Any use of the device requires full understanding and strict observation of the Instructions for
Use supplied with the device.
The device is only to be used for the purposes specified here.
Maintenance
The maintenance intervals and measures stated in this Technical Handbook as well as the
specifications in the Instructions for Use of the used DrägerSensors
Repair of the instrument may only be carried out by trained service personnel.
Accessories
Do not use accessory parts other than those listed in the order list.
Safe coupling with electrical devices
Devices which are not mentioned in the Instruction for Use or in this Technical Handbook can
only be coupled electronically after consultation with the manufacturers or an expert.
Use in areas subject to explosion hazards
Devices or components for use in explosion-hazard areas which have been tested and
approved according to national, European or international Explosion Protection Regulations
may be used only under the conditions explicitly specified in the approval and with
consideration of the relevant legal regulations. The equipment or components may not be
modified in any manner. The use of faulty or incomplete parts is forbidden. The appropriate
regulations must be observed at all times when carrying out repairs on these devices or
components.
*)
must be observed.
Liability for proper function or damage
The liability for the proper function of the device is irrevocably transferred to the owner or
operator to the extent that the device was serviced or repaired by personnel not employed or
authorized by Dräger Safety or when the device was used in a manner not conforming to its
intended use.
Dräger Safety cannot be held responsible for damage caused by non-compliance with the
above recommendations.
The warranty and liability provisions of the terms of sale and delivery of Dräger Safety are
likewise not modified by the recommendations given above.
Dräger Safety AG & Co. KGaA
*)
The Instructions for Use of the DrägerSensors are supplied with the device on CD.
4
Intended Use
Intended Use
Portable gas measuring device for the continuous monitoring of the concentration of several
gases in the ambient air in the working area.
X-am 1100, X-am 1700:
independent measurement of four gases.
X-am 1100: 120 days service life from activation of the device (refer to page 11).
X-am 1700: 2 years service life from activation of the device (refer to page 11).
X-am 2000
independent measurement of one gas up to four gases.
Caution:
The device may only be used in mixtures of flammable gases and vapours with air – not
for use in oxygen-enriched atmospheres. Danger of explosion!
Particular care must be taken, for example, when using the device in the vicinity of
ethine-oxygen welding operations (ethine = acetylene).
If an unlit oxygen-enriched ethine jet is inadvertently directed towards the
DrägerSensor CatEx 125, an uncontrolled ignition may occur.
When using the DrägerSensor CatEx 125 in X-am 1100/1700/2000, a fresh air calibration
must be performed after an extreme impact load in the event that the zero point has changed
by more than 3 %LEL.
, depending on the device type:
5
Tests and Approvals
Tests and Approvals
Marking
X-am 1100, 1700 and 2000; approved as type LQG 00xx
I M2 / II 2G
0158
Ex ia d I/IIC T4/T3
DEMKO 06 ATEX 06.140055X
IECEx ULD 06.0001X
Um = 4.6 V, Im = 1.3 A
Only as to intrinsic safety for use in haz. loc.
Intrinsically safe / Securite Intrinseque Ex ia
Class I&II, Div. 1, Gr. A,B,C,D,E,F,G TC T4/T3
o
C ≤ Ta ≤ +50 / +40 oC
–20
Temperature class and ambient temperature: refer to the power pack!
ATTENTION:
Read the safety measures in the Instructions for Use.
Do not replace or charge batteries in potentially explosive areas.
Dräger Safety, 23560 Lübeck, Germany
Serial no.
Power pack 83 18 703; approved as type ABT 0000
Temperature class T4
–20
use with alkaline batteries
Energizer No. E91
Energizer No. EN91 (industrial)
Varta Type 4106 (power one)
Temperature class T3
–20
use with NiMH batteries
GP 180AAHC (1800 mAh)
*)
o
C ≤ Ta ≤ +50 oC
o
C ≤ Ta ≤ +40 oC
Power pack 83 18 704; approved as HBT 0000
Temperature class T4
o
–20
C ≤ Ta ≤ +50 oC
*)
Year of construction is coded by the third capital letter of the serial number: T = 2003, U = 2004, W = 2005,
X = 2006, Y = 2007, Z = 2008, A = 2009, B = 2010, C = 2011, etc.
Example: Serial number ARUH-0054: the third letter is U, so the year of construction is 2004.
6
Tests and Approvals
Intended Operating Area and Operating Conditions
Hazardous areas classified by zones
The device is intended to be used in hazardous areas or mines susceptible to firedamp
classified zone 1 or zone 2, within a temperature range of –20
o
C (depending on battery pack and batteries), where gases of explosion groups IIA, IIB
+50
o
C to +50 oC or –20 oC to
or IIC and temperature class T3 or T4 (depending on battery pack and batteries) may be
present.
If used in mines, the device is only to be used in areas known to have a low risk of mechanical
impact.
Hazardous areas classified by divisions
The device is intended to be used in hazardous areas or mines susceptible to firedamp
classified Class I&II, Div. 1 or Div. 2, within a temperature range of –20
o
C to +50 oC (depending on battery pack and batteries), where gases or dusts of groups
–20
o
C to +50 oC or
A, B, C, D or E, F, G and temperature class T3 or T4 (depending on battery pack and
batteries) may be present.
Safety Instructions
ATTENTION:
To reduce the danger of explosion, do not mix new batteries with old batteries and do not
mix batteries made by different manufacturers.
ATTENTION:
Always disconnect the device from the power pack before carrying out any maintenance
operations.
ATTENTION:
Substitution of components may impair intrinsic safety.
CAUTION:
Not tested in an oxygen-enriched atmosphere (>21% O
).
2
7
What is What
+
What is What
Front panel
1
Gas entry
2
Alarm LED
3
Buzzer
OK
on
4
key
M
5
key
6
Display
1
2
2
3
6
Rear panel
1
IR interface
2
Fastening clip
3
Type plate
4
Charging contacts
5
Power pack
Display
1
Measured gas display
2
Measured value display
3
Special symbols
5
0
4
00223854_01.eps
1
2
3
5
4
00323854_01.eps
123
8
00423854_01_en.eps
Special Symbols
Fault message, refer to page 14
Warning message, refer to page 14
The peak value display for all measuring gases, refer to page 14
What is What
The exposure evaluation display (TWA) for measuring gas H
refer to page 14
S and CO,
2
The exposure evaluation display (STEL) is switched on for measuring rangasge H
and CO, refer to page 14
The device is set to function test with gas (bump test),
refer to page 28
The device is set to the fresh air calibration function,
refer to page 35
The device is set to the 1-button calibration function,
refer to page 37
The device is set to the single-gas calibration function,
refer to page 39
The function for password entry is active, refer to page 16
Battery / rechargeable battery 100 % full
Battery / rechargeable battery 2/3 full
Battery / rechargeable battery 1/3 full
Battery / rechargeable battery empty
S
2
9
Configuration
Configuration
Standard Gas Configuration
DrägerSensorCatEx 125
[%LEL]
Measuring range
Alarm A1
- setpoint
- can be acknowledged
- self-latching
Alarm A2
- setpoint
- can be acknowledged
- self-latching
2)
2)
0 to 1000 to 250 to 20000 to 200
20
yes
no
2)
40
no
yes
Standard Device Configuration on Delivery
XXS O
2
[Vol.-%]
1)
19
no
yes
23
no
yes
X-am 1100 / X-am 1700 / X-am 2000
XXS CO
[ppm]
30
yes
no
60
no
yes
XXS H2S
[ppm]
10
yes
no
20
no
yes
Function test with gas (bump test) in
Quick Menu
2)
Fresh air calibration in Quick Menu
Life sign
Switch off
LEL factor
2)
2)
2)
2)
4.4 (4.4 vol. % corresponds to 100 %LEL)
off
off
on
not disabled
Changing the standard configuration: See “Configuring the Device” on page 20.
Attention!
After a basic initialization has been carried out with the PC software "CC Vision",
individual alarm settings may have been changed.
___________________
1)
In the case of O2 A1 is the lower alarm setpoint: an alarm is issued if the value is too low.
2)
Different settings can be selected to meet customer requirements on delivery. The current setting can be
checked and changed with the software "CC Vision".
A version of Dräger CC-Vision which can be used for Dräger X-am 1100/1700/2000 is supplied with the
device on CD.
10
Activating the Device
+
_
_
p
Before using the device for the first time,
insert the supplied batteries or a charged
NiMH power pack T4 (order no. 83 18 704),
refer to Replacing the Batteries, page 41.
Charge the rechargeable batteries if
necessary, page 42.
The Dräger X-am 2000 is ready for
operation.
Only in the case of Dräger X-am 1100 and
Dräger X-am 1700:
The following activation sequence also has to
be carried out:
●
Press and hold the key for approx. 3
seconds until the countdown
shown in the display has elapsed.
M
» 3 . 2 . 1 «
Activating the Device
— The remaining operating time is
displayed, e.g.
» d 730 « (remaining
operating time 730 days).
— After that, the device is automatically
switched off again. You can switch it on
again at any time by pressing the key.
OK
on
s
de.e
01
02223854
11
Operation
+
Operation
Switching on the device
OK
● Press and hold the key for approx. 3 seconds until the countdown » 3 . 2 . 1 « shown in
the display has elapsed.
— All the display segments, including the visual, audible and vibration alarms, are activated for
a short time.
— The software version is displayed.
— The device performs a self test.
— The remaining operating time (not in the case of X-am 2000) is displayed, e.g. » d 730 «
(remaining operating time 730 days).
— All alarm setpoints A1 and A2 as well as TWA
in succession.
— The sensor which is next due for calibration/adjustment is displayed with the days remai-
ning until the next calibration/adjustment e.g. » Ex %UEG CAL 20 «.
— During the warm-up time of the sensors, the respective display of the measured value
flashes and the special symbol » « (for warning) is displayed.
● Press the key to cancel the display of the activation sequence.
OK
on
on
*)
and STEL*) for H2S and CO are displayed
Switching off the device
OK
● Press and hold the key and the key at the same time until the countdown » 3 . 2 . 1 «
shown in the display has elapsed.
— Before the device is switched off, the visual, audible and vibration alarms are activated for
a short time.
*)
Only when activated in the device configuration. Delivery status: not activated.
12
on
M
Operation
Before entering the workplace
CAUTION:
Check and, if necessary, adjust the calibration before carrying out safety-relevant
measurements.
A function test with gas (bump test) must be carried out in accordance with local
regulations.
● Switch on the device. The current measured values are shown in the display.
● Observe any warning » « or fault » « messages.
The device can be operated normally. If the warning message does not go out
automatically during operation, the device must be maintained after the end of
use.
The device is not ready to measure and requires maintenance.
If one of these special symbols is displayed, appropriate measures, refer to page 22 to
page 24, must be taken.
● Check that the gas inlet opening on the device is not covered.
CAUTION:
The presence of catalyst poisons in t he measured gas (e.g.volatile silicone, sulphur, heavy
metal compounds or halogenated hydro-carbons) can damage the DrägerSensor CatEx
125. If the CAT Ex sensor cannot be calibrated to the target concentration anymore, the
sensor must be replaced.
The display of the Cat Ex sensor may be incorrect in an oxygen-poor atmosphere.
Electrical operating safety (Ex protection) is not guaranteed in an oxygen-enriched
atmosphere.
During operation
During operation, the measured values for every measured gas are displayed.
If a measuring range is exceeded or a negative drift occurs, the following displays are shown
instead of the measured value display:
» « (too high concentration) or
» « (negative drift).
In the event of an alarm, the corresponding displays, including the visual, audible and vibration
alarms are activated – refer to “Identifying Alarms” on page 18.
13
Operation
Calling the Info Mode
OK
● In measuring mode, press the key for approx. 3 seconds.
● The peak values as well as the TWA
succession (press the key for the next display).
The peak values = the maximum measured values in the case of e.g. CO, H2S, ... or
the minimum measured values in the case of O
displayed
OK
⇓⇓
on
⇓⇓
key
The average values of the exposures based on a shift of e.g. 8 hours (TWA) of all
the active sensors for the exposure evaluation are displayed
OK
⇓⇓
on
⇓⇓
key
The short-term values (STEL) = average values of the concentrations over the
average value duration of all the active sensors for the exposure evaluation are
displayed
OK
⇓⇓
on
⇓⇓
key
The device is in measuring mode again
— If no key is pressed for 10 seconds, the device returns automatically to measuring mode.
on
*)
OK
on
and STEL*) exposure values are displayed in
within the storage interval are
2
*)
Only when activated in the device configuration. Delivery status: not activated.
14
Calling the Quick Menu
+M+
+
— If functions were activated for the quick menu with the PC software "CC Vision"
functions are activated in the quick menu on delivery):
(no
Operation
● In measuring mode, press the key three times.
M
If no functions have been activated in the quick menu, the device remains in measuring
mode.
— You can select the activated functions of the quick menu by pressing the key.
OK
● Press the key to call the selected function.
on
Possible functions of the quick menu
Function test with gas (bump test), refer to page 28
Fresh air calibration, refer to page 35
● Press the key to cancel the active function and to switch to measuring mode.
M
— If no key is pressed for 60 seconds, the device returns automatically to measuring mode.
15
Operation
+
_
p
+
+
+
+
Calling the Calibration Menu
— The calibration menu can only be
accessed by entering a password.
Password on delivery: » 001 «
— The default password on delivery can be
changed using the PC software "CC
Vision".
● In measuring mode, press the key for
M
at least 5 seconds.
— The function for entering the password is
selected.
— The special symbol » « (for the "enter
password" function) is displayed.
— The display shows » 000 « with the first
digit flashing.
● Use the key to set the flashing digit.
● Press the key, the second digit starts
M
OK
on
flashing.
● Use the key to set the flashing digit.
● Press the key, the third digit starts
M
OK
on
flashing.
● Use the key to set the flashing digit.
● Press the key to confirm the password
M
OK
on
once it has been set completely.
— The calibration menu functions can now
be selected by pressing the key.
OK
● Press the key to call the selected
on
M
function.
s
01.e
02323854
16
Calibration menu functions
+
Fresh air calibration, refer to page 35
1-button calibration, refer to page 37
Single gas calibration, refer to page 39
Operation
● Press the key to cancel the active function.
M
— If no key is pressed for 60 seconds, the device returns automatically to measuring mode.
17
Identifying Alarms
Identifying Alarms
An alarm is displayed visually, audibly and through vibration in a specific pattern.
Concentration pre-alarm A1
The alarm is indicated by an intermittent alarm message:
Display » A1 « and measured value alternating: not for O
— The pre-alarm A1 is not self-latching and stops when the concentration has dropped below
the alarm setpoint A1.
— In the case of A1 a single tone is audible and the alarm LED flashes.
Acknowledging a pre-alarm:
● Press the key. Only the audible alarm and the vibration alarm are switched off.
OK
on
Concentration main alarm A2
The alarm is indicated by an intermittent alarm message:
Display » A2 « and measured value alternating:
In the case of A2 a double tone is audible and the alarm LED flashes twice.
: » A1 « and measurement value alternating = oxygen deficiency
For O
2
» A2 « and measurement value alternating = oxygen surplus
CAUTION:
Leave the area immediately. You are in mortal danger! A main
alarm is self-latching and cannot be acknowledged or cancelled.
!
2
After leaving the area, if the concentration is less than the alarm setpoint A2:
● Press the key. The alarm messages are switched off.
OK
on
STEL / TWA exposure alarm
The alarm is indicated by an intermittent alarm message:
Display » A2 « and » « or » « and measured value alternating:
CAUTION:
Leave the area immediately. After this alarm, the deployment of
personnel is subject to the relevant national regulations.
— The STEL and TWA alarm cannot be acknowledged or cancelled.
● Switch off the device. The values for the exposure evaluation are deleted after the device
is switched on again.
18
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.