PSS® 7000 C/W SENTINEL® 7000 ELECTRONIC
MONITORING UNIT (EMU)
User Instructions
Page 2
Page 3
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
1
Index Page No:
Section 1 Approvals, For Your Safety and Liability 04
Section 2 Description and Intended Use 06
Section 3 Technical Data 08
Section 4 Preparation for Use 10
Section 5 Pre Operational Checks 13
Section 6 Putting on the SCBA 17
Section 7 During Use 18
Section 8 After Use 20
Section 9 Cleaning the SCBA 21
Section 10 Storage - Ready for Use 21
Section 11 Special Instructions 22
Section 12 Cautions and Limitations 22
Section 13 CBRN Use 23
Section 14 S – Special or Critical Users’ Instructions 23
Section 15 RIC UAC Use 24
Section 16 Contact Details 24
Section 17 Equipment Maintenance Procedures 25
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 4
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
2
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 5
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
Liability Statement
Dräger, while endeavouring to ensure correctness of statements of fact and advice contained
in this publication; gives no guarantee or warranty in respect thereof, and accepts no liability
for any mis-statement or inaccuracy in the publication, or for any omission therefrom.
3
The liability for the proper function of the
(SCBA) referred to in this publication
the SCBA is serviced, or repaired, by personnel not employed or authorised by Dräger, or
if the SCBA is used in a manner not conforming to its intended use. Dräger cannot be held
responsible for damage caused by non-compliance with the recommendations given above.
The warranty and liability provisions of the terms of sale and delivery of Dräger are likewise
not modified by the recommendations given above.
Terms and Conditions of warranty for the Dräger PSS® 7000 Series of SCBA can be obtained
from Dräger on request.
Dräger Safety Limited
self contained compressed air breathing apparatus
is irrevocably transferred to the owner or operator if
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 6
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
4
1 Approvals, For Your Safety, Warranty and Liability
This User Instruction details the Preparation for Use, Operation and After Use
procedures required to use the Dräger PSS® 7000 Series of SCBA fitted with the
following:
1. Dräger Sentinel 7000 - Electronic Monitoring Unit and PASS
2. Wireless HUD System – Head Up Display (HUD)
It is recommended that this manual is read, and fully understood, before attempting
to use the equipment.
been thoroughly trained in the correct use of the device.
Approvals
The Dräger PSS® 7000 Series are open-circuit, pressure-demand, self-contained
compressed air respiratory apparatus (SCBA) certified by NIOSH and in certain
combinations for CBRN use, and also by SEI to meet the requirements of NFPA
1981:2007. The apparatus must only be used in conjunction with compressed air
cylinders approved by NIOSH. For non-CBRN use refer to the separate NIOSH Approval
Label matrix table – 3356261 for allowable configurations. For CBRN use refer to
the separate NIOSH CBRN Approval Label insert – 3356226 for allowable respirator
configurations. For CBRN use – the user must also refer to the additional CBRN User’s
Instruction 3356259.
This device may only be worn by expert personnel who have
For Your Safety
General safety information for the Dräger PSS® 7000 Series of self-contained breathing
apparatus (SCBA):
Use of this apparatus requires relevant equipment training; observance of this user
instruction (UI); and compliance with national regulations, laws and standards governing
the use of respiratory apparatus.
Use of this apparatus must be only for the purpose specified in this UI, or as confirmed
in writing by Dräger.
Only trained and competent personnel should inspect and service the apparatus at
specified intervals. Records of inspections and servicing are to be maintained in line
with NFPA 1852 – Standard on Selection, Care, and Maintenance of Open-Circuit SelfContained Breathing Apparatus.
Use of the apparatus should be consistent with NFPA 1500 – Standard on Fire Department
Occupational Safety and Health Program.
The automatic alarm function of the Sentinel 7000 detects movement of the wearer and
movement or vibration to which the wearer may be subjected.
Do not use any form of chemical marking or paint on the equipment.
Dräger recommend that a service contract be obtained from your Dräger branch or
agent.
Contact Dräger for details of service contracts and training courses.
Use only original Dräger spare parts and test equipment for service, repair and
maintenance.
Notify Dräger if there is component fault or failure.
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 7
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
All approved respiratory equipment shall be selected, fitted, used, and maintained in
accordance with Mine Safety and Health Administration (MSHA), Occupational Safety
and Health Administration (OSHA), and other applicable regulations.
5
Warranty and Liability Statement
Terms and conditions of warranty for the Dräger PSS® 7000 Series of SCBA and
associated accessories can be found in the original Dräger invoice document, or
can be obtained from Dräger on request.
The Dräger guarantee is void if the original sealing caps on the apparatus are tampered
with, removed or broken
Responsibility for the reliable function of the apparatus transfers to the owner or
operator when it is serviced or repaired by personnel not authorised by Dräger, or
when it is used in a manner not conforming to its intended use.
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 8
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
6
2 Description and Intended Use
This User Instruction (UI) describes the Dräger PSS® 7000 Series of SCBA fitted
with the following:
A Dräger Sentinel 7000 electronic monitoring system, with integral personal
alert safety system (PASS).
A Dräger wireless head-up display (HUD) unit.
The Dräger PSS® 7000 is a compact and lightweight apparatus that provides the
wearer with respiratory protection when working in contaminated or oxygen-deficient
conditions. The system can be used as a self-contained system or with an independent
air supply to increase effective operation time. The system is fully compatible with a wide
range of compressed-air cylinders, face masks and lung demand regulators (LDRs).
The carrying system has a height-adjustable articulating backplate (6, Fig 1) that
provides improved comfort and ease of use, resulting in increased manoeuvrability
and reduced fatigue. The carbon-composite backplate incorporates a sliding heightadjustable yoke that has three preset positions. A flexible pivot joint located at the
base of the backplate (in line with the hip of the wearer) moves in response to the
twisting and bending of the wearer, improving the weight distribution and freedom of
movement. The system includes adjustable shoulder straps (2) and waist belt (11).
The Sentinel 7000 electronic monitoring system is a multi-function system that
provides continuous monitoring of the SCBA status including remaining cylinder
pressure, movement of the wearer, main battery condition, end-of-service time
(EOST) and PASS. It provides visual indications of system status and audible and
visual alarms in warning conditions. User control and monitoring of the system is
through a user interface (1) that incorporates switches, LEDs, a liquid crystal display
(LCD) display screen and an alarm sounder. A backlight illuminates the user interface
display screen when required. Additional alarm sounders (5), with warning LEDs top
and bottom, are mounted on the backplate and operate only during PASS alarms.
The operating settings are preset and non-adjustable by the user.
The PASS function of Sentinel 7000 electronic monitoring system is an alarm system
that can be activated manually or automatically. The automatic alarm uses a motion
sensor to detect movement and activate a pre-alarm and main alarm at timed intervals
when no movement is sensed. The manual alarm is activated by a press button on
the user interface. A limitation of the PASS is that it detects movement or vibration
to which the wearer is subjected.
Power supplies include a main battery located in the backplate, a back-up battery
located the pressure module and a HUD battery located in the HUD. The system is
switched on by a press button on the user interface or by cylinder pressure felt at a
pressure module, with cylinder pressure transmitted to the pressure module through
a high-pressure hose (7). At switch on, a self test is performed and then the system
adopts the active mode where the PASS function is operational.
All variants utilise the same high-performance, first-stage, regulator (9) fitted with an EOST
mechanical whistle (10). The pressure regulator supplies medium-pressure breathing
air to a quick-release coupling (4) via a medium-pressure hose (3). Incorporated in the
first-stage regulator is a rapid intervention crew universal air connection (RIC UAC) (8).
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 9
D
The RIC UAC is a male coupling that allows emergency replenishment of breathing air
to the air cylinder while wearer is breathing from apparatus.
A wireless HUD unit (refer to instruction no. 3356234) locates in the face mask and
provides visual indications to the wearer of a number of system conditions. Signals
from the Sentinel 7000 to the HUD are from a pressure transducer/transmitter
incorporated in the pressure module.
Page 10
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
3 Technical Data
High Pressure (HP) Connections
The following SCBA HP connections are available for associated compressed air cylinder
assemblies.
8
2216 psig connection to CGA 346
4500 psig connection to CGA 347
Quick Connect – 2216psig or 4500psig (Option) – Refer to User Instruction.
Power Supplies
Main battery – 7.5 V
Back-up battery – 3 V
Head-up display battery – 3 V.
RIC UAC Connection
2216 psi or 4500 psi male, quick-release coupling with pressure relief valve
(PRV).
Dräger push-in connector.
Medium pressure – 87 psi to 130 psi
Air flow – In excess of 1000 litres/minute
Air flow at 290 psi – In excess of 500 litres/minute.
Activation commencement range (mechanical and electronic):
2216 psi cylinder – 600 psi to 510 psi
4500 psi cylinder – 1215 psi to 1035 psi.
LDR to Face mask Connection
Pressure and Flow Details
EOST Alarms
General Information
Batteries
Danger of explosion. Do not change the batteries in an explosive or flammable
!
atmosphere.
Do not dispose of batteries in a fire. Batteries must be disposed of in line with
!
local regulations.
Do not attempt to recharge any non-rechargeable battery.
!
Battery information:
Main battery 7.5 V (5 x 1.5 V AA alkaline batteries)
Back-up battery 3 V (CR123 lithium battery)
HUD battery 3 V (CR123 lithium battery).
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 11
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
Use only the following approved battery types:
Duracell MN1500 (1.5 V)
Energizer LR6 (1.5 V)
Panasonic CR123AL/1BP (3 V).
Batteries are supplied with the equipment but are not fitted. Dräger
i
i
recommend that the batteries be removed when the system is not used
for long periods as a small amount of discharge still occurs. The normal
operating life of the batteries is dependent on usage time, frequency of alarms,
backlight illumination and ambient temperature.
The back-up battery will only supply power when the main battery is
i
i
disconnected or discharged. When this occurs, the back-up battery will
only supply power for HUD functions.
9
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 12
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
4 Preparation for Use
Equipment configurations for non-CBRN use are detailed in a NIOSH Approval Table
(refer to instruction no. 3356261). Configurations for CBRN use are detailed later
in this instruction (refer to CBRN Use).
Installing the Back-Up Battery
Orientate the SCBA to access the battery compartment.
Unscrew and remove the battery cap (1, Fig 2), using a suitable coin.
Insert the battery, +ve terminal end first, into the battery compartment.
Refit and secure the battery cap. Do not overtighten.
10
1
Fig. 2
2199
Fig. 3
2195
Installing the Main Battery
Orientate the SCBA to access the pressure module (1, Fig 3).
Inspect the terminals on the battery and pressure module, and the sealing rim around
the battery terminals (all must be clean and undamaged).
Locate the two slots in the top of the battery pack (2) on the tabs in the battery
housing.
Firmly press the battery into position until an audible click is heard, to indicate a
secure attachment. The Sentinel 7000 will emit a single tone and will commence
the self-check sequence (refer to Sentinel 7000 Self Check).
Press and hold the RH and LH buttons of the user interface until the display clears,
then immediately release the buttons.
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 13
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
A battery check is performed
i
i
during the self-check
sequence. If the battery
voltage is below a preset voltage
the backlight will illuminate; the
low battery icon (Fig 4) will be
displayed; the Sentinel 7000
will emit a series of tones for
approximately four seconds; and
then the display will switch off. When
this occurs replace the battery.
To remove the battery insert
i
i
and press the two pronged
key (supplied with the SCBA)
into the two slots at the base of the
battery. This will open the locking
latch allowing the battery to be
removed.
11
Fig. 4
2580
Installing the Cylinder
The following instructions are for a screw-on cylinder coupling. Installation
i
i
of a quick-connect cylinder coupling is detailed in a separate UI (refer to
instruction no. 3356260).
Set the carrying-frame backplate to the short (S) position (refer to Adjusting the
Backplate Height).
Inspect and check the following:The external thread of the cylinder valve port.
The internal thread of the first-stage regulator handwheel.
The O-ring seal in the pressure connector is in position and not damaged.
The bore to the sintered filter in the HP connector of the first-stage regulator is clean
and free from dirt and contamination.
Lay the carrying frame horizontal and fully extend the cylinder strap.
Take care to prevent impact damage as the cylinder valve is aligned with the first-
stage regulator handwheel.
Insert the cylinder (valve end first) through the cylinder strap, to align the cylinder
valve with the handwheel.
Lift the cylinder and backplate into the vertical position (supported on the end of the
cylinder).
Align and fully tighten the handwheel (clockwise). Do not use tools or overtighten.
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 14
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
Place the unit back into the horizontal
position.
Take up the slack of the cylinder
strap and activate the cam-lock
mechanism by pulling the free end
of the strap back over the cylinder
(Fig 5).
Secure the strap using the Velcro
fastening.
Adjusting the Backplate
Height
12
Fig. 5
2186
Lift the SCBA into the vertical
position.
Simultaneously press the two spring-
loaded buttons (Fig 6) of the locking
catch.
Slide the yoke to the required
position (short (S), medium (M) or
long (L)) and release the springloaded buttons.
Fig. 6
Grasp the yoke and the frame and
attempt to raise and lower the yoke
to confirm that the locking catch is
fully engaged.
2006
Connecting the LDR and HUD
Insert and push the male coupling of the LDR hose into the quick-release coupling
of the SCBA medium-pressure hose until it latches into position. Check the security
of the connection (do not connect the LDR to the face mask at this stage).
Install a serviceable HUD into the face mask (refer to instruction no. 3356234).
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 15
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
5 Pre Operational Checks
Sentinel 7000 Self Check
The self check can be performed with or without a pressurised cylinder
i
i
fitted.
Press the LH button of the user interface, or open the cylinder valve of a pressurised
cylinder. The following self-check sequence will commence:
First stage – The Sentinel 7000 will emit a single tone from the user interface and
the additional alarm sounders, and the display backlight will illuminate.
Second stage – The display will show a tick symbol (Fig 7); the blue, red and green
LEDs (Fig 8) will illuminate.
13
bar
psi
MPa
Fig. 7
2191
Third stage – The display will show
the cylinder type (Fig 9) (2216 psi
or 4500 psi)
Fourth stage – The display will
show the normal operating screen
(Fig 10); the blue, red and green
LEDs (Fig 8) will illuminate; the
blue and red LEDs (1, Fig 11) will
illuminate.
BRGRB
Fig. 8
2200
psi
End of sequence – Two ‘trill’ alarms
will sound; the display will show the
normal operating screen (Fig 10);
the green LED (Fig 8) will flash at
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Fig. 9
2190
Page 16
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
i
i
i
i
psi
14
Fig. 10
2192
approximately one second intervals
to confirm that the Sentinel 7000
has passed the self check and is in
the active mode.
If a cross icon (Fig 12) with
i
i
a fault code is displayed, the
Sentinel 7000 has failed the
self check. Contact Dräger
for repair.
Press and hold the RH and LH
buttons of the user interface until
the display clears, then immediately
release the buttons.
High Pressure Leak Test
and EOSTI Warning Test
The following is not intended as
a test of HUD functionality. If the
HUD fails to operate as described,
refer to the HUD UI (instruction no.
3356234).
Fig. 11
2210
Fig. 12
2201
1
Place the face mask next to the
SCBA (within three feet of the
pressure module).
Press the reset button (1, Fig 13)
of the LDR to switch off the positive
pressure.
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Fig. 13
2117
Page 17
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
Open the cylinder valve slowly, but
fully, to pressurise system. The
Sentinel 7000 and HUD systems
will activate as follows:
The Sentinel 7000 will emit a single
tone and will commence a self-check
sequence (refer to Sentinel 7000
Self Check).
Up to approximately 45 seconds after
the start of the self check, all six
HUD LEDs (Fig 14) will flash twice
to indicate that the Sentinel 7000 is
communicating with the HUD.
Depending on cylinder pressure,
some of the line of four HUD LEDs
(red/amber/green/green) will flash
(on for 15 seconds/off for 45
seconds).
Close the cylinder valve and observe the user interface display. The pressure reading
shall not decrease more than 200 psi in 1 minute.
If the SCBA fails this test, or an immediate leak is evident, do not continue with the
check. Vent the pressure, switch off the system and contact Dräger for repair.
R
A
G
G
B
G/Y
LED
Fig. 14
2189
15
Cover the outlet of the LDR with the ball of the hand and press the front button (2,
Fig 13).
Carefully lift the ball of the hand to slowly vent the system – observe the pressure
displayed on the user interface.
Note the pressures for the electronic EOST activation (indicated by an audible alarm
tone; flashing red and blue LEDs on the user interface; flashing red LED on the HUD)
and the mechanical whistle operation. Electronic and Mechanical activation does
not need to be simultaneous but both must be within the acceptable range (refer to
Technical Data – EOST Alarms). Refer to the Service Manual if the pressures are
outside the acceptable range.
Allow the system to vent to zero pressure. The display will show zero pressure and
all of the user interface LEDs (Fig 8) will flash at approximately one second intervals.
The green LED indicates that the system is still in active mode.
Press and hold the RH and LH buttons of the user interface until the display clears,
then immediately release the buttons.
The HUD red LED (Fig 14) will flash intermittently.
After approximately 180 seconds all six HUD LEDs will flash twice to indicate that
the unit has logged off from the Sentinel 7000.
Press the reset button of the LDR to switch off the positive pressure.
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 18
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
Check that the face mask port and the connector O-ring of the LDR are clean and
undamaged.
Align and push the LDR into face mask port until it latches in position.
Check the attachment by attempting to pull the LDR away from the face mask.
Normal Operation
All safety information, including the preliminary safety details, must be read and
!
understood before commencing any user task. The procedures detailed in this
UI may only be carried out by trained and competent SCBA users.
Before use checks
Carry out a visual inspection of the complete SCBA including any ancillary equipment that
will be used during the operation. The tasks detailed in the preparation for use and the preoperational checks sections must also be completed satisfactorily before release of the SCBA
for operational use. Components that are worn or damaged must be repaired or replaced
before proceeding. Do not use the equipment if any of the following is found:
The SCBA fails to meet any of the test specifications.
Any casing is cracked or broken.
The LCD display of the user interface cannot be read, or one or more of the display
segments is inoperative.
Audible and/or visual alarms are not functioning.
Any fault icon shown.
16
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 19
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
6 Putting on the SCBA
This apparatus may only be
!
worn by trained and competent
personnel.The user must
ensure that any accessories,
ancillary equipment, turnout gear
and other clothing items do not
interfere with the SCBA and do not
create a safety hazard.
If necessary, adjust the backplate
height (refer to Adjust the Backplate
Height).
Loosen the shoulder straps and waist
belt and put on the SCBA.
Check that the shoulder pads are not twisted and take the weight of the system on
the shoulders by pulling the shoulder straps. Do not fully tighten at this stage.
i
s
p
Fig. 15
17
2193
Close the waist belt buckle and pull the ends of the waist belt forward until the strap
padding fits securely and comfortably over the hips (Fig 15). Tuck the belt ends behind
the hip pads.
Pull the shoulder straps until the SCBA rests securely and comfortably on the hips.
Do not overtighten. Tuck the strap ends behind the waist belt.
Fully loosen the head straps of the face mask and place the neck strap over the back
of the neck.
Press the reset button (1, Fig 13) of the LDR to switch off the positive pressure.
Open the cylinder valve slowly, but fully, to pressurise system. The Sentinel 7000 and HUD
systems will activate.
Putting on the Face Mask
Use only the mask sizes that have been confirmed by a quantitative fit test
!
(QNFT).
Select the correct size of mask and inner mask.
Put on the face mask (for non-CBRN use, refer to instruction no. 9021649; for CBRN use,
refer to instruction no. 3356259).
Press the front button (2, Fig 13) of the LDR and check that air flow is delivered into the
face mask.
Press the reset button (1) to stop the air flow.
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 20
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
Close the cylinder valve and breathe normally to empty the system of pressure.
When the system is empty, inhale and hold breath for approximately 8 seconds. Readjust
and retest until the face mask contracts and holds on to the face with no air leaks.
Open the cylinder valve slowly, but fully, to pressurise system.
7 During Use
Do not commence any operation (including supplied-air respirator (SAR)
operations) using a cylinder that is less than 80% full. The effective working
duration of the apparatus is dependent on the capacity of the air cylinder
!
and the breathing rate of the wearer.
On activation if the unit fails to operate, or a low battery alarm activates,
i
i
then the relevant batteries should be renewed.
Regularly observe warning and
display signals, and note the
following:
18
For emergency help or assistance,
press the yellow button in the centre
of the user interface to activate the
psi
PASS alarm (see below).
To illuminate the display backlight,
press and release the LH or RH
button of the user interface.
Remaining cylinder pressure is
indicated on the user interface and
the HUD LEDs. The user interface
shows the cylinder pressure digitally
Fig. 16
2202
(numeric) and as an analogue segment display (Fig 16). The HUD shows the cylinder
pressure with LEDs (refer to the Table 1). The user interface digital display is the
most accurate indicator.
Remaining time to EOST alarms is indicated on the user interface as a digital (numeric)
display.
React to the following alarm and warning signals as necessary:EOST – The user interface will emit an audible alarm tone, and red and blue LEDs
will flash; the red LED on the HUD will flash; the mechanical whistle on the first-stage
regulator will sound.
PASS pre-alarm – A repeating audible alarm tone will be emitted from the user
interface sounder and the additional alarm sounders. Move the user interface to
cancel the alarm (do not attempt to use the buttons to switch off the pre-alarm.
PASS alarm – A high-level sweeping alarm will be emitted from the user interface
sounder and the additional alarm sounders; red and blue LEDs on the user interface
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 21
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
(Fig 8) and top and bottom of the
additional alarm sounders (Fig 11)
will flash intermittently; the user
interface will show the alarm icon
(Fig 17). Simultaneously press and
hold the RH and LH buttons of the
user interface to cancel the alarm.
Low main battery – A low battery icon
(Fig 4) will be displayed on the user
interface or the G/Y battery LED (Fig
14) will flash yellow.
Low HUD battery – The G/Y battery
LED will flash green.
Loss of HUD communication – The
HUD blue LED flash
Additional air flow can be delivered into the face mask as follows:Press and release the bypass button (3, Fig 13) to deliver a single jet of air into the
face mask.
Additional air flow required – Press and rotate the bypass button to deliver a
sustained air supply (80 to 130 litres/minute) into the face mask
Excessive or loss of air flow – Close the cylinder valve then immediately begin to
slowly reopen the valve. Use the cylinder valve as a regulating valve to set the air
flow to meet the user requirement. This procedure can also be used with ratchettype cylinder valves.
psi
Fig. 17
2198
19
The air flow procedures above are emergency measures that may greatly
!
reduce the operating duration of the air supply. When activated the
user must immediately evacuate to a safe area. The reason for using
the procedure must be investigated and repaired before reusing the
apparatus.
Table 1 Head-up display LEDs
Key On Flashing R Red A Amber G Green
LED's
Cylinder
Contents
Full to 3/4
3/4 to 1/2
1/2 (for 20 secs once)
1/2 to 1/4
1/4 to approx. 100 psig
R
E
D
A
M
B
E
R
G
R
N
G
R
E
E
E
E
N
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 22
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
8 After Use
Do not remove the SCBA until in a safe area. Do not drop or throw down
!
the SCBA as damage could occur.
Release and loosen all of the face mask straps.
Lift and remove the face mask.
Fully extend all of the straps of the head harness.
Close the cylinder valve.
Release the waist belt buckle, lift shoulder strap buckles to loosen and remove the
apparatus.
Press the front button (2, Fig 13) of the LDR to vent system.
Press the reset button (1) to switch off the positive pressure.
If the LDR has been set to bypass, press and rotate the bypass button (3) to switch
off the bypass.
20
Press and hold the RH and LH buttons of the user interface until the display clears,
then immediately release the buttons. After approximately 180 seconds, all six HUD
LEDs will flash twice to indicate that the unit has logged off.
Carry out a visual inspection of the complete SCBA including any ancillary equipment
used during the operation.
If required, lightly lubricate the LDR O-ring (recommended lubricant is Molykote 111).
Pass the SCBA to the service department, with details of any faults/damage that
occurred during use.
Removing the Cylinder
Do not attempt to remove the cylinder with the system pressurised.
!
Close the cylinder valve and fully vent the system.
Lift the free end of the cylinder strap from the Velcro fastening and lift the strap
against the cam-lock mechanism to release the strap tension. Loosen the strap.
Unscrew the handwheel of the first-stage regulator from cylinder valve.
Lift and carefully slide the cylinder away from first-stage regulator towards the top of
the backplate.
Remove the cylinder from the apparatus.
Inspect and recharge the cylinder (refer to Charging the Cylinder).
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 23
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
21
Charging the Cylinder
Air quality for compressed air cylinders must conform to the minimum grade
!
requirements for Type 1 gaseous air as defined in the CGA Commodity
Specification for Air, G-7.1 (Grade D or higher quality) and where appropriate
be in accordance with: NFPA 1989 Standard on Breathing Air Quality for
Emergency Services Respiratory Protection.
Recharge cylinders to the indicated working pressure marked on the
!
cylinder.
If the moisture content exceeds the recommended levels, ice particles can
!
form, reducing or blocking the airflow.
9 Cleaning the SCBA
Use only recommended cleaning agents. Do Not use organic solvents, such
!
as acetone, alcohol, white spirit, trichloroethylene or similar. Do not use
any form of mechanical, electrical or ultrasonic agitation in cleaning baths.
Refer to manufacturers’ instructions when using cleaning and disinfecting agents.
Particular attention should be paid to concentration and reaction times.
Do not exceed 86 °F for washing and rinsing solutions. Do not exceed 140
!
°F for drying. Do not immerse the LDR.
Recommended cleaning and disinfecting agent is liquid AirKem A-33. Full
i
i
details of cleaning and disinfecting agents are available from Dräger on
request.
Carefully clean SCBA components (use suitable cleaning baths, and immerse and
manually agitate components in solution as necessary).
Thoroughly remove any cleaning solutions by rinsing in clean water.
Dry all components, including internal parts
10 Storage - Ready for Use
Fully extend the shoulder straps, waist belt and head harness straps.
Store apparatus in a cool dry environment, free from dust and dirt (do not expose to
direct sunlight).
When the SCBA is stored with the cylinder fitted, the apparatus must be supported by
the cylinder and not the backplate.
Remove the batteries if the equipment is not used for long periods (a small amount of
discharge occurs during storage).
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 24
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
22
11 Special instructions
Use of an Independent Air Supply (SAR Airline Connection)
Air quality must conform to the statutory requirements.
!
Use of an airline connection by a second person (buddy-breather) voids NIOSH
approval.
The time required for the wearer to escape to a safe area must be within the remaining
breathing time of the cylinder, taking into account the remaining air content in the
cylinder and the breathing rate of the wearer.
Independent air supplies must meet the following standards:
Type-1 gaseous air as defined in: CGA Commodity Specification for Air, G-71 (grade
D or higher)
NFPA 1989 Standard on Breathing Air Quality for Fire and Emergency Services
Respiratory Protection
Air supply pressure – 87 psi to 125 psi Airline hose length – 5 feet to 300 feet (maximum working hose length must not exceed
12 individual hose lengths)
Airline flow rate – 550 litres/minute Approved for use at temperatures above minus 25 degrees Fahrenheit (minus 31.7
degrees Centigrade).
Sentinel 7000 alarms and warning signals will operate as normal.
i
i
The user interface display will show cylinder pressure when the cylinder valve
is open.
Turn on the independent air supply.
Connect the independent air supply coupling to the secondary supply hose (refer to
instruction no. 3355853) and breathe normally.
Close the cylinder valve (if the EOST warning signal activates, silence by taking several
short deep breaths or momentarily operating the LDR bypass button (3, Fig 13).
If any air supply problems are encountered, proceed as follows:Open the cylinder valve to return to breathing from the attached cylinder.
Disconnect the independent air supply coupling.
Leave the hazardous area by the shortest and safest escape route, if necessary.
12 Cautions and Limitations
D – Air-line respirators can be used only when the respirators are supplied with respirable
air meeting the requirements of CGA G – 7.1, Grade D or higher quality.
E – Use only the pressure ranges and hose lengths specified in the User’s
Instructions.
I – Contains electrical parts that may cause an ignition in flammable or explosive
atmospheres.
J – Failure to properly use and maintain this product could result in injury or death.
M – All approved respirators shall be selected, fitted, used, and maintained in accordance
with MSHA, OSHA, and other applicable regulations.
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 25
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
N – Never substitute, modify, add, or omit parts. Use only exact replacement parts in the
configuration as specified by the manufacturer.
O – Refer to User’s Instructions and/or maintenance manuals for information on use and
maintenance of these respirators.
S – Special or critical User’s Instructions and/or specific use limitations apply. Refer to
User’s Instructions before donning.
13 CBRN Use
The Dräger PSS® 7000 Series is certified by National Institute for Occupational Safety
and Health (NIOSH), for limited chemical, biological, radiological or nuclear (CBRN)
use and by the Safety Equipment Institute (SEI) to meet the requirements of NFPA
1981. Approvals are only valid when the apparatus is used with compressed air cylinders
approved by NIOSH. Equipment configurations for CBRN use are detailed in a CBRN
Approval Label (refer to instruction no. 3356226).
If an FPS NFPA LDR, Dräger recommend that a quantitative fit test (QNFT) be performed
on the face mask before use in a CBRN environment. The fit test must be conducted
strictly in accordance with the requirements outlined in the OSHA Respiratory Protection
Standard 29 CFR, Section 1910.134.
23
Q – Use in conjunction with personal protective ensembles that provide appropriate
levels of protection against dermal hazards.
R – Some CBRN agents may not present immediate effects from exposure, but can
result in delayed impairment, illness or death.
T – Direct contact with CBRN agents requires proper handling of the SCBA after each
use and between multiple entries during the same use. Decontamination and disposal
procedures must be followed. If contaminated with liquid chemical warfare agents,
dispose of the SCBA after decontamination.
U – The respirator should not be used beyond 6 hours after initial exposure to chemical
warfare agents to avoid possibility of agent permeation.
14 S – Special or Critical Users’ Instructions
Approved for use at temperatures above minus 25 degrees Fahrenheit (minus 31.7
degrees Centigrade).
When used as a combination supplied-air respirator/self-contained breathing
apparatus (SAR/SCBA), not more than 20 per cent of the air supply can be used during
entry.
During supplied air use, the cylinder valve must remain closed. If the supplied air
fails, open the cylinder valve and immediately proceed to fresh air.
Supplied air source must meet the following criteria: pressure 87 – 125 psi, air
flow rate at least 550 litres/minute.
Important Note: If it is decided to exit the working area with the airline disconnected
or, in an emergency, if the air supply fails, breathe normally and immediately proceed
as follows:
Open the cylinder valve (counterclockwise) slowly, but fully and breathe normally.
Sentinel 01: Edition 1: July 2008 - Subject to Modification
Page 26
PSS® 7000 SERIES c/w SENTINEL 7000 EMU
Disconnect the hose of the independent air supply from the male coupling of the airline
hose connection. Breathe normally and immediately leave the hazardous area by the
shortest and safest route.
Safety Warning: The remaining duration begins from the time of opening the cylinder
valve and disconnecting the independent air supply. The time required to allow the
wearer to escape to a safe area must be within the remaining air capacity (volume)
of the cylinder taking into account the breathing rate of the wearer.
15 RIC UAC Use
Use of the RIC UAC should be by trained and competent personnel only.
The RIC UAC filling hose is a required component of the NFPA 1981 certification.
Only use a filling hose which has been certified to NFPA 1981 for use in immediately
dangerous to life or health (IDLH) atmospheres.
Do not use the RIC UAC connection for second person (buddy-breather).
Do not use the RIC UAC support pack to transfer air from one compressed air breathing
apparatus to another.
Do not allow oil, grease or other contaminants to contact the RIC UAC connection.
Do not attempt to disassemble or repair the RIC UAC connection.
Caution: The secondary air supply pressure to the RIC UAC must not exceed maximum
rated working pressure of the cylinder(s) being filled.
Caution: If the pressure relief valve of the RIC UAC is activated, the SCBA must be
returned to the nearest Dräger branch or agent.
Caution: If a leak is detected while refilling in a contaminated or oxygen-deficient
gaseous atmosphere, stop refilling and immediately leave the hazardous area.
24
16 Contact Details
Any issues with the equipment, including damage, malfunction, or failure of the
breathing apparatus that may present a hazard to the user should be reported to: