Dräger X-zone 5000 Instructions For Use Manual

de
Gebrauchsanweisung 2
en
Instructions for Use
fr
es
Instrucciones de uso
it
Istruzioni per l'uso
nl
Gebruiksaanwijzing
da
Brugsanvisning
no
Bruksanvisning
Dräger X-zone 5000
DRAFT 02 - 08/02/10
2
For Your Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
What is what . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Switching on the instrument . . . . . . . . . . . . . . . . . .5
Dräger X-am 5x00 info mode . . . . . . . . . . . . . . . . .6
Switching off the device . . . . . . . . . . . . . . . . . . . .6
Power supply for the Dräger X-am 5x00 with NiMH battery in the Dräger X-zone 5000: . . .6
Perform a function test with gas . . . . . . . . . . . . . .7
Perform a connection test . . . . . . . . . . . . . . . . . . .7
Safety housing . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Alarm-attenuation ring . . . . . . . . . . . . . . . . . . . . . .7
Establishing a wireless connection . . . . . . . . . . . .8
Connecting devices via a cable connection . . . . .10
Relay output: . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
During operation . . . . . . . . . . . . . . . . . . . . . . . . .12
Alarms (default settings) . . . . . . . . . . . . . . . . . . . .12
Concentration pre-alarm A1 . . . . . . . . . . . . . . . . .13
Concentration main alarm A2 . . . . . . . . . . . . . . .13
Battery pre-alarm . . . . . . . . . . . . . . . . . . . . . . . . .13
Battery main alarm . . . . . . . . . . . . . . . . . . . . . . . .13
Device alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Operation with pump (optional) . . . . . . . . . . . . . .14
Commissioning and performing the
measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Observe the following during measuring
mode with pump . . . . . . . . . . . . . . . . . . . . . . . . .15
Device configuration . . . . . . . . . . . . . . . . . . . . . . .15
Faults, Cause, Remedy . . . . . . . . . . . . . . . . . . . . .16
Fault messages . . . . . . . . . . . . . . . . . . . . . . . . . 18
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Maintenance intervals . . . . . . . . . . . . . . . . . . . . .19
Replacing the batteries . . . . . . . . . . . . . . . . . . . .19
Charging the batteries . . . . . . . . . . . . . . . . . . . . .20
Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Disposing of the device . . . . . . . . . . . . . . . . . . . . .20
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Key allocation of Dräger X-zone 5000 . . . . . . . . . .22
Status LED Overview . . . . . . . . . . . . . . . . . . . . . . .22
Overview of LED Ring and Horn Signals . . . . . . .24
Order List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Contents
DRAFT 02 - 08/02/10
3
For Your Safety
Strictly follow the Instructions for Use
Any use of the device requires full understanding and strict observation of these Instructions for Use. The device is only to be used for the purposes specified here.
Maintenance
The device must be inspected and serviced regularly by trained service personnel. Repair of the device may only be carried out by trained service personnel. We recommend that a service contract be obtain ed with Dräger and that all repairs also be carried out by Dräger. Only genuine Dräger parts should be used for maintenance. Strictly follow the instructions in the chapter "Maintenance intervals" on Page 18.
Accessories
Do not use accessory part s other than those specified in the order list Page 24.
Safe coupling with electrical device s
Electrical connections to devices which are not listed in these Instructions for Use should only be made following consultation with the respective manufacturers or an expert.
Use in areas subject to explosion hazards
Devices or components for use in explosion-hazard areas which have been tested and approved according to national, European or international Explosion Protection Regulations may only be used under the conditions specified in the approval and with consideration of the relevant legal regulations. The equipment or components may not be modified in any manner. The use o f faulty or incomple te parts is forbidden. The appropriate regulations must be observed at all times when carrying out repairs on these devices or components.
Safety symbols used in these Instructions for Use
These Instructions for Use contain a number of warnings for risks and hazards which might occur when using the instrument. These warnings contain signal words which will alert you to the degree of hazard you may encounter . These signal words and correspo nding hazards are as follows:
Intended Use
The Dräger X-zone 5000 is an explosion-proof, portable alarm amplifier for the quasi-stationary monitoring of hazard areas. The Dräger X-zone 5000 can be used in combination with the Dräger X-am 5000/5600 gas detection instruments. Several Dräger X-zone 5000 devices can operate in a self-crosslinking network.
Description
The Dräger X-zone 5000 alarm amplifier is intended for use under industrial conditions within a specified temperature range for an uninterrupted o perating period of up to 5 days. Via a wireless connection and/or communication ca ble, several Dräger X-zone 5000 devices can be grouped to form an alarm chain. In the event of an alarm, all connected Dräger X-zone 5000 devices will issue the alarm. The alarm is generated using the gas detection instrument Dräger X-am 5x00. The Dräger X-am 5x00 is connected via an infrared interface to the Dräger X-zone 5000. In addition to alarm generation, the Dräger X-am 5x00 is used as the user interface for the Dräger X-zone 5000. When the Dräger X-am 5x00 generates a gas alarm, this is transmitted to the Dräger X-zone 5000 and amplified acousticall y and optically.
DANGER
Indicates an immediate hazardous situatio n whic h, if not avoided, could result in death or serious injury.
WARNING
Indicates a potential hazardous situation which, if no t avoided, could result in death or serious injury.
CAUTION
Indicates a potential hazardous situation w hic h, if not avoided, could result in injury or damage to property. It may also be used to alert against unsafe practices.
NOTICE
Additional information on how to use the device.
NOTICE
The Dräger X-zone 5000 is intended for detection in ambient air. Any increased exposure to certain hydrocarbons can result in restrictions in the detection quality of the electrochemical sensors.
DRAFT 02 - 08/02/10
4
What is what
Front 1 Carrying handle
2 LED indicators 3 LED alarm ring 4 Inductive charging station 5 Lock 6 Device receptacle 7 Dräger X-am 5x00 holder diffusion mode 8 Dräger X-am 5x00 holder pump mode (optional)
Rear side 9 Quick-reference guide 10 Switch relay conn ec tio n / RS48 5 co nn ec tio n 11 Charging port connection / RS485 connection
Display 12 IR interface 13 Battery LED 14 Data transmission LED 15 Pump LED 16 Grouping LED 17 /wireless network key 18 key 19 Power contacts for Dräger X-am 5x00 20 Pump inlet (optional) 21 Pump outlet (optional)
OK
1
2
3
4
5
6
7
x-zone 5000
8
00233112.eps
10 11
9
00333112.eps
12
17
13 14 15 16
19
20
18
21
DRAFT 02 - 08/02/10
5
Operation
Switching on the instrument
STANDBY mode
z Place the Dräger X-am 5x00 into the device
receptacle.
z Place the holder onto the housing. z Fasten the lock at the holder.
ON mode
z Switch on the Dräger X-am 5x00 in accordance with
the instructions for use.
– The visual and the audible alarm will be activated for
a short time.
– The Dräger X-am 5x00 switches to the X-zone mode
(refer to “X-zone mode:” on page 5).
– The visual and acoustic life signal (green LED ring
and single tone) is issued depending on the configuration (1 - 60 seconds; default setting: 2 seconds).
– The Dräger X-zone 5000 is in the ON mode and
ready for operation.
– In ON mode, the alarm signals of the Dräger X-am
5x00 are evaluated, processed, and, if applicable, transmitted to other Dräger X-zone 5000 devices.
X-zone mode:
– The battery symbol on the Dräger X-am 5x00 is
replaced by the X-zone mode symbol
on the
Dräger X-zone 5000.
– The Dräger X-am 5x00 is supplied via the Dräger
X-zone 5000 battery.
– If the activated Dräger X-am 5x00 is removed from
the Dräger X-zone 5000, the device will exit X-zone mode again after max. 10 seconds.
NOTICE
Only Dräger X-am 5x00 devices of software version
3.5 and later are compatible with Dräger X-zon e 5000.
z Release the lock on
the holder.
z Remove the holder. z Hold down the key
on the Dräger X-zone 5000 for approx. 3 seconds.
– The battery LED will be
lit in green, red/green or red (refer to “Status LED Overview” on page 21) depending on the battery capacity.
– The Dräger X-
zone 5000 is in the STANDBY mode.
NOTICE
The Dräger X-am 5x00 must be equipped with a NiMH battery. The clip on the Dräger X-am 5x00 must be connected.
00433112.eps
OK
CAUTION
When switching on the Dräger X-zone 5000, Dräger recommends wearing hearing protection or using the alarm-attenuation ring (order no. 83 20 1 10) as the acoustic alarm is activate d for a brief period.
DRAFT 02 - 08/02/10
6
Dräger X-am 5x00 info mode
z Press and hold the key of the Dräger X-am 5x00
for approx. 3 seconds in detection mode.
– Press the key successively for the next display.
The peak values and the exposition values TW A and STEV as well as additional X-zone information will be displayed. If there are warnings or faults, the corresponding notes or error codes (e. g. X01) are displayed (refer to “Fault messages” on page 17).
Info window for Dräger X-am 5x00:
1 Station number in grouping mode:
The station number is not currently supported by the Dräger X-zone 5000.
2 Wireless status: ON/OFF:
Displays whether the wireless function of the Dräger X-zone 5000 is switched on or off. The wireless function can be switched on and off using the Dräger CC-Vision PC software (refer to “Device configuration” on page 14).
3 Network number:
If independent radio networks are required, dif ferent network numbers must be assigned (refer to “Establishing a wireless connection” on page 8). The network number can be set using the Dräger CC-Vision PC software (refer to “Device configuration” on page 14).
4 Grouping ID:
The grouping ID is not currently supported by the Dräger X-zone 5000.
5 X-zone mode symbol
Displays whether or not there is a connection between the Dräger X-zone 5000 and the Dräger X­am 5x00.
– If no key is pressed for 10 seconds, the Dräger X-am
5x00 returns automatically to detection mode.
Switching off the device
STANDBY mode
z Switch off the Dräger X-am 5x00 in the Dräger X-
zone 5000 in accordance with the instructions for use.
– The visual and the audible alarm will be activated for
a short time before the Dräger X-zone 5000 switches off.
– The Dräger X-zone 5000 switches to the ST ANDBY
mode.
OFF mode
z Release the lock at the holder. z Remove the holder. z If necessary , remove the Dräger X-am 5x00 from the
device receptacle.
z Press the key and the key on the Dräger X-
zone 5000 and hold down for approx. 3 seconds. – The battery LED goes off. – The Dräger X-zone 5000 is switched off
(OFF mode).
Power supply for the Dräger X-am 5x00 with NiMH battery in the Dräger X-zone 5000:
OK
OK
RF: NET: P\N:
ON
--
005 FFFF
1
2 3 4
5
CAUTION
When switching off the Dräger X-zone 5000, Dräger recommends wearing hearing protection or using the alarm-attenuation ring (order no. 83 20 110) as the acoustic alarm is activated for a brief period.
Device mode
Power supply for the Dräger X-am 5x00
Dräger X-zone 5000: ON mode Dräger X-am 5x00: Switched on
Dräger X-am 5x00 is permanently supplied with power.
Dräger X-zone 5000: STANDBY mode Dräger X-am 5x00: Switched off
Dräger X-am 5x00 is supplied with power via trickle charging.
Dräger X-zone 5000: OFF mode Dräger X-am 5x00: Switched off
Dräger X-am 5x00 is not supplied with power.
OK
DRAFT 02 - 08/02/10
7
Perform a function test with gas
The function test can be carried out in two ways.
Function test for the Dräger X-am 5x00:
z Perform a function test in accordance with the
instructions for use for the gas detection instrument before using with the Dräger X-zone 5000.
Function test on the Dräger X-am 5x00 in combination with the Dräger X-zone 5000:
z Switching on Dräger X-zone 5000 (refer to
“Switching on the instrument” on page 5).
z Place adapter (1) (order no. 83 20 108) on the holder
(diffusion).
z Connect the test gas cylinder to the adapter (2). z Open the test gas cylinder valve to let test gas flow
over the sensors.
z Wait until the instrument displays the test gas
concentration with sufficient tolerance: Ex: ±20 %
1)
O2: ±0.8 Vol.-%
1)
TOX: ±20 %1).
– Depending on the test gas concentration, alarm A1
or A2 is issued when the alarm thresholds are exceeded.
z Close the test gas cylinder valve and remove the
adapter from the holder. If the displays are outside of the above-mentioned ranges: Have the Dräger X-am 5x00 calibrated by service personnel.
Perform a connection test
The connection test can be used to check the correct connection to all Dräger X-zone 5000 devices. z Press the key on of the the Dräger X-am 5x00
devices being used three times in succession. A signal via a horn and LED ring is issued three times on every device connected wirelessly or via a cable.
Safety housing
In exceptionally rough environments, and in ver y strong sunlight and high temperatures (> +40 °C), the use of the safety housing (order no. 83 21 519) is recommended.
1 Hose/cable holder
Alarm-attenuation ring
When switching on and off and during the function test of the Dräger X-zone 5000, Dräger recommends wearing hearing protection or using the alarm­attenuation ring (order no. 83 20 110) as the acoustic alarm is sounded for a short time.
CAUTION
A function test must be carried out on every device before use.
NOTICE
The function test can only be carried out with the Dräger X-am 5x00 holder diffusion mode (ord er no. 83 20 636).
1) Upon application of the Dräger mixed gas (order no. 68 11130) the displays should be within this range. Different concentrations can be set using the Dräger CC-Vision PC software provided.
02733112.eps
1
2
WARNING
The alarm-attenuation ring must not be used in explosion-hazard areas!
OK
1
03033112.eps
DRAFT 02 - 08/02/10
8
Establishing a wireless connection
Up to 25 devices can be connected on a wireless network via a wireless connection. The typical radio range is up to 100 m in industrial surroundings (environmental factors may affect the range). The devices can be operated in a chain, star or ring topology. It is possible to establish an open radio network or independent radio networks. All devices within radio range are automatically connected.
Open wireless network (standard configuration):
With an open wireless network, any number of Dräg er X-zone 5000 devices (up to 25) can be added to or removed from a wireless network number.
Independent radio networks:
If independent radio networks are required, different network numbers need to be assigned for the relevant radio networks (refer to “Device configuration” on page 14). Example: Two independent radio networks each with three Dräger X­zone 5000 devices should be created. To do this, the Dräger CC-Vision PC software should be used to set NET:001 with three Dräger X-zone 5000 devices and NET:002 for three further Dräger X-zone 5000 devices.
Sub-networks:
1 Sub-network 1 2 Sub-network 2 3 e. g. steel wall, wall, HGV, etc.
Sub-networks are created via an unintentional division of the open network into two or more radio networks. This can occur if the user is running two activated Dräger X-zone 5000 (with the same network number). This means that the devices have permanently interconnected wireless connection (data transmission LED is lit in green). This means that the user can no longer tell whether or not the wireless connection is already set up with the devices already installed. To avoid sub-networks, the Dräger X-zone 5000 devices should always be set up and switched on in succession.
Positioning the device:
z Switch on the first Dräger X-zone 5000 (refer to
“Switching on the instrument” on page 5) and position at the relevant location.
WARNING
A maximum of 25 Dräger X-zone 5000 devices may be connected on a wireless network as otherwise it is not possible to ensure a reliable connection and alarm forwarding.
CAUTION
FCC and IC: This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
– this device may not cause harmful interference,
and
– this device must accept any interference received,
including interference that may cause undesired operation.
Modifications not expressl y ap pr ov ed by this company could void the user's authority to operate the equipment. The internal / external antennas used for this mobile transmitter must provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjun ction with any other antenna or transmitter. This device and its antenna must not be co-located or operating in conjunction with any other antenna or transmitter.
NOTICE
If the devices are in ON mode, they automatically establish the wireless connection.
CAUTION
Before positioning the devices, a function test (refer to “Perform a function test with gas” on page 7) must be conducted on every device.
NOTICE
The gas supply must be connected at an angle of 360°. If required, use base (order no. 83 20 645) to raise the detection position by approx. 30 cm.
NOTICE
When positioning the devices, ensure that sub­networks are avoided.
z Switch on the second Dräger X-
zone 5000 and position at a distance so that the data transmission LED is lit in green.
– The wireless connection is
established and the data transmission LED indicates the connection to at least one device within radio range.
03533112.eps
1
3
2
DRAFT 02 - 08/02/10
Loading...
+ 18 hidden pages