DAKOTA ULTRASONICS MODEL VX ULTRASONIC VELOCITY GAUGE
Page 2
Copyright 2008 Dakota Ultrasonics. All rights reserved.
No part of this publication may be reproduced, translated into another
language, stored in a retrieval system, or transmitted in any form or by any
means; electronic, mechanical, photocopying, recording, or otherwise,
without the prior written consent of Dakota Ultrasonics.
Every precaution has been taken in the preparation of this publication.
Dakota Ultrasonics assumes no responsibility for errors or omissions.
Neither is any liability assumed for damages resulting from the use of
information contained herein.
Any brand or product names mentioned herein are used for identification
purposes only, and are trademarks or registered trademarks of their
respective holders.
1500 Green Hills Road, #107 Scotts Valley, CA 95066 USA
Tel (831) 431-9722 Fax (831) 431-9723
www.dakotaultrasonics.com
Page 3
CONTENTS
VX Ultrasonic Velocity Gauge
INTRODUCTION
OPERATION
THE KEYPAD
THE DISPLAY
THE TRANSDUCER
MAKING MEASUREMENTS
CONDITION AND PREPARATION OF SURFACES
PROBE ZERO
CALIBRATION
APPENDIX C:SOUND VELOCITIES OF COMMON MATERIALS
WARRANTY INFORMATION
27
29
DISCLAIMER
Inherent in ultrasonic measurement is the possibility that the instrument
will use the second rather than the first echo from the back surface of the
material being measured. This may result in a reading that is TWICE what
it should be. This typically occurs when the material being measured is
thinner than the minimum capability of the transducer being used.
Responsibility for proper use of the instrument and recognition of this
phenomenon rests solely with the user of the instrument.
Page 4
Dakota Ultrasonics
blank page
Page 5
VX Ultrasonic Velocity Gauge
1
INTRODUCTION
The Dakota Ultrasonics model VX is an Ultrasonic Velocity Gauge. A
velocity gauge displays the speed of sound through materials of a known
thickness by measuring one-half the time-of-flight of a sound wave
transmitted and received at a single point. The sound velocity is
represented in units of Inches-per-microsecond or Meters-per-second. A
velocity gauge is useful when a material type or consistency is not known.
This manual is presented in three sections. The first section covers
operation of the VX, and explains the keypad controls and display. The
second section provides guidelines in selecting a transducer for a specific
application. The last section provides application notes and a table of
sound velocity values for various materials.
Dakota Ultrasonics maintains a customer support resource in order to
assist users with questions or difficulties not covered in this manual.
Customer support may be reached at any of the following:
• Dakota Ultrasonics, 1500 Green Hills Road, #107
Scotts Valley, CA 95066 USA
• Telephone: (831) 431- 9722
• Facsimile: (831) 431-9723
• www.dakotaultrasonics.com
Page 6
Dakota Ultrasonics
2
blank page
Page 7
VX Ultrasonic Velocity Gauge
3
OPERATION
The VX interacts with the operator through the membrane keypad and
the LCD display. The functions of the various keys on the keypad are
detailed below, followed by an explanation of the display and its various
symbols.
The Keypad
This key is used to turn the VX on and off. When the gauge is turned
ON, it will first perform a brief display test by illuminating all of the
segments in the display. After one second, the gauge will displ ay the
internal software version number. After displaying the version number, the
display will show .0000 IN / µs (or 0000 M /s if using metric units), indicating
the gauge is ready for use.
Pressing the ON/OFF key turns OFF the VX. The gauge has a special
memory that retains all of its settings even when the power is off. The
gauge also features an auto-powerdown mode designed to conserve
battery life. If the gauge is idle for 5 minutes, it will turn itself off.
Page 8
Dakota Ultrasonics
4
The PRB- 0 key is used to "zero" the VX in much the same way that a
mechanical micrometer is zeroed. If the gauge is not zeroed correctly, all
of the measurements that the gauge makes may be in error by some fixed
value. Refer to page 12 for an explanation of this important procedure.
The CAL key is used to enter and exit the VX's calibration mode. This
mode is used to adjust the material thickness value that the VX will use
when calculating sound velocity. The gauge will either calculate the soundvelocity from a sample of the material being measured, or allow a known
velocity value to be entered directly. Refer to page 13 for an explanation of
the two CAL functions available.
The IN/MM key is used to switch back and forth between English and
metric units. This key may be used at any time, whether the gauge is
displaying a thickness (IN or MM) or a velocity value (IN/µs or M/s).
Page 9
VX Ultrasonic Velocity Gauge
5
The UP arrow key has two functions. When the VX is in calibration
mode, this key is used to increase numeric values on the display. An autorepeat function is built in, so that when the key is held down, numeric
values will increment at an increasing rate. When the VX is not in
calibration mode, the UP arrow key switches the SCAN measurement
mode on and off. Refer to page 16 for an explanation of the SCAN
measurement mode.
The DOWN arrow key has two functions. When the VX is in the CAL
mode, this key is used to decrease numeric values on the display. An
auto-repeat function is built in, so that when the key is held down, numeric
values will decrement at an increasing rate. When the VX is not in
calibration mode, the DOWN arrow key switches the display backlight
between three available settings. OFF will be displayed when the backlight
is switched off. AUTO will be displayed when the backlight is set to
automatic mode, and ON will be displayed when the backlight is set to stay
on. In the AUTO setting, the backlight will illuminate when the VX is
actually making a measurement.
Page 10
Dakota Ultrasonics
6
The Display
The numeric portion of the display consists of 4 complete digits
preceded by a leading "1", and is used to display numeric values, as well
as occasional simple words, to indicate the status of various settings.
When the VX is displaying velocity measurements, the display will hold the
last value measured, until a new measurement is made. Additionally,
when the battery voltage is low, the entire display will begin to flash.
When this occurs, the batteries should be replaced.
These eight vertical bars form the Stability Indicator. When the VX is
idle, only the left -most bar and the underline will be on. When the gauge is
making a measurement, six or seven of the bars should be on. If fewer
than five bars are on, the VX is having difficulty achieving a stable
measurement, and the value displayed may not be accurate.
Page 11
VX Ultrasonic Velocity Gauge
7
When the IN symbol is on, the VX is displaying a thickness value in
inches. The maximum thickness that can be displayed is 19.999 inches.
When the MM symbol is on, the VX is displaying a thickness value in
millimeters. If the displayed thickness exceeds 199.99 millimeters, the
decimal point will shift automatically to the right, allowing values up to
1999.9 millimeters to be displayed.
When the IN symbol is on, in conjunction with the /µs symbol, the VX is
displaying a sound-velocity value in inches-per-microsecond.
When the M symbol is on, in conjunction with the /s symbol, the VX is
displaying a sound-velocity value in meters-per-second.
Page 12
Dakota Ultrasonics
8
The Transducer
The transducer is the "business end" of the VX. It transmits and
receives the ultrasonic sound waves that the VX uses to calculate the
sound velocity of the material being measured. The transducer connects to
the VX via the attached cable, and two coaxial connectors. When using
transducers manufactured by Dakota Ultrasonics, the orientation of the
dual coaxial connectors is not critical: either plug may be fitted to either
socket in the VX.
The transducer must be used correctly in order for the VX to produce
accurate, reliable measurements. Below is a short description of the
transducer, followed by instructions for its use.
This is a bottom view of a typical transducer. The two semicircles of the
wearface are visible, as is the barrier separating them. One of the
semicircles is responsible for conducting ultrasonic sound into the material
being measured, and the other semicircle is responsible for conducting the
echoed sound back into the transducer. When the transducer is placed
against the material being measured, it is the area directly beneath the
center of the wearface that is being measured.
Page 13
VX Ultrasonic Velocity Gauge
9
This is a top view of a typical transducer. Press against the top with the
thumb or index finger to hold the transducer in place. Moderate pressure is
sufficient, as it is only necessary to keep the transducer stationary, and the
wearface seated flat against the surface of the material being measured.
Making Measurements
In order for the transducer to do its job, there must be no air gaps
between the wear-face and the surface of the material being measured.
This is accomplished with the use of a "coupling" fluid, commonly called
"couplant". This fluid serves to "couple", or transmit, the ultrasonic sound
waves from the transducer, into the material, and back again. Before
attempting to make a measurement, a small amount of couplant should be
applied to the surface of the material being measured. Typically, a single
droplet of couplant is sufficient.
After applying couplant, press the transducer (wearface down) firmly
against the area to be measured. The Stability Indicator should have six or
seven bars darkened, and a number should appear in the display. If the
VX has been properly "zeroed" (see page 12) and calibrated to a known
thickness (see page 13), the number in the display will indicate the actual
sound velocity of the material directly beneath the transducer.
Page 14
Dakota Ultrasonics
10
If the Stability Indicator has fewer than five bars darkened, or the
numbers on the display seem erratic, first check to make sure that there is
an adequate film of couplant beneath the transducer, and that the
transducer is seated flat against the material. If the condition persists, it
may be necessary to select a different transducer (size or frequency) for
the material being measured. See page 17 for information on transducer
selection.
While the transducer is in contact with the material being measu red, the
VX will perform four measurements every second, updating its display as it
does so. When the transducer is removed from the surface, the display will
hold the last measurement made.
IMPORTANT
Occasionally, a small film of couplant will be drawn out between the
transducer and the surface as the transducer is removed. When this
happens, the VX may perform a measurement through this couplant film,
resulting in a measurement that is larger or smaller than it should be. This
phenomenon is obvious when one value is observed while the transducer
is in contact with the material, and another value is observed after the
transducer is removed.
Page 15
VX Ultrasonic Velocity Gauge
11
Condition and Preparation of Surfaces
In any ultrasonic measurement scenario, the shape and roughness of
the test surface are of paramount importance. Rough, uneven surfaces
may limit the penetration of ultrasound through the material, and result in
unstable, and therefore unreliable, measurements. The surface being
measured should be clean, and free of any small particulate matter, rust, or
scale. The presence of such obstructions will prevent the transducer from
seating properly against the surface. Often, a wire brush or scraper will be
helpful in cleaning surfaces. In more extreme cases, rotary sanders or
grinding wheels may be used, though care must be taken to prevent
surface gouging, which will inhibit proper transducer coupling.
Extremely rough surfaces, such as the pebble-like finish of some cast
irons, will prove most difficult to measure. These kinds of surfaces act on
the sound beam like frosted glass on light, the beam becomes diffused and
scattered in all directions.
In addition to posing obstacles to measurement, rough surfaces
contribute to excessive wear of the transducer, particularly in situations
where the transducer is "scrubbed" along the surface. Transducers should
be inspected on a regular basis, for signs of uneven wear of the wearface.
If the wearface is worn on one side more than another, the sound beam
penetrating the test material ma y no longer be perpendicular to the material
surface. In this case, it will be difficult to exactly locate tiny irregularities in
the material being measured, as the focus of the soundbeam no longer lies
directly beneath the transducer.
Page 16
Dakota Ultrasonics
12
Probe Zero
Setting the Zero Point of the VX is important for the same reason that
setting the zero on a mechanical micrometer is important. If the gauge is
not "zeroed" correctly, all of the measurements the gauge makes will be in
error by some fixed number. When the VX is "zeroed", this fixed error
value is measured and automatically corrected for in all subsequent
measurements. The VX may be "zeroed" by performing the following
procedure:
Performing a Probe-Zero
1) Make sure the VX is on.
2) Plug the transducer into the VX. Make sure that the connectors are
fully engaged. Check that the wearface of the transducer is clean
and free of any debris.
3) On the top of the VX, above the display, is the metal probe-disc.
Apply a single droplet of ultrasonic couplant to the face of this disc.
4) Press the transducer against the probe- disc, making sure that the
transducer sits flat against the surface of the probe-disc. The
display should show some value, and the Stability Indicator should
have nearly all its bars illuminated.
5) While the transducer is firmly coupled to the probe-disc, press the
PRB-0 key on the keypad. The VX will display "Prb0" while it is
calculating its zero point.
6) Remove the transducer from the probe- disc.
Page 17
VX Ultrasonic Velocity Gauge
13
At this point, the VX has successfully calculated it's internal error factor,
and will compensate for this value in any subsequent measurements.
When performing a "probe-zero", the VX will always use the sound-velocity
value of the built-in probe- disc, even if some other velocity value has been
entered for making actual measurements. Though the VX will remember
the last "probe-zero" performed, it is generally a good idea to perform a
"probe-zero" whenever the gauge is turned on, as well as any time a
different transducer is used. This will ensure that the instrument is always
correctly zeroed.
Calibration
In order for the VX to make accurate measurements, it must be set to
the exact thickness the material being measured. Different types of
material have different inherent sound-velocities. For ex ample, the velocity
of sound through steel is about 0.233 inches- per-microsecond, versus that
of aluminum, which is about 0.248 inches- per-microsecond. If the gauge is
not set to the correct thickness, all of the measurements the gauge makes
will be erron eous.
Page 18
Dakota Ultrasonics
14
Calibration to a known thickness
NOTE: This procedure requires a sample piece of the specific
material to be measured, the exact thickness of which is known,
e.g. from having been measured by some other means.
1) Make sure the VX is on.
2) Perform a Probe-Zero (refer to page 12)
3) Apply couplant to the sample piece.
4) Press the transducer against the sample piece, making sure that
the transducer sits flat against the surface of the sample. The
display sh ould show some (probably incorrect) sound velocity
value, and the Stability Indicator should have nearly all its bars on.
5) Having achieved a stable reading, remove the transducer. If the
displayed value changes from the value shown while the
transducer was coupled, repeat step 4.
7) Press the CAL key. The IN / µs (or M /s) symbol should begin
flashing.
8) Press the CAL key again. The IN (or MM ) symbols should begin
flashing and a thickness value is displayed. Use the UP and
DOWN arrow keys to adjust the displayed thickness up or down,
until it matches the thickness of the sample piece.
9) Press the CAL key once more to exit the calibration mode. The VX
will now display sound velocities of materials of the same thickness
as entered.
Page 19
VX Ultrasonic Velocity Gauge
15
Calibration to a known velo city
NOTE: This procedure requires that the operator know the sound-
velocity of the material to be measured. A table of common
materials and their sound-velocities can be found in Appendix C.
1) Place the transducer on the material to be measured and obtain a
stable reading.
2) Press the CAL key to enter calibration mode. The IN /µs (or M /s)
symbol will begin flashing.
3) Use the UP and DOWN arrow keys to adjust the displayed velocity
up or down, until it matches the known sound-velocity of the
material.
4) Press the CAL key again and the display will show the thickness
value calculated for the material.
4) Press the CAL key once more to exit the calibration mode. The VX
is now ready to perform measurements.
NOTE: During the calibration procedure when IN /µs, or M /s is
flashing, pressing the PRB-0 key will restore the gauge to the factory
default sound-velocity for steel (.2330 IN / µs or 5920 M /s).
Page 20
Dakota Ultrasonics
16
Scan Mode
While the VX excels at making single point measurements, it is
sometimes desirable to examine a larger region to check for material
consistency. The VX includes a feature, called Scan Mode, which allows it
to do just that.
In normal operation, the VX performs and displays four measurements
every second, which is quite adequate for single measurements. In Scan
Mode, however, the gauge performs sixteen measurements every second,
but does not display them. While the transducer is in contact with the
material being measured, the VX is keeping track of the fastest velocity it
finds. The transducer may be "scrubbed" across a surface, and any brief
interruptions in the signal will be ignored. When the transducer loses
contact with the surface for more than a second, the VX will display the
fastest sound velocity it found.
When the VX is not in calibration mode, press the UP arrow key to turn
Scan Mode on and off. A brief message will appear in the display
confirming the operation. While scanning, the display will show a moving
series of dashes instead of a sound value.
Page 21
VX Ultrasonic Velocity Gauge
17
TRANSDUCER SELECTION
The VX is inherently capable of performing measurements on a wide
range of materials, from various metals to glass and plastics. Different
types of material, however, will require the use of different transducers.
Choosing the correct transducer for a job is critical to being able to easily
perform accurate and reliable measurements. The following paragraphs
highlight the important properties of transducers, which should be
considered when selecting a transducer for a specific job.
Generally speaking, the best transducer for a job is one that sends
sufficient ultrasonic energy into the material being measured such that a
strong, stable echo is received by the VX. Several factors affect the
strength of ultrasound as it travels. These are outlined below:
• Initial Signal Strength
The stronger a signal is to begin with, the stronger its return echo
will be. Initial signal strength is largely a factor of the size of the
ultrasound emitter in the transducer. A large emitting area will send
more energy into the mater ial being measured than a small emitting
area. Thus, a so-called "1/2-inch" transducer will emit a stronger signal
than a "1/4-inch" transducer.
• Absorption and Scattering
As ultrasound travels through any material, it is partly absorbed. If
the material through which it travels has any grain structure, the sound
waves will also experience scattering. Both of these effects reduce the
Page 22
Dakota Ultrasonics
18
strength of the waves, and thus, the VX's ability to detect the returning
echo.
Higher frequency ultrasound is absorbed and scattered more than
ultrasound of a lower frequency. While it may seem that using a lower
frequency transducer might be better in every instance, low frequencies
are less directional than high frequencies. Thus, a higher frequency
transducer would be a better choice for detecting the exact location of
small pits or flaws in the material being measured.
• Geometry of the Transducer
The physical constraints of the measuring environment sometimes
determine a transducer's suitability for a given job. So me transducers
may simply be too large to be used in tightly confined areas. Also, the
surface area available for contacting with the transducer may be limited,
requiring the use of a transducer with a small wearface. Measuring on
a curved surface, such as an engine cylinder wall, may require the use
of a transducer with a matching curved wearface.
• Temperature of the Material
When it is necessary to measure on surfaces that are exceedingly
hot, high temperature transducers must be used. These transducers
are built using special materials and techniques that allow them to
withstand high temperatures without damage. Additionally, care must
be taken when performing a "Probe-Zero" or "Calibration to Known
Thickness" with a high temperature transducer. Se e Appendix B for
more information on measuring materials with a high temperature
transducer.
Page 23
VX Ultrasonic Velocity Gauge
19
Selection of the proper transducer is often a matter of tradeoffs between
various characteristics. It may be necessary to experiment with a variety of
transducers in order to find one that works well for a given job. Dakota
Ultrasonics can provide assistance in choosing a transducer, and offers a
broad selection of transducers for evaluation in specialized applications.
Page 24
Dakota Ultrasonics
20
Physical
Weight: 10 ounces
Size: 2.5W x 4.75H x 1.25D inches
(63.5W x 120.7H x 31.8D mm).
Operating Temperature: -20 to 120 °F (-20 to 50 °C)
Case: Extruded aluminum body / nickel plated aluminum end
caps.
APPENDIX A
Product Specifications
Keypad
Sealed membrane, resistant to water and petroleum products.
Power Source
Two “AA” size, 1.5 volt alkaline or 1.2 volt NiCad cells. 200 hours
typical operating time on alkaline, 120 hours on NiCad.
Display
Liquid-Crystal-Display, 4.5 digits, 0.500-inch high numerals. LED
backlight.
Measuring
Range: 0.025 to 19.999 inches (0.63 to 500 millimeters)
Resolution: 0.001 inch (0.01 millimeter)
Accuracy: ±0.001 inch (0.01 millimeter), depends on material
and conditions
Sound Velocity Range: .0492 to .3930 IN/ µs (1250 to 10000 M/s)
Page 25
VX Ultrasonic Velocity Gauge
21
blank page
Page 26
Dakota Ultrasonics
22
APPENDIX B
Application Notes
• Measuring hot surfaces
The velocity of sound through a substance is dependant upon its
temperature. As materials heat up, the velocity of sound through them
decreases. In most applications with surface temperatures less than about
200°F (100 °C), no special procedures must be observed. At temperatures
above this point, the change in sound velocity of the material being
measured starts to have a noticeable effect upon ultrasonic measurement.
At such elevated temperatures, it is recommended that the user perform
a calibration procedure (refer to page 11) on a sample piece of known
thickness, which is at or near the temperature of the material to be
measured. This will allow the VX to correctly calculate the velocity of
sound through the hot material.
When performing measurements on hot surfaces, it may also be
necessary to use a specially constructed high -temperature transducer.
These transducers are built using materials that can withstand high
temperatures. Even so, it is recommended that the probe be left in contact
with the surface for as short a time as needed to acquire a stable
measurement. While the transducer is in contact with a hot surface, it will
begin to heat up itself, and through thermal expansion and other effects
may begin to adversely affect the accuracy of measurements.
Page 27
VX Ultrasonic Velocity Gauge
23
• Measuring laminated materials
Laminated materials are unique in that their density (and therefore
sound-velocity) may vary considerably from one piece to another. Some
laminated materials may even exhibit noticeable changes in sound-velocity
across a single surface. The only way to reliably measure such materials is
by performing a calibration procedure on a sample piece of known
thickness. Ideally, this sample material should be a pa rt of the same piece
being measured, or at least from the same lamination batch. By calibrating
to each test piece individually, the effects of variation of sound-velocity will
be minimized.
An additional important consideration when measuring laminates, is that
any included air gaps or pockets will cause an early reflection of the
ultrasound beam. This effect will be noticed as a sudden increase in sound
speed in an otherwise regular surface. While this may impede accurate
measurement of the material, it does provide the user with positive
indication of air gaps in the laminate.
Dakota Ultrasonics warrants the VX against defects in materials and workmanship
for a period of five years from receipt by the end user. Additionally, Dakota Ultrasonics
warrants transducers and accessories against such defects for a period of 90 days from
receipt by the end user. If Dakota Ultrasonics receives notice of such defects during the
warranty period, Dakota Ultrasonics will either, at its option, repair or replace products
that prove to be defective.
Should Dakota Ultrasonics be unable to repair or replace the product within a
reasonable amount of time, the customer's alternative exclusive remedy shall be refund
of the purchase price upon return of the product.
• Exclusions •
The above warranty shall not apply to defects resulting from: improper or
inadequate maintenance by the customer; unauthorized modification or misuse; or
operation outside the environmental specifications for the product.
Dakota Ultrasonics makes no other warranty, either express or implied, with respect
to this product. Dakota Ultrasonics specifically disclaims any implied warranties of
merchantability or fitness for a particular purpose. Some states or provinces do not
allow limitations on the duration of an implied warranty, so the above limitation or
exclusion may not apply to you. However, any implied warranty of merchantability or
fitness is limited to the five-year duration of this written warranty.
This warranty gives you specific legal rights, and you may also have other rights,
which may vary from state to state or province to province.
• Obtaining Service During Warranty Period •
If your hardware should fail during the warranty period, contact Dakota Ultrasonics
and arrange for servicing of the product. Retain proof of purchase in order to obtain
warranty service.
For products that require servicing, Dakota Ultrasonics may use one of the following
methods:
- Repair the product
- Replace the product with a re-manufactured unit
- Replace the product with a product of equal or greater performance
- Refund the purchase price.
• After the Warranty Period •
If your hardware should fail after the warranty period, contact Dakota Ultrasonics for
details of the services available, and to arrange for non-warranty service.
Page 32
Dakota Ultrasonics
28
blank page
Page 33
MATERIAL SAFETY DATA SHEET
N/A = not applicable or not available(To comply with 29 CFR 1910.1200)
SECTION 1 – PRODUCT IDENTIFICATION
Product Name: SOUNDSAFE
Generic Name: Ultrasonic Couplant
Manufacturer: Sonotech, Inc.
774 Marine Dr., Bellingham, WA 98225
(360) 671-9121
SECTION 2 – HAZARDOUS INGREDIENTS
This material does not contain any ingredients having
known health hazards in concentrations greater than 1%.
This material does not contain any known or suspected
carcinogens.
SECTION 4 – FIRE AND EXPLOSION
HAZARD DATA
Flash Point : none
Upper Exposure Limit: none
Lower Exposure Limit: none
Special Fire Fighting Procedures : N/A
Extinguishing media: N/A
Unusual Fire and Explosion Hazards : none
Boiling Point: >220 °FpH: 7.35 – 7.9
Freezing Point: <20°FAcoustic Imp.: 1.726x10
Vapor Pressure: N/AVapor Density: N/A
Evaporation Rate: N/ASpecific Gravity: >1.02
Solubility in Water: complete
Appearance and Odor: water white, opaque gel;
bland odor
6
SECTION 5 – REACTIVITY DATA
Stability: Stable
Conditions to Avoid: none
Incompatibility (Materials to Avoid): none known
Hazardous Polymerization: will not occur
Hazardous Decomposition or Byproducts: none known
SECTION 6 – HEALTH HAZARD AND FIRST AID DATA
Routes of Entry :
Skin: not likelyIngestion: not normally
Eyes: not normallyInhalation: no
Effects of Overexposure: Acute: May cause temporary
1
eye irritation
Chronic: none expected
SECTION 7 – STORAGE AND HANDLING
INFORMATION
Precautions to be taken in handling and storage: Store
between 20°F and 120 °F. Spills are slippery and should
be cleaned up immediately.
Steps to be taken in case material is released or spilled:
Pick up excess for disposal. Clean with water.
Waste disposal method: Dispose of in accordance with
federal, state, and local regulations.
1
SOUNDSAFE contains only food grade and cosmetic grade ingredients.
SONOTECH, INC.
Toll Free: 1-800-458-4254
774 Marine Dr., Bellingham, WA 98225
Telephone: (360) 671-9121
First Aid Procedures :
Skin: Remove with water if desired.
Eyes: Flush with water for 15 minutes.
Ingestion: For large quantities, induce vomiting and
call a physician.
Inhalation: N/A
SECTION 8 – CONTROL MEASURES
Respiratory Protection: not required
Ventilation: not required
Protective Gloves : on individuals demonstrating
sensitivity to SOUNDSAFE
Eye Protection: as required by working conditions
Other Protective Equipment : not required
Fax: (360) 671-9024
Page 34
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.