3
Operation Overview
Operation Overview
Upon power-up, the following 5 parameters are 
displayed and updated automatically each second.
RH - The relative humidity sensor is located inside
the white plastic housing. The sensor may require a 
longer period to stabilize in the event of extreme 
humidity changes.
Ta - The air temperature sensor is also located
inside the white plastic housing. 
Ts - A surface temperature thermocouple is located
on the tip of the silver probe. It is spring loaded and 
designed to quickly self-align on a surface. Press it 
against a surface and allow a few seconds for the 
value to stabilize. Do not drag the sensor over the 
surface. Wet the surface for a faster response.
For surface temperature measurement over 
a long period of time, consider using the optional 
Magnetic Surface Temperature kit (pg. 11).
Td - The dew point temperature is a function of air
temperature and relative humidity (see ISO 8502-4 
Annex A). It is the temperature to which a volume of 
air must be cooled in order to reach saturation.
- The delta value is the surface minus dew point
temperature (
Ts - Td). This value is important in
judging the risk of condensation, or dew, forming on 
a surface. For example, ISO 8502-4 recommends 
that, unless otherwise agreed, the temperature of a 
steel surface generally should be at least 3°C (5°F) 
above the dew point when paints are applied.
The instrument may require a period of time 
before the above climatic parameters stabilize. This 
is most noticeable when the instrument is relocated 
from one extreme environment to another (i.e. from 
a warm location to a cold location). A period of up to 
30 minutes should provide adequate time for 
stabilization.
NOTE:
NOTE: