Dräger SAM 3100, SAM 3200 Instructions For Use Manual

Dräger SAM 3100 / 3200 Sampling system
Instructions for Use
Contents
1 For your safety . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
1.2 Definitions of alert icons . . . . . . . . . . . . . . . . . . . . . .4
2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
3 Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
4 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
4.1 Air quantity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
4.2 Standard system without delivery unit . . . . . . . . . . .5
4.3 System with ejector unit . . . . . . . . . . . . . . . . . . . . . .5
4.4 System with gas sampling pump . . . . . . . . . . . . . . .5
4.5 Circuit breaker (option) . . . . . . . . . . . . . . . . . . . . . . .5
4.6 Motor circuit protector (option) . . . . . . . . . . . . . . . . .6
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
5.1 Mechanical fastening . . . . . . . . . . . . . . . . . . . . . . . .6
5.2 Installing gas lines . . . . . . . . . . . . . . . . . . . . . . . . . . .6
5.3 Electrical connections . . . . . . . . . . . . . . . . . . . . . . . .7
6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
7 Wiring diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . .8
7.1 Wiring diagram for PEX 3000 . . . . . . . . . . . . . . . . . .9
7.2 Wiring diagram for Polytron 3000 / Polytron 7000 . .10
7.3 Wiring diagram for PIR 7000 / PIR 7200 . . . . . . . . . 11
8 List of materials . . . . . . . . . . . . . . . . . . . . . . . . . .12
Contents
Dräger SAM 3100/3200 3
For your safety
!
!
!

1 For your safety

1.1 General safety notes

Before using this product, carefully read the Instructions
for Use.
Strictly follow the Instructions for Use. The user must fully
understand and strictly observe the instructions. Use the product only for the purposes specified in the Intended use section of this document.
Do not dispose of the Instructions for Use. Ensure that they
are retained and appropriately used by the product user.
Only trained and competent users are permitted to use
this product.
Comply with all local and national rules and regulations
associated with this product.
Only trained and competent personnel are permitted to
inspect, repair and service the product as detailed in these Instructions for Use (see section 6 on page 8). Further maintenance work that is not detailed in these Instructions for Use may only be carried out by Dräger or personnel qualified by Dräger. Dräger recommends a Dräger service contract for all maintenance activities.
Use only genuine Dräger spare parts and accessories,
or the proper functioning of the product may be impaired.
Do not use a faulty or incomplete product. Do not modify
the product.
Notify Dräger in the event of any component fault or failure.
Safe connection to electrical devices
Electrical connections to devices which are not listed in
these Instructions for Use should only be made following consultation with the respective manufacturers or an expert.
Use in potentially explosive atmospheres
WARNING
The equipment is only approved for use in potentially explosive atmospheres when using components suitable for such areas, complete with test certificate. Risk of explosion!
Devices or components for use in potentially explosive
atmospheres, which have been tested and approved according to national, European or international Explosion Protection Regulations, may only be used under the conditions specified in the approval and with consideration of the relevant legal regulations. The devices or components may not be modified in any manner. The use of faulty or incomplete parts is forbidden. The appropriate regulations must be observed at all times when carrying out repairs on these devices or components.

1.2 Definitions of alert icons

The following alert icons are used in this document to identify and highlight areas of the associated text that require a greater awareness by the user. A definition of the meaning of each icon is as follows:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
!
Indicates a potentially hazardous situation which, if not avoided, could result in physical injury, or damage to the product or environment. It may also be used to alert against unsafe practices.
NOTICE
i
i
Indicates additional information on how to use the product.

2 Intended use

The Dräger SAM 3100/SAM 3200 is an aspirator system for continual intake of gas/air mixtures for a gas transmitter (SAM 3100) or for two gas transmitters (SAM 3200), in order to carry out measurements of toxic and/or explosive substances in hard-to-access places.
In the event of insufficient process overpressure, a partial flow
of the process gas is siphoned off from the process using a gas sampling pump with electric drive, or alternatively using an ejector.
In order to separate solid particles from the gas flow,
the gas is passed across a filter.
The sample gas is passed to the flow meter with ring
initiator (for flow monitoring) via a 3/2-way spherical valve (option for simplified feeding of test gas) and then flows into the sample chamber of the sensor or sensors.
The air is either discharged to atmosphere or fed back
to the intake point by means of the air outlet on the transmitter.
WARNING
During installation without return feed for the sample gas back to the intake point, the sample gas should be guided to a vent or outside. Risk of poisoning or explosion! The return feed must not be pressurised.
4 Dräger SAM 3100/3200
Specification
!
Example arrangements
Dräger SAM 3100 without gas sampling pump:
1 Dräger PIR 7000 gas transmitter with process adapter 2 Sample gas outlet 3 Flow meter with ring initiator (flow monitoring at 25 to 250 l/h) 4 3/2-way calibrating gas valve 5 Analysis filter (filter element 1 µm PTFE) 6 Sample gas inlet 7 Dräger Polytron 7000 gas transmitter with adapter for
gas exposure
8 Gas sampling pump
Dräger SAM 3200 with gas sampling pump:

3 Specification

Ambient temperature: +10 to +40 °C Protection class: IP 20 (standard
Flow meter with ring initiator
Flow monitoring:
Standard: 10–100 l/h Optional: 6–60 / 25–250 l/h
Material: PVDF
Sample gas lines
Material: Screw connections:
Gas sampling pump (option)
Power supply: 230 V / 50 Hz Pumping rate: approx. 5.5 l/min.
Ejector (option)
Material: Stainless steel Compressed air supply: 4 bar to 6 bar
configuration without housing)
PVDF Hose: PTFE

4 Settings

4.1 Air quantity

The flow should be set according to the sensor type:
Transmitter with electrochemical sensor ~60 l/hTransmitter with electrochemical sensor + IR transmitter
in series ~80 l/h
IR transmitter ~180 l/h
The flow should be matched to the circumstances in the system as a whole.

4.2 Standard system without delivery unit

The sample air leaves the process with excess pressure!
The sample gas flow should be set to 60 l/h at the fine
regulating valve on the flow meter. The top of the ball is the read mark for the scale.
Notice on handling: see the Instructions for Use for the
flow meter.

4.3 System with ejector unit

To function correctly, the regulating valve needs an input
pressure of 1 bar to 10 bar. The control range of the regulating valve is 0.2 bar to 10 bar.
Setting the sample gas flow:Open the fine regulating valve on the flow meter (turn anti-
clockwise to the stop, then 1 turn back).
Set the flow approximately at the regulating valve, then carry
out the fine regulation using the regulating valve on the flow meter . The top of the ball is the read mark for the scale.
Notice on handling: see the Instructions for Use for the
flow meter and for the regulating valve.

4.4 System with gas sampling pump

The sample gas flow should be set to 60 l/h at the fine
regulating valve on the flow meter. The top of the ball is the read mark for the scale.
Setting the sample gas flow:Open the fine regulating valve on the flow meter (turn anti-
clockwise to the stop, then 1 turn back), set the flow approximately at the bypass valve on the gas sampling pump, then carry out the fine regulation using the regulating valve on the flow meter. The top of the ball is the read mark for the scale.
Notice on handling: see the Instructions for Use for the
flow meter and for the gas sampling pump.
Regulating valve (option)
Input pressure: 1 bar to 10 bar Control range: 0.2 bar to 10 bar
3/2-way spherical valve (option)
Material: Stainless steel
WARNING
The equipment is only approved for use in potentially explosive atmospheres when using components suitable for such areas, complete with test certificate. Risk of explosion!
Test certificates are included in the equipment documentation.

4.5 Circuit breaker (option)

The circuit breaker is delivered together with the aspirator
system as a separate item. Installation is carried out on site by the operator in accordance with the assembly instructions. Information is included in the documentation.
The circuit breaker makes it possible to monitor line breaks
and short circuits. Furthermore, the action of the relay can be reversed by means of a switch. Settings should be adjusted in accordance with the operator's requirements during commissioning. Information is included in the documentation.
Standard: Auxiliary power 24 V DCOptional: Auxiliary power 230 V, 50 Hz
Dräger SAM 3100/3200 5
Loading...
+ 11 hidden pages