Danfoss PFM 100 User Manual

User Guide
PFM 100 Measuring Instrument
1 Application
The PFM 100 is used for measuring the differential pressure on both sides of a valve in hydronic systems.
The set consists of :
One PFM 100 Measuring Instrument.
Two hoses with couplings.
Two needles incl. ball valves.
The PFM 100 can communicate with you in one of 10 possible lan­guages.
Please note that the pressure unit must not be exposed to sub­zero temperatures. The pressure unit must at all times be stored at temperatures above 0° C.
2 Functions
3 Usage
1. Connect the hoses to the valve.
2. Air the hoses.
3. Connect the valve to the PFM 100. Zero adjust.
4.
From the main menu select the Zeroing option in order
to correct the static pressure influence. To reset the device to zero follow the instruction on the
display.
ON / OFF button.
Menu button.
Used for navigating through the menu options. Press to enter the menu. Press and hold to exit the menu.
Confirmation button.
Press to select the marked item.
5. Type in the Kv-value. If the kV value of the measured valve is known, it can be en­tered in the PFM 100 to display the valve flow.
From the main menu select the Kv value option.
Press the Confirmation button to go through the se-
quence of numbers. Press the Menu button to increase / decrease the num-
bers. Press and hold the Confirmation button to save and
exit the menu.
6.
Read off the flow in the display of the PFM 100.
Danfoss Heating Solutions VUE2D202
1
User Guide PFM 100 Measuring Instrument
4 Technical Specifications
Type Specifications Pressure sensor Piezo resistive true differential Pressure range Max. overpressure
Non-linearity and hysteresis error 0.15% of pressure range Temperature error 1.5% over ambient and medium temperature Medium temperature -5 to 90°C Ambient temperature -5 to 50°C Storage temperature -5 to 50°C Pressure connection R21 quick couplers Batteries 2 x AA NiMH rechargeable Power consumption Max. 55 mA Display 128 x 64 monochromatic, backlight Keyboard 3 keys Pressure units 11 Flow units 11, incl. US units kV-range 0 to 99999, with 0.1 step
Dimensions 94 x 218 x 35 mm Weight 600 g, incl. batteries Cover IP65 Calibration validity 12 months
10 bar Positive side: 15 bar
Negative side: 10 bar
5 Ordering PFM 100
No. Type Set / pcs Danfoss code no. 1 FPM 100 Measuring Instrument 1 003L8260
6 Spare Part List
No. Type Set / pcs Danfoss code no. 1 Measuring hose, 2 x 1.5 m 1 set 003L8261 2 Needle 1 pcs 003L8262 3 Quick coupling 1 pcs 003L8263 4 Filters 1 set 003L8264 5 Strap 1 pcs 003L8265
6
Danfoss A/S Heating Solutions Haarupvaenget 11 8600 Silkeborg Denmark Phone:+45 7488 8000 Fax: +45 7488 8100 Email: heating.solutions@danfoss.com www.heating.danfoss.com
ASV-I/M measuring KIT for PFM100
1 set 003L8274
Danfoss can accept no responsibility for possible errors in catalogues, brochures and other printed material. Danfoss reserves the right to alter its products without notice. This also applies to products already on order provided that such alterations can be made without subsequential changes being necessary in specifications already agreed. All trademarks in this material are property of the respective companies. Danfoss Heating Solutions and the Danfoss Heating Solutions logotype are trademarks of Danfoss A/S. All rights reserved.
VUE2D202 Produced by Danfoss Heating Solutions © 08/2016