Danfoss GDA SC 10000, GDA EC 300, GDA SC 1000, GDA EC 1000, GDC IR 10000 Installation Manual

...
Installation Guide
Making Bump test on Danfoss gas detecting sensors, types GDA and GDC
148 R95 44
148 R95 44
4 DKRCI.PI.S00.B2.02 / 520H2572 © Danfoss A/S (MWA), 2015-02 © Danfoss A/S (MWA), 2015-02 DKRCI.PI.S00.B2.02 / 520H2572 1
Beaker
EC/SC/CT­Adaptor
M35 Adaptor
BUMP TEST equipment
EC/SC/CT-Adaptor Ampoule Breaker
Sensor PCB
Gas ows through the back and front of the sensor head
Bump test on a wall mounted CO2 detector
1
2
3
4
5
ENGLISH
Appendix 1
Checking the zero. (0.0 V output)
EC sensor
To adjust the zero, Pot VR201 must be operated. It is located on the Sensor PCB
Connect a voltmeter to TP0 on the mother PCB and Con3, pin 3 on the mother PCB
Pot VR201 (on the Sensor PCB) is used to adjust the zero of the range (span). Measure the Voltage output between TP0 (negative) and Con3 pin 3 (positive) at the 0V signal and adjust the Pot to 0.0 V or slightly positive (0.01 V is acceptable).
SC sensor
To adjust the zero, Pot RV2 must be operated. It is located on the Sensor PCB.
Connect a voltmeter to TP0 on the mother PCB and Con3, pin 3 on the mother PCB (see g. A1)
Pot RV2 (on the Sensor PCB) is used to adjust the zero of the range (span). Measure the Voltage output between TP0 (negative) and Con3 pin 3 (positive) at the 0V signal and adjust the Pot to 0.0 V or slightly positive (0.01 V is acceptable).
CT sensor
To adjust the zero, Pot R3 must be operated. It is located on the Sensor PCB.
Connect a voltmeter to TP0 on the mother PCB and Con3, pin 3 on the mother PCB (see g. A1).
Pot R3 (on the Sensor PCB) is used to adjust the zero of the range (span). Measure the Voltage output between TP0 (negative) and Con3 pin 3 (positive) at the 0V signal and adjust the Pot to 0.0 V or slightly positive (0.01 V is acceptable).
© Danfoss A/S (MWA), 2015-02 DKRCI.PI.S00.B2.02 / 520H2572 32 DKRCI.PI.S00.B2.02 / 520H2572 © Danfoss A/S (MWA), 2015-02
Table 1 - Bump Test
Pass test criteria
Models
Sensor
Requirement for sensor response to pass Bump Test. Measure Voltage Output on Con3,(mother
PCB)
pin 1 and 3
Standard Basic Standard Basic with
LCD display
IP 65 for High RH and
Fast response
IP 56 enclosure IP 56 enclosure
Low Temperature
EExd model EExd model
Low Temperature
IP 66 enclosure 5 m remote
IP 65 sensor
IP 66 enclosure 5 m remote
IP 65 EExd sensor
1
GDA EC 100 >= 5 .0 V
Beaker + EC/SC/CT-Adaptor Beaker Beaker + EC/SC/CT-Adaptor Beaker
GDA EC 300 >= 1.67 V
Bump test method 100 ppm Ammonia Ampoule
2
GDA EC 1000 >= 5 V Beaker + EC/SC/CT-Adaptor Beaker Beaker + EC/SC/CT-Adaptor Beaker
Bump test method 1000 ppm Ammonia Ampoule
3
GDA SC 1000 >= 5 .0 V Beaker + EC/SC/CT-Adaptor Beaker Beaker + EC/SC/CT-Adaptor Beaker + M35 Adaptor Beaker Beaker + M35 Adaptor
Bump test method 1000 ppm Ammonia Ampoule
4
GDA SC 10000 >= 0.5 V Beaker + EC/SC/CT-Adaptor Beaker Beaker + EC/SC/CT-Adaptor Beaker + M35 Adaptor Beaker Beaker + M35 Adaptor
Bump test method 1000 ppm Ammonia Ampoule
5
GDA CT 30000 >= 5 .0 V Beaker + EC/SC/CT-Adaptor Beaker + M35 Adaptor Beaker + EC/SC/CT-Adaptor Beaker + M35 Adaptor
Bump test method App. 5-8 % Ammonia water in a cloth
1
)
6
GDC IR 10000 >= 1.0 V Beaker Beaker not available Beaker Beaker Beaker Beaker not available not available
Bump test method 2000 ppm CO
2
Ampoule
1
) Drop a cloth to the bottom of the beaker, hold it at the bottom with the screw. Moisten the cloth (5-10 drops) with 5-8 % ammonia water.
Restricted bump test
Models
Sensor
Requirement for sensor response to pass Bump Test. Measure Voltage Output on Con3,(mother
PCB)
pin 1 and 3
Standard Basic Standard Basic with
LCD display
IP 65 for High RH and
Fast response
IP 56 enclosure IP 56 enclosure
Low Temperature
EExd model EExd model
Low Temperature
IP 66 enclosure 5 m remote
IP 65 sensor
IP 66 enclosure 5 m remote
IP 65 EExd sensor
7
GDC IR 20000 >= 2 .0 V
Sensor is self calibrating. Breathe on sensor to see responseGDC IR 40000 >= 2 .0 V
Bump test method
8
GDHC SC 1000 >= 5 V
Long life sensor. Use a cigarette lighter to see response
GDHF SC 1000 >= 5 V
GDHF-R3 SC 1000 >= 5 V
Bump test method
9
GDH SC 5000 >= 5 .0 V
Long life sensor. Use a cigarette lighter to see response
Bump test method
ENGLISH
Introduction
The frequency of testing or calibration is generally determined by local regulation or standards. There are two concepts that need to be dierentiated: bump test and calibration
Bump Test: This consists of exposing the sensor to a gas. The objective is to establish if the sensor is reacting to the gas and all the sensor outputs are working correctly. A qualied bump test is a test carried out using ampoules or cylinders of known concentration.
Calibration: This consists of exposing the sensor to a calibration gas setting the “zero” or “Standby voltage”, the span or range, and checking/adjusting all the outputs, so that they are activated at a specied gas concentration.
Danfoss recommends annual checks by qualied bump test and replacement of the sensor PCB with a Danfoss pre-calibrated certied sensor PCB every two years. The mother PCB is at the same time tested with the GD tester.
The alternative to this is a full on site calibration. Sensor replacement should be more cost eective, eliminates end of life concerns, and constantly renews the detection system.
Procedures for bump test and calibration vary depending on the sensor technology used and the gas in question.
The GD is available in four sensor versions: Semiconductor (SC), Electrochemical (EC), Catalytic (CT), and Infrared (IR) for CO
2
.
Before you carry out the test or calibration:
1. Advise occupants, plant operators, and supervisors.
2. Check if the relays (Con6 and Con7) are connected to external systems such as sprinkler systems, plant shut down, external sirens and beacons, ventilation, etc. and disconnect as appropriate.
3. Deactivate alarm delays with JP1 and JP2. See g. 1. For this the GD must be powered o.
4. For Bump Test the GD should be powered up overnight.
NOTE: If unit has been installed and running for about 24 hrs, and you need to power it o to set the delay at 0 min, then the warm-up is 2 min and you can begin the testing or calibration.
ENGLISH
Bump test (every year)
Ideally bump tests are conducted on site in a clean air atmosphere.
The bump test can be carried out on following types of GD (see table 1) GDA (qualied bump test) GDC (qualied bump test/ restricted bump test) GDHC (restricted bump test) GDHF (restricted bump test) GDHF-R3 (restricted bump test)
See table 1, to nd the GD type and the corresponding Bump test method.
Checking the zero (0 Voltage output). See Appendix 1. (not required for IR sensor)
1. Remove the cover of enclosure of the GD
2. Make sure that both the ampoules and the calibration beaker are clean and dry (see g. 5).
3. Unscrew the beaker hold screw and place the ampoule so that is sits in the base of the beaker (see g. 2).
4. Tighten on the screw ampoule without breaking it.
5. For the models: Standard Basic Standard Basic with LCD display IP 56 enclosure IP 56 enclosure Low Temperature Screw in: EC/SC/CT-Adaptor or M35-Adaptor into the Beaker according to table 1.
Place the beaker with the EC/SC/CT­ Adaptor over the sensor. It should be as tight tting as possible to allow maximum exposure to the gas (see g. 5).
For the GDC (CO2)models: Standard Basic Standard Basic with LCD display IP 56 enclosure IP 56 enclosure Low Temperature
Place the beaker over the sensor in a 45° degree (see g. 3). This allows gas to ow through the back of the sensor and through the calibration ports.
For the models: IP 65 for High RH and Fast response EExd model EExd model Low Temperature IP 66 enclosure 5 m remote IP 65 sensor IP 66 enclosure 5 m remote IP 65 EExd sensor
Screw/t the beaker (possibly with M35 Adaptor) on the remote sensorhead (see g. 4). It should be as tight tting as possible to allow maximum exposure to the gas.
6. Connect volt meter to monitor sensor response. (Con3 pin 1 and 3 - see g. 1).
7. Tighten on the ampoule until it shatters allowing the content to diuse in the beaker. It should be left in place for approximately 5 min.
8. Voltage output will increase. This conrms that the sensor is responding. See Pass test criteria in table 1.
9. Carefully remove any ampoule remains from the gas detector, and reassemble the sensor enclosure.
Fill out the GD Bump Test Certicate.
Loading...