DeFelsko PosiTector UTG User Manual

PosiTector UTG
INSTRUCTION MANUAL v. 2.2
for UTG ME models
Simple. Durable. Accurate.
Ultrasonic Thickness Gage
Introduction
Introduction
The PosiTector UTG ME is a hand-held Ultrasonic Thickness Gage that uses the non-destructive ultra­sonic pulse-echo principle to measure the wall thick­ness of a wide variety of materials. The default and most common setting for the PosiTector UTG is multiple-echo (ME) mode. In ME mode the Gage measures the metal thickness of a painted structure, disregarding the paint thickness. The user can switch the gage to single-echo mode by selecting SE Mode from the Gage menu (see pg.15).
The probe (a single element contact transducer) transmits an ultrasonic pulse into the material to be measured. This pulse travels through the material towards the other side. When it encounters an inter­face such as air (back wall) or another material, the pulse is echoed back to the probe. The time required for the pulse to propogate through the material is measured by the Gage, represented as t
1
and t
2
below.
In single-echo mode (see Figure 1) the Gage determines thickness by measuring t1(uncoated) or
t2 (coated), dividing it by two and then multiplying by
the velocity of sound for that material (steel).
steel
coating
steel
2
t
>
1
t
For uncoated materials t1relates directly to material thickness. When a material is coated the propaga-
tion time is increased and is shown above as t
2
.
Coatings such as paint have a slower velocity of
Figure 1
2
t
1
t
Probe
Probe
In multiple-echo mode the Gage determines thickness by measuring the time between at least three consecutive back wall echoes.
steel
In the figure above, multiple-echo mode measures only the time between echoes. Regardless of whether the steel is coated or not, all times between echoes are the same. In multiple-echo mode the Gage determines thickness by measuring
t
1+t2+t3
, dividing it by six and then multiplying by the
velocity of sound for that material. The resultant thickness calculation made by the instrument is therefore an accurate measurement of the steel thickness only, disregarding the coating thickness.
The velocity of sound is expressed in inches per microsecond or meters per second. It is different for all materials. For example, sound travels through steel faster (~0.233 in/µs) than it travels through plastic (~0.086 in/µs).
Probe
1
t
2
t
3
t
1
t
2
t
3
t
Coating
Figure 2
Operation Overview
Operation Overview
1. Turn the Gage ON (see Power-Up - pg. 5)
2. Verify accuracy (see Verification - pg. 9)
3. Select the correct velocity of sound (see pg. 10)
4. Measure the part (see How to Measure - pg. 5)
Probe
sound than that of metal. Thus the single-echo tech­nique will produce a thickness result greater than the actual combined coating+metal thickness.
Zero Plate
Finger Grip
Wear Face
(Measurement Head)
USB/IR Port
Probe connector
Bi-color indicator light
Display
Probe (transducer)
Typical Display
Current Measurement
Currently selected material
Unit of Measurement
The Coupled Symbol appears when the probe is ultrasonically connected with a surface.
inch
Steel
Battery Icon
Measurement Mode ­Currently set to ME
How to Measure
How to Measure
1. Remove black rubber cap from probe if supplied. Couplant (glycol gel ­included) must be applied to the surface to be tested to eliminate air gaps between the wear face and the surface. A single drop of couplant is sufficient when taking a spot measurement.
2. Place the probe flat on the surface. Use moder­ate pressure to press against the top of the probe with the thumb or index finger. When the probe sens­es echoed ultrasound, a coupled symbol will appear on the display and thickness values will be displayed. While the probe is coupled, the PosiTector UTG will perform 3 readings per second, updating the display each time.
3. When the probe is removed from the surface, the last measurement will remain on the display.
Occasionally, excess couplant will remain on the probe when the probe is lifted from the surface. This may cause the PosiTector UTG to display a final measurement value different from those observed when the probe was on the surface. Discard this value and repeat the measurement.
Power-up / Power-down
Power-up / Power-down
The PosiTector UTG powers-up when any button is pressed. To preserve battery life, the Gage powers­down after approximately 3 minutes of no activity. All settings are retained.
Throughout this manual, the symbol indicates more information about the particular topic or feature is available on our website.
Go to: www.defelsko.com/manuals
NOTE:
Surface Conditions
Surface Conditions
Ultrasonic measurements are affected by the condition, roughness and contour of the surface to be tested.
Measurement results may vary on coarse surfaces. Where possible, it is recommended to seat the transducer on a smooth flat surface that is parallel to the opposite side of the material.
On rough surfaces, the use of a generous amount of couplant minimizes the surface effects and serves to protect the transducer from wear, particularly when dragging the probe across a surface.
On smooth, uncoated metal surfaces the Gage (in ME mode) may occasionally be unable to give a measurement result even when the "coupled" symbol appears. Use additional couplant and lighter pressure on the probe when measuring. Alternatively, laying a plastic shim on the surface with couplant applied to both sides to simulate a painted surface will help produce a steel­only thickness measurement (multiple-echo mode). Switching the Gage to SE mode (see pg.15) will also help produce a steel-only thickness measurement.
Zero Menu Option
Zero Menu Option
The PosiTector UTG probe can be "zeroed" to com- pensate for temperature and wear. This action is not required when operating in ME (multiple-echo) mode except when the Gage is new and after a Reset (pg.
16). However it ensures best accuracy when operat­ing in SE (single-echo) mode. Therefore it is a good idea to "zero" the probe before each work shift and when the temperature of the part changes. During zeroing, it does not matter what the current velocity setting is.
Zero
ME Mode Note:
Menu Operation
Menu Operation
Gage functions are menu controlled. To access the Menu, turn the Gage on, then press the button.
3. Select the Zero menu option
(LCD will display image shown in Figure A). Hold the probe in the air and press the button to select OK to continue.
Figure A
4. When the image shown in
Figure B is displayed, press and hold the probe against the zero plate.
5. When the Gage beeps and the
image shown in Figure C appears, remove the probe from the zero plate.
Figure C
Figure B
The Gage will display 0.00.
DO NOT place probe on the Zero plate
until Step 4.
Important:
1. Make sure the Gage is on and the probe is
wiped clean.
2. Apply a single drop of couplant onto the zero
plate located on the underside of the Gage. DO NOT apply couplant directly onto the probe face.
Current selection is displayed with darkened background
Reset
Gage Info
Flip Display
Units
Set Clock
Language
Memory
Zero Cal Settings SE Mode
HiLo Alarm
Setup
Some menu items have a tick box to their right to indicate current status. An empty box indicates that feature in not active.
“Focus” is currently at this unselected (empty) Radio button
“Radio” buttons. Only one can be selected at a time.
“List“ box
a “tick” indicates this
square box has
been selected
Calibration, Verification
Calibration, Verification
and Adjustment
and Adjustment
Three steps ensure best accuracy…
1.Calibration - typically done by the manufacturer or a qualified lab
2.Verification of Accuracy - as done by the user
3.Adjustment - to a known thickness or sound velocity for the material to be measured
Calibration
Calibration is the controlled and documented process of measuring traceable calibration
To navigate, press (-) to scroll DOWN, (+) to scroll UP and to SELECT. Press both (-)(+) buttons at any time to exit any menu or select Exit from the Menu.
List boxes have a down arrow on the right-hand side. Use the (-) and (+) buttons until your desired choice appears, then press to select this choice and move focus onto the next item.
Verification
Gage accuracy can and should be verified using known reference standards of the material to be tested.
Verification is an accuracy check performed by the user using known reference standards. A successful verification requires the Gage to read within the combined accuracy of the Gage and the reference standards.
Calibration step blocks are available for this purpose. But while they are valuable for verifying gage accuracy to tight, traceable tolerances, they may not represent the actual material to be measured.
Therefore a reference standard can also be a small piece of the material to be measured. If the actual material is not available, material of the same composition (sound velocity) can be used. Thickness can be determined with alternate means such as a
Certification: All PosiTector UTG instruments ship with a Certificate of Calibration. For organizations with re-certification requirements, gages may be returned at regular intervals for calibration. DeFelsko recommends that customers establish gage calibration intervals based upon their own experience and work environment. Based on our product knowledge, data and customer feedback, a one year calibration interval from either the date of calibration, date of purchase, or date of receipt is a typical starting point.
standards and verifying that the results are within the stated accuracy of the Gage. Calibrations are typically performed by the Gage manufacturer or by a certified calibration laboratory in a controlled environment using a documented process.
The PosiTector UTG allows four simple adjustment choices. All four methods are based on the simple
10
Adjustment
Adjustment, or Calibration Adjustment is the act of aligning the Gage's thickness readings to match that of a known reference sample.
The PosiTector UTG is factory calibrated. But in order for it to take accurate thickness measurements of a particular material it must be set to the correct sound velocity for that material. Be aware that material composition (and thus its sound velocity) can vary from stated tables and even between lots from a manufacturer. Adjustment to a sample of known thickness of the material to be measured ensures that the Gage is adjusted as close as possible to the sound velocity of that specific material. Samples should be flat, smooth and as thick as the maximum expected thickness of the piece to be tested.
Cal Settings
micrometer. If the Gage does not measure the sample within the combined tolerances of the Gage and the micrometer, the Gage can be adjusted (see below). For best accuracy this process should be repeated if the temperature of the part changes substantially.
To guard against measuring with an improperly adjusted Gage, verify the Gage at the beginning and the end of each work shift. During the work shift, if the Gage is dropped or suspected of giving erroneous readings, its accuracy should be re­verified. In the event of physical damage, wear, high usage, or after an established calibration interval; the Gage should be returned to the manufacturer for repair or calibration.
11
Thickness
1.Apply a drop of couplant onto the reference standard.
2.Measure the reference standard.
3.Lift the probe. Select the Cal Settings Thickness menu option.
4.Adjust the display down (-) or up (+) to the reference standard thickness.
5.Press to store the adjustment and exit.
premise of adjusting the sound velocity. The first three adjustment methods make 1­point calibration adjustments to optimize the linearity of the Gage over small ranges. The
Thickness
Material
Velocity
2 Pt Adjust
fourth method makes a 2-point calibration adjustment to allow for greater accuracy over a large range.
The most common method of adjustment is to meas­ure a sample of known thickness. Select a reference standard of material as close as possible in compo­sition to the intended application. For best results, the thickness of the reference standard should be equal to or slightly greater than the thickness of the part to be measured.
Material
If a known thickness of the material is not available, but the material is known, this quick adjustment allows the user to load one of several preprogrammed material velocities.
1. Select the Cal Settings Material menu option.
2. Scroll to the desired material.
3. Press to store the adjustment and exit.
12
2 Pt Adjust
A 2-Point adjustment allows for greater accuracy while simultaneously adjusting probe zero. Select two reference standards as close as possible in composition to the intended application. For best results, the thickness of the thicker reference standard should be equal to, or slightly greater than the thickest part to be measured. The thickness of the thinner reference standard should be as close as possible to the lower end of the expected measurement range.
1. Select the Cal Adjust 2 Pt Adjust
menu option.
2. Measure the thinner reference sample.
3. Lift the probe from the sample and adjust the dis­play down (-) or up (+) to the expected thickness.
4. When the expected thickness is reached, press
5. Measure the thicker reference sample.
6. Lift the probe from the reference sample and adjust the display down (-) or up (+) to the expected thickness.
7. Press to store the adjustment and exit.
Velocity
If the sound velocity for the test material is known, the Gage can be adjusted to that specific sound velocity by performing the following steps.
1.Select the Cal Adjust Velocity menu option.
2.Adjust the display down (-) or up (+) to the desired velocity. Holding the (-) or (+) buttons increases the rate of change.
3.When the desired velocity is reached, press
to store the adjustment and exit.
13
Memory Management
Memory Management
The PosiTector UTG can record 10,000 measurements in up to 1000 groups (batches) for on-screen statistical purposes, printing to an optional IR printer, or for downloading to a personal computer (optional PosiSoft software and USB cable required).
Memory
New Batch
When a batch is open, create a new batch
by pressing (+)
Shortcut:
Closes current batch and creates a new batch name using the next higher number. For example, if only Batch 1 and Batch 3 exist, then Batch 4 would be created and made the current batch. The icon appears and statistics are displayed and each measurement will now be simultaneously shown on the display and stored into this new batch. On screen statistics are immediately updated with each measurement. New batch names are date stamped at the time they are created.
Open
Selects a previously created batch name to open and make current. If it contains measurements, on­screen statistics will immediately reflect values calculated from this batch.
Mean
(average)
Standard
Deviation
No. of
measurements
Max and Min
values
Current
Batch
Last
reading
Memory Icon
Print
Prints all stored measurements to the optional IR printer or to a PC’s default Windows printer via the optional USB cable and PosiSoft software. Press (-)(+) simultaneously to cancel printing.
Downloading Measurements Stored in Memory Measurements stored in the Gage's memory (in batches) can be downloaded to a computer using optional PosiSoft software and USB cable. Measurements are not erased from memory after downloading.
PosiSoft® allows entry of notes and annotations, prints histograms and basic charts, manages data, and readings can be exported to a document or spreadsheet.
Remove the last reading from the current
open batch by pressing (-).
NOTE:
14
To delete a value, scroll to that value (align the “+” symbol beside it) then either take another measurement to change it, or press to delete it or exit. Statistics are updated.
View
Lists all readings on the display from the current or most recently used batch. It begins by showing the last 10 measurement values. Scroll using the (-) or (+) buttons. Hold for 1 second to scroll a page at a time.
Delete
Removes a batch completely from memory. The name is deleted and all measurements are erased.
Close
Stops the recording process, closes the current batch, and removes the statistics from the display.
15
When HiLo Alarm is first selected, the current Lo setting is displayed. Adjust down (-) or up (+). Alternatively, measure a material with a thickness close to the required value and make final adjustments with the buttons. Select to accept this value. The current Hi setting is now displayed.
This mode allows the Gage to visibly and audibly alert the user when measurements exceed user­specified limits.
Hi Lo Limits (Alarm)
Hi Lo Limits (Alarm)
HiLo Alarm
The default and most common setting for the PosiTector UTG is multiple-echo mode
The user can switch the gage to single-echo mode by selecting SE Mode from the Gage menu
- To detect pits and flaws
- To increase the measurement range
- To obtain thickness measurements in circumstances where multiple-echo can not
Single Echo Mode
Single Echo Mode
SE Mode
Press (-) to quickly toggle between ME
and SE measurement modes.
Shortcut:
Follow the same procedure to adjust this setting. The icon will appear on the display.
All measurements will now be compared to your defined limits. The Gage LED blinks green if results are within those limits or red if readings are outside set limits. A single low tone will display if it is below the Lo limit, and a HIGH tone if it is above the Hi limit.
All batches are date-stamped when created, and all measurements are time-stamped (24 hour format) when stored into these batches. It is therefore impor­tant to keep both the date and time current using this menu button. Alternatively, the date and time can be automatically updated when the gage is connected to PosiSoft using the Gage Utilities -> Set Clock func­tion in PosiSoft.
Set Clock
This menu button converts the display and all stored readings from inch to mm or vice versa.
Units
16
A more thorough Reset can be performed by holding the (+) button when the Gage is powered down until the Reset symbol appears. It performs the same function as a menu Reset with addition of Units = inch, and Language = English.
Date and Time are not affected by any Reset.
NOTE:
Reset restores factory settings and returns the Gage to a known, out-of-the-box condition. It is handy when you want to “start all over”.
The following occurs:
- all batches are closed and stored measurements are erased.
- calibration adjustments are cleared and returned to the Gage’s factory calibration settings.
- menu setting are returned to the following:
Memory = OFF HiLo Alarm = OFF SE Mode = OFF
Reset
Setup Menu
Setup Menu
Setup
17
Troubleshooting
Troubleshooting
Some common reports received by our Service Department along with possible causes. Most condi­tions however can be cleared with a Reset (pg. 16).
Probe continues to measure after lifted from surface
Wipe away any excess couplant on probe tip.
Gage fails to power down
Ensure the probe is clean and free of couplant. The Gage will not turn off if coupled symbol is displayed on LCD.
Gage is coupled, but not measuring
See Surface Conditions, pg. 6.
The surface temperature of the test piece should not exceed (150° C / 300° F)
Application Notes
Application Notes
Measuring on hot surfaces
Measurements taken at higher temperatures (above 100° C / 212° F) require special consideration. Both material sound velocity and probe zero will change with temperature. For maximum accuracy at high temperatures, adjustment should be performed using a material of known thickness heated to the temperature where measurements are to be performed. The probe should remain on the surface
only as long as it takes to get a measurement.
Gage readings appear to be double the expect­ed thickness
This sometimes occurs near the minimum measur­ing range of the instrument (0.100" or 2.5mm). The first return echo is unmeasurable, so the gage meas­ures the second return echo. The resultant calcula­tion is double the actual thickness.
18
This image appears if the memory of the
gage has become corrupt. This can occur in the event that the gage batteries were removed while the instrument was powered-on and the gage was not able to self power-down. If this message appears it will be followed by a full reset. All readings in memory will be erased and gage settings will be reset to "out-of-the-box" settings.
To retain all user settings and stored mem­ory readings, only replace the batteries after the Gage has automatically powered-down.
-Nickel-cadmium and nickel-metal hydride recharge­able batteries will work but the Gage may appear to have weak batteries.
Changing The Batteries
Changing The Batteries
The battery icon displays four bars with fresh alkaline batteries installed. As the batteries weaken, the number of bars will be reduced. When the bat­teries become very weak the Power Warning image will display and the Gage will automatically power-down. To main­tain user settings and stored meas­urements, the batteries must be replaced. USE ONLY “AAA” ALKALINE BATTERIES
NOTES:
Power Warning
Measurement jumps as probe is lifted from surface
Occasionally, excess couplant will remain on the probe when the probe is lifted from the surface. This may cause the PosiTector UTG to display a final measurement value different from those observed when the probe was on the surface. Discard this value and repeat the measurement.
Gage displays -.-- (dashed lines) when measur­ing on an smooth uncoated surface in ME mode
See ME Mode Note, pg. 6
Technical Data
Technical Data
Conforms to: ASTM E797
19
Returning for Service
Returning for Service
Before returning the Gage for service…
1.Install new Alkaline batteries in the proper
alignment as shown within battery compartment.
2.Examine the probe tip for dirt or damage.
3.Perform a Reset (pg.16) and a Zero (pg.6)
If you must return the Gage for service, describe the problem fully and include measurement results, if any. Be sure to also include the probe, your company name, company contact, telephone number and fax number or email address.
Website: www.defelsko.com/support
Measurement Range*
Single-Echo Multiple-Echo
0.100" to 5.000"
0.100" to 2.500"
2.50 to 125 mm
2.50 to 60 mm
Resolution
0.001 in. 0.01 mm
Accuracy
+ 0.001 in. + 0.03 mm
Velocity Range
0.0492 to 0.393 in/µs 1250 to 10,000 m/s
Measurement Rate
3 readings/second
Gage: 0 to 50° C (+32° to +120° F) Probe: -20° to 55° C (-4° to +131° F) continuous Material Surface Temp -40° to +150° C (-40° to +300° F)
Operating Range:
*Measurement range is for carbon steel and depends upon surface condition, temperature and material.
Gage body dimensions: 146 x 64 x 31 mm (5.75" x 2.5" x 1.2")
Battery Life: 80 hours continuous ( >100 hours idle)
www.defelsko.com
© DeFelsko Corporation USA 2007
All Rights Reserved
This manual is copyrighted with all rights reserved and may not be reproduced or transmitted, in whole or part, by any means, without written permission from DeFelsko Corporation.
DeFelsko, PosiTector and PosiSoft are trademarks of DeFelsko Corporation regis­tered in the U.S. and in other countries. Other brand or product names are trade­marks or registered trademarks of their respective holders.
Every effort has been made to ensure that the information in this manual is accu­rate. DeFelsko is not responsible for printing or clerical errors.
20
Limited Warranty, Sole Remedy
Limited Warranty, Sole Remedy
and Limited Liability
and Limited Liability
DeFelsko's sole warranty, remedy, and liability are the express limited warranty, remedy, and limited liability that are set forth on its website:
www.defelsko.com/terms
Available Options
Available Options
A variety of accessories are available to help you get the most from your PosiTector UTG ultrasonic thickness gage.
Loading...