DeFelsko PosiTest PC powder Cheker Instruction Manual

Instruction Manual v. 3.0
Powder Checker
Uncured Coating Thickness Gage
Introduction
The PosiTest PC Powder Checker is a hand-held, non-contact
coating thickness gage that uses high-frequency ultrasound to predict the cured thickness of uncured powder coatings.
1
1. POWER-UP the PosiTest PC by pressing the button.
2. ZERO the gage using the supplied Zero Fixture (pg. 4).
3. MEASURE the applied powder (pg. 4).
4. ADJUST to the specific powder if necessary (pg. 8).
Quick Start
PosiTest PC Contents
n
PosiTest PC gage
n
Zero Fixture
n
Replacement probe screen
n
4 AAA alkaline batteries
n
Instruction manual
n
Hard shell carrying case
n
USB Cable
n
PosiSoft Desktop reporting software
n
Two (2) year warranty
To preserve battery life, the PosiTest PC will power
down after approximately 5 minutes of no activity. All settings are retained.
NOTE:
The USB Port has several functions:
n Gage updates (pg. 10) n USB Keyboard Mode /
live stream readings (pg. 10)
n Download stored readings to
PosiSoft Desktop (pg. 10)
n Allows the gage to be operated
without batteries when connected to a suitable USB power supply
Sensor/Screen
Display
Buttons
USB Port
Memory Mode: Turns Memory Mode ON.
The gage will store and maintain a running average for up to 999 readings. Stored measurements can be downloaded to PosiSoft
Desktop Software (pg. 10). When Memory Mode
is ON, the following menu options appear:
2
Press the and buttons simultaneously to enter the gage menu. Navigate between menu items using the or buttons. Use the to select the highlighted m
enu item.
The following menu items are displayed:
Zero: Zeroes the instrument using the supplied Zero
Fixture (pg. 4)
1-Point Adjustment: Adjusts the gage to a known thickness of powder (see Adjustment Techniques, pg. 8)
2-Point Adjustment: Adjusts the gage to two known
thicknesses of powder if the 1-Point Adjustment does not improve accuracy (see , pg. 9)
Display Brightness: Adjusts the gage display
brightness using the and buttons.
RESET Memory Mode. All stored readings and the
running average will be cleared from gage memory.
Turns Memory Mode OFF
Memory Mode is unavailable when USB Keyboard Mode is active
While in Memory Mode, press the button to discard
the last reading
NOTES:
Menu Navigation
3
Units: Toggles the display units between mils and
microns.
Sound: Turns the built-in speaker ON or OFF .
When set to OFF, a red speaker icon will appear in the gage display.
Scan Mode: In this mode, the gage will continuously
take readings, and only add a reading to memory when the button is released. Ideal for quickly analyzing
large areas. (See Scan Mode, pg. 6) USB Keyboard Mode: When enabled and connected to
a computer, the PosiTest PC will be recognized as a
keyboard. Readings are sent to the computer as they are taken, emulating keystrokes. (pg. 10)
Memory Mode is unavailable when in USB Keyboard Mode.
NOTE:
Gage Reset: Restores the gage to default settings,
clears stored memory and calibration adjustments.
Power Off: Powers down the gage. Alternatively, the
gage will automatically power down after approximately 5 minutes of no activity. All settings are retained.
Exit Menu: Exits the gage menu and returns to the main
measurement screen.
4
For best accuracy, perform a Zero at the
beginning of every shift to compensate for temperature and humidity changes. It is also required after a Gage Reset or whenever the “Zero Probe” graphic is displayed.
1. Select Zero from the Gage menu.
2. Place the probe tip in the supplied Zero
Fixture.
3. Press the button and hold the Zero Fixture
on the probe until the hourglass icon disappears and a green check mark is
displayed. If the zero fails, a red X will display. Repeat the Zero process.
The Zero process can take up to 10 seconds
NOTE:
How to Measure
Power-up the PosiTest PC by pressing the button. With the
gage at least 10 cm (4 inches) away from the coated part, press and hold the button. A target grid will appear on the display:
Target Grid
Distance Target Bar
Distance Paddle
Zero Fixture
Zeroing
5
While watching the gage display, move the probe tip closer to the
target, aligning the Distance Paddle with the Distance Target Bar. The optimum distance for measurement is 2 cm (3/4 inch)
between the probe tip and the surface of the applied powder.
Ensure the PosiTest PC is held perpendicular to the surface.
The instrument will take a measurement when:
a. The Distance Paddle is near the Distance Target Bar. This
indicates that the instrument is the correct distance away from the part.
b. The Distance Paddle is as wide as possible. This indicates
that the instrument is being held perpendicular to the surface.
For best results use Memory Mode (pg. 2) to generate averages.
The average of a series of measurements is often a more meaningful prediction of cured thickness than a single reading.
If a thickness result does not appear after 5-6 seconds, release the measurement button and try again.
Small and/or erratically swinging parts may present measurement challenges. Steady the part whenever possible. If erroneous readings continually result, change the measurement strategy by rotating the gage, altering its alignment to the part, or by
2 cm (3/4 inch)
90°
Correct
Incorrect
Correct
Incorrect
6
0 A good reading, but the predicted cured thickness is
less than the minimum range.
+++ A good reading, but the predicted cured thickness is
greater than the maximum range.
--- An indeterminate or invalid reading.
measuring a different location on the part. If a valid measurement cannot be calculated, one of the following messages will be displayed:
Single Reading Mode (default)
The gage will automatically display the predicted cured thickness after analyzing the powder. To take another reading, release the
button, then press and hold the again.
Scan Mode
Press and hold the button. The gage continuously analyzes the powder and displays the predicted cured thickness until the
button is released. The last valid measurement is added to
memory when the button is released if Memory mode is on.
Maintenance
The probe screen protecting the sensor of the PosiTest PC can
be unscrewed for cleaning. Accumulated powder should be carefully removed from the screen by no more than 0.2 MPa (20 ­30 psi) of compressed air.
- Do not blow air directly into the sensor.
- Replacement screens are available for purchase.
7
Calibration and Adjustment
The PosiTest PC probe is factory calibrated and performs an
automatic self-check each time it calculates a measurement. For most powder measurement applications, no further adjustment is
necessary. Simply Zero the gage (pg. 4) at the beginning of
each shift, then measure.
The PosiTest PC’s factory calibration accurately predicts cured
powder thickness for a wide range of powder coatings. However,
some powder formulations may require a Calibration Adjustment. Begin by performing a Verification of Accuracy to determine if
there is a need for an adjustment. If the average of several thickness results is outside the combined tolerances of the two
measuring instruments (PosiTest PC and dry film thickness
instrument) then two adjustment techniques are available to
correct the variance (see Adjustment Techniques).
Verification of Accuracy
Before first use on a new powder and periodically thereafter, an accuracy check should be performed. This ensures the gage measures properly by comparing predicted thickness values with actual cured thickness values.
1. Zero the probe (pg. 4).
2. Spray powder onto a flat, smooth metal sample.
3. Measure the uncured powder with the PosiTest PC.
4. Cure the powder.
5. Measure cured thickness with a magnetic or eddy current
coating thickness gage such as the PosiTest DFT or PosiTector 6000.
6. Compare results.
A successful verification requires that the average of a series of measurements by both instruments be within their combined
tolerances. If not, adjust the PosiTest PC using one of the methods described in Adjustment Techniques (pg 8).
8
Adjustment Techniques
If the average of a series of measurements by both instruments
(PosiTest PC and dry film thickness instrument) is not within combined tolerances, the PosiTest PC can be adjusted using
one of two methods:
1-Point Adjust – The first and most common choice. 2-Point Adjust – Used only if a 1 Pt Adjust does not
sufficiently improve accuracy.
1-Point Adjust
1. Spray approximately 75 microns (3 mils) of powder onto a
flat, smooth metal sample.
2. Select 1-Pt Adjust from the gage menu.
3. Take at least 3 uncured powder
measurements with the PosiTest PC. Press
the button to discard the last reading if it should not be included in the average of the measurements.
4. Select OK to confirm and store the average of the
measurements.
5. Cure the powder. The instrument will power off after 5
minutes of inactivity, but the uncured powder measurements will be retained.
6. Measure cured thickness with a magnetic or eddy current
dry-film thickness gage. Record the results.
7. Power-up the instrument, and select 1-Pt
Adjust from the gage menu and then
select the option.
8. Adjust the displayed reading lower or
higher to match the cured thickness value.
Select OK to store the adjustment.
9. Verify accuracy (pg. 7) to ensure the instrument is
measuring within tolerance.
This only appears if 1 pt adjustment readings have been stored
NOTE:
9
2-Point Adjust
1. Spray powder onto two flat, smooth metal samples. Apply
approximately 40 microns (1.5 mils) on the first, and 75 microns (3 mils) on the second.
2. Select 2-Pt Adjust from the gage menu.
3. Take at least 3 uncured powder measurements on the first
(thin) powder sample with the PosiTest PC. Press the
button to discard the last reading if it should not be included in the average of the measurements.
4. Select OK to confirm and store the average of the
measurements.
5. Take at least 3 uncured powder measurements on the
second (thick) powder sample with the PosiTest PC. Press
the button to discard the last reading if it should not be included in the average of the measurements.
6. Select OK to confirm and store the average of the
measurements.
7. Cure the powder on both panels. The instrument will power
off after 5 minutes of inactivity, but the uncured powder measurements will be retained.
8. Measure cured thickness on both panels with a magnetic or
eddy current coating thickness gage. Record the results.
9. Select 2-Pt Adjust from the gage menu and then select
the option.
10. Adjust the displayed reading for the first (thin) sample lower
or higher to match the cured thickness
values. Select OK to store the adjusted
measurement value.
11. Adjust the displayed reading for the second
(thick) sample lower or higher to match
the cured thickness values. Select OK to
store the adjusted measurement value.
12. Verify accuracy (pg. 7) to ensure the
instrument is measuring within tolerance.
DO NOT unplug the Gage during the update operation.
10
PosiSoft Desktop
PosiSoft Desktop - Powerful desktop software (PC/Mac) for
downloading, viewing, printing and storing measurement data. Includes a customizable, templated PDF Report Generator. Learn more at www.defelsko.com/posisoft
Gage Update
The PosiTest PC gage can be updated by the user.
To perform an update, the Gage must be connected with the USB cable to an internet connected computer running PosiSoft Desktop.
If an update is available, a prompt will appear in PosiSoft for the user to preform an update.
Ensure that stored measurements are downloaded to
PosiSoft Desktop before updating the gage.
NOTE:
USB Keyboard Mode / Live Stream Measurements
When enabled, the PosiTest PC will be recognized as a USB
keyboard device when connected to a computer. This mode is compatible with any software that accepts keystrokes. Readings are sent to the computer as they are taken, emulating keystrokes.
Can be used in both Single Reading Mode and Scan Mode.
Memory Mode is unavailable when in USB Keyboard Mode.
11
Technical Data
Measuring Range
20-100 microns
0.8-4.0 mils
Accuracy
+ 5 microns
+
0.2 mil
Resolution
1 micron
0.1 mil
Measurement Time 2-5 seconds
Measurement Distance
to Powder
19 mm
0.75 inch
Measurement Area
2 mm
0.08 inch
Temperature Range
0 to +50°C
+32° to +120°F
Gage Size
241 x 50 x 50 mm
9.5 x 2.0 x 2.0 in.
Gage Weight
(without batteries)
140 g
4.9 oz
Conforms to: ASTM D7378
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
12
www.defelsko.com
© 2018 DeFelsko Corporation USA
All Rights Reserved
This manual is copyrighted with all rights reserved and may not be reproduced or transmitted, in whole or part, by any means, without written permission from DeFelsko Corporation.
DeFelsko, PosiTest, PosiTector and PosiSoft are trademarks of DeFelsko Corporation registered in the U.S. and in other countries. Other brand or product names are trademarks or registered trademarks of their respective holders.
Every effort has been made to ensure that the information in this manual is accurate. DeFelsko is not responsible for printing or clerical errors.
Returning for Service
Before returning the instrument for service…
1. Install new or newly recharged batteries in the proper
alignment as shown within battery compartment.
2. Examine the probe tip screen for dirt or damage.
Replacement screens are available for purchase.
3. Perform a Reset (pg. 3) and Zero (pg. 4).
IMPORTANT:
If you must return the Gage for service, describe the problem fully and include measurement results, if any. Be sure to also include the probe, your company name, company contact, telephone number and fax number or email address. Website: www.defelsko.com/support
Limited Warranty, Sole Remedy
and Limited Liability
DeFelsko's sole warranty, remedy, and liability are the express limited warranty, remedy, and limited liability that are set forth on its website: www.defelsko.com/terms
Manual de instrucciones v. 3.0
Powder Checker
Medidor manual de espesor de recubrimientos sin contacto
Introducción
El PosiTest PC Powder Checker es un medidor manual de espesor de recubrimientos sin contacto que utiliza ultrasonidos por alta frecuencia para determinar espesores compactos de recubrimientos de polvos no curados.
1
1. ENCIENDA el PosiTest PC pulsando el botón .
2. AJUSTE A CERO el medidor utilizando el accesorio de
cero (Pág. 4).
3. MIDA el polvo aplicado (Pág. 4).
4. AJUSTE al polvo específico si fuera necesario (Pág. 8).
Inicio rápido
Para ahorrar energía, el PosiTest PC se apagará tras unos 5 minutos de inactividad. Se conservarán todas las configuraciones.
NOTA:
El puerto USB realiza varias funciones:
n Actualización del medidor (Pág. 10) n Modo teclado USB / transmisión
directa de lecturas (Pág. 10)
n Descarga de lecturas almacenadas
en PosiSoft Desktop (Pág. 10)
n Permite que el medidor funcione
directamente conectado a una fuente de alimentación USB
Contenidos PosiTest PC
n
Medidor PosiTest PC
n
Accesorio de cero
n
Pantalla de sonda de recambio
n
Pilas alcalinas: 4 AAA
n
Manual de instrucciones
n
Maleta rígida
n
Cable USB
n
Aplicación de informes PosiSoft Desktop
n
Garantía de dos (2) años
Sensor/pantalla
Pantalla
Botón
Puerto USB
El modo de memoria se deshabilitará si está activo el modo de teclado USB.
En modo memoria, pulse el botón para descartar la última lectura
Modo memoria: Activa el modo de memoria.
El medidor almacenará y mantendrá un promedio de ejecución de hasta 999 lecturas. Las mediciones almacenadas pueden descargarse al software PosiSoft Desktop (Pág. 10). Si está activado el modo memoria, aparecerán las siguientes o
pciones de menú:
2
Menú Navegación
Pulse al tiempo los botones y para acceder al menú del medidor. Desplácese por las diferentes opciones de menú utilizando los botones o . Utilice para acceder a la opción de menú seleccionada.
Estas son las opciones de menú que se muestran:
Cero: Ajusta el cero del instrumento utilizando el acce­sorio de cero (Pág. 4)
Ajuste de 1 punto: Ajusta el medidor a un espesor conocido de polvo (consulte Ajustes técnicos, Pág. 8)
Ajuste de 2 puntos: Ajusta el medidor con dos espe­sores conocidos de polvo si el ajuste de 1 punto no mejo­ra la precisión (consulte Ajustes técnicos, Pág. 9)
Brillo de la pantalla: Ajusta el brillo de la pantalla del medidor utilizando los botones y .
RESTABLECE el modo memoria. Todas las lecturas almacenadas y el promedio de ejecución se borrarán de la memoria del medidor.
Desactiva el modo memoria
NOTAS:
3
Unidades: Alterna las unidades entre mils y micras.. Sonido: Activa el encendido o apagado del
altavoz integrado. Si está apagado, aparecerá en la pantalla del medidor un icono con un altavoz rojo.
Modo captura: En este modo, el medidor realizará lecturas continuas y solo agregará una lectura a la memoria si el botón se libera. Ideal para analizar rápidamente grandes áreas. (Consulte Modo captura, Pág. 6)
Modo teclado USB: Cuando está activado y conectado a un ordenador, el PosiTest PC se reconocerá como un teclado. Las lecturas se enviarán al equipo a medida que se tomen, emulando pulsaciones de teclas. (Pág. 10)
El modo de memoria se deshabilitará si está en
modo de teclado USB.
NOTA:
Reinicio del medidor: Restablece la configuración predeterminada del medidor y borra la memoria almacenada y los ajustes de calibración.
Apagar: Apaga el medidor. De otra manera, el medidor se apagarán automáticamente tras unos 5 minutos de inactividad. Se conservarán todas las configuraciones.
Cerrar menú: Cierra el menú del medidor y vuelve a la pantalla principal de mediciones.
Loading...
+ 37 hidden pages