Catalytic Bead DrägerSensor®
Sensors for Fixed Gas Detectors
The DrägerSensor® ... DQ detects ammable gases and vapours such as
hydrogen. Due to the double-detector compensation method, the catalytic
sensor is particularly long-term stable. The wire mesh at the gas inlet
serves as a ame barrier. So it ensures explosion protection at the same
time short response time.
Reference chamber with reference pellistorMeasuring pellistor
Flameproof stainless steel housing
Flame barrier
D-10088-2018
02 | Catalytic Bead DrägerSensor®
Benets
The catalytic bead principle
The DrägerSensors ... DQ are based on the catalytic bead detection principle. The most important component
are the hot, catalytically active pearls on platinum coils (pellistors). These pellistors burn the gas and additional
heat of reaction is created. The temperature increases depending on the gas concentration and thus also
changes the electrical resistance of the platinum coils. This change is passed on as a measuring signal. There
are ve different catalytic bead sensors from Dräger, which are used in different transmitters and measuring
heads.
DrägerSensor® PR M DQ
PR stands for poison resistant. The pellistors of the DrägerSensor PR M DQ have a longer service life under
the inuence of poisonous substances. The sensor has a metric thread (M) on the back.
DrägerSensor® PR NPT DQ
The DrägerSensor PR NPT DQ also withstands poisoning substances for a long time. This sensor is equipped
with a NPT thread on the back.
DrägerSensor® HT M DQ
The DrägerSensor HT M DQ can be used at ambient temperatures of up to 150 ° C and is also approved
accordingly (HT = high temperature). This sensor also has a metric thread.
DrägerSensor® LC M
LC stands for >low concentration<. The DrägerSensor LC M detects very low gas concentrations in the
measuring range from 0 to 10% LEL. This is possible by a special arrangement of the PR pellistors, as well as
an amplier electronics. The sensor has a metric thread.
DrägerSensor® LC NPT
The LC NPT sensor also detects gas concentrations in the measuring range of 0 to 10% LEL. It is equipped
with a NPT thread.
System Components
D-13692-2010
Catalytic Bead DrägerSensor® | 03
Dräger Polytron® SE Ex
The Dräger Polytron® SE Ex ... DQ sensing heads are gas detectors for
the continuous monitoring of ammable gases and vapours in the
ambient air. Measurement is based on the heat of reaction principle
where a chemical reaction takes place in a catalytic bead (also known
as a pellistor) inside the sensor.
Dräger PEX 3000
D-11158-2011
D-32407-2011
The transmitter Dräger PEX 3000 detects ammable gases and vapours
in concentrations below their lower explosive limit. Its DQ-sensor
provides a long-term stable measuring signal and responds to gas
within a few seconds.
Dräger Polytron® 5200 CAT
The Dräger Polytron® 5200 CAT is a cost-effective explosion-proof
transmitter for the detection of ammable gases in the lower explosion
limit (LEL). It uses a catalytic bead DrägerSensor® Ex … DQ that will
detect most ammable gases and vapours. A 3-wire 4 to 20 mA
analogue output with relays makes it compatible with most control
systems.
Dräger Polytron® 8200 CAT
The Dräger Polytron® 8200 CAT is an advanced explosion-proof
transmitter for the detection of ammable gases in the lower explosion
®
limit (LEL). It uses a catalytic bead DrägerSensor
Ex … DQ that will
detect most ammable gases and vapours. In addition to a 3-wire 4 to
20-mA analogue output with relays it also offers Modbus and Fieldbus
protocols, making it compatible with most control systems.
D-15042-2010
04 | Catalytic Bead DrägerSensor®
System Components
D-11951-2016
Dräger PointGard 2200
The Dräger PointGard 2200 series is a self-contained gas detection
system for the continuous area monitoring of ammable gases and
vapours in ambient air. PointGard 2200’s rugged, water-resistant
housing comes complete with a horn and strobes, a built-in power
supply, and reliable DrägerSensor®.
Related Products
ST-8821-2005
DrägerSensor IR
Upgrade from catalytic bead to infrared technology with ease using the
DrägerSensor IR. The Sensor IR can replace catalytic ex-sensors
(pellistors) from the majority of manufacturers without replacing
controllers, cables, junction boxes or control systems.
Technical Data
DrägerSensor® PR M DQ, PR NPT DQ, HT M DQ
Catalytic Bead DrägerSensor® | 05
Environmental conditionsTemperature (PR M/NPT DQ)
Temperature (HT M DQ)
Pressure
Relative humidity
StorageTemperature
Pressure
Relative humidity
Gas or vapourCAS-No.LEL in Vol.-%Typical sensitivity
II 2G Ex db IIC T6...T4 Gb, II 2G Ex tb IIIC T130°C...T195°C Db
IP6x (-50°C ≤ Ta ≤ +40°C/55°C/85°C)
Ex db IIC T6...T4 Gb, Ex tb IIIC T130°C...T195°C Db IP6x (-50°C ≤
Ta ≤ +40°C/55°C/85°C)
II 2G Ex db IIC T6...T4 Gb, II 2G Ex tb IIIC T130°C...T195°C Db
IP6x (-50°C ≤ Ta ≤ +40°C/55°C/85°C)
Ex db IIC T6...T4 Gb, Ex tb IIIC T130°C...195°C Db IP6x (-50°C ≤
Ta ≤ +40°C/55°C/85°C)
Class I, Groups A,B,C,D Class II, Groups E,F,G (-50°C ≤ Ta ≤
+85°C)
II 2G Ex db IIC T3 Gb, II 2GD Ex tb IIIC T195°C Db IP6x (-50°C ≤
Ta ≤ +155°C)
Ex db IIC T3 Gb, Ex tb IIIC T195°C Db IP6x (-50°C ≤ Ta ≤ +155°C)
DrägerSensor® LC M, LC NPT
Environmental conditionsTemperature
Pressure
Relative humidity
StorageTemperature
Pressure
Relative humidity
GasCAS-LEL in Vol.-%Test gas
concentration in
ppm
Methane
Ammonia
Ethene
Propane
Propene
Hydrogene
Approvals
DrägerSensor® LC MATEX
DrägerSensor® LC NPTATEX
74-82-84.02,56.07.06.0
7664-41-715.064.09.05.0
74-85-12.014.05.03.0
74-98-62.016.04.03.0
115-07-12.07004.05.02.0
1333-74-04.025.06.04.0
IECEx
IECEx
Test gas
concentration in
% LEL
-40 to +85 °C
900 to 1100 hPa
5 bis 95 % r. h.
-40 to +65 °C
70 to 1300 hPa
10 to 90 % r. h.
Sensitivity in mV
/ % LEL
II 2G Ex db eb IIC T6...T4, II 2D Ex tb IIIC
T80...130°C Db IP6x (-40°C ≤ Ta ≤
+40/50/85°C)Ex db eb IIC T6...T4, Ex tb IIIC T80...130°C
Db IP6x (-40°C ≤ Ta ≤ +40/50/85°C)
II 2G Ex db IIC T6...T4, II 2D Ex tb IIIC
T80...130°C Db IP6x (-40°C ≤ Ta ≤
+40/50/85°C)Ex db IIC T6...T4, Ex tb IIIC T80...130°C Db
IP6x (-40°C ≤ Ta ≤ +40/50/85°C)
Reading of a
methane-
calibrated
instrument in %
LEL
Ordering Information
Catalytic bead DrägerSensor®
DrägerSensor® PR M DQ
DrägerSensor® HT M DQ
DrägerSensor® PR NPT DQ
DrägerSensor® LC M
DrägerSensor® LC NPT
Catalytic Bead DrägerSensor® | 07
68 14 140
68 14 145
68 14 150
68 10 350
68 10 675
Calibration / adjustment accessories
Test gas cylinder methane apprx. 40 %LEL, 150 bar
Pressure reduceron request
Calibration adapter for PR M DQ, HT M DQ and PR NPT DQ
Remote calibration adapter for PR M DQ, HT M DQ and PR NPT DQ
1
1
Process adapter for PR M DQ, HT M DQ and PR NPT DQDQ
Calibration adapter for LC M and LC NPT
1
metrologically tested according to EN 60079-29-1 (in connection with the respective measuring heads or transmitters)
on request
68 06 978
68 12 480
68 12 470
68 06 978