RETURN TO TECHNICAL SERVICE BULLETIN TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Vapor 2000 w/ Selectatec Mount
TECHNICAL SERVICE BULLETIN # 1
Product: Vapor 2000 w/ S-2000 (Selectatec mount)
Date: October 16, 2002
Re: 6 Month and 24 Month Maintenance Interval procedure
Reference Doc: See attached: “Service Instructions for Vapor S-2000 (M 34821) on Vapor
2000”
Reason/Rational: Some Vapor 2000 vaporizers w/ S-2000 mounts can be hard to turn on.
Solution/Procedure: Perform the attached procedure at the suggested interval(s).
Devices affected: Vapor 2000 w/ S-2000 mount (Selectatec mount) distributed to Canada.
Vapor 2000 w/ S-2000 mounts are not distributed in the U.S. and are
therefore not affected.
Timing/When: Every 6 months and every 24 months as prescribed in the attached
procedure.
Additional Info: Please contact Mike Munsell (x. 2285) to obtain the specialized tool
needed for the attached service.
It is recommended that the Dow Corning Gn-Plus Lubricant specified in the
attached procedure, be obtained locally from this suggested supplier:
A.E. Blake Sales LTD.
Montreal: Toronto:
3596 boulevard Poirier 31 Estate Drive
Montreal, QC H4R 2J5, Canada Toronto, ON M1H 2Z2, Canada
Tel: 514-332-4214 Tel: 416-431-0440
Fax: 514-332-2860 Fax: 416-431-0450
Toll Free: 1-800-879-5748 Toll Free: 1-888-664-2643
montrealsales@aeblake.com torontosales@aeblake.com
Cost: No reimbursement for performing procedure. To be performed as part of
routine maintenance.
Ordering Info/Parts: None. See “Additional Info” above.
Distribution: Lifetronics Medical Inc., Brathwaites Olivier Medical Inc.
DrägerService If you have any questions, please contact Technical Support by phone at
3122 Commerce Drive 1-800-543-5047 or by fax at 215-721-5784.
Telford, PA 18969
Tel: (215) 721-5402
(800) 543-5047
Fax: (215) 723-5935
Web: www.draegermedical.com
DrägerService is a division of
Draeger Medical, Inc.
Page 1 of 3
RETURN TO TECHNICAL SERVICE BULLETIN TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Dräger Medizintechnik
D
1 Service Instructions for Vapor S-2000 (M 34821) on Vapor 2000
1.1 Six-Month Maintenance Intervals
Inspection at 6-month intervals including report carried out by qualified service personnel in
connection with Vapor 2000.
Inspection:
− The lever can be easily pressed through the interlock pins.
− The control dial can be opened easily. The control dial can be turned evenly and easily
over the full adjustment range.
If the lever or the control dial is sluggish, refer to the procedure described in the 24-month
maintenance intervals section.
1.2 24-Month Maintenance Intervals
Apart from the 6-month inspection intervals, the plug-in adapter S-2000 (M 34821) must be
maintained at 24-month intervals.
Required material and spare parts:
− Molykote grease G-n plus, tube 100 g (1554166)
− Loctite 221 (1195727) only after removal/replacement of the lever and/or the bar.
Procedure:
• Remove the two bearing cases.
• Clean the interlock pins and bearing cases (inside).
•Inspect the sliding area of the interlock pins (A and B seeFig. 1:). If there are visible
signs of wear, replace the interlock pins.
• Apply a thin layer of Molykote grease on the sliding surface of the interlock pins (A and B
seeFig. 1:).
Fig. 1: Sliding area of the interlock pins
For internal use only. Copyright reserved.
GBR5327400T02.fm 06.07.00
5327.400 Vapor 2000 06/2000 Ser vice Instructions Page 1