Dräger Savina Service Manual

5664.900 Savina

Index

Contents

Navigation

Test Certificates

Repair Instructions

Technical Service Bulletins

Conversions

Periodic Manufacturer’s

Service Strategy

TSBs

List of Conversions

Service (PMS)

 

 

 

US Electrical Safety Test

Maintenance Procedure

 

 

Electr. Safety Test Form

Service Mode

 

 

 

Error List

 

 

 

Alarm Messages

 

 

 

Repair Information/Error Events

 

 

 

Repair Procedure

 

 

 

Software Download

 

 

 

Software Versions

 

 

Spare Parts Lists

Schematics and Diagrams

Function Descriptions

Operating Instructions

US Spare Parts List

Schematics and Diagrams

Function Description

US Operating Instructions

 

 

 

For information only:

 

 

 

English Instructions for Use

 

 

 

French Instructions for Use

 

 

 

Spanish Instructions for Use

A

B

C

O5664900T03.fm 27.08.01

For internal use only. Copyright reserved.

 

 

1DYLJDWLRQ

1DYLJDWLRQ WLSV

 

 

1DYLJDWLRQ

1DYLJDWLRQ WLSV

Words in blue are hyperlinks (except for the Dräger logo). Clicking on these hyperlinks will open the document at the corresponding text or view.

The icons on the left margin of the page are also hyperlinks. Clicking on them will have the following effects:

Opens the start page.

Opens the table of contents.

A

B

C Opens the keyword index.

If you do not see these icons, then you have opened a "separate" file. If you want to return to the start page, just close this "separate" file. Such "separate" files are, for example, "Instructions for Use/Operating Manuals" and the "Spare Parts Lists".

Dräger Medizintechnik GmbH

Navigation Savina Version 2.0

Page 2

A

B

C

 

 

1DYLJDWLRQ

1DYLJDWLRQ WLSV

 

 

If an hyperlink takes you to a figure, it may happen that only the figure number and the title appears on the upper edge of the screen. In this case you need to scroll up or select "Fit in window" from the Acrobat Reader "display" menu.

If you want to open several files in the same Acrobat Reader window (that is, a file is not closed automatically if you open another one), XQVHOHFW the "Open Cross-Doc Links in Same Window"

box in the "File/Preferences/General..." control page.

For further information on how to use the Acrobat Reader, please refer to the Acrobat Reader Online Guide in the Acrobat Reader "Help" menu.

O5664900T03.fm 27.08.01

For internal use only. Copyright reserved.

Dräger Medizintechnik GmbH

Navigation Savina Version 2.0

Page 3

A

B

C

27.08.01

only. Copyright reserved.

D5664900TECIVZ.fm

For internal use

&RQWHQWV

1DYLJDWLRQ

1DYLJDWLRQ WLSV

 

3HULRGLF 0DQXIDFWXUHU¶V 6HUYLFH 306

 

'HYLFH LQ JHQHUDO

 

5HSODFHPHQW RI ZHDU DQG WHDU SDUWV

 

(OHFWULFDO VDIHW\ WHVW

 

)XQFWLRQDO WHVWV

 

3ODFH IXOO\ IXQFWLRQDO XQLW DW XVHU V RZQHU V

GLVSRVDO

 

&RQILUPDWLRQ RI WHVW

 

5HSRUW

 

$QQH[

 

6HUYLFH 6WUDWHJ\

 

,WHPV WR EH UHSDLUHG 5HSDLU ORFDWLRQV

 

*HQHUDO QRWHV

 

0DLQWHQDQFH 3URFHGXUH 5HSODFLQJ ILOWHU DQG VHDOLQJ ULQJ LQ 22 FRP

SUHVVHG JDV LQOHW

 

22 SUHVVXUH UHGXFHU

 

5HSODFLQJ WKH 22 SUHVVXUH UHGXFHU ................... 57

Dräger Medizintechnik GmbH

Savina Version 2.0

Page 4

A

B

C

27.08.01

only. Copyright reserved.

D5664900TECIVZ.fm

For internal use

&RQWHQWV

 

$GMXVWLQJ WKH 22 SUHVVXUH UHGXFHU ...................

59

5HSODFLQJ WKH EORZHU XQLW

 

5HSODFLQJ WKH LQWHUQDO UHFKDUJHDEOH EDWWHULHV

 

 

5HSODFLQJ WKH UHDO WLPH FORFN RQ WKH

 

&RQWURO 3&%

 

6HUYLFH 0RGH

 

*HQHUDO

 

,QWHUQDO 6HUYLFH 0RGH

 

(QWHULQJ WKH ,QWHUQDO 6HUYLFH 0RGH ...................

74

/LVW RI WHVW VWHSV ..............................................

78

'HVFULSWLRQ RI WKH WHVW VWHSV ............................

80

Test Step 0 .......................................................

80

Test Step 1 .......................................................

80

Test Step 2 .......................................................

80

Test Step 3 .......................................................

81

Test Step 4 .......................................................

81

Test Step 5 .......................................................

82

Test Step 6 .......................................................

83

Test Step 7 .......................................................

84

Test Step 8 .......................................................

84

Test Step 9 .......................................................

85

Test Step 10 .....................................................

86

Test Step 11 .....................................................

87

Test Step 12 .....................................................

87

Test Step 13 .....................................................

88

Test Step 14 .....................................................

88

Test Step 15 .....................................................

89

Dräger Medizintechnik GmbH

Savina Version 2.0

Page 5

A

B

C

27.08.01

only. Copyright reserved.

D5664900TECIVZ.fm

For internal use

&RQWHQWV

 

Test Step 16 .....................................................

89

Test Step 17 .....................................................

90

Test Step 18 .....................................................

92

Test Step 19 .....................................................

94

Test Step 20 .....................................................

96

Test Step 21 .....................................................

96

Test Step 22 .....................................................

98

Test Step 23 .....................................................

98

Test Step 24 .....................................................

99

Test Step 25 .....................................................

99

Test Step 26 ...................................................

100

Test Step 27 ...................................................

100

3& 6HUYLFH 0RGH

 

5HTXLUHG HTXLSPHQW ......................................

101

(QWHULQJ WKH 3& 6HUYLFH 0RGH ........................

102

([LWLQJ 6HUYLFH 0RGH .....................................

107

(UURU /LVW

6WUXFWXUH RI HUURU OLVW

 

6RIWZDUH PDVWHU SURFHVVRU 03 HUURU JURXS

0DVWHU SURFHVVRU 03 UHDG RQO\ PHPRU\ HU URU JURXS

0DVWHU SURFHVVRU 03 5$0 HUURU JURXS

,QWHUQDO $' FRQYHUWHU VRIWZDUH 03 HUURU JURXS

,QWHUQDO FRPPXQLFDWLRQ VRIWZDUH 03 HUURU

Dräger Medizintechnik GmbH

Savina Version 2.0

Page 6

A

B

C

27.08.01

only. Copyright reserved.

D5664900TECIVZ.fm

For internal use

&RQWHQWV

JURXS

 

6RIWZDUH )3 DQG 03 HUURU JURXS

 

:DWFKGRJ 03 HUURU JURXS

 

57& HUURU JURXS

 

)URQW SURFHVVRU )3 VRIWZDUH HUURU JURXS

3RZHU SDFN HUURU JURXS

 

 

 

 

 

 

 

2SHUDWLQJ YROWDJHV HUURU JURXS

 

520 IURQW SURFHVVRU )3 HUURU JURXS

 

 

 

 

%ORZHU HUURU JURXS

 

5$0 IURQW SURFHVVRU )3 HUURU JURXS

 

 

 

 

&RROLQJ HUURU JURXS

 

,QWHUQDO $' FRQYHUWHU VRIWZDUH )3 HUURU

 

 

JURXS

 

6\VWHP UHOLHI HUURU JURXS

 

 

 

,QWHUQDO FRPPXQLFDWLRQ VRIWZDUH 03 HUURU

/RXGVSHDNHU HUURU JURXS

 

JURXS

 

 

 

:DWFKGRJ )3 HUURU JURXS

 

*ROG&DS HUURU JURXS

 

 

 

Dräger Medizintechnik GmbH

Savina Version 2.0

Page 7

A

B

C

27.08.01

only. Copyright reserved.

D5664900TECIVZ.fm

For internal use

&RQWHQWV

$X[LOLDU\ DODUP HUURU JURXS

 

$PELHQW SUHVVXUH PHDVXUHPHQW ZLWK 6 HU

 

 

URU JURXS

 

$LUZD\ SUHVVXUH PHDVXUHPHQW HUURU JURXS

([SLUDWRU\ IORZ PHDVXUHPHQW HUURU JURXS

,QVSLUDWRU\ IORZ PHDVXUHPHQW HUURU JURXS

2 FRQFHQWUDWLRQ PHDVXUHPHQW HUURU JURXS

 

 

$LUZD\ WHPSHUDWXUH PHDVXUHPHQW HUURU

 

JURXS

 

$PELHQW SUHVVXUH PHDVXUHPHQW ZLWK 6 HU

URU JURXS

 

Dräger Medizintechnik GmbH

Savina Version 2.0

2 VXSSO\ SUHVVXUH PHDVXUHPHQW HUURU

JURXS

 

6L6: 03 HUURU JURXS

 

6L6: )3 HUURU JURXS

 

$ODUP 0HVVDJHV

 

9HQWLODWLRQ PRQLWRULQJ

 

8SSHU DLUZD\ SUHVVXUH OLPLW ...........................

168

First pressure limit ..........................................

168

Second pressure limit (Paw high + 5 mbar) .........

170

/RZHU DLUZD\ SUHVVXUH OLPLW ...........................

172

0LQXWH YROXPH ..............................................

175

 

Page 8

A

B

C

27.08.01

only. Copyright reserved.

D5664900TECIVZ.fm

For internal use

&RQWHQWV

 

MV low ..........................................................

175

MV high .........................................................

177

,QVSLUDWRU\ WLGDO YROXPH .................................

179

VTi low ..........................................................

179

VTi high .........................................................

181

2[\JHQ FRQFHQWUDWLRQ ...................................

183

FiO2 low ........................................................

183

FiO2 high .......................................................

185

FiO2 limit values .............................................

187

&KDQJH LQ EUHDWKLQJ F\FOH .............................

188

Apnea ...........................................................

188

Fail to Cycle ...................................................

191

Apnea ventilation ............................................

193

+LJK UHVSLUDWRU\ IUHTXHQF\ ............................

194

$6% ! VHFRQGV ............................................

195

,QVSLUDWRU\ DLUZD\ WHPSHUDWXUH ......................

197

8VHU IXQFWLRQV ..............................................

199

Standby activation ...........................................

199

Flow monitoring ..............................................

200

O2 monitoring ................................................

201

Mask ventilation (NIV) .....................................

203

Nebulization ...................................................

204

Inspiration hold ...............................................

205

0RQLWRULQJ RI VHQVRUV

 

$LUZD\ SUHVVXUH VHQVRUV ...............................

206

([SLUDWRU\ IORZ VHQVRU ..................................

209

Expiratory flow sensor faulty .............................

209

Expiratory flow sensor not positioned correctly .....211

,QVSLUDWRU\ IORZ VHQVRU ..................................

212

2[\JHQ VHQVRUV ............................................

213

%UHDWKLQJ JDV WHPSHUDWXUH VHQVRU DW < SLHFH . 215

Dräger Medizintechnik GmbH

Savina Version 2.0

Page 9

A

B

C

27.08.01

only. Copyright reserved.

D5664900TECIVZ.fm

For internal use

&RQWHQWV

 

Breathing-gas temperature sensor faulty ............

215

Breathing-gas temperature sensor disconnected

. 216

$PELHQW SUHVVXUH VHQVRUV .............................

217

'HYLFH LQWHUQDO WHPSHUDWXUH VHQVRUV ..............

218

0RQLWRULQJ RI DFWXDWRUV

 

([SLUDWRU\ YDOYH ............................................

219

0RQLWRULQJ RI 3((3 .......................................

221

([SLUDWRU\ UHVLVWDQFH 3((3 KLJK ..................

223

*DV VXSSOLHV .................................................

226

Air intake .......................................................

226

Blower ...........................................................

228

Speed adaptation ............................................

229

No oxygen .....................................................

231

Oxygen supply high .........................................

233

0RQLWRULQJ RI FHQWUDO V\VWHP

 

%RRW SURFHGXUH .............................................

234

Testing of watchdogs .......................................

234

5-V voltage monitoring test ...............................

236

System relief ..................................................

237

Timer ............................................................

238

Settings .........................................................

239

Boot sequence ...............................................

240

0HPRU\ ........................................................

241

RTC battery ...................................................

241

Data retention in EEPROM and RTC RAM .........

242

RAM .............................................................

243

ROM .............................................................

244

2SHUDWLQJ YROWDJHV ........................................

246

3RZHU SDFN ...................................................

248

Charge state of internal battery .........................

248

Status of internal battery ..................................

250

Dräger Medizintechnik GmbH

Savina Version 2.0

Page 10

A

B

C

27.08.01

only. Copyright reserved.

D5664900TECIVZ.fm

For internal use

&RQWHQWV

 

Power failure ..................................................

252

Main switch ....................................................

254

$PELHQW SUHVVXUH ..........................................

256

'HYLFH LQWHUQDO WHPSHUDWXUH ...........................

258

'HYLFH IDQ .....................................................

259

.H\SDG .........................................................

260

/RXGVSHDNHU .................................................

262

$X[LOLDU\ DODUP ..............................................

263

'HYLFH UHVHW ..................................................

265

5HSDLU ,QIRUPDWLRQ (UURU (YHQWV

 

.QRZQ HUURUV

 

5HZRUNLQJ WKH FRYHU RI WKH 2 VHQVRUV ..........

268

:DUQLQJ PHVVDJH 1R LQW EDWWHU\ .............

269

"Starting aid" for charging of internal battery ........

271

Dräger Medizintechnik GmbH Savina Version 2.0

 

8QLW FDQQRW EH VZLWFKHG RII ...........................

272

Remedy 1 ......................................................

272

Remedy 2 ......................................................

272

(UURU ZKHQ XSGDWLQJ IURP 6: WR

...... 273

6: WR XSGDWH SURFHGXUH ..................

274

3RVVLEOH HUURUV GXULQJ 2 FDOLEUDWLRQ .............

275

5HSDLU 3URFHGXUH

 

2SHQLQJ WKH XQLW

 

2SHQLQJ WKH RSHUDWRU FRQWURO SDQHO ................

278

5HPRYLQJ WKH UHDU SDQHO ................................

281

3RZHU SDFN

 

0HDVXULQJ WKH VXSSO\ YROWDJHV ......................

285

0RWRU &RPPXWDWRU 3&%

 

 

Page 11

A

B

C

27.08.01

only. Copyright reserved.

D5664900TECIVZ.fm

For internal use

&RQWHQWV

9LHZ RI WKH 0RWRU &RPPXWDWRU 3&% ...............

287

6RIWZDUH 'RZQORDG

 

3LQ DVVLJQPHQW RI WKH 0RWRU &RPPXWDWRU 3&% 288

 

 

 

 

5HTXLUHG HTXLSPHQW

 

)DQ

 

 

 

3LQ DVVLJQPHQW RI WKH IDQ

289

'RZQORDG

 

 

 

 

 

'RZQORDGLQJ PHGLFDO HTXLSPHQW VRIWZDUH WR 3& ..

2 'LDSKUDJP 3&%

 

305

 

 

305

5HSODFLQJ WKH 2 'LDSKUDJP 3&%

290

'RZQORDGLQJ ................................................

 

 

Pin assignment of the O2 Diaphragm PCB .........

293

%RRWVWUDS GRZQORDG

 

 

 

%\SDVV YDOYH 9

 

5HTXLUHG WHVW HTXLSPHQW ................................

311

 

 

 

 

,QVWDOODWLRQ DQG XVH RI WKH ERRWVWUDS GRZQORDG SUR

([SLUDWRU\ YDOYH 9

 

JUDP ......................................................

312

 

 

5HSODFLQJ WKH ((3520

 

6RIWZDUH 9HUVLRQV

 

5HSODFLQJ WKH &RQWURO 3&%

 

2YHUYLHZ RI WKH GLIIHUHQW VRIWZDUH 6: YHUVL

 

 

RQV

 

Dräger Medizintechnik GmbH

Savina Version 2.0

Page 12

A

B

C

27.08.01

only. Copyright reserved.

D5664900TECIVZ.fm

For internal use

&RQWHQWV

 

'HVFULSWLRQ RI FKDQJHV

 

6FKHPDWLFV DQG 'LDJUDPV

 

)URQW YLHZ 6DYLQD

 

5HDU YLHZ RI 6DYLQD

 

,GHQWLILFDWLRQ RI 3&%V

 

22 9DOYH 3&% ................................................

328

&RQWURO 3&% ..................................................

330

)URQW 3DQHO 3&% ............................................

331

22 'LDSKUDJP3&% ........................................

333

3QHXPDWLF GLDJUDP

 

6DYLQD WXELQJ GLDJUDP

 

Dräger Medizintechnik GmbH Savina Version 2.0

 

)XQFWLRQ 'HVFULSWLRQ

 

*HQHUDO

 

:RUNLQJ SULQFLSOH ..........................................

336

0RQLWRULQJ ....................................................

337

3ULQFLSDO FRPSRQHQWV

 

(OHFWURQLFV ...................................................

339

Power pack ....................................................

340

Control PCB ...................................................

343

Motor Commutation PCB .................................

346

O2 Valve PCB .................................................

348

O2 Diaphragm PCB .........................................

349

Fan ...............................................................

350

2SHUDWRU &RQWURO 3DQHO ..................................

351

Front Panel PCB ............................................

352

Display ..........................................................

353

 

Page 13

A

B

C

27.08.01

only. Copyright reserved.

D5664900TECIVZ.fm

For internal use

&RQWHQWV

 

Membrane keypad ..........................................

353

Control knob ...................................................

353

3QHXPDWLF V\VWHP .........................................

354

Ventilation function ..........................................

358

O2 mix ...........................................................

359

Pneumatic safety devices .................................

360

Nebulizer function ...........................................

360

O2 calibration function ......................................

361

Sensors .........................................................

361

&RQYHUVLRQ ,QVWUXFWLRQV

/LVW RI &RQYHUVLRQV

 

Dräger Medizintechnik GmbH

Savina Version 2.0

Page 14

D

Periodic

 

 

 

 

Manufacturer’s

 

 

 

 

Service (PMS)

DrägerService

Savina

 

 

 

 

Explanation of Symbols

 

 

 

 

OK

Z = Check condition

 

 

 

 

 

 

Defect/error/fault

F = Check function

 

 

 

 

 

Spare parts used

D = Check for leaks

 

 

 

Report

P = Enter value

Accessories missing

For internal use only. Copyright reserved.

File no.:

Edition

Installation site:

Serial no.:

Date of delivery/ startup:

Invoice no. or delivery no.:

Other:

5664.900

08.2001

A

B

C

SW 01.n

SW 02.n

1.Device in general

Composition of the inspection price:

Basic unit with one expiratory valve. Additional expiratory valves must be charged separately.

1.1Software version

P

Enter software version of Savina.

1.2

Options as per rating plate

 

 

 

 

 

 

 

 

 

 

 

 

 

BIPAP/PCV+

 

 

P

 

 

 

 

 

 

 

 

 

 

 

 

Autoflow

 

 

 

 

 

 

 

 

 

 

 

 

 

P

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Central alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

P

 

 

 

 

 

 

 

 

 

 

 

 

NIV (mask ventilation, SW 02.n or later)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

P

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.3Operating hours

The operating hours can be read out under "Config" in the Savina menu.

1.3.1Total hours of operation

P

Enter total operating hours.

1.3.2

P

Service operating hours

Operating hours since last service.

15

Dräger Savina Service Manual

A

B

C

1.4General condition

 

 

Z

1.4.1

Accompanying documents

 

 

Instructions for use and medical product

 

 

logbook available according to operators.

Z

1.4.2

Device rear panel

 

1.4.2.1Dust filter set

Replace dust filter, if necessary. Dust filter

must not be pressed together.

Z

1.4.2.2Ambient air filter

Replace ambient-air filter, if necessary.

Z

1.4.3Power pack

1.4.3.1DC connection for external battery.

Connection has no visible damage.

1.4.3.2Cooling air fan

Switch on the unit. Check condition and functioning of cooling air fan. Switch off the unit.

1.4.4O2 inlet block.

Inlet block is fitted with NIST, DIN or DISS connectors.

Check condition and tightness of inlet block.

1.4.5Inspiration block

Z

Z F

Z D

Check condition of inspiration block.

Z

1.4.5.1Inspiratory socket

Check condition of inspiratory socket.

Z

1.4.6Nebulizer connection

The nebulizer connection is marked with a nebulizer symbol.

1.4.7Flow sensor mount (flow sensor)

The flow sensor can be inserted into its mount and moved to the right and to the left without effort.

1.4.7.1Flow sensor

Z

Z F

The flow sensor is undamaged.

Z

16

A

B

C

1.4.8Expiratory valve mount

Check locking function.

Z F

1.4.9Expiratory valve(s)

Enter serial number of valves: Valve 1:

P

Valve 2:

P

Valve 3:

P

1.4.9.1Expiratory valve, complete See also test item 2.1.

Inspect the following components. To do so, remove the respective expiratory valve:

Diaphragm, complete Non-return valve

Flow sensor sleeve, complete

Lip seal (for pressure measurement) Expiratory socket

Water trap container

 

Valve 1:

Z

 

Valve 2:

Z

 

Valve 3:

Z

 

Assemble each expiratory valve ready for use.

 

1.4.10

Savina front panel

 

1.4.10.1

Keypad, control knob, and display

 

 

Keypad, control knob, and display are

 

 

undamaged (for example, have no scratches

 

 

that could affect the view).

Z

17

A

B

C

1.4.11"Savina Mobil" trolley

1.4.11.1Castors

Wheels, of which 2 locking. Check screw

fittings for secure seating to 25 Nm.

Z F

1.4.11.2Trolley accessories

Check condition and function of the following accessories (if fitted):

Humidifier holder (optional)

 

Cylinder holder (optional)

 

 

 

 

 

 

 

 

 

 

 

 

 

"Multiple socket outlet" retrofit kit (option)

 

 

 

 

 

 

 

 

 

 

 

 

 

Battery pack (option, in trolley pedestal)

 

 

 

 

 

 

 

 

 

 

 

 

 

External battery connecting cable

 

 

Z F

 

 

 

 

 

 

 

 

 

 

 

1.4.12

Unit accessories

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Check condition and function of the following

 

 

 

 

 

 

 

 

 

 

 

 

 

accessories (if fitted):

 

 

 

 

 

 

 

 

 

 

 

 

 

Hinged arm (if fitted)

 

 

Z F

 

 

 

 

 

 

 

 

 

 

 

 

Temperature sensor (if fitted)

 

 

 

 

 

 

 

 

 

 

 

 

 

Z

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.4.13Compressed gas connecting tubes

Compressed gas connecting hoses comply with national regulations.

 

For example, compressed gas connecting

 

 

hoses with sealing rings, neutral color

 

 

for O2 NIST connectors, 3 m,

 

 

for O2 NIST connectors, 5 m,

Z D

1.4.14

Hose systems

 

 

Hose systems correspond to specifications in

 

 

the Instructions for Use/Operating

 

 

Instructions.

Z

1.4.15Adult test lung

Check condition and completeness of adult test lung.

The adult test lung comprises the following components:

Face mask angle piece

Catheter connector, ISO, size 7

 

Breathing bag, 2 L

Z

The breathing bag must not be overstretched.

18

1.4.16Drug nebulizer (white housing)

Additional note:

The drug nebulizer, 84 05 000, (black housing), must not be used in Savina.

A

B

C

1.4.16.1Condition test

Check the condition of the following components:

1

2 3 4

5

6

9

7

1 Patient connection

2 Nebulizer housing

3 Gasket

4 O ring

5 Nozzle

6 O ring

7 Atomizer

8 Container

9 Nebulizer hose

8

1.4.16.2Functional test

Fill nebulizer housing with water up to the "3"

mark. Supply drug nebulizer with 2.0 bar ± 0.2 bar.

Test: Vapor must be produced.

1.4.17Special accessories (if provided)

Check condition and function of the following special accessories (if available):

Resutator 2000 Child Resutator 2000

1.5Unit options and configuration

Not applicable —

Z

F

Z F

19

A

B

C

2.Replacement of wear and tear parts

2.1 * Replace diaphragm, complete, 84 13 661, in the expiratory valve.

Next replacement:___________________

2.2Replace internal batteries (rechargeable), 18 * 41 416, (x2) every 2 years

(see also chapter "Maintenance Procedure Replacing the internal rechargeable batteries").

Next replacement:___________________

2.3Replace batteries (rechargeable), 18 43 303,

*(if fitted, x2) in Savina Mobile every 2 years. Next replacement:___________________

2.4

*

Replace O2 filter, 84 08 208, and sealing

 

 

ring, M 09 257, in valve block, complete,

 

 

every 6 years.

 

 

(see also chapter "Maintenance Procedure

 

 

Replacing filter and sealing ring in O2

 

 

compressed gas inlet").

 

 

Next replacement:___________________

2.5

*

Replace O2 pressure reducer, 84 13 666, in

 

 

valve block, complete, every 6 years.

 

 

(see also chapter "Maintenance Procedure

 

 

O2 pressure reducer").

 

 

Next replacement:___________________

2.6

*

Replace real-time clock, 18 45 527, on

 

 

Control PCB, every 6 years.

 

 

(see also chapter "Maintenance Procedure

 

 

Replacing the real-time clock on the

 

 

Control PCB").

 

 

Next replacement:___________________

 

 

Note: After changing the real-time clock re-

 

 

enter the date and time.

2.7

*

Replace the blower unit with the "motor

blower unit" spares set, 84 13 643, after 20000 hours’ operation.

(see also chapter "Maintenance Procedure Replacing the blower unit").

20

A

B

C

3.Electrical safety test

The following steps describe the safety

Seech cksdocumentaccording" Electricalto IEC 60601SafetyandTestVDEin the USA/Canada".

0751. Any decision about performing safety checks according to VDE 0751 or IEC 60601 must be made under consideration of

Recordapplicablemeasuredn tional regvalueslatioinsthe. Electricaldocument "Electrical Safety Test Form Savina". safety checks should be carried out either

according to IEC 60601 or VDE 0751.

Savina conforms to regulations of protection class I, type B; with plugged-in airway temperature sensor to regulations of protection class I, type BF.

3.1

Visual inspection of basic unit

 

 

Check the following parts for possible

 

 

damage:

 

 

Power cable

 

 

Power switch

 

 

Compare the values of the power pack fuses

 

 

for mains power supply and internal

 

 

rechargeable battery with the rated values on

 

 

the rear panel.

Z

3.2Test of electrical safety to IEC 60-601

3.2.1Protective earth conductor test

Tester

Savina

 

L

 

N

6 V

 

 

PE

ohm

Test probe

PS

Protective earth conductor test

Connect the test probe in sequence to the following test points:

Power pack ground stud

Screws on housing

Rails, at the sides of Savina (if fitted)

Oxygen connection

Test value R≤ 0.2 ohms

P

21

A

B

C

3.2.2Earth leakage current test

Tester

Savina

L/N

 

Mains voltage

 

N/L

 

PE

 

PS

 

Earth leakage current test

 

Normal condition (N.C.):

 

I 500 A

P

Single fault condition (S.F.C.):

 

Neutral conductor interrupted:

 

I 1000 A

P

In the following steps the test is repeated, but

 

with the power plug turned over. This

 

condition can be created internally with some

 

types of test devices.

 

Normal condition (N.C.):

 

I 500 A

P

Single fault condition (S.F.C.):

 

Neutral conductor interrupted:

 

I 1000 A

P

22

A

B

C

3.2.3Patient leakage current test

Tester

Savina

L/N

Mains voltage

N/L

PE

PS

(1)

mA

Patient leakage current test

Connect airway temperature sensor (1) to the socket of the airway temperature sensor of Savina and to the tester.

Enter serial number of the airway temperature sensor:

P

 

Normal condition (N.C.):

 

 

I 100 A

P

 

Single fault condition (S.F.C.):

 

 

Neutral conductor interrupted:

 

 

I 500 A

P

 

In the following steps the test is repeated, but

 

 

with the power plug turned over. This

 

 

condition can be created internally with some

 

 

types of test devices.

 

 

Normal condition (N.C.):

 

 

I 100 A

P

 

Single fault condition (S.F.C.):

 

 

Neutral conductor interrupted:

 

 

I 500 A

P

3.2.4

Multiple socket outlet on trolley (if fitted)

 

3.2.4.1Compare fuses with rated values on the

socket-outlet.

Z

3.2.4.2Protective earth conductor test

Check all sockets.

 

Test value R 0.2 ohms

P

23

A

B

C

3.2.4.3Earth leakage current test

The test is carried out without Savina or other devices.

Normal condition (N.C.):

 

I ≤ 50 µ A

P

Single fault condition (S.F.C.):

 

Neutral conductor interrupted:

 

I ≤ 100 µ A

P

In the following steps the test is repeated, but

 

with the power plug turned over. This

 

condition can be created internally with some

 

types of test devices.

 

Normal condition (N.C.):

 

I ≤ 50 µ A

P

Single fault condition (S.F.C.):

 

Neutral conductor interrupted:

 

I ≤ 100 µ A

P

Continue Test Certificate at test item 4..

 

3.3Test of electrical safety to VDE 0751

3.3.1Protective earth conductor test

Tester

Savina

 

L

 

N

6 V

 

 

PE

ohm

Test probe

PS

Protective earth conductor test

Connect the test probe in sequence to the following test points:

Power pack ground stud

Screws on housing

Rails, at the sides of Savina (if fitted)

Oxygen connection

Test value R ≤ 0.3 ohms

P

24

A

B

C

3.3.2Equivalent device leakage current test

Tester

Savina

Rv

L

Mains voltage

N

 

 

PE

mA

PS

 

 

Test probe

Equivalent device leakage current test

Connect airway temperature sensor (1) to the socket of the airway temperature sensor of Savina and to the tester. Connect test probe to Savina.

Enter serial number of the airway temperature sensor:

P

Subsequent measurements may exceed the first-measured value by max. 50% but must at the same time be 1000 A.

Initial value: _______ A

Note:

Always enter the initial value in a new PMS.

I 1000 A

P

3.3.3equivalent patient leakage current test

Tester

Savina

 

L

Mains voltage

N

 

 

PE

mA

PS

Test probe

 

 

equivalent patient leakage current test

(1)

(1)

25

A

B

C

Connect airway temperature sensor (1) to the socket of the airway temperature sensor of Savina and to the tester. Connect test probe to Savina.

Enter serial number of the airway temperature sensor:

P

Subsequent measurements may exceed the first-measured value by max. 50% but must at the same time be 5000 A.

Initial value: _______ A

Note:

Always enter the initial value in a new PMS.

I 5000 A

P

3.3.4Multiple socket outlet on trolley (if fitted)

3.3.4.1Compare fuses with rated values on the

socket-outlet.

Z

3.3.4.2Protective earth conductor test Check all sockets.

Test value R 0.3 ohms

P

3.3.4.3Equivalent device leakage current test

The test is carried out without Savina or other devices.

Subsequent measurements may exceed the

first-measured value by max. 50% but must at the same time be 100 A.

Initial value: _______ A

Note:

Always enter the initial value in a new PMS.

I 100 A

P

26

A

B

C

4.Functional tests

4.1Power pack

Connect mains power and O2 supply. Switch on the unit.

4.1.1Testing the "electrical supply" display

The LEDs on the front of Savina show the following operating states:

Mains power supply: green External battery: off Internal battery:

Green: Internal rechargeable batteries are fully charged.

Yellow: Internal rechargeable batteries are

 

being charged.

F

4.1.2Testing the internal supply voltage Cut mains power.

The unit continues to ventilate. Audible alarm

sounds and display message "!! Int. battery

 

activated" appears.

F

The LEDs on the front of Savina show the

 

following operating states:

 

Mains power supply: off

 

External battery: off

 

Internal battery: green

F

4.1.3Testing the supply voltage failure alarm Connect device to the mains power supply.

Remove the fuse for the internal rechargeable battery. The fuse is located on the rear panel of the unit.

Disconnect the device from the mains power supply.

Ventilation off. All LEDs are off. Audible alarm

 

sounds.

F

Switch off the unit.

 

Restore mains power.

 

Re-insert fuse for internal rechargeable

 

battery.

 

Switch unit back on.

 

27

A

B

C

4.1.4Testing the external d.c. voltage supply

4.1.4.1Battery operation simulation

Notice: Maximum test duration = 1 minute!

“Connect Charge tester (Evita), complete" to d.c. voltage input of Savina (use the two lower sockets).

Left side terminal:

Right side terminal:

(-) black

I

(+) red

0

 

 

Using "DC voltage source, 3 -18 V" apply a voltage of 14 V to the charge tester.

After 6 seconds, at the latest, the LEDs on the

 

front of Savina show the following operating

 

states:

 

Mains power supply: green

 

External battery: green

 

Internal battery:

 

Green: Internal rechargeable batteries are

 

fully charged.

 

Yellow: Internal rechargeable batteries are

 

being charged.

F

Reduce the voltage from the DC source from

 

14 V to 11 V.

 

The LEDs on the front of Savina show the

 

following operating states:

 

Mains power supply: green

 

External battery: yellow

 

Internal battery:

 

Green: Internal rechargeable batteries are

 

fully charged.

 

Yellow: Internal rechargeable batteries are

 

being charged.

F

28

A

B

C

4.1.5On-board electrical system simulation

Short-circuit the two upper sockets on the d.c. voltage input of Savina using the "Cable for charge tester".

Set the voltage from the d.c. source to 14 V and wait for the "External battery" LED to light up green.

Reduce the voltage from the DC source to 11 V.

The LEDs on the front of Savina show the following operating states:

Mains power supply: green External battery: green Internal battery:

Green: Internal rechargeable batteries are fully charged.

Yellow: Internal rechargeable batteries are

 

being charged.

F

Remove the charge tester. Switch off the unit.

 

4.2Tests in internal service mode

Switch on Savina and, at the same time, press and hold the backlighting dim-up/dim-down key and the "Config" key until the service mode is activated (see also chapter "Service Mode Entering the Internal Service Mode").

4.2.1Checking the RS232 interface Test Step 0 : call up this test step.

Connect PC to Savina using "RS232 extension" and "RS232 adapter, crossed". Start terminal program and adjust the following:

COM 1 (or corresponding) Baud rate: 19200

Data bits: 8 Parity: None Stop bits: 1 Protocol: None

The settings on the PC must match those in the Medibus log of the Savina in the "Configure" menu.

Carry out "Send Logbook" function in internal service mode.

Test value:

The content of the error log is displayed on

 

the PC (terminal program).

F

29

A

B

C

4.2.2Testing the keys

Call up test steps 1 and 2 in sequence (as of SW 1.10)

Check the functioning of all keys.

Test value:

All keys should function properly.

F

4.2.3Nebulizer flow test

Test Step 13: call up this test step.

Connect a flowmeter tube with an oxygen scale of up to 16 L/min to the nebulizer connection. Switch on the nebulizer valve.

Test value:

Nebulizer flow = 8 L/min to 14 L/min.

P

4.2.4Comparing measurement functions of the flow sensors

Test Step 14: call up this test step.

Connect (short-circuit) the inspiratory and the expiratory sockets using a ventilation hose. Set the "PRESSURE BYPASS" value to

40 mbar. Test value:

The measured inspiratory and expiratory flows

 

must not differ by more than 13%.

P

4.2.5Offset voltages of the airway pressure sensors

Call up test 19. Test values:

"V_PAW_INSP" = 730 mV ± 100 mV

P

"V_PAW_EXP" = 730 mV± 100 mV

P

30