Drive Medical Design 13605, 13606 User Manual

tamper proof chair and bed patient alarm
item # 13605 item # 13606
Directions For Application, Use and Care
Application
The Drive Sensor Basic Alarm is used with a sensor pad, lap cushion or seat belt to alert sta that a resident is about to wander. When sensor is activated an 85dB or 100dB alarm tone sounds notifying caregiver. User can choose 0 or 2 second delay seing. LED indicates low baery. Alarm features convenient ON/OFF switch. 9V baery included.
Installation
1. Insert one 9V baery into baery compartment of the alarm. To open compartment, remove screw and slide compartment door open.
2. With baery door open, select delay seing for alarm. Delay can be set to respond immediately (zero second delay) or to sound aer weight has been removed from sensor pad for a full 2 seconds. When used with sensor pads, 2-second delay is particularly helpful with active patients who produce false alarms due to frequent weight shis. Zero second delay seing is recommended for use with lap cushion and seat belt (Figure 1).
3. Also located in baery compartment is the tone selector switch. Choose from either ALARM tone or CHIME tone (Figure 1).
4. If using mesh bag to aach alarm to chair or bed, place alarm in bag and close zipper. Alarm cord should exit bag at zipper opening. Use strap on
bag to hang unit from push handles of wheelchair, or secure to chair or
bed using hook and loop straps. If using mounting bracket to aach alarm box to chair or bed, remove belt clip from back of alarm unit. Peel backing from pressure sensitive hook (included), and aach to back of alarm box. Press rmly to aach alarm to mounting bracket.
5. BED: Mount alarm box toward foot of bed. Route cord toward footboard of bed. Keep cord clear of siderails and all other moving parts. Secure excess cord for patient/sta safety.
6. CHAIR: Mount alarm box out of patient’s reach. Route cord toward back of chair. Keep cord clear of wheels and all other moving parts. Secure excess cord for patient/sta safety (Figure 2).
figure 1
Tone Selector
CHIME or ALARM
Power Switch
Off, Low 85dB or Hi 100dB
Delay Selector 0 sec or 2 sec
Use
1. Move volume switch to Low or Hi (85 dB or 100 dB), depending on required alarm volume for environment (Figure 1).
2. Select alarm tone, alarm or chime. Select delay seing, 0-seconds or 2-seconds. Switches are located inside baery compartment (Figure 1).
3. Plug sensor pad, lap cushion or seat belt into telephone-type jack on top of alarm box. Inspect wires for damage before connecting. Replace if damage to wiring or device is present. Once initial pressure or contact
is detected by alarm unit, unit will beep once to alert caregiver that unit
is active.
4. If resident leaves the chair or bed, or unbuckles the seat belt, alarm will sound and LED will ash. Unit will sound continuously until patient returns to bed or chair, or power switch is moved to the OFF position.
For lap cushion, unit will sound continuously until activation strap is
properly re-secured or power is turned o. Seat belt will continue to sound until buckled or power switch is turned OFF.
Note: Baery power is low if LED ashes continuously and sounds a single beep every 30 seconds. Replace baery immediately. If alarm tone decreases in volume, replace baery immediately.
locate mounting bracket out of patient’s reach, low on back of wheelchair
figure 2
Testing the System
1. Test pad by applying pressure to pad for 3 seconds and releasing. Alarm will sound when pressure is released. Test lap cushion by connecting activation strap and releasing. Alarm will sound when strap is released. Test seat belt by fastening buckle and releasing. Alarm will sound when belt is unbuckled. Note: If using alarm unit with optional delay seings, pressure must be removed from pad for entire delay cycle before alarm will sound.
2. Always test alarm unit and pad, cushion or seat belt together before each use. Never use if alarm fails to sound when tested. Always retest if alarm unit is dropped.
3. Inspect pad, lap cushion or seat belt wires, and all connections for damage. Do not use if damaged. Replace immediately.
Caution
• This device is not a substitute for routine visual monitoring of patients. It will not prevent falls. Sta must make frequent routine visual inspections to ensure sensor pad, lap cushion or seat belt is properly placed and alarm unit is powered on.
• Use of alarm unit may not be suitable for all “High Risk” patients. Other safety measures may still be necessary to ensure patient safety.
Optional Nurse Call Feature for Bedside Use
Plug one end of Nurse Call Cable (sold separately) into top of alarm unit and opposite end of cable into a compatible nurse call plug located at resident’s bedside. When alarm sounds, a signal will activate nurse call station just as it would if emergency buon were pressed.
Care
• Wipe clean using standard cleaning disinfectants.
• Do not immerse in uids.
• Check function before each use and if alarm unit is dropped. Do not use equipment that has been damaged.
Precautions
• Avoid contact with water: remove alarm unit from wheelchair before washing.
• Alarm emits extremely loud signal. Do not point unit directly at yourself or anyone else unless you need immediate assistance. Do not place unit near resident’s ear as possible hearing loss can occur.
Maintenance
Check alarm before each use to ensure alarm function. If alarm fails to sound, replace baery. If alarm fails to work with a new baery, inspect and clean baery terminals.
6 Month Limited Warranty
Your Drive brand product is warranted to be free of defects in materials and workmanship for 6 months of the original consumer
purchaser. This device was built to exacting standards and carefully inspected prior to shipment. This 6 Month Limited Warranty is an expression of our condence in the materials and workmanship of our products and our assurance to the consumer of years of dependable service. In the event of a defect covered by this warranty, we will, at our option, repair or replace the device. This warranty does not cover device failure due to owner misuse or negligence, or normal wear and tear. The warranty does not extend to non-durable components, such as rubber accessories, casters, and grips, which are subject to normal wear and need periodic replacement. If you have a question about your Drive device or this warranty, please contact an authorized Drive dealer.
item # 13605
www.drivemedical.com
item # 13606
© 2006 Medical Depot, Inc. All rights reserved.
Drive is a trademark of Medical Depot, Inc.
Port Washington N.Y. 11050 USA Made in China
Loading...