Datacolor 9020-0129, 9120-0129, 9020-0190, 9120-0190 Operating Manual

Operating Manual
Original operating manual
Datacolor CONDITIONER™
Sample conditioning cabinet with display controller MB2
9020-0129, 9120-0129 without lights 9020-0190, 9120-0190 with lights
Issue 06/2017 Art. No. 7001-0346
Content
1. SAFETY .................................................................................................................. 6
1.1 Legal considerations ........................................................................................................................... 6
1.2 Structure of the safety instructions ...................................................................................................... 6
1.2.1 Signal word panel ...................................................................................................................... 6
1.2.2 Safety alert symbol .................................................................................................................... 7
1.2.3 Pictograms ................................................................................................................................. 7
1.2.4 Word message panel structure ................................................................................................. 8
1.3 Localization / position of safety labels on the cabinet ......................................................................... 8
1.4 Type plate ........................................................................................................................................... 9
1.5 General safety instructions on installing and operating the cabinets ............................................... 10
1.6 Intended use ..................................................................................................................................... 12
1.7 Operating instructions ....................................................................................................................... 12
1.8 Measures to prevent accidents ......................................................................................................... 13
1.9 Resistance of the humidity sensor against harmful substances ....................................................... 14
2. CABINET DESCRIPTION ..................................................................................... 15
2.1 Cabinet overview ............................................................................................................................... 16
2.2 Instrument panel ............................................................................................................................... 17
2.3 Lateral control panel ......................................................................................................................... 17
2.4 Rear view with water connections .................................................................................................... 18
3. COMPLETENESS OF DELIVERY, TRANSPORTATION, STORAGE, AND
INSTALLATION .................................................................................................... 19
3.1 Unpacking, and checking equipment and completeness of delivery ................................................ 19
3.2 Guidelines for safe lifting and transportation .................................................................................... 19
3.3 Storage .............................................................................................................................................. 20
3.4 Location of installation and ambient conditions ................................................................................ 21
4. INSTALLATION AND CONNECTIONS................................................................ 22
4.1 Spacer for wall distance .................................................................................................................... 22
4.2 Wastewater connection ..................................................................................................................... 23
4.3 Freshwater supply ............................................................................................................................. 23
4.3.1 Automatic fresh water supply via water pipe ........................................................................... 24
4.3.2 Manual fresh water supply via external freshwater can (option) ............................................. 26
4.4 Optional water treatment system BINDER Pure Aqua Service ........................................................ 26
4.5 Optional external freshwater and wastewater cans .......................................................................... 26
4.5.1 Mounting the freshwater can ................................................................................................... 27
4.5.2 Mounting the wastewater can.................................................................................................. 28
4.5.3 Mounting with wastewater recycling ........................................................................................ 29
4.6 Placement of the light cassettes (cabinet with lights) ....................................................................... 30
4.7 Electrical connection ......................................................................................................................... 30
4.8 Connection of the optional voltage changer ..................................................................................... 31
5. FUNCTIONAL OVERVIEW OF THE MB2 CHAMBER CONTROLLER ............... 33
5.1 Operating functions in normal display ............................................................................................... 35
5.2 Display views: Normal display, program display, chart-recorder display ......................................... 36
5.3 Controller icons overview .................................................................................................................. 37
5.4 Operating modes ............................................................................................................................... 39
5.5 Controller menu structure ................................................................................................................. 40
5.5.1 Main menu ............................................................................................................................... 41
5.5.2 “Settings” submenu ................................................................................................................. 42
5.5.3 “Service” submenu .................................................................................................................. 42
5.6 Principle of controller entries ............................................................................................................. 43
5.7 Performance during and after power failures .................................................................................... 43
5.8 Performance when opening the door ................................................................................................ 44
Datacolor CONDITIONER™ (MB2) 06/2017 page 2/126
6. START UP ............................................................................................................ 44
6.1 Turning on the cabinet ...................................................................................................................... 44
6.2 Controller settings upon start up ....................................................................................................... 44
6.3 Turning on/off humidity control .......................................................................................................... 45
7. SET-POINT ENTRY IN “FIXED VALUE” OPERATING MODE ........................... 46
7.1 Set-point entry for temperature, humidity, light intensity, and fan speed through the “Setpoints”
menu ................................................................................................................................................. 47
7.2 Direct setpoint entry for temperature, humidity, and light int ensity via Normal display .................... 48
7.3 Special controller functions via operation lines ................................................................................. 48
8. TIMER PROGRAM: STOPWATCH FUNCTION .................................................. 49
8.1 Starting a timer program ................................................................................................................... 49
8.1.1 Performance during program delay time ................................................................................. 49
8.2 Stopping a running timer program .................................................................................................... 50
8.2.1 Pausing a running timer program ............................................................................................ 50
8.2.2 Cancelling a running timer program ........................................................................................ 50
8.3 Performance after the end of the program ........................................................................................ 50
9. TIME PROGRAMS ............................................................................................... 51
9.1 Starting an existing time program ..................................................................................................... 51
9.1.1 Performance during program delay time ................................................................................. 52
9.2 Stopping a running time program ...................................................................................................... 52
9.2.1 Pausing a running time program ............................................................................................. 52
9.2.2 Cancelling a running time program ......................................................................................... 52
9.3 Performance after the end of the program ........................................................................................ 52
9.4 Creating a new time program ............................................................................................................ 53
9.5 Program editor: program management ............................................................................................. 53
9.5.1 Deleting a time program .......................................................................................................... 54
9.6 Section editor: section management ................................................................................................ 55
9.6.1 Add a new program section ..................................................................................................... 56
9.6.2 Copy and insert or replace a program section ........................................................................ 56
9.6.3 Deleting a program section ..................................................................................................... 57
9.7 Value entry for a program section ..................................................................................................... 58
9.7.1 Section duration....................................................................................................................... 58
9.7.2 Set-point ramp and set-point step ........................................................................................... 59
9.7.3 Special controller functions via operation lines ....................................................................... 60
9.7.4 Setpoint entry .......................................................................................................................... 61
9.7.5 Tolerance range ...................................................................................................................... 61
9.7.6 Repeating one or several sections within a time program ...................................................... 62
9.7.7 Saving the time program ......................................................................................................... 63
10. WEEK PROGRAMS ............................................................................................. 64
10.1 Starting an existing week program.................................................................................................... 64
10.2 Cancelling a running week program ................................................................................................. 64
10.3 Creating a new week program .......................................................................................................... 65
10.4 Program editor: program management ............................................................................................. 66
10.4.1 Deleting a week program ........................................................................................................ 67
10.5 Section editor: Section management ................................................................................................ 68
10.5.1 Add a new program section ..................................................................................................... 69
10.5.2 Copy and insert or replace a program section ........................................................................ 69
10.5.3 Deleting a program section ..................................................................................................... 70
10.6 Value entry for a program section ..................................................................................................... 70
10.6.1 Set-point ramp and set-point step modes ............................................................................... 70
10.6.2 Weekday .................................................................................................................................. 71
10.6.3 Start time ................................................................................................................................. 71
10.6.4 Setpoint entry .......................................................................................................................... 72
10.6.5 Special controller functions via operation lines ....................................................................... 72
Datacolor CONDITIONER™ (MB2) 06/2017 page 3/126
11. NOTIFICATION AND ALARM FUNCTIONS ........................................................ 73
11.1 Notification and alarm messages overview ...................................................................................... 73
11.1.1 Notifications ............................................................................................................................. 73
11.1.2 Alarm messages ...................................................................................................................... 74
11.1.3 Messages concerning the humidity system ............................................................................ 74
11.2 State of alarm .................................................................................................................................... 75
11.3 Resetting an alarm, list of active alarms ........................................................................................... 75
11.4 Tolerance range settings .................................................................................................................. 76
11.5 Activating / deactivating the audible alarm (alarm buzzer) ............................................................... 77
12. TEMPERATURE SAFETY DEVICES ................................................................... 77
12.1 Over temperature protective device (class 1) ................................................................................... 77
12.2 Overtemperature safety controller class 3.1 ..................................................................................... 77
12.2.1 Safety controller modes ........................................................................................................... 78
12.2.2 Setting the safety controller ..................................................................................................... 78
12.2.3 Message and measures in the state of alarm ......................................................................... 79
12.2.4 Function check ........................................................................................................................ 79
13. USER MANAGEMENT ......................................................................................... 80
13.1 Authorization levels and password protection .................................................................................. 80
13.2 Log in................................................................................................................................................. 83
13.3 Log out .............................................................................................................................................. 84
13.4 User change ...................................................................................................................................... 84
13.5 Password assignment and password change .................................................................................. 85
13.5.1 Password change .................................................................................................................... 85
13.5.2 Deleting the password for an individual authorization level .................................................... 87
13.5.3 New password assignment for “service” or “admin” authorization level when the password
function was deactivated ........................................................................................................ 88
13.6 Activation code .................................................................................................................................. 89
14. GENERAL CONTROLLER SETTINGS ................................................................ 90
14.1 Selecting the controller’s menu language ......................................................................................... 90
14.2 Selecting the temperature unit .......................................................................................................... 90
14.3 Setting date and time ........................................................................................................................ 91
14.4 Display configuration ......................................................................................................................... 92
14.4.1 Adapting the display parameters ............................................................................................. 92
14.4.2 Touchscreen calibration .......................................................................................................... 93
14.5 Network and communication ............................................................................................................. 94
14.5.1 Ethernet ................................................................................................................................... 94
14.5.2 Web server .............................................................................................................................. 95
14.5.3 E-Mail ...................................................................................................................................... 96
14.6 USB menu ......................................................................................................................................... 97
15. GENERAL INFORMATION .................................................................................. 98
15.1 Service contact page ........................................................................................................................ 98
15.2 Current operating parameters ........................................................................................................... 98
15.3 Event list ............................................................................................................................................ 99
15.4 Counter of operating hours (cabinet with lights) ............................................................................. 100
15.5 Technical information on the cabinet .............................................................................................. 100
16. CHART RECORDER DISPLAY ......................................................................... 101
16.1 Views ............................................................................................................................................... 101
16.1.1 Show and hide legend ........................................................................................................... 101
16.1.2 Switch between legend pages .............................................................................................. 101
16.1.3 Show and hide specific indications ....................................................................................... 102
16.1.4 History display ....................................................................................................................... 102
16.2 Setting the parameters .................................................................................................................... 105
Datacolor CONDITIONER™ (MB2) 06/2017 page 4/126
17. HUMIDIFICATION / DEHUMIDIFICATIO N SYSTEM ......................................... 106
17.1 Function of the humidifying and dehumidifying system .................................................................. 108
18. DEFROSTING AT REFRIGERATING OPERATION (CABINET WITHOUT
LIGHTS) ............................................................................................................. 109
19. ILLUMINATION SYSTEM (CABINET WITH LIGHTS) ....................................... 110
20. MAINTENANCE, CLEANING, AND SERVICE .................................................. 111
20.1 Maintenance intervals, service ........................................................................................................ 111
20.2 Replacement of the fluorescent tubes ............................................................................................ 112
20.3 Cleaning and decontamination ....................................................................................................... 112
20.3.1 Cleaning ................................................................................................................................ 113
20.3.2 Decontamination ................................................................................................................... 115
21. DISPOSAL.......................................................................................................... 116
21.1 Disposal of the transport packing ................................................................................................... 116
21.2 Decommissioning ............................................................................................................................ 116
21.3 Disposal of the cabinet in the member states of the EU ................................................................. 116
21.4 Disposal of the cabinet in non-member states of the EU ............................................................... 117
22. TROUBLESHOOTING ....................................................................................... 118
23. TECHNICAL DESCRIPTION .............................................................................. 120
23.1 Factory calibration and adjustment ................................................................................................. 120
23.2 Over current protection ................................................................................................................... 120
23.3 Definition of usable volume ............................................................................................................. 120
23.4 Technical Data ................................................................................................................................ 121
23.5 Equipment ....................................................................................................................................... 122
23.6 Spare parts (extract) ....................................................................................................................... 123
23.7 Dimensions size 115 ....................................................................................................................... 124
24. EU DECLARATION OF CONFORMITY ............................................................. 125
All efforts have been made to ensure the accuracy of the information presented in this format. However, should any errors be detected, Datacolor appreciates your efforts to notify us of these oversights.
Changes are periodically made to this information and are incorporated into forthcoming versions. Data­color reserves the right to make improvements and/or changes in the product(s) and/or program(s) de­scribed in this material at any time.
©2010 Datacolor. Datacolor, and other Datacolor product trademarks are the property of Datacolor.
Datacolor CONDITIONER™ (MB2) 06/2017 page 5/126
Dear customer,
For the correct operation of the Datacolor CONDITIONER™ sample conditioning cabinet, it is important that you read this operating manual completely and carefully and observe all instructions as indicated. Failure to read, understand and follow the instructions may result in personal injury. It can also lead to damage to the cabinet and/or poor equipment performance.

1. Safety

This operating manual is part of the components of delivery. Always keep it handy for reference. The device should only be operated by laboratory personnel especially trained for this purpose and familiar with all precautionary measures required for working in a laboratory. Observe the national regulations on minimum age of laboratory personnel. To avoid injuries and damage observe the safety instructions of the operating manual.
WARNING
Failure to observe the safety instructions. Serious injuries and cabinet damage.
Observe the safety instructions in this operating manual. Carefully read the complete operating instructions of the cabinets.

1.1 Legal considerations

This operating manual is for informational purposes only. It contains information for installing, start-up, operation and maintenance of the product. Note: the contents and the product described are subject to change without notice.
Understanding and observing the instructions in this operating manual are prerequisites for hazard-free use and safety during operation and maintenance. In no event shall Datacolor be held liable for any dam­ages, direct or incidental arising out of or related to the use of this manual.
This operating manual cannot cover all conceivable applications. If you would like additional information, or if special problems arise that are not sufficiently addressed in this manual, please ask your dealer or contact us directly by phone at the number located on page one of this manual
Furthermore, we emphasize that the contents of this operating manual are not part of an earlier or exist­ing agreement, description, or legal relationship, nor do they modify such a relationship. All obligations on the part of Datacolor derive from the respective purchase contract, which also contains the entire and exclusively valid statement of warranty administration. The statements in this manual neit her augm ent nor restrict the contractual warranty provisions.

1.2 Structure of the safety instructions

In this operating manual, the following safety definitions and symbols indicate dangerous situations fol­lowing the harmonization of ISO 3864-2 and ANSI Z535.6.

1.2.1 Signal word panel

Depending on the probability of serious consequences, potential dangers are identified with a signal word, the corresponding safety color, and if appropriate, the safety alert symbol.
DANGER
Indicates an imminently hazardous situation that, if not avoided, will result i n death or serious (irreversi­ble) injury.
Datacolor CONDITIONER™ (MB2) 06/2017 page 6/126
Warning signs
Hot surface
Danger of frost
or chemical burns
Mandatory action signs
instructions
plug
Wear protective gloves
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious (irre­versible) injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in m oderat e or m inor (re­versible) injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in dam age to the product and/or its functions or of a property in its proximity.

1.2.2 Safety alert symbol

Use of the safety alert symbol indicates a risk of injury. Observe all measures that are marked with the safety alert symbol in order to avoid death or
injury.

1.2.3 Pictograms

Electrical hazard
Lifting hazard
Risk of corrosion and /
Scalding hazard
Harmful substances
Explosive atmosphere
High humidity
Biohazard
Stability hazard
Pollution Hazard
Mandatory regulation
Environment protection
Datacolor CONDITIONER™ (MB2) 06/2017 page 7/126
Read operating
Disconnect the power
Wear safety goggles
Lift with several persons
Prohibition signs
water
Pictograms (warning signs)
Do NOT touch
Do NOT spray with
Information to be observed in order to ensure optimum function of the product.
Do NOT climb

1.2.4 Word message panel structure

Type / cause of hazard. Possible consequences.
Instruction how to avoid the hazard: prohibition
Instruction how to avoid the hazard: mandatory action.
Observe all other notes and information not necessarily emphasized in the same way, in order to avoid disruptions that could result in direct or indirect injury or property damage.

1.3 Localization / position of safety labels on the cabinet

The following labels are located on the cabinet:
Hot surface (inner glass door above the glass door handle)
Burning and scalding hazard (cabinet rear)
Datacolor CONDITIONER™ (MB2) 06/2017 page 8/126
Figure 1: Position of labels on the cabinet rear
Indications of the type plate (example)
Information
datacolor
Distributor: Datacolor
BINDER GmbH
Manufacturer: BINDER GmbH
Datacolor CONDITIONER™
Model designation
Sample conditioning cabinet
Device name
Built
2017
Serial no. of the cabinet
Serial No.
00000000000000
Year of construction
Nominal temperature
70 °C / 158 °F
Nominal temperature
IP protection
20
IP type of protection acc. to standard EN 60529
Temp. safety device
DIN 12880
Temperature safety device acc. to standard DIN 12880:2007
Class
3.1
Class of temperature safety device
Art. No.
9020-0129
Art. no. of the cabinet
Project No.
---
Optional: Special application acc. to project no.
2,0 kW
Nominal power
8,4 A
Nominal current
200-240 V / 50 Hz
200-240 V / 60 Hz
2 ~
Current type
Max operating pressure in the refrigerating system (15 bar / 218 PSI)
R 134A - 0,180 kg
Refrigerant type and filling weight
Contains fluorinated greenhouse gases covered by the Kyoto Protocol
Symbol on the type plate
Information
Nominal temp.
70 °C
2,00 kW / 8,4 A
Max. operating pressure 15 bar
158 °F
200-240 V / 50 Hz
R 134A – 0,180 kg
IP protection
20
200-240 V / 60 Hz
Contains fluorinated
Safety device
DIN 12880
2 ~
greenhouse gases covered
Class
3.1 by the Kyoto Protocol
Art. No.
9020-0129
Built
2017
Datacolor CONDITIONER™ Sample conditioning cabinet
BINDER GmbH
www.binder-world.com
KBF115 - DC
Serial No. 00000000000000
Keep safety labels complete and legible.
Replace safety labels that are no longer legible. Contact your local Datacolor service representative for these replacements.

1.4 Type plate

The type plate sticks to the left side of the cabinet, bottom right-hand.
Project No.
Made in Germany
Im Mittleren Ösch 5 78532 Tuttlingen / Germany
Figure 2: Type plate (example: cabinet without lights)
Nominal voltage range +/-10% at the indicated power frequency
Max. operating pressure 15 bar
CE conformity marking Electrical and electronic equipment manufactured / placed on the mar-
ket in the EU after 13 August 2005 and be disposed of in separate collection according to Directive 2012/19/EU on waste electrical and electronic equipment (WEEE).
Datacolor CONDITIONER™ (MB2) 06/2017 page 9/126

1.5 General safety instructions on installing and operating the cabinets

With regard to operating the cabinets and to the installation location, please observe the local national safety regulations that apply for laboratories or in a room similar to a laboratory.
Datacolor is only responsible for the safety features of the cabinet provided skilled electricians or qualified personnel authorized by Datacolor perform all maintenance and repair, and if components relating to cabinet safety are replaced in the event of failure with original spare parts.
When the cabinet is installed, operated, cleaned, decontaminated, adjusted or set up incorrectly, there is a risk of malfunction which could cause harm to human beings and material damage to the equipment and samples.
Therefore the cabinet should only be installed, operated, cleaned, decontaminated, adjusted and set up by suitably qualified persons.
Persons qualified to install, operate, clean, and decontaminate the cabinet only include laboratory
personnel who have carefully read the operating manual and are especially trained for this purpose. Observe the national regulations on minimum age of laboratory personnel.
Persons qualified to repair and adjust the cabinet only include electricians and service engineers au-
thorized by the Datacolor, who have undergone appropriate electrical training and who have carefully read the cabinet service and operating manuals.
To operate the cabinet, use only original Datacolor accessories or accessories from third-party suppliers authorized by Datacolor. The user is responsible for any risk caused by using unauthorized accessories.
CAUTION
Danger of overheating. Damage to the cabinet.
Do NOT install the cabinet in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
Do not operate the cabinets in hazardous locations.
DANGER
Explosion hazard. Danger of death.
Do NOT operate the cabinet in potentially ex pl osiv e areas. KEEP explosive dust or air-solvent mixtures AWAY from the cabinet.
The cabinets do not dispose of any measures of explosion protection.
DANGER
Explosion hazard. Danger of death.
Do NOT introduce any substance into the cabinet which is combustible or explosiv e at
working temperature.
NO explosive dust or air-solvent mixture in the inner chamber.
Any solvent contained in the charging material must not be explosive or inflammable. I.e., irrespective of the solvent concentration in the steam room, NO explosive mixture with air must form. The temperature inside the chamber must lie below the flash point or below the sublimation point of the charging material. Familiarize yourself with the physical and chemical properties of the charging material, as well as the contained moisture constituent and its behavior with the addition of heat energy and hum i dity.
Datacolor CONDITIONER™ (MB2) 06/2017 page 10/126
Familiarize yourself with any potential health risks caused by the charging material, the contained mois­ture constituent or by reaction products that may arise during the temperature process. Take adequate measures to exclude such risks prior to putting the cabinet into operation.
DANGER
Electrical hazard. Danger of death.
The cabinet must NOT become wet during operation or maintenance.
The cabinets were produced in accordance with VDE regulations and were routinely tested in accordance to VDE 0411-1 (IEC 61010-1).
During and shortly after operation, the temperature of the inner surfaces almost equals the set-point.
CAUTION
The glass doors, the glass door handles, the inner chamber, and the light cassettes (cabinet with lights) will become hot during operation.
Danger of burning.
Do NOT touch the glass doors, the inner surfaces, the light cassettes or the charging
material during operation.
WARNING
Stability hazard. Danger of injury. Damage to the cabinet and the charging material. Housing cover breakaway.
Do NOT climb on the lower housing cover. Do NOT load the lower housing cover and the door with heavy objects while the cabi-
net door is open.
Datacolor CONDITIONER™ (MB2) 06/2017 page 11/126

1.6 Intended use

Datacolor CONDITIONER™ sample conditioning cabinets are suitable for exact conditioning of harmless materials. A mixture of any component of the charging material with air must NOT be explosive. The op­erating temperature must lie below the flash point or below the sublimation point of the charging material. Any component of the charging material must NOT be able to release toxic gases.
Other applications are not approved. The cabinets are not classified as medical devices as defined by the Medical Device Directive
93/42/EEC.
Following the instructions in this operating manual and conducting regular maintenance work (chap. 20) are part of the intended use.
DANGER
Explosion or implosion hazard. Danger of poisoning. Danger of death.
Do NOT introduce any substance combustible or explosive at working temperature into
the cabinet, in particular no energy sources such as batteries or lithium-ion batteries.
NO explosive dust or air-solvent mixture in the inner chamber. Do NOT introduce any substance which could lead to release of toxic gases.
The charging material shall not contain any corrosive ingredients that may damage the ma­chine components made of stainless steel, aluminum, and copper. Such ingredients include in particular acids and halides. Any corrosive damage caused by such ingredients is excluded from liability by Datacolor.
WARNING: If customer should use a cabinet running in non-supervised continuous opera­tion, we strongly recommend in case of inclusion of irrecoverable specimen or samples to split such specimen or samples and store them in at least two cabinets, if this is feasible.
In case of foreseeable use of the cabinet there is no risk for the user through the integration of the cabi­net into systems or by special environmental or operating conditions in the sense of EN 61010-1:2010. For this, the intended use of the cabinet and all its connections must be observed.

1.7 Operating instructions

Depending on the application and location of the cabinet, the operator of the cabinet must provide the relevant information for safe operation of the cabinet in a set of operating instructions.
Keep these operating instructions with the cabinet at all times in a place where they are clearly visible. They must be comprehensible and written in the language of the employees.
Datacolor CONDITIONER™ (MB2) 06/2017 page 12/126

1.8 Measures to prevent accidents

The operator of the cabinet must observe the following rule: “Betreiben von Arbeitsmitteln. Betreiben von Kälteanlagen, Wärmepumpen und Kühleinrichtungen“ (Operation of work equipment. Operation of refrig­eration systems, heat pumps and refrigeration equipment) (GUV-R 500 chap. 2.35) (for Germany).
The manufacturer took the following measures to prevent ignition and explosions:
Indications on the type plate
See operating manual chap. 1.4.
Operating manual
An operating manual is available for each cabinet.
Overtemperature monitoring
The cabinet is equipped with a temperature display, which can be read from outside. The cabinet is equipped with an additional safety controller (temperature safety device class 3.1 acc.
to DIN 12880:2007). Visual and audible (buzzer) signals indicate temperature exceeding.
Safety, measurement, and control equipment
The safety, measuring, and control equipment is easily accessible.
Electrostatic charge
The interior parts are grounded.
Non-ionizing radiation
Non-ionizing radiation is not intentionally produced, but released only for technical reasons by electri­cal equipment (e.g. electric motors, power cables, solenoids). The machine has no permanent mag­nets. If persons with active implants (e.g. pacemakers, defibrillators) keep a safe distance (distance of field source to implant) of 30 cm, an influence of these implants can be excluded with high probabil i ty.
Protection against touchable surfaces
Tested according to EN ISO 13732-1:2008.
Floors
See operating manual chap. 3.4 for correct installation
Cleaning
See operating manual chap. 20.3.
Examinations
The chamber has been inspected by the “Deutsche Gesetzliche Unfallversicherung e.V. (DGUV) (German Social Accident Insurance (DGUV)” (German Social Accident Insurance (DGUV), Testing and Certification Body for Foodstuffs and Packaging Industry in DGUV Test).
Datacolor CONDITIONER™ (MB2) 06/2017 page 13/126
Maximum work place threshold limit value
Tolerated concentration with permanent load
Substance
Formula
ppm
mg/m3
ppm
mg/m3
Ammonia
NH3
20
14
5500
4000
Acetone
CH3COCH3
500
1200
3300
8000
Benzene
300
1200
150000
Chlorine
Cl2
0.5
1.5
0.7 2 Acetic acid
CH3COOH
10
25
800
2000
Ethyl acetate
CH3COOC2H5
400
1400
4000
15000
Ethanol
C2H5OH
500
960
3500
6000
Ethylene glycol
HOCH2CH2OH
10
26
1200
3000
Formaldehyde
HCHO
0.3
0.37
2400
3000
Isopropanol
(CH3)2CHOH
200
500
4800
12000
Methanol
CH3OH
200
260
3500
6000
Methyl ethyl ketone
C2H5COCH3
200
590
3300
8000
Ozone
O3
0.1
0.2
0.5 1 Hydrochloric acid
HCl 2 3
300
500
Hydrogen sulphide
H2S
10
15
350
500
Nitrogen oxides
NOx
5 9 5 9 Sulphur dioxide
SO2
5
13 5 13
Toluol
C6H5CH3
100
380
1300
5000
Xylene
C6H4(CH3)2
100
440
1300
5000

1.9 Resistance of the humidity sensor against harmful substances

The following list of harmful substances refers only to the humidity sensor and does not include any other materials incorporated in the cabinet or prohibited substances in relation to explosion protection.
Some gases - especially clean gases - do not have any influence on the humidity sensor. Others do have a very small influence, whereas others may influence the sensor to a larger extent.
The following gases do not influence the sensor and the humidity measurement: Argon (Ar), carbon
dioxide (CO
The following gases do not or to a minor extent influence the sensor and the humidity measurement:
Butane (C
The following gases do not, or to a minor extent influence the sensor and the humidity measurement,
provided that the indicated loads are not exceeded:
),helium (He), hydrogen (H2), neon (Ne), nitrogen (N2), nitrous oxide (N2O), oxygen (O2)
2
), ethane (C2H6), methane (CH4), natural gas propane (C3H8)
4H10
These values are to be considered as approximate values. The sensor resistance largely depends on the temperature and humidity conditions during the time of exposure to harmful substances. Avoid simultaneous condensation. Tolerated error of measurement: ± 2 %r.H. The maximum work place threshold limit value is one that can be regarded as harmless for humans.
Vapors of oil and fat are dangerous for the sensor because they may condensate at the sensor and
thus prevent its function (insulating layer). For similar reasons it is not possible to measure smoke gases.
Datacolor CONDITIONER™ (MB2) 06/2017 page 14/126

2. Cabinet description

The Datacolor CONDITIONER™ sample conditioning cabinet is equipped with a multifunctional micro­processor display controller with 2-channel technology for temperature and humidity plus a digital display accurate to one-tenth of a degree resp. 0.1% r.H. With its comprehensive program control functions, the display program controller MB2 permits the high precision performance of temperature and humidity cy­cles.
The sample conditioning cabinet with multifunctional microprocessor display controller MB2 is available in two different types:
without light unit
with light unit
The sample conditioning cabinet is designed for exact conditioning of harmless materials such as paper, textiles, plastics, building materials, to condition the samples at fixed temperature and humidity, so as ensure no change in color of the samples occurs due to varying temperature and humidity
The APT.line™ preheating chamber system guarantees high level of spatial and time-based temperature precision, thanks to the direct and distributed air circulation into the interior. The fan supports exact at­tainment and maintenance of the desired temperature accuracy. Furthermore, it permits simulating exact­ly and over long periods constant conditions for other applications such as sample conditioning for mate­rial testing of paper, textiles, plastics, building materials, etc.
A resistance humidifying system humidifies the air. For this purpose, use deionized (demineralized) wa­ter. The option BINDER Pure Aqua Service allows using the cabinet with any degree of water hardness.
The cabinet with lights unit is equipped with “Bio Vital” fluorescent tubes installed in two light cassettes. They illuminate very homogeneously the racks below them. Both light cassettes are permanently con­nected. The cable length permits easily removal to replace the fluorescent tubes. Control is effected via the MB2 controller. Light is dimmable in 1% steps.
The inner chamber, the pre-heating chamber and the interior side of the doors are all made of stainless steel V2A (German material no. 1.4301, US equivalent AISI 304). The housing is RAL 7035 powder­coated. All corners and edges are also completely coated.
The inner chamber is closed by a glass door that enables the samples to be viewed for a short time with­out disturbing the climate in the inner chamber.
All cabinet functions are easy and comfortable to use thanks to their clear arrangement. Major features are easy cleaning of all cabinet parts and avoidance of undesired contamination.
The efficient program controller is equipped with a multitude of operating functions, in addition to recorder and alarm functions. Programming of test cycles is easily accomplished via the modern MB2 touch screen controller. The cabinet comes equipped with an Ethernet serial interface for computer communi­cation. In addition, the BINDER communication software APT-COM™ 3 permits networking up to 30 chambers and connecting them to a PC for controlling and programming, as well as recording and repre­senting temperature and humidity data. For further options, see chap. 23.5.
The cabinet is equipped with two safety devices to protect the unit, its environment and the samples in­side from forbidden temperature excesses:
Electric safety controller
The sample conditioning cabinet is equipped with an electric over temperature safety device class 3.1 acc. to DIN 12880:2007. It is called a “safety controller”. It serves to protect the charging material against extensive high temperatures. The safety controller is electrically independent from the main temperature controller and takes over control when a selectable set point is reached.
Temperature fuse
The sample conditioning cabinet is equipped with a temperature fuse, class 1 acc. to DIN 12880:2007. It serves to protect the unit and prevents danger caused by considerable defects. If a temperature of about 110 °C is reached, the temperature fuse switches off the unit permanently.
Datacolor CONDITIONER™ (MB2) 06/2017 page 15/126
(A)
Temperature range Humidity range Cabinet without lights 0 °C / 32 °F up to 70 °C / 158 °F 10% r.H. to 80% r.H. Cabinet with lights 0 °C / 32 °F up to 50 °C / 122 °F 10% r.H. to 70% r.H.
For the control ranges of temperature and humidity, see climatic diagram (chap. 17). Standard temperature and humidity settings for sample conditioning are typically 21 °C and 65 % r.H.

2.1 Cabinet overview

(B)
(C)
(D)
Figure 3: Datacolor CONDITIONER™ sample conditioning cabinet
(A) Instrument box (B) Door handle (C) Outer door (D) Lower front cover: refrigerating machine and humidity generation module
Datacolor CONDITIONER™ (MB2) 06/2017 page 16/126

2.2 Instrument panel

5,7" controller display with touchscreen
USB interface
Pilot lamp
Figure 4: Instrument panel with MB2 program controller and USB interface

2.3 Lateral control panel

(1)
(2)
Figure 5: Lateral control panels at the right side of the refrigerating / humidity generation module
(1) Main power switch (2) Ethernet interface
Datacolor CONDITIONER™ (MB2) 06/2017 page 17/126
(3)

2.4 Rear view with water connections

(4)
Figure 6: Water connections on the cabinet rear
(3) Socket for optional freshwater can (chap. 4.5) (4) Power cable (5) Purging outlet for humidifying module – for service purpose only (6) Freshwater connection “IN” with screw thread ¾’’ for hose ½“, with union nut (7) Wastewater connection “OUT” with hose olive for hose ½“
(7)
(6)
(5)
Datacolor CONDITIONER™ (MB2) 06/2017 page 18/126
3. Completeness of delivery, transportation, storage, and installa-
tion

3.1 Unpacking, and checking equipment and completeness of delivery

After unpacking, please check the cabinet and its optional accessories, if any, based on the delivery re­ceipt for completeness and for transportation damage. Inform the carrier immediately if transportation damage has occurred.
The final tests of the manufacturer may have caused traces of the shelves on the inner surfaces. This has no impact on the function and performance of the cabinet.
Please remove any transportation protection devices and adhesives in/on the cabinet and on the doors and remove the operating manuals and accessory equipment.
CAUTION
Sliding or tilting of the cabinet. Damage to the cabinet. Risk of injury by lifting heavy loads.
Do NOT lift or transport the cabinet using the door, t he handl e, or the lower housing.
Lift the cabinet from the pallet at the four lower corners with the aid of four people
If you need to return the cabinet, please use the original packing and observe the guidelines for safe lift­ing and transportation (chap. 3.2).
For disposal of the transport packing, see chap. 21.1.

3.2 Guidelines for safe lifting and transportation

After operation, please observe the guidelines for temporarily decommissioning the cabinet (chap. 21.2).
CAUTION
Sliding or tilting of the cabinet. Damage to the cabinet. Risk of injury by lifting heavy loads.
Transport the cabinet in its original packaging only. For moving or shipping, secure the cabinet with transport straps.
Do NOT lift or transport the cabinet using the door, t he handl e, or the lower housing.
Lift the cabinet at the four lower corners with the aid of 4 people
You can order transport packing for moving or shipping purposes from your local Datacolor service rep­resentative.
Datacolor CONDITIONER™ (MB2) 06/2017 page 19/126
CAUTION
midifying system must be activated (deactivated operation line “Humidity off”, chap.
) and close the tap of
Permissible ambient temperature range during transport:
If the steam humidifying system has NOT been emptied: +3 °C / 37.4 °F to +60 °C / 140 °F.
After your local Datacolor service representative has emptied the steam humidifying system: -10 °C /
14 °F to +60 °C / 140 °F.
At temperatures below +3 °C / 37.4 °F, water must be completely removed from the hum i difying system.
CAUTION
Transport below +3 °C / 37.4 °F with filled steam humidifying system. Freezing in the steam generator. Damage to the cabinet.
Contact your local Datacolor service representative before any transportation below +3
°C / 37.4 °F.

3.3 Storage

Intermediate storage of the cabinet is possible in a closed and dry room. Observe the guidelines for tem­porary decommissioning (chap. 21.2).
Permissible ambient temperature range during storage:
If the steam humidifying system has NOT been emptied: +3 °C / 37.4 °F to +60 °C / 140 °F.
After your local Datacolor service representative has emptied the steam humidifying system: -10 °C /
14 °F to +60 °C / 140 °F
With temperatures below +3 °C / 37.4 °F, water must be completely removed from the humidifying sys­tem.
CAUTION
Storage below +3 °C / 37.4 °F with filled steam humidifying system. Freezing in the steam generator. Damage to the cabinet.
Contact your local Datacolor service representative before any transportation below +3
°C / 37.4 °F.
Permissible ambient humidity: max. 70 % r.H., non-condensing
Condensation by excess humidity. Danger of corrosion on the housing after operating at humidity values > 70 % r.H.
for a long period.
Dry the cabinet completely before shut-down:
Set the humidity to 0 % r.H. To enable dehumidification, the humidifying and dehu-
7.3 and setting “Control on”, chap. 6.3).
Set the temperature set point to 60 °C / 140 °F for approx. 2 hours (Manual mode).
Only then, shut down the cabinet at the main power switch (1
the water supply.
When after storage in a cold location you transfer the cabinet to its warmer installation site, condensation may form. Before start-up, wait at least one hour until the cabinet has attained ambient temperature and is completely dry.
Datacolor CONDITIONER™ (MB2) 06/2017 page 20/126
CAUTION

3.4 Location of installation and ambient conditions

Set up the cabinet on a flat, even surface, and in a well-ventilated, dry location and align it using a spirit level. The site of installation must be capable of supporting the cabinet’s weight (see technical data, chap.
23.4). The cabinets are designed for setting up inside a building (indoor use).
Danger of overheating. Damage to the cabinet.
Do NOT set up cabinet in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
Permissible ambient temperature range during operation: +18 °C / 64.4 °F to +32 °C / 89.6 °F. At
elevated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem­perature of +22 °C +/- 3 °C / 71.6 °F ± 5.4 °F to which the specified technical data relate. De­viations from the indicated data are possible for other ambient conditions. Lower values of the temperature range indicated in the technical data are valid at an ambient temperature of m ax. 25 °C / 77 °F.
With each degree of ambient temperature >25 °C / 77 °F, the refrigeration power decreases by 1.5 K.
Permissible ambient humidity: 70 % r.H. max., non-condensing
When operating the cabinet at temperature set-points below ambient temperature, high ambient humidity may lead to condensation on the cabinet.
Installation height: max. 2000 m / 6562 ft. above sea level.
A water tap (1 bar to 10 bar) is necessary for the installation of the humidificat i on sy st em (chap. 4.3). If no suitable in-house water connection is available, you can manually supply water by filling the freshwater can (option, chap. 4.5).
To avoid any possible water damage, provide a floor drain at the location of the device. Select a suitable installation site to avoid any consequential damage by splashing water.
When placing several cabinets of the same size side by side, maintain a minimum distance of 250 mm /
9.84 in between each cabinet. Wall distances: rear 100 mm / 3.9 in, sides 160 mm / 6.29 in. Spacing
above the cabinet of at least 100 mm / 3.9 in must also be accounted for.
CAUTION
Danger by stacking. Damage to the cabinets.
Do NOT place the cabinets on top of each other.
To completely separate the cabinet from the power supply, you must disconnect the power plug. Install the cabinet in a way that the power plug is easily accessible and can be easily pulled in case of danger.
For the user there is no risk of temporary overvoltages in the sense of EN 61010-1:2010.
With an increased amount of dust in the ambient air, clean the condenser fan (by suction or blowing) several times a year.
Avoid any conductive dust in the ambiance according to the cabinet layout complying with pollution de­gree 2 (IEC 61010-1).
Datacolor CONDITIONER™ (MB2) 06/2017 page 21/126
Do not install or operate the cabinet in potentially explosive areas.
DANGER
Explosion hazard. Danger of death.
Do NOT operate the cabinet in potentially ex pl osiv e areas.
KEEP explosive dust or air-solvent mixtures AWAY from the v i cinity of the cabinet.
After turning off the cabinet, you must close the tap of the water supply. Install the cabinet in a way that the freshwater supply is easily accessible.
In case of a prolonged temporal decommissioning: Leave the cabinet door open or remove the access port plugs.

4. Installation and connections

4.1 Spacer for wall distance

Please fix both spacers with the supplied screws at the cabinet rear. This serves to ensure the prescribed minimum distance to the rear wall of 100 mm / 3.94 in.
Figure 7: Spacer for wall distance
Figure 8: Cabinet rear with mounted spacers
Datacolor CONDITIONER™ (MB2) 06/2017 page 22/126

4.2 Wastewater connection

Fasten the wastewater hose to the wastewater connection “OUT” (7) on the rear of the cabinet (olive 14 mm). Observe the following points:
You can use a part of the supplied water hose as a drainage hose. In case another hose is used, it
has to be permanently resistant against at least 95 °C / 203 °F.
Mount the wastewater hose with a maximum positive inclination of 1 m and a maximum total length of
3 m.
Protect the cabinet end of the drainage hose with one of the supplied hose clamps.
Reliably prevent sucking back of wastewater. The end of the wastewater hose must not be immersed
in liquids. This can be ensured e.g., by free discharge.
Waste water is collected in an internal can with a volume of approx. 0.5 liters. It i s pumped off only when required, thus there is no continuous waste water flow.
Protect the waste water supply with the supplied hose clamps.

4.3 Freshwater supply

Connect the waste water pipe before connecting the cabinet to a freshwater pipe or filling the freshwater can (option, chap. 4.5).
You can supply the cabinet with freshwater via a water pipe or by manually filling a freshwater can (op­tion, chap. 4.5).
Water intake temperature NOT below +5 °C / 41 °F and not exceeding 40 °C / 104 °F.
CAUTION
Calcification of the humidifying system. Damage to the cabinet.
Operate the cabinet with deionized (demineralized) water only.
Types of suitable water quality:
Deionized water from a water treatment installation already existing at the customer's site. Conductivi-
ty from 1 µS /cm up to a maximum of 20 µS/cm. (Water, which is in equilibrium with the CO and has a conductivity below 1 µS/cm (ultrapure water), may cause acid corrosion due to its low pH.)
in the air,
2
Water treated by the optional water treatment system BINDER Pure Aqua Service (disposable sys-
tem). A reusable measuring equipment to assess the water quality is included (chap. 4.4).
The manufacturer is NOT responsible for the water quality at the user’s site. Any problems and malfunctions that might arise following use of water of deviat i ng qual i ty is
excluded from liability by the manufacturer. The warranty becomes void in the event of use of water of deviating quality.
Datacolor CONDITIONER™ (MB2) 06/2017 page 23/126
For the water supply, fix the delivered adapter with hose olive on the thread at the rear of

4.3.1 Automatic fresh water supply via water pipe

An enclosure inside the cabinet contains the connection kit for freshwater and wastewater. Install the freshwater connection using either the enclosed water hose or another pressure-resistant one. To ac­complish this, remove the cover of the freshwater connection “IN” (6) on the rear of the cabinet. Protect both ends of the hose with two of the four supplied hose clamps.
Before turning on the cabinet, check the connection for leaks. Water supply is automatically effected via the freshwater connection “IN” (6).
As the appliance only lets in water when required, there is no continuous water flow.
Supply pressure 1 to 10 bar when connecting to a water pipe
Water type: deionized (demineralized) water
Water intake temperature NOT below +5 °C / 41 °F and not exceeding 40 °C / 104 °F.
The water intake should be provided with a shut-off slide or water-tap.
the cabinet.
Protect the water supply at one side with the supplied hose clamp.
The connection kit for connecting the cabinet to the water main enclosed with the cabinet includes a protective device against flooding caused by burst water hoses. Items supplied:
Hose burst protection device.
2 hose nozzles with screwing
4 hose clamps
6m water hose (PVC black, 12 mm (½’’) inner diame­ter), which should be cut into 2 accordingly, for the feed hose and drain.
Pressure-resistant up to 10bar, heat resistant up to 95 °C / 203 °F.
Datacolor CONDITIONER™ (MB2) 06/2017 page 24/126
Assembly:
Screw the hose burst protection device onto a water tap with a G¾ inch right turning thread connection. The connection is self-sealing. Establish the connection between the safety kit and the cabinet with a part of the supplied hose. Protect both ends of the hose by the supplied hose clamps.
We recommend connecting the hose as the last step in order to avoid twisting the hose while screwing on the safety kit.
Open the water tap slowly in order to avoid actuating the hose burst protection device.
Figure 9: Assembly of the connection kit
Protection principle of the hose burst protection device: Whenever a strong water flow of about 18 l / min. occurs, e.g. caused by a burst water hose, a valve au-
tomatically cuts off the water supply, which can be heard as a clicking noise. The water supply now re­mains shut until it is manually released.
Release of the reflux protection device:
In case the burst protection device has interrupted the water supply, first find the reason and remove it as necessary. Close the water tap. Release the valve by a half left-turn of the upper knurled part. You can hear the release of the valve as a clicking noise. Tighten the burst protection device against the water tap by a right turn. Open the water tap slowly afterwards.
Maintenance of the assembly of the hose burst protection device:
Calcification can impair valve function. We recommend an annual inspection by a skilled plumber. The plumber should demount the safety kit to check the valve by hand for functi on, calcification or blockage.
CAUTION
Danger of calcification. Impairment of valve function.
Have a plumber inspect the valve annually. Remove calcifications by citric acid or acetic acid solutions. Continue by testing the function and tightness of the mounted cabinet
Check: Quickly open the water tap while there is no cabinet connected – the valve should cut off the wa­ter flux without any delay.
Datacolor CONDITIONER™ (MB2) 06/2017 page 25/126

4.3.2 Manual fresh water supply via external freshwater can (option)

If no house water connection with suitable water is available, you can manually supply water by filling a freshwater can (volume: 20 liters / 0.71 cu.ft), which is part of the option “External freshwater and wastewater cans” (chap. 4.5). You can attach the freshwater can on the rear of the cabinet or place it next to the cabinet.
To guarantee humidification during 24 hours even at high humidity set-points with manual water supply, we recommend filling the freshwater can (option) daily at the end of the day.

4.4 Optional water treatment system BINDER Pure Aqua Service

The optional water treatment system (disposable unit) is available to treat tap water. The lifetime of the unit depends on water quality and the amount of treated water used. The measuring equipment to assess the water quality is reusable.
For detailed information on installing and operating the water treatment system BINDER Pure Aqua Service, please refer to the operating manual Art. No. 7001-0269, coming with BINDER Pure Aqua Service.

4.5 Optional external freshwater and wastewater cans

If no suitable in-house water connection is available, you can manually supply water by filling the optional external freshwater can. There is an additional external water can for the waste water. Volume: 20 liters /
0.71 cu.ft.
The cans are placed in holding devices. You can affix them directly at the rear of the cabinet or place them next to the cabinet.
Figure 10: Cabinet rear view with installed external water cans (option)
Datacolor CONDITIONER™ (MB2) 06/2017 page 26/126
The socket (3) is marked with a sticker:

4.5.1 Mounting the freshwater can

(1) Fixing (if required)
Hang the can with its holding device on its 4 carriers. You can install it either at the left or the right side.
Cable connection (8)
Freshwater hose connection (9)
Figure 11: Freshwater can (option)
(2) Cable connections
Connect the plug of the cable to the socket (3) at the rear of the cabinet.
Socket (3)
(7)
Datacolor CONDITIONER™ (MB2) 06/2017 page 27/126
(6)
Figure 12: Connections at the cabinet rear
(3) Hose connections
Plug the freshwater hose into the hose connection (9) above the freshwater can and secure it with a hose clamp. You can use a part of the standard supplied water hose.
Screw the hose nozzle (brass) to the free edge of the hose and screw it directly onto the freshwa­ter connection “IN” (6) at the rear of the cabinet.
When the freshwater can is empty, the message “Freshwater supply” will be displayed on the controller (chap. 11.1.3), the buzzer sounds, and the humidification module turns off. After acknowledging the alarm, the humidification module tries to fill up and start operating.
To guarantee humidification during 24 hours even at high humidity set-points with manual water supply, we recommend filling the freshwater can (option) at the end of each day.

4.5.2 Mounting the wastewater can

(1) Fixing (if required)
Hang the can with its holding device on its 4 carriers at the free space next to the freshwater can.
(2) Hose connections
Plug the wastewater hose to the hose connection (10) of the wastewater can and secure it with a hose clamp. You can use a part of the standard supplied water hose.
Hose connection of the wastewater can (10)
Plug the free hose edge to the wastewater connection “OUT” (7) at the rear of the cabinet and se­cure it with a hose clamp.
Figure 13: Wastewater can (option)
You can remove the wastewater can with its holding device for emptying (disconnect the hose first before emptying).
CAUTION
Overflow of the wastewater can. Damage to the surrounding.
Empty the wastewater can in a timely manner before it is full.
Bringing a source of humidity into the inner chamber may increase wastewater production. Regularly check the filling level of the wastewater can.
Datacolor CONDITIONER™ (MB2) 06/2017 page 28/126
CAUTION

4.5.3 Mounting with wastewater recycling

When the cabinet interior is clean, you can reuse the wastewater from the cabinet. Connect the wastewater connection “OUT” (7) of the cabinet with the freshwater hose connection (11) of the freshwa­ter can. The wastewater can is not used in this case.
Soiling of the vapor humidification system. Damage to the cabinet.
Reuse wastewater ONLY with a clean cabinet interior. In case of soiling / contamination of the interior, conduct the wastewater to the
wastewater connection or use the wastewater can.
(1) Fixing of the freshwater can (if required)
Hang the can with its holding device on its 4 carriers. You can install it either at the left or the right side.
(2) Cable connections of the freshwater can
Connect the plug of the cable to the socket (3) at the rear of the cabinet as described in chap. 2.4.
(3) Hose connections
Plug the wastewater hose into the hose connection (11) of the freshwater can and secure it with a hose clamp. You can use a part of the standard supplied water hose.
Plug the free hose edge to the wastewater connection “OUT” (7) at the rear of the cabinet and se­cure it with a hose clamp.
Hose connection of the freshwater can (11)
Figure 14: Freshwater can (option)
Bringing a source of humidity into the inner chamber may increase wastewater production. Regularly check the filling level of the freshwater can.
Datacolor CONDITIONER™ (MB2) 06/2017 page 29/126

4.6 Placement of the light cassettes (cabinet with lights)

The light cassettes are permanently connected. The cable length permits easily removal to replace the fluorescent tubes or to change the position of the light cassette inside the cabinet.
You can insert the light cassettes in different heights onto the beads at the lateral walls of the chamber. Insert and pull out the light cassettes only by the handles.
Do not place any charging material directly onto the light cassettes because those are heated by the fluo­rescent tubes which would lead to exposing the charging material to undefined temperatures. The tem­perature directly below or on the light cassettes is not equal to the temperature displayed at the tempera­ture controller.
Place the charging material on the supplied racks below the light cassettes.
CAUTION
The light cassettes will become hot with temperature set-points >40 °C. Danger of burning.
Do NOT touch the light cassettes during operation.
Let the light cassettes cool down before changing their position.

4.7 Electrical connection

The cabinets are supplied ready for connection. They come with a fixed power connection cable of at least 1800 mm / 70.87 in in length.
Power plug: NEMA 6-20P
Voltage +/-10% at the indicated power frequency: 200-240 V at 50Hz, 200-240 V at 60Hz
Current type: 2~
Cabinet fuse: 16 Amp
The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the cabinet’s protective conductor meets the latest technology. The protective conductors of the socket and plug must be compatible!
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the cabinet’s type plate (left cabinet side, bottom right -hand, see chap. 1.4).
When connecting, please observe the regulations specified by the local electricity supply company.
We recommend the use of a residual current circuit breaker.
Pollution degree (acc. to IEC 61010-1): 2
Installation category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage. Damage to the equipment.
Check the power supply voltage before connection and start-up. Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 23.4).
Datacolor CONDITIONER™ (MB2) 06/2017 page 30/126
Loading...
+ 97 hidden pages