• Use only Dräger sensors which are designed to be used with the
Polytron 2 XP Tox transmitter.
• When the sensor is installed, the electronics of the transmitter automatically
adjusts to the operating parameters of the sensor.
A sensor which has been calibrated within the calibration interval need not be
re-calibrated at this time. See sensor data sheet.
When the calibration interval has expired, a warning message is displayed.
If the newly installed sensor is the same type (identical part number to the one
before), the transmitter will keep its configuration. Otherwise, the transmitter will
show the message ‘connect sensor xxxxxxx’, where xxxxxxx is the part number of
the previously connected sensor. (See also Section 5.5, Switching Sensor Lock
On/Off).
• Remove bayonet ring from sensor enclosure; remove cover plate.
• Pull out cable with connector.
• Take sensor out of packaging. Make sure that the sensor is compatible to
Polytron 2 XP Tox.
18
cable with connector
cover plate
bayonet ring
Page 4
Technical Information
Signal Transmission to Central Control Unit
Voltage of Power Supply
Physical Specifications
6Technical Information
6.1Approvals
UL, CSA ......................................................................................................... Class I, Div 1, Groups B, C, D
ATEX ...................................................................................... II 2G EEx d [ia] IIC T6 (Ta = -40 to +65°C)
II 3D EEx d [ia] IIC T85 (Ta = -40 to +65°C): IP66
CE marking .................................................... Electromagnetic Compatibility (Directive 89/336/EEC)
6.2Signal Transmission to Central Control Unit
AnalogTransmission by 3-core shielded cable
– Measurement current ........................................................................................................ 4 mA to 20 mA
– Measuring range not reached ........................................................................................ 3.8 mA to 4 mA
– Measuring range exceeded ....................................................................................... 20 mA to 20.5 mA
– Transmitter fault ............................................................................................................................. < 3.2 mA
Calibration System #2 - for calibration with gas cylinder
6806978Calibration Adapter
Consult your domestic Dräger subsidiary, agent, or distributor for availability.
8.2Replacement Parts
Order #Description
4543196Polytron 2 XP Tox Bucket Assembly
4543197Polytron 2 XP Tox Bucket Assembly with relays
4509314Calibration Adapter
4543198Galvanic Isolator
4543199Analog Front End
4543201Sensor Housing Assembly