Danfoss Everis 4240, Everis 4280 Installation guide

INSTRUCTIONS
Thermostatic sensor
type RTS-R EverisTM 4240 & 4280
Application
013R9392
Operation
Setting the temperature
Thermostatic sensor type RTS-R EverisTM 4240 and 4280 can be fitted without adapter on valve radiators (compact radiators) from the manufacturers: Baufa, Brötje, Brugman (Piano, VK), Buderus, CICH
The thermostatic sensor with a liquid-filled bellows system provides proportional regulation. The sensor is actuated by ambient temperature. The calibrated pressure in the bellows corresponds to the tempera­ture of the charge. This pressure is balanced by the force of a regulating spring. On a rise in ambient temperature the liquid pressure
The required room temperature is set by turning the setting dial. The temperature scales show the correlation between scale values and the room temperature. The temperature values stated are for guidance only as the obtained room temperature will often be influenced by installation conditions.
(Europanel), De Longhi (Linea, Platella), Jaga (Linea Plus), Northor, Ocean, Potterton - Myson, Schäfer, Termoteknik, Vogel & Noot (Cosmo - Compact).
013R9392
rises in the bellows, moving the valve cone towards the closed position until equilibrium exists between the bellows and the spring. On a drop in ambient temperature the liquid pressure falls, allowing the bellows to contract and the valve cone to move towards the open position until a state of equilibrium is re-established.
The temperature scales are stated according to European standards at Xp = 2 °C. This means that the radiator thermostats close at a sensor temperature which is 2 °C higher than stated on the temperature scales.
Different temperatures are recommended depending on the purpose of the room. The illustration shows the most frequently used.
Mounting
Adjusting temperature control range
Room
temporarily
not in use
Remove the protective cap of the valve.
Before mounting, slack the connection nut of the sensor and turn the setting dial on position 5.
Then press the sensor on the valve and secure the connection ring by hand (5 Nm) to firmly connect the sensor with the valve.
Setting the upper limit
e.g. 3 as the highest possible temperature selection:
1. Turn the handle to max. position.
2. Keep the limiter tab to the right of the scale pointer pressed down with a small screwdriver.
3. Turn the handle to 3 on scale and release the tab.
Cellar Corridor,
bedroom
Setting the lower limit
e.g. 2 as the lowest possible temperature selection:
1. Turn the head to. min. position.
2. Keep the limiter tab to the left of the scale pointer pressed down with a small screwdriver.
3. Turn the head to 2 and release the tab.
Living room,
kitchen
Bath room
5 Nm
DKCD VIVFM102 03/2004
Loading...