Installation, Operation and Care Manual
EverLight® 1
Section I Introduction
EverLight
®
This manual contains installation, operation and care
instructions and user service information for the
EverLight® LED light.
The EverLight is designed to provide trouble-free
service when installed, operated and maintained
according to the procedures set forth in this manual.
To ens ure correct track, ceiling, wa ll or ca binet
mounting installation, carefully read all the instructions contained in this manual paying close attention
to all notes, notices, cautions and warnings.
After the EverLight is installed, please review the
operation procedures and care guidelines with the
doctor's staff. Then!"#$%#!&'()!*$+,$"!(+!&'#
-./&.01)!.22(/# for future reference.
Specifications
3.4#0!5,66"78 Primary: 100-240 V AC 50/60 Hz.
Secondary: 12V DC
96#0$&(+:!;*6#0$:#8! .4 max. - .16 min. Amp
<./$"!=#+:&'8 27"
C Type of protect ion against electri c shock: Cl ass 1
Equipment
C Degree of protection against the ingress of water:
Ordinary
C Equipment not suitable for use in the presence of
a flammable anesthetic mixture with air or with
oxygen or nitrous oxide.
C Mode of operation: Continuous
C Environmental conditions recommended ranges for:
Transport and Storage:
D Temperature range of - 10º C to + 70º C
D Relative humidity range of 10% to 100%
D Atmospheric pressure range of 500 to 1060 hPa
Normal Use:
D Temperature range of + 10º C to + 40º C
D Relative humidity range of 30% to 75%
D Atmospheric pressure range of 700 to 1060 hPa
Explanation of Symbols used on Equipment:
=Attention, Consult Accompanying Documents
>.".0!?#*6#0$&,0#8 High: 5300° Kelvin
(Color Matching)
Medium: 4300° Kelvin
(Operative Setting)
Low: 3200° Kelvin
(Composite Setting)
=(:'&!@+&#+)(&78High: 2300 Footcandles
(25,000 Lux)
Medium: 1375 Footcandles
(15,000 Lux)
Low: 725 Footcandles
(8,000 Lux)
=(:'&!3$&�+8!!6-1/2" x 3-1/2"
A+(&!B#(:'&8 Track Mount:87 lbs.
Ceiling Mount: 41 lbs
.
Wall Mo un t: 59 lbs.
abinet Mount: 51 lbs.
C
Classifications
Class 1, Medical Equipment
for Ordinary Use
with Respect to Electric Shock,
Fire and Mechanical Hazards
53HN
Only in Accordance with
UL 60601-1, IEC 60601-1,
CAN/CSA 22.2 No. 601.1
equipment was tested and complies with
Class A digital device limits. These limits are
designed to give equipment reasonable
protection against detrimental interference
when operated in a commercial environment.
precautions regarding electromagnetic compatibility (EMC) and needs to be installed
according to EMC information. (See EMC
Information Pgs. 27 and 28.)
equipment can effect medical electrical
equipment.
2Installation, Operation and Care Manual
Section I Introduction
Dimensions
24"
1.61"
1.25"
15"
17"
14.375"
9"
3.5"
Tr a ck M o u n t
1.375"
24"
7.25"
14.375"
15"
17"
9"
31.875"
Ceiling Mount
EverLight® 3
Section I Introduction
24"
EverLight
®
14.375"
15"
17"
9"
4.5"
Wall/Cabinet Mount
1.375"
56"
4Installation, Operation and Care Manual
Section I Introduction
Light Head Range of Motion
145º
145º
150º
180º
Packaging
Verify the EverLigh t compone nts packagi ng conten ts
against the appropriate mounting configuration's
packing list.
!"#$:Be aware that some sections of the carton
may be packed as fillers and might be empty (even if
optional components are ordered).
— NOTICE —
For any questions about an order, please
contact a DentalEZ Equipment customer
service representative at %&'((&)#$&*!+"
,%&'((&-'-&.(-(/.
Carefully unpack the light. Inspect all of the items
received to assure there has been no shipping damage.
*0 123435617812789:;6<62=19:>?39?1?@6193<<76<
7446273?6ABC1#@6>1<6D:<?13>2107A61?@613DD<:D<7&
3?619A37481E7?@1?@6193<<76<1E7?@7>1F.1@:G<8130?6<
26A7;6<B. (See Product Support Services, Page 25.)
— IMPORTANT NOTE —
All parts supplied are vital for proper installation.
):1>:? discard any hardware or components
until installation is checked and complete.
EverLight® 5
Section II Installation
EverLight
®
Tr ack M oun t A s se m bly
To o l s R e q u i r e d
H Phillips-head ScrewdriverH Torpedo Level
H Set of Allen WrenchesH Power Drill
H Stripping/Crimping toolH Drill Bits
H Pipe Cutter/HacksawH Tape Measure
H Lubricant: Silicon/Grease Product
Track Assembly Installation
— NOTICE —
Because of the size and weight of the track
mount assembly, two people are required
for installation.
!"#$I1The ceiling structure
supporting 200 pounds. Additional bracing to the
ceiling structure may need to be provided.
!"#$I1The mounting lag bolts
minimum of 2-1/2" of solid wood.
4G8?
be capable of
4G8?
go into a
%.After determining the location of the track
light (see illustration below), arrange for the
hole for the power cord to be above the head
of the chair.
F.Drill four 5/16" mounting holes and a 1-1/2"
power supply hole.
— NOTICE —
A full-size track mount template is available.
Order PN:1F(JJ&KLK.
-. Remove the track light assembly from the
packaging and lay it flat with mounting side
down on the floor or sturdy table.
1..Using a Phillips-head
screwdriver, remove
and retain the four
screws that hold the
side covers to each
end of the wood
mounting pallet.
Screw
Side
Cover
Power Connection
This End
Floor
Suggested Track Mount Location
End of Track
77"
53"
71"
Extended-length
post assembly
required for
ceiling heights:
8' 1" to 10'
8' 0" Min.
6Installation, Operation and Care Manual
Section II Installation
!"#$I1Refer to the exploded illustration below
during track light installation.
L.Loosen, but do not remove, two Phillips-
head screws from one end cap of the track
mount assembly.
(C After loosening the end cap (do not remove
end cap), take off the side covers and set
them aside.
Side Covers
Power Supply
Housing Cover
Side
Cover
Support Hub
End Cap
Phillips-head
Screws
Wood
Mounting
Pallet
End Cap
Tro l ley
Trolley Cover
Cut End
Machined End
Stop Collar
Drop Post
Swivel Key
Swivel Sleeve
Rigid Arm
Spindle
Light Arm Assembly
EverLight® 7
Section II Installation
EverLight
®
J.Remove the four screws from the power
supply housing cover, then take off the cover
and set the screws and cover aside.
Power Supply
Housing Cover
Screws
'.Remove the template from the ceiling.
M.Mount the track assembly to the ceiling using
the four 1/4 x 3" lag bolts supplied. (Only if
needed, slide the supplied spacers over the four
lag bolts before mounting the track.)
Power Supply Wiring
%.Make sure the appropriate electrical line
(100-240 VAC, 50/60 Hz) is properly
supplied to the power supply housing.
— NOTICE —
All wiring 4G8? be in accordance with local
electrical codes.
!"#$I1If possible, install a wall switch in the
power source to the track light. This can be turned off
to eliminate wear of the power supply and protect
against electrical surges.
— WARNING —
Before proceeding, verify the power at the
main circuit is "++.
1/4 x 3"
Lag Bolt
If trolley removal is necessary:
%.Remove the end cover and the stops on the end
of the track opposite the power supply
housing.
F.While supporting the trolley, roll it off the track
and remove the strain relief from the bracket.
-.To reassemble, reverse the procedure making
sure the rounded wheel goes into the grooved
side of the track.
!"#$I1There is a hole in the wood pallet and the
power supply housing for routing the electrical supply.
F.Pull the power supply cord through the
ceiling and hole in the wood mounting pallet.
Power Supply
Housing
-.Connect the supply to the black white and
green leads in the power supply housing.
Woo d Mo unti ng
Pallet
8Installation, Operation and Care Manual
!"#$I1When connecting to a 115V power source,
connect the wires accordingly: ground to green, neutral
to white and black to hot.
Section II Installation
4.Tuck the wires in neatly and replace the power
supply housing cover using the retaining
screws previously removed.
5.Replace the side covers using the four
Phillips-head screws previously removed.
6.Tighten the two Phillips-head screws on the
loosened end cap.
Drop Post Installation
NOTE: The track light is designed to accommodate
up to a 10' ceiling. The drop post
length
chart below for the correct post length for various
ceiling heights.
for the appropriate ceiling height. Refer to the
Ceiling Height 8'8'-6" 9'9'-6"10'
Post Len g th12" 18"24" 30"36"
must be cut to
NOTE:This hole is for the 5/16" bolt that secures
the drop post to the support hub.
4.Route the pigtail from the trolley through the
drop post
5.Insert the drop post into the support hub and
tighten the 5/16" bolt through the drop post
and let the post rest on the bolt.
— CAUTION —
Failure to install the bolt as described may
allow the light to fall.
6.Tighten the eight set screws in the support
hub and make sure the post is plumb from
front to back and side to side.
7.Slide the cover and cover support up the
drop post and tighten the set screw in the
cover support to prevent the cover from
falling.
1.Remove the three screws from the lower end
of the drop post assembly. Then take out the
swivel sleeve and swivel key.
— CAUTION —
Do not cut the post at the end containing
the three machined screw holes.
2.Measure the post from the end containing the
three machined screw holes and cut to length
according to the chart above.
NOTE:Remove any sharp edges.
— CAUTION —
Do not drill into the end of the post
containing the three machined screw holes.
3.At a distance of 1-5/8" from the cut end of
the post, drill an 11/32" (.343) diameter hole
on the center line of the post.
8. Lightly lubricate the inside of the swivel
sleeve.
NOTE:The swivel sleeve and swivel key must face
towards the toeboard of the chair to prevent the light
arm from rotating a full 360°. (See illustration below.)
Button-head Screws
Swivel
Key
Swivel
Sleeve
9.Correctly orient and place the swivel sleeve on
the rigid arm spindle and insert the swivel
key in the slot to secure the sleeve and
spindle.
10.After determining the proper wire length,
push any excess wiring into the post and
connect the harnesses.
EverLight® 9
Section II Installation
Drop Post
Spindle with
Swivel Sleeve
and Swivel Key
11.Insert the spindle with the swivel and swivel
key into the drop post with the key facing
towards the foot of the chair.
12.Align the three holes of the swivel sleeve and
the drop post and secure with three
button-head screws.
EverLight
Flex Arm Height Adjustment
1.An adjustment can be made to the flex arm
travel in the event the flex arm contacts the
track or the ceiling.
2.To access the adjustable stop, remove the
four screws that hold the arm cover.
3.Remove one of the arm end caps. These have
been attached with a small amount of glue
and require care in removing.
4.Slide the arm cover toward one end of the
arm to expose the compression spring
retaining mechanism.
5.The adjustable stop is located on the spring
rod directly behind the friction block and
between the spring stopper.
Stop
®
Secure with
three button-
head screws
13.Check the level of the post and readjust if
necessary.
14.See "Testing" section on this page.
6.Loosen the two retaining set screws in the
top of the block and position the stop to
limit the vertical travel of the arm to the
desired position. Then tighten the set screws.
7. Slide the arm cover back in position and
attach using four retaining screws.
8.Replace the end cap using a small amount of
adhesive.
Te s t i n g
1.Before proceeding, verify that the power to
the main circuit breaker panel is ON.
2Move the toggle switch, located immediately
behind the light head, UP to turn the light ON.
10Installation, Operation and Care Manual
Section II Installation
Ceiling Mount Assembly
To o l s R e q u i r e d
H Phillips-head ScrewdriverH Bubble Level
H Set of Allen WrenchesH Power Drill & Bits
H Stripping/Crimping toolH Hammer
H Pipe Cutter/HacksawH Tape Measure
H Lubricant: Silicon/Grease Product
— NOTICE —
For any questions about an order, please contact
a DentalEZ Equipment customer service repre-
sentative at %&'((&)#$&*!+".
Suggested Ceiling Mount Location
25-1/2"
Framing & Mounting
— IMPORTANT NOTE —
Before any framing is done, 3AE3B8 obtain the
doctor's approval.
%.Adjust the chair back to its full1)"N! posi-
tion and the seat to normal operating position.
F.For a suggested ceiling mount location, refer
to the illustration below.
!"#$I1Frame locations may be varied to suit
individual preferences. But, be aware of the following
framing considerations.
H Check ceiling structure and assure it will
support a minimum of 200 pounds.
25-1/2"
6"
Left-Hand Operation
6"
Right-Hand Operation
EverLight® 11
Section II Installation
EverLight
®
H Ensure no lateral movement of the light
column is possible.
H Check for an electrical circuit near the
support area of the light.
-.Using the full-size ceiling mount template in
the back of this manual, drill three holes for
the ceiling mount plate.
Ceiling Joists
Nails or
Lag Screws
..Attach the ceiling mount plate to the framing
joist using three bolts that are 3/8" diameter
or larger and appropriate for the framing
construction. (The bolts are not supplied.)
!"#$I1For suspended ceiling installation, spacers are
provided to use between the frame and the ceiling mount plate.
Power Supply Wiring
%.Make sure the appropriate electrical line
(100-240 VAC, 50/60 Hz) is properly
supplied to the junction box.
F.Remove the junction box cover.
— NOTICE —
All wiring 4G8? be in accordance with local
electrical codes.
— WARNING —
Before proceeding, verify the power at the
main circuit breaker panel is "++.
-.Route the power supply wiring through the
hole in the ceiling mount plate.
Wiring
Hole
Power
Supply
Mounting Light Support Above
Suspended Ceiling
Steel Bar
Framing
Mounting Light Support At
Suspended Ceiling Level
Steel Bar
Joist
Framing & Cross
Brace
Joist
Junction
Box
..Secure the power supply wiring to the
junction box with strain relief according to
local electrical codes.
Power Supply
Ground
L.Secure the incoming ground wire to the
junction box ground screw using a ring
terminal from the bag of supplies.
Ceiling
Mount Plate
Junction Box
Ground
Screw
12Installation, Operation and Care Manual
Section II Installation
Drop Post Installation
!"#$I1The ceiling light is designed to accommo-
up to an 10' ceiling. The drop post
date
9G?1?:1A6>5?@
Refer to the chart below for the correct post length for
various ceiling heights.
%.Remove the three screws from the lower end
of the drop post assembly. Then take out the
swivel sleeve and swivel key.
for the appropriate ceiling height.
CEILING DROP POST
HEIGHTLENGTH
10'36"
9' 6"32"
9'26"
8' 6"20"
8'14"
— CAUTION —
):1>:?19G? the drop post at the machined end
that contains the swivel sleeve screws.
F.Measure the post from the end containing
the three machined screw holes and cut to
length according to the chart above.
!"#$I1Remove any sharp edges.
4G8?1O6
L.Insert the cut end of the drop post into the
support hub and tighten the 5/16" bolt
through the drop post and let the post rest on
the bolt.
— CAUTION —
+37AG<61?:17>8?3AA1?@61O:A?13812689<7O62143B
3AA:E1?@61A75@?1?:103AAC
Ceiling Mount Plate
Ceiling Mount Collar
Drop Post
Swivel Key
Swivel Sleeve
Ceiling Mount Cover
Ceiling Stop Collar
Cut End
Machined
End
— CAUTION —
):1>:?12<7AA into the machined end of the
tube that contains the swivel sleeve screws.
-.At a distance of 1-3/4" from the cut end of
the post, drill an 11/32" (.343) diameter hole
on the center line of the post.
!"#$I1This hole is for the 5/16" bolt that secures
the drop post to the support hub.
..Route the light cord from the power supply
through the cut end of the drop post . Then
pull the cord through the machined end.
Light Arm
(.Using the bubble level make sure the post is
plumb from front to back and side to side. If
needed, adjust the eight set screws in the ceiling
mount collar.
— CAUTION —
All 8 set screws 4G8?1O61?75@? to prevent the
light from falling!
J.Tighten the eight set screws in the ceiling
mount collar.
EverLight® 13
Section II Installation
EverLight
®
8. Slide the cover and cover support up the drop
post and tighten the set screw in the cover
support to prevent the cover from falling.
9.Lightly lubricate the inside of the swivel sleeve.
NOTE:The swivel sleeve and swivel key must face
towards the toeboard of the chair to prevent the light
arm from rotating a full 360°. (See illustration below.)
Button-head Screws
Swivel
Key
Swivel
Sleeve
10.Correctly orient and place the swivel sleeve on
the rigid arm spindle and insert the swivel
key in the slot to secure the sleeve and
spindle.
Flex Arm Height Adjustment
An adjustment can be made to the flex arm to limit
the vertical travel of the arm.
NOTE: If flex arm height adjustment is needed for
the ceiling mount, refer to the adjustment instructions
for the track light found on Page 10.
Te s t i n g
1.Before proceeding, verify that the power to
the main circuit breaker panel is ON.
2Move the toggle switch, immediately behind
the light head, to the left and turn the light ON.
11.Connect the wiring harness from the rigid arm
to the light cord harness in the drop post.
12.Insert the spindle with the swivel and swivel
key into the drop post with the key facing
towards the foot of the chair.
13.Align the three holes of the swivel sleeve and
the drop post and secure with three buttonhead screws.
14.Check the level of the post and readjust if
necessary.
15.Pull the excess cord through the junction box
and secure with strain relief bushing to the
junction box.
16.After the wiring connections are complete,
replace the junction box cover.
17.See "Testing" section on this page.
14Installation, Operation and Care Manual
Section II Installation
Wall Mount Assembly
To o l s R e q u i r e d
H Power Drill & Bits
H Slotted Screwdriver
H Phillips-head Screwdriver
H Wire Crimping Tool
H Tape Measure
H Wire Stripper
H Set of Allen Wrenches
H Lubricant: Silicon/Grease Product
Wall Mount Positioning
*PQ"R#S!#1!"#$I1Check wall structure and
assure it will support a minimum of 200 pounds.
%.Place the chair back in its full )"N! position
and the seat in the normal operating position.
F. For a suggested wall mount location, refer to
the illustration below and follow this guide:
3.First identify the center point of the long
arm rotation. This point should align with
the chair's center line and off-set 25-1/4"
from the front edge of the chair's base
plate in the direction of the headrest.
O.From this point, swing a 60" radius arc,
representing the long arm rotation, until
an intersection with the wall is found.
Wall Mount Installation
%.After determining the proper position of the
wall mount, use the wall mount template in
the back of this manual to align mounting
holes on two wall studs.
F.Depending on the size of wall fastener
chosen, drill four mounting holes into the
wall studs. Then drill a 1-1/2" center hole for
the incoming power source.
!"#$I1Make sure the required power supply wiring
runs from inside the wall to the proper location.
Mounting
Bracket
Mounting
Bracket
Right-hand
Operation
Left-hand
Operation
Suggested Wall Mount Location
WALL
R60"
Optional
Right-hand
Position
25.25"
WALL
Optional
Left-hand
Position
WALL
63"
FLOOR
EverLight® 15
Section II Installation
EverLight
®
— NOTICE —
All wiring 4G8? be in accordance with local
electrical codes.
Wall Mount
Cover
Mounting Holes
-.On the wall mount, remove and retain the
four screws that secure the cover. Then take
off the cover.
To p H o l e f o r
Light Arm
Screws
Nut
Black
Wire
Harness
From Light
Mounting
Bracket
Power
Supply
Harness
White
Wire
Green/Yellow
Wire
%K.Connect the white wire from the power
supply to the white wire.
%%.Insert the harness from the light into the hole
provided in the mounting bracket and attach
the nut.
%F.Connect the power supply harness to the
harness from the light.
— WARNING —
Before proceeding, verify the power at the main
circuit breaker panel is "++.
..Feed the incoming power supply cord
through the center hole of the wall mount.
L.Secure the wall mount to the wall using the
four chosen fasteners.
(.Place a small amount of lubricant on the
light arm pivot pin.
J.Insert the light arm assembly, light cord first,
into the top hole of the wall mount.
'.Connect the ground wire from the power
supply to the green/yellow wire.
M.Connect the black wire from the power
supply to the black wire.
%-.After connecting the power leads, replace the
cover on the wall mount making sure no
wires interfere with the closing of the cover.
%..Secure the cover to the wall mount using the
four screws previously removed.
Te s t i n g
%.Before proceeding, make sure the power at
the main circuit breaker panel is1"!.
F.Move the toggle switch, immediately behind
the light head, to the "! (up) position..
16Installation, Operation and Care Manual
Cabinet Mount Assembly
To o l s R e q u i r e d
H Power Drill & Bits
H Slotted Screwdriver
H Phillips-head Screwdriver
H Wire Crimping Tool
Cabinet Mount Installation
Before proceeding with the cabinet mount assembly
installation, the following important information regarding placement PTU# be taken into consideration.
It is imperative that the light assembly's upper
primary arm is:
H mounted parallel with the ceiling,
H Tape Measure
H Wire Stripper
H Set of Allen Wrenches
H Magnetic Level
H Lubricant: Silicon/Grease Product
Power Supply Wiring
Make sure the required power supply wiring is
available in the sub-base of the cabinet.
Suggested Cabinet
Mount Location
Ceiling Line
1.5"
H situated at least 1-1/2" above an X-ray arm's
highest folded position, and
H >:? in contact with or interferes with any
obstruction in the ceiling, especially sprinkler
heads, smoke alarms, etc.!
The illustration below gives the suggested mounting
location for the DentalEZ cabinet mount assembly
and the HDX Intraoral X-ray.
!"#$I1The cabinet assembly will accommodate
either one-light or two-light assemblies.
Typical Sprinkler Head
8"
Typical Clearance
6"
96"
Typical
Clearance
for X-ray
Arm
25"
Adjustable
Height Post
EverLight® 17
EverLight
®
%.Secure the vertical mounting post(s) to the
metal frame inside the cabinet assembly using
the metal post support brackets supplied.
(Refer to the illustration below.)
!"#$I1 The location of the metal angle
brackets at the lower end of the vertical
mounting post(s) determines the height of the
light assembly.
F.Using the supplied 1/4" lag bolts, attach the
metal angle brackets to the plywood center
panel.
Upper Primary
Arm
Metal Post
Support Bracket
!"#$I1The metal post support brackets are
designed to facilitate perpendicular and
parallel adjustments.
-.Using a magnetic level, ensure the light post
is plumb.
..Apply a thin coating of lubricant to the
bearing surfaces.
L.To achieve anti-rotation of the vertical
mounting posts, tighten the set screws in the
post support brackets.
(. Route the light assembly electrical power
cord through the metal post support brackets
and cabinet metal frame to the metal
subbase.
J.Insert the light arm in the bushing.
'.Connect the cord to the power supply. Then
plug the power supply into an electrical
receptacle.
switch as described in Section III Operation
(Page 19).
End View
18Installation, Operation and Care Manual
Section III Operation
The EverLight dental light is designed to provide
illumination of the oral cavity for use by dental
professionals in the dental operatory.
Light Switch
The light ON/OFF paddle is located on the elbow
section of the light arm assembly..
ON
➜
➜
OFF
Light
Paddle
Procedural Settings
! A - Sun - Color Matching - 5300° Kelvin
! B - Sun/Bulb - Operative Setting - 4300° Kelvin
! C - Bulb - Composite Setting - 3200° Kelvin
Sun - 5300° Kelvin
(Color Matching)
Sun/Bulb - 4300° Kelvin
(Operative Setting)
Bulb - 3200° Kelvin
(Composite Setting)
Intensity Settings
! 1 - Low - 8,000 Lux
! Flip the paddle to the UP to turn the light ON.
! Flip the paddle DOWN to shut the light OFF.
To u c h p a d s
Designed with three procedural and three intensity
settings EverLight offers a total of nine settings to
assure the correct light in every procedure.
Procedural
Settings
Intensity
Settings
! 2 - Medium - 15,000 Lux
! 3 - High - 25,000 Lux
Low
(8,000 Lux)
High
(25,000 Lux)
Medium
(15,000 Lux)
EverLight® 19
Section III Operation
Positioning Light
Using the handles on either side of the LED head,
position the light to illuminate the oral cavity. (See
Light Head Range Of Motion, Page 5.)
Handles
EverLight
®
— HELPFUL NOTE —
To take full advant ag e o f the Ever Ligh t's
intensity and color correction, position
the light between FKX and .LX away from
the oral cavity.
20Installation, Operation and Care Manual
Section III Operation
Eliminating Adjustable Arm Drift
The E3AA14:G>?Y8 adjustable arm is factory-adjusted
to counter-balance the weight of the light head. If it
becomes necessary to adjust the mechanism because
of added weight, such as a mirror, perform the
following steps:
To o l s R e q u i r e d
H 5/64 Allen wrench
H 9/64 Allen wrench
H Large Flat-head (slotted) Screwdriver
%.Using a 5/64 Allen wrench, loosen the set
screw under the cap at the end of the
adjustable arm nearest the light head.
Adjustable
End Cap
Set
Screw
Arm
-. Insert a 9/64 Allen wrench into the two
small holes in the side of the adjustable arm
and completely loosen the friction block.
9/64 Allen
Wrench
Friction Block
Adjustable
Arm
Tw o
Small
Holes
..Insert a large flat-head screwdriver into the
large slot in the side of the adjustable arm
and tighten the slotted nut.
Large
Slot
Adjustable Arm
F. Using a 5/64 Allen wrench, remove the four
Allen-head screws from the underside of the
adjustable arm (two on each end). Then slide
off the top cover along with the end cap.
5/64 Allen-head Screws (2)
Adjustable Arm
To p C o v e r
5/64 Allen-head
Screws (2)
End Cap
Slotted
Nut
Large Flat-head
Screwdriver
L.Adjust the friction block to provide one to
two pounds maximum of break-away force.
(.Replace the adjustable arm cover and secure
using the four 5/64 Allen-head screws
previously removed.
J.Replace the end cap and tighten the set screw
using a 5/64 Allen wrench.
EverLight® 21
Section IV Care
Cleaning & Disinfecting
— CAUTION —
Avoid spraying anything directly on the light.
Spray cloth first, then wipe.
%.To clean all surfaces, use a typical
SZ["\"Z&+R$$, commercial cleaner or a
solution of mild liquid detergent and water.
F.After washing, a disinfectant may be used
following the manufacturer's instructions.
!"#$I Daily cleaning with soap and water or a
neutralizer for iodophors may reduce discoloration.
— CAUTION —
All surface disinfectants cause varying degrees
of discoloration to light-colored surfaces.
EverLight
®
-.The following disinfectants are acceptable to
use:
• Caviwipes
• Banicide
• Coecide
22Installation, Operation and Care Manual
Section V User Service Information
Tr ou b l e s h oo t i n g
The EverLight is designed to provide trouble-free
service when installed, operated and maintained
according to the procedures set forth in this manual.
Under normal use, there is no routine user maintenance
required.
— WARNING —
Exercise extreme caution when trouble-
shooting the electrical components of the
EverLight. When electrical power is re-
quired, safety precautions 4G8?1be observed.
The chart below should be used when troubleshooting EverLight problems.
If these suggested troubleshooting procedures do not
resolve the problem, see the service instruction that
follows.
Service Instruction
If the area of concern is not addressed in this
manual, contact your local DentalEZ full-service
dealership. (See Product Support Services, Page 25.)
Please have the following product information
available:
H Model Name
H Model Number
H Serial Number
H Installation Date
H Dealer
Light Arm
Assembly
Model/Serial
Number
Plate
— NOTICE —
For convenient referencing, a color-coded wiring
schematic is located at the end of this manual.
SymptomPossible Cause(s)Solution
No lightMaster/light switch is in "++ position Turn "! master/light switch.
No powerCheck circuit breaker.
Power supply failureReplace power supply.
Broken arm wire harnessCheck arm wire harness for
continuity.
Faulty switchReplace switch.
Low light intensityIntensity set on lowSet intensity to high.
Plastic light shield dirtyClean light shield.
Light arm drifts verticallyCounterbalance out of adjustmentAdjust arm drift. (See Page 21.)
Light arm drifts horizontallyLight post not levelLevel light post.
EverLight® 23
Section VI Parts List/Diagrams
1
(Wall light arm assembly shown)
Light Arm
#ema N t i K / t r a P#tiK/traP
1paCd nEm r A118 - 1083
2paCm r Ad i g i R418 - 1083
3gnihsuBtsoP848 - 1083
EverLight
®
2
Light Head
#ema N t i K / t r a P#tiK/traP
4tooB&r e veLhc t i wS353 - 2083
5hct iw Sf fO/nO153 - 2083
6lor tnoCthg i LDELediSthg iR923 - 2083
7lor tnoCthg i LDELediSt feL823 - 2083
8laceDl o r t noCr o l oC733 - 2083
9laceDl o r t noCy t i sne t n I633 - 2083
01slaceD/wt iK gnisuoHlor tnoCthgiL833 - 2083
11teSr e v oCr a e l Ct n o r F133 - 2083
tiKelb aCn obbiR233 - 2083
tiKylp p uSrewoP433 - 2083
EC , t i Ky l p p u S r e wo P253 - 2083
4
1
3
5
7
6
8
9
12
(Drop Post for
Ceiling and Track
mount only)
24Installation, Operation and Care Manual
#ema N t i K / t r a P#tiK/traP
21tsoPpo rD)gni l ieC&kcarT(911 - 6852
10
11
Drop Post
Product Support Services
DentalEZ Group
DentalEZ Equipment Division
EverLight
DentalEZ Group and its employees are proud of the products we provide to the dental community. We stand behind these
products with a warranty against defects in material and workmanship as provided below.
In the event that you experience difficulty with the application or operation of any of our products, please contact our Technical
Service Department at our expense at 1-866-DTE-INFO (1-866-383-4636).
If we cannot resolve the issue by telephone, we will arrange for a representative to contact you or suggest that the product be
repaired by a dealer service technician.
If product repair or return is required, we will provide you with a R6?G<>1SG?@:<7]3?7:>1number and shipping instructions to
return the product to the proper facility. If the product is under warranty, we will ask you to provide proof of purchase, such as a
copy of your invoice. Please be sure to include the R6?G<>1SG?@:<7]3?7:>1number on the package you are returning. Q<:2G9?8
Freight costs for product returns are the responsibility of the customer. Products under warranty will be repaired or replaced at our
sole discretion and returned at our expense. Products outside the warranty limits will be repaired and returned with costs invoiced
to the customer. We are not responsible for shipping damages. We will, however, help you file a claim with the freight carrier.
Written repair estimates are available.
DentalEZ warrants the EverLight to be free of defects in material and workmanship, under normal usage, under the following
terms:
®
Z75@?1[:4D:>6>?8IN3<<3> ?B1Q6 <7:2 I
Structural Members: Fixed Arm, Adjustable Arm and Handles3 Years from date of installation*
Switches, LED Driver, Lens/LED6 Years from date of installation*
Please note the following additional terms of our warranty and return policy:
H Warranties cover manufacturing defects only and do not cover defects resulting from abuse, improper handling, cleaning, care
or maintenance, normal wear and tear or non-observance of operating, maintenance or installation instructions. +37AG<61?:
H THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NO EMPLOYEE, REPRESENTATIVE OR DEALER IS AUTHORIZED TO CHANGE THIS WARRANTY IN ANY WAY
OR TO GRANT ANY OTHER WARRANTY.
WARR ANTY REPA IRS: Pa rts repair ed or repla ced on a pr oduct th at is in warranty will be warranted for the duration of that
product's original warranty.
NON-WARRANTY REPAIRS: The warranty on parts either repaired or replaced on an out-of-warranty product will cover the
repaired part only and will be for the time frame of a new parts warranty period.
PRODUCT RETURN: Opened products or product returns more than a year old cannot be returned for credit. There will be a
15% ($25.00 minimum) restocking charge on all items authorized for return.
* Provided conditions defined in the Instruction Manual are met.
Guidance and manufacturer’s declaration - electromagnetic emissions
The Everlight Model EVLCE is intended for use in the electromagnetic environment specified below.
The customer or the user of the Model EVLCE should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The Model EVLCE uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
CISPR 11
Class A
The Model EVLCE is suitable for use in all
establishments, other than domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A,
Not Applicable
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Class A
Not Applicable
●
Use only replacement cables listed in Parts Section, Page 24. Other
cables and accessories may negatively affect EMC performance.
●
When the Everlight is used adjacent to other equipment, observe the
light to verify normal operation.
— WARNING —
EverLight® 27
EMC Information
Guidance and manufacturer’s declaration – electromagnetic immunity
The Everlight Model EVLCE is intended for use in the electromagnet environment specified
below. The customer or the end user of the Model EVLCE should assure that it is used in such
an environment.
Immunity test
Electromagnetic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
IEC 60601
Test level
+6 kV contact
+
8 kV air
+2 kV for power
supply lines
+1 kV for input/output
lines
Compliance level
6 kV contact
+
+
8 kV air
+2 kV for power
supply lines
+1 kV for input/output
lines
EverLight
Electromagnetic environment
- guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30 %.
Mains power quality should
be that of a typical
commercial or hospital
environment.
®
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
+1 kV differential
mode
+
2 kV common mode
<5 % U
(>95 % dip in U
T
)
T
for 0.5 cycle
40 % UT
(60 % dip in U
)
T
for 5 cycles
70 % UT
(30 % dip in U
)
T
for 25 cycles
<5 % U
(>95 % dip in U
T
)
T
for 5 sec
1 kV differential
+
mode
+
2 kV common mode
<5 % U
(>95 % dip in U
for 0.5 cycle
40 % UT
(60 % dip in U
for 5 cycles
70 % UT
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
3 A/m 3 A/m
Mains power quality should
be that of a typical
commercial or hospital
environment.
T
)
T
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the Model EVLCE requires
)
T
continued operation during
power mains interruptions, it
is recommended that the
Model EVLCE be powered
)
T
T
)
T
from an uninterruptible
power supply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.