Danfoss PFM 5000 User guide

User Guide: PFM 5000 Measuring Instrument - Standard Version
VUE2E102
© Danfoss HEC
VUE2E102
Contents
PFM 5000 main features.............................................................................................................................................................................5
Introduction.................................................................................................................................................................................................5
PFM 5000 Standard components................................................................................................................................................................7
Switching on and off...................................................................................................................................................................................8
Introduction.................................................................................................................................................................................................8
Getting started ............................................................................................................................................................................................9
Installation of PFM 5000 application.......................................................................................................................................................10
Delivery with Smartphone.........................................................................................................................................................................12
Bluetooth settings......................................................................................................................................................................................13
Run PFM 5000 application.......................................................................................................................................................................15
PFM 5000 application settings..................................................................................................................................................................16
Start of measuring.....................................................................................................................................................................................17
Sensor zero setting.....................................................................................................................................................................................20
Data recording...........................................................................................................................................................................................21
Data recording...........................................................................................................................................................................................22
(continued).................................................................................................................................................................................................22
Data recording...........................................................................................................................................................................................23
(continued).................................................................................................................................................................................................23
Calculation of valve presetting for a required flow in one branch............................................................................................................23
Calculation of valve presetting for a required flow in one branch (continued)........................................................................................24
Calculation of pressure loss on the valve for required flow......................................................................................................................25
Working with projects...............................................................................................................................................................................26
Starting with projects.................................................................................................................................................................................27
Creating new project ................................................................................................................................................................................28
Creating new project ................................................................................................................................................................................29
Creating new project ................................................................................................................................................................................30
Preparing for project balancing - measuring.............................................................................................................................................31
Preparing for project balancing - measuring.............................................................................................................................................32
Balancing of basic project ........................................................................................................................................................................33
HEC © Danfoss
VUE2E102
Balancing calculation of a project with central pressure input - procedure..............................................................................................36
Replacement of sintered filters..................................................................................................................................................................37
Technical specifications............................................................................................................................................................................38
© Danfoss HEC
VUE2E102

PFM 5000 main features

Separated main pressure unit and computing unit
based on Smart Phone with Android OS connected by Bluetooth
Multisource version includes Converter Unit which
converts Bluetooth to wireless communication in ISM band
Possible connection up to 10 pressure units
Accurate pressure measuring with true differential
pressure sensor and 24 bit ADC
Hydraulic by-pass for exact small differential pressure
measuring
External PT-100 thermometer
Working with projects
Programmable autonomous recording mode
Main unit powered by rechargeable Li-Ion battery

Introduction

PFM 5000 Standard has been designed to create hydraulic balance in heating and cooling systems.It enables measur­ing of pressure flow and temperatures in a system. Using differences in pressure, PFM 5000 is capable of turning these two pressures on both sides of a valve into flow.Flow through individual branches of the system can be meas­ured, creating perfect balance in the whole system. PFM 5000 Standard has a number of key features, that makes it easy to use. PFM 5000 Standard consists of two separate units: a main pressure unit for measuring pressure, flow and temperat­ure, and a computing unit for displaying results and data analysis.
The main pressure unit is extremely robust, with a sturdy frame and sound cover.Inside the pressure unit is a differential pressure meter with an integrated true differential pressure sensor for accurate digital data processing.
The flow meter automatically corrects the flow for different types of mediums being measured, such as antifreeze ad­ditives in cooling systems.
PFM 5000 Standard has the advantage of high class digital technology, which enables it to compensate for inaccuracies, normally associated with pressure measurements, such as temperature dependency and non-linearity.In order to in­crease the accuracy of low pressure measuring and to enable pressure connection deaeration, the main pressure unit is de­signed with incoming by-pass couplers for hydraulic zero set­ting. This forms the basis for the accuracy of measurements. The PFM 5000 Standard can be supplemented with an ex­ternal thermometer connected via a coaxial connector.The temperature of the working medium can easily be measured by inserting the thermometer into the measuring nipples of the valve, were you normally insert 3 mm measuring needles.
Specifications for other valve manufacturers than Danfoss are pre-programmed in the memory of the PFM 5000 Standard.
The autonomous measuring mode of PFM 5000 Standard allows independent data recording based on a pre­programmed period.The acquired data are initially saved in
HEC © Danfoss
VUE2E102
5
PFM 5000 Standard can calculate complicated multi-branch heating systems by simulating the hydraulic system with balancing calculations based on two measuring in each branch.
During balancing calculations, the PFM 5000 Standard utilises a sophisticated method for calculating hydraulic resistances within the system. As a result, the PFM 5000 Standard comes up with a proposal with the lowest energy loss.This function dramatically decreases the time required for balancing.
the main pressure unit and can subsequently be analysed in computing unit .
© Danfoss HEC
VUE2E102
6

PFM 5000 Standard components

HEC © Danfoss
VUE2E102
7

Switching on and off

1. Press ON/OFF button to switch on PFM 5000 Standard. The sensor switches off automatically 60 minutes after the end of communication with PDA.

Introduction

.
© Danfoss HEC
VUE2E102
8

Getting started

1. Smartphone has touch-sensitive display and usually three basic buttons:
Home – it serves for return to Smartphone main window.
Back – it returns the current window of a back. While
application is running, you can return up to the phone main window and application stops.
Menu – phone or application menu.
2. Insert SIM card to your Smartphone. Unlock SIM card by standard procedure for mobile phones.
Make sure that you have arranged internet services to your SIM card due to downloading about 50 MB of data during Smartphone activation. Next amount of data will be charged during PFM5000 application download.
3. Create Google Account.
4. Fill requested fields.
HEC © Danfoss
VUE2E102
9
Getting started
(continued)
5. You have created Google account after some standard steps.

Installation of PFM 5000 application

1. Enter Market 2. Search for PFM 5000 application.
© Danfoss HEC
VUE2E102
10
Installation of PFM 5000 application
(continued)
3. Choose PFM 5000 ST. 4. Click Download.
5. Continue Accept &
download
6. Continue Open.
HEC © Danfoss
VUE2E102
11
Installation of PFM 5000 application
(continued)
7. PFM 5000ST main menu.
8. Return to Smartphone Main Window by Back button

Delivery with Smartphone

PFM 5000 application is installed on the Smartphone. You need only insert SIM card to phone and setup call and internet services. It is necessary to create Google account if you want download upgrades of PFM5000 application.
© Danfoss HEC