This manual must be given to the Owner/User of this bed and
should be read carefully before putting this product into use.
Keep manual in an accessible location for future reference.
15561/15571 Series Manual REV1.1.29.18
TABLE OF CONTENTS
Standard Symbols & Warning Labels ................................................................................................ 1-3
Click here to view detailed product information and pricing. Call 800-397-5899 to order today!
STANDARD SYMBOLS USED IN THIS MANUAL
This manual includes important information about the safety of personnel and equipment. As you read through this
manual be aware of the symbols and their meanings.
DANGER
!
Information that appears under the DANGER symbol concerns the protection of personnel from direct and
pending hazards that, if not avoided, will result in immediate, serious personal injury or death in addition to
damage of the equipment.
DANGER SHOCK HAZARD
Information that appears under the DANGER SHOCK HAZARD symbol concerns the protection of personnel
from possible hazards related to electrical shock.
FIRE HAZARD
Information that appears under the FIRE HAZARD symbol concerns the protection of personnel from possible
hazards related to re and ammability.
!
WARNING
Information that appears under the WARNING symbol concerns the protection of personnel from possible
hazards that can result in injury or death in addition to damage of the equipment.
CAUTION
!
Information that appears under the CAUTION symbol concerns the protection of personnel from possible
hazards that may result in minor injury or damage of the equipment.
CRUSH HAZARD
Information that appears under the CRUSH HAZARD symbol concerns the protection of personnel from
possible hazards related to reduce clearances.
NOTE:
Information that appears with the NOTE text gives added information, which helps in understanding the item
being described.
1
PRODUCT SYMBOLS
SymbolsDescriptionSymbolsDescription
Type BF applied part equipment.
IEC 60601-2-52: The applied part includes all
parts of the medical bed that are within reach
of the patient, even if they are underneath
the mattress support surface. Applied parts
list: Mattress, Bed frame, Side Rail, Head
Board and Foot Board, Mattress retainer.
According to IEC 60529. Rating for dust-
IPX4
protection and identied as equipment that is
protected against splashing water.
Class II equipment
The owner’s manual needs to be read before
operating the bed.
A product bearing the IEC
listed Mark is determined
to have met the minimum
requirements of prescribed
product safety standards.
Maximum patient weight and
safe working load
Do not remove motor cover or repair yourself. Contact
your local Provider or Dealer for technical assistance.
Do not stack more than 5 cartons high
5
104oF
40oC
Do not step
Keep Dry
Do not exceed 104 degrees F
Do not leave in direct sun light for extended period of
time
Handle with care
2
WARNING LABELS
3
INTENDED USE
The Drive #15561/15571 series beds are intended for use within a Homecare environment, Application Environment 4
(i.e.: care provided in a domestic area where ME Equipment is used to alleviate or compensate for an injury, disability or
disease).
This bed is NOT intended for patient/resident transport. Wheels are provided only to allow movement within the patient/
resident room for cleaning or patient/resident access.
CONTRA-INDICATIONS
This bed should not be operated (control the movements of the bed through the control interfaces)by persons who do
not have the cognitive skills to understand the information in the owner’s manual or cannot understand and perform the
proper operation of the bed. This bed should not be used or operated by children.
GENERAL SAFETY
NOTE: In this manual, the terms patient and resident refer to a resident in a homecare environment.
!
WARNING
• NEVER allow anyone under the bed at any time.
• DO NOT let any body parts protrude over the side or between parts, especially when the bed is being operated.
• This bed should not be placed in an oxygen enriched environment.
• Unplug bed or stop pressing pendant button when bed does not function as expected.
• DO NOT operate this product without rst reading and understanding this user manual. Damage or injury may result from improper use of this product.
• Possible injury or Death may occur if accessories are not provided by Drive DeVilbiss Healthcare. Please contact
Drive DeVilbiss Healthcare for accessories that are compatible with this bed.
• Possible injury or Death may occur if replacement parts are not provided by Drive DeVilbiss Healthcare.
Please contact Drive DeVilbiss Healthcare for replacement parts compatible with this bed.
WARNING
!
• Power cord must be plugged into appropriate wall outlet, ensuring it can be unplugged easily in case of an emergency.
• Never permit more than one (1) person in/on the bed at any time. Never exceed the weight capacity of this bed. The
weight capacity of this bed is 450lbs (205kg) including accessories and options (not all accessories and options can
be used at the same time). Once resident’s weight is known (X), subtract that from 450 lb total weight capacity (Y) and
remainder is allowable weight for mattress, accessories and optional equipment (Z). (Z = Y – X)
Assumes: Max Patient/resident Weight 350 lbs (158.76 kg)
Full or Half Rails (avg/pair) 25 lbs (11.34 kg)
Mattress weight (avg) 25 lbs (11.34 kg)
Trapeze 26 lbs (11.79 kg)
• The medical bed should be left in its lowest position except when care is being provided in order to reduce risk of injury
due to falls.
• The bed MUST be connected to an appropriate power source. Unless equipped with battery backup, this bed will not
operate when not plugged in. Mains cable and proper function must be checked regularly.
• Possible Injury or Death may occur if bed is pushed over abrupt thresholds while bed is occupied. This bed is not intended for patient/resident transport. Please use an approved resident transport device when moving a resident.
4
DANGER SHOCK HAZARD
• DO NOT plug anything into the control box of bed (i.e. pendants and actuators) while mains cable (power cord) is
plugged into the wall outlet.
• Any cords or tubing used on or with this bed MUST be routed and secured properly to ensure that they do NOT
become entangled, kinked or severed during normal operation of the bed.
• DO NOT roll the bed over any power or pendant cords.
• DO NOT open any actuators, control boxes, or pendants. Service and repair must only to be performed by authorized
service personnel. If unauthorized service is performed on any components the warranty is void.
• Inappropriate handling of power supply cords can create a shock hazard.
• When routing cables from other equipment in the medical bed, precautions shall be taken to avoid squeezing those
between parts of the medical bed.
• DO NOT allow the cord, electrical outlets, and electrical control box or hand pendant to become wet or submerged.
• DO NOT operate the bed if any electrical component such as the power cord, electrical outlet, connections, motor/
actuator or mechanical component has malfunctioned or has been damaged in any way.
• The bed MUST be connected to an appropriate power source. Unless equipped with the optional battery backup,
this bed will not operate when not plugged in. Mains cable and proper function must be checked regularly.
• Possible injury or Death may occur due to pendant cord being a source for entangling patient/resident. Patients/
residents with decreased mental acuity should NOT have access to pendant.
FIRE HAZARD
• DO NOT use near open ame or explosive gases.
• Possible Fire Hazard if used with nasal mask in 1/2 bed tent 02 administering equipment. If 02 tent is being used it
should not fall below the mattress deck. The pendant should not be placed in an oxygen enriched environment such
as an 02 tent.
• This bed should not be placed in an oxygen enriched environment.
CAUTION
!
• This medical bed is not intended to work with patient lifts other than those specied by Drive DeVilbiss Healthcare as being compatible.
• Do NOT use side rails as handles for moving the bed.
• Keep all moving parts, including the mattress deck (sleep surface), main frame, all drive actuators and moving
components free of obstructions.
• Ensure that the individual using this bed is properly positioned, particularly when the bed is being operated.
• Body weight should be evenly distributed over the sleeping surface of the bed. Avoid situations where entire body
weight is on a raised head or foot surface. This includes while assisting the resident in repositioning or transferring in
or out of bed.
• Supervision is required when this product is operated by or near children or people with disabilities.
• Do not modify this bed without authorization of the manufacturer. If unauthorized service is performed on any
components the warranty is void.
• This bed frame complies with EMC requirements of IEC 60601-1-2. Radio transmitting equipment, cell phones or
similar electronic devices, used in proximity of the bed, may affect the beds performance.
5
CRUSH HAZARD
!
• Possible crush hazard exists between the bed deck and frame. Keep hands and feet out from under the bed deck at
all times while operating the bed.
• Possible Injury may occur when lowering or raising the bed. Keep feet and hands clear of bed frame.
• Do not put your hands/ngers into the opening of head or thigh deck edges, crush point may cause injury to your
hands/ngers.
ENTRAPMENT WARNING
!
• The entrapment issues can still arise when components and accessories are not properly installed.
• If a resident/patient’s mental or physical condition could lead to resident/patient entrapment, the mattress deck
(sleep surface) should be left in the at position when unattended. Failure to do so could result in injury or death.
• Incompatible mattress and side rails can create hazards. Only compatible Drive DeVilbiss Healthcare side rails may
be used on this bed. Make sure mattress is the correct size for bed frame and the Side rails are secured to frame to
decrease the risk of entrapment.
• If bed frames have been serviced or any other adjustments have been made, you must ensure all parts are securely
back in place before operating the bed.
WARNING
!
Incompatible mattress and side bars/rails can create hazards. Read instructions for use.
MATTRESS SPECIFICATIONS
WARNING
!
Possible ENTRAPMENT Hazard may occur if you do not use the recommended specication mattress.
Resident entrapment may occur leading to injury or death.
• A mattress may not be included with this bed. It is recommended that a standard 36” wide mattress that is made
to t an 80”or 84” (extension) length bed frame is used, such as a Drive DeVilbiss Healthcare standard inner
spring, foam, ber, or pressure redistribution mattress. Mattress height must be a minimum of 5-1/2 inches and
a maximum of 7 inches.
• Also available are Drive DeVilbiss Healthcare’s assortment of mattress overlays and Low Air Loss otation mattress systems.
6
ELECTROMAGNETIC EMMISSION AND IMMUNITY
This MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided in the table below.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT
WARNING
!
The use of ACCESSORIES, transducers and cables other than those specied, with the exception of transducers and
cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replacement parts for internal components, may
result in increased EMISSION or decreased IMMUNITY of the EQUIPMENT or SYSTEM.
Below cables information are provided for EMC reference
CableMax. cable length,
Shielded/unshielded
AC Power Line3.8mUnshielded1 SetAC Power
Signal Line
(from remote control part
to be main body)
WARNING
!
Important information regarding Electro Magnetic Compatibility (EMC)
SEMI ELECTRIC BED WITH MANUAL LEG ADJUST requires special precautions regarding EMC when put into service
according to the EMC information provided in the user manual; SEMI ELECTRIC BED WITH MANUAL LEG ADJUST
conforms to this IEC 60601-1-2:2014 standard for both immunity and emissions. Nevertheless, special precautions need
to be observed:
SEMI ELECTRIC BED WITH MANUAL LEG ADJUST with no ESSENTIAL PERFORMANCE is intended for use in a
professional healthcare facility environment and homecare environment.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the SEMI ELECTRIC BED WITH MANUAL LEG
ADJUST, including cables specied by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.”
3.2mUnshielded1 SetDC Power
NumberCable classication
The use of accessories and cables other than those specied by Drive DeVilbiss Healthcare, with the exception of
accessories and cables sold by Drive DeVilbiss Healthcare for use with the SEMI ELECTRIC BED WITH MANUAL LEG
ADJUST as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of
the SEMI ELECTRIC BED WITH MANUAL LEG ADJUST.
WARNING: Use of this equipment SEMI ELECTRIC BED WITH MANUAL LEG ADJUST adjacent to or stacked with other
equipment should be avoided because it could result in improper operation. ”
7
EMI COMPLIANCE TABLE
Table 1 - Emisson
PhenomenonComplianceElectromagnetic environment
RF emissionsCISPR 11
Group 1, Class B
Table 2 - Enclosure Port
PhenomenonBasic EMC standardImmunity test levels
Electrostatic
Discharge
Radiated RF EM eldIEC 61000-4-310V/m
Proximity elds from RF
wireless communications
equipment
Rated power frequency
magnetic elds
Table 3 - Proximity elds from RF wireless communications equipment
IEC 61000-4-2±8 kV contact
IEC 61000-4-3Refer to table 3
IEC 61000-4-830A/m
Professional healthcare facility environment and
homecare environment
Professional healthcare facility environment and
homecare environment
±2kV, ±4kV, ±8kV, ±15kV air
80MHz-2.7GHz
80% AM at 1kHz
50Hz or 60Hz
Test frequency
(HHz)
385380-390Pulse modulation 18Hz, 27V/m
450430-470FM, ±5kHz deviation, 1kHz sine, 28V/m
710704-787Pulse modulation 217Hz, 9V/m
745
780
810800-960Pulse modulation 18Hz, 28V/m
870
930
17201700-1990Pulse modulation 217Hz, 28V/m
1845
1970
24502400-2570Pulse modulation 217Hz, 28V/m
52405100-5800Pulse modulation 217Hz, 9V/m
5500
5785
Band
(MHz)
Immunity test levels
Professional healthcare facility environment and
homecare environment
8
Table 4 - Input AC power Port
PhenomenonBasic EMC standardImmunity test levels
Professional healthcare facility environment and
homecare environment
Electrical fast
transients/burst
Surges
Line-to-line
Conducted disturbances
induced by RF elds
Voltage dipsIEC 61000-4-11
Voltage interruptionsIEC 61000-4-11
IEC 61000-4-4±2 kV
IEC 61000-4-5±0.5 kV, ±1 kV
IEC 61000-4-63V, 0.15MHz-80MHz
100kHz repetition frequency
6V in ISM bands and amateur radio bands
between 0.15MHz and 80MHz
80%AM at 1kHz
0% UT; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and 315º
0% UT; 1 cycle and
70% UT; 25/30 cycles
Single phase: at 0º
0% UT; 250/300 cycles
Table 5 – Signal input/output parts Port
PhenomenonBasic EMC standardImmunity test levels
Professional healthcare facility environment and
homecare environment
Electrical fast transients/
burst
Conducted disturbances
induced by RF elds
IEC 61000-4-4±1 kV
100kHz repetition frequency
IEC 61000-4-63V, 0.15MHz-80MHz
6V in ISM bands and amateur radio bands
between 0.15MHz and 80MHz
80%AM at 1kHz
9
ELECTRICAL SAFETY WARNINGS
!
A safety feature of this product includes protection against overheating caused by excessive or extended periods of
operation. Depending on the duration, this includes multiple or repeated adjustments or the use of multiple functions at
one time. To ensure trouble free operation, always allow a slight pause between multiple adjustments and avoid pressing
more than one function button at a time. If thermal protection activation should occur, the bed will not respond to pendant
commands. Given this situation, release the pendant button and allow the bed unit to sit for several minutes. This will allow
the protection function time to reset and restore bed function. Depending on severity of the initial overheating, this could
take up to 30 minutes.
Allow a slight pause between adjustments and avoid pressing multiple buttons at the same time. If pendant buttons are
depressed too rapidly or multiple buttons are pressed at the same time, the desired feature may not activate. Simply
release the pendant button, permit a slight pause and then activate the next operation.
Always unplug the bed from the electrical outlet before cleaning.
DO NOT open assemblies such as the motors, motor housing or pendant. No user serviceable parts are inside. Only
qualied technicians are permitted to repair these parts. If unqualied individuals perform any work on these beds, the
warranty is void.
DO NOT place pendant under or between objects. This may unintentionally press the buttons and may cause injury or
damage.
Electronic equipment may be inuenced by Radio Frequency Interference (RFI). Caution should be exercised with regard to
the use of portable communications equipment in the area around such equipment. If RFI causes erratic behavior, unplug
the electric bed IMMEDIATELY. Leave unplugged while transmission is in progress.
Ensure all cables and cords are routed such that they will not become entangled or pinched. Otherwise damage or injury
may result.
If a button on the pendant sticks or does not release and/or one of the functions will not stop moving. Unplug the power
cord from the electrical supply outlet and take the bed out of service for repair.
If a liquid is spilled in or around the electric bed, unplug the electric bed before cleaning. Clean up the spill and allow the
electric bed or the area around the electric bed to dry thoroughly before using the electric controls again.
Inspect the covering of the hand pendant panel to assure that the covering is not cracked or damaged.
Keep all electrical cords away from heated or hot surfaces.
Keep the cord away from heated surfaces.
NEVER operate this bed if it has a damaged cord or plug, if it is not working properly, if it has been dropped, damaged, or
dropped into water. Return the bed to a service center for examination and repair.
On Full Electric beds, the Hi/Lo crank MUST be removed before the bed is used. Failure to remove the crank may cause
damage or personal injury.
Refer servicing to qualied personnel only. Grounding reliability depends upon a properly grounded wall outlet.
The pendant and power cords must be routed and secured properly to ensure that the cords DO NOT become entangled,
pinched and/or severed during operation of the electric bed.
Unplug from the outlet before servicing. Unplug the power cord from its power source before performing any maintenance
on the manual/electric bed.
Use caution when disconnecting the pendant. DO NOT press pendant buttons.
When bed is not to be used for an extended period, unplug electric bed from the wall outlet.
When using an extension cord, use an extension cord having at least 16 AWG (American Wire Gauge) wire and the same
or higher electrical rating as the device being connected. Use of improper extension cord could result in a risk of re and
electric shock.
When using nasal or masked type oxygen administering equipment, the oxygen or air tubing MUST be routed and secured
properly to ensure that the tubing does not become entangled and/or severed during normal operation of the bed.
DO NOT use near explosive gases.
Possible re hazard when used with oxygen administering equipment other than nasal or masked type.
Use masked or nasal type oxygen administering equipment only in conjunction with this bed. The use of ANY other type of
oxygen administering equipment can result in a re hazard.
10
TECHNICAL SPECIFICATIONS
Overall Length87.2”
Overall Width w/o Rails35.63”
Overall Width w/Rails37”
Length & Width of Mattress Deck79.73” x 34.65”
Maximum Weight Capacity450lbs. includes all accessories
Head Deck Angle0°to 68°
Foot Deck Angle0°to 30°
Height Range
Manual Bed Ends11” to 20”
Crank Bed Ends15.1” to 22.5”
Under Bed Clearance
Manual Bed Ends4.4”
Crank Bed Ends8.3”
Total Product Weight
Head Section41lbs
Foot Section41lbs
Manual Bed Ends44lbs
Crank Bed Ends52lbs
Noise Level≤65dB(A)
Electrical Requirements120VAC,60Hz
Rated Current Input1.8A max
Fuse Specication2A 250V
Duty CycleT_On 2 Min / T_Off 18 Min
Battery Backup: Handset (1) & Motor (2)9V
Relative Humidity15% to 93% relative humidity, non-condensing
Recommended Storage &Shipping Temperature-22°F to +158°F
Recommended Operating Temperature41°F to 104°F
Atmospheric Pressure Range700hPa to 1060hPa
OPERATING TEMPERATURE
If bed has been stored in an area where the temperature is above or below the Recommended Operating Temperature
range of 41°F to 104°F. You must wait at least 30 minutes for the bed temperature to adjust to the Recommended
Operating Temperature range of 41°F to 104°F before use.
BED SERIAL NUMBERS
When ordering parts or when contacting Drive DeVilbiss Healthcare Customer Service Department, please include bed’s
model and serial numbers, found on the identication labels as below. The identication labels are located under the sleep
deck on the either side of the bed frame.
COMPLIANCE INFORMATION
Matching the correct bed components to meet regulatory specications can be complicated. Drive DeVilbiss Healthcare
offers a wide variety of compliance options and we can assist your facility in selecting components or accessories that are
recommended for the specic bed model.