Datacolor AHIBA IR, AHIBA IR Pro Installation Manual

AHIBA IR & IR Pro
Quick Installation Guide
IMPORTANT: ALWAYS REFER TO USER MANUAL FOR MORE COMPLETE & DETAILED INSTRUCTIONS
Power: 230V ( 10%), 15A, 50/60 Hz, single phase with earth ground.
It is recommended to use surge protected and dedicated line. Water (AHIBA IR Pro only): Use only filtered and softened water. Pressure: 2-3 bar (30-45 psi) Hardness: <10 dH Temperature: Ideally <15C Hose: Min. diameter 3/8” (10mm), Hi Temp. Unit Foot Print: 864 x 597 x 648mm (H x W x D) minimum
(For detailed external dimensions and footprint – Refer to Technical Data section in User Manual)
2. Open top of shipping box.
Remove system accessories box carefully. Caution: Accessories box contains sensitive and fragile parts; handle with care to
prevent damage.
Check that the shipment is complete (Refer to included Packing Slip). The unit’s controller and
mounting bracket are contained within a separate box. Remove the controller and mounting bracket from this box and mount the controller on the unit using the screws mounted in the top of the unit.
Be sure to connect controller grounding wire as shown in the diagram (Fig. 01). Connect the I/O cable from the unit to the controller’s rear panel. Ensure screws are tight. Remove the unit from its pallet (caution: unit weighs approx. 74Kg (163 lb), a minimum of 2
people are required to lift).
Place unit of level and stable table/bench top with a minimum of 140 Kg load capacity.
3. Provide sufficient space between the back of unit and the wall. The rear of the unit must be a minimum of 6 inches from any surface. This is required for adequate airflow for the instrument fans and will prevent kinking of hoses if the unit is an AHIBA IR Pro.
4. During shipment the unit might be stored where ambient temperature is low. This could cause the thermostats to trip. Press the Red Reset button located on the rear panel of the unit. (See fig. 02)
5. Ensure IR lamp circuit breaker is “on”. (See fig. 02)
6. Refer to labels on the back of unit for proper connection.
(AHIBA IR Pro Only) Caution: Improper installation of hoses can cause damage to the machine.
Important: The water outlet hose must always be channeled to a lower
elevation than machine water outlet.
Important!!! (AHIBA IR Pro only)
Before powering up the unit Slowly turn
inlet water on and check for any leaks.
Doc. No. 72-990120-English
Fig. 01
Fig. 02
November 23, 2017
No.288 Shengpu Road, Suzhou Industrial Park, Export Processing Zone B, Suzhou, Jiangsu, P.R. China P +011 86.512.625.80338 F +011 86.512.625.80339
www.datacolor.com
7. Loading the beakers
For best results adhere to the following:
Do not mix beakers sizes: Be consistent.
o Only install beakers of the same size. (Ex. Do not mix 150cc with 300cc) o Do not mix black coated beakers with stainless steel beakers. o Do not over tighten the beaker lids
Over tightening can lead to o-ring wear and damage to the
beaker lid.
Beaker Tightening Instructions
o Hand tighten the beaker lid closing screw firstly o Then using the supplied Hex driver turn the beaker lid screw
approximately ¼ - ½ turn should be sufficient closing and sealing pressure for the beaker lid to avoid leaking of the solution.
If additional tightening is needed to seal it could mean and issue with the O-ring or other beaker components and may need
replacement.
8. Preparing samples
When performing a job, always fill the beakers with equal amounts of sample and dye liquor. Do not under- fill or over fill a beaker. Refer to applications section of User Manual
for detailed information. (Note: The o-ring in the beaker lid can push out to release pressure caused by an improperly filled beaker. If this happens re-install the o-ring and check sample-to-liquor ratio).
Samples should be wet out, rung and folded Always ensure that the reference beaker contains a sample. Never perform a
job without the
o reference beaker as part of the dyeing process.
9. Sample Folding
Fabric should be folded as shown in this example.
Do not crumple fabric and stuff into beaker.
10. Beaker Placement
Always install the reference beaker in a position where the sensor can be easily connected to the
connection port on the beaker lid. Avoid heavy bending of the sensor cable.
Fill the rest of the wheel evenly, always maintaining wheel balance no matter how many beakers
are installed.
Sensor is mounted to the beaker wheel hub with a locking collar; it should not be removed from
this point.
o Always connect sensor after the reference beaker is installed on the beaker wheel and
always remove the sensor cable from the reference beaker before removing the beaker from the wheel.
Unnecessary stress and bending of the sensor can lead to premature failure.
11. Cleaning
Refer to user manual for more detailed information on cleaning. The unit should be cleaned as needed with a mild soap and clean cloth to ensure inner and outer casing remains clean. ALL accessories must be thoroughly cleaned and dried after each dye process to remove any residual dyestuffs and chemical auxiliaries to
avoid dye cross contamination and chemical sedimentation on the accessories surface.
Loading...
+ 1 hidden pages