Delta RetroSign GR1, RetroSign GR3 Quick Manual

GR3 & GR1 Quick Guide (UK) rev 08-07-2010.doc
Quick guide - RetroSign GR1 & GR3®
Getting started
Turn the RetroSign on by pressing. The display will show some information about the instrument. Repeat to show the results of the last measurement.
The instrument is calibrated and ready if the icon is displayed, if not, the instrument must first be calibrated. Press and follow the instructions to do this.
Start a measurement by pressing the trigger knob. The display shows "WAIT" while measuring then the measured values is shown.
If errors are detected during a measurement or calibration sequence, a message will be displayed and one or more error icons appear.
Keyboard functions:
On/Clear Push shortly to activate the instrument.
When switched on it clears all selected operations.
Displays the last measurement result.
Clears the mean calculation when displaying this
Off Terminates all activity, saves changed values and powers
off.
Scroll up Open the menu, scroll through menus
Increment/scroll edit values.
Scroll down Open the menu, scroll through menus
Decrement/scroll edit values.
Calibration Open Instrument calibration menu, select/change settings.
Enter Activate selected function or accept changed settings.
Switch between the different results and status display screens
Trig Start R' measurement.
Start quick calibration.
Icons:
The instrument status and operational mode are signalled by the use of icons.
Instrument is calibrated and ready.
Calib. Monitor Active for normal aperture.
Menu selected. Calib. Monitor Disabled for reduced aperture. Calibration selected. Battery fully charged. Reference calibration in progress. Battery getting low. Zero calibration in progress. Battery empty, recharge. TAG Reader function enabled.
(option)
Display backlight is enabled.
GPS On. Position FIX OK
(option)
GPS On. No FIX
(option)
Calculating mean values
Bluetooth Disconnected/Connected
(option)
GR3 & GR1 Quick Guide (UK) rev 08-07-2010.doc
Calibration:
RetroSign features two levels of calibration: quick and full.
Quick calibration.
Quick calibration is initiated by first pressing and then the trigger. Mount the reference cap (reflective surface pointing in) before triggering. The instrument will be calibrated to the last used values.
Full calibration.
Full calibration is initiated by pressing and then . Follow the on-screen instructions.
Zero Measurement. Mount the zero cap (dark surface pointing in). Press when ready. WAIT is displayed during the
measurement. The measured zero values and a short status for each geometry are shown, then the display shows that the instrument is ready for the next step in the calibration procedure.
Reference Measurement. Mount the reference cap (reflective surface pointing in). Edit the reference values one by one by using
the and keys until the values are the same as printed on the calibration unit. Press when ready. WAIT is displayed once more during the calibration process; the display shows the calibration results and then asks the operator for acceptance, press to accept or to discard the calibration, the previous values will then be used.
After a successful calibration the icon is shown.
Menu:
Enter the menu by repeatedly pressing or until the desired function displays and then follow
the on-screen instructions.
1
) All log entries are marked as having been read, the log status will only show entries since this operation.
2
) Calculates the mean value of all measurements since the last ON/C action, press once from the main screen to display
the calculated mean values. Press On/C twice to restart the calculation.
3
) Instrument require recalibration if changed.
4
) Always disable the Bluetooth function if using the Extender Box
5
) NA if option not installed.
Tag Reader (optional)
Connect the Reader unit to the front connector and enable the TAG Reader in the menu. When doing a measurement the instrument requests to read a TAG on the first trigger action and then needs a second trigger to actually do the measurement. This also applies when using the remote trigger.
Battery:
To remove the battery release the retaining spring and pull the battery out of the handle.
Temperatures in excess of 50°C (122°F) can damage the battery. Do not short circuit and do not dispose together with household waste.
USB PC connection:
Important! Follow the instructions found on the Instrument CD before using the USB connection; drivers must be installed
on the PC prior connecting to a PC-USB port.
Seq ID: Name a measurement series Seq ID count: Running number for the Seq ID. can be edited Clear ID: Delete the Seq ID string Sound control: Enable/Disable system sounds
Clear Work Log: Reset user log
1
Remove last Measurement:
Delete last measurement from log
Mean calculation. Enable/Disable Mean calculations
2
Calib. monitor: Enable/Disable the calibration monitor
3
Remote Box display Select result for display on remote-box 4LCD backlight: Enable/Disable LCD backlight GPS state
Enable/Disable the GPS unit
5
DGPS mode: Select GPS position correction system
5
Reader state
Enable/Disable the TAG Reader option 5Off Time: Set time for auto turn off
* RetroSign GR3 is an optical precision instrument, handle with care. * Store it the carrying case in a clean and dry environment. * Do not recharge the battery unnecessary as this will reduce battery life. * Carefully check and clean the reference cap with a clean, soft and dry cloth before each calibration.