Danfoss SonoMeter 40 Installation guide [ro]

Ghid de instalare
SonoMeter 40
Contor de energie ultrasonic pentru aplicații de încălzire și de răcire
www.danfoss.com
This doc. is managed by 500B0577
1
of
2
Ghid de instalare SonoMeter 40
EU DECLARATION OF CONFORMITY
Danfoss A/S
6430 Nordborg, Denmark | CVR nr.: 20 16 57 15 | Telephone: +45 7488 2222 | Fax: +45 7449 0949
declares under our sole responsibility that the
Product category: Energy Meters Type designation(s): SonoMeter 40
Covered by this declaration is in conformity with the following directive( s), standard(s) or other normative document(s), provided that the product is used in accorda
[RED] – Radio Equipment Directive 2014/53/EU Article 3.1a (LVD)
EN 61010-1:2010+A1:2019 EN 62368-1:2014 EN 62311: 2008
Article 3.1b (EMC)
EN 301 489-3 V2.1.1 EN 301 489-1 V1.9.2
Article 3.2 (Radio)
EN 300 220-1 V3.1.1 EN 300 220-2 V3.1.1
[EMC] – Electromagnetic Compatibility Directive 2014/30/EU
EN 55032:2015+A11:2020
[LVD] – Low Voltage Directive 2014/35/EU
EN 61010-1:2010+A1:2019 EN 62368-1:2014 EN 62311: 2008
[MID] – Measuring Instruments Directive 2014/32/EU
EN 1434-1:2015+A1:2018 EN 1434-2:2015+A1:2018 EN 1434-3:2015 EN 1434-4:2015+A1:2018EN 1434-5:2015+A1:2019 EN 1434-6:2015+A1:2019 WELMEC 7.2:2015 - Software Guide
Notiied Body: Lithuanian Energy Institute, 1621, performed type approval and issued certiicate LT-1621-MI004-047.
[RoHS] – Restriction of Hazardous Substances Directive 2011/65/EU+A:2015/863
EN IEC 63000:2018
Danfoss Energy Metering
nce with our instructions.
2
1
1, 3
4
1 For variants with Radio Module the declaration for EMC & LVD shall be ignor ed 2 For variants without Radio Module the declaration for RED shall be ignored. 3 For variants without Radio Module and without Mains power supply (24V AC /DC) the declaration for RED and LVD shall be ignored. 4 Not within MID – National Type Approval Certiicate for cooling energy DE-21- M-PTB-0069 is issued by “Physikalisch-T
Date:
2021.09.22
Place of issue:
8811 Scheifling Austria
Danfoss only vouches for the correctness of the English version of this declaration. In the event of the declaration being translated into any other language, the translator concerned shall be liable for the correctness of the translation
ID No: 014R2944
Signature:
Name: Norbert Spreitzer
Title: Product Portfolio Manager
Revision No: 01
2021.09.22
Place of issue:
1210 Ljubljana, Slovenia
Approved by
Signature:
Name: Gasper Benedik
Title: Energy Meters Director
echnische Bundesanstalt”.
Page
2 | © Danfoss | Contoare de energie | 2021.11 AN384648824392ro-000102
This doc. is managed by 500B0577
2
of
2
Ghid de instalare SonoMeter 40
UK DECLARATION OF CONFORMITY
Danfoss A/S
6430 Nordborg, Denmark | CVR nr.: 20 16 57 15 | Telephone: +45 7488 2222 | Fax: +45 7449 0949
declares under our sole responsibility that the
Product category: Energy Meters Type designation(s): SonoMeter 40
Covered by this declaration is in conformity with the following directive( s), regulation(s), standard(s) or other normative document(s), provided that the product
Radio Equipment Regulations 2017 Article 3.1a (LVD)
BS EN 61010-1:2010+A1:2019 BS EN 62368-1:2014 BS EN 62311:2008
Article 3.1b (EMC)
EN 301 489-3 V2.1.1 EN 301 489-1 V1.9.2
Article 3.2 (Radio)
EN 300 220-1 V3.1.1 EN 300 220-2 V3.1.1
Electromagnetic Compatibility Regulations 2016
BS EN 55032:2015+A11:2020
Electrical Equipment (Safety) Regulations 2016
BS EN 61010-1:2010+A1:2019 BS EN 62368-1:2014 BS EN 62311:2008
Danfoss Energy Metering
is used in accordance with our instructions.
2
1
1, 3
The Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment Regulations 2012 (as amended)
BS EN IEC 63000:2018
Heat Metering and Billing Regulations 2014
Heat meter accuracy cl. 7.9:
Notiied Body: Lithuanian Energy Institute, 1621, performed type approval and issued certiicate LT-1621-MI004-047.
1 For variants with Radio Module the declaration for EMC & LVD shall be ignor ed 2 For variants without Radio Module the declaration for RED shall be ignored. 3 For variants without Radio Module and without Mains power supply (24V AC /DC) the declaration for RED and LVD shal
Date:
2021.09.22
Place of issue:
8811 Scheifling Austria
Danfoss only vouches for the correctness of the English version of this declaration. In the event of the declaration being translated into any other language, the translator concerned shall be liable for the correctness of the translation
ID No: 014R2944
Issued by Date:
Signature:
Name: Norbert Spreitzer
Title: Product Portfolio Manager
Revision No: 01
2021.09.22
Place of issue:
1210 Ljubljana, Slovenia
Approved by
Signature:
Name: Gasper Benedik
Title: Energy Meters Director
l be ignored.
Page
AN384648824392ro-000102 © Danfoss | Contoare de energie | 2021.11 | 3
Ghid de instalare SonoMeter 40
1. Montarea
1.1. Pregătirea
Echipamentul poate fi instalat numai de personal calificat, respectând cerințele enumerate în acest document. La www.heating.danfoss.com puteți găsi instrucțiuni mai detaliate.
Notă! Acest produs este aprobat pentru o temperatură ambiantă cuprinsă între 5 și 55 °C, dar pentru a se asigura condiții optime pentru baterie, se recomandă instalarea calculatorului la max. 45 °C. A se evita tensionarea instalației prin conducte și fitinguri. Spălați sistemul.
1.2. Identificarea instalației: Instalație cu conductă de retur/tur și direcție de curgere
Flow direction
Direcția de curgere
Instalație cu conductă de retur
Flow direction
Direcția de curgere
Instalație cu conductă de tur
AN384648824392ro-000102 © Danfoss | Contoare de energie | 2021.11 | 4
Ghid de instalare SonoMeter 40
1.3. Montarea senzorului de debit
Poziția conductei: Nu există limite, dar evitați
pozițiile în care se poate acumula aer.
Rotație pe axa conductei: Senzorul de debit trebuie să fie înclinat între 45 și 315° pentru a se evita acumularea aerului în acesta.
Condiții de intrare/ieșire (numai pentru DN 65 – 100) Pentru maximizarea performanței, este necesară existența unor condiții de curgere în linie dreaptă pentru intrare și pentru ieșire, în amonte și în aval de senzorul de debit: 5 × DN la intrarea și 3 × DN la ieșirea senzorului de debit.
5 x DN 3 x DN
flow
debit
1.4. Montarea și sigilarea calculatorului
Calculatorul contorului de energie termică poate fi instalat în unități încălzite, iar temperatura ambiantă de lucru nu trebuie să depășească 55 °C. Nu poate fi expus la lumina directă a soarelui. Calculatorul este montat pe un suport auxiliar (poate fi orientat în direcția necesară, la un unghi de 90°:
Modurile posibile de montare a calculatorului (suport auxiliar): – Montare directă pe carcasa senzorului de debit, rotind la fiecare 90°
(numai când temperatura pe tur nu depășește 90 °C):
5 | © Danfoss | Contoare de energie | 2021.11 AN384648824392ro-000102