Danfoss PFM 3000A User guide [pl]

Instrument for the Balancing of Heating Systems
PFM 3000
User Guide
VUE1V102
PFM 3000
PFM 3000
Content
Content
Introduction
Invaluable Advantages of PFM 3000
Important User Notice
Putting the Instrument into Operation
Battery Charging
Deaeration of Connecting Tubes
PFM 3000 Zero Measuring Correction
Functions of Keys on the Instrument's
Keyboard
Measuring with PFM 3000
Measuring Flow in the Fittings Not Saved in the
PFM 3000 Memory ..................................
............................................
........................................
.................
...............................
...............
....................................
....................
................
...........................................
..........................
...............................
Calculation of the Fitting Presetting for Requested
Flow ...............................................
Calculation of the Pressure Drop for Requested
3
flow ................................................
5
Data Record
6
6
6
6
7
Processing Projects
7
7
Measurements Necessary for the Calculation of
8Language Changing ..................................
Project Balancing ...................................
8Real Time Clock Setting ..............................
9Automatic Shut OFF ..................................
9Display Backlight Time ...............................
10Pressure Units ......................................
Balance Calculation of the Project with a Pressure
10Flow Units ..........................................
Input in the Middle ...................................
10Measuring Filter .....................................
Communication with a Computer
10Valve Selection .....................................
Data Processing and Displaying Program - User
11Valve Presetting Change .............................
software PFM 3000 ..................................
11Measured Medium ..................................
11
11Measuring Without Project ...........................
11System Overpressure Measuring .....................
12Differential Pressure and Flow Measuring .............
Maintenance Instructions
Parts of the Delivery
12
Technical Specifications
13Temperature Measuring .............................
PFM 3000 Menu
13
.......................................
...............................
..................
..........................
...............................
...........................
....................................
13
13
13
13Record Preparing ...................................
14Start of Recording ...................................
15Full Data Memory and Low Battery ....................
15End of Recording ...................................
15
16Creating a New Project ..............................
17Project without a Common Valve .....................
17
18Balance Calculation of the Project ....................
18Results of Project Balancing ..........................
18Project Balancing Check .............................
20
21
21
21Minimum PC Requirements ..........................
21Software Installation .................................
21Operating the Program ..............................
21Measured Data Processing ..........................
23
23
25
27
3
PFM 3000
4
PFM 3000
Introduction
PFM 3000 has been designed for the servicing and hydraulic balancing of heating and cooling systems. It is equipped with a number of added built-in functions that make your work enjoyable. The instrument is small. Its hard metal case lets you use it even under severe outdoor conditions. Even if you spill water on the case, PFM 3000 stays intact. A delivery includes standard RS232 interface and user software for the processing of gathered data. A basic modulus of the instrument is a differential manometer with digital pressure processing. It is equipped with a built-in pressure sensor. The manometer combines the advantages of digital technology and compensates temperature dependence and the non-linear character of measuring. It makes possible the measuring of overpressure or underpressure in a system and the measuring of differential pressure. Using pressure difference at a balancing or measuring component you can measure flow at particular branches of a system and balance the heating system. Excellent features of the manometer enable further accurate measuring. An external thermometer modulus is a part of the basic measuring equipment of the instrument. It enables to measure the temperature of a medium. The thermometer connects to the RS232 interface connector. One finds it easy to work with, as its dimensions are compatible with the dimensions of the measuring needles of balancing valves. The measuring spike of the thermometer should be plugged into an output for the measuring of differential flow. Along with other data temperature is being recorded in the
instrument’s memory. A built-in flow meter modulus is an indispensable instrument for the balancing of hydraulic systems. It calculates flow at a particular branch of a heating system out of pressure difference at a measuring element. The memory of the instrument keeps the characteristics of balancing and measuring elements of established producers. You can choose a particular type in the instrument’s menu. There is an automatic flow correction function for flow measuring in systems with non-freezing additives or cooling systems. A valve presetting calculation modulus is an revelsal part of the flow meter modulus. The output of the valve presetting calculation function is valve presetting for required flow. The function calculates valve presetting using valve characteristics stored in the memory. Therefore you always receive the right value of presetting. An integrated recording modulus including a real time circuit is another positive feature of the instrument. The time circuit is all time powered from apparatus lithium battery. The modulus helps keep records from project realisation or diagnose time actions in a heating system. The memory stores pressure, flow, temperature, chosen valve, valve presetting and measurement identification data, which helps you process and evaluate results on your PC. While recording measured data with a longer recording period, the instrument switches automatically to the sleep mode. This is enabled by a long-term recording function with energy coming from an internal source. You can record data at different points of the heating system and have the current state of the system processed in your office. You can also choose a periodical recording. The
5
PFM 3000
analysis of time actions will help you choose an optimum system solution. To download recorded data into your PC you can use software included in the delivery. The software makes it possible to display data in a chart or graph. They can also be printed. Data can be exported in standard PC formats and processed by a text editor, chart processor or database programs. A heating system balancing modulus seems to be the most important component for the user. The modulus calculates the balancing of the system with many branches. It makes it possible to regulate even a complicated heating system including a common valve. A sophisticated calculation method optimises hydraulic ratios in the system and produces a project aiming at minimum energy loss in the system. The method reduces substantially the time needed for balancing. The function calculates valve presetting directly using the characteristics stored in the memory. Therefore you always receive the right presetting value.
Invaluable Advantages of PFM 3000
y Mechanical design y Integrated pressure sensor y Robust metal case y Water resistant y Big 4-line display y Illuminated display y Well-arranged 16-key foil keyboard y Chargeable lithium battery
Important User Notice
If you have previously used the instrument to measure liquid medium, do not expose it to temperature lower than the point of congelation of measured liquid. The pressure sensor may be damaged. Inputs for measuring tubing are equipped with sintered filters
- filtering capacity 50 µm. When measuring in systems with a
lot of dirt, filters get fouled. If the instrument's display indicates high pressure or one of pressure inputs does not work after measuring tubing have been disconnected, replace the filters.
Putting the Instrument into Operation
PFM 3000 is delivered with precharged lithiun battery and is ready to run immediately. Please check apparatus clock after first switch on.
Battery Charging
During measuring you can see battery state after hot key
pushing. The battery capacity baragraph appears on
display
6
PFM 3000
Message Low Battery! flashes on the display every 15 sec bellow battery capacity 5%. You can work in this state about 6 hours, after it PFM 3000 automatically switches off. Connect battery charger into conector under the cover at PFM 3000 bottom. Charging time is about 7 hours.
Deaeration of Connecting Tubes
To minimize measurement errors, deaerate the connecting tubing of the instrument. Take a quick coupler R20 off the tubing by pressing a safety ring (black collar) towards the quick coupler. Do not disconnect the quick coupler merely by pulling or force. The surface of the tubing or screw joint might be damaged. Connect the tubing to the measuring valve by a quick coupler R21 or to a discharge cock by means of a measuring adaptor. Flush out the tubing with water until air bubbles disappear. Put back the quick coupler R20.
hoses. While it has not requested effect, special maintenance is needed. Zero is not stored in PFM 3000 configuration and not corrected value is measured after new switch on.
Functions of Keys on the Instrument's Keyboard
Keys with standard numbers or letters are used to enter data into computer. Following keys have special functions:
Switches the instrument on and off. To switch on the instrument, hold the key for one second.
Resets a reading about pressure on the display. The key also corrects possible zero instability and the dependence of the differential pressure on static pressure.
If in the measuring function, it opens the menu. It is used for the upward movement within the menu. If you press the key while entering data, the new value will not be saved. The previous data will be kept.
Serves to accept a value or a selected command in menu.
PFM 3000 Zero Measuring Correction
When measuring zero pressure will take an error after a time
of PFM 3000 using, you can correct this effect by ZERO function. In the instant that zero error is grater some kPa, we recommend clean or change sintered filters in apparatus
Hot key Valve. While in the measuring mode, it opens a selection of valves.
Hot key Preset. While in the measuring mode, it opens a valve presetting.
Hot key Battery. While in the measuring mode, it opens battery capacity display.
7
PFM 3000
Hot key Project Change. While in the Project making mode, it changes displayed projects.
Deletes a sign from the actual position and deletes a whole line if held.
Use the keys to move within the menu or the value entry. While in the measuring mode, it
switches displaying modes. In the Project
making mode, it changes displayed branches.
PFM 3000 Configuration
Before measuring, configure the instrument according to a particular system. The configuration will be saved in the memory until a next change regardless of shutting down the instrument.
Press MENU key to open menu. The first line shows an entry name. The other lines show the list of sub entries. A selected entry is marked at the beginning of the line. If the entry has sub entries, it is marked by at the very end of the line.
Move in menu and confirm requested item.
Return one level up, by holding 1 sec you will return into the measuring from arbitrary menu level.
Language Changing
Main Menu Select item Configuration and confirm it.
Menu Configuration. Select Language item and confirm it.
Menu Language. Select requested Language and confirm it.
Real Time Clock Setting
You can check Date and Time in the Time mode
Mode hot key
Mode offer
Select Date and Time and press OK.
If PFM 3000 clock setting is incorrect set the appropriate parameter
Main menu. Select Configuration item and confirm.
Open menu
8
PFM 3000
Configuration menu. Select Date and Time item and
confirm.
Menu Date and Time. Select Minute item and confirm.
Enter minutes.
Move cursor to requested position.
...
There is no need to enter every decade and/or decimal point of the number in question, only the desired changes
Rewrite digit at cursor position.
If needed, erase digit at cursor position by DEL key. Holding DEL by 1s you will erase whole number.
Confirm correct number.
Automatic Shut OFF
PFM 3000 shuts down automatically 10 min after pressing the keyboard for the last time. In case of measuring for a longer period of time without using the keyboard switch this
function off in the Configuration/Automatic OFF menu. While in the recording mode, this function is switched off automatically. It is reactivated after the end of recording. The automatic shut down command is not saved in the
instrument's configuration. Next time when you switch the instrument on the function is active again.
Main menu Select Display item and confirm.
Menu Configuration. Select Automatic OFF item and confirm.
Menu Automatic OFF. Select requested function and confirm.
Display Backlight Time
This is a function to be appreciated when you measure in dark. When you set the backlight time to 0, the function is not active.
Main menu. Select Display item and confirm.
Display menu. Select Backlight Time item and confirm.
Enter Backlight Time value and confirm.
9
PFM 3000
Pressure Units
Flow Units
Main menu. Select Display item and confirm.
Display menu. Select Units item and confirm.
Units menu. Select Pressure Units item and confirm.
Pressure units choice. Select requested unit and confirm.
Back one menu level.
Units menu. Select Flow Units item and confirm.
Flow units choice. Select requested unit and confirm.
a change of damping (time constant) of the measured value. Display reading achieve 99,9% input pressure value after
0,75 s for the Fast filter, 5 s for the Medium Filter and 20 s
for the Slow Filter. There is additional filter for the flow measuring, which adds 16 s more at arbitrary filtering choice.
Main menu. Select Display item and confirm.
Menu Display. Select Measuring Filter item and confirm.
Choose requested filter and confirm.
Valve Selection
Valve choice hot key
Valve manufacturer selection. Choose requested manufacturer and confirm.
Valve type selection. Choose requested valve and confirm.
Measuring Filter
The filtration of measured values makes it possible to achieve a stabilised display of the value despite the fluctuation in the measured value. Filtration is manifested as
Valve presetting input.
10
PFM 3000
Enter valve presetting.
... ...
..
Decimal point you will enter by 1 s holding. There is only one decimal point allowed in PFM 3000.
Confirm. PFM 3000 returns back into measuring.
Valve Presetting Change
Presetting hot key.
Enter new presetting.
Confirm. PFM 3000 returns back into measuring.
Measured Medium
PFM 3000 uses a built-in measuring correction function for heat-carrying mediums. To use measuring correction, we need to know the temperature of the medium.
Enter Main Menu.
Select Medium Select. item.
Medium Select. menu. Choose medium and confirm.
Enter medium concentration in % and confirm.
When the temperature probe is connected, measured temperature appears on the display. You have to connect temperature probe to the measured medium in this case. In other case do not connect the thermometer to PFM 3000 and enter temperature manually.
Confirm.
Measuring with PFM 3000
Measuring Without Project
System Overpressure Measuring
Measuring Mode hot key
Measuring Mode choice.
11
PFM 3000
Select Pressure Meas. and confirm.
Connect static pressure to the higher pressure input (red). Do not connect the lower pressure input (blue). The instrument measures overpressure in the system compared with atmospheric pressure.
Differential Pressure and Flow Measuring
PFM 3000 can measure pressure difference at any point in the system. Flow can be measured only in specified regulation fittings saved in the memory. Use connection tubing to connect the measuring outputs of the balancing valve to pressure inputs. Higher pressure to the positive (red) input, lower pressure to the negative (blue) input. To suppress the effect of static pressure in the system, set PFM 3000 zero.
Completely open the regulation valve and stop flow of the medium.
Press ZERO key.
If the zero error exceeds a few kPa, clean the sintered filters at the pressure input. Call a professional if the error persists.
Neither zero-setting is saved in the instrument's
configuration. Next time you switch the instrument on the zero is not being corrected.
After you turn on the flow, PFM 3000 measures pressure difference at the regulation valve.
Measuring mode hot key.
Measuring mode choice
Select Flow Measuring and confirm.
PFM 3000 computes flow at the selected valve.
Measuring Flow in the Fittings Not Saved in the PFM 3000 Memory
To measure flow, enter the respective fitting and its presetting. You may want to measure flow in the fitting that is not saved in the PFM 3000 memory.
Main menu. Select Direct Kv Input and confirm.
Enter Kv value and confirm.
12
PFM 3000
PFM 3000 will calculate flow using the given Kv value. Press
the VALVE or the PRESET key and the instrument returns to standard flow measuring.
Temperature Measuring
Switch Off PFM 3000 and connect the temperature probe. Switch ON PFM 3000.
Measuring Mode choice.
Select Temp. Measuring and confirm.
Heating Calculations
Calculation of the Fitting Presetting for Requested Flow
This function makes it possible to calculate the fitting presetting for requested flow. Select the type of fitting and connect PFM 3000 to the measured fitting.
Main menu. Select Heating Calcul. item and confirm.
Select Valve Preset item and confirm.
You will be asked to enter requested flow. Measure actual flow and confirm with key OK. Then close the fitting and confirm the pressure value under the closed state. PFM 3000 will calculate presetting for requested flow.
Calculation of the Pressure Drop for Requested flow
This function will show you, if it is possible to adjust the branch, when you know available branch pressure. Choose actual valve type and presetting.
Select Pressure Drop item from Heating Calcul. menu.
Enter requested flow. PFM 3000 will compute pressure drop across actual valve with actual presetting. When the computed result is greater then the available pressure, you simply know, that this requested flow is not genuine.
Leave off the pressure drop computing by ESC key.
Data Record
Record Preparing
You have to check respectively set up record parameters
from the Record menu.
13
PFM 3000
Period/Hours = 0 Period/Minutes = 0
Period/Hours = X Period/Minutes = Y
Check free record capacity from Record/Capacity menu.
The display indicates filled memory space in the form of a baragraph. One baragraph unit equals appr. 100 records. If there is not enough vacant memory space, delete the memory in Record/Erase menu.
Record period may be from 1 s up to 24 hours.
PFM 3000 records data with period from Period/Seconds input.
Seconds of period are ignored
and PFM 3000 records data with
period X:Y. While waiting for the record the instrument switches into the energy-saving sleep mode. The display is not active at that time. After pre-programmed period elapses, PFM 3000 wakes up and measures time necessary for parameters fixation. After that record shoots and apparatus go to the sleep mode again.
Select menu item Record/Place
or Record/Description.
Enter the identification chains of the record made of letters from A to Z, numbers from 0 to 9, space, dash and point (max. 16 characters). Thus each record includes a 32-character identification.
The characters are written in blink cursor position. You can move cursor by arrow keys. Alphanumeric key pressing overwrite the character at cursor position. Every alphanumeric key has some values. The key in left, for instance, writes characters A, B, C, 1 by sequel at the same position. After you press another key, the cursor moves right one position. When you need the same character twice, move the cursor by arrow after entering the first character and enter the character again.
This key deletes character in cursor position and whole text right from the cursor moves one character left. Holding the key by 1 sec erases complete line.
You will confirm the input by the OK key. If you want leave input without writing, press ESC key.
Start of Recording
Start the record from
Record/Start menu.
At the beginning of each recording the memory saves a location and description (entered in the previous paragraph), selected valve and presetting, time, pressure,
14
PFM 3000
flow and temperature. During the data record flashes before the time reading.
*
Full Data Memory and Low Battery
When the data memory fills up during record, following display will appear.
The record will stop and PFM 3000 switches off after 10 minutes. Low battery state during data record works the same as in normal measuring. With the battery capacity less then 5 %
message Low Battery! blinks on the display every 15 s. When zero battery capacity PFM 3000 switches off.
End of Recording
Record continues.
Processing Projects
PFM 3000 has a built-in modulus for the calculation of project balancing with one horizontal distribution and a maximum of 32 vertical branches. The calculation method presumes there is constant pressure at the input of the project and the project includes no hydraulic elements with a feedback (regulators of differential pressure at branches, thermostatic valves). You can store two this projects into the instrument memory. Also it is able calculation of project balancing powered in the middle by binding two simple projects.
Project schematic
End of record.
During record with Sleep apparatus wakes up by any key pressing and apparatus ask you for the end of record as the previous case.
Note: The Zero function (the correction of the zero shift of measuring) is not active in the sleep mode.
15
PFM 3000
Creating a New Project
For new heating system may be projected value. For old systems this presetting value, which will allow measurable pressure drop at all branch balancing valves.
Menu Project/New project
Enter Project Name, max. 12 characters.
Common valve choice. For the project without Common Valve select No Common Valve.
Select Common Valve producer.
Select Common Valve type.
Enter Common Valve initial presetting from Project/Initial Preset menu.
Menu Branch/New branch.
Enter name of the branch, max. 12 characters.
Choose balancing valve in branch.
Enter initial presetting of the branch balancing valve. Use the same rules as in the common valve.
Enter requested flow.
Repeat New branch steps for all branches in the project.
The branches has to be added in the order of distance from heating medium input, so they are in the real project topology. When you create new project in PFM 3000 at one blow, each additional branch adds behind previous branch. When you need add a branch somewhere in the middle of the system, you need to select branch before the new
branch from menu Branch/Selection.
16
PFM 3000
Project without a Common Valve
System without a common valve schematic
Occasionally we come across systems without a common valve on input. PFM 3000 makes it possible to balance this
kind of system by selecting Project/Common valve/
Without common valve. If you create the project on
personal computer then cancel marking Common valve.
Connect PFM 3000 to the input (into the measured system) of the heat-carrying medium.
Menu Project/Input Pressure
Go to the first project branch and connect PFM 3000 to the balancing valve. Select a respective branch.
Choose Branch/Measure item.
Menu Measure/Initial Preset
Measurements Necessary for the Calculation of Project Balancing
Set all fittings in the project including common valve to initial presetting.
Select a respective project from menu
Project/Selection.
Menu Measure/Valve Closed.
Close the balancing valve.
Set back initial presetting.
Repeat branch measuring at all project branches.
You can also create project in PC and download it into the
17
PFM 3000
PFM 3000. Make use the User software PFM 3000.
Balance Calculation of the Project
Select Project/Balance calculation/Start to start calculation. PFM 3000 checks whether all values needed for calculation have been measured. In case some values are missing, the instrument will indicate branches that still have to be measured. The result of calculation is the presetting all valves including the common valve for requested flow. During the balance calculation the display indicates progress the procedure in the baragraph.
Results of Project Balancing
When calculation is over, display the results under
Project/Balance calculation/Results menu.
Pressure behind common valve computed from project compensatory scheme.
Particular branch presetting including common valve.
It is not possible to balance the branch. It must be higher requested flow value for particular valve or smaller dimension of the valve.
It is not possible to balance the branch. It must be smaller requested flow value for particular valve or bigger dimension of the valve.
Project Balancing Check
y Set the calculated presetting at all project branches
including the common valve. Actual flow should correspond to requested flow.
y If necessary, correct small flow deflections from
requested values by changing fitting presetting.
y Store actual flow into the project in the
Branch/Measure Branch/Actual flow menu. Do not forget to select a respective branch under
Branch/Select branch.
For balancing check is very expedient Projects Mode, so that you can easily change branches with branch parameters for displaying.
PFM 3000 mode selection hot key.
Mode selection proposal.
18
Select Project and confirm.
There are all significant project data on this display. Project name at the line 1. Selected branch at the line 2. Requested flow at the line 3. Real flow at the line 4.
Hot key switches two PFM 3000 projects, you can list
between branches by arrow keys. When you change the branch, PFM 3000 automatically chooses appropriate valve and presetting for real flow calculation. When the project is balanced in PFM 3000, apparatus chooses balanced presetting, in other case initial presetting.
PFM 3000
19
PFM 3000
Balance Calculation of the Project with a Pressure Input in the Middle
We often come across systems with one pressure input and distribution into both sides of the input. PFM 3000 makes it possible to measure this kind of system by dividing it into two projects with simple distribution. You measure and balance the
projects separately. Then using the command Bind projects you balance the whole system with the input in the middle.
Scheme of the System with a Pressure Input in the Middle
Balance calculation of the project with a pressure input in the middle - Procedure
y Measure project Right Side.y Create two projects - marked as the Left Side and
Right Side in the picture.
y Start the balance calculation of the Left Side project.y Close the Right Side by the right stop valve.
y Pre-adjust all valves in the left side to initial presetting.
y Close the Left Side. Open the Right Side.
y Start the balance calculation of the Right Side project.
y Start the balance calculation of the Left Side project.y Measure project Left Side.
y Select Project/Balance calculation/Bind projects to
calculate the balance of the whole system.
20
PFM 3000
Communication with a Computer
PFM 3000 can be easily connected to the computer by a serial RS232 line.
Data Processing and Displaying Program ­User software PFM 3000
User software provide easily work with data recorded in instrument and easily prepare projects in PC. Recorded data can be processed by the PC program that is a part of the apparatus delivery. Data can be displayed in a chart and graph. They can be printed and also exported for further processing by chart editors e.g. Data may be saved in PC in the internal format of PFM 3000 (*.txt) for further processing. For the balancing of hydraulic systems the program makes it possible to prepare two projects with a maximum of 32 branches of each. During measuring data can viewed directly on the computer monitor.
Minimum PC Requirements
PC 486, 16 MB RAM, min. 2 MB space left on HDD, Windows 95 or higher
y Insert the installation disc into the disc drive.
y In Windows click setup.exe under Start/Run. The
program installs the software automatically.
Operating the Program
Main toolbar
The toolbar includes the icons of the most frequent operations.
From the left: open file, save file, download data from the instrument, print and work with projects.
Selecting units
The program allows various units of pressure and flow to be
entered and displayed. Choose Options/Units Selection to select the units of pressure and flow. You can select from the following units: Pressure
y Pa, hPa, kPa, Mpa y mbar, bar, psi, at y mWs, mmWs, mmHg
Flow
y m3/hod, m3/min, m3/s y l/hod, l/min, l/s
Software Installation
An installation CD-ROM is a part of the instrument accessory. It contents software needed for communication with PC. Installation procedure:
Measured Data Processing
Downloading data from the instrument
Use a communication cable to connect the instrument.
Switch PFM 3000 on. Click Read to download data from
21
PFM 3000
apparatus. Unless PFM 3000 is connected or switched on, the message PFM 3000 not connected will be displayed. If there are no recorded data, the message No recorded data will be displayed. If data have been recorded, they will be displayed in the user software.
Data classification
Data are classified into records from particular places. Select a requested place to receive measured pressure, flow and temperature. Measured data can be viewed also in graphs.
Graph view toolbar
The toolbar includes buttons for the scale adjustment and the graph heading adjustment.
The functions of the icons (from the left): original scale, scale up, scale
down, heading adjustment. To move along the graph, use the right mouse-button. To blow up an area, create a rectangle by clicking the left mouse-button in the left top corner of the area and move to the right bottom corner.
Using data
You can save data to the disc in the PFM 3000 format from
menu File/Save as choose export from File/Export menu to export them in the database format. Exported data you can again open in a spreadsheet programs, for instance Excel.
To browse previously saved data, click File/Open.
Printing
You can print recorded data and graphs indicating pressure, flow and temperature change within a time period with an adjustable recording period. Select a printer type under
Print/Printer set-up…. You can print either table and graphs of stored values. To print a graph, use a landscape page orientation. Select requested data from measured points.
Project preparation and project saving
For the balancing of hydraulic systems the program makes it possible to prepare two projects with a maximum of 32 branches of each.
Project tool bar
Creating a project
Each project includes information about the common valve, input pressure and the common valve set-up. In the Project section fill in the name, the common valve type from the valve list and input pressure. For project without common
valve deselect Shared Valve box. Hit Add button to attach the project to the tree of projects and branches in the left part of the window. To enter branches, select a respective project (click the mouse button at the respective project). In the Branch section fill in the name, requested flow and the valve from the valve list. To attach a branch to the tree of
project and branches, select Add in the Branch section. After entering all branches, save the projects to the disc.
22
PFM 3000
Working with projects
Click Open and Save at the tool bar to download project files from/to the disc. The file can consist of max. of two projects.
Click Read from the instrument and Write to download projects. Projects and their branches may get modified also in the PFM 3000. Save data in the instrument carefully to avoid overwriting previously saved projects or measured data. Therefore before downloading a project from PFM 3000 you have to confirm a requested operation in a displayed dialogue box.
Click Print to print the project on a selected printer.
Maintenance Instructions
Calibrate the instrument every 12 months. Replace sintered filters at pressure inputs every 12 months. Sintered filter is located in the R20 quick coupling (smaller one). You can replace sintered filter by the following drawing:
Parts of the Delivery
PFM 3000 measuring computer 1 pcs battery charger 1 pcs thermometer PT100 digital 1 pcs connecting tubing 1 pair quick coupling R21 1 pair quick coupling R20 1 pair hoses Din 7 mm connection 1 pair measuring adapter G3/4" 1 pair reducer G3/4-G1/2 1 pair measuring nozzle 1 pair CD with the user software 1 pcs RS232 communication cable 1 pcs detailed manual, warranty certificate 1 pcs plastic case 1 PCs
23
PFM 3000
24
Technical Specifications
PFM 3000
CommentUnitValueItem
kPa1 000Pressure range
kPa1 500Maximum overpressure
% of range0,15Linearity, repeability and hysteresis error
% of range0,25Temperature error
% of range0,06Static pressure effect
°C-5 až 90Medium temperature
°C-5 až 50Ambient temperature
°C-10 až 70Storage temperature
PT 100 digitalTemperature probe
°C-20 up to 120Range of Temperature probe
°C+/- 1Temperature error
Li-Ion batteryPower
1 sec up to 24 hoursPeriod of record
RS 232Interface
7,2 V 1250 mAh
hours7Charging time
mm77x192x25Dimensions w x h x d
g390Weight
IP65Cover
months12Calibration validity
at the end of a connecting tubes 1.5 m
max.hours120Operating time
max.1 800No of records
max.valves400No of pre-programmed valves
max.2No of projects
max.32No of branches in each project
25
PFM 3000
26
PFM 3000 MenuPFM 3000 Menu
PFM 3000
Project
Selection
Balanc. Solve
Start
Bind Projects
Results
Source Pressure
Common Valve
New Project
Branch
Selection
Measure
New Branch
Valve Closed
Initial Preset.
Real Flow
Display
Units
Pressure Units
Flow Units
Meas. Filter
Backlight Time
Pa hPa kPa MPa mbar bar psi
at mWs mmWs mmHg
m3/s m3/min m3/hod l/s l/min l/hod
Configuration
Date and Time
Automatic OFF
Language
Minute
Hour
Day
Month
Year
Medium Select. Record
Watter
Ethylenglykol
Propylenglykol
Capacity used
Start
Erase
Period
Place
Hours
Minutes
Seconds
Direct KV Input
Heating Calcul.
Valve Preset
Pressure Drop
Rename Project
Delete Project
Initial Preset.
Edit Branch
Delete Branch
Description
27
29
www.danfoss.com
Loading...