5664.900 Savina
Index
Contents
Navigation
Test Certificates |
Repair Instructions |
Technical Service Bulletins |
Conversions |
Periodic Manufacturer’s |
Service Strategy |
TSBs |
List of Conversions |
Service (PMS) |
|
|
|
US Electrical Safety Test |
Maintenance Procedure |
|
|
Electr. Safety Test Form |
Service Mode |
|
|
|
Error List |
|
|
|
Alarm Messages |
|
|
|
Repair Information/Error Events |
|
|
|
Repair Procedure |
|
|
|
Software Download |
|
|
|
Software Versions |
|
|
Spare Parts Lists |
Schematics and Diagrams |
Function Descriptions |
Operating Instructions |
US Spare Parts List |
Schematics and Diagrams |
Function Description |
US Operating Instructions |
|
|
|
For information only: |
|
|
|
English Instructions for Use |
|
|
|
French Instructions for Use |
|
|
|
Spanish Instructions for Use |
A
B
C
O5664900T03.fm 27.08.01 |
For internal use only. Copyright reserved. |
|
|
|
1DYLJDWLRQ |
1DYLJDWLRQ WLSV |
|
|
|
1DYLJDWLRQ
1DYLJDWLRQ WLSV
Words in blue are hyperlinks (except for the Dräger logo). Clicking on these hyperlinks will open the document at the corresponding text or view.
The icons on the left margin of the page are also hyperlinks. Clicking on them will have the following effects:
Opens the start page.
Opens the table of contents.
A
B
C Opens the keyword index.
If you do not see these icons, then you have opened a "separate" file. If you want to return to the start page, just close this "separate" file. Such "separate" files are, for example, "Instructions for Use/Operating Manuals" and the "Spare Parts Lists".
Dräger Medizintechnik GmbH |
Navigation Savina Version 2.0 |
Page 2 |
A
B
C
|
|
|
1DYLJDWLRQ |
1DYLJDWLRQ WLSV |
|
|
|
If an hyperlink takes you to a figure, it may happen that only the figure number and the title appears on the upper edge of the screen. In this case you need to scroll up or select "Fit in window" from the Acrobat Reader "display" menu.
If you want to open several files in the same Acrobat Reader window (that is, a file is not closed automatically if you open another one), XQVHOHFW the "Open Cross-Doc Links in Same Window"
box in the "File/Preferences/General..." control page.
For further information on how to use the Acrobat Reader, please refer to the Acrobat Reader Online Guide in the Acrobat Reader "Help" menu.
O5664900T03.fm 27.08.01 |
For internal use only. Copyright reserved. |
Dräger Medizintechnik GmbH |
Navigation Savina Version 2.0 |
Page 3
A
B
C
27.08.01 |
only. Copyright reserved. |
D5664900TECIVZ.fm |
For internal use |
&RQWHQWV
1DYLJDWLRQ
1DYLJDWLRQ WLSV |
|
3HULRGLF 0DQXIDFWXUHU¶V 6HUYLFH 306 |
|
'HYLFH LQ JHQHUDO |
|
5HSODFHPHQW RI ZHDU DQG WHDU SDUWV |
|
(OHFWULFDO VDIHW\ WHVW |
|
)XQFWLRQDO WHVWV |
|
3ODFH IXOO\ IXQFWLRQDO XQLW DW XVHU V RZQHU V |
|
GLVSRVDO |
|
&RQILUPDWLRQ RI WHVW |
|
5HSRUW |
|
$QQH[ |
|
6HUYLFH 6WUDWHJ\ |
|
,WHPV WR EH UHSDLUHG 5HSDLU ORFDWLRQV |
|
*HQHUDO QRWHV |
|
0DLQWHQDQFH 3URFHGXUH 5HSODFLQJ ILOWHU DQG VHDOLQJ ULQJ LQ 22 FRP
SUHVVHG JDV LQOHW |
|
22 SUHVVXUH UHGXFHU |
|
5HSODFLQJ WKH 22 SUHVVXUH UHGXFHU ................... 57
Dräger Medizintechnik GmbH |
Savina Version 2.0 |
Page 4
A
B
C
27.08.01 |
only. Copyright reserved. |
D5664900TECIVZ.fm |
For internal use |
&RQWHQWV |
|
$GMXVWLQJ WKH 22 SUHVVXUH UHGXFHU ................... |
59 |
5HSODFLQJ WKH EORZHU XQLW |
|
5HSODFLQJ WKH LQWHUQDO UHFKDUJHDEOH EDWWHULHV |
|
|
|
5HSODFLQJ WKH UHDO WLPH FORFN RQ WKH |
|
&RQWURO 3&% |
|
6HUYLFH 0RGH |
|
*HQHUDO |
|
,QWHUQDO 6HUYLFH 0RGH |
|
(QWHULQJ WKH ,QWHUQDO 6HUYLFH 0RGH ................... |
74 |
/LVW RI WHVW VWHSV .............................................. |
78 |
'HVFULSWLRQ RI WKH WHVW VWHSV ............................ |
80 |
Test Step 0 ....................................................... |
80 |
Test Step 1 ....................................................... |
80 |
Test Step 2 ....................................................... |
80 |
Test Step 3 ....................................................... |
81 |
Test Step 4 ....................................................... |
81 |
Test Step 5 ....................................................... |
82 |
Test Step 6 ....................................................... |
83 |
Test Step 7 ....................................................... |
84 |
Test Step 8 ....................................................... |
84 |
Test Step 9 ....................................................... |
85 |
Test Step 10 ..................................................... |
86 |
Test Step 11 ..................................................... |
87 |
Test Step 12 ..................................................... |
87 |
Test Step 13 ..................................................... |
88 |
Test Step 14 ..................................................... |
88 |
Test Step 15 ..................................................... |
89 |
Dräger Medizintechnik GmbH |
Savina Version 2.0 |
Page 5
A
B
C
27.08.01 |
only. Copyright reserved. |
D5664900TECIVZ.fm |
For internal use |
&RQWHQWV |
|
Test Step 16 ..................................................... |
89 |
Test Step 17 ..................................................... |
90 |
Test Step 18 ..................................................... |
92 |
Test Step 19 ..................................................... |
94 |
Test Step 20 ..................................................... |
96 |
Test Step 21 ..................................................... |
96 |
Test Step 22 ..................................................... |
98 |
Test Step 23 ..................................................... |
98 |
Test Step 24 ..................................................... |
99 |
Test Step 25 ..................................................... |
99 |
Test Step 26 ................................................... |
100 |
Test Step 27 ................................................... |
100 |
3& 6HUYLFH 0RGH |
|
5HTXLUHG HTXLSPHQW ...................................... |
101 |
(QWHULQJ WKH 3& 6HUYLFH 0RGH ........................ |
102 |
([LWLQJ 6HUYLFH 0RGH ..................................... |
107 |
(UURU /LVW
6WUXFWXUH RI HUURU OLVW |
|
6RIWZDUH PDVWHU SURFHVVRU 03 HUURU JURXS
0DVWHU SURFHVVRU 03 UHDG RQO\ PHPRU\ HU URU JURXS
0DVWHU SURFHVVRU 03 5$0 HUURU JURXS
,QWHUQDO $' FRQYHUWHU VRIWZDUH 03 HUURU JURXS
,QWHUQDO FRPPXQLFDWLRQ VRIWZDUH 03 HUURU
Dräger Medizintechnik GmbH |
Savina Version 2.0 |
Page 6
A
B
C
27.08.01 |
only. Copyright reserved. |
D5664900TECIVZ.fm |
For internal use |
&RQWHQWV
JURXS |
|
6RIWZDUH )3 DQG 03 HUURU JURXS |
|
|
:DWFKGRJ 03 HUURU JURXS |
|
57& HUURU JURXS |
|
|
)URQW SURFHVVRU )3 VRIWZDUH HUURU JURXS |
3RZHU SDFN HUURU JURXS |
|
||
|
|
|
|
|
|
|
2SHUDWLQJ YROWDJHV HUURU JURXS |
|
|
520 IURQW SURFHVVRU )3 HUURU JURXS |
|
|
||
|
|
%ORZHU HUURU JURXS |
|
|
5$0 IURQW SURFHVVRU )3 HUURU JURXS |
|
|
||
|
|
&RROLQJ HUURU JURXS |
|
|
,QWHUQDO $' FRQYHUWHU VRIWZDUH )3 HUURU |
|
|
||
JURXS |
|
6\VWHP UHOLHI HUURU JURXS |
|
|
|
|
|||
,QWHUQDO FRPPXQLFDWLRQ VRIWZDUH 03 HUURU |
/RXGVSHDNHU HUURU JURXS |
|
||
JURXS |
|
|||
|
|
|||
:DWFKGRJ )3 HUURU JURXS |
|
*ROG&DS HUURU JURXS |
|
|
|
|
Dräger Medizintechnik GmbH |
Savina Version 2.0 |
Page 7
A
B
C
27.08.01 |
only. Copyright reserved. |
D5664900TECIVZ.fm |
For internal use |
&RQWHQWV
$X[LOLDU\ DODUP HUURU JURXS |
|
$PELHQW SUHVVXUH PHDVXUHPHQW ZLWK 6 HU |
|
|
|
URU JURXS |
|
$LUZD\ SUHVVXUH PHDVXUHPHQW HUURU JURXS
([SLUDWRU\ IORZ PHDVXUHPHQW HUURU JURXS
,QVSLUDWRU\ IORZ PHDVXUHPHQW HUURU JURXS
2 FRQFHQWUDWLRQ PHDVXUHPHQW HUURU JURXS
|
|
$LUZD\ WHPSHUDWXUH PHDVXUHPHQW HUURU |
|
JURXS |
|
$PELHQW SUHVVXUH PHDVXUHPHQW ZLWK 6 HU |
|
URU JURXS |
|
Dräger Medizintechnik GmbH |
Savina Version 2.0 |
2 VXSSO\ SUHVVXUH PHDVXUHPHQW HUURU
JURXS |
|
6L6: 03 HUURU JURXS |
|
6L6: )3 HUURU JURXS |
|
$ODUP 0HVVDJHV |
|
9HQWLODWLRQ PRQLWRULQJ |
|
8SSHU DLUZD\ SUHVVXUH OLPLW ........................... |
168 |
First pressure limit .......................................... |
168 |
Second pressure limit (Paw high + 5 mbar) ......... |
170 |
/RZHU DLUZD\ SUHVVXUH OLPLW ........................... |
172 |
0LQXWH YROXPH .............................................. |
175 |
|
Page 8 |
A
B
C
27.08.01 |
only. Copyright reserved. |
D5664900TECIVZ.fm |
For internal use |
&RQWHQWV |
|
MV low .......................................................... |
175 |
MV high ......................................................... |
177 |
,QVSLUDWRU\ WLGDO YROXPH ................................. |
179 |
VTi low .......................................................... |
179 |
VTi high ......................................................... |
181 |
2[\JHQ FRQFHQWUDWLRQ ................................... |
183 |
FiO2 low ........................................................ |
183 |
FiO2 high ....................................................... |
185 |
FiO2 limit values ............................................. |
187 |
&KDQJH LQ EUHDWKLQJ F\FOH ............................. |
188 |
Apnea ........................................................... |
188 |
Fail to Cycle ................................................... |
191 |
Apnea ventilation ............................................ |
193 |
+LJK UHVSLUDWRU\ IUHTXHQF\ ............................ |
194 |
$6% ! VHFRQGV ............................................ |
195 |
,QVSLUDWRU\ DLUZD\ WHPSHUDWXUH ...................... |
197 |
8VHU IXQFWLRQV .............................................. |
199 |
Standby activation ........................................... |
199 |
Flow monitoring .............................................. |
200 |
O2 monitoring ................................................ |
201 |
Mask ventilation (NIV) ..................................... |
203 |
Nebulization ................................................... |
204 |
Inspiration hold ............................................... |
205 |
0RQLWRULQJ RI VHQVRUV |
|
$LUZD\ SUHVVXUH VHQVRUV ............................... |
206 |
([SLUDWRU\ IORZ VHQVRU .................................. |
209 |
Expiratory flow sensor faulty ............................. |
209 |
Expiratory flow sensor not positioned correctly .....211 |
|
,QVSLUDWRU\ IORZ VHQVRU .................................. |
212 |
2[\JHQ VHQVRUV ............................................ |
213 |
%UHDWKLQJ JDV WHPSHUDWXUH VHQVRU DW < SLHFH . 215 |
Dräger Medizintechnik GmbH |
Savina Version 2.0 |
Page 9
A
B
C
27.08.01 |
only. Copyright reserved. |
D5664900TECIVZ.fm |
For internal use |
&RQWHQWV |
|
Breathing-gas temperature sensor faulty ............ |
215 |
Breathing-gas temperature sensor disconnected |
. 216 |
$PELHQW SUHVVXUH VHQVRUV ............................. |
217 |
'HYLFH LQWHUQDO WHPSHUDWXUH VHQVRUV .............. |
218 |
0RQLWRULQJ RI DFWXDWRUV |
|
([SLUDWRU\ YDOYH ............................................ |
219 |
0RQLWRULQJ RI 3((3 ....................................... |
221 |
([SLUDWRU\ UHVLVWDQFH 3((3 KLJK .................. |
223 |
*DV VXSSOLHV ................................................. |
226 |
Air intake ....................................................... |
226 |
Blower ........................................................... |
228 |
Speed adaptation ............................................ |
229 |
No oxygen ..................................................... |
231 |
Oxygen supply high ......................................... |
233 |
0RQLWRULQJ RI FHQWUDO V\VWHP |
|
%RRW SURFHGXUH ............................................. |
234 |
Testing of watchdogs ....................................... |
234 |
5-V voltage monitoring test ............................... |
236 |
System relief .................................................. |
237 |
Timer ............................................................ |
238 |
Settings ......................................................... |
239 |
Boot sequence ............................................... |
240 |
0HPRU\ ........................................................ |
241 |
RTC battery ................................................... |
241 |
Data retention in EEPROM and RTC RAM ......... |
242 |
RAM ............................................................. |
243 |
ROM ............................................................. |
244 |
2SHUDWLQJ YROWDJHV ........................................ |
246 |
3RZHU SDFN ................................................... |
248 |
Charge state of internal battery ......................... |
248 |
Status of internal battery .................................. |
250 |
Dräger Medizintechnik GmbH |
Savina Version 2.0 |
Page 10
A
B
C
27.08.01 |
only. Copyright reserved. |
D5664900TECIVZ.fm |
For internal use |
&RQWHQWV |
|
Power failure .................................................. |
252 |
Main switch .................................................... |
254 |
$PELHQW SUHVVXUH .......................................... |
256 |
'HYLFH LQWHUQDO WHPSHUDWXUH ........................... |
258 |
'HYLFH IDQ ..................................................... |
259 |
.H\SDG ......................................................... |
260 |
/RXGVSHDNHU ................................................. |
262 |
$X[LOLDU\ DODUP .............................................. |
263 |
'HYLFH UHVHW .................................................. |
265 |
5HSDLU ,QIRUPDWLRQ (UURU (YHQWV |
|
.QRZQ HUURUV |
|
5HZRUNLQJ WKH FRYHU RI WKH 2 VHQVRUV .......... |
268 |
:DUQLQJ PHVVDJH 1R LQW EDWWHU\ ............. |
269 |
"Starting aid" for charging of internal battery ........ |
271 |
Dräger Medizintechnik GmbH Savina Version 2.0 |
|
8QLW FDQQRW EH VZLWFKHG RII ........................... |
272 |
Remedy 1 ...................................................... |
272 |
Remedy 2 ...................................................... |
272 |
(UURU ZKHQ XSGDWLQJ IURP 6: WR |
...... 273 |
6: WR XSGDWH SURFHGXUH .................. |
274 |
3RVVLEOH HUURUV GXULQJ 2 FDOLEUDWLRQ ............. |
275 |
5HSDLU 3URFHGXUH |
|
2SHQLQJ WKH XQLW |
|
2SHQLQJ WKH RSHUDWRU FRQWURO SDQHO ................ |
278 |
5HPRYLQJ WKH UHDU SDQHO ................................ |
281 |
3RZHU SDFN |
|
0HDVXULQJ WKH VXSSO\ YROWDJHV ...................... |
285 |
0RWRU &RPPXWDWRU 3&% |
|
|
Page 11 |
A
B
C
27.08.01 |
only. Copyright reserved. |
D5664900TECIVZ.fm |
For internal use |
&RQWHQWV
9LHZ RI WKH 0RWRU &RPPXWDWRU 3&% ............... |
287 |
6RIWZDUH 'RZQORDG |
|
3LQ DVVLJQPHQW RI WKH 0RWRU &RPPXWDWRU 3&% 288 |
|
|
|
|
|
5HTXLUHG HTXLSPHQW |
|
)DQ |
|
|
|
3LQ DVVLJQPHQW RI WKH IDQ |
289 |
'RZQORDG |
|
|
|
||
|
|
'RZQORDGLQJ PHGLFDO HTXLSPHQW VRIWZDUH WR 3& .. |
|
2 'LDSKUDJP 3&% |
|
305 |
|
|
305 |
||
5HSODFLQJ WKH 2 'LDSKUDJP 3&% |
290 |
'RZQORDGLQJ ................................................ |
|
|
|
||
Pin assignment of the O2 Diaphragm PCB ......... |
293 |
%RRWVWUDS GRZQORDG |
|
|
|
||
%\SDVV YDOYH 9 |
|
5HTXLUHG WHVW HTXLSPHQW ................................ |
311 |
|
|
||
|
|
,QVWDOODWLRQ DQG XVH RI WKH ERRWVWUDS GRZQORDG SUR |
|
([SLUDWRU\ YDOYH 9 |
|
JUDP ...................................................... |
312 |
|
|
||
5HSODFLQJ WKH ((3520 |
|
6RIWZDUH 9HUVLRQV |
|
5HSODFLQJ WKH &RQWURO 3&% |
|
2YHUYLHZ RI WKH GLIIHUHQW VRIWZDUH 6: YHUVL |
|
|
|
RQV |
|
Dräger Medizintechnik GmbH |
Savina Version 2.0 |
Page 12
A
B
C
27.08.01 |
only. Copyright reserved. |
D5664900TECIVZ.fm |
For internal use |
&RQWHQWV |
|
'HVFULSWLRQ RI FKDQJHV |
|
6FKHPDWLFV DQG 'LDJUDPV |
|
)URQW YLHZ 6DYLQD |
|
5HDU YLHZ RI 6DYLQD |
|
,GHQWLILFDWLRQ RI 3&%V |
|
22 9DOYH 3&% ................................................ |
328 |
&RQWURO 3&% .................................................. |
330 |
)URQW 3DQHO 3&% ............................................ |
331 |
22 'LDSKUDJP3&% ........................................ |
333 |
3QHXPDWLF GLDJUDP |
|
6DYLQD WXELQJ GLDJUDP |
|
Dräger Medizintechnik GmbH Savina Version 2.0 |
|
)XQFWLRQ 'HVFULSWLRQ |
|
*HQHUDO |
|
:RUNLQJ SULQFLSOH .......................................... |
336 |
0RQLWRULQJ .................................................... |
337 |
3ULQFLSDO FRPSRQHQWV |
|
(OHFWURQLFV ................................................... |
339 |
Power pack .................................................... |
340 |
Control PCB ................................................... |
343 |
Motor Commutation PCB ................................. |
346 |
O2 Valve PCB ................................................. |
348 |
O2 Diaphragm PCB ......................................... |
349 |
Fan ............................................................... |
350 |
2SHUDWRU &RQWURO 3DQHO .................................. |
351 |
Front Panel PCB ............................................ |
352 |
Display .......................................................... |
353 |
|
Page 13 |
A
B
C
27.08.01 |
only. Copyright reserved. |
D5664900TECIVZ.fm |
For internal use |
&RQWHQWV |
|
Membrane keypad .......................................... |
353 |
Control knob ................................................... |
353 |
3QHXPDWLF V\VWHP ......................................... |
354 |
Ventilation function .......................................... |
358 |
O2 mix ........................................................... |
359 |
Pneumatic safety devices ................................. |
360 |
Nebulizer function ........................................... |
360 |
O2 calibration function ...................................... |
361 |
Sensors ......................................................... |
361 |
&RQYHUVLRQ ,QVWUXFWLRQV
/LVW RI &RQYHUVLRQV |
|
Dräger Medizintechnik GmbH |
Savina Version 2.0 |
Page 14
D |
Periodic |
|||
|
|
|
|
Manufacturer’s |
|
|
|
|
Service (PMS) |
DrägerService |
Savina |
|||
|
|
|
|
|
Explanation of Symbols |
|
|||
|
|
|
OK |
Z = Check condition |
|
|
|
||
|
|
|
Defect/error/fault |
F = Check function |
|
|
|||
|
|
|
Spare parts used |
D = Check for leaks |
|
|
|
Report |
P = Enter value |
∩Accessories missing
For internal use only. Copyright reserved.
File no.:
Edition
Installation site:
Serial no.:
Date of delivery/ startup:
Invoice no. or delivery no.:
Other:
5664.900
08.2001
A
B
C
SW 01.n
SW 02.n
1.Device in general
Composition of the inspection price:
Basic unit with one expiratory valve. Additional expiratory valves must be charged separately.
1.1Software version
P
Enter software version of Savina.
1.2 |
Options as per rating plate |
|
|
|
|
|
|
|
|
|
|
|
|
|
BIPAP/PCV+ |
|
|||||||||||
|
P |
|
|
|
|
|
|
|
|
|
|
|
|
|
Autoflow |
|
|
|
|
|
|
|
|
|
|
|
|
|
P |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||
|
Central alarm |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||||||||
|
P |
|
|
|
|
|
|
|
|
|
|
|
|
|
NIV (mask ventilation, SW 02.n or later) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||||||||
|
P |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1.3Operating hours
The operating hours can be read out under "Config" in the Savina menu.
1.3.1Total hours of operation
P
Enter total operating hours.
1.3.2 |
P |
Service operating hours |
Operating hours since last service.
15
A
B
C
1.4General condition
|
|
Z |
1.4.1 |
Accompanying documents |
|
|
Instructions for use and medical product |
|
|
logbook available according to operators. |
Z |
1.4.2 |
Device rear panel |
|
1.4.2.1Dust filter set
Replace dust filter, if necessary. Dust filter
must not be pressed together. |
Z |
1.4.2.2Ambient air filter
Replace ambient-air filter, if necessary. |
Z |
1.4.3Power pack
1.4.3.1DC connection for external battery.
Connection has no visible damage.
1.4.3.2Cooling air fan
Switch on the unit. Check condition and functioning of cooling air fan. Switch off the unit.
1.4.4O2 inlet block.
Inlet block is fitted with NIST, DIN or DISS connectors.
Check condition and tightness of inlet block.
1.4.5Inspiration block
Z
Z F
Z D
Check condition of inspiration block. |
Z |
1.4.5.1Inspiratory socket
Check condition of inspiratory socket. |
Z |
1.4.6Nebulizer connection
The nebulizer connection is marked with a nebulizer symbol.
1.4.7Flow sensor mount (flow sensor)
The flow sensor can be inserted into its mount and moved to the right and to the left without effort.
1.4.7.1Flow sensor
Z
Z F
The flow sensor is undamaged. |
Z |
16
A
B
C
1.4.8Expiratory valve mount
Check locking function. |
Z F |
1.4.9Expiratory valve(s)
Enter serial number of valves: Valve 1:
P
Valve 2:
P
Valve 3:
P
1.4.9.1Expiratory valve, complete See also test item 2.1.
Inspect the following components. To do so, remove the respective expiratory valve:
Diaphragm, complete Non-return valve
Flow sensor sleeve, complete
Lip seal (for pressure measurement) Expiratory socket
Water trap container
|
Valve 1: |
Z |
|
Valve 2: |
Z |
|
Valve 3: |
Z |
|
Assemble each expiratory valve ready for use. |
|
1.4.10 |
Savina front panel |
|
1.4.10.1 |
Keypad, control knob, and display |
|
|
Keypad, control knob, and display are |
|
|
undamaged (for example, have no scratches |
|
|
that could affect the view). |
Z |
17
A
B
C
1.4.11"Savina Mobil" trolley
1.4.11.1Castors
Wheels, of which 2 locking. Check screw
fittings for secure seating to 25 Nm. |
Z F |
1.4.11.2Trolley accessories
Check condition and function of the following accessories (if fitted):
Humidifier holder (optional)
|
Cylinder holder (optional) |
|
|
|
|
|
|
|
|
|
|
|
|
|
"Multiple socket outlet" retrofit kit (option) |
|
|
|
|
|
|
|
|
|
|
|
|
|
Battery pack (option, in trolley pedestal) |
|
|
|
|
|
|
|
|
|
|
|
|
|
External battery connecting cable |
|
|||||||||||
|
Z F |
|
|
|
|
|
|
|
|
|
|
|
|
1.4.12 |
Unit accessories |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||
|
Check condition and function of the following |
|
|
|
|
|
|
|
|
|
|
|
|
|
accessories (if fitted): |
|
|
|
|
|
|
|
|
|
|
|
|
|
Hinged arm (if fitted) |
|
|||||||||||
|
Z F |
|
|
|
|
|
|
|
|
|
|
|
|
|
Temperature sensor (if fitted) |
|
|
|
|
|
|
|
|
|
|
|
|
|
Z |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1.4.13Compressed gas connecting tubes
Compressed gas connecting hoses comply with national regulations.
|
For example, compressed gas connecting |
|
|
hoses with sealing rings, neutral color |
|
|
for O2 NIST connectors, 3 m, |
|
|
for O2 NIST connectors, 5 m, |
Z D |
1.4.14 |
Hose systems |
|
|
Hose systems correspond to specifications in |
|
|
the Instructions for Use/Operating |
|
|
Instructions. |
Z |
1.4.15Adult test lung
Check condition and completeness of adult test lung.
The adult test lung comprises the following components:
Face mask angle piece
Catheter connector, ISO, size 7 |
|
Breathing bag, 2 L |
Z |
The breathing bag must not be overstretched.
18
1.4.16Drug nebulizer (white housing)
Additional note:
The drug nebulizer, 84 05 000, (black housing), must not be used in Savina.
A
B
C
1.4.16.1Condition test
Check the condition of the following components:
1
2 3 4
5
6
9
7
1 Patient connection
2 Nebulizer housing
3 Gasket
4 O ring
5 Nozzle
6 O ring
7 Atomizer
8 Container
9 Nebulizer hose
8
1.4.16.2Functional test
Fill nebulizer housing with water up to the "3"
mark. Supply drug nebulizer with 2.0 bar ± 0.2 bar.
Test: Vapor must be produced.
1.4.17Special accessories (if provided)
Check condition and function of the following special accessories (if available):
Resutator 2000 Child Resutator 2000
1.5Unit options and configuration
—Not applicable —
Z
F
Z F
19
A
B
C
2.Replacement of wear and tear parts
2.1 * Replace diaphragm, complete, 84 13 661, in the expiratory valve.
Next replacement:___________________
2.2Replace internal batteries (rechargeable), 18 * 41 416, (x2) every 2 years
(see also chapter "Maintenance Procedure Replacing the internal rechargeable batteries").
Next replacement:___________________
2.3Replace batteries (rechargeable), 18 43 303,
*(if fitted, x2) in Savina Mobile every 2 years. Next replacement:___________________
2.4 |
* |
Replace O2 filter, 84 08 208, and sealing |
|
|
ring, M 09 257, in valve block, complete, |
|
|
every 6 years. |
|
|
(see also chapter "Maintenance Procedure |
|
|
Replacing filter and sealing ring in O2 |
|
|
compressed gas inlet"). |
|
|
Next replacement:___________________ |
2.5 |
* |
Replace O2 pressure reducer, 84 13 666, in |
|
|
valve block, complete, every 6 years. |
|
|
(see also chapter "Maintenance Procedure |
|
|
O2 pressure reducer"). |
|
|
Next replacement:___________________ |
2.6 |
* |
Replace real-time clock, 18 45 527, on |
|
|
Control PCB, every 6 years. |
|
|
(see also chapter "Maintenance Procedure |
|
|
Replacing the real-time clock on the |
|
|
Control PCB"). |
|
|
Next replacement:___________________ |
|
|
Note: After changing the real-time clock re- |
|
|
enter the date and time. |
2.7 |
* |
Replace the blower unit with the "motor |
blower unit" spares set, 84 13 643, after 20000 hours’ operation.
(see also chapter "Maintenance Procedure Replacing the blower unit").
20
A
B
C
3.Electrical safety test
The following steps describe the safety
Seech cksdocumentaccording" Electricalto IEC 60601SafetyandTestVDEin the USA/Canada".
0751. Any decision about performing safety checks according to VDE 0751 or IEC 60601 must be made under consideration of
Recordapplicablemeasuredn tional regvalueslatioinsthe. Electricaldocument "Electrical Safety Test Form Savina". safety checks should be carried out either
according to IEC 60601 or VDE 0751.
Savina conforms to regulations of protection class I, type B; with plugged-in airway temperature sensor to regulations of protection class I, type BF.
3.1 |
Visual inspection of basic unit |
|
|
|
Check the following parts for possible |
|
|
|
damage: |
|
|
|
− |
Power cable |
|
|
− |
Power switch |
|
|
Compare the values of the power pack fuses |
|
|
|
for mains power supply and internal |
|
|
|
rechargeable battery with the rated values on |
|
|
|
the rear panel. |
Z |
3.2Test of electrical safety to IEC 60-601
3.2.1Protective earth conductor test
Tester |
Savina |
|
L |
|
N |
6 V |
|
|
PE |
ohm |
Test probe |
PS |
Protective earth conductor test
Connect the test probe in sequence to the following test points:
−Power pack ground stud
−Screws on housing
−Rails, at the sides of Savina (if fitted)
−Oxygen connection
Test value R≤ 0.2 ohms |
P |
21
A
B
C
3.2.2Earth leakage current test
Tester |
Savina |
L/N |
|
Mains voltage |
|
N/L |
|
PE |
|
PS |
|
Earth leakage current test |
|
Normal condition (N.C.): |
|
I ≤ 500 A |
P |
Single fault condition (S.F.C.): |
|
Neutral conductor interrupted: |
|
I ≤ 1000 A |
P |
In the following steps the test is repeated, but |
|
with the power plug turned over. This |
|
condition can be created internally with some |
|
types of test devices. |
|
Normal condition (N.C.): |
|
I ≤ 500 A |
P |
Single fault condition (S.F.C.): |
|
Neutral conductor interrupted: |
|
I ≤ 1000 A |
P |
22
A
B
C
3.2.3Patient leakage current test
Tester |
Savina |
L/N
Mains voltage
N/L
PE
PS
(1)
mA
Patient leakage current test
Connect airway temperature sensor (1) to the socket of the airway temperature sensor of Savina and to the tester.
Enter serial number of the airway temperature sensor:
P
|
Normal condition (N.C.): |
|
|
I ≤ 100 A |
P |
|
Single fault condition (S.F.C.): |
|
|
Neutral conductor interrupted: |
|
|
I ≤ 500 A |
P |
|
In the following steps the test is repeated, but |
|
|
with the power plug turned over. This |
|
|
condition can be created internally with some |
|
|
types of test devices. |
|
|
Normal condition (N.C.): |
|
|
I ≤ 100 A |
P |
|
Single fault condition (S.F.C.): |
|
|
Neutral conductor interrupted: |
|
|
I ≤ 500 A |
P |
3.2.4 |
Multiple socket outlet on trolley (if fitted) |
|
3.2.4.1Compare fuses with rated values on the
socket-outlet. |
Z |
3.2.4.2Protective earth conductor test
Check all sockets. |
|
Test value R ≤ 0.2 ohms |
P |
23
A
B
C
3.2.4.3Earth leakage current test
The test is carried out without Savina or other devices.
Normal condition (N.C.): |
|
I ≤ 50 µ A |
P |
Single fault condition (S.F.C.): |
|
Neutral conductor interrupted: |
|
I ≤ 100 µ A |
P |
In the following steps the test is repeated, but |
|
with the power plug turned over. This |
|
condition can be created internally with some |
|
types of test devices. |
|
Normal condition (N.C.): |
|
I ≤ 50 µ A |
P |
Single fault condition (S.F.C.): |
|
Neutral conductor interrupted: |
|
I ≤ 100 µ A |
P |
Continue Test Certificate at test item 4.. |
|
3.3Test of electrical safety to VDE 0751
3.3.1Protective earth conductor test
Tester |
Savina |
|
L |
|
N |
6 V |
|
|
PE |
ohm |
Test probe |
PS |
Protective earth conductor test
Connect the test probe in sequence to the following test points:
−Power pack ground stud
−Screws on housing
−Rails, at the sides of Savina (if fitted)
−Oxygen connection
Test value R ≤ 0.3 ohms |
P |
24
A
B
C
3.3.2Equivalent device leakage current test
Tester |
Savina |
Rv |
L |
Mains voltage |
N |
|
|
|
PE |
mA |
PS |
|
|
|
Test probe |
Equivalent device leakage current test
Connect airway temperature sensor (1) to the socket of the airway temperature sensor of Savina and to the tester. Connect test probe to Savina.
Enter serial number of the airway temperature sensor:
P
Subsequent measurements may exceed the first-measured value by max. 50% but must at the same time be ≤ 1000 A.
Initial value: _______ A
Note:
Always enter the initial value in a new PMS.
I ≤ 1000 A |
P |
3.3.3equivalent patient leakage current test
Tester |
Savina |
|
L |
Mains voltage |
N |
|
|
|
PE |
mA |
PS |
Test probe |
|
|
equivalent patient leakage current test
(1)
(1)
25
A
B
C
Connect airway temperature sensor (1) to the socket of the airway temperature sensor of Savina and to the tester. Connect test probe to Savina.
Enter serial number of the airway temperature sensor:
P
Subsequent measurements may exceed the first-measured value by max. 50% but must at the same time be ≤ 5000 A.
Initial value: _______ A
Note:
Always enter the initial value in a new PMS.
I ≤ 5000 A |
P |
3.3.4Multiple socket outlet on trolley (if fitted)
3.3.4.1Compare fuses with rated values on the
socket-outlet. |
Z |
3.3.4.2Protective earth conductor test Check all sockets.
Test value R ≤ 0.3 ohms |
P |
3.3.4.3Equivalent device leakage current test
The test is carried out without Savina or other devices.
Subsequent measurements may exceed the
first-measured value by max. 50% but must at the same time be ≤ 100 A.
Initial value: _______ A
Note:
Always enter the initial value in a new PMS.
I ≤ 100 A |
P |
26
A
B
C
4.Functional tests
4.1Power pack
Connect mains power and O2 supply. Switch on the unit.
4.1.1Testing the "electrical supply" display
The LEDs on the front of Savina show the following operating states:
Mains power supply: green External battery: off Internal battery:
Green: Internal rechargeable batteries are fully charged.
Yellow: Internal rechargeable batteries are |
|
being charged. |
F |
4.1.2Testing the internal supply voltage Cut mains power.
The unit continues to ventilate. Audible alarm
sounds and display message "!! Int. battery |
|
activated" appears. |
F |
The LEDs on the front of Savina show the |
|
following operating states: |
|
Mains power supply: off |
|
External battery: off |
|
Internal battery: green |
F |
4.1.3Testing the supply voltage failure alarm Connect device to the mains power supply.
Remove the fuse for the internal rechargeable battery. The fuse is located on the rear panel of the unit.
Disconnect the device from the mains power supply.
Ventilation off. All LEDs are off. Audible alarm |
|
sounds. |
F |
Switch off the unit. |
|
Restore mains power. |
|
Re-insert fuse for internal rechargeable |
|
battery. |
|
Switch unit back on. |
|
27
A
B
C
4.1.4Testing the external d.c. voltage supply
4.1.4.1Battery operation simulation
Notice: Maximum test duration = 1 minute!
“Connect Charge tester (Evita), complete" to d.c. voltage input of Savina (use the two lower sockets).
Left side terminal: |
Right side terminal: |
||
(-) black |
I |
(+) red |
|
0 |
|||
|
|
Using "DC voltage source, 3 -18 V" apply a voltage of 14 V to the charge tester.
After 6 seconds, at the latest, the LEDs on the |
|
front of Savina show the following operating |
|
states: |
|
Mains power supply: green |
|
External battery: green |
|
Internal battery: |
|
Green: Internal rechargeable batteries are |
|
fully charged. |
|
Yellow: Internal rechargeable batteries are |
|
being charged. |
F |
Reduce the voltage from the DC source from |
|
14 V to 11 V. |
|
The LEDs on the front of Savina show the |
|
following operating states: |
|
Mains power supply: green |
|
External battery: yellow |
|
Internal battery: |
|
Green: Internal rechargeable batteries are |
|
fully charged. |
|
Yellow: Internal rechargeable batteries are |
|
being charged. |
F |
28
A
B
C
4.1.5On-board electrical system simulation
Short-circuit the two upper sockets on the d.c. voltage input of Savina using the "Cable for charge tester".
Set the voltage from the d.c. source to 14 V and wait for the "External battery" LED to light up green.
Reduce the voltage from the DC source to 11 V.
The LEDs on the front of Savina show the following operating states:
Mains power supply: green External battery: green Internal battery:
Green: Internal rechargeable batteries are fully charged.
Yellow: Internal rechargeable batteries are |
|
being charged. |
F |
Remove the charge tester. Switch off the unit. |
|
4.2Tests in internal service mode
Switch on Savina and, at the same time, press and hold the backlighting dim-up/dim-down key and the "Config" key until the service mode is activated (see also chapter "Service Mode Entering the Internal Service Mode").
4.2.1Checking the RS232 interface Test Step 0 : call up this test step.
Connect PC to Savina using "RS232 extension" and "RS232 adapter, crossed". Start terminal program and adjust the following:
COM 1 (or corresponding) Baud rate: 19200
Data bits: 8 Parity: None Stop bits: 1 Protocol: None
The settings on the PC must match those in the Medibus log of the Savina in the "Configure" menu.
Carry out "Send Logbook" function in internal service mode.
Test value:
The content of the error log is displayed on |
|
the PC (terminal program). |
F |
29
A
B
C
4.2.2Testing the keys
Call up test steps 1 and 2 in sequence (as of SW 1.10)
Check the functioning of all keys.
Test value:
All keys should function properly. |
F |
4.2.3Nebulizer flow test
Test Step 13: call up this test step.
Connect a flowmeter tube with an oxygen scale of up to 16 L/min to the nebulizer connection. Switch on the nebulizer valve.
Test value:
Nebulizer flow = 8 L/min to 14 L/min. |
P |
4.2.4Comparing measurement functions of the flow sensors
Test Step 14: call up this test step.
Connect (short-circuit) the inspiratory and the expiratory sockets using a ventilation hose. Set the "PRESSURE BYPASS" value to
40 mbar. Test value:
The measured inspiratory and expiratory flows |
|
must not differ by more than 13%. |
P |
4.2.5Offset voltages of the airway pressure sensors
Call up test 19. Test values:
"V_PAW_INSP" = 730 mV ± 100 mV |
P |
"V_PAW_EXP" = 730 mV± 100 mV |
P |
30