Dräger Savina 300 Technical Manual

Technical Documentation IPM
Savina 300 Intensive Care Ventilator
Warning
All servicing and/or test procedures on the device require detailed knowledge of this docu­mentation. Use of the device requires detailed knowledge and observance of the relevant Instructions for Use.
– This page has been intentionally left blank –
2
No.1309-0000001227_Publication
Savina 300
Table of contents
Table of contents
General
General notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Function descriptions
Block diagram and functional principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Electronic assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Control panel with TFT colour display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pneumatic assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance instructions
Disassembling/assembling the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing the microfilter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing the dust filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing the O2 sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing the diaphragm of the expiratory valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing the internal batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing the external batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing the consumables in the O2 gas inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Annex
5
6
9
10
12
18
20
31
32
38
40
42
47
50
54
65
71
No.1309_0000001227_Publication
Savina 300
3
– This page has been intentionally left blank –
4
No.1309-0000001227_Publication
Savina 300
General
This chapter contains general notes and definitions that are important for the
use of this documentation.
General notes ....................................................................................................................................
6
No.1309_0000001227_Publication
Savina 300
5
General
General notes
General notes
Notes on use Read through the following notes thoroughly before applying this documenta-
tion.
Dräger reserves the right to make changes to the device and/or to this docu­mentation without prior notice. This documentation is intended solely as an information resource for experts.
Copyright and other pro­tected rights
Definitions
The content of this documentation, in particular its design, text, software, technical drawings, configurations, graphics, images, data and their selection and its composition and any amendments to it ("content") are protected by copyright. The content must not (in whole or in part) be modified, copied, dis­tributed, reproduced, republished, displayed, transmitted or sold without the prior written consent of Dräger.
WARNING!
A WARNING statement provides important information about a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
A CAUTION statement provides important information about a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or in damage to the medical device or other property.
NOTE
A NOTE provides additional information intended to avoid inconvenience during operation and/or servicing.
6
Inspection = Identification of actual condition
Maintenance = Measures to maintain the speci-
fied condition
Repair = Measures to restore specified
condition
Servicing = Inspection, maintenance, repair
No.1309_0000001227_Publication
Savina 300
General
General notes
General safety precautions Read through each section thoroughly before beginning servicing. Always
use the correct tools and the specified test equipment. Otherwise the device may not work correctly or may be damaged.
WARNING!
Improper servicing If the medical product is not properly serviced, the safety of the patient and/or the operator may be put at risk.
Have the medical product checked and maintained on a regular basis
by appropriately qualified experts, otherwise the proper functioning of the device may be compromised.
Have repairs to the medical product carried out only by personnel who
have undergone product-specific Dräger training.
NOTE
Dräger recommends entering into a service contract with DrägerService and having all repairs likewise carried out by DrägerService.
WARNING!
Replacement parts not certified by Dräger standards Dräger cannot guarantee or confirm the operational safety of third-party replacement parts used on the device.
Use only replacement parts certified to Dräger standards for servicing
of the device, otherwise the proper functioning of the device may be compromised.
Pay attention to the „Servicing“ section of the Instructions for Use.
WARNING!
Non-conforming test values If test values do not conform to specifications, the safety of the patient may be put at risk.
Do not put the device into operation if test values do not conform to
specifications.
Contact your local service organization.
WARNING!
Impermissible modifications to the device If impermissible modifications are made to the device, the safety of the patient may be put at risk.
Do not modify the device without Dräger's permission.
WARNING!
Risk of infection The device may transmit pathogens following use on the patient.
Before carrying out any servicing, ensure that the device and its com-
Service only cleaned and disinfected devices and device components.
No.1309_0000001227_Publication
Savina 300
ponents have been handed over by the user cleaned and disinfected.
7
General
General notes
WARNING!
Risk to patients
Ensure that no patient is connected to the device before starting main-
tenance or repair work.
NOTE
Where reference is made to legislation, regulations and standards, in respect of devices used and serviced in Germany they are based on the laws of Germany. Users and technicians in other countries must comply with their national laws and/or international standards.
8
No.1309_0000001227_Publication
Savina 300
Function descriptions
This chapter contains descriptions of the device's technical functions.
Block diagram and functional principle ..............................................................................................
Electronic assembly ...........................................................................................................................
Control panel with TFT colour display................................................................................................
Pneumatic assembly..........................................................................................................................
10
12
18
20
No.1309_0000001227_Publication
Savina 300
9
Function descriptions
Block diagram and functional principle
Block diagram and functional principle
Introduction The following text sets out the block diagram and the functional principle of
the Savina 300.
Block diagram
Fig. 1 Block diagram Savina 300
10
No.1309_0000001227_Publication
Savina 300
Function descriptions
Block diagram and functional principle
Functional principle The blower generates the necessary compressed air for ventilation. A con-
trollable valve (non-return valve) is switched in parallel with the blower to reg­ulate the pre-set ventilation parameter. The non-return valve opens or closes according to the pre-set ventilation parameters.To increase the oxygen con­centration in the ventilation gas, an external oxygen source can be connected to the device. Sensors, the electronics and the valve block meter the oxygen concentration.
Main components The device consists of the following main components:
Elektronic assembly
Control unit
Pneumatic assembly
No.1309_0000001227_Publication
Savina 300
11
Function descriptions
Electronic assembly
Electronic assembly
Introduction The electronic assembly contains the following subassemblies:
Power supply unit
Central Control Board
Motor actuator
–O2 Valve PCB
O2 Diaphragm PCB
Fan
Power supply unit The power supply unit delivers the supply voltages for the device. The input
voltage range is 100 V to 240 V AC and 50 Hz to 60 Hz. The power supply unit can also be operated with an external rechargeable battery (12 V or 24 V).
The connection to the alternating voltage is made by a power cable. The con­nection for the external rechargeable battery is made by an encoded connec­tor.
The power supply unit actuates the „Mains power“, „External battery or on­board power supply“ and „Internal battery“ LEDs. The LEDs are mounted on the membrane keypad of the control panel and indicate the respective operat­ing status.
The device includes two internal rechargeable batteries (2 x 12 V) which enable uninterrupted operation in the event of a mains power failure. The internal rechargeable batteries supply the O2 sensors with power even when the device is switched off. When the device is switched on the valid O2 val­ues are present.
The power supply unit generates the following supply voltages:
–+5 V
–-15 V
–+15 V
–+24 V
–+48 V
The output voltages are short-circuit-proof and stable at no-load.
12
The output voltages are generated according to the following priority, depen­dent on the input voltages:
Input voltage Priority Action
AC voltage 1 Charge external and internal batteries, and maintain the
charge.
External battery/batteries 2 Charge internal batteries, and maintain the charge.
Internal batteries 3 -
No.1309_0000001227_Publication
Savina 300
Function descriptions
Electronic assembly
The fan cools the power pack.
Central Control Board The Central Control Board is the device's central „control and monitoring
unit“. It has three separate processor systems (master processor, front pro­cessor and display processor).
Three processor systems save the changeable, non-volatile data to EEPROMs.
Fig. 2 Block diagram of the Central Control Board
Master processor
The master processor has the following tasks:
Activation of the actuators (valves, valve block, blower)
Reading of the measured values from the sensors
Control of the ventilation
Monitoring of the front processor
The master processor data are cached in the RAM.
The EEPROM of the master processor system stores the calibration data of the sensors and set values such as volume.
No.1309_0000001227_Publication
Savina 300
13
Function descriptions
Electronic assembly
Front processor
The front processor has the following tasks:
Interface with the display processor
Monitoring of the master processor and display processor
Monitoring of the input logic and the data e.g. the pixel sum
The RAM caches data from the front processor.
The EEPROM with socket of the front processor stores safety parameters, set values such as volume, software options and the operating hours.
Display processor
The display processor has the following tasks:
Presentation of curves and parameters on the display, LEDs
Input and operator data (keyboard, rotary encoder, touchscreen)
Collation of data such as the pixel sum.
The EEPROM of the display processor stores user-specific display settings, such as measured value positions.
Flash EEPROMs
The device's operating system program (software) is stored in Flash EEPROMs (rewritable memory modules).
Real-time clock
The real-time clock generates the time and date information. The real-time clock has an internal battery.
The Central Control Board incorporates the following functions:
Processing of the signals from the sensors (O2, flow, pressure, tempera-
ture)
Control of the blower and valves
Monitoring of the unit functions and the supply voltages
Actuation of the displays
Keypad interpretation
Provision of the internal and external interfaces
Set values such as the volume
14
Mains power failure alarm
If the mains power fails, the device switches to the internal batteries which supply the device with power. If the internal batteries are discharged and the device has no external batteries, an acoustic alarm is sounded with a Gold­Cap and a horn.
No.1309_0000001227_Publication
Savina 300
Function descriptions
Electronic assembly
Temperature hybrid
The temperature hybrid reads the data of the AWT01 sensor and converts these into digital signals. An electrical isolation occurs in this process.
Internal temperature measurement
The internal temperature measurement is a safety function. In the event of overheating (if the internal temperature of the device is too high), an alarm is sounded.
Fan actuation and fans
The fan rotates quickly after the device has been switched on to remove any residual oxygen from the device. The rotation of the fan is detected and con­trolled in three stages (slow, medium and fast).
Three temperature sensors control the fan.
CAN
The CAN interface is a fast, serial interface. Via the CAN interface the control unit can communicate with the electronic and pneumatic assemblies. The transmission rate is 800 kbit/s.
Motor actuator The motor actuator controls the blower motor. The motor actuator is located
in a self-contained housing. The supply voltage for the motor actuator is +48 V and is protected by a fuse (6.3AT).
The input voltage range of the motor actuator is 12 to 52.5 V. The rotation speed is set by the Central Control Board. The control voltage for the rotation speed is 0 V to +5.00 V, corresponding to a rotation speed of 0 to 12,000 rpm. The rotation speed range is 4,000 to 12,000 rpm.
The motor actuator acquires the „actual speed signal“ and forwards it to the Central Control Board. The „actual value signal“ is 6 pulses per rotation. In the event of discrepancies in the rotation speed the Central Control Board adjusts the speed according to the deviation.
Fig. 3 Block diagram of the motor actuator
No.1309_0000001227_Publication
Savina 300
15
Function descriptions
Electronic assembly
O2 Valve PCB The O2 Valve PCB holds the pressure sensors (absolute pressure S6 and S7
and O2 supply pressure S5), the actuator for the O2 calibrating valve and the nebulizer valve and the actuator for the valve block.
The signals of the pressure sensors are amplified and routed to the Central Control Board. The supply voltage (+5 V) for the pressure sensors is gener­ated by the O2 Valve PCB.
The valve block valves, the O2 calibrating valve and the nebulizer valve can be operated separately by an electronic switch. They are actuated by the Central Control Board.
Fig. 4 Block diagram of the O2 Valve PCB
O2 Diaphragm PCB The O2 Diaphragm PCB amplifies the signals from the O2 sensors and mea-
sures the temperature of the O2 sensors and of the respiratory gas in the inspiration block. The temperature of the O2 sensors is required to compen­sate for the temperature-sensitive O2 measurements. The EEPROM on the Central Control Board stores the calibration data of the sensors. The refer­ence voltage for the O2 sensors is generated from the voltage of the rechargeable batteries.
The supply voltages for the O2 Diaphragm PCB are +5 V and +15 V.
16
Fig. 5 O2 Diaphragm PCB block diagram
No.1309_0000001227_Publication
Savina 300
Function descriptions
Electronic assembly
Fan The fan takes in ambient air through the cooler and cools the blower motor.
The air flow removes excess oxygen from the device.
The supply voltage for the fan is +24 V. The Central Control Board regulates the speed of the fan.
Fig. 6 Fan control
Item Designation
1 Supply voltage
2 Target speed
3 Speed signal
4 Sensor supply voltage
5 Ground
No.1309_0000001227_Publication
Savina 300
17
Function descriptions
Control panel with TFT colour display
Control panel with TFT colour display
Control unit The control panel is the interface between the unit and the user. The control
panel is used to enter and display the ventilation parameters.
Fig. 7 Block diagram of the control panel with TFT colour display
Membrane keypad The keypad features the control keys and the associated LEDs.
TFT colour display The 12.1” TFT colour display comprises the actual display and the backlight
converter. The TFT colour display has a resolution of 800 x 600 pixels.
The backlight converter generates a high voltage for the display backlighting.
Rotary encoder The rotary knob is used to set and acknowledge the ventilation parameters.
The shaft encoder transmits square signals to the processor system as it rotates, and the signals are then evaluated by the Central Control Board. The voltage supply is +5 V.
No.1309_0000001227_Publication
18
Savina 300
Touch-panel Analog-resistive touch-screen
Function descriptions
Control panel with TFT colour display
Fig. 8 Analog-resistive touch-screen
Analog systems consist of two opposing conductive indium tin oxide (ITO) layers (x and y layer) which are actuated with a constant direct voltage. Indium tin oxide is a transparent semiconducting material.
Between the two ITO layers there are a large number of small, barely visible so-called spacer dots which ensure the two layers are kept separate.
In 8-wire systems the touch-screen has eight wires routed to the controller – four for each axis.
When the touch-screen is touched at a certain point where the two ITO layers are located an electrical contact is produced. The resistance of this contact results in a different voltage at each point. The change in voltage is then used to define the x and y coordinates.
The Central Control Board controls the communication between the proces­sor system and the touch-screen. The correct position is determined with the aid of the relevant software drivers. The analog touch-screen works very pre­cisely, and provides a high resolution.
No.1309_0000001227_Publication
Savina 300
19
Function descriptions
Pneumatic assembly
Pneumatic assembly
Functional diagram
Fig. 9 Functional diagram
Item Designation
F1 Micro-filter (AIR)
F2 Filter (O2)
F4 Filter (inspiratory flow sensor)
F5 Filter (Filter element)
Volume Mixing chamber
SD1 Suction sound insulator
SD2 Sound insulator
DR1 Supply pressure regulator (Oxygen)
Gebläse Blower
Kühler Cooler
V1 Reduction valve
V3 Expiratory valve
20
No.1309_0000001227_Publication
Savina 300
Function descriptions
Pneumatic assembly
Item Designation
V4 Switching valve (oxygen compensation)
V5 Switching valve (nebulizer)
V6.1 Calibration valve (inspiratory airway pressure sensor)
V6.2 Calibration valve (expiratory airway pressure sensor)
V7.1 - V7.8 Oxygen metering valves
V8 Pilot valve for emergency vent valve
V9 Emergency vent valve
S1 Inspiratory flow sensor
S2 Expiratory flow sensor
S3.1 O2 sensor 1 (measurement and control)
S3.2 O2 sensor 2 (Monitoring)
S4.1 Inspiratory airway pressure sensor (located in inspiratory branch)
S4.2 Expiratory airway pressure sensor (located in expiratory branch)
S5 Supply pressure sensor (oxygen)
S6 Pressure sensor 1 (absolute pressure)
S7 Pressure sensor 2 (absolute pressure)
D1 Safety pressure-limitting valve (passive, approx. 120 mbar)
D2 Emergency air valve (-3 mbar to -6 mbar)
D3 Expiratory non-return valve
D4 Inspiratory non-return valve
D5 Flush flow non-return valve
D6 Non-return valve (LPO)
R1 Flush flow metering unit (0,1 L/min at 30 mbar)
R2 O2 calibration metering unit for (0,1 L/min, integrated in valve block)
R3 Metering unit for O2 measurement (0,2 L/min at 30 mbar, Sensor 3.1)
R4 Metering unit for nebulizer (10 L/min, integrated in valve block)
R5.1 - R5.8 Metering units for the oxygen metering valves
R6 Metering unit for O2 measurement (0,2 L/min at 30 mbar, Sensor 3.2)
Main components The pneumatic assembly consists of the following components:
Plug-in unit
Valve block
Inspiratory block
Pressure measuring block
Patient system
Flow sensors
No.1309_0000001227_Publication
Savina 300
21
Function descriptions
Pneumatic assembly
Device functions
Ventilation function When the device is switched on the power supply unit supplies the blower
motor with operating voltage. The blower motor draws in ambient air through the microfilter F1, the volume and the sound insulators SD1 and SD3. The blower motor compresses the gas intake to an overpressure up to max. 140 mbar at a delivery rate of up to 180 L/min. The air compressed by the blower motor passes through the sound insulator SD4, the cooler, the sound insula­tor SD2, the filter F4 and the inspiratory flow sensor S1 to the inspiratory non­return valve D4. The controller operates the blower motor during a breath at a constant speed. The bypass valve regulates the inspiratory pressure. The combination of the blower motor and bypass valve V1 provides a pressure source.
Fig. 10 Detail view of functional diagram; ventilation function
The sound insulators SD1 and SD2 on the inlet and outlet sides of the blower motor reduce the sound level.
The bypass valve V1 is operated such that the desired respiratory pressure is applied to the blower outlet, and thus to the patient. If the patient needs a high flow during the inspiratory phase, the gas flows in part or in its entirety from the blower motor outlet to the patient and the gas flow through the bypass valve V1 is reduced. During the expiratory phase all the blower gas flows through the bypass valve V1.
The cooler reduces the respiratory gas temperature down to a permissible range.
22
No.1309_0000001227_Publication
Savina 300
Function descriptions
Pneumatic assembly
O2 mixture with O2 high pressure
In order to be able to ventilate with an increased O2 concentration, the unit must be supplied with 2.7 to 6.0 bar O2. The oxygen is filtered by the filter F2. With the aid of the digital valve block consisting of 8 digital solenoid valves, oxygen is metered into the volume (mixing chamber). The amount of metered oxygen depends on the pre-set O2 concentration and on the inspiratory flow rate measured by the flow sensor S1. The oxygen is metered in a closed con­trol loop. In the process, the inspiratory O2 concentration is measured by the O2 sensor S 3.1.
Fig. 11 Detail view of functional diagram; O2 mixture with O2 high pres-
sure
No.1309_0000001227_Publication
Savina 300
23
Function descriptions
Pneumatic assembly
O2 mixture with O2 low pressure („LPO“ option)
NOTE
Connect only O2 low pressure sources without humidifier to the device!
An O2 low pressure source without humidifier feeds the oxygen into the “LPO” connection on the back of the unit. The filter (filter element) F5 protects the non-return valve D6 from coarse particles. The oxygen flows from the non-return valve D6 into the volume (mixing chamber). In the volume (mixing chamber) it is mixed with the drawn-in and filtered fresh air.
Fig. 12 Detail view of functional diagram; O2 mixture with O2 low pressure
When no O2 low pressure source is connected to the unit, the non-return valve D6 prevents gas from escaping during normal operation.
NOTE
In „LPO“ mode, the valve block in the O2 supply is not actuated.
24
No.1309_0000001227_Publication
Savina 300
Function descriptions
Pneumatic assembly
Pneumatic safety devices The pneumatic safety valve D1 ensures that the ventilation pressure cannot
rise above 120 mbar. In the event of inspiratory stenosis the pressure is lim­ited by opening the expiratory valve. The mechanical vacuum valve D2 ensures (except in the case of inspiratory stenosis) that the patient can breathe ambient air in case of a fault.
The pneumatic emergency vent valve V9 relieves the pressure in the breath­ing system in a case of expiratory stenosis if the pressure cannot be relieved through the expiratory valve. To do so, the electric emergency vent valve (pilot valve) V8 actuates the emergency vent valve V9 accordingly.
Fig. 13 Detail view of functional diagram; pneumatic safety devices
Drug nebulizer The drug nebulizer is operated with 100% oxygen. The pressure regulator
DR1 ensures in the case of widely varying supply pressure (2.7 to 6.0 bar) that the pneumatic drug nebulizer receives a constant supply pressure of 2 bar.During the nebulizing phase the solenoid valve V5 operates in an “inspi­ration” (open position) and “expiration” (closed position) cycle. When the neb­ulizer function is inactive the nebulizer switching valve V5 is closed.
Fig. 14 Detail view of functional diagram; drug nebulier
No.1309_0000001227_Publication
Savina 300
25
Function descriptions
Pneumatic assembly
O2 calibration function During operation, the switching valve (oxygen compensation) V4 is set to
“measurement” – that is, the connection between the inspiratory side and the oxygen sensor is open. During oxygen sensor calibration oxygen passes to the oxygen sensor. This layout permits “online” calibration of the oxygen sen­sor S3.1 during ventilation. The oxygen sensor S3.2 must be calibrated man­ually (patient disconnected).
Fig. 15 Detail view of functional diagram; O2 calibration function
O2 sensor detection The device has an oxygen sensor detector which is necessary for the „LPO“
option.
26
No.1309_0000001227_Publication
Savina 300
Function descriptions
Pneumatic assembly
Sensors The pressure sensor (absolute pressure) S6 measures the atmospheric pres-
sure necessary for oxygen measurement and for volume application. The pressure sensor (absolute pressure) S7 monitors the pressure sensor (abso­lute pressure) S6.The oxygen sensor S3.1 generates the signal for the dis­played „FiO2“ measured value and the signal to control the inspiratory oxygen concentration. The oxygen sensor S3.2 monitors the oxygen sensor S3.1
Fig. 16 Detail view of functional diagram; pressure sensors and oxygen
sensor
The airway pressure sensor S4.1 measures the pressure in the inspiratory branch. The airway pressure sensor S4.2 measures the pressure in the expi­ratory branch. The output signals of the airway pressure sensors are needed
No.1309_0000001227_Publication
Savina 300
27
Function descriptions
Pneumatic assembly
to determine the airway pressure and for control and monitoring purposes. The airway pressure is measured on the basis of the measured value from the airway pressure sensor in the respective no-flow branch.
Fig. 17 Detail view of functional diagram; airway pressure sensors
The calibration valves V6.1 and V6.2 enable calibration of the inspiratory and expiratory airway pressure sensors. During calibration, the corresponding calibration valve interrupts the connection to the ventilation circuit and switches the airway pressure sensor to ambient pressure.
28
Fig. 18 Detail view of functional diagram; calibration valves
No.1309_0000001227_Publication
Savina 300
Function descriptions
Pneumatic assembly
Flow sensor S1 measures the inspiratory gas flow. The measurement vari­able is used to calculate the necessary oxygen flow and to actuate the oxy­gen metering valves V7.1 to V7.8 in order to control the breaths and monitor the device functions. The flow sensor includes a temperature measurement function to measure the inspiratory gas temperature.
Fig. 19 Detail view of functional diagram; flow sensor
The flow sensor S2 measures the gas flow through the expiratory valve. The flow sensor is a temperature-compensated hot-wire anemometer with no flow direction detector. With this signal the patient is monitored (e.g. the minute volume).
Fig. 20 Detail view of functional diagram; flow sensor
Expiration is executed with the directly actuated expiratory valve V3.
The expiratory valve has the following functions:
PEEP control during the expiratory phase
Close the breathing system during the inspiratory phase.
No.1309_0000001227_Publication
Savina 300
29
– This page has been intentionally left blank –
30
No.1309-0000001227_Publication
Savina 300
Loading...
+ 136 hidden pages