Before using this product, carefully read the associated Instructions for Use. This
document does not replace the Instructions for Use.
Definitions of alert icons
The following alert icons are used in this document to provide and highlight areas of
the associated text that require a greater awareness by the user. A definition of the
meaning of each icon is as follows:
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death
or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death
or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in
physical injury, or damage to the product or environment. It may also be used to alert
against unsafe practices.
NOTICE
Indicates additional information on how to use the product.
Intended Use
Portable gas detection instrument for the continuous monitoring of the concentration of
several gases in the ambient air within the working area and in explosion-hazard areas.
X-am 5000, depending on the device type and configuration of DrägerSensors:
independent measurement of one up to five gases.
4
Tests and Approvals
Tests and Approvals
Copies of the name plate and the declaration of conformity are provided in the
enclosed supplementary documentation (order no. 90 33 890).
The BVS 10 ATEX E 080 X technical suitability test is based on the adjustment with the
target gas.
Do not stick anything on the name plate on the gas detector.
The technical suitability tests are valid for the X-am 5000 gas detector and the
calibration cradle. The explosion-protection approvals are only valid for the X-am 5000
gas detector; the calibration cradle must not be used in the Ex zone.
5
Tests and Approvals
Intended operating area and operating conditions
Areas subject to explosion hazards, classified by zones
The instrument is intended for the use in explosion-hazard areas of Zone 0, Zone 1 or
Zone 2 or in mines at risk due to fire damp. It is intended for use within a temperature
range of -20 °C to +50 °C, and for areas in which gases of explosion groups IIA, IIB or
IIC and temperature class T3 or T4 (depending on the batteries and rechargeable
battery) may be present. If used in mines, the instrument is only to be used in areas
known to have a low risk of mechanical impact.
Areas subject to explosion hazards, classified by divisions.
The instrument is intended for use in explosion-hazard areas according to Class I&II,
Div. 1 or Div. 2 within a temperature range of -20 °C to +50 °C, and for areas where
gases or dusts of groups A, B, C, D, E, F, G and temperature class T3 or T4 may be
present (depending on the rechargeable battery and batteries).
WARNING:
CSA requirement: The sensitivity must be tested on a daily basis before first use with
a known concentration of the gas to be measured in accordance with 25 to 50 % of
the concentration limit value. The accuracy must be 0 to +20 % of the actual value.
The accuracy can be corrected via calibration.
Safety instructions
WARNING
To reduce the danger of explosion, do not mix new batteries with old batteries and do
not mix batteries made by different manufacturers.
WARNING
Always disconnect the device from the power pack before carrying out any
maintenance operations.
WARNING
Substitution of components may impair intrinsic safety.
CAUTION
Not tested in an oxygen-enriched atmosphere (>21 % O
WARNING
High off-scale readings may indicate an explosive concentration.
6
).
2
Tests and Approvals
CAUTION
Use only supply units ABT 0100 (83 22 237), HBT 0000 (83 18 704) or HBT 0100
(83 22 244). Check the supply unit for approved batteries and applicable temperature
class.
Note the following for CSA (Canadian Standards Association) applications:
For the CSA approval only the functions of the device component that is used to
measure flammable gases are tested. The device is not approved by CSA for use in
mining.
WARNING:
Before daily use, test the sensitivity with a known concentration of the applicable gas
corresponding to 25 to 50% of the maximum concentration. The accuracy must be
within a range of 0 to +20% of the actual value. Perform a calibration to correct the
accuracy if necessary.
Only applies for class II certification. CSA standard C22.2 no. 152 does not contain
any requirements for class II hazard areas. Therefore this device was not tested for
class II by the CSA. The sensor may become blocked and not measure the gas
correctly or not warn the user that gas measurement is not possible.
WARNING
CSA requirement: Measured values over the full scale value are displayed as an explosive concentration.
NOTICE
CSA requirement: a measurement suitability test must be conducted for flammable
gases.
7
What is What
00223999_02.eps
0
1
2
6
5
4
3
2
X-am 5000
7
OK
00323999_02.eps
2
1
4
6
3
5
456
CH4
O
2
CO
%LEL
Vol %
ppm
123
CH4
%LEL
O
2
Vol %
CO
ppm
H
2
S
ppm
NH
3
ppm
What is What
Front panel
1 Gas entry
2 Alarm LED
3 Buzzer
4 key
5 key
6 Display
7 Tool for replacing
the sensor
Rear panel
1 IR interface
2 Fastening clip
3 Type plate
4 Charging contacts
5 Power pack
6 Serial no.
Display
for 5 measuring channels
only:
1 Measured gas
display with unit
2 Measured value
display
3 Special symbols
other:
4 Measured gas
display
5 Measured gas
display with unit
6 Special symbols
The following only shows the device version with 5 measuring channels.
8
Special symbols
Fault message, refer to page 17
Warning message, refer to page 17
The peak value display for all measuring gases,
refer to page 17
The exposure evaluation display (TWA) for measuring gases, e.g., H
CO,
refer to page 17
The exposure evaluation display (STEL) for measuring gases, e.g,. H
CO,
refer to page 17
The device is set to function test with gas (bump test),
refer to page 40
The device is set to the fresh air calibration function,
refer to page 46
The device is set to the 1-button calibration function,
refer to page 47
The device is set to the single gas calibration function,
refer to page 50
The function for password entry is active,
refer to page 20
Battery / rechargeable battery 100 % full
Battery / rechargeable battery 2/3 full
Battery / rechargeable battery 1/3 full
Battery / rechargeable battery empty
What is What
S and
2
S and
2
Marking offset channels:
FeatureReading on the display
ToxicTwinsHCN+
compensationCO+
CO H
2
9
Configuration
Configuration
Standard gas configuration
DrägerSensorMeasuring
range
1)
Alarm A1
1)
Alarm A2
1)
CatEx 125 PR
0 to 10020YesNo40NoYes
setpoint
can be
acknowledged
self-latching
setpoint
can be
acknowledged
[%LEL]
CatEx 125 PR Gas
[%LEL]
XXS O2 [vol. %]0 to 2519
XXS O2 100 [vol. %] 0 to 10018.5
XXS O2 LS / CO-LC
[vol. %], [ppm]
XXS O2 / H2S-LC
[Vol%], [ppm]
0 to 10020YesNo40NoYes
2)
NoYe s23NoYes
2)
NoYe s24NoYes
0 to 25 O
0 to 2000 CO
0 to 25 O
0 to 100 H2S
2
2
19 O
30 CO
19 O
5H2SNoYes
No
2
Yes
2
YesNo23 O
60 CO
YesNo23 O
10 H2SNoNo
2
2
NoNoYes
Yes
Yes
Yes
XXS CO [ppm]0 to 2,00030YesNo60NoYes
XXS CO LC [ppm]0 to 2,00030Ye sNo60NoYes
XXS CO-LC/H2S-LC
[ppm]
XXS CO-LC/O
[ppm], [vol.-%]
2
0 to 2,000
0 to 100
0 to 2,000 CO
0 to 25 O
2
30
5
30 CO
19 O
Yes
YesNoNo6010
YesNoNo
2
Yes
60 CO
23 O
2
NoNoYes
Yes
NoNoYes
Yes
XXS CO HC [ppm]0 to 10,000600Ye sNo1,200NoYes
XXS CO H2-CP [ppm] 0 to 2.00030YesNo60NoYes
XXS H2 [ppm]0 to 2.000200Ye sNo400NoYes
XXS H2S [ppm]0 to 2005YesNo10NoYes
XXS H2S LC [ppm]0 to 1005YesNo10NoYes
XXS H2S HC [ppm]0 to 1,00010YesNo20NoYes
XXS H2S/CO [ppm]0 to 200 H2S
XXS H2S-LC / CO-LC
[ppm]
0 to 2,000 CO
0 to 100 H2S
0 to 2000 CO
5 H2S
30 COYesNo
5 H2S
30 CO
Yes
YesNoNo
10 H2S
60 CONoYes
10 H2S
60 CONoNo
Yes
Yes
XXS NO [ppm]0 to 20025YesNo50NoYes
XXS NO2 [ppm]0 to 505Ye sNo10NoYes
XXS SO2 [ppm]0 to 1000,5Ye sNo1NoYes
XXS PH3 [ppm]0 to 200.1YesNo0.2NoYes
XXS PH3 HC [ppm]0 to 2,0005Ye sNo10NoYes
XXS HCN [ppm]0 to 501.9Ye sNo3.8NoYes
XXS HCN PC [ppm]0 to 505Ye sNo10NoYes
XXS NH3 [ppm]0 to 30020YesNo40NoYes
XXS CO2 [vol. %]0 to 50.5Ye sNo1NoYes
XXS Cl2 [ppm]0 to 200.5Ye sNo1NoYe s
XXS H2 HC [vol. %]0 to 40.8Ye sNo1.6NoYe s
self-latching
10
Configuration
DrägerSensorMeasuring
range
1)
Alarm A1
setpoint
can be
acknowledged
1)
self-latching
Alarm A2
setpoint
can be
acknowledged
1)
self-latching
XXS OV [ppm]0 to 20010Ye sNo20NoYe s
XXS OV A [ppm]0 to 20010Ye sNo20NoYe s
XXS Odorant [ppm] 0 to 4010YesNo20NoYe s
XXS Amine [ppm]0 to 10010Ye sNo20NoYe s
XXS COCI2 [ppm]0 to 100.1Ye sNo0.2NoYes
XXS O3 [ppm]0 to 100.1Ye sNo0.2NoYes
XXS NO2 LC [ppm]0 to 500,5YesNo1,0NoYes
1) Different settings can be selected to meet customer requirements on delivery. The current setting can
be checked and changed with the Dräger CC Vision software. A version of Dräger CC-Vision suitable
for the Dräger X-am 5000/2000 can be downloaded on the product page of the X-am 5000:
www.draeger.com.
2) In the case of O
A1 is the lower alarm setpoint: an alarm is issued if the value is too low.
2
11
Configuration
Standard device configuration
NOTICE
Only trained personnel are permitted to make changes to the device configuration.
Dräger X-am® 5000
Bump test mode
Fresh air calibration
Operating signal
Capture rangeON
Switch off
LEL factor
2)
2)
(ch4)
2) 4) 5)
STEL
(short-term average)
2) 5) 6)
TWA
(shift average)
Alarm A1
7)
Alarm A1 at O
Alarm A2
1) X-am® is a registered trademark of Dräger.
2) Different settings can be selected to meet customer requirements on delivery. The current setting can
3) A periodic short signal indicates the operating capacity of the instrument. If there is no operating signal,
4) STEL: average value of an exposure over a short period, generally 15 minutes.
5) Interpretation only if the sensor is designed for this.
6) TWA: shift averages are workplace limit values for generally eight hours per day of exposure for five
7) Latching and acknowledgement of alarms A1 and A2 can be configured with the Dräger CC Vision PC
7)
be checked and changed with the Dräger CC Vision software.
correct operation cannot be guaranteed.
days a week during a working life.
software.
Changing the standard configuration: See “Configuring the Device” on page 26.
2)
2) 3)
sensor
2
1)
Extended bump test
2)
ON
ON
allowed
4.4 (vol. %)
(4.4 vol. % corresponds to 100 %LEL)
STEL function - disabled
Average value duration = 15 minutes
TWA function - disabled
Average value duration = 8 hours
can be acknowledged, non-latching, pre-alarm,
rising flank
7)
cannot be acknowledged, latching, like main alarm,
falling flank
cannot be acknowledged, latching, main alarm,
rising flank
WARNING
After a basic initialization has been carried out with the PC software Dräger CC
Vision, individual alarm settings may have been changed.
Selecting or disabling the capture ranges (only applies for the measuring mode):
The capture range is selected in the measuring mode (factory setting) and permanently
disabled in calibration mode.
The CC-Vision PC software can be used to select or disable the capture ranges for the
measuring mode.
12
Operation
OK
Operation
Preparations for operation
Before using the instrument for the first time, insert a charged NiMH T4 power pack
or batteries approved by Dräger, see "Changing the batteries" on page 55.
The instrument is now ready for operation.
WARNING
To reduce the risk of ignition of a flammable or explosive atmosphere, strictly adhere
to the following warning statements:
Only use power pack types ABT 01xx, HBT 00xx or HBT 01xx. See the marking on the
rechargeable battery for permitted rechargeable batteries and the corresponding
temperature class.
Substitution of components may impair intrinsic safety.
Switching on the device
Press and hold the key for approx. 3 seconds until the countdown »3.2.1«
shown in the display has elapsed.
All the display segments, including the visual, audible and vibration alarms, are
activated for a short time.
The software version is displayed.
The device performs a self test.
The next sensor which is next due for calibration is displayed with the days
remaining until the next calibration, e.g., » Ex %LEL CAL 20 «.
The time until the bump test interval elapses is displayed in days, e.g., » bt 123 «.
All A1 and A2 alarm thresholds and »« (TWA)
gases (e. g. H
During the sensor warm-up phase:
The display for the measured value flashes
The special symbol »« is displayed.
No alarms are issued during the warm-up phase.
The red LEDs flash.
The gas detector is ready to measure when the measured values no longer flash
and the red LEDs are no longer illuminated. The special symbol »« may continue to be displayed if corresponding warnings (e.g. not yet ready for calibration)
are active (to view the warnings, see the technical manual).
Press the key to cancel the display of the activation sequence.
OK
S or CO) are displayed consecutively.
2
1)
and »« (STEL)1) for all toxic
1) Only when activated in the instrument configuration. Delivery condition: not activated.
13
Operation
Switching off the device
Press and hold the key and key at the same time until the countdown
»3.2.1« shown in the display has elapsed.
Before the device is switched off, the visual, audible and vibration alarms are activated
for a short time.
OK
Before entering the workplace
WARNING
Before any measurements relevant to safety are made, check the adjustment with a
bump test, adjust if necessary and check all alarm elements. If national regulations
apply, a bump test must be performed according to the national regulations. Faulty
adjustment may result in incorrect measuring results, with possible serious
consequences.
WARNING
In an oxygen enriched atmosphere (>21 vol. % O
be guaranteed; remove instrument from the explosion-hazard area.
CAUTION
The CatEx sensor is intended for measurements of flammable gases and vapours
mixed with air (i.e. O
content ≈ 21 vol.%). Incorrect measured values may be displayed
2
in the case of oxygen deficient or oxygen enriched environments.
NOTICE
If the gas detector is used for offshore applications, a distance of 5 m to a compass
must be complied with.
), the explosion protection cannot
2
Switch on the device. The current measured values are shown in the display.
Observe any warning »« or fault messages »«.
The device can be operated normally. If the warning message does not go out
automatically during operation, the device must be maintained after the end of
use.
The device is not ready to measure and requires maintenance.
If one of these special symbols is displayed, appropriate measures, refer to page 30
to page 33, must be taken.
14
Operation
Check that the gas inlet opening on the device is not covered.
WARNING
Explosion hazard! To reduce the risk of flammable or explosive atmospheres igniting,
it is essential that the warning notices below are observed:
Fractions of catalytic poisons in the measuring gas (e.g. volatile silicon, sulphur,
heavy metal compounds or halogenated hydrocarbon) can damage the Cat Ex
sensor. If the CatEx sensor can no longer be calibrated to the target concentration,
the sensor must be replaced.
In case of measurements in oxygen-deficient atmosphere (<12 vol. % O
CatEx sensor may show incorrect displays; in this case, a reliable measurement
) the
2
with a CatEx sensor is not possible.
In an oxygen enriched atmosphere (>21 vol. % O
cannot be guaranteed; remove instrument from the Ex area.
), the explosion protection
2
High values outside the display area indicate an explosive concentration where
applicable.
15
Operation
During operation
During operation, the measured values for every measured gas are displayed.
In the event of an alarm, the corresponding displays, the visual, audible and vibration
alarms are activated – see chapter “Identifying Alarms”.
If a measuring range is exceeded or a negative drift occurs, the following displays
are shown instead of the measured value display:
»«(concentration too high) or
»«(measuring range not reached) or
»«
If an O2 sensor is fitted and this sensor measures an O2 concentration of below
After the measuring range of the TOX measuring channels has been exceeded
In the event of an alarm, the corresponding displays, including the visual, audible and
vibration alarms, are activated – see “Identifying Alarms” on page 21.
Special states in which there is no measuring operation (quick menu, calibration
menu, warm-up of sensors, password input) are indicated by a visual signal (slow
flashing of the alarm LED ).
(blocking alarm).
12 vol. %, an error is indicated with » « on the ex-channel instead of a
measured value if the measured value is below the pre-alarm threshold (not with the
setting CH4 with measuring range >100 %LEL).
temporarily (up to one hour), checking the measuring channels is not necessary.
NOTICE
If the measuring range is exceeded significantly on the CatEx channel (very high
concentration of flammable materials), a blocking alarm is triggered. This CatEx
blocking alarm can be acknowledged manually by switching the device off and back on
again in fresh air.
In configuration setting CH
triggered as the heat conductance measurement principle is used.
The measuring range 0 to 100 vol. % CH4 is not suitable for monitoring explosive
mixtures in the measuring range from 0 to 100%LEL.
If the DrägerSensor CatEx 125 PR is used in the gas detector, a zero point and
sensitivity calibration must be carried out after experiencing an impact load that
results in a non-zero display when exposed to fresh air. This warning does not apply if
the DrägerSensor CatEx 125 PR Gas is used.
16
with measuring range 100 vol. %, no blocking alarm is
4
CAUTION
WARNING
Operation
OK
OK
OK
OK
OK
OK
OK
Calling the Info Mode
In measuring mode, press the key for approx. 3 seconds.
If any warning or fault messages exist, the corresponding information or error codes
are displayed (page 30 to page 37).
Press the key successively for the next display.
The peak values and the exposition values TWA
Warning messages are displayed. Numerical codes of warning messages:
see page 30.
key
Fault messages are displayed. Numerical codes of fault messages: see
page 33.
key
The peak values = the maximum measured values in the case of, e.g., CO,
H
S, ... or the minimum measured values in the case of O2 within the storage
2
interval are displayed
key
The average values of the exposures based on a shift of, e.g., 8 hours (TWA)
of all the active sensors for the exposure evaluation are displayed
key
The short-term values (STEL) = average values of the concentrations over
the average value duration of all the active sensors for the exposure
evaluation are displayed
key
The device is in measuring mode again
OK
1)
and STEL1) are displayed.
If no key is pressed for 10 seconds, the device returns automatically to measuring
mode.
Calling the Info-Off Mode
When the device is in a deactivated state, press the key.
The name of the gas, measuring unit and measuring range limit value are displayed for
all channels.
Pressing the key again exits the Info Off mode (or via timeout).
1) Only when activated in the device configuration. Delivery status: not activated.
17
Operation
Calling the Quick Menu
Only the fresh air calibration is activated in the quick menu on delivery. The PC
software Dräger CC Vision can be used to activate the bump test for the quick menu
and/or the function for displaying and deleting peak values.
In measuring mode, press the key three times.
If no functions have been activated in the quick menu, the device remains in
measuring mode.
You can select the activated functions of the quick menu by pressing the key.
Press the key to call the selected function.
OK
Possible functions of the quick menu
Bump test, refer to page 40
Fresh air calibration, refer to page 46
Delete peak values, refer to page 19
Display pump information, refer to page 23
Activate or deactivate pump, refer to page 23
Press the key to cancel the active function and to switch to measuring mode.
If no key is pressed for 60 seconds, the device returns automatically to measuring
mode.
18
Quick menu "Delete peak val-
04823999_01_en.eps
ch4
%LEL
O
2
Vol%
CO
ppm
H
2
S
ppm
NH
3
ppm
04923999_01_en.eps
ch4
%LEL
O
2
Vol%
CO
ppm
H
2
S
ppm
NH
3
ppm
OK
OK
ues"
After the function has been selected, the
current peak values are displayed; the
peak values special symbol appears in
the display at the same time.
The peak values can be deleted by
pressing the key for 5 sec.
The adjacent display appears.
Operation
Press the key to end the function.
19
Operation
02323999_01.eps
OK
Calling the Calibration Menu
The calibration menu can only be accessed by entering a password.
Password on delivery: »001«
The default password on delivery can be changed using the PC software
Dräger CC Vision.
In measuring mode, press the key for at least 4 seconds.
The function for entering the password is selected.
The »« special symbol (for the "enter password" function) is displayed.
The display shows »000«, with the
first digit flashing.
Use the key to set the flashing digit.
Press the key, the second digit
starts flashing.
Use the key to set the flashing digit.
Press the key, the third digit starts
flashing.
Use the key to set the flashing digit.
Press the key to confirm the
password once it has been set
completely.
The calibration menu functions can
now be selected by pressing the
key.
Press the key to call the selected
function.
OK
OK
OK
Calibration menu functions
Fresh air calibration, refer to page 46
1-button calibration, refer to page 47
Single gas calibration, refer to page 50
Press the key to cancel the active function.
If no key is pressed for 10 minutes, the device automatically returns to measuring
mode.
20
Identifying Alarms
OK
OK
Identifying Alarms
An alarm is displayed visually, audibly and through vibration in a specific pattern.
NOTICE
At low temperatures the legibility of the display can be improved by switching on the
backlight.
Concentration pre-alarm A1
The alarm is indicated by an intermittent alarm message:
Display »A1« and measured value alternating: not for O
The pre-alarm A1 is not self-latching and stops when the concentration has dropped
below the alarm setpoint A1.
In the case of A1 a single tone is audible and the alarm LED flashes.
Acknowledging the pre-alarm:
Press the key. Only the audible alarm and the vibration alarm are switched off.
Concentration main alarm A2
The alarm is indicated by an intermittent alarm message:
Display »A2« and measured value alternating:
In the case of A2 a double tone is audible and the alarm LED flashes
twice.
For O
:»A1« and measured value alternating = oxygen deficiency
2
»A2« and measured value alternating = oxygen surplus
!
2
DANGER
Leave the area immediately. Danger to life! A main alarm is self-latching and cannot
be acknowledged or cancelled.
After leaving the area, if the concentration is less than the alarm setpoint A2:
Press the key. The alarm messages are switched off.
If the measuring range is exceeded significantly on the CatEx channel (very high
concentration of flammable materials), a blocking alarm is triggered. This CatEx
blocking alarm can be acknowledged manually by switching the device off and back on
again in fresh air.
In configuration setting CH
triggered as the heat conductance measurement principle is used.
with measuring range 100 vol. %, no blocking alarm is
4
CAUTION
The measuring range 0 to 100 vol. % CH
mixtures in the measuring range from 0 to 100 %LEL.
is not suitable for monitoring explosive
4
21
Identifying Alarms
STEL / TWA exposure alarm
The alarm is indicated by an intermittent alarm message:
Display »A2« and »« (TWA) or »« (STEL) and measured
value alternating:
CAUTION
Health hazard!
personnel is subject to the relevant national regulations.
STEL and TWA alarms cannot be acknowledged or canceled.
Switch off the device. The values for the exposure evaluation are deleted after the
device is switched on again.
The STEL alarm can be triggered with a maximum delay of one minute.
Leave the area immediately. After this alarm, the deployment of
NOTICE
Battery pre-alarm
The alarm is indicated by an intermittent alarm message:
Flashing special symbol »« on the right side of the display:
Acknowledging the pre-alarm:
Press the key. Only the audible alarm and the vibration alarm are switched off.
The battery still lasts approx. 20 minutes after the first battery pre-alarm.
OK
Battery main alarm
The alarm is indicated by an intermittent alarm message:
Flashing special symbol »« on the right side of the display:
The battery main alarm cannot be acknowledged or canceled:
The device is automatically switched off again after 10 seconds.
Before the device is switched off, the visual, audible and vibration alarms are
activated for a short time.
Device alarm
The alarm is indicated by an intermittent alarm message:
Special symbol »« on the right side of the display:
The device or one or several sensor channels are not ready for operation.
For remedies, refer to page 30 to page 37.
If necessary, commission the Dräger Service Center to eliminate the error.