the sensors to warm up before the calibration!Warming-up time: see instructions for use / data sheets for
the Dräger sensors installed (product page for X-am 2500 at www.draeger.com).
Improving the zero point accuracy – perform fresh air
calibration, page 16.
Set the sensitivity of all sensors to the value of the
test gas – carry out the 1-button calibration, page 17.
Set the sensitivity of a sensor to the value of the test gas –
span calibration/adjustment, page 18.

5.2 Run fresh air calibration

To improve the zero-point accuracy, a fresh air calibration can be carried out.
Calibrate the instrument to fresh air, free of measured
gases or other interfering gases.
Sensors which have not warmed up or which are faulty
prevent a calibration.
In the case of sensors which are in the warm-up phase,
the message » 159 « is displayed with the special symbol » « (for warning message).
In the case of a sensor or instrument error,
the message » 109 « is displayed with the special symbol » « (for a fault message).
The message is cleared after 5 seconds and the
function is available again in the menu.
During the fresh air calibration the zero point of all sensors (with the exception of the DrägerSensor XXS O
In the case of the DrägerSensor XXS O
to 20.9 vol. %.
1. Switch on instrument.
2. Depending on instrument configuration:
Open the Quick menu and select the fresh air
calibration function, page 15.
or
Open the Calibration menu and select the fresh air
calibration function, page 15.
The current gas
concentration values flash.
When the measured values
have stabilized:
3. Press the key to carry out the fresh air calibration.
) are set to 0.
2
, the display is set
2

5.1 Adjustment interval

Observe the relevant specifications in the Instructions for
Use/data sheets of the Dräger Sensors installed.
For critical applications, observe the recommendations in
EN 60079-29-2
1)
or EN 45544-42) and national regulations. We recommend that you adjust all the channels after 6months.
CAUTION
Never inhale the test gas. Health hazard! Observe the hazard warnings of the relevant Safety Data Sheets.
1) EN60079-29-2 – Guidelines for selection, installation, use and maintenance of instruments for the detection and measurement of flammable gases and oxygen.
2) EN 45544-4 – Electrical instruments for the direct detection and direct concentration measurement of toxic gases and vapours – Part 4: Guide for selection, installation, use and maintenance.
16 Dräger X-am 2500
The display containing the
current gas concentration changes with the display »OK«.
4. Press the key to confirm the calibration or wait for approx. 5 seconds.
Calibrate instrument
01533366_en.eps
CH
4
%LEL
O
2
Vol% CO
ppm H
2
S
ppm
i
i
i
i
i
i
00833366.eps
0,5 L/min
0
01633366_en.eps
CH4 %LEL
O
2
Vol % CO
ppm H
2S
ppm
OK
01733366_en.eps
CH
4
%LEL O
2
Vol% CO
ppm H
2
S
ppm
If a fault has occurred during the fresh air calibration:
The fault message » «
appears and » « is displayed for the respective sensor instead of the measured value.
In this case, repeat the
fresh air calibration.
Replace the sensor if
necessary, page 20.
NOTICE
Automatic surrogate calibration If the corresponding gas combination and the sensor are approved to do so, an automatic surrogate calibration and tests can be performed using the PC software Dräger CC-Vision
1)
. A gas for the bump test, for the adjustment and the measured gas can be set in the gas change wizard in Dräger CC-Vision. Conversions are performed automatically and no longer need to be made manually. The settings are also used by the Dräger X-dock.
1) The free of charge PC software Dräger CC-Vision can be downloaded from the following web address: www.draeger.com/software
5.3 Automatic fresh air calibration of the
CatEx sensor in the charging cradle
This feature is used to perform an automatic fresh air calibration of the CatEx sensor after placing it in the charging cradle. The feature can be adjusted using the Dräger CC-Vision PC software.
NOTICE
If the automatic fresh air calibration feature of the CatEx sensor is activated, the user must ensure that the charging cradle remains in normal atmosphere (21 vol.% O duration of the process.
) free of contaminants for the entire
2

5.4 1-button calibration

NOTICE
If no sensors are approved for the 1-button adjustment via the PC software Dräger CC-Vision, the 1-button adjustment menu function will not be available.
All sensors approved by the Dräger CC-Vision PC software
take part in the 1-button adjustment.
In the case of the 1-button calibration, the sensitivity of all
sensors is set to the value of the test gas. When using test gas cylinder 68 11 130 = mixed gas with 50 ppm CO, 15 ppm H
S, 2.5 vol. % CH4, 18 vol. % O2.
2
If a mixed gas with another
composition is used, the specified concentration values in the instrument must be changed to the target values of the mixed gas used using the PC software Dräger CC­Vision.
1. Connect the test gas cylinder with the calibration cradle.
2. Vent the test gas into a fume cupboard or into the open air (with a hose connected to the second connector of the calibration cradle).
CAUTION
!
Never inhale the test gas. Health hazard! Observe the hazard warnings of the relevant Safety Data Sheets.
3. Switch on the instrument and insert it into the calibration cradle until it engages.
4. Call the calibration menu, enter the password and select the 1-button calibration function, page 15.
5. Press the key to start the 1-button calibration.
6. Open the test gas cylinder valve to let test gas flow over the sensor. The currently displayed measured values start to flash. The flashing stops after a static measured value has been reached. The calibration is now carried out automatically. The displayed measured values change to the values according to the gas supplied.
7. The automatic stability monitoring can be overridden by pressing the key. A calibration is carried out immediately.
OK
If it is detected that no test gas has been supplied, the 1-button calibration is cancelled. The channels then display »n/a«. If only one sensor is included in the 1-button calibration, an adjustment is carried out in any case when
OK
the key is pressed.
When the calibration is completed and the displayed measured values have stabilised:
The display containing the
current gas concentration changes with the display »OK«.
8. Press the key or wait
OK
for 5 seconds to quit the calibration.
The instrument changes to
the measuring mode
9. Close the test gas cylinder valve and remove the instrument from the calibration cradle.
Dräger X-am 2500 17
Calibrate instrument
01833366_en.eps
CH4 %LEL
O
2
Vol% CO
ppm H
2S
ppm
00833366.eps
0,5 L/min
0
!
01933366_en.eps
CH4 %LEL
CO
2
ppm
O
2
Vol% CO
ppm
H
2S
ppm
NH
3
ppm
OK
02033366.eps
ch
4
%LEL
CO
2
ppm
O
2
Vol% CO
ppm H
2
S
ppm
NH
3
ppm
OK
02433366_en.eps
CH
4
%LEL
OK
02533366_en.eps
CH4 %LEL
CO
2
ppm
O
2
Vol% CO
ppm H
2S
ppm
NH
3
ppm
OK
If a fault occurs during the 1-button calibration:
The fault message » «
appears and » « is displayed for the respective sensor instead of the measured value.
In this case, repeat
the 1-button calibration or carry out a single gas calibration, see section
5.4.1 on page 18.
Replace the sensor if
necessary, page 20.

5.4.1 Calibrating the sensitivity for an individual measuring channel

The span calibration can be carried out specifically for
individual sensors.
In the case of the span calibration, the sensitivity of the
selected sensor is set to the value of the test gas used.
Use a standard test gas.
Allowed test gas concentration:
Ex: 40 to 100 %LEL O
10 to 25 vol. %
2
CO: 20 to 999 ppm
S: 5 to 99 ppm
H
2
Test gas concentration of other gases: see Instructions for Use for the respective DrägerSensors.

5.4.2 Sensitivity calibration for CatEx

Display on channel selection:
1. Press the key to start the calibration for the heat tinting or press the -key to select the next sensor.
If the displayed measurement value is stable:
2. Press the key to
OK
perform the calibration. The display containing the current gas concentration changes with the display
»OK«.
3. Press the key or wait for approx. 5 seconds to end the calibration of this measuring channel.
The next measuring channel is offered for calibration.
After the calibration of the last measuring channel, the device changes to measuring mode.
4. Close the test gas cylinder valve and remove the device from the calibration cradle.
1. Connect the test gas cylinder with the calibration cradle.
2. Vent the test gas into a fume cupboard or into the open air (with a hose connected to the second connector of the calibration cradle).
CAUTION
Never inhale the test gas. Health hazard! Observe the hazard warnings of the relevant Safety Data Sheets.
If a fault occurred during the span calibration.
The fault message » «
appears and » « is displayed for the sensor instead of the measured value.
In this case, repeat the
calibration.
If necessary, replace the
sensor, page 20.
3. Switch on the instrument and insert it into the calibration cradle.
4. Press the [+] key and keep it pressed for 5 seconds to open the calibration menu, enter the password and select the single gas calibration function, page 15.
5. Press the key to start the channel selection.
The display flashes the gas
OK
of the first measuring channel, e.g. » Ex %LEL «.
6. Press the key to start the calibration function of this measuring channel, or use the key to select another measuring channel (O ppm or CO - ppm).
18 Dräger X-am 2500
2
- vol. %, H2S -
5. Press the key to confirm the calibration gas concentration or use the [+] key to change the calibration gas concentration and complete the process by pressing the key. The measurement value flashes.
6. Open the test gas cylinder valve to let test gas flow over the sensor. The displayed, flashing measurement value changes to the value according to the supplied test gas.
7. Press the -key to select the next sensor.
OK
Operation with pump
Notice for the adjustment of the ex-channel to nonane as a measuring gas:
During the calibration of the ex-channel, propane can be
used as a substitute calibration gas.
When using propane to adjust the ex-channel to nonane, the display must be set to twice the used test gas concentration.
Notice for the use in subsurface mining:
For the calibration of the ex-channel to the measuring gas
methane, the display of the instrument must be set to a value of 5 % (relative) higher than the used test gas concentration.
Automatic fresh air calibration in the charging cradle (CatEx sensor only):
Calibrate the gas detector to fresh air, free of measured gases or other interfering gases. If the function is selected, a fresh air calibration of the CatEx sensor is performed automatically as soon as the gas detector is inserted in the charging cradle. This function can be selected or disabled using the CC-Vision PC software.
No calibration takes place if the warm-up is not yet complete:
Alarm LED is illuminated red.The acoustic signal sounds twice followed by three short
tones and the gas detector switches off.
Once the fresh air calibration has been successfully completed:
Alarm LED is illuminated red.The acoustic signal sounds once followed by three short
tones and the gas detector switches off.
The duration of the flushing phase depends on factors such as type and concentration of the gas or vapour to be measured, material, length, diameter and age of the sampling hose or probe. As a rule of thumb, a typical flushing time of 3 seconds per metre can be assumed for a sampling hose (factory-new, dry, clean). This flushing time applies in addition to the sensor response time (see instructions for use of the gas detector used).
Example: The flushing time for a 10 m hose is approx. 30 seconds. The sensor response time is approx. 60 seconds in addition. The overall time before reading the gas detector therefore is approx. 90 seconds. The flow alarm is delayed by 10 to 30 seconds, depending on the hose length.

6.1 Performing a measurement with the Dräger X-am Pump

Required accessories (see section 14 on page 28):
Dräger X-am PumpSampling hose and probes
Pump symbols:
Pump battery 100 % charged
Remaining charge of pump battery: 2/3
Remaining charge of pump battery: 1/3
Warning for pump (Gas detector can no longer detect pump.)
Leak test: Block suction inlet
Leak test: Release suction inlet
If a fault has occurred during the fresh air calibration:
The fault message appears and is displayed for the
respective sensor instead of the measured value.
In this case, repeat the fresh air calibration. If necessary, have the sensor replaced by qualified personnel.

5.5 Calibration with the X-dock maintenance station

The modular X-dock 5300/6300/6600 maintenance station (see order list) can be used to automatically perform calibrations and bump tests of several gas warning devices in parallel and independently of each other. More detailed information can be found in the corresponding instructions for use (see the X-dock product page at www.draeger.com).

6 Operation with pump

Observe the following when performing measurements using the pump
Perform visual inspection of the probe, if necessary.Wait for the flushing time to end.
Flush the Dräger sampling hose or Dräger probes prior to each measurement with the air sample to be measured.
The flushing phase is necessary to minimize or eliminate any effects associated with the use of a sampling hose or a probe, e.g. memory effects, dead volume.
Pump battery discharged
Commissioning and performing the measurement:
See instructions for use of the Dräger X-am Pump.
Viewing pump information:
1. Open the quick menu (see “Opening the Quick Menu” on page 15).
2. Select and confirm with the button.
OK
The following pump information will be displayed:
serial numberpump runtimepump battery chargepump status (activated or deactivated)
3. Press the button to return to measuring mode.
WARNING
!
No measurement! If the pump is connected but deactivated, the gas detector is not ready to measure. The red LEDs on the gas detector flash.
WARNING
!
¡Impairment of measuring accuracy! After measuring high nonane concentrations (>20 %LEL), the measuring accuracy for nonane is impaired. The pump is not suited for the continuous measuring of high nonane concentrations.
Dräger X-am 2500 19
Replacing the sensors
OK
i
i
IR
Calibration cradle (order no. 83 18 752) with inserted USB DIRA with USB cable (order no. 83 17 409)
USB DIRA with USB cable (order no. 83 17 409)
USB 2.0
USB 2.0
00733366_en.eps
0
0
X-am 2500
Activating or deactivating the pump:
1. Open the quick menu (see “Opening the Quick Menu” on page 15).
Select or and activate or deactivate the pump by
2. pressing the button.
3. Press the button to return to measuring mode.
6.2 Performing a measurement with a
manual pump adapter and rubber ball pump
Required accessories (see section 14 on page 28):
Manual pump adapterRubber ball pumpSampling hoseProbes
Commissioning and performing the measurement:
See instructions for use of the accessories used.

7 Replacing the sensors

CAUTION
Damage to components! There are components in the instrument that are sensitive to electric charge. Before opening the instrument to replace the sensor, ensure that the person performing the work is earthed to avoid damage to the device. Earthing can be safely ensured, e. g. via an ESD workstation (electrostatic discharge).
NOTICE
To open the instrument, unfasten the casing screws using a screwdriver (Torx T6).
To replace the sensors of the instrument, connect the instrument with a PC.
Replace the sensors using the PC program Dräger CC Vision.
Next:
Conduct the fresh air calibration page 16.
and then:
Calibrating sensitivity: either perform 1-button calibration, page 17 or run sensitivity calibration, see page 18.
20 Dräger X-am 2500
Troubleshooting

8 Troubleshooting

Fault Cause Remedy
Not possible to switch on the instrument Discharged power pack Charge the power pack, page 6.
Discharged alkaline batteries Insert new alkaline batteries, page 20.
Not possible to switch off the instrument The instrument is not set to measuring
mode
The instrument is configured to "Disable prohibited"
Display »––« Measuring range calibrated/adjusted
incorrectly
Electronics or sensors defective Must be repaired by Service.
List of the numerical codes of the warning and fault messages in the info mode, see page 15.

8.1 Warning messages

Select measuring mode.
Configure the instrument to "Disable allowed" with Dräger CC Vision.
Recalibrate/adjust the measuring range, page 20.
Special symbol » « and displayed numerical code:
152 Customer's service life counter about to elapse Reset the service life counter using Dräger
153 Database 90 % full Read the database soon and clear memory
154 Database full Read the database and clear memory.
155 Interval for bump test elapsed Conduct the bump test page 20.
156 Battery pre-alarm of X-am Pump Recharge the battery soon. The battery will last for
159 Calibration not possible. The menu function
251 DrägerSensor CatEx 125 PR warming up Wait until warm-up time is complete.
252 DrägerSensor CatEx 125 PR warming up Wait until warm-up time is complete.
253 Ex concentration has drifted into the
254 The temperature is too high Operate the instrument within the allowed
255 The temperature is too low Operate the instrument within the allowed
256 Calibration interval for DrägerSensor CatEx
257 Alarm threshold A2 is set to greater than
271 Heat conductance calibration interval for
272 Sensor is switched off due to excess gas Restart the device
Cause Remedy
CC Vision.
afterwards.
at least 20 minutes after the first battery pre-alarm.
cannot be carried out because of a message which is preventing the function (e.g. sensors in warm-up phase).
negative range
125 PR has expired
60 %LEL
DrägerSensor CatEx 125 PR has expired
Determine the message code via the info menu and switch it off, if necessary.
Conduct the fresh air calibration page 16.
temperature range.
temperature range.
Perform sensitivity calibration for DrägerSensor CatEx 125 PR, page 18.
Set alarm threshold to less than 60 %LEL.
Perform sensitivity calibration for DrägerSensor CatEx 125 PR, page 18.
351 DrägerSensor XXS EC1 is warming up Wait until warm-up time is complete.
352 DrägerSensor XXS EC1 is warming up Wait until warm-up time is complete.
353 EC1 concentration has drifted into the
negative range
Dräger X-am 2500 21
Conduct the fresh air calibration page 16.
Troubleshooting
Special symbol » « and displayed numerical code:
354 The temperature is too high Operate the instrument within the allowed
355 The temperature is too low Operate the instrument within the allowed
356 The calibration interval for DrägerSensor XXS
357 Alarm threshold A2 is set to greater than
451 DrägerSensor XXS EC2 in the warm-up phase Wait until warm-up time is complete.
452 DrägerSensor XXS EC2 in the warm-up phase Wait until warm-up time is complete.
453 EC2 concentration has drifted into the
454 The temperature is too high Operate the instrument within the allowed
455 The temperature is too low Operate the instrument within the allowed
456 The calibration interval for DrägerSensor
457 Alarm threshold A2 is set to greater than
Cause Remedy
temperature range.
temperature range.
EC1 has elapsed
60 %LEL
negative range
XXS EC2 has elapsed
60 %LEL
Run sensitivity calibration for DrägerSensor XXS EC1, page 18.
Set alarm threshold to less than 60 %LEL.
Conduct the fresh air calibration page 16.
temperature range.
temperature range.
Run sensitivity calibration for DrägerSensor XXS EC 2, page 18.
Set alarm threshold to less than 60 %LEL.
551 DrägerSensor XXS EC3 in the warm-up phase Wait until warm-up time is complete.
552 DrägerSensor XXS EC3 in the warm-up phase Wait until warm-up time is complete.
553 EC3 concentration has drifted into the
negative range
554 The temperature is too high Operate the instrument within the allowed
555 The temperature is too low Operate the instrument within the allowed
556 The calibration interval for DrägerSensor
XXS EC3 has elapsed
557 Alarm threshold A2 is set to greater than
60 %LEL
575 Calibration interval for the compensation
channel has elapsed
576 Calibration required because of overgassing. Adjust the sensitivity of the compensation channel.
651 DrägerSensor XXS EC 4 in the warm-up phase Wait until warm-up time is complete.
652 DrägerSensor XXS EC 4 in the warm-up phase Wait until warm-up time is complete.
653 EC 4 concentration has drifted into the
negative range
654 The temperature is too high Operate the instrument within the allowed
655 The temperature is too low Operate the instrument within the allowed
656 The calibration interval for DrägerSensor
XXS EC 4 has elapsed
657 Alarm threshold A2 is set to greater than
60 %LEL
Conduct the fresh air calibration page 16.
temperature range.
temperature range.
Run sensitivity calibration for DrägerSensor XXS EC 3, page 18.
Set alarm threshold to less than 60 %LEL.
Adjust the sensitivity of the compensation channel.
Conduct the fresh air calibration page 16.
temperature range.
temperature range.
Run sensitivity calibration for DrägerSensor XXS EC 4, page 18.
Set alarm threshold to less than 60 %LEL.
22 Dräger X-am 2500

8.2 Fault message

Troubleshooting
Special symbol »« and displayed numerical code:
102 The customer's service life counter has elapsed Reset the service life counter using Dräger
103 The instrument is defective The instrument must be repaired by Service.
104 Check sum error program code The instrument must be repaired by Service.
105 Bump test interval elapsed Run the bump test page 13.
106 The calibration interval has elapsed (at least
107 Bump test error (at least 1 channel has a bump
108 The instrument is defective The instrument must be repaired by Service.
109 The menu function cannot be carried out
111 Faulty alarm element test: Alarm light Repeat alarm element test using X-dock.
112 Faulty alarm element test: Alarm horn Repeat alarm element test using X-dock.
113 Faulty alarm element test: Vibration motor Repeat alarm element test using X-dock.
114 Defective parameter check Correct parameters and repeat test using X-dock
115 Instrument deactivated by X-dock Instrument activated using X-dock.
116 Faulty software update The instrument must be repaired by Service.
117 User parameters not feasible Check configuration of user parameters and adjust
118 Flow alarm of X-am Pump Check the gas circuit for obstructions and replace
121 Overvoltage on X-am Pump Contact DrägerService.
122 Battery main alarm of X-am Pump Charge pump.
Cause Remedy
CC Vision.
Run sensitivity calibration, see page 17 and/or
1 calibration interval has elapsed)
test error)
because of an error
page 18.
Run bump test, page 13 or run sensitivity calibration, page 17 and/or page 18.
Determine the error code via the info menu and switch it off, if necessary.
filters if necessary.
201 No valid zero point calibration of the
DrägerSensor CatEx 125 PR
202 No valid sensitivity calibration of the
DrägerSensor CatEx 125 PR
203 Measurement from DrägerSensor CatEx
125 PR is in negative range
204 DrägerSensor CatEx 125 PR not plugged in
or faulty
205 Error during bump test of DrägerSensor
CatEx 125 PR
207 Faulty rise time test Repeat rise time test using X-dock.
208 User parameters not feasible Check configuration of user parameters and adjust
218 Blocking alarm not plausible. Calibrate the sensor.
221 Too little oxygen to operate the DrägerSensor
CatEx 125 PR
222 No valid zero point calibration of the
DrägerSensor CatEx 125 PR for heat conduction
223 No valid sensitivity calibration of the
DrägerSensor CatEx 125 PR for heat conduction
224 Instrument incorrectly configured by Dräger
CC-Vision.
Conduct the fresh air calibration page 16.
Run sensitivity calibration, see page 17 and/or page 18.
Conduct the fresh air calibration page 16.
Check DrägerSensor CatEx 125 PR, page 20
Repeat bump test, where necessary, calibrate or replace the DrägerSensor CatEx 125 PR, page 20.
Operate sensor in an environment with at least 8vol. % O
Conduct the fresh air calibration page 16.
Run heat conduction sensitivity calibration, see page 17 and/or page 18.
Change sensor for applicable channel with Dräger CC-Vision.
.
2
Dräger X-am 2500 23
Troubleshooting
Special symbol »« and displayed numerical code:
301 No valid zero point calibration of the Dräger
302 No valid sensitivity calibration of the Dräger
303 The measured value of DrägerSensor XXS EC
304 DrägerSensor XXS EC1 is not inserted or faulty Check Dräger Sensor XXS EC1, page 20.
305 Error in bump test of Dräger Sensor XXS EC1 Repeat bump test, calibrate or replace
307 Faulty rise time test Repeat rise time test using X-dock.
308 User parameters not feasible Check configuration of user parameters and adjust
324 Instrument incorrectly configured by Dräger
326 Error during warm-up acceleration
401 No valid zero point calibration of the Dräger
402 No valid sensitivity calibration of the Dräger
403 The measured value of DrägerSensor XXS EC2
404 DrägerSensor XXS EC2 is not inserted or faulty Check Dräger Sensor XXS EC2, page 20.
405 Error in bump test of Dräger Sensor XXS EC2 Repeat function test, calibrate or replace Dräger
406 Faulty filter test Repeat filter test using X-dock.
407 Faulty rise time test Repeat rise time test using X-dock.
408 User parameters not feasible Check configuration of user parameters and adjust
424 Instrument incorrectly configured by Dräger
426 Error during warm-up acceleration Dräger
Cause Remedy
Conduct the fresh air calibration page 16.
Sensor XXS EC1
Sensor XXS EC1
1 is in the negative range
CC-Vision.
Dräger Sensor XXS EC1
Sensor XXS EC2
Sensor XXS EC2
is in the negative range
CC-Vision.
Sensor XXS EC2
Run sensitivity calibration, see page 18 and/or fresh air calibration, page 16.
Conduct the fresh air calibration page 16.
DrägerSensor XXS EC1, if necessary page 20.
Change sensor for applicable channel with Dräger CC-Vision.
Disconnect and reconnect power pack or replace the sensor. Sensor must not be loaded with gas within the first 5 minutes.
Conduct the fresh air calibration page 16.
Run sensitivity calibration, page 18.
Conduct the fresh air calibration page 16.
Sensor XXS EC2, if necessary page 20.
Change sensor for applicable channel with Dräger CC-Vision.
Disconnect and reconnect power pack or replace the sensor. Sensor must not be loaded with gas within the first 5 minutes.
501 No valid zero point calibration of the Dräger
Sensor XXS EC3
502 No valid sensitivity calibration of the Dräger
503 The measured value of DrägerSensor XXS EC3
504 DrägerSensor XXS EC3 is not inserted or faulty Check Dräger Sensor XXS EC3, page 20.
505 Error in bump test of Dräger Sensor XXS EC3 Repeat bump test, calibrate or replace
506 Faulty filter test Repeat filter test using X-dock.
507
508 User parameters not feasible Check configuration of user parameters and adjust
524 Instrument incorrectly configured by Dräger
525 No valid sensitivity calibration for the
24 Dräger X-am 2500
Sensor XXS EC3
is in the negative range
Faulty rise time test Repeat rise time test using X-dock.
CC-Vision.
compensation channel
Conduct the fresh air calibration page 16.
Run sensitivity calibration, page 18.
Conduct the fresh air calibration page 16.
DrägerSensor XXS EC3, if necessary page 20.
Change sensor for applicable channel with Dräger CC-Vision.
Carry out span calibration for compensation electrode.
Troubleshooting
Special symbol »« and displayed numerical code:
526 Error during warm-up acceleration Dräger
601 No valid zero point calibration of the Dräger
602 No valid sensitivity calibration of the Dräger
603 The measured value of Dräger Sensor XXS EC4
604 DrägerSensor XXS EC4 is not inserted or faulty Check Dräger Sensor XXS EC4, page 20.
605 Error in bump test of Dräger Sensor XXS EC4 Repeat bump test, calibrate or replace
606 Faulty filter test Repeat filter test using X-dock.
607 Faulty rise time test Repeat rise time test using X-dock.
608 User parameters not feasible Check configuration of user parameters and adjust
624 Instrument incorrectly configured by Dräger
626 Error during warm-up acceleration Dräger
Cause Remedy
Disconnect and reconnect power pack or replace
Sensor XXS EC3
Sensor XXS EC4
Sensor XXS EC4
is in the negative range
CC-Vision.
Sensor XXS EC4
the sensor. Sensor must not be loaded with gas within the first 5 minutes.
Conduct the fresh air calibration page 16.
Run sensitivity calibration, page 18.
Conduct the fresh air calibration page 16.
DrägerSensor XXS EC4, if necessary page 20.
Change sensor for applicable channel with Dräger CC-Vision.
Disconnect and reconnect power pack or replace the sensor. Sensor must not be loaded with gas within the first 5 minutes.
Dräger X-am 2500 25
Maintenance
i
i

9 Maintenance

9.1 Maintenance table

The instrument should be inspected and maintained by suitably qualified persons annually. Consult:
EN 60079-29-2 – Guide for the selection, installation, use
and maintenance of apparatus for the detection and measurement of combustible gases or oxygen
EN 45544-4 – Electrical apparatus used for the direct
detection and direct concentration measurement of toxic gases and vapours - Part 4: Guide for selection, installation, use and maintenance
national regulations
Recommended calibration interval for measuring channels Ex, O
, H2S, SO2, NO2 and CO: 6 months.
2
NOTICE
Calibration intervals of other gases: see Instructions for Use of the respective DrägerSensors.
Depending on instrument configuration:
Replace the alkaline batteries or charge the battery –
see section 3.1.2 on page 7 – after each use, at the latest after the battery alarm has been triggered or after 2 weeks.
Calibrating the instrument – see section 5 on page 16.
At regular intervals, according to the sensors used and
the operating conditions. For sensor-specific calibration data, refer to the Instructions for Use/data sheets of the sensors used
Before you carry out safety-related relevant
measurements, the zero point and sensitivity of the instruments should be tested in accordance with national regulations.
Inspection by suitably qualified persons – every year.
The inspection intervals must be established in each
individual case and shortened if necessary, depending on technical safety considerations, engineering conditions, and the technical requirements of the equipment.
We recommend that a service agreement be concluded
with Dräger and that repairs also be carried out by them.
Replace the sensors, page 20 – if necessary, when it is not
possible to calibrate the sensors any more.
1)
.

10 Cleaning

CAUTION
!
Abrasive cleaning tools (brushes etc.), cleaning agents and cleaning solvents can destroy the sensor diaphragm of the upper cradle.
For information on suitable cleaning agents and disinfectants and their specifications, see document 9100081 at www.draeger.com/IFU
The instrument does not need any special care.Dirt and deposits can be removed from the instrument by
.
washing it with cold water. A sponge can be used for wiping if necessary.
Carefully dry the instrument with a cloth.

11 Storage

Dräger recommends storing the instrument in the charger
module (order no. 83 18 639).
Dräger recommends checking the charge of the power
supply at least every three weeks if the instrument is not stored in the charger module.

12 Disposal

This product must not be disposed of as municipal waste. This is indicated by the adjacent icon. You can return this product to Dräger free of charge. For information please contact the national sales organisations and Dräger.
Batteries and rechargeable batteries must not be disposed of as municipal waste. This is indicated by the adjacent icon. Collect and dispose of batteries and rechargeable batteries at battery collection centres, in accordance with applicable regulations.
Electrochemical sensors
WARNING
!
Acid burn risk! Do not throw into fire or open with force.
As with batteries, dispose of as special waste in line with local waste disposal regulations. Further information can be obtained from the relevant local authority and from appropriate waste disposal companies.
The DrägerSensor CatEx 125 PR should be disposed of as electronics waste.
1) Instructions for use/data sheets for the Dräger sensors can be downloaded from the product page for the X-am 2500 on the following website: www.draeger.com. See also the enclosed instructions for use and data sheets for the sensors used.
26 Dräger X-am 2500

13 Technical data

13.1 X-am 2500

Ambient conditions:
during operation and storage Temperature class T4 (–20 to +50 °C):
Temperature across a short period: –40 to +50 °C
Air pressure 700 to 1300 hPa
Humidity 10 to 90 % (short-term up to 95 %) relative humidity
Storage time
X-am 2500 Sensors
Position of use any
Instrument data
Protection class IP 67 for instruments with sensors
Alarm volume Typically 90 dB (A) at 30 cm distance
Operating time:
Alkaline battery Typically 12 hours under normal conditions
NiMH power pack: T4 (type HBT 0000) T4 HC (type HBT 0100)
Dimensions approx. 130 mm x 48 mm x 44 mm (H x W x D)
Weight approx. 220 g to 250 g
Refresh interval for display and signals
1) Not part of the BVS10 ATEX E 080X and PFG 10 G 001X technical suitability test.
NiMH power packs type: HBT 0000, HBT 0100 Power pack type: ABT 0100
with alkaline single cell type: Duracell Procell MN 15001), Duracell Plus Power MN 1500
1)
Temperature class T3 (–20 to +40 °C): Power pack type: ABT 0100
with NiMH single cell type: GP 180AAHC
1)
with alkaline single cell type: Panasonic LR6 Powerline
Temperature class T3 (0 to +40 °C): Power pack type: ABT 0100
with alkaline single cell type: Varta 4006
1)
, Varta 4106
Maximum of 15 minutes with NiMH power pack T4 (HBT 0000) or T4 HC (HBT 0100) Prerequisite: storage of the instrument at room temperature (+20 °C) for at least 60 minutes in advance.
1 year 1 year
Typically 12 hours under normal conditions Typically 13 hours under normal conditions
1s
Technical data
1)
Dräger X-am 2500 27
Order list

14 Order list

Name and description
Dräger X-am 2500
Unlimited multi gas detector for 1 to 4 gases with replaceable sensors. With selectable special calibration. Default calibration for the ex-sensor: methane. Order no. 83 23 900 with default alarm thresholds that can be adjusted specifically for each country, all other versions have a default setting that is valid for all countries.
Dräger X-am 2500 standard instrument:
Dräger X-am 2500 Ex 83 23 910
Dräger X-am 2500 Ex, O
2
Dräger X-am 2500 Ex, O2, H2S LC 83 23 914
Dräger X-am 2500 Ex, O2, CO 83 23 916
Dräger X-am 2500 Ex, O2, CO, H2S LC 83 23 918
Basic instrument with selectable special calibrations, including a calibration certificate
Power packs:
NiMH power pack T4 (type HBT 0000) 83 18 704
NiMH power pack T4 HC (type HBT 0100)
Battery holder ABT 0100 83 22 237
Alkaline batteries T3 (2x)
Alkaline batteries T4 (2x)
1)
1
Battery and charger set (includes NiMH power pack T4, charger module for Dräger X-am 1/2/
5000 and plug-in power supply unit)
5)
Chargers:
Charging adapter for Dräger X-am 1/2/5000 83 26 101
Charger module for Dräger X-am 1/2/5000 83 18 639
Plug-in power supply unit in connection with adapter 83 25 736 for a maximum of 20 X-am 1/2/5000 charging modules
Plug-in power supply unit (worldwide) for max. 5 Dräger X-am 1/2/5000 charger modules
Plug-in power supply unit (worldwide) for 1 Dräger X-am 1/2/5000 charger module
Vehicle connection, 12 V/24 V for Dräger X-am 1/2/5000 charger module
Vehicle mount for 1 Dräger X-am 1/2/5000 charger module
Accessories
The accessories are not part of BVS10 ATEX E 080X and PFG 10 G 001X.
Order no.
83 23 912
83 23 900
83 22 244
83 22 239
83 22 240
83 18 785
83 21 850 83 25 736
83 16 994
83 16 997
45 30 057
83 18 779
Name and description
Order no.
Pump accessories:
Dräger Pump X-am 1/2/5000 83 19 400
Case for Dräger Pump X-am 1/2/5000 83 19 385
Dräger X-am Pump 83 27 100
Case for Dräger X-am Pump 83 27 104
USB power supply unit (for Dräger X-am
83 27 102
Pump)
USB cable (for Dräger X-am Pump) 83 27 108
Rubber ball pump 68 01 933
Manual pump adapter 83 19 195
Filter set for X-am 1/2/5000 83 19 364
Extension hoses and probes (excerpt):
Measuring probe2), 1.5 m
Telescopic probe2), plug-in
Telescopic probe2) 100 with accessories
Telescopic probe2) 150 stainless steel
Bar probe2) 90
3 mm tube connection set
Floating probe (3 mm), EPP, 3 m hose
Floating probe (3 mm), EPP, 10 m hose
2)
2)
2)
Fluoroelastomer hose (3 mm) incl. adapter,
2)
5m
Fluoroelastomer hose (3 mm) incl. adapter,
2)
10 m
Fluoroelastomer hose (3 mm) incl. adapter,
2)
20 m
Fluoroelastomer hose3) (3 mm), sold by the
64 08 239
68 01 954
83 16 530
83 16 533
83 16 532
83 27 641
83 25 831
83 25 832
83 25 705
83 25 706
83 25 707
83 25 837
metre, specify length when ordering
Hose (3 mm), PVC, sold by the metre, specify
83 25 838
length when ordering
Hose (rubber, 3 mm), CR-NR, sold by the
83 25 839
metre, specify length when ordering
Accessories for recording the measured values and for configuration:
Dräger CC-Vision (free full version available at www.draeger.com/software)
Dräger GasVision (trial version available at www.draeger.com/software)
Dräger GasVision licence key
83 25 646
(to upgrade from trial version to full version)
USB DIRA with USB cable (USB infrared
83 17 409 adapter for communication between Dräger X-am 1/2/5000 and PC)
28 Dräger X-am 2500
Order list
Name and description
Order no.
Calibration accessories:
Dräger X-dock, e.g. X-dock 5300 X-am 125 83 21 880
Bump test station, incl. mixed-gas cylinder 83 19 130
Calibration cradle for Dräger X-am 1/2/5000 83 18 752
Mixed gas cylinder
2.5 vol.% CH
, 18 vol.% O2, 15 ppm H2S,
4
68 11 130
50 ppm CO
Test gas cylinder, propane, 0.9 vol.% C3H8 in
68 11 118
air
On-demand controller 83 16 556
Standard controller 68 10 397
Other accessories:
Additional rubber housing boot 83 21 506
Leather carrying case 83 18 755
Spare parts
DrägerSensor CatEx 125 PR, 0 to 100 %LEL 68 12 950
DrägerSensor CatEx 125 PR Gas,
68 13 080
0 to 100 %LEL
DrägerSensor XXS O2, 0 to 25 vol.%
4)
DrägerSensor XXS CO LC, 0 to 2000 ppm
DrägerSensor XXS H2S LC, 0 to 100 ppm
4)
4)
68 10 881
68 13 210
68 11 525
DrägerSensor XXS NO2, 0 to 50 ppm 68 10 884
DrägerSensor XXS SO2, 0 to 100 ppm 68 10 885
1) Not part of the BVS 10 ATEX E 080X and PFG 10 G 001X technical suitability tests.
2) This accessory is optimised for the Dräger X-am Pump (market launch in 2017) (for hoses with 3 mm inner diameter).
3) 3 mm tube connection set (order no. 83 27 641) required.
4) Expected sensor life time: O CatEx > 3 years. Older type sensors (order no. 68 108 82 for CO or order no. 68 10 883 for H configured.
5) Sale of these parts to be discontinued at the end of 2020
S) can be registered in the instrument, but not
2
and CO >5 years,
2
Dräger X-am 2500 29
© Dräger Safety AG & Co. KGaA
Revalstraße 1 23560 Lübeck, Germany Tel +49 451 882 0 Fax +49 451 882 20 80 www.draeger.com
90 33 366 - TH 4638.205 en © Dräger Safety AG & Co. KGaA Edition 08 - December 2020 (Edition 01 - November 2012) Subject to alteration
Loading...