Dräger Polytron 5200, Polytron 53X0 Instructions For Use Manual

Page 1
WARNING
Strictly follow the Instructions for Use. The user must fully understand and strictly observe the instructions. Use the product only for the purposes specified in the Intended use section of this document.
Dräger Polytron 5200/53X0
Instructions for Use
www.norrscope.com
!
Page 2
www.norrscope.com
Page 3
Content
www.norrscope.com
Content
1 For your safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 General safety statements. . . . . . . . . . . . . . . . . . . . 4
1.2 Definition of alert icons. . . . . . . . . . . . . . . . . . . . . . . 4
2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1 Product overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1.1 Explosion proof instrument . . . . . . . . . . . . . . . . . . . 5
2.1.2 Explosion proof instrument with increased safety
wiring compartment (e-box) . . . . . . . . . . . . . . . . . . . 5
2.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Intended operating area and operating conditions. . 6
2.4 Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1.1 Installation restrictions . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 Electrical installation without e-box . . . . . . . . . . . . . 8
3.2.1 Power and signal wiring . . . . . . . . . . . . . . . . . . . . . . 8
3.2.2 Relay option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2.3 Remote sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.3 Electrical installation with e-box . . . . . . . . . . . . . . . . 9
3.3.1 Field wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.3.2 Field wiring: power only version. . . . . . . . . . . . . . . . 9
3.3.3 Field wiring: power and relay, or power, relay and re-
mote sensor version . . . . . . . . . . . . . . . . . . . . . . . 10
3.3.4 Attaching main instrument to e-box . . . . . . . . . . . . 11
3.3.5 Instrument wiring . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3.6 Instrument wiring: power only version . . . . . . . . . . 12
3.3.7 Instrument wiring: power and relay, or power, relay
and remote sensor version . . . . . . . . . . . . . . . . . . 12
3.4 Connecting the instrument to a controller
from Dräger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4.1 Electrical connections at the controller . . . . . . . . . 13
3.5 Normal operation . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.5.1 Analog signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.5.2 The display and LEDs . . . . . . . . . . . . . . . . . . . . . . 13
3.6 Menu navigation. . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.6.1 Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.6.2 Changing parameter values / status . . . . . . . . . . . 14
3.6.3 Exiting the menu . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.7 Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.7.1 -0- adj. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.7.2 Spn adj. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.7.3 A1 test / A2 test / Fail test . . . . . . . . . . . . . . . . . . . 16
3.7.4 A1 set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.7.5 A2 set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.7.6 A1 ris or A1 fall. . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.7.7 A2 ris or A2 fall. . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.7.8 A1 lat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.7.9 A2 lat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.7.10 A1 ack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.7.11 A2 ack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.7.12 Explanation of combining latching and
acknowledgment of alarms . . . . . . . . . . . . . . . . . . 17
3.7.13 A1 energ on or off . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.7.14 A2 energ on or off. . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.7.15 Set gas cat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.7.16 Cal at dSIR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.7.17 Cal set 3-5 or stdy . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.7.18 SNR set dd LC dSIR . . . . . . . . . . . . . . . . . . . . . . . 18
3.7.19 Pass set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.7.20 LCD on or LCD off . . . . . . . . . . . . . . . . . . . . . . . . . 18
4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.1 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.1.1 Zero calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.1.2 Span calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.2 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.2.1 Replacing the sensor . . . . . . . . . . . . . . . . . . . . . . . 21
4.3 Replacing the main electronics . . . . . . . . . . . . . . . 21
5 Default settings . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.1 Settings which can be changed via the menu. . . . 22
5.2 Fixed settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6 Sensor principle. . . . . . . . . . . . . . . . . . . . . . . . . . 23
6.1 Operating principle for DrägerSensor DD, LC . . . . 23
6.2 Operating principle for DrägerSensor DSIR . . . . . . 23
6.3 Contaminating gases for DrägerSensor DD, LC . . 23
7 Disposing of the instrument . . . . . . . . . . . . . . . . 23
8 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.1 Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.2 Marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.3 Signal transmission to central control unit . . . . . . . 24
8.4 Voltage of power supply. . . . . . . . . . . . . . . . . . . . . 24
8.5 Tightening torque . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.6 Tightening torque and wire size for field
wiring terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.7 Physical specifications . . . . . . . . . . . . . . . . . . . . . . 25
8.8 Environmental parameters. . . . . . . . . . . . . . . . . . . 25
8.9 Ambient influences. . . . . . . . . . . . . . . . . . . . . . . . . 25
9 Order list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
9.1 Detectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
9.1.1 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
9.2 Separate sensing head (not included)
when ordering . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
9.2.1 Polytron 5200 DD remote e . . . . . . . . . . . . . . . . . . 27
9.2.2 Polytron 5200 LC remote e . . . . . . . . . . . . . . . . . . 27
9.2.3 Polytron 5310 remote e . . . . . . . . . . . . . . . . . . . . . 27
9.2.4 Polytron 5300 remote e . . . . . . . . . . . . . . . . . . . . . 27
9.3 Replacement sensors
(all versions except remote e) . . . . . . . . . . . . . . . . 27
9.4 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
9.4.1 Accessories DD, LC . . . . . . . . . . . . . . . . . . . . . . . . 27
9.4.2 Accessories DSIR, PIR 3000 . . . . . . . . . . . . . . . . . 27
9.5 Spare parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Dräger Polytron 5200 / Dräger Polytron 53X0 3
Page 4
For your safety
www.norrscope.com
1 For your safety
1.1 General safety statements
Before using this equipment, carefully read the Instructions
z
for Use (IfU). Strictly follow the Instructions for Use. The user must fully
z
understand and strictly observe the instructions. Use the equipment only for the purposes and under the conditions specified in this document.
Comply with all local and national laws, rules and regula-
z
tions associated with this equipment.
z
Only trained and competent personnel are permitted to in­spect, repair and service the product as detailed in these Instructions for Use. Further maintenance work that is not detailed in these Instructions for Use must only be carried out by Dräger or personnel qualified by Dräger. Dräger rec ommends a Dräger service contract for all maintenance ac­tivities.
Use only genuine Dräger spare parts and accessories, oth-
z
erwise the proper functioning of the equipment may be im­paired.
The flameproof/explosion proof joints are not in accor-
z
dance with the relevant minimum or maximum values of EN/IEC 60079-1. The joints are not intended to be re­worked by the user.
Do not dispose of the Instructions for Use. Ensure that they
z
are retained and appropriately used by the equipment user. The measuring function of the gas detection transmitter for
z
explosion protection, according to Annex II, clauses 1.5.5,
1.5.6 and 1.5.7 of Directive 94/9/EC is currently not covered.
Substitution of components may impair Intrinsic Safety.
z
Only if intrinsic safety is involved.
1.2 Definition of alert icons
To reduce the risk of ignition of a flammable or explosive atmo­sphere, strictly adhere to the following Caution and Warning statements.
The following alert icons are used in this document to provide and highlight areas of the associated text that require a greater awareness by the user. A definition of the meaning of each icon is as follows:
!
!
-
!
i
i
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in physical injury, or damage to the product or environment. It may also be used to alert against unsafe practices.
NOTICE
Indicates additional information on how to use the instrument.
Safe connection of electrical devices
z
Never connect this instrument to other electrical devices before consulting the manufacturer or an expert.
Using the product in areas subject to explosion hazards:
z
Instruments or components for use in explosion-hazard ar­eas which have been tested and approved according to na­tional, European or international Explosion Protection Regulations may only be used under the conditions speci­fied in the approval and with consideration of the relevant legal regulations.
z
The instruments or components may not be modified in any manner. The use of faulty or incomplete parts is forbidden. The appropriate regulations must be observed at all times when carrying out repairs on these instruments or compo­nents.
4 Dräger Polytron 5200 / Dräger Polytron 53X0
Page 5
Description
www.norrscope.com
2 Description
2.1 Product overview
2.1.1 Explosion proof instrument
1 Enclosure lid 2 Bezel with main
electronics 3 Relay board (optional) 4 Enclosure bottom 5 Sensor 6 Assembled instrument
2.1.2 Explosion proof instrument with increased safety wiring compartment (e-box)
1 Enclosure lid 2 Bezel with main
electronics 3 Relay board (optional) 4 Enclosure bottom 5 Sensor 6 Feed-through 7 Field wire terminals
(not shown) 8 e-box 9 Assembled instrument
Dräger Polytron 5200 / Dräger Polytron 53X0 5
Page 6
Description
www.norrscope.com
2.2 Intended use
The Dräger Polytron 5000 is an explosion-proof instrument for the continuous monitoring of combustible gases and vapors containing hydrocarbons. The instrument is housed in a rug­ged, stainless steel or aluminum enclosure for indoor and out­door applications. The instrument can be connected through a sealed conduit or approved cable gland to a Dräger monitoring system or a Programmable Logic Controller (PLC). With the optional alarm relay configuration, the instrument can be oper ated as a stand-alone instrument. The instrument is designed to be installed in permanent locations and is approved for use in hazardous, classified areas, see Section 8.1 on Page 24.
Caution: Not tested in oxygen enriched atmospheres (>21%
). High off-scale readings may indicate an explosive con-
O
2
centration.
Dräger Polytron 5200 with DrägerSensor
Catalytic bead sensor for monitoring of combustible gases and vapors containing hydrocarbons.
{
Measuring range: 0 to 100 %LEL
Dräger Polytron 5200 with DrägerSensor® Ex LC
Catalytic bead sensor for monitoring of combustible gases and vapors containing hydrocarbons.
{
Measuring range: 0 to 10 %LEL
® 3)
Ex DD
2)
2)
2.3 Intended operating area and operating conditions
Hazardous areas classified by zones:
The device is intended to be used only in hazardous areas classified zone 1 or zone 2, within a temperature range as marked on the device, where gases of explosion groups IIA, IIB or IIC and temperature class T4 or T6 (depending on the
-
maximum ambient temperature) or dusts of groups IIIA, IIIB or IIIC may be present.
Hazardous areas classified by divisions:
The device is intended to be used only in hazardous areas Class I&II, Div. 1 or Div. 2, within a temperature range as marked on the device, where gases or dusts of groups A, B, C, D or E, F, G and temperature class T4 or T6 (depending on the maximum ambient temperature) may be present.
2.4 Design
The instrument is powered by 10 to 30 VDC. Gas concentra­tions, status messages, and menu choices are displayed on a 4 digit LCD display and 3 colored LEDs. The measured gas concentration is converted to a 4 to 20 mA analog output signal. The instrument detects whether it is operating in current source versus sink mode when power is applied.
Dräger Polytron 5310 with DrägerSensor® IR
IR sensor for monitoring of combustible gases and vapors con­taining hydrocarbons.
{
Measuring range: 0 to 100 %LEL
Dräger Polytron 5300 with Dräger PIR 3000
2)
1)
IR sensor for monitoring of combustible gases and vapors con­taining hydrocarbons.
{
Measuring range: 0 to 100 %LEL
1) not yet available
2) Lower Explosion Limit, depending on the substance and the regu­lations applicable at the respective location.
3) DrägerSensor® is a registered trademark from Dräger
2)
WARNING
!
Explosive. Not to be used in oxygen enriched atmo­spheres. None of the Polytron 5000 instruments are certified and approved to be operated in oxygen enriched atmospheres.
Navigation through the menu is done by tapping a magnetic wand on the glass at the Up and Down arrows or OK.
The instrument can be configured, calibrated and maintained non-intrusively without declassifying the area.
6 Dräger Polytron 5200 / Dräger Polytron 53X0
Page 7
3 Operation
www.norrscope.com
3.1 Installation
To ensure overall system performance and effectiveness, the selection of an installation site for the instrument is the most important factor. Considerable thought must be given to every detail of installation, particularly:
{
the local, state, federal codes and requirements that govern the installation of gas monitoring equipment
{
the electrical codes that govern the routing and connec­tion of electrical power and signal cables to gas monitoring equipment.
{
for non-conduit installations, an approved cable gland must be used (e.g. Hawke A501/421/A/¾” NPT or equivalent) see Section 8.1 on Page 24. It might be necessary to connect the shield of the cable to the ca­ble gland and to the controller in order to improve RFI immunity
{
the full range of environmental conditions to which the instruments will be exposed to.
{
the physical data of the gas or vapor to be detected.
{
the specifics of the application, (e.g. possible leaks, air movement/draft, etc.)
{
the degree of accessibility required for maintenance purposes
{
the types of optional and accessory equipment that will be used with the system
{
any other limiting factors or regulations that would affect system performance or installations.
{
The flameproof/explosion proof enclosure provides three ¾" NPT openings, which can be used for field wir­ing, direct attachment of a sensor or wiring of a remote sensor. For correct tightening torques of conduit hubs, plugs and sensor
{
Secondary circuit intended to be supplied from an isolating source (N/A for relay circuits).
{
The optional increased safety terminal box provides up to four 20 mm openings, which can be used for field wir­ing or wiring of a remote sensor. The permissible cable diameter range is 7-12 mm. When installed at locations exceeding ambient temper-
{
atures of 55 °C, use only appropriate wiring, specified for at least 25 °C above the maximum ambient temperature.
{ Strip wire insulation by 5-7 mm.
see Section 8.5 on Page 24.
Operation
Connect the wires as indicated in wiring figure in sec-
{
tion 3.2.2 (also showing grounding conductor terminal).
3.1.1 Installation restrictions
The instrument must have between 10 to 30 VDC at the instru­ment. This ultimately determines the distance the instrument can be mounted from the controller or power supply. The in strument accepts wire sizes of 12 to 24 AWG (0.2 to 2.5 mm2). Use at least a three-conductor, shielded cable.
The instrument must not be exposed to radiant heat that will cause the temperature to rise beyond the limits stated in see Section 8.8 on Page 25. The use of a reflecting shield is rec­ommended.
The enclosure is weatherproof and suitable for outdoor instal­lation. The use of the optional splash guard is recommended to protect the sensor from water and dust.
Each instrument must be installed and operated in an environ­ment that conforms to the specifications, see Section 8 on Page 24.
NOTICE
i
i
The instrument may be equipped with a dust plug at the conduit entry. This plug is not explosion-proof nor meant to be watertight, and must be removed before connecting the instrument to a sealed conduit.
-
Dräger Polytron 5200 / Dräger Polytron 53X0 7
Page 8
Operation
www.norrscope.com
3.2 Electrical installation without e-box
CAUTION
!
!
Ensure wiring for relays and connections for sensor are made before applying power.
3.2.1 Power and signal wiring
z
Loosen set-screw and unscrew lid from instrument.
1. Pull out the bezel, by grasping the notches on either side of the display with your fingers and pull it up.
3.2.2 Relay option
If the relay option has been installed, the wires for the alarm devices will be connected to the 9-pin connector.
Turn bezel over and pull off the 9-pin connector.
z
Connect the wires for alarm 1, alarm 2 and fault to the ter-
z
minals, as indicated in the following wiring table.
z
With factory default settings and during normal operation, the relays are energized. This provides “fail-safe” opera­tion. The terminal designators indicated in the following wiring table are shown as factory default and normal oper ation mode, see Section 5 on Page 22.
z
Plug connector back into socket.
NOTICE
i
i
To ensure that a fault is recognized - without having to look at the instrument - an alarm device must be con­nected to the fault relay.
-
z
Turn bezel over and pull off the 3-pin connector.
Connect the three wires for power and signal to the appropriate terminal as indicated in the following wiring table and figure. Fasten terminal screws with the correct torque according to the table Tightening torque and wire size for field wiring terminal see Section 8.6 on Page 24.
If operated as a stand-alone instrument, wire pin 1 to pin 3.
z
z
Plug connector back into socket.
z
The cable shielding should only be connected at the con­troller.
z
Place bezel back into the enclosure.
z
Screw the lid back on, until it is seated (correct torque) see Section 8.5 on Page 24, and tighten set-screw.
3 Pin Connector (Power and Signal)
Pin 1 2 3 Mark + none S
Function V + V -
4-20 mA
signal
1 Connector for DrägerSensor DD, LC, and DSIR
9 Pin Connector (Relays)
Fault Relay A2 Relay A1 Relay Pin 1 2 3 4 5 6 7 8 9 Mark NO C NC NO C NC NO C NC
3.2.3 Remote sensor
The remote junction box (part number 4544098 for stainless steel or 4544099 for aluminum) allows sensor installation at a location away from the instrument electronics. This remote in­stallation permits easy setup when the sensor must be located in a difficult to reach or awkward position, see Polytron 5000 Junction Box Instructions for Use 4544286 for more details.
8 Dräger Polytron 5200 / Dräger Polytron 53X0
Page 9
Operation
www.norrscope.com
3.3 Electrical installation with e-box
Installing this configuration is a two step process.
First, the e-box is mounted and connected to the field wires. Second, the main instrument enclosure with the electronics and sensor is attached to the e-box.
The e-box can be pre-mounted, wired and sealed with the sup­plied cover. Once the site is ready for commissioning, the instrument is then hooked up to the e-box and taken into oper­ation; avoiding that the instrument is damaged during the construction phase.
The connection between the e-box and the main instrument is realized via a ‘feed-through’. Depending on the instrument selected, there are 3 types of feed-through.
{
3 wire for power (part number 4544182)
{
9 wire for power and relay (part number 4544169)
{
14 wire for power, relay and remote sensor (part number 4544168)
In some e-boxes, the field wires are terminated in spring clamp terminals.
Screw driver or special tool
3.3.1 Field wiring
Connect all applicable field wires to the respective terminals.
3.3.2 Field wiring: power only version
Solid Oval (Power Only)
Terminal 1 (top) Terminal 2 Terminal 3
V + V - 4 to 20 mA Signal
1. Insert screw driver (width 3 mm) into the spring clamp terminal.
2. Press spring down to open the clamp in the lower part. Insert stripped cable end or ferrule (for stranded wires) into the lower part.
3. Remove screw driver. The electrical connection is ensured via the constant pressure of the spring.
or
1. Insert special tool (part number 8318376) into the spring clamp terminal.
2. Press spring down to open the clamp in the lower part. Insert stripped cable end or ferrule (for stranded wires) into the lower part.
3. Remove special tool. The electrical connection is ensured via the constant pressure of the spring.
WARNING
!
Bare cables must not stick out of the spring terminals. The method of explosion protection during mainte­nance is based on the condition that it is not possible to contact bare cable parts by a probe of 2.5 mm diam­eter (definition of IP 30).
Dräger Polytron 5200 / Dräger Polytron 53X0 9
Page 10
Operation
www.norrscope.com
3.3.3 Field wiring: power and relay, or power, relay and remote sensor version
Solid Oval (Power) X1
Terminal 1 (top)
V + V - V -
Terminal 2 Terminal 3 Te rmin a l 4 Terminal 5
4 to 20 mA Signal
Ground
Dashed Oval (Relay) X2
Terminal 1 (left) 2 3 4 5 6 7 Fault Fault Alarm 2 Alarm 2 Alarm 1 Alarm 1 Ground
Solid Rectangle (Remote) X3
Sensing Head Polytron SE Ex PR M1 DD 6812711 Sensing Head Polytron SE Ex PR M2 DD 6812710 Sensing Head Polytron SE Ex PR M3 DD 6812718 Sensing Head Polytron SE Ex HT M DD 6812720
Sensing Head Polytron SE Ex LC M1 DD 6812722 Sensing Head Polytron SE Ex LC M2 DD 6812721 Sensing Head Polytron SE Ex LC M3 DD 6812719
Sensing Head DrägerSensor IR complete Set e 6811165 Sensing Head DrägerSensor IR complete Set e 6811265
Sensing Head Dräger PIR 3000 complete Set e 6811160 Sensing Head Dräger PIR 3000 complete Set e 6811270
1) not yet available
Sensor Wire Color
S1 S2 S3 S4 S5
Wire S2
Black
to
Yell ow
S1
Wire S2
Black
to
Yell ow
S1
Wire S2
Black
to
Yell ow
S1
1 1
----- ----- Brown Black Red
Wire S4
to
S5
Wire S4
to
S5
Wire S4
to
S5
Brown
Brown
Brown
10 Dräger Polytron 5200 / Dräger Polytron 53X0
Page 11
3.3.4 Attaching main instrument to e-box
www.norrscope.com
To attach the main instru­ment to the e-box, pull the hinge pin out of the e-box. Align the boss of the instru­ment with the boss of the e-box and push the hinge pin back in. The instrument is now supported and can swivel freely to give access to the wiring.
1 Hinge pin
Operation
Dräger Polytron 5200 / Dräger Polytron 53X0 11
Page 12
Operation
www.norrscope.com
3.3.5 Instrument wiring
3.3.6 Instrument wiring: power only version
Connect the instrument wires from the feed-through to the respective terminal in the e-box.
Solid Oval (Power Only)
Terminal 1 (top) Terminal 2 Terminal 3
V + V - 4 to 20 mA Signal
Red Black Brown
Rectangle (Ground Lug Connection)
Connect ground wire from d-box to ground lug connection
3.3.7 Instrument wiring: power and relay, or power, relay and remote sensor version
Plug the connectors of the feed-through into the sockets of the e-box interface PCB (X11, X12, X13).
.
Rectangle (Ground Lug Connection)
Connect ground wire from d-box to ground lug connection
If a relay option is used, and the default configuration for NO and NC does not fit for the application, the wiring must be changed at the relay board,
z
To rewire the Alarm 1 relay, move the gray wire from A1-
see Section 5 on Page 22.
NO to A1-NC.
z
To rewire the Alarm 2 relay, move the blue wire from A2­NO to A2-NC.
To rewire the Fault relay, move the violet wire from FLT-
z
NO to FLT-NC.
z
The wires to A1-C, A2-C and FLT-C should not be moved.
12 Dräger Polytron 5200 / Dräger Polytron 53X0
After all connections are made, swing instrument onto e-box (ensuring that no wires are pinched and the seal is not compro­mised) and tighten all four screws with correct torque see Section 8.5 on Page 24
Page 13
3.4 Connecting the instrument to a controller
www.norrscope.com
from Dräger
For hook-up information, please refer to the Instructions for Use included with the Dräger controller (e.g. Regard, Quad Gard).
3.4.1 Electrical connections at the controller
Connect the shield of the wires to the instrument earth ground of the controller (e.g. chassis, ground busbar, etc.)
3.5 Normal operation
NOTICE
i
i
Before leaving the instrument for normal operation, check the configuration and calibration for the proper settings.
Switch power supply on.
z
The instrument will go through a start-up sequence (LCD / LED test, software version, and initialization) and start the warm-up period. The display shows
Operation
for the DD and LC sensor an overrange has to be ac-
{
knowledged with OK.
{
when a fault has been detected the display toggles between ‘Err’ and a number and the yellow LED is lit,
-
see Section 4.2 on Page 20
If the optional relay board is installed:
{
when the first alarm has been triggered the red LED blinks in single mode
{
when the second alarm has been triggered the red LED blinks in double mode
If an alarm is acknowledgeable, and it is acknowledged, the blinking of the red LED changes to steady lit and remains lit until the alarm condition is not present any more, Section 3.7.10 on Page 17.
The segments of the display and LED symbols.
see
and the instrument emits the maintenance signal on the analog output see Section 4.2 on Page 20. After the warm-up period, the instrument goes into normal operation.
Pressing the Down arrow during the warm-up period will dis­play the current sensor selection for example SNR dd.
Pressing the Up arrow will exit the function.
Changing the sensor selection during warm-up will cause
Cycling the power will initialize the instrument to accept the change. All configurations must be checked, and the instru­ment must be calibrated.
3.5.1 Analog signals
The current output of the instrument during normal operation is between 4 and 20 mA and is proportional to the detected gas concentration.
Polytron 5000 uses different current values to indicate various modes of operation, see Section 8.3 on Page 24. This follows the NAMUR NE43 standard.
3.5.2 The display and LEDs
Symbol LED Description
Red Alarm Triggered
Yell ow Fault / Warning
Green
Power ON
Normal Operation
In normal operation, the display shows the measured gas con­centration and unit of measurement. The green LED is lit.
The following special symbols may also be displayed:
{ when the measuring range of the sensor has been
exceeded
Dräger Polytron 5200 / Dräger Polytron 53X0 13
Page 14
Operation
www.norrscope.com
3.6 Menu navigation
Ta p OK when the full password is displayed.
z
Tapping the magnetic wand (part number, 4544101, blue body) over the Up and Down arrows scrolls through the menu selections.
When the last menu item is reached, the menu will bottom-out, and the Up arrow has to be used to scroll back up through the menu.
The active menu item as well as its current value or status will flash on the display as it scrolls.
If the displayed value matches the set password, access will be given to the rest of the menu. If an incorrect password is en tered, the instrument will return to the measurement mode, see Section 3.7 on Page 15.
3.6.2 Changing parameter values / status
Select the menu item to be accessed using the Up and Down arrows.
z
When the desired menu item is displayed, tap OK with the magnetic wand.
The current value or status will flash.
z
The Up and Down arrows allow adjusting the value of a nu-
z
merical parameter or scrolling through preset choices. Once the display shows the desired value or choice, tap
z
OK to validate the new parameter. This returns to the main menu, where another menu item can be selected.
3.6.3 Exiting the menu
To get back into the measurement mode, tap Up until the menu is exited. The actual gas concentration will be displayed.
-
NOTICE
i
i
The instrument is designed for the magnetic wand to be used with the enclosure lid in place. If the enclosure lid is not in place, the magnetic wand may activate two or more buttons at once (cross-talk).
3.6.1 Password
The use of a password is optional on the Polytron 5000. A password consists of a 4-digit number from 0000 to 9999; a value of 0000 disables password protection and allows anyone to access the menu.
z If the password is enabled, from measurement mode, tap
the Down arrow. The 4-digit LCD will then show ‘0000’, with the first zero on
z
the left blinking. Use the Up and Down arrows to increment or decrement
z
this digit, then tap OK.
z
The second digit will blink; set the correct value using the Up and Down arrows.
z
Repeat the process for the other two digits.
14 Dräger Polytron 5200 / Dräger Polytron 53X0
Page 15
3.7 Menu
www.norrscope.com
Operation
1 Only for relay version 2Only for DSIR
Dräger Polytron 5200 / Dräger Polytron 53X0 15
Page 16
Operation
www.norrscope.com
3.7.1 -0- adj
Allows adjusting the zero reference point of the sensor, see Section 4.1.1 on Page 19
3.7.2 Spn adj
Allows adjusting the sensitivity to match the known concentra­tion of an applied calibration gas, see Section 4.1.2 on Page 19.
NOTICE
i
i
Sections 3.7.3 Relay test to 3.7.14 A2 energized are only for users who have installed the relay option. All other users continue with section 3.7.17. Cal set.
3.7.3 A1 test / A2 test / Fail test
With these functions the status of the relays and LED can be changed for test purposes (e.g. to check the function of alarm devices connected to the relays). After exiting this function, the status of the relays, if changed, will automatically return to their original status.
3.7.4 A1 set
Configuring the low level alarm set point.
Leaving the magnet on the arrow will be interpreted as multiple tapping.
An alarm set point of zero disables the alarm.
3.7.5 A2 set
Configuring the high level alarm set point.
Leaving the magnet on the arrow will be interpreted as multiple tapping.
An alarm set point of zero disables the alarm.
Alarm hierarchy:
A2 overrides A1 on the LED and display. However, the A1 and A2 relays operate independently, i.e. if A1 is acknowledgeable, A2 is not~, and the gas concentration is such that it triggers A2: Acknowledging will cause the A1 relay to release. However, the red LED will still double blink as long as the A2 condition continues to exists.
NOTICE
i
i
When the relays are activated, alarm devices will be switched on.
Alarm hysteresis:
In order to avoid chatter at an alarm threshold, a fixed hyster­esis of 5% of full scale is programmed.
3.7.6 A1 ris or A1 fall
Configuring whether the alarm should be triggered by a rising or falling gas concentration.
3.7.7 A2 ris or A2 fall
Configuring whether the alarm should be triggered by a rising or falling gas concentration.
16 Dräger Polytron 5200 / Dräger Polytron 53X0
Page 17
Operation
www.norrscope.com
3.7.8 A1 lat
Configuring the A1 alarm level to status latching or non­latching.
Latching means that once the alarm level is reached, the in­strument will trigger the alarm. It will remain in alarm status even if the gas concentration subsequently does not meet the alarm condition any more. To clear a latching alarm it has to be acknowledged with OK.
In non-latching mode, the alarm status clears if the gas con­centration does not meet the alarm condition any more.
3.7.9 A2 lat
Configuring the high level A2 alarm status to latching or non­latching.
3.7.10 A1 ack
3.7.12 Explanation of combining latching and acknowledgment of alarms
Since the concepts of latching status and acknowledgment can be confusing, the following four combinations are offered for clarification:
Latching and Acknowledgeable
Latching and Non-Acknowledgeable
Non-Latching and Acknowledgeable
Non-Latching and Non-Acknowledgeable
3.7.13 A1 energ on or off
Configuring the A1 alarm relay as energized.
Energized ON means that the alarm relay will change the sta­tus in alarm condition or power off (fail safe).
Relay must be reset manually and can be reset before the alarm condition clears.
Relay must be reset manually. Relay cannot be reset before the alarm condition clears.
Relay will reset automatically when the alarm condition clears or can be reset manu­ally.
Relay will reset automatically when the alarm condition clears. Relay cannot be reset manually before the alarm condition clears.
Configuring the A1 alarm as acknowledgeable or non­acknowledgeable.
Acknowledgeable means that the alarm relay can be reset before alarm conditions clear.
Non-acknowledgeable means that the alarm relay can not be reset until the alarm conditions clear.
3.7.11 A2 ack
Configuring the A2 alarm as Acknowledgeable or Non­acknowledgeable.
3.7.14 A2 energ on or off
Configuring the A2 alarm relay as energized.
The fault relay is always energized ON (fail safe).
3.7.15 Set gas cat
Only for DSIR. Allows selecting the gas category.
The actual selection is made at the DSIR. While in this menu, the Polytron 5000 will display gas category as it is output from the DSIR. See Instructions for Use for DrägerSensor IR.
Dräger Polytron 5200 / Dräger Polytron 53X0 17
Page 18
Operation
www.norrscope.com
3.7.16 Cal at dSIR
Only for DSIR. Allows the user to calibrate directly at the DSIR.
The actual calibration is done at the DSIR. This is recommend­ed for the first time the DSIR is connected to the Polytron 5000. While in the menu, the Polytron 5000 acts as a display device for the DSIR. The Up and Down arrows have no effect on the reading while in this menu. See Instructions for Use Dräger­Sensor.
3.7.17 Cal set 3-5 or stdy
Configuring the maintenance signal
The maintenance signal is transmitted on the 4 to 20 mA ana­log output anytime the menu is accessed. It is user-selectable between:
{
a steady 3.4 mA output signal
{
an oscillating 3 to 5 mA signal with a frequency of 1 Hz
3.7.19 Pass set
Configuring the password
Only numbers 0 to 9 are allowed for the password.
The use of a password is optional on the Polytron 5000. A password consists of a 4-digit number from 0000 to 9999; a value of 0000 disables password protection and allows anyone to access the menu.
To set/change the password, enter the function
z
the 4-digit LCD will show ‘0000’ with the first digit on the left blinking
use the Up and Down arrows to increment or decrement
z
this digit, then tap OK
z
the second digit will blink; set the correct value using the Up and Down arrows
z
repeat the process for the other two digits tap OK when the full password is displayed
z
z
once a password is set, it has to be entered to gain access to the menu.
3.7.18 SNR set dd LC dSIR
Configuring the sensor to be used
Polytron 5000 can use one of the following sensors:
1. DrägerSensor DD catalytic bead sensor for 0 to 100 %LEL monitoring
2. DrägerSensor LC catalytic bead sensor for 0 to 10 %LEL monitoring
3. DrägerSensor IR infrared sensor for 0 to 100 %LEL monitoring
CAUTION
!
!
The sensor used with Polytron 5000 has to be selected in this menu item for proper linearization of the sensor signal. After changing the selected sensor, the instrument will re-start. When exchanging a sensor for a different type, the in­strument must be calibrated.
3.7.20 LCD on or LCD off
This feature allows turning the LCD off in measurement mode, effectively turning the Polytron 5000 into a non-display instru ment.
The functionality of the instrument remains active, indepen­dently of the LCD state.
When in measurement mode and the LCD is set to OFF, press­ing the Down arrow still gives complete access to the menu.
In case of an alarm, the red LED will blink. In case of a fault, the yellow LED will lit and the display will toggle between ‘Err’ and a number.
-
18 Dräger Polytron 5200 / Dräger Polytron 53X0
Page 19
Maintenance
www.norrscope.com
4 Maintenance
4.1 Calibration
Calibration of the instrument must be performed at regular intervals as detailed in the sensor data sheet.
WARNING
!
Calibration gas must not be inhaled! See appropriate Material Safety Data Sheets. For proper operation, never adjust the span before completing zero adjustment. Performing these opera­tions out of order will cause the calibration to be faulty.
General procedure
1. Attach the pressure regulator to the calibration gas cylin­der.
2. Fit calibration adapter to the sensor.
3. The gas flow should be between 0.5 and 2.0 L/min. At higher altitudes, a flow greater than 0.5 l/min might be nec­essary (reduced partial pressure).
4. Connect the tubing to the barbed fitting.
NOTICE
i
i
In some cases (e.g. combustible vapors) a more ac­curate calibration can be obtained when the Dräger calibration chamber is used instead of a calibration gas cylinder.
1 Pressure regulator 2 Calibration adapter 3 Gas cylinder 4Tubing
Only for DSIR
Place the magnetic wand onto the sensor surface area
z
marked by and hold down for at least 5 seconds. The display will change to a negative value (e.g. -10% LEL) as long as the magnetic wand is held in place. This initiates the auto-zeroing of the sensor.
Remove the magnetic wand. After 30 seconds, the sensor
z
will exit the auto-zeroing. As confirmation the display will change again to a negative value (e.g. -10% LEL).
After this procedure the zero at the transmitter is adjusted for eventual offsets.
4.1.1 Zero calibration
1. Enter the function with OK.
2. Apply synthetic air or Nitrogen.
z
The display will show the current value blinking (e.g. ‘2’).
z
Wait for the value to stabilize.
3. Use the Up / Down arrow to adjust the value to 0.
4. Acknowledge with OK; the display will then jump back to
the main menu.
5. Turn off the gas flow and remove the calibration adapter from the sensor, or disconnect tubing.
NOTICE
i
i
Ambient air can be used to zero the sensor instead of Nitrogen or Synthetic Air if the area is known to be free of the target gas or any gas to which the sensor may be cross-sensitive (as listed on the sensor data sheet). In this case, no cylinder or calibration adapter is needed for the zero calibration.
4.1.2 Span calibration
1. Check the set gas category (only for DSIR).
2. Enter the function with OK.
3. Apply span gas (e.g. 50 %LEL Methane). The display will show the current value blinking (e.g. ‘48’).
z
Wait for the value to stabilize.
z
4. Use the Up / Down arrow to adjust the value to 50.
5. Acknowledge with OK; the display will then jump back to the main menu.
6. Turn off the gas flow and remove the calibration adapter from the sensor, or disconnect tubing.
Dräger Polytron 5200 / Dräger Polytron 53X0 19
Page 20
Maintenance
www.norrscope.com
4.2 Troubleshooting
Display Sequence Fault Meaning Remedies
PLS CONN SNR Please connect sensor Remove power; Attach sensor; Reapply power
Reading is above full scale
Errors
Err 100 X Instrument self test error Call Dräger Service Err 101 X Firmware CRC error Call Dräger Service Err 102 X RAM error Call Dräger Service Err 103 X Flash memory error Call Dräger Service Err 104 X EEPROM memory error Call Dräger Service Err 105 X Analog signal converter error Call Dräger Service Err 106 X Incorrect software installed Call Dräger Service
Err 108 X 4-20 output error
Err 109 X Sensor failure Remove power; reconnect or replace sensor; reapply power Err 112 X Zero calibration failed Calibrate sensor Err 113 X Span calibration failed Calibrate sensor
Err 115 X Gas value is under range
Err 117 X
Err 198 X Memory error Call Dräger Service Err 199 X Factory calibration required Call Dräger Service
Warnings
Info 301 Sensor is warming up Wait for warm-up phase to end
Magnetic key is stuck “on” for more than 1 minute
Messages in the 100 range are faults; messages in the 300 range are warnings which can easily be cleared.
Display remains as long as reading is above full scale. For the DD and LC sensors an overrange has to be acknowledged with OK.
Remove power; check the field wiring from the Polytron 5000 to the control system; Reapply power
Calibrate the sensor zero point Check environment for possible gases with negative cross­sensitivity to the measured gas
Call Dräger Service
20 Dräger Polytron 5200 / Dräger Polytron 53X0
Page 21
Maintenance
!
www.norrscope.com
4.2.1 Replacing the sensor
To replace the sensor:
1. Turn off power to the instrument or declassify the area ac­cording to local procedures.
2. Loosen set-screw and unscrew lid from instrument.
3. Remove the bezel, by grasping the notches on either side of the display with your fingers and pull it up.
4. Unplug the sensor cable.
5. Unscrew the sensor.
6. Insert the sensor wires through the threaded port in the housing.
7. Screw the sensor into the port.
CAUTION
!
Five threads must be engaged to ensure explosion­proof status.
8. Plug the sensor wires into the socket as indicated.
4.3 Replacing the main electronics
To replace the main electronics:
1. Turn off power to the instrument or declassify the area ac­cording to local procedures.
2. Loosen set-screw and unscrew lid from instrument.
3. Remove the bezel, by grasping the notches on either side of the display with your fingers and pull it up.
4. Unplug the sensor cable, power and relay wires (if applica­ble).
5. Replace the main electronics.
6. Plug the sensor cable, power and relay wires (if applicable) into the appropriate sockets.
7. Re-install sensor electronics by pushing the bezel back into place.
8. Screw the lid back on, until it is seated, and tighten set­screw.
9. Apply power to the instrument.
10. Review and modify the configuration if necessary.
11. Calibrate instrument, see Section 4.1 on Page 19.
12. Always test a newly-installed sensor with target gas to verify proper operation.
9. Re-install sensor electronics by pushing the bezel back into place.
10. Screw the lid back on, until it is seated, and tighten set­screw.
11. Apply power to the instrument.
12. Calibrate instrument, see Section 4.1 on Page 19.
13. Always test a newly-installed sensor with target gas to verify proper operation.
Dräger Polytron 5200 / Dräger Polytron 53X0 21
Page 22
Default settings
www.norrscope.com
5 Default settings
5.1 Settings which can be changed via the menu
Menu Default Setting
DD, DSIR LC
A1 Set 20 %LEL 2.0 %LEL A2 Set 40 %LEL 4.0 %LEL A1 Rising/Falling rising rising A2 Rising/Falling rising rising A1 Lat / nLat nLat nLat A2 Lat / nLat Lat Lat A1 Ack / nAck Ack Ack A2 Ack / nAck nAck nAck A1 Enrg / wired in e-box energized / NO energized / NO A2 Enrg / wired in e-box energized / NO energized / NO Cal Set steady steady Pass Set 0000 0000 LCD Set on on
5.2 Fixed settings
Fault Meaning
Fault Relay (can­not be changed)
Yellow Fault LED Lit when a Warning or Fault condition is
Red Alarm LED Single blink when A1 condition is present
energized / NO
present, see Section 4.2 on Page 20.
Double blink when A2 condition is present
If an alarm is configured acknowledge­able and the alarm is acknowledged, single/double blink changes into continuous lit
Alarm Hierarchy: A2 overrides A1 on the LED. However, the A1 and A2 relays op­erate independently, i.e. if A1 is acknowl­edgeable, A2 is not, and the gas concentration is such that it triggers A1 and A2: Acknowledging will cause the A1 relay to release. However, the red LED will still double blink as long as the A2 condition continues to exist.
22 Dräger Polytron 5200 / Dräger Polytron 53X0
Page 23
Sensor principle
www.norrscope.com
6 Sensor principle
6.1 Operating principle for DrägerSensor DD, LC
The DrägerSensor is a transducer for measuring the partial pressure of combustible gases and vapors contained in ambi­ent air. It uses the heat-of-combustion principle.
The monitored air diffuses through the sintered metal disc into the sensor. The mixture of combustible gases and vapors are catalytically combusted at a heated detector element (pelli­stor). The monitored air supplies the Oxygen required for the combustion (must be greater than 12 Vol%). Due to the result­ing heat-of-combustion, the detector element gets hotter. This increase in heat causes a change of resistance in the detector element, which is proportional to the concentration of the mix­ture of combustible gases and vapors in the monitored air.
In addition to the catalytically active detector element, there is an compensator element. Both elements are parts of a Wheat­stone bridge. Thus environmental effects like changes in tem­perature or humidity are almost entirely compensated.
6.2 Operating principle for DrägerSensor DSIR
The DrägerSensor IR infrared gas sensor is a gas transmitter designed to determine the concentration of gases and vapors in the ambient air. The principle of measurement is based on the concentration-dependent absorption of infrared radiation in measured gases.
The monitored ambient air diffuses through sintered material into the flameproof housing of a measuring cuvette. The broad band light emitted by the radiator passes through the gas in the cuvette and is reflected by the cuvette walls from where it is di­rected towards the inlet window of a dual element detector. One channel of the detector measures the gas-dependent light transmission of the cuvette (measuring channel), the other channel is used as reference. The ratio between measuring and reference signal is used to determine the gas concentra tion in the cuvette. The cuvette is heated to avoid condensation of the atmosphere’s moisture content.
Internal electronics and software are used to calculate the con­centration. As an output signal, the gas sensor emulates the half bridge of a catalytic bead sensor.
6.3 Contaminating gases for DrägerSensor DD, LC
NOTICE
i
i
Certain substances in the atmosphere to be monitored may impair the sensitivity of the sensors which are in stalled in the sensing heads:
a) polymerizing substances such as acrylonitrile, buta­diene and styrene,
b) corrosive compounds such as halogenated hydro­carbons (releasing halogens such as bromine, chlor­ine or fluorine when being oxidized) and halogen hy­dride acids as well as acidic gaseous compounds such as sulphur dioxide and nitrogen oxides,
c) catalyst poisons such as sulphurous and phospho­rous compounds, silicon compounds (especially silicones), and metal-organic vapors. The sensors used are based on measuring elements (pellistors) which are “poison-resistant” (PR) and so have a longer lifetime than conventional sensors if catalyst poisons occur.
A calibration check might be necessary, if the sensor was ex­posed to a high concentration of combustible gases or vapors for an extended period of time or to contaminants as listed above.
7 Disposing of the instrument
Disposing of electric and electronic equipment:
EC-wide regulations for the disposal of electric and
-
-
electronic appliances which have been defined in the EC Directive 2002/96/EC and in national laws have been effective since August 2005 and apply to this instrument. Special collecting and recycling options have been established for households. However, as this instrument has not been registered for household usage, it must not be disposed of through these means. The instrument can be returned to your national Dräger Sales Organization for disposal. Please do not hesitate to contact the above if you have any further questions on this issue.
-
Dräger Polytron 5200 / Dräger Polytron 53X0 23
Page 24
Technical data
Sample
www.norrscope.com
8 Technical data
8.1 Approvals
See printout of approval label.
8.2 Marking
The marking is reproduced on a separate piece of paper shipped with the transmitter.
Serial No. key: The year of manufacture is indicated by the third letter in the serial number: A = 2009, B = 2010, C = 2011, D = 2012, E = 2013, F = 2014, H = 2015, J = 2016, K = 2017, etc.
8.4 Voltage of power supply
Operating Voltage 3-core shielded cable, 10 to 30 VDC
In-rush Current
Operating Current (maximum)
Relay Rating
300 mA for 140 msec @ 24 VDC 350 mA for 140 msec @ 24 VDC with relay option
95 mA @ 24 VDC w/o relay non­remote sensor 145 mA @ 24 VDC with relay remote sensor
SPDT 5 A @ 230 VAC, 5 A @ 30 VDC, resistive load
8.5 Tightening torque
Part TQ Lb. In. TQ Nm
lid min. 266 min. 30
sensors min. 266 min. 30
plugs min. 266 min. 30
conduit hubs min. 443 min. 50
instrument to e-box 71 8
8.6 Tightening torque and wire size for field wiring terminals
Example: Serial No. ARBH-0054: the third letter is B, which means that the unit was manufactured in 2010.
8.3 Signal transmission to central control
Current Meaning
4 mA Zero point 20 mA Full scale deflection < 1.2 mA Fault
3.8 mA to 4 mA Sensor drift below zero 20 mA to 20.5 mA Measuring range exceeded > 21 mA Fault on the analog output
3.4 mA steady or 1 Hz modulation between 3 and 5 mA (user selectable)
unit
Signal for maintenance mode
Electronic
TQ Lb. In. Wire Size AWG Wire Size mm
All field wiring
terminals
4.4 - 7.0
(0.5 - 0.8 Nm)
In case of optional increased safety terminal box, securely screw the transmitter onto the e-box using 4 screws with a tightening torque of 8 Nm.
24 - 12 0.2 - 2.5
2
24 Dräger Polytron 5200 / Dräger Polytron 53X0
Page 25
8.7 Physical specifications
www.norrscope.com
Enclosure NEMA 4 x, IP 65/66/67
Size L x W x D approx.
Weight approx.
without e-box with e-box
without e-box, aluminum without e-box, stainless steel 316 with e-box, aluminum with e-box, stainless steel 316
7” x 5.8” x 5.1” (180 x 150 x 130 mm)
7.3” x 7.1” x 7.4” (185 x 180 x x190 mm)
3.9 lbs (1.8 kg)
8.0 lbs. (3.6 kg)
6.9 lbs. (3.1 kg)
11.0 lbs. (5.0 kg)
8.8 Environmental parameters
Technical data
Pressure
Humidity 0% to 100% RH, non-condensing
Temperature Sensor without relays with relays
DD / LC
DSIR / PIR 3000
1)
not yet available
20.7 to 38.4 in. of hg (700 to 1300 mbar)
-40 to 176 °F (-40 to 80 °C)
-40 to 149 °F
1)
(-40 to 65 °C)
-40 to 158 °F (-40 to 70 °C)
-40 to 149 °F (-40 to 65 °C)
8.9 Ambient influences
See sensor data sheets.
Dräger Polytron 5200 / Dräger Polytron 53X0 25
Page 26
Order list
www.norrscope.com
9 Order list
9.1 Detectors
9.1.1 Transmitter
Description
Polytron 5200 DD d A 4544150 Polytron 5200 DD d A Relay 4544151 Polytron 5200 DD d S 4544152 Polytron 5200 DD d S Relay 4544153 Polytron 5200 DD e A 4544154 Polytron 5200 DD e A Relay 4544155 Polytron 5200 Remote DD e A 4544156 Polytron 5200 Remote DD e A Relay 4544157 Polytron 5200 DD e S 4544158 Polytron 5200 DD e S Relay 4544159 Polytron 5200 Remote DD e S 4544160 Polytron 5200 Remote DD e S Relay 4544161
Description
Polytron 5200 LC d A 4544130 Polytron 5200 LC d A Relay 4544131 Polytron 5200 LC d S 4544132 Polytron 5200 LC d S Relay 4544133 Polytron 5200 LC e A 4544134 Polytron 5200 LC e A Relay 4544135 Polytron 5200 Remote LC e A 4544136 Polytron 5200 Remote LC e A Relay 4544137 Polytron 5200 LC e S 4544138 Polytron 5200 LC e S Relay 4544139 Polytron 5200 Remote LC e S 4544140 Polytron 5200 Remote LC e S Relay 4544141
Part Number
Part Number
Description
Part Number
Polytron 5310 d A 4544370 Polytron 5310 d A Relay 4544371 Polytron 5310 d S 4544372 Polytron 5310 d S Relay 4544373 Polytron 5310 e A 4544374 Polytron 5310 e A Relay 4544375 Polytron 5310 Remote e A 4544376 Polytron 5310 Remote e A Relay 4544377 Polytron 5310 e S 4544378 Polytron 5310 e S Relay 4544379 Polytron 5310 Remote e S 4544380 Polytron 5310 Remote e S Relay 4544381
Description
Polytron 5300 d A
Polytron 5300 d A Relay
Polytron 5300 d S
Polytron 5300 d S Relay
Polytron 5300 e A
Polytron 5300 e A Relay
Polytron 5300 Remote e A
Polytron 5300 Remote e A Relay
Polytron 5300 e S
Polytron 5300 e S Relay
Polytron 5300 Remote e S
Polytron 5300 Remote e S Relay
1) not yet available
Part Number
4544170
4544171
4544172
4544173
4544174
4544175
4544176
4544177
4544178
4544179
4544180
4544181
1
1
1
1
1
1
1
1
1
1
1
1
26 Dräger Polytron 5200 / Dräger Polytron 53X0
Page 27
Order list
www.norrscope.com
9.2 Separate sensing head (not included) when ordering
9.2.1 Polytron 5200 DD remote e
Description
Sensing Head Polytron SE Ex PR M1 DD 6812711 Sensing Head Polytron SE Ex PR M2 DD 6812710 Sensing Head Polytron SE Ex PR M3 DD 6812718 Sensing Head Polytron SE Ex HT M DD 6812720
9.2.2 Polytron 5200 LC remote e
Description
Sensing Head Polytron SE Ex LC M1 DD 6812722 Sensing Head Polytron SE Ex LC M2 DD 6812721 Sensing Head Polytron SE Ex LC M3 DD 6812719
9.2.3 Polytron 5310 remote e
Description
Sensing Head DrägerSensor IR complete Set e
Sensing Head DrägerSensor IR complete Set e2
Part Number
Part Number
Part Number
68 11165
6811265
9.4 Accessories
Description
D Junction Box Aluminum 4544099 D Junction Box Stainless Steel 316 4544098 Duct Mount Kit Polytron 5000 for DD, LC,
DSIR, PIR 3000 Pipe Mount Kit 4544198 IRDA Interface 4544197 Magnetic Wand with Key Chain 4544101
9.4.1 Accessories DD, LC
Description
Splash Guard 6812510 Calibration Adapter PE 4509314 Calibration Adapter PE, Europe 6806978 Calibration Adapter Viton 6810536 Process Adapter (stainless steel, with
lock nut M30 x 1.5) for DD Process Adapter (stainless steel, with
lock nut M36 x 1.5) for LC Calibration Chamber 6802206
Part Number
6812725
Part Number
6812470
6812465
9.2.4 Polytron 5300 remote e
Description
Sensing Head Dräger PIR 3000 complete Set e
Sensing Head Dräger PIR 3000 complete Set e2
1) not yet available
Part Number
6811160
6811270
9.3 Replacement sensors (all versions except remote e)
Description
DrägerSensor Ex DD NPT 6812380 DrägerSensor Ex LC NPT 6810675 DrägerSensor IR NPT 6811901 Dräger PIR 3000 NPT 6811080
Part Number
9.4.2 Accessories DSIR, PIR 3000
Description
1
1
Splash Guard 6810796 Calibration Adapter 6810859 Process Adapter 6811330 Protection Labyrinth 68 11135
Part Number
9.5 Spare parts
Description
Bezel Polytron 5200 / 53X0 / 57X0 4544183 PCB Main Polytron 5200 / 53X0 4544186 PCB Relay 4544297 PCB e-box 6812839 Feed-through 3 Wires 4544182 Feed-through 9 Wires 4544169 Feed-through 14 Wires 4544168 Hardware Kit Enclosure 4544167 Hardware Kit e-box 6812838 Cable Gland e-box 6812868
Part Number
Dräger Polytron 5200 / Dräger Polytron 53X0 27
Page 28
www.norrscope.com
Page 29
www.norrscope.com
Page 30
www.norrscope.com
Page 31
www.norrscope.com