Strictly follow the Instructions for Use.
The user must fully understand and strictly observe the
instructions. Use the product only for the purposes specified
in the Intended use section of this document.
!
Dräger Polytron 8100
Instructions for Use
Page 2
Page 3
Contents
1For your safety .........................................................5
Before using this product, carefully read the Instructions for
Use (IfU).
Strictly follow the Instructions for Use. The user must fully
understand and strictly observe the instructions. Use the
product only for the purposes and under the conditions
specified in this document.
Do not dispose of the Instructions for Use. Ensure that they
are retained and appropriately used by the product user.
Only fully trained and competent users are permitted to use
this product.
Comply with all local and national laws, rules and
regulations associated with this product.
Only trained and competent personnel are permitted to
inspect, repair and maintain the product as detailed in
these Instructions for Use. Further maintenance work that
is not detailed in these Instructions for Use must only be
carried out by Dräger or personnel qualified by Dräger.
Dräger recommends a Dräger service contract for all
maintenance activities.
Maintenance must be performed as described, see
Section 6 on Page 28.
Use only genuine Dräger spare parts and accessories.
Otherwise the proper functioning of the product may be
impaired.
The flameproof / explosion proof joints are not in
accordance with the relevant minimum or maximum values
of EN/IEC 60079-1. The joints are not intended to be reworked by the user.
WARNING
Substitution of components may impair Intrinsic
Safety. This only applies to the instrument or parts of it
which are classified as intrinsically safe.
Only operate the product within the framework of a risk-
based alarm signaling concept.
1.2Definitions of alert icons
The following alert icons are used in this document to provide
and highlight areas of the associated text that require a greater
awareness by the user. A definition of the meaning of each
icon is as follows:
DANGER
!
Indicates an imminently hazardous situation which, if
not avoided, will result in death or serious injury.
WARNING
!
Indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
!
Indicates a potentially hazardous situation which, if not
avoided, could result in physical injury, or damage to
the product or environment.
It may also be used to alert against unsafe practices.
NOTICE
i
i
Indicates additional information on how to use the
product.
Safe connection of electrical devices
Before connecting this instrument to electrical devices not
mentioned in the IfU consult the manufacturer or an expert.
Using the product in areas subject to explosion hazards:
Instruments or components for use in explosion-hazard
areas which have been tested and approved according to
national, European or international Explosion Protection
Regulations may only be used under the conditions
specified in the approval and with consideration of the
relevant legal regulations.
The instruments or components may not be modified in any
manner. The use of faulty or incomplete parts is forbidden.
The appropriate regulations must be observed at all times
when carrying out repairs on these instruments or
components.
Dräger Polytron 81005
Page 6
Description
4
6
1
3
5
2
00133298.eps
4
8
1
6
7
3
5
2
00633298.eps
2Description
2.1Product overview
2.1.1Explosion proof instrument
2.1.2Explosion proof instrument with increased safety
wiring compartment (Docking Station)
1 Enclosure lid
2 Handle
3 Bucket with main electronics (and optional relay)
4 Enclosure bottom
5 Sensor
6 Assembled instrument
1 Enclosure lid
2 Handle
3 Bucket with main electronics (and optional relay)
4 Enclosure bottom
5 Sensor
6 Feed-through cable
7 Docking Station
8 Assembled instrument
6Dräger Polytron 8100
Page 7
Description
!
!
2.2Intended use
The Dräger Polytron 8000 is an explosion proof instrument for
continuous monitoring of toxic gases or oxygen in the ambient
air. The instrument is housed in a rugged stainless steel or
aluminum enclosure for indoor and outdoor applications. The
instrument can be connected through a sealed conduit or
approved cable gland to a Dräger monitoring system or a
Programmable Logic Controller (PLC). With the optional alarm
relay configuration, the instrument can be operated as a standalone unit. The instrument is designed to be installed in
permanent locations and is approved for use in hazardous,
classified areas, see Section 10.1 on Page 37. With the
optionally integrated relay module, the instrument can be
operated without a central controller (with additional local
alarm signaling).
For safe operation, the connection to a central controller is
necessary. This also applies if all relays are connected
because the fault relay is not activated during maintenance. In
any case, the analog output must be monitored because it
might indicate a special status by a current between 0 and 3.5
mA.
CAUTION
Not tested in oxygen enriched atmospheres
(>21% O2). High off-scale readings may indicate an
explosive concentration.
2.3Intended operating area and conditions
Hazardous areas classified by zones:
The device is intended to be used only in hazardous areas
classified Zone 1 or Zone 2, within a temperature range as
marked on the device, where gases of explosion Groups IIA,
IIB or IIC and temperature class T4 or T6 (depending on the
maximum ambient temperature) or dusts of Groups IIIA, IIIB or
IIIC may be present.
Hazardous areas classified by divisions:
The device is intended to be used only in hazardous areas
Class I or II, Div. 1 or Div. 2, within a temperature range as
marked on the device, where gases or dusts of Groups A, B,
C, D or E, F, G and temperature class T4 or T6 (depending on
the maximum ambient temperature) may be present.
2.4Design
The instrument is powered by 10 to 30 VDC (see Section 10.4
on Page 41). Gas concentrations, status messages, and menu
choices are displayed on a backlit graphic LC-display and 3
colored LEDs. The measured gas concentration is converted
to a 4 to 20 mA analog (per NAMUR recommendation NE43)
or a digital HART output signal. The instrument can be
operated as current source or sink.
Navigation through the menu is done by taping a magnetic
wand on the glass at the appropriate indicator.
WARNING
Danger of explosions. Not to be used in oxygen
enriched atmospheres. None of the Polytron 8000
instruments are certified and approved to be operated
in oxygen enriched atmospheres.
The instrument can be configured, calibrated and maintained
non-intrusively without declassifying the area.
Dräger Polytron 81007
Page 8
Installation
3Installation
3.1General information for installation
To ensure overall system performance and effectiveness, the
selection of an installation site for the instrument is one of the
most important factors. Strict compliance where possible and
considerable thought must be given to every detail of
installation, particularly:
the local, state, national codes and regulations that govern
the installation of gas monitoring equipment.
the electrical codes that govern the routing and connection
of electrical power and signal cables to gas monitoring
equipment.
for non-conduit installations, an approved cable gland must
be used, see Section 10.1 on Page 37 (e.g. Hawke A501/
421/A/¾” NPT or equivalent). It might be necessary to
connect the shield of the cable to the cable gland and to the
controller in order to improve RFI immunity.
the full range of environmental conditions to which the
instruments will be exposed to.
the physical data of the gas or vapor to be detected.
the specifics of the application, (e.g. possible leaks, air
movement/draft, etc.).
the degree of accessibility required for maintenance
purposes.
the types of optional and accessory equipment that will be
used with the system.
any other limiting factors or regulations that would affect
system performance or installations.
the flameproof / explosion proof enclosure provides three
¾" NPT openings, which can be used for field wiring, direct
attachment of a sensor or wiring of a remote sensor.
unused openings must be closed with a plug. For correct
tightening torques of conduit hubs, cable glands, plugs and
sensor see Section 10.8 on Page 42.
secondary circuit intended to be supplied from an isolating
source (N/A for relay circuits).
the optional increased safety terminal box provides up to
four 20 mm openings, which can be used for field wiring or
wiring of a remote sensor. The permissible cable diameter
range is 7 to 12 mm.
when installed at locations exceeding ambient
temperatures of 55 °C, use only appropriate wiring,
specified for at least 25 °C above the maximum ambient
temperature.
strip wire insulation by 5 to 7 mm.
connect the wires as indicated in wiring figure Section 3 on
the wiring for the optional relay module must be selected
and fused according to the rated voltages, currents and
environmental conditions.
if stranded conductors are used, a ferrule must be used.
3.2Installation restrictions
The instrument must have between 10 to 30 VDC at the
instrument. This ultimately determines the distance the
instrument can be mounted from the controller or power
supply. The instrument accepts wire sizes of 12 to 24 AWG
(0.2 to 2.5 mm2). Use at least a three-conductor, shielded
cable.
The instrument must not be exposed to radiant heat that
will cause the temperature to rise beyond the limits stated
in see Section 10.6 on Page 41. The use of a reflecting
shield is recommended.
The enclosure is weatherproof within environmental
parameters and suitable for outdoor installation. The use of
the optional splash guard (6812510) is recommended to
protect the sensor from water and dust.
Each instrument must be installed and operated in an
environment that conforms to the specifications, see
Section 10 on Page 37.
To insure proper operation of the instrument, the
impedance of the 4 to 20 mA signal loop must not exceed
500 Ohms. Depending on the operating voltage and
according to the application (e.g. HART operation), certain
minimum impedances must be observed (see Section 3.3
on Page 8). The conductors for the power supply must
have an adequately low resistance to insure the correct
supply voltage at the instrument.
NOTICE
i
i
The instrument may be equipped with a dust plug at
the conduit entry. This plug is not explosion proof nor
meant to be watertight, and must be removed before
connecting the instrument to a sealed conduit or
installation of a flameproof cable gland.
3.3Impedance of signal loop
.
Operating mode
Operation
without HART
communication
Operation with
HART
communication
(HART mode)
HART Multidrop
operation
Impedance range of the
signal loop
0 to 230 Ωat 10 V DC
Rising linearly with the
supply voltage from:
0 to 230 Ω at 10 V to
0 to 500 Ω at 16 V
0 to 500 Ω18 to 30 V DC
230 to 270 Ωat 13 V DC
Rising linearly with the
supply voltage from:
230 to 270 Ω at 11 V to
230 to 500 Ω at 16 V
230 to 500 Ω18 to 30 V DC
230 to 500 Ω10 to 30 V DC
Supply
voltage range
10 to 18 V DC
11 to 16 V DC
8Dräger Polytron 8100
Page 9
Installation
!
01133300.eps
VDC
+
-
mA
01033300.eps
VDC
+
-
mA
!
3.4Mechanical installation
Use the drilling template for mounting on a flat surface.
The mounting surface should be even.
Dräger recommends using M6, 1/4” bolts with hex socket
caps.
The instrument must be accessible for maintenance (e.g.
calibration).
The future use of accessories and maintenance equipment
must be kept in mind.
The access of the gas or vapor to the sensor must not be
obstructed.
3.5Electrical installation without Docking
Station
CAUTION
Ensure wiring for relays and connections for sensor
are made before applying power.
Connection diagram for operation as a current source
3.5.1Power and signal wiring
Loosen set-screw and un-
screw lid from instrument.
Lift the handle and pull-out
the bucket with the main
electronics.
Turn bucket over and pull
off the 5-pin connector.
Connect the three wires for
power and signal to the
appropriate terminal as
indicated in the following
wiring table and figure.
Fasten terminal screws
with the correct torque
according to the table
00933300.eps
tightening torque and wire size for field wiring terminal, see
Section 10.9 on Page 42.
5-way Pin connector (power and signal)
Pin12345
MarkPWR+ PWR- PWR- 4-20 mAPE
Function V+V-V-4 to 20 mA signal
(Operation as
source or sink)
PE
Connection diagram for operation as a current sink
If operated as a stand-alone instrument, wire pin 3 to pin 4.
Plug connector back into socket and tighten screws.
The cable shielding should only be connected at the
controller.
Place bucket back into the enclosure.
Screw the lid back on, until it is seated (see Section 10.8
on Page 42) and tighten set-screw.
3.5.2Relay option
WARNING
At voltages >30 V AC or >42.4 V DC, the relay cables
must be enclosed in protective tubing, or doubleinsulated cables must be used.
If the relay option has been installed, the wires for the alarm
devices will be connected to the 9-pin connector. A piece of
wire insulation (heat shrink tube) and rubber boot are supplied
for extra protection for relay wires.
Turn bucket over and pull off the 9-pin connector.
Cut the heat shrink tube as needed and slide over the relay
wires before insertion into the 9-pin connector.
Position the heat shrink tube at the edge of wire insulation
and use a heat gun to shrink the tubing securely onto wire
insulation.
Slide rubber boot over the wires.
Dräger Polytron 81009
Page 10
Installation
i
i
3
4
5
2
00733298.eps
1
Connect the wires for alarm 1 (pre alarm), alarm 2 (main
alarm) and fault alarm to the terminals, as indicated in the
following wiring table. Fasten terminal screws with the
correct torque according to the table tightening torque and
wire size for field wiring terminals, see Section 10.9 on
Page 42.
With factory default settings and during normal operation,
the relays are energized. This provides “fail-safe”
operation. The terminal designators indicated in the
following wiring table are shown as factory default and
normal operation mode, see Section 7 on Page 34.
Plug connector back into socket and tighten screws.
Put rubber boot over connector.
NOTICE
To ensure that a fault is recognized - without having to
look at the instrument - an alarm device must be
connected to the fault relay.
Loosen set-screw (2), 2mm Allen screw.
Unscrew bayonet ring (3)
and remove blank.
Remove sensor from
packaging.
Insert sensor (4) into the
opening. The Dräger logo
on the sensor must point
to the mark on the sensing
head housing (5).
Lock sensor with bayonet
ring.
Tighten set-screw (2).
Mandatory for Zone 22
installations.
NO = Normally open, NC = Normally closed, C = Common
3.6Electrical installation with Docking
Station
Installing this configuration is a two-step process.
First, the Docking Station is mounted and connected to the
field wires. Second, the main instrument enclosure with the
electronics and sensor is attached to the Docking Station.
The Docking Station can be pre-mounted, wired and sealed
with the supplied cover. Once the site is ready for
commissioning, the instrument is then hooked up to the
Docking Station and taken into operation; avoiding that the
instrument is damaged during the construction phase.
The connection between the Docking Station and the main
instrument is realized via a ‘feed-through’. Depending on the
instrument selected, there are 2 types of feed-through.
3 wire for power (part number 4544182)
9 wire for power and relay (part number 4544169)
3.6.1Field wiring
For field wiring of Docking Station refer to Instructions for use
9033242.
10Dräger Polytron 8100
Page 11
Installation
!
i
i
i
i
3.8Installation of EC sensing head remote
3.8.1Wall or pipe mounting kit
CAUTION
Observe the general installation information, see
Section 3 on Page 8.
Multiple EC sensing heads remote must not be daisychained.
NOTICE
The EC sensor wall or pipe mount kit (part
number. 45 44 213) is required to mount the EC
sensor on a wall or pipe.
The sensing head should be mounted at a lowvibration location at even temperature near a potential
leakage point (avoid direct sunlight).
3.8.2Installing sensor
Loosen set-screw (2), 2 mm Allen screw.
Unscrew bayonet ring (3) and remove blank.
Remove sensor from packaging.
Insert sensor (4) into the opening. The Dräger logo on the
sensor must point to the mark on the sensing head housing
(5).
Lock sensor with bayonet ring.
Tighten set-screw (2). Mandatory for Zone 22 installations.
Install wall or pipe mount set (8) as specified in the
assembly instructions.
Screw the EC sensing head to the wall or pipe mount
bracket (8) using screw and washer (6).
NOTICE
Pay attention to the orientation of the cable
connector (9)!
Select the appropriate protection cap (7) and place it on the
screw.
Dräger Polytron 810011
Page 12
Installation
00333298.eps
2
2
3
5
11
1413
12
4
10
15
i
i
!
3.9Connecting the EC sensing head remote
to Polytron 8000
Connect remote cable plug (11), to the EC sensing head
remote (12) and lock.
Attach the shield wire (13) to the grounding point (14) of the
bracket (M5 thread).
Loosen set-screw (2), 2 mm Allen screw.
Unscrew bayonet ring (3) and remove blank. Either
bayonet ring can be used: black or silver.
Insert the remote adapter (4) in the opening. The Dräger
logo on the sensor adapter must point to the mark on the
sensing head housing (5).
Lock sensor adapter with bayonet ring.
Attach the shield wire (15) to the ground lug of the
enclosure.
Tighten set-screw (2). Mandatory for Zone 22 installations.
NOTICE
The EC sensing head remote is automatically
recognized by the instrument.
3.10 Connecting the instrument to a controller
from Dräger
NOTICE
i
i
For hook-up information, please refer to the
Instructions for Use which was included with the
Dräger controller (e.g. Regard, QuadGard).
Electrical connections at the controller
Connect the shield of the wires to the instrument earth ground
of the controller (e.g. chassis, ground busbar, etc.).
3.11 Connecting the instrument to a PC
The separately available Polytron 8000 IR interface
connection kit is intended for use with the Polytron 8000 and
enables communication between the Polytron 8000 and a PC
using the PolySoft 8000 software.
3.12 PolySoft 8000 PC software (optional)
The PolySoft 8000 PC software is used to display instrument
information, edit configuration settings and download the
instrument memory.
CAUTION
After modifying parameters with the PolySoft 8000
PC-software or another software, verify all parameters
by downloading or by checking them on the
Polytron 8000.
12Dräger Polytron 8100
Page 13
3.13 Installing software dongles
00833300.eps
The following software dongles are available for the
Polytron 8000:
Installation
Sensor test dongle
83 17 619
Diagnostic dongle
83 17 860
Activates the sensor self-test
(only for certain sensors)
Activates the sensor self-test, the
display of the sensor vitality and the
sensor diagnostic functions
(only for certain sensors and functions)
To install a software dongle:
Switch off power to the instrument or declassify the area
according to the local regulations.
Loosen set-screw and unscrew lid from instrument.
Lift the handle and pull-out the bucket with the main
electronics.
Insert the dongle in the slot with the Dräger logo facing
upwards.
Place bucket back into the enclosure.
Screw the lid back on, until it is seated (see Section 10.8
on Page 42), and tighten set-screw.
Switch power back on.
Dräger Polytron 810013
Page 14
Operation
!
i
i
4Operation
4.1Normal operation
WARNING
Before leaving the instrument for normal operation,
check the configuration and calibration for the proper
settings.
Switch power supply on.
The instrument will go through a start-up sequence (LCD / LED
test, software version, and initialization) and start the warm-up
period. The display shows that the sensor will be ready for
measurement in hh:mm:ss (countdown) and the instrument
transmits the maintenance signal, see Section 10.3 on
Page 41.
After the warm-up period, the instrument goes into normal
operation. The display shows the current gas concentration,
the selected gas and the units of measurement. The green
LED is lit.
NOTICE
The correct date and time settings are important for
many functions see Section 5.7.3 on Page 20.
4.1.1Analog signals
The current output of the instrument during normal operation is
between 4 and 20 mA and is proportional to the detected gas
concentration.
Polytron 8000 uses different current values to indicate various
modes of operation. The factory default settings are user
adjustable for application specific requirements, see Section 7
on Page 34. This follows the NAMUR recommendation NE43.
4.1.2The display and LEDs
In normal operation, the display shows the measured gas
concentration, the selected gas and the unit of measurement.
The green LED is lit.
The following symbols may also be displayed:
" ", when the measuring range is exceeded
" - - - - " and " X " when a fault has been detected. The
yellow LED is lit, see Section 6.3 on Page 31.
"SIL", when SIL status is activated.
If the optional relay board is installed:
When the first alarm (pre alarm) has been triggered the red
LED blinks in single mode. The A1 relay is asserted.
When the second alarm (main alarm) has been triggered
the red LED blinks in double mode. The A2 relay is
asserted.
If the A2 is configured as acknowledgeable, and the A1 is
configured as non-acknowledgeable, then, when an A2
alarm occurs, the red LED will blink in double blink A2
mode. After the alarm is acknowledged, the red LED will
blink in single blink mode to indicate A1 is still active.
4.1.3The LED symbols
NOTICE
i
i
The alarm triggering function is only available when
the optionally integrated relay module is used.
SymbolLEDDescription
redAlarm triggered
yellowFault
green
4.1.4Status indicators in the display
The following status indicators may appear on the right side of
the display.
Warning message available – to display warnings, see
Section 5.2.1 on Page 18.
Error message available – to display errors, see
Section 5.2.2 on Page 18.
Maintenance signal is transmitted, see Section 5.8.2 on
Page 22.
Measuring range of analog interface exceeded
Measurement less than range of analog interface
Analog interface is set to a fixed value (e.g. Multidrop HART
communication) and is not transmitting any measurement
signal.
"Preventive" maintenance: The sensor is ready for operation.
"Preventive" maintenance: The sensor is ready for operation
but is close to the end of its life cycle.
"Preventive" maintenance: The sensor is still ready for
operation but should be changed as soon as possible.
The data-logger is active in Roll mode.
To activate/deactivate, see Section 5.10.2 on Page 27.
The data-logger is active in Stack mode.
To activate/deactivate, see Section 5.10.2 on Page 27.
SIL
SIL is activated (refer to the Polytron 8000 Safety Manual
part number 9033307).
Power ON
Normal operation
4.2Activating the Info mode
No password required.
The Info mode is used to show instrument relevant information.
This will not interrupt the normal operation of the instrument.
Tap and hold [UP] for more than 3 seconds.
The instrument information is shown on several screens.
No changes can be made.
Tap [UP] / [DOWN] to switch between the screens.
The Info mode can be exited at any time by tapping [OK].
If no key is tapped within 30 seconds, the instrument will
automatically return to normal operation.
14Dräger Polytron 8100
Page 15
Operation
A001.eps
07.11.2003 12:34
SW Version
: 1
Part No.
: 8317778
Serial No.
: ARUA0001
DeviceCode: 00006317
Instrument Info
1/3
A002.eps
2/3
Sensorname: O
2
Part No. : 6809630
Serial No.
: XXXXXXX
EEPROM Typ
: 1
EEPROM Vers.: 1
Sensor Info
A003.eps
3/3
Gasname : O
2
Range : 25.00 Vo
4–20 SP
: 25.00 Vo
Alarm A1
: 19.00 Vo
Alarm A2
: 23.00 Vo
Sensor Config.
00733300.eps
Example of the Info mode
Depending on sensor the screens may vary:
1. Screen - Instrument information:
Date and Time
Software version
Part number of the instrument electronics
Serial number of the instrument electronics
Instrument code
2. Screen - Sensor information:
Installed sensor
Part number of the sensor
Serial number of the sensor
Software version
Software revision
4.4Menu navigation
Tapping the magnetic wand (part number 4544101: blue body,
white logo) over [UP] / [DOWN] scrolls through the menu
selections.
Tapping [OK] confirms a selection.
NOTICE
i
i
The instrument is designed for the magnetic wand to
be used with the enclosure lid in place. If the enclosure
lid is not in place, the magnetic wand may activate two
or more buttons at once (cross-talk).
3. Screen - Sensor configuration:
Selected gas
Maximum measuring range and units for the sensor
Measuring range and units for the analog interface. Only
shown if the 4 to 20 mA interface is enabled.
A1 Alarm set point and units
A2 Alarm set point and units
1
1
4.3Activating the Quick Menu
No password required.
Tap and hold [DOWN] for 1 to 2 seconds.
The selected information about status or instrument
settings can be retrieved (e.g. warning messages, error
messages, installed modules and calibration data, time/
concentration graph, see Section 5.7.5 on Page 21.
No changes can be made.
4.4.1Password
A password consists of a 4-digit alphanumeric value.
From normal operation, tap and hold [DOWN] for more
than 3 seconds.
The message “Enter the password” will be displayed.
Confirm with [OK].
The screen to enter the password will be displayed with the
first digit on the left blinking.
Tap [UP] / [DOWN] to increment or decrement this digit,
then tap [OK].
The second digit will blink; set the correct value by tapping
[UP] / [DOWN], then tap [OK].
Repeat the process for the other two digits.
Tap [OK] when the full password is displayed.
If the displayed value matches the set password, access will
be given to the menu. If an incorrect password is entered, the
instrument will indicate “wrong password” and show the
dashes and blinking digit again, see Section 5.7.2 on Page 20.
Graphic symbols facilitate navigation in the various menus.
Combined with the text »Back«, »Menu«, etc., exit
the menu or go back one step.
Closed folder
More functions or sub-menus are available.
Opened folder
The available functions and sub-menus are shown.
Function
When activated, functions can be executed in one or
more steps.
Selection activated
For functions that can be selected and activated,
the activation is executed by tapping [OK].
List above closed / complete
There are no other functions, menus or sub-menus listed
above.
List can be scrolled upwards
There are more functions, menus or sub-menus
listed above.
List below closed / complete
There are no other functions, menus or sub-menus listed
below.
List can be scrolled downwards
There are more functions, menus or sub-menus
listed below.
Next
Perform context-related action.
Number / total number of steps (screens) within
the function.
Entry
Enter data with [UP] / [DOWN].
5Menu
The menu is grouped into three main segments.
Information
Specific data about the instrument, the sensor and the
measured gas are summarized in this menu. The menu can be
accessed without a password.
No changes can be made.
Calibration
Calibration of the instrument must be performed at regular
intervals as detailed in the sensor data sheet.
Settings
This sub-menu contains all functions to configure the
instrument to the application specific requirements.
4.4.3Changing parameter values or statuses
Select the menu item to be accessed by tapping [UP] /
[DOWN].
When the intended menu item is displayed, tap [OK].
The current value or status will be displayed.
[UP] / [DOWN] allow adjusting the value of a numerical
parameter or scrolling through preset choices.
Once the display shows the intended value or choice, tap
[OK] to validate the new parameter and return to the main
menu.
When exiting via “Back to menu” or “Previous”, all
changes will be discarded.
4.4.4Exiting the menu
Tap [UP] to “Back to measurement” and tap [OK].
16Dräger Polytron 8100
Page 17
5.1Menu overview
Back to
◄
measuring
Menu
Informationpage 18Instrumentpage 18
Sensorpage 18
Calibrationpage 18Dataloggerpage 18
Settingspage 19
Zero calibrationpage 29
Span calibrationpage 29
Auto CALpage 30
This function displays warning messages in plain text and the
corresponding number see Section 6.3.2 on Page 32. The [ ! ]
symbol is displayed if there are warning messages available.
Select Information > Instrument > Warnings and
confirm.
If there are several warning messages available, there is
an indicator in the upper right-hand corner of the display,
e.g. 1/3 = Screen 1 of 3.
5.2.2Fault messages
This function displays fault messages in plain text and the
corresponding number see Section 6.3.1 on Page 31. The [ X ]
symbol is displayed if there are fault messages available.
Select Information > Instrument > Faults and confirm.
If there are several fault messages available, there is an
indicator in the upper right-hand corner of the display e.g.
1/2 = Screen 1 of 2.
5.2.3Device codes
This function displays warning and fault codes as a table
(hexadecimal).
Select Information > Instrument > Device codes and
confirm.
If all codes are 00 there are no warning or fault messages
available. The instrument is working to specification.
5.4Information data-logger
These functions contain information about the data-logger.
5.4.1Display the data-logger status
Select Information > Data logger > Logger status and
confirm.
To change the status see Section 5.10 on Page 27.
5.4.2Display graph
Display the history of the past 15-minutes on a time/
concentration graph.
Select Information > Data logger > Graph and confirm.
5.5Calibration
5.5.1Zero calibration
Allows adjusting the zero reference point of the sensor, see
Section 6.2.1 on Page 29.
5.5.2Span calibration
Allows adjusting the sensitivity to match the known
concentration of an applied calibration gas, see Section 6.2.2
on Page 29.
5.5.3Auto calibration
Automated sequence reducing the number of steps during
calibration; see Section 6.2.3 on Page 30.
NOTICE
The device code is essential for technical support. It is
important to identify the exact location and values of
the codes.
5.2.4Installed modules
This function displays an overview of the installed hardware
modules.
Select Information > Instrument > Modules and confirm.
The list contains modules. The installed modules are
identified by the symbol. The symbol means that the
module is not installed.
Highlight a module and access detailed information with
[OK].
5.3Information sensor
This function contains information on calibration dates.
5.3.1Display last calibration date
Select Information > Sensor > Last cal. date and
confirm.
5.3.2Display next calibration due date
Select Information > Sensor > Next cal. date and
confirm.
5.6Settings SIL activation
NOTICE
i
i
SIL lock is only available for certain sensors (refer to
the Polytron 8000 Safety Manual part number
9033307).
SIL lock can only be activated if the sensor test is
activated see Section 5.9.9 on Page 26.
This function protects the instrument against unauthorized
changes. Changing the configuration (e.g., changing the
measuring range from 100 %LEL to 50 %LEL) will force the
user to confirm all safety-relevant parameters before the
instrument returns to normal operation.
Select Settings > Instrument > SIL activation.
Select Enable or Disable and confirm.
NOTICE
i
i
When SIL is activated, the instrument displays all
safety-relevant parameters and settings before
returning to normal operation. Carefully review all
parameters and settings and confirm.
18Dräger Polytron 8100
Page 19
Menu
!
!
i
i
5.7Settings instrument
5.7.1Alarm settings
CAUTION
This function is only available if the relay module is
installed.
With factory default settings and during normal operation, the
relays are energized. This provides a “fail-safe” operation.
Switching the alarm On or Off
Select Settings > Instrument > Alarm > Alarm on/off.
Select Enable or Disable and confirm.
CAUTION
If the alarm is set to Off, an alarm condition will not be
shown by the LEDs or assert the relay!
NOTICE
The maintenance signal is transmitted and the symbol
[ ] is displayed.
Configuring Relay A1 or A2
This function defines whether the alarm relay is energized
during normal operation or energized at an alarm condition.
If the configuration of the relay is set to “Normally energized”,
the relay contact is asserted during normal operation and
changes if an alarm is triggered. At a loss of power this will
trigger an alarm (fail-safe). The fault relay is factory-configured
to “Normally energized” and cannot be changed.
Select Normally energ. or Energ. on alarm and confirm.
Configuring Alarm A1 or A2
Latching means that once the alarm level is reached, the
instrument will trigger the alarm. It will remain in alarm status
even if the gas concentration subsequently does not meet the
alarm condition any more. To clear a latching alarm it has to be
acknowledged with [OK].
In non-latching mode, the alarm status clears if the gas
concentration does not meet the alarm condition anymore.
Acknowledgeable means that the alarm relay can be reset
before alarm condition clears.
Non-acknowledgeable means that the alarm relay can not be
reset until the alarm condition clears.
NOTICE
i
i
The hysteresis function defines an interval where a
triggered relay maintains its status until the gas
concentration is outside the defined interval. This
prevents relays from chattering at an alarm set-point.
Example: A2 at 40 % LEL Methane
Hysteresis: 3 % LEL
Alarm asserts at 40 % LEL
Alarm de-asserts at
<37%LEL(=40%LEL-3%LEL)
Set the alarm set-point and confirm.
Select Next and confirm.
The current alarm direction is displayed.
Select Rising or Falling and confirm.
Select Next and confirm.
The current alarm latch mode is displayed.
Select Latching or Non Latching and confirm.
Select Next and confirm.
The current acknowledgement mode is displayed.
Select Acknowledgeable or NonAcknowledgeabl or
PreAcknowledgeabl and confirm.
Select Next and confirm.
The current hysteresis is displayed.
Set the hysteresis and confirm.
Select Next and confirm.
A confirmation screen shows all settings.
Select Confirm and confirm with [OK].
The new settings are now saved.
Repeat settings for A1 and A2.
Dräger Polytron 810019
Page 20
Menu
i
i
i
i
Explanation of combining latching, acknowledgment and
pre-acknowledgment of alarms:
Since the concepts of latching and acknowledgment can be
confusing, the following six combinations are offered for
clarification:
Latching and
acknowledgeable
Latching and nonacknowledgeable
Latching and preacknowledgeable
Non-latching and
acknowledgeable
Non-latching and
nonacknowledgeable
Non-latching and
preacknowledgeable
NOTICE
For safe operation:
The relays must be set to “Normally energized” and
the A2 alarm (main alarm) must be set to “latching”
and “not acknowledgeable” or “preacknowledgeable”.
The A1 alarm (pre alarm) may only be set to
“acknowledgeable” if it is used for triggering an
acoustic alarm device.
Relay Test
These functions change the status of a relay and LED for test
purposes (e.g. to check the function of alarm devices connected
to the relay) and the symbol [ ] is displayed. After exiting this
function, the status of the relay and LED will automatically return
to their previous status.
NOTICE
When the relays are activated, alarm devices will be
switched on.
Relay must be reset manually and can
be reset before the alarm condition
clears.
Relay must be reset manually. Relay
cannot be reset before the alarm
condition clears.
Relay must be reset manually. Relay
can be reset while the alarm condition
is still present but relay will not reset
until the alarm condition clears.
Relay will reset automatically when
the alarm condition clears or can be
reset manually.
Relay will reset automatically when
the alarm condition clears. Relay
cannot be reset manually before the
alarm condition clears.
Relay will reset automatically when
the alarm condition clears. Relay can
be reset while the alarm condition is
still present but relay will not rest until
the alarm condition clears.
Select Settings > Instrument > Alarm > Set fault and
confirm.
Select Enable or Disable and confirm.
NOTICE
i
i
If the function is activated, the fault relay is asserted and
the 4 to 20 mA interface is set to the fault current. The
yellow LED is lit, the symbol for maintenance is
shown [ ].
5.7.2Passwords
These functions change the passwords for calibration (zero
and span calibration) and settings (access to all configuration
parameters).
Select the line for editing the password and confirm.
Set the password and confirm.
The setting for the password is displayed.
Select Confirm and confirm with [OK].
5.7.3Date and time
NOTICE
i
i
The correct date and time settings are important for
many functions.
Select Settings > Instrument > Date and time and
confirm.
Select the line for editing the date or time and confirm.
Set each digit and confirm.
Select Confirm and confirm with [OK].
5.7.4Language
Select Settings > Instrument > Language and confirm.
Select the language from the list and confirm.
20Dräger Polytron 8100
Page 21
Menu
5.7.5Function key
This function activates the Quick Menu.
The selected information about status or instrument settings
can be retrieved (e.g. warning messages, fault messages,
installed modules, calibration data and time/concentration
graph.)
Select Settings > Instrument > Function key and
confirm.
Select a function and confirm.
Repeating the confirmation will deactivate the selected
function.
In normal operation, the selected function can be activated by
briefly tapping [UP] for 1 to 2 seconds.
If no function key is selected, the Quick Menu is deactivated.
Selection options:
Graph (only with
data-logger
dongle)
The measurements of the past 15
minutes are displayed as a time/
concentration graph.
FaultsFault messages are displayed in plain
text see Section 5.2.1 on Page 18.
WarningsWarning messages are displayed in
plain textsee Section 5.2.2 on
Page 18.
Bump-testThe bump-test allows applying gas to
the sensor without generating an
alarm. The maintenance signal is
transmitted.
Sensor vitality
(only with
Shows the remaining sensor vitality
(only available for certain sensors).
diagnostic dongle)
If Bump-test is selected:
The maintenance symbol [ ] is shown on the display and the
maintenance signal is transmitted.
After 2 minutes or tapping [OK] again will end the bump test
and the instrument returns to normal operation.
5.7.7Software dongle
This function deactivates a SW dongle to safely remove the
dongle or if the dongle is faulty.
A dongle can only be reactivated by cycling the power to the
instrument.
If "non display" has been selected, the display shows the start
screen and the applicable symbols.
If the relay option is installed and an alarm is triggered:
regardless of the selected display mode, the display will show
the current gas concentration and the red LED will blink.
When activated, the display is shown inverted and all the LEDs
are lit.
5.7.6Device initialization
This function resets the instrument to factory default settings
(see Section 7 on Page 34).
Alarm settings
Passwords
Language
Function key
HART interface
Data-logger
Analog interface
Relay configuration
Select Settings > Instrument > Device init. and confirm.
Select Confirm and confirm with [OK].
Dräger Polytron 810021
Page 22
Menu
5.8Settings communication
5.8.1HART interface
These functions change the settings of the HART interface.
Polling address
The polling address configures the instrument for analog
operation (4 to 20 mA) or multidrop operation. Setting the
polling address to 0 activates analog operation (4 to 20 mA).
Setting the polling address to a value between 1 and 15
activates multidrop operation. In multidrop operation the
analog interface is deactivated and set to a constant current of
approximately 1 mA.
To allow the central controller to request the Unique Identifier
(unique HART address) with HART command #0, all the
devices on a line must be configured with different polling
addresses. It is advised to choose a sequence starting with 1
and continuous increments.
This function corresponds to HART command #6 (Write Polling
Address).
Select Settings > Communication > HART interface >
Polling address and confirm.
Select the line for editing the address and confirm.
Set the polling address and confirm.
Select Confirm and confirm with [OK].
Unique Identifier
This function reads the Unique Identifier (unique HART
address). It has to be known for almost all HART addressing
commands. However, this information is only required for
systems which are not able to read back the Unique Identifier
using the HART command #0 in Short-Frame Format or HART
command #11. The display corresponds to the address for
HART command #0 (Read Unique Identifier) or #11 (Read
Unique Identifier associated with Tag).
Select Settings > Communication > HART interface >
Unique Identifier and confirm.
The Unique Identifier is displayed.
Tag
The tag can be used to identify an instrument. It can contain up
to 8 alphanumeric characters. The tag can also be used as an
address for reading the unique identifier using HART
command #11 (Read Unique Identifier Associated with Tag),
even if the polling address is unknown. This requires that a
unique tag has been configured for each instrument.
Select Settings > Communication > HART interface >
Tag and confirm.
Select the line for editing the tag and confirm.
Set the tag and confirm.
Select Confirm and confirm with [OK].
Dräger REGARD
This function switches between different HART protocols.
NOTICE
i
i
REGARD SW 2.0 is the default setting. REGARD SW
1.4 must be selected for REGARD controllers with
SW 1.4.
Select Settings > Communication > HART interface >
Dräger REGARD and confirm.
Select REGARD SW 2.0 or REGARD SW 1.4 and confirm.
5.8.2Analog interface
These functions set currents for special signals on the analog
interface and send test currents for validation at the central
controller.
Fault current
This function defines the current for the fault signal.
Select Settings > Communication > Analog interface >
Fault current and confirm.
Select the line for editing the current and confirm.
Set the current and confirm.
The setting for the fault current is displayed.
Select Confirm and confirm with [OK].
Warning signal
WARNING
!
This function switches the warning signal On or Off.
If turned On, the warning signal will be transmitted on the
analog interface if a warning message is available. The default
setting for the warning signal is “warning current” for one
second and the “measurement current” for 9 seconds. These
time intervals and the warning current can be changed.
To switch the warning signal on or off:
Select Settings > Communication > Analog interface >
Warning and confirm.
Select Enable or Disable and confirm.
22Dräger Polytron 8100
Page 23
Menu
0533300.eps
t
[s]
T
1
Messwert
4
20
[mA]
T
2
3
Value
i
i
0633300.eps
t
[s]
1,1
[mA]
0,7
5
3
Warning interval
This function defines the time interval for the warning signal.
Select Settings > Communication > Analog interface >
Warning interval and confirm.
Select the line for editing T1 and confirm.
Set the time and confirm.
The setting for Time T1 is displayed.
Select Next and confirm.
Select the line for editing T2 and confirm.
Set the time and confirm.
The setting for Time T2 is displayed.
Select Confirm and confirm with [OK].
Warning current
This function defines the current for the warning signal.
Select Settings > Communication > Analog interface >
Warning current and confirm.
Select the line for editing the current and confirm.
Set the current and confirm.
The setting for the warning current is displayed.
Select Confirm and confirm with [OK].
Maintenance signal
This function selects the type of maintenance signal.
Select Settings > Communication > Analog interface >
Maint. signal and confirm.
Select static or dynamic signal type and confirm.
NOTICE
The static signal type is a constant current. However,
the current value can be configured.
Maintenance current
This function defines the current for the maintenance signal.
Select Settings > Communication > Analog interface >
Maint. current and confirm.
Select the line for editing the current and confirm.
Set the current and confirm.
The setting for the maintenance current is displayed.
Select Confirm and confirm with [OK].
NOTICE
i
i
The maintenance current can only be set if the
maintenance signal has been set to static.
Beam block on or off
This function is only available for Polytron 87X0 (using the
sensor PIR 7X00).
Analog offset
This function adds an offset to the analog output. The offset is
constant over the entire range of the analog signal.
This function is essential for an installation where the analog
current at the instrument differs from the current at the central
controller.
Select Settings > Communication > Analog interface >
Analog offset and confirm.
Select the line for editing the offset (range: -0.2 to 1.5 mA)
and confirm.
Set the current and confirm.
The setting for the analog offset is displayed.
Select Confirm and confirm with [OK].
Test functions for the analog interface
NOTICE
i
i
These functions change the current of the analog
interface for test purposes (e.g. to check the
programming of the central controller). It might be
necessary to inhibit the alarms at the central controller
to avoid false alarms. After exiting these functions, the
currents, if changed, will automatically return to the
maintenance signal.
During the test, the maintenance symbol [ ] is
displayed.
The dynamic signal type is a square wave signal with
the following characteristics.
Dräger Polytron 810023
Page 24
Menu
Set current
This function sets the current to any value between 0 and
22 mA.
Select Settings > Communication > Analog interface >
Set current and confirm.
Once the alarms at the central controller are inhibited,
confirm the message Inhibit all alarms.
Select the line for editing the current and confirm.
Set the current and confirm.
The setting for the current is displayed.
Select Next and confirm.
Select Set current out or Current off and confirm.
Select Next and confirm.
The function is terminated.
Once the alarms at the central controller are enabled
again, confirm the message Enable all alarms.
Set concentration
This function sets the current corresponding to any
concentration value between 0 and full-scale.
Select Settings > Communication > Analog interface >
Set concentr. and confirm.
Once the alarms at the central controller are inhibited,
confirm the message Inhibit all alarms.
Select the line for editing the concentration and confirm.
Set the concentration and confirm.
The setting for the concentration is displayed.
Select Next and confirm.
Select Set concentr. out or Concentration off and
confirm.
Select Next and confirm.
The function is terminated.
Once the alarms at the central controller are enabled
again, confirm the message Enable all alarms.
Set fault
This function sets the current to the fault current.
Select Settings > Communication > Analog interface >
Set fault and confirm.
Select Enable or Disable and confirm.
Set warning
This function sets the current to the warning current.
Select Settings > Communication > Analog interface >
Set warning and confirm.
Select Enable or Disable and confirm.
Set maintenance
This function sets the current to the maintenance current.
Select Settings > Communication > Analog interface >
Set mainten. and confirm.
Select Enable or Disable and confirm.
Set beam block signal
This function is only available for Polytron 87X0 (using the
sensor PIR 7X00).
5.8.3Profibus address
Not yet active - for future functions.
5.9Settings sensor
5.9.1Sensor change
With this function, a sensor can be changed during operation
without triggering a fault signal at the central controller. It also
ensures that all sensor data in the microprocessor can be
saved in the sensor memory (EEPROM) before disconnecting
the sensor.
A sensor can be changed at any time. However, for safety
reasons a fault alarm is generated when the sensor is
disconnected until a new sensor is connected.
1. Select Settings > Sensor > Sensor change and confirm.
The maintenance signal is generated on the analog
interface, the following display shows: “Please remove the
sensor”.
2. Replace the old sensor with a new sensor, see Section 6.4
on Page 33.
When the new sensor is installed:
Display: “Loading data, please wait”.
When the sensor data are loaded:
Display: “Data loaded”.
3. Select Back to menu and confirm.
The maintenance signal on the analog interface remains
pending until the sensor is warmed-up.
The warm-up time depends on the type of sensor and its
history. For example, if the sensor has already warmed-up on
another instrument and the operation was only interrupted for
a short time, the warm-up time can be shorter than the time
specified in the Instructions for Use for the sensor.
Maximum warm-up time of a new sensor: see the Instructions
for Use for the sensor.
NOTICE
i
i
If a sensor of the same type (same part number) was
previously installed, the instrument specific
configuration is retained (gas type, measurement
range, test gas, calibration interval, etc.)
Otherwise the factory default settings of the new
sensor (see Instructions for Use for the sensor) are
uploaded and will overwrite the instrument specific
configuration.
This can be prohibited if the sensor lock function (see
Section 5.9.4 on Page 25) is activated.
24Dräger Polytron 8100
Page 25
Menu
i
i
!
5.9.2Auto CAL
NOTICE
This function is only active if enabled here.
Select Settings > Sensor > Set auto calibration and
confirm.
Select Enable or Disable and confirm.
5.9.3Sensor test
This function is only active for Polytron 8000 using the EC
sensor.
This function is only active if the sensor test or diagnostics
dongle is installed.
This function activates the sensor test. The instrument
routinely checks the sensor for proper function. If the sensor
does not pass the self-test, a warning or fault message will be
issued.
Select Settings > Sensor > Sensor test and confirm.
Select Enable or Disable and confirm.
5.9.4Sensor lock
This function is only active for Polytron 8000 using the EC
sensor and Polytron 87X0 using the PIR 7X00 sensor.
This function activates the sensor lock feature.
Select Settings > Sensor > Sensor lock and confirm.
Select On or Off and confirm
CAUTION
If
the instrument will reject any sensor whose part
turned
turned
number does not match the part number of the
On
previously installed sensor.
If
the instrument will accept any suitable sensor,
uploading the default settings of this sensor.
Off
However, this will overwrite any customized
settings with the default settings of the new
sensor.
5.9.5Sensor type
This function is only active for Polytron 8200 or Polytron 8310
using the DD, LC or DSIR sensor.
5.9.6Gas settings
This function sets the gas type, measuring range and units of
measurement.
Not all sensors offer a selection of values.
Select Settings > Sensor > Gas setting and confirm.
The current measured gas is displayed.
Select a measured gas from the list and confirm.
The current unit of measurement is displayed.
Select the unit of measurement from the list and confirm.
The current full scale deflection is displayed.
Set the full scale deflection and confirm (only for certain
sensors, see sensor data sheet).
The new full scale deflection is displayed.
Select Next and confirm.
An overview of the new gas settings is displayed.
After review: Select Previous for changes and confirm with
[OK].
Select Confirm to accept the settings and confirm with
[OK].
Information about units of measurement
Depending on the target gas, the units of measurement can be
switched between:
The instrument automatically converts the sensor signal into
the selected unit of measurement. Units which are not
available for a particular sensor or selected substance will not
be shown.
Information about the full scale deflection (measuring
range) and the analog output of 20 mA.
Some sensors offer a min/max range for the full scale
deflection.
The selected full scale deflection determines the output of the
20 mA signal on the analog interface.
Example:
Required range 0 to 500 ppm CO (e.g. part number 6809605
default 300 ppm, range min/max = 50/1000 ppm).
Select full scale deflection as 500 ppm. The analog output will
be linear between 4 mA = 0 ppm and 20 mA = 500 ppm.
5.9.7Reset sensor
This function resets all sensor parameters to the factory default
settings (see Section 7 on Page 34).
Select Settings > Sensor > Reset sensor required! and
confirm.
Select Confirm and confirm with [OK].
Dräger Polytron 810025
Page 26
Menu
i
i
5.9.8Calibration interval
NOTICE
The calibration interval can be set between 0 and
720 days.
After the calibration interval expires, the instrument will issue a
warning that the calibration is past due.
Select Settings > Sensor > Calibration interval
and confirm.
Set the calibration interval and confirm.
5.9.9Set sensor test
This function is only active for Polytron 8000 using the EC
sensor.
This function is only active it the sensor test or diagnostics
dongle is installed.
This function periodically initiates the sensor self-test.
Select Settings > Sensor > Set sensor-test and confirm.
Select Enable or Disable and confirm.
5.9.10 Fast Response
This function is only activated for Polytron 87X0 using the
PIR 7X00 sensor.
26Dräger Polytron 8100
Page 27
Menu
5.10 Data-logger
The data-logger can store up to 3000 values. At a sampling
interval of one measurement per minute the data-logger stores
the measurement history of approximately 50 hours. This time
can be significantly increased if the Trigger function is enabled.
The data-logger can only be downloaded using the PolySoft
8000 PC software. The last 15 minutes as a time/concentration
graph can be viewed entering the menu Information >Datalogger > Graph (see Section 5.7.5 on Page 21) or via the
function key.
5.10.1 Data-logger on or off
Select Settings > Datalogger > Datalogr. on/off and
confirm.
Select Enable or Disable and confirm.
5.10.2 Set data-logger
Sampling time
This function defines how often a value will be stored.
This function determines which value will be stored.
Select Settings > Datalogger > Set datalogger > Peak/
Average and confirm.
Select the Peak or Average and confirm.
PeakThe maximum value (if monitoring falling
concentrations, the minimum value) of the
measured concentrations within the selected
sampling time is stored.
Average The average value of all the measured
concentrations within the selected sampling
time is stored.
Trigger value
This function defines the trigger threshold to store a value.
The trigger value is defined as a percentage of the full scale
deflection.
Example: A trigger value of 2 % at full scale deflection of
500 ppm will only store values if they deviate by 10 ppm
(relative to the previously stored value).
This function defines whether values are overwritten or top out.
Select Settings > Datalogger > Set Datalogger > Stack/
Roll and confirm.
Select Stack or Roll and confirm.
RollOnce the capacity of the data-logger is
reached, the old data will be overwritten by
new values.
StackOnce the capacity of the data-logger is
reached, no more data can be stored. The
instrument will issue a warning.
5.10.3 Clear data-logger
This function deletes the stored data.
Select Settings > Datalogger > Clear Datalogr.
and confirm.
To clear the data-logger, select Confirm and confirm with
[OK].
Trigger on or off
Select Settings > Datalogger > Set datalogger > Trigger
on/off and confirm.
Select On or Off and confirm.
If
turned
On
turned
Measurements will be stored if they exceed a
threshold beyond the trigger value (relative
to the last stored value).
If
Measurements within the sampling time will
be stored.
Off
Dräger Polytron 810027
Page 28
Maintenance
!
i
i
00533298.eps
1
5
4
3
2
6Maintenance
6.1Maintenance intervals
6.1.1During commissioning
Check calibration.
Check signal transmission to the central controller, LEDs
and triggering of alarm devices.
6.1.2Every 6 months
Inspection by trained service personnel.
Check signal transmission to the central controller, LEDs
and triggering of alarm devices.
The maintenance intervals must be established for each
individual installation. Depending on safety considerations
and application specific conditions the instrument is used
in, these might need to be shortened.
Contact Dräger Service to contract regular calibration and
maintenance services.
6.1.3Change sensor on demand
To change the sensor, see Section 6.4 on Page 33.
6.2Calibration
Calibration of the instrument must be performed at regular
intervals as detailed in the sensor data sheet.
General procedure
Attach the pressure regulator to the calibration gas
cylinder.
Fit calibration adapter to the sensor.
The gas flow should be approximately 0.5 L/min.
If the intended operation is at high altitudes, the reading will
be lower than the reading at sea level (reduced partial
pressure). A new span calibration is recommended if the
altitude or the ambient pressure is changed. The factory
calibration is set to sea level.
Connect the tubing to the barbed fitting.
WARNING
Calibration gas must not be inhaled! See appropriate
Material Safety Data Sheets. Calibration gas should be
vented into a fume hood or to the outside of the
building.
For proper operation, never adjust the span before
completing zero adjustment. Performing these
operations out of order will cause the calibration to be
faulty.
NOTICE
Calibration is not possible if the date and time are not
set (see Section 5.7.3 on Page 20).
Dräger recommends calibrating instruments with target gas.
This method of target gas calibration is more accurate than a
surrogate gas calibration. A surrogate gas calibration may only
be performed as an alternative if a target gas calibration is not
possible.
The sensor should be fully warmed-up (see sensor data
sheet).
1 Pressure regulator
2 EC sensing head
3 Calibration adapter
4Tubing
5 Calibration gas cylinder
28Dräger Polytron 8100
Page 29
Maintenance
!
6.2.1Zero Calibration
WARNING
Ambient air can be used to zero the sensor instead of
Nitrogen or Synthetic Air only if the area is known to be
free of the target gas or any gas to which the sensor
may be cross-sensitive (as listed on the sensor data
sheet). In this case, no cylinder or calibration adapter
is needed for the zero calibration.
For Oxygen (O2) sensors the zero cannot be calibrated. For
test purposes only Nitrogen (N2) must be used.
Select Calibration > Zero Calibration and confirm.
The Maintenance signal is transmitted and the symbol [ ]
is displayed.
The message Apply zero gas is displayed.
Apply Synthetic Air or Nitrogen.
Select Next and confirm.
The current value is displayed.
After the displayed value is stable:
Select Calibrate and confirm.
The message Please wait... is displayed.
The new current value is displayed.
Select Next and confirm.
Turn off gas flow and remove the calibration adapter from
the sensor or disconnect tubing.
If the current value is not within the alarm range:
Select Next and confirm.
The instrument returns to the calibration menu.
6.2.2Span calibration
Select Calibration > Span calibration and confirm.
The Maintenance signal is transmitted and the symbol [ ]
is displayed.
The parameters for the calibration gas are displayed, e.g.:
GasH2S
Unitppm
Conc.25
Depending on the sensor these parameters can be changed:
Select Gas and confirm.
Select the calibration gas from the list and confirm.
Select Unit and confirm.
Select the unit of measurement from the list and confirm.
Select Conc. and confirm.
Set the concentration of the calibration gas.
If the settings are correct:
Select Next and confirm.
A message like Gas flow ON H2S is displayed.
Apply calibration gas.
Select Next and confirm.
The current value is displayed.
After the displayed value is stable:
Select Next and confirm.
The message Please wait... is displayed.
The new current value is displayed.
Select Next and confirm.
Turn off gas flow and remove the calibration adapter from
the sensor or disconnect tubing.
If the current value is not within the alarm range:
Select Next and confirm.
The instrument returns to the calibration menu.
A calibration can be aborted at any time.
Dräger Polytron 810029
Page 30
Maintenance
i
i
!
6.2.3Auto Calibration
The instrument offers the option of an automatically
sequenced calibration. This reduces the number of manual
interactions with the instrument. The auto calibration is only
suitable for experienced users, because calibration errors can
happen due to procedural mistakes.
NOTICE
i
i
After 10 minutes without confirmation, the instrument
reverts to the calibration menu without performing
a calibration.
NOTICE
Not all of the available sensors and gases support auto
calibration. If the function is not available, a manual
calibration must be performed (see 6.1.1 on Page 28
and 6.1.2 on Page 28).
WARNING
Ambient air can be used to zero the sensor instead of
Nitrogen or Synthetic Air if the area is known to be free
of the target gas or any gas to which the sensor may
be cross-sensitive (as listed on the sensor data sheet).
In this case, no cylinder or calibration adapter is
needed for the zero calibration.
Dräger recommends performing a second,
independent measurement. Some sensors may only
be calibrated with Synthetic Air or Nitrogen (see
sensor data sheet)
Select Settings > Sensor > Autocal. set. and confirm.
Select Enable or Disable and confirm.
Apply Synthetic Air or Nitrogen.
Select Calibration > Auto cal. and confirm.
The Maintenance signal is transmitted and the symbol [ ]
is displayed.
The message Please wait... is displayed and the
instrument automatically performs the zero calibration.
After a successful zero calibration, the span calibration is
started.
Once the current value is stable, the instrument
automatically performs the span calibration.
After a successful calibration, the new current value and
the message Value OK? are displayed.
If not, select Redo and confirm to perform the calibration
again.
If the value is OK, turn off gas flow and remove the
calibration adapter from the sensor or disconnect tubing.
If the current value is not within the alarm range:
Select Accept value and confirm to finish the calibration.
The instrument returns to the calibration menu.
The parameters for the calibration gas are displayed e.g.:
GasH2S
Unitppm
Conc.25
Depending on the sensor these parameters can be changed:
Select Gas and confirm.
Select the calibration gas from the list and confirm.
Select Unit and confirm.
Select the unit of measurement from the list and confirm.
Select Conc. and confirm.
Set the concentration of the calibration gas.
If the settings are correct:
Apply calibration gas.
Select Next and confirm to start the calibration.
Select Back to menu to abort the calibration.
Serious data error in the instrument,
various causes.
Have the instrument checked by DrägerService
Reset the instrument to factory default settings (see Section 7
on Page 34).
If this fault occurs again:
Have the instrument checked by DrägerService.
Maintenance
®1
.
#0104 to 20 mA interface cable not connected.Disconnect the power supply, connect the 4 to 20 mA interface
and reconnect the power supply.
If operated as a stand-alone instrument without a central
controller, wire pin 3 to pin 4 (see Section 3.5 on Page 9).
#045Instrument not detecting any sensor.Check connections.
If this fault occurs again:
Have the instrument checked by DrägerService.
#051
Faulty zero calibration.Perform zero calibration (see Section 6.2.1 on Page 29).
#054
#055Faulty span calibration.Perform span calibration (see Section 6.2.2 on Page 29).
#076
#080 - #084
Instrument fault.Check electrical connections.
If this fault occurs again:
Have the instrument checked by DrägerService.
#090Sensor lock function is activated. A sensor
with a different part number has been
Use a sensor with the same part number or disable the sensor
lock function (see Section 5.9.3 on Page 25).
connected.
#105Instrument fault.Have the instrument checked by DrägerService.
#137Instrument fault.Cycle power.
If this fault occurs again:
Have the instrument checked by DrägerService.
1 DrägerService® is a registered trademark of Dräger.
Dräger Polytron 810031
Page 32
Maintenance
6.3.2Warning reference
Warning
number
#101Data-logger in Stack mode is 100 % full and
#102Data-logger in Stack mode is 90 % full.Download the data as soon as possible and clear the data-
#103
#106
#104Date or time setting not valid.Set date and time (see Section 5.7.3 on Page 20).
#107Battery for data memory empty.Have the instrument checked by DrägerService.
#110
#111
#112
#113Alarms inhibited.Enable alarms (see Section 5.7.1 on Page 19).
#164
#165
#167
#170
CauseRemedy
Download the data, clear the data-logger.
is not logging any more data.
logger.
Data error in the instrument. Some dongle
functions like data-logger, sensor test, etc.
may no longer be available.
SW dongle has been removed without being
deactivated.
Sensor warm-up phase has not ended.
Increased measurement error must be
expected.
Calibration interval expired.Recalibrate the instrument (see Section 6.2 on Page 28).
Reset the instrument to factory default settings (see Section 7
on Page 34).
If this error occurs again:
Have the instrument checked by DrägerService.
Deactivate SW dongle (see Section 5.10 on Page 27).
Wait until the sensor has warmed up. Do not calibrate before
sensor is fully warmed-up.
32Dräger Polytron 8100
Page 33
Maintenance
!
i
i
i
i
3
4
5
2
00433243.eps
1
6.4Replacing the sensor
CAUTION
Always test a newly installed sensor with target gas to
verify proper operation
If the sensor is replaced, all settings and parameters
must be checked for correctness.
NOTICE
The sensor can be replaced in the hazardous,
classified area, without interrupting the power supply.
NOTICE
If a sensor of the same type (same part number) was
previously installed, the instrument specific
configuration is retained (gas type, measurement
range, test gas, calibration interval, etc.)
Otherwise the factory default settings of the new
sensor (see Instructions for Use for the sensor) are
uploaded and will overwrite the instrument specific
configuration.
This can be prohibited if the sensor lock function (see
Section 5.9.4 on Page 25) is activated.
To replace the sensor:
Select Settings > Sensor
> Sensor change and
confirm.
Loosen set-screw (2)
2 mm Allen screw.
Unscrew bayonet ring (3).
Remove sensor from
packaging.
Insert sensor (4) into the
opening. The Dräger logo
on the sensor must point
to the mark on the sensing
head housing (5).
Lock sensor with bayonet
ring.
Tighten set-screw (2).
Mandatory for Zone 22
installations.
If necessary, apply the
label included with the
sensor to the instrument.
This identifies the gas type
from a distance, even if the power has failed.
Calibrate instrument, if necessary (see Section 6.2 on
Page 28).
Check the installation requirements and instrument for SIL
status (see Section 5.6 on Page 18).
6.5Replacing the main electronics
CAUTION
!
Always test a newly installed main electronics with
target gas to verify proper operation
If the main electronics is replaced, all settings and
parameters must be checked for correctness.
To replace the main electronics:
Turn off power to the instrument or declassify the area
according to local regulations.
Loosen set-screw and unscrew lid from instrument.
Lift the handle and pull out the bucket with the main
electronics.
Turn bucket over and pull off the power, relay and sensor
connectors.
Replace the main electronics.
Plug the power, relay and sensor connectors back into
appropriate sockets and tighten screws.
Place bucket back into the enclosure.
Screw the lid back on until it is seated (see Section 10.8 on
Page 42) and tighten set screw.
Apply power to the instrument.
Check all settings and parameters.
Calibrate instrument, if necessary (see Section 6.2 on
Page 28).
Check the installation requirements and instrument for SIL
status (see Section 5.6 on Page 18).
Dräger Polytron 810033
Page 34
Factory default settings
7Factory default settings
7.1Settings which can be changed via the menu
MenuFactory Default SettingRange
Relay active on alarm / No alarmActive on alarmOn / Off
A1 AlarmDepending on the sensor
A2 AlarmDepending on the sensor
A1 alarm: direction
(falling for DrägerSensor O2)
A2 alarm: directionRising
A1 latch modeNon-latchingLatching / Non-Latching
A2 latch modeLatching
A1 acknowledgement modeAcknowledgeableAcknowledgeable / Not acknowledgeable
A2 acknowledgement modeNot acknowledgeable
Normal operation A1 relayEnergized Energized / Not energized
Normal operation A2 relayEnergized
Hysteresis for A1 alarm at direction rising 00 to A1
Hysteresis for A2 alarm at direction rising 00 to A2
Calibration password_ _ _1
Settings password_ _ _2
LCD settingOnOn / Off
SIL statusOffOn / Off
LanguageENDE / EN / FR / ES / RU / ZH
Fault current1.2 mA0 to 3.5 mA
WarningOffOn / Off
Warning current3.0 mA0 to 3.5 mA
Warning cycle time T110 sec5 to 60 sec.
Warning cycle time T21 sec1 to 59 sec.
MaintenancestaticStatic / Dynamic
Maintenance current3.4 mA0 to 3.5 mA
Analog offset0 mA- 0.2 to 1.5 mA
HART address00 to 15
Auto-calibrationOffOn / Off
Calibration intervalDepending on sensor0 to 720
Function keyFaultsGraph, fault, warning, vitality, bump test
RisingFalling / Rising
/ Pre-acknowledgeable
34Dräger Polytron 8100
Page 35
7.2Fixed settings
FaultMeaning
Fault relayEnergized / NO
Yellow Fault LED Lit if a fault message is available (see
Section 6.3 on Page 31).
Single blink if A1 condition is present.
Double blink if A2 condition is present.
Red Alarm LED
Alarm Hierarchy
If an alarm is configured as
acknowledgeable and the alarm is
acknowledged, single/double blink
changes into continuously lit.
A2 overrides A1 on the LED. However, the
A1 and A2 relays operate independently.
I.e. if A1 is acknowledgeable and A2 is
not, and the gas concentration is such that
it triggers A1 and A2: acknowledging will
cause the A1 relay to release. However,
the red LED will still double blink as long
as the A2 condition continues to exist.
Factory default settings
Dräger Polytron 810035
Page 36
Sensor principle
8Sensor principle
Dräger electrochemical sensors are measuring the partial
pressure of gases under atmospheric conditions. The
monitored ambient air diffuses through a membrane into the
liquid electrolyte in the sensor. The electrolyte contains a
measuring electrode, a counter-electrode and a reference
electrode. An electronic potentiostat circuit ensures a constant
electrical voltage between measuring electrode and reference
electrode. Voltage, electrolyte and electrode material are
selected to suit the gas being monitored so that it is
transformed electrochemically on the measuring electrode and
a current flows through the sensor. This current is proportional
to the gas concentration.
At the same time, Oxygen from the ambient air reacts at the
counter-electrode electrochemically.
The current flowing through the sensor is amplified
electrochemically, digitized and corrected for several
parameters (e.g. the ambient temperature).
9Disposing the instrument
Disposal of electrical and electronic equipment:
EC-wide regulations governing the disposal of
electrical and electronic appliances which have been
defined in the EC Directive 2002/96/EC and in
national laws have been effective since August 2005
and apply to this device.
Common household appliances can be disposed of
using special collecting and recycling facilities.
However, this device has not been registered for
household usage. Therefore it must not be disposed
of through these channels. The device can be
returned to your national Dräger Sales Organization
for disposal. Please do not hesitate to contact Dräger
if you have any further questions on this issue.
Disposal of electrochemical sensors:
To be disposed in accordance with the material
safety data sheet of the corresponding sensor.
WARNING
!
Danger of explosions! Do not dispose sensors in fire,
risk of chemical burns!
Do not open with force.
Observe the applicable local waste disposal
regulations. For information, consult your local
environmental agency, local government offices or
appropriate waste disposal companies.
36Dräger Polytron 8100
Page 37
10Technical data
!
D
Polytron 8000
ETR 0401
Dräger Safety 23560 Lübeck, Germany
II 2G
II 2D
WARNING: Do not open when energized
I
S
F
S
I
E
A
D
L
C
9N54
0158
Class I, Div 1, Groups A,B,C,D | Class II, Div 1, Groups E,F,G
Class I, Zone 1, Group IIC | T-Code T6/T4 | Type 4X
Supply: 10...30 VDC, 0.1...0.35 A | Relay: 5 A, 30 VDC or 230 VAC
CAUTION: Do not open cover. Opening
cover with circuits alive provokes risk of
Ignition of Hazardous Atmospheres.
W
ARNING: To reduce the risk of Ignition
of Hazardous Atmospheres, the conduit
must be sealed within 18’’ of the enclosure.
WARNING: Read Manual before operating.
Gas Detector for Use in Hazardous Locations as to
Fire, Electrical Shock and Explosion Hazards only
Part No: 4544404
Serial No. XXXX-9999
Ex d [ia] IIC T6/T4 Gb
Ex tb IIIC T80/130°C Db
-40°C ≤ Ta ≤ +40/+80°C
PTB 11 AT E X 1005X
IECEx PTB 11.0005X
IP6x | P ≤ 5 W | UL only: -40°C °C≤ Ta ≤ +70
Sample
CAUTION
Specifications and restrictions in the Instructions for
Use and/or data sheets for the sensors used must be
observed. For SIL applications the Dräger Polytron
8X00 Safety Manual must be observed.
The measuring range and performance
characteristics are dependent on the installed sensor
(see the Instructions for Use and/or data sheet for the
installed sensors).
10.1 Approvals
10.1.1 ATEX, IECEx, UL, CSA
Technical data
Response time
t
0...20
t
0...90
< 12 seconds
< 30 seconds
NOTICE
i
i
At temperatures below -5 °C, the response time may
increase.
Calibration
Flow rate0.5 L/min
Zero gasNitrogen (99.9 Vol% N
)
2
Target gasOxygen / Nitrogen gas mixture
Warm-up time
Operation< 15 minutes
Calibration<
2 hours
See printout of approval label inside the shipping box of the
instrument.
10.1.2 DIN EN 50104 for DrägerSensor O
(6809720)
2
Sensor Principle
The DrägerSensor O
electrode sensor for measuring Oxygen (O
(6809720) is an electrochemical two-
2
) in ambient air.
2
Reaction at electrodes
Measuring electrode
Counter electrode
O2 + 2 H2O + 4 e– => 4 OH
2 Pb => 2 Pb
+ + 4 e
2
–
–
Environmental parameters
Pressure20.7 to 38.4 in. Hg (700 to1300 hPa)
Humidity10 to 95 %RH, non-condensing
Temperature-5 to +40 °C, short term -20 to +55 °C
Storage humidity30 to 70 %RH, non-condensing
Storage temperature0 to +40 °C
Dräger Polytron 810037
Page 38
Technical data
i
i
Cross sensitivities for DrägerSensor O
(6809720)
2
Gas / vaporChemical symbolGas concentrationMeasurement value deviation
The table shows the response of the sensor to other
gases than the target gas (cross sensitivities). The
listed values are typical and are valid for new sensors.
The table does not claim to be complete. Gas mixtures
can be displayed as the sum of all components.
There are no known cross sensitivities against
interfering gases with a concentration up to 100 ppm
(see table for concentrations higher than 1 Vol%).
in Vol% O2
The influence of O
account (partial pressure measurement).
displacement is not taken into
2
Increased sensor drift could be caused by high
concentrations of acid gases (> 1 Vol%), which might
reduce the calibration interval. The sensor service life
is reduced depending on the exposure time and the
concentration of the acid gases (e.g. lifetime in
atmospheres containing CO
hours).
:5000 Vol% CO2x
2
Organic solvents (e.g. acetone, propyl alcohol, etc.)
can penetrate into the plastic parts of the sensor. If the
sensor is exposed to higher concentrations (> 1 Vol%)
over several days, these substances could cause a
sensor drift which might reduce the calibration interval.
However, these will not reduce the sensor service life.
38Dräger Polytron 8100
Page 39
Technical data
i
i
10.1.3 DIN EN 50104 for DrägerSensor O
Sensor Principle
The DrägerSensor O
three-electrode sensor for measuring Oxygen (O
air.
NOTICE
The sensor cannot be used for Oxygen measurements
in the presence of Helium!
Reaction at electrodes
Measuring electrode
Counter electrode
Environmental parameters
Pressure20.7 to 38.4 in. Hg (700 to1300 hPa)
Humidity5 to 95 %RH, non-condensing
Temperature-40 to +60 °C, short term up to +65 °C
Storage temperature0 to +40 °C
Response time
t
0...20
t
0...90
LS (6809630) is an electrochemical
2
+ 4 H+ + 4 e– => 2 H2O
O
2
2 H2O => O2 + 4 H+ + 4 e
< 12 seconds
< 30 seconds
LS (6809630)
2
) in ambient
2
–
Calibration
Flow rate0.5 L/min
Zero gasNitrogen (99.9 Vol% N
Target gasOxygen / Nitrogen gas mixture
Warm-up time
Operation< 90 minutes
Calibration<
6 hours
)
2
Dräger Polytron 810039
Page 40
Technical data
Cross sensitivities for DrägerSensor O
LS (6809630)
2
Gas / vaporChemical symbolGas concentrationMeasurement value deviation
The table shows the response of the sensor to other
gases than the target gas (cross sensitivities). The
listed values are typical and are valid for new sensors.
The table does not claim to be complete. Gas mixtures
can be displayed as the sum of all components.
Gases with a negative cross sensitivity can cancel out
a positive sensor signal. The influence of the O
displacement is not taken into account (partial
pressure measurement).
Example:
2 Vol% ethylene in the air - cross sensitivity (table
value) = -1 Vol% O
displacement caused by 2 Vol% ethylene (2% of 20
O
2
Vol% O
Instrument display (20.9 - 1 - 0.4) = 19.5 Vol% O
CAUTION
The exposure to higher concentrations of unsaturated
hydrocarbons, alcohols or hydrogen over an extended
period of time (dose approx. 100,000 ppm x hours)
could cause the sensor to fail.
) = -0.4 Vol% O2
2
.
2
.
2
10.2 Marking
The marking is reproduced on a separate piece of paper
shipped with the instrument.
Serial Number key: The third letter of the serial number
specifies the manufacturing year: A = 2009, B = 2010,
C = 2011, D = 2012, E = 2013, F = 2014, H = 2015, J = 2016,
K = 2017, etc.
Example: Serial Number ARCB-0103: the third letter is C,
which means that the unit was manufactured in 2011.
10.3 Signal transmission to central controller
10.4 Power supply and relays
Operating voltage10 to 30 VDC at the instrument
Cable specification3-core shielded cable
Inrush current2.3 A for 2 ms at 24 VDC, 10 Ω
resistor
Operating current
2
(max.)
Relay rating (option) SPDT, 0.1 A - 5 A at 230 VAC,
80 mA at 24 VDC, without relay, nonremote sensor
100 mA at 24 VDC, with relay, remote
sensor
0.1 A - 5 A at 30 VDC, resistive load.
For safety-related applications (SIL 2),
the maximum contact rating is reduced
(see Dräger Polytron 8X00 Safety
Manual).
10.5 Physical specifications
Enclosure
material
Enclosure
protection
DisplayResolution 128 x 64 pixel, back-lit
Size (LxWxD, approx.)
Weight (approx.)
copper free aluminum or
316 stainless steel
NEMA 4X and IP 65/66/67
Without Docking Station
7.1” x 5.8” x 5.1” (180 x 150 x 130 mm)
With Docking Station
7.3” x 7.1” x 7.4” (185 x 180 x 190 mm)
Without Docking Station
aluminum4.9 lbs. (2.2 kg)
stainless steel 3168.8 lbs. (4.0 kg)
NOTICE
Frequency of measurement calculation: 1 x per
second (update of display, analog interface and
relays).
10.3.1 Analog signal
DescriptionAnalog output
Normal operation4 to 20 mA
Drift below zero3.8 to 4 mA
Measuring range exceeded20 to 20.5 mA
Instrument fault1.2 mA
Fault on analog interface> 21 mA
Maintenance signal3.4 mA steady signal or 1 Hz
modulation between 3 and
5 mA (selectable)
10.3.2 HART digital communication
Dräger Polytron 810041
10.6 Environmental parameters
Pressure20.7 to 38.4 in. Hg (700 to 1300 hPa)
Humidity0 to 100 %RH, non-condensing
Temperature-40 to +149 °F (-40 to +65 °C)
Storage temperature -4 to +149 °F (-20 to +65 °C)
See sensor data sheet for sensor specifications.
10.7 Environmental influences
For influences on the measurement performance and
restrictions of a particular sensor see sensor data sheet.
With Docking Station
aluminum 7.7 lbs. (3.5 kg)
stainless steel 31611.9 lbs (5.4 kg)
Page 42
Technical data
10.8 Tightening torque for instrument threads
The values are valid for aluminum and stainless steel
316 versions.
PartTorque Lb. In.Torque Nm
Enclosure lidmin. 44,3min. 5
Sensormin. 266min. 30
Blind plugmin. 266min. 30
Feed-through
Conduit hub
Instrument to
Docking Station
(4 screws)
min. 443min. 50
718
10.9 Tightening torque and size for field wiring
terminals