Cylinder high-pressure connectors:
● 2216 psi connector to CGA 346
● 4500 psi connector to CGA 347
● Quick-connect cylinder coupling (2216 psi or 4500 psi)
RIC UAC Connector:
● 2216 psi or 4500 psi, male, quick-release coupling with pressure relief
valve
Lung demand regulator to face mask connector:
● Dräger push-in connector
EOST alarms – Activation commencement range (mechanical and
electronic):
● 2216 psi cylinder: 819 psi to 732 psi
● 4500 psi cylinder: 1665 psi to 1485 psi
Gauge power supply:
● Battery: Panasonic CR123A (3 V lithium)
● Voltage: 6 Vdc
9 Special instructions
9.1 Use of an independent air supply (supplied airline
respirator (SAR) connection)
!
WARNING
Air quality must conform to the statutory requirements.
Use of an Emergency Breathing Support System (EBSS) must
comply with NIOSH and NFPA 1981 requirements.
The time required for the wearer to escape to a safe area must
be within the remaining breathing time of the cylinder, taking into
account the remaining air content in the cylinder and the breathing
rate of the wearer.
Independent air supplies must meet the following standards:
● Type-1 gaseous air as dened in: CGA Commodity Specication for
Air, G-71 (grade D or higher)
● NFPA 1989 Standard on Breathing Air Quality for Fire and Emergency
Services Respiratory Protection
● Air supply pressure: 87 psi to 125 psi
● Airline hose length: 5 feet to 300 feet (maximum working hose length
must not exceed 12 individual hose lengths)
● Airline ow rate: 550 liters/minute
● Cold temperature approval of -25 °F (-31.7 °C)
i
i
NOTICE
The TX Gauge warning signals will operate as normal. The
TX Gauge will show cylinder pressure when the cylinder valve is
open.
1. Turn on the independent air supply.
2. Connect the independent air supply coupling to the secondary supply
hose (see the EBSS/SAR Connection Instructions for Use) and
breathe normally.
3. Close the cylinder valve (if the whistle sounds, silence it by taking
several short deep breaths or momentarily operating the lung demand
regulator bypass button (Fig 3, Item 3).
4. If any air supply problems are encountered, proceed as follows:
a. Open the cylinder valve to return to breathing from the attached
cylinder.
b. Disconnect the independent air supply coupling.
c. Leave the hazardous area by the shortest and safest escape
route, if necessary.
9.2 CBRN use
The Dräger PSS 5000 Series is certied by National Institute for
Occupational Safety and Health (NIOSH), for chemical, biological,
radiological or nuclear (CBRN) use and by the Safety Equipment Institute
(SEI) to meet the requirements of NFPA 1981. Approvals are only valid
when the apparatus is used with compressed-air cylinders approved
by NIOSH. Equipment congurations for CBRN use are detailed in a
CBRN Approval Label (see CBRN Approval Label 3361124).
Dräger recommend that a quantitative t test (QNFT) be performed on
the face mask before use in a CBRN environment. The t test must be
conducted strictly in accordance with the requirements outlined in the
OSHA Respiratory Protection Standard 29 CFR, Section 1910.134.
9.3 Cautions and limitations
CAUTIONS AND LIMITATIONS
D – Air-line respirators can be used only when the respirators are supplied
with respirable air meeting the requirements of CGA G – 7.1, Grade
D or higher quality.
E – Use only the pressure ranges and hose lengths specied in the
USER'S INSTRUCTIONS.
I – Contains electrical parts that may cause an ignition in ammable or
explosive atmospheres.
J – Failure to properly use and maintain this product could result in injury
or death.
M – All approved respirators shall be selected, tted, used, and maintained
in accordance with MSHA, OSHA, and other applicable regulations.
N – Never substitute, modify, add, or omit parts. Use only exact replacement
parts in the conguration as specied by the manufacturer.
O – Refer to User’s Instructions and/or maintenance manuals for
information on use and maintenance of these respirators.
S – Special or critical User’s Instructions and/or specic use limitations
apply. Refer to User’s Instructions before donning.
CAUTIONS AND LIMITATIONS – CBRN
Q – Use in conjunction with personal protective ensembles that provide
appropriate levels of protection against dermal hazards.
R – Some CBRN agents may not present immediate effects from exposure,
but can result in delayed impairment, illness or death.
3362443
© Dräger Safety UK Limited
Edition 10 – February 2017 (Edition 01 – July 2013)
Subject to alteration
Draeger Safety UK Limited
Ullswater Close Tel +44 1670 352891
Blyth, NE24 4RG Fax +44 1670 356266
United Kingdom www.draeger.com
T – Direct contact with CBRN agents requires proper handling of the
SCBA after each use and between multiple entries during the same
use. Decontamination and disposal procedures must be followed.
If contaminated with liquid chemical warfare agents, dispose of the
SCBA after decontamination.
U – The respirator should not be used beyond 6 hours after initial exposure
to chemical warfare agents to avoid possibility of agent permeation.
EBSS – EBSS Activation or engagement of EBSS in either the donor
or receiver mode changes the SCBA use to Escape-Only, approved
service time for either the donor, or the receiver is no longer applicable.
Additional critical cautions and limitations apply. Refer to the section
EBSS in the users' instructions.
9.4 S – Special or Critical Users instructions
● The EOSTI alarm set point of this SCBA is 35 ± 2 % of the rated
cylinder pressure. For the activation range see Section 8 (Technical
Data).
● Cold temperature approval of -25 °F (-31.7 °C).
● When used as a combination supplied-air respirator/self-contained
breathing apparatus (SAR/SCBA), not more than 20 percent of the air
supply can be used during entry.
● During supplied air use, the cylinder valve must remain closed. If the
supplied air fails, open the cylinder valve and immediately proceed to
fresh air.
● Supplied air source must meet the following criteria: pressure 87 psi to
125 psi, air ow rate at least 550 liters/minute.
Important Note: If it is decided to exit the working area with the airline
disconnected or, in an emergency, if the air supply fails, breathe
normally and immediately proceed as follows:
● Open the cylinder valve (counterclockwise) slowly, but fully, and
breathe normally.
● Disconnect the hose of the independent air supply from the male
coupling of the airline hose connection. Breathe normally and
immediately leave the hazardous area by the shortest and safest
route.
The remaining duration begins from the time of opening the
cylinder valve and disconnecting the independent air supply. The
time required to allow the wearer to escape to a safe area must be
within the remaining air capacity (volume) of the cylinder taking into
account the breathing rate of the wearer.
● Use of the RIC UAC should be by trained and competent personnel
only.
● The RIC UAC must only be used to recharge a cylinder in emergency
situations as dened in NFPA 1981.
● The RIC UAC lling hose is a component of the NFPA 1981 certication.
Only use a lling hose which has been certied to NFPA 1981 for use
in immediately dangerous to life or health (IDLH) atmospheres.
● Can not use the RIC UAC connection for second person (EBSS).
● Can not use the RIC UAC to transfer air from one compressed-air
breathing apparatus to another.
● Do not allow oil, grease or other contaminants to contact the RIC UAC
connection.
● Do not attempt to disassemble or repair the RIC UAC connection.
● Caution: The secondary air supply pressure to the RIC UAC must not
exceed maximum rated working pressure of the cylinder(s) being lled.
● Caution: If the pressure relief valve of the RIC UAC is activated, the
SCBA must be returned to the nearest Dräger branch or agent.
● Caution: If a leak is detected while refilling in a contaminated or
oxygen-deficient gaseous atmosphere, stop refilling and immediately
leave the hazardous area.
9.5 EBSS – Special or Critical Users instructions
Safety Warning: Use of an Emergency Breathing Support System
(EBSS) must comply with NIOSH and NFPA 1981 requirements.
Refer also to the Instructions for Use for the EBSS.
● EBSS may not be engaged or activated in donor mode after the donor
End-of-Service-Time-Indicator (EOSTI) has activated.
● Users must be fully trained in the operation of EBSS in accordance
with a training program conforming to the requirements of NFPA
Standards 1404, Fire Service Respiratory Protection Training and
1500, Fire Department Occupational Safety and Health Program.
● Simultaneous connection of more than two users, one donor, and one
receiver, is not permitted.
The time required to allow the wearers to escape to a safe area must
be within the remaining air capacity (volume) of the cylinder taking into
account the breathing rate of the wearers.
10 Warranty information
Unless otherwise agreed between Dräger and the customer, the following
shall apply in the event of defects of the product in material or workmanship:
The customer shall contact the company where he bought the product
(“Seller”). The warranty conditions agreed between the customer and the
Seller shall apply. The product must be used in strict accordance with the
Instructions for Use. Any use disregarding the Instructions for Use may
void warranty.
11 Contact details
Any issues with the equipment, including damage, malfunction, or failure
of the breathing apparatus that may present a hazard to the user should
be reported to Dräger US Customer Service – Phone 1-800-437-2437.
Contact with the certication organizations may be reached at:
NIOSH, NPPTL - Phone 1-412-386-4000
SEI (NFPA) - 1307 Dolley Madison Blvd, Suite 3A, McLean, VA 22101
Phone 1-703-442-5732