Dakota VX User Manual

Page 1
OPERATION MANUAL
P/N P
138
0002
Rev 1.
9, January
200
8
DAKOTA ULTRASONICS MODEL VX ULTRASONIC VELOCITY GAUGE
Page 2
Copyright 2008 Dakota Ultrasonics. All rights reserved.
No part of this publication may be reproduced, translated into another language, stored in a retrieval system, or transmitted in any form or by any means; electronic, mechanical, photocopying, recording, or otherwise, without the prior written consent of Dakota Ultrasonics.
Every precaution has been taken in the preparation of this publication. Dakota Ultrasonics assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from the use of information contained herein.
Any brand or product names mentioned herein are used for identification purposes only, and are trademarks or registered trademarks of their respective holders.
1500 Green Hills Road, #107 Scotts Valley, CA 95066 USA
Tel (831) 431-9722 Fax (831) 431-9723
www.dakotaultrasonics.com
Page 3
CONTENTS
VX Ultrasonic Velocity Gauge
INTRODUCTION OPERATION
THE KEYPAD THE DISPLAY THE TRANSDUCER MAKING MEASUREMENTS CONDITION AND PREPARATION OF SURFACES PROBE ZERO CALIBRATION
SCAN MODE TRANSDUCER SELECTION APPENDIX A: PRODUCT SPECIFICATIONS APPENDIX B: APPLICATION NOTES
1 3 3 6 8
9 11 12 13 17 18 21 23
APPENDIX C: SOUND VELOCITIES OF COMMON MATERIALS WARRANTY INFORMATION
27 29
DISCLAIMER
Inherent in ultrasonic measurement is the possibility that the instrument will use the second rather than the first echo from the back surface of the material being measured. This may result in a reading that is TWICE what it should be. This typically occurs when the material being measured is thinner than the minimum capability of the transducer being used. Responsibility for proper use of the instrument and recognition of this phenomenon rests solely with the user of the instrument.
Page 4
Dakota Ultrasonics
blank page
Page 5
VX Ultrasonic Velocity Gauge
1
INTRODUCTION
The Dakota Ultrasonics model VX is an Ultrasonic Velocity Gauge. A velocity gauge displays the speed of sound through materials of a known thickness by measuring one-half the time-of-flight of a sound wave transmitted and received at a single point. The sound velocity is represented in units of Inches-per-microsecond or Meters-per-second. A velocity gauge is useful when a material type or consistency is not known.
This manual is presented in three sections. The first section covers operation of the VX, and explains the keypad controls and display. The
second section provides guidelines in selecting a transducer for a specific application. The last section provides application notes and a table of sound velocity values for various materials.
Dakota Ultrasonics maintains a customer support resource in order to assist users with questions or difficulties not covered in this manual. Customer support may be reached at any of the following:
Dakota Ultrasonics, 1500 Green Hills Road, #107 Scotts Valley, CA 95066 USA
Telephone: (831) 431- 9722
Facsimile: (831) 431-9723
www.dakotaultrasonics.com
Page 6
Dakota Ultrasonics
2
blank page
Page 7
VX Ultrasonic Velocity Gauge
3
OPERATION
The VX interacts with the operator through the membrane keypad and the LCD display. The functions of the various keys on the keypad are detailed below, followed by an explanation of the display and its various symbols.
The Keypad
This key is used to turn the VX on and off. When the gauge is turned ON, it will first perform a brief display test by illuminating all of the
segments in the display. After one second, the gauge will displ ay the internal software version number. After displaying the version number, the
display will show .0000 IN / µs (or 0000 M /s if using metric units), indicating the gauge is ready for use.
Pressing the ON/OFF key turns OFF the VX. The gauge has a special memory that retains all of its settings even when the power is off. The gauge also features an auto-powerdown mode designed to conserve battery life. If the gauge is idle for 5 minutes, it will turn itself off.
Page 8
Dakota Ultrasonics
4
The PRB- 0 key is used to "zero" the VX in much the same way that a mechanical micrometer is zeroed. If the gauge is not zeroed correctly, all
of the measurements that the gauge makes may be in error by some fixed value. Refer to page 12 for an explanation of this important procedure.
The CAL key is used to enter and exit the VX's calibration mode. This mode is used to adjust the material thickness value that the VX will use
when calculating sound velocity. The gauge will either calculate the sound­velocity from a sample of the material being measured, or allow a known velocity value to be entered directly. Refer to page 13 for an explanation of the two CAL functions available.
The IN/MM key is used to switch back and forth between English and metric units. This key may be used at any time, whether the gauge is
displaying a thickness (IN or MM) or a velocity value (IN/µs or M/s).
Page 9
VX Ultrasonic Velocity Gauge
5
The UP arrow key has two functions. When the VX is in calibration mode, this key is used to increase numeric values on the display. An auto­repeat function is built in, so that when the key is held down, numeric
values will increment at an increasing rate. When the VX is not in calibration mode, the UP arrow key switches the SCAN measurement mode on and off. Refer to page 16 for an explanation of the SCAN measurement mode.
The DOWN arrow key has two functions. When the VX is in the CAL mode, this key is used to decrease numeric values on the display. An auto-repeat function is built in, so that when the key is held down, numeric values will decrement at an increasing rate. When the VX is not in calibration mode, the DOWN arrow key switches the display backlight between three available settings. OFF will be displayed when the backlight is switched off. AUTO will be displayed when the backlight is set to automatic mode, and ON will be displayed when the backlight is set to stay on. In the AUTO setting, the backlight will illuminate when the VX is actually making a measurement.
Page 10
Dakota Ultrasonics
6
The Display
The numeric portion of the display consists of 4 complete digits preceded by a leading "1", and is used to display numeric values, as well as occasional simple words, to indicate the status of various settings.
When the VX is displaying velocity measurements, the display will hold the last value measured, until a new measurement is made. Additionally,
when the battery voltage is low, the entire display will begin to flash. When this occurs, the batteries should be replaced.
These eight vertical bars form the Stability Indicator. When the VX is idle, only the left -most bar and the underline will be on. When the gauge is making a measurement, six or seven of the bars should be on. If fewer
than five bars are on, the VX is having difficulty achieving a stable measurement, and the value displayed may not be accurate.
Page 11
VX Ultrasonic Velocity Gauge
7
When the IN symbol is on, the VX is displaying a thickness value in inches. The maximum thickness that can be displayed is 19.999 inches.
When the MM symbol is on, the VX is displaying a thickness value in millimeters. If the displayed thickness exceeds 199.99 millimeters, the decimal point will shift automatically to the right, allowing values up to
1999.9 millimeters to be displayed.
When the IN symbol is on, in conjunction with the /µs symbol, the VX is displaying a sound-velocity value in inches-per-microsecond.
When the M symbol is on, in conjunction with the /s symbol, the VX is displaying a sound-velocity value in meters-per-second.
Page 12
Dakota Ultrasonics
8
The Transducer
The transducer is the "business end" of the VX. It transmits and receives the ultrasonic sound waves that the VX uses to calculate the sound velocity of the material being measured. The transducer connects to the VX via the attached cable, and two coaxial connectors. When using transducers manufactured by Dakota Ultrasonics, the orientation of the dual coaxial connectors is not critical: either plug may be fitted to either socket in the VX.
The transducer must be used correctly in order for the VX to produce accurate, reliable measurements. Below is a short description of the
transducer, followed by instructions for its use.
This is a bottom view of a typical transducer. The two semicircles of the wearface are visible, as is the barrier separating them. One of the semicircles is responsible for conducting ultrasonic sound into the material being measured, and the other semicircle is responsible for conducting the echoed sound back into the transducer. When the transducer is placed against the material being measured, it is the area directly beneath the center of the wearface that is being measured.
Page 13
VX Ultrasonic Velocity Gauge
9
This is a top view of a typical transducer. Press against the top with the thumb or index finger to hold the transducer in place. Moderate pressure is sufficient, as it is only necessary to keep the transducer stationary, and the wearface seated flat against the surface of the material being measured.
Making Measurements
In order for the transducer to do its job, there must be no air gaps between the wear-face and the surface of the material being measured. This is accomplished with the use of a "coupling" fluid, commonly called "couplant". This fluid serves to "couple", or transmit, the ultrasonic sound waves from the transducer, into the material, and back again. Before attempting to make a measurement, a small amount of couplant should be applied to the surface of the material being measured. Typically, a single droplet of couplant is sufficient.
After applying couplant, press the transducer (wearface down) firmly against the area to be measured. The Stability Indicator should have six or seven bars darkened, and a number should appear in the display. If the VX has been properly "zeroed" (see page 12) and calibrated to a known thickness (see page 13), the number in the display will indicate the actual sound velocity of the material directly beneath the transducer.
Page 14
Dakota Ultrasonics
10
If the Stability Indicator has fewer than five bars darkened, or the numbers on the display seem erratic, first check to make sure that there is an adequate film of couplant beneath the transducer, and that the transducer is seated flat against the material. If the condition persists, it may be necessary to select a different transducer (size or frequency) for the material being measured. See page 17 for information on transducer selection.
While the transducer is in contact with the material being measu red, the VX will perform four measurements every second, updating its display as it does so. When the transducer is removed from the surface, the display will hold the last measurement made.
IMPORTANT
Occasionally, a small film of couplant will be drawn out between the transducer and the surface as the transducer is removed. When this happens, the VX may perform a measurement through this couplant film, resulting in a measurement that is larger or smaller than it should be. This phenomenon is obvious when one value is observed while the transducer is in contact with the material, and another value is observed after the transducer is removed.
Page 15
VX Ultrasonic Velocity Gauge
11
Condition and Preparation of Surfaces
In any ultrasonic measurement scenario, the shape and roughness of the test surface are of paramount importance. Rough, uneven surfaces may limit the penetration of ultrasound through the material, and result in unstable, and therefore unreliable, measurements. The surface being measured should be clean, and free of any small particulate matter, rust, or scale. The presence of such obstructions will prevent the transducer from seating properly against the surface. Often, a wire brush or scraper will be helpful in cleaning surfaces. In more extreme cases, rotary sanders or grinding wheels may be used, though care must be taken to prevent surface gouging, which will inhibit proper transducer coupling.
Extremely rough surfaces, such as the pebble-like finish of some cast irons, will prove most difficult to measure. These kinds of surfaces act on the sound beam like frosted glass on light, the beam becomes diffused and scattered in all directions.
In addition to posing obstacles to measurement, rough surfaces contribute to excessive wear of the transducer, particularly in situations where the transducer is "scrubbed" along the surface. Transducers should be inspected on a regular basis, for signs of uneven wear of the wearface. If the wearface is worn on one side more than another, the sound beam penetrating the test material ma y no longer be perpendicular to the material surface. In this case, it will be difficult to exactly locate tiny irregularities in the material being measured, as the focus of the soundbeam no longer lies directly beneath the transducer.
Page 16
Dakota Ultrasonics
12
Probe Zero
Setting the Zero Point of the VX is important for the same reason that setting the zero on a mechanical micrometer is important. If the gauge is not "zeroed" correctly, all of the measurements the gauge makes will be in
error by some fixed number. When the VX is "zeroed", this fixed error value is measured and automatically corrected for in all subsequent measurements. The VX may be "zeroed" by performing the following procedure:
Performing a Probe-Zero
1) Make sure the VX is on.
2) Plug the transducer into the VX. Make sure that the connectors are fully engaged. Check that the wearface of the transducer is clean and free of any debris.
3) On the top of the VX, above the display, is the metal probe-disc. Apply a single droplet of ultrasonic couplant to the face of this disc.
4) Press the transducer against the probe- disc, making sure that the transducer sits flat against the surface of the probe-disc. The display should show some value, and the Stability Indicator should have nearly all its bars illuminated.
5) While the transducer is firmly coupled to the probe-disc, press the
PRB-0 key on the keypad. The VX will display "Prb0" while it is calculating its zero point.
6) Remove the transducer from the probe- disc.
Page 17
VX Ultrasonic Velocity Gauge
13
At this point, the VX has successfully calculated it's internal error factor,
and will compensate for this value in any subsequent measurements. When performing a "probe-zero", the VX will always use the sound-velocity
value of the built-in probe- disc, even if some other velocity value has been entered for making actual measurements. Though the VX will remember
the last "probe-zero" performed, it is generally a good idea to perform a "probe-zero" whenever the gauge is turned on, as well as any time a different transducer is used. This will ensure that the instrument is always correctly zeroed.
Calibration
In order for the VX to make accurate measurements, it must be set to the exact thickness the material being measured. Different types of material have different inherent sound-velocities. For ex ample, the velocity of sound through steel is about 0.233 inches- per-microsecond, versus that of aluminum, which is about 0.248 inches- per-microsecond. If the gauge is not set to the correct thickness, all of the measurements the gauge makes will be erron eous.
Page 18
Dakota Ultrasonics
14
Calibration to a known thickness
NOTE: This procedure requires a sample piece of the specific
material to be measured, the exact thickness of which is known, e.g. from having been measured by some other means.
1) Make sure the VX is on.
2) Perform a Probe-Zero (refer to page 12)
3) Apply couplant to the sample piece.
4) Press the transducer against the sample piece, making sure that the transducer sits flat against the surface of the sample. The display sh ould show some (probably incorrect) sound velocity value, and the Stability Indicator should have nearly all its bars on.
5) Having achieved a stable reading, remove the transducer. If the displayed value changes from the value shown while the transducer was coupled, repeat step 4.
7) Press the CAL key. The IN / µs (or M /s) symbol should begin
flashing.
8) Press the CAL key again. The IN (or MM ) symbols should begin
flashing and a thickness value is displayed. Use the UP and DOWN arrow keys to adjust the displayed thickness up or down, until it matches the thickness of the sample piece.
9) Press the CAL key once more to exit the calibration mode. The VX
will now display sound velocities of materials of the same thickness as entered.
Page 19
VX Ultrasonic Velocity Gauge
15
Calibration to a known velo city
NOTE: This procedure requires that the operator know the sound-
velocity of the material to be measured. A table of common materials and their sound-velocities can be found in Appendix C.
1) Place the transducer on the material to be measured and obtain a stable reading.
2) Press the CAL key to enter calibration mode. The IN /µs (or M /s)
symbol will begin flashing.
3) Use the UP and DOWN arrow keys to adjust the displayed velocity
up or down, until it matches the known sound-velocity of the material.
4) Press the CAL key again and the display will show the thickness
value calculated for the material.
4) Press the CAL key once more to exit the calibration mode. The VX is now ready to perform measurements.
NOTE: During the calibration procedure when IN /µs, or M /s is
flashing, pressing the PRB-0 key will restore the gauge to the factory default sound-velocity for steel (.2330 IN / µs or 5920 M /s).
Page 20
Dakota Ultrasonics
16
Scan Mode
While the VX excels at making single point measurements, it is
sometimes desirable to examine a larger region to check for material consistency. The VX includes a feature, called Scan Mode, which allows it
to do just that.
In normal operation, the VX performs and displays four measurements every second, which is quite adequate for single measurements. In Scan Mode, however, the gauge performs sixteen measurements every second, but does not display them. While the transducer is in contact with the
material being measured, the VX is keeping track of the fastest velocity it finds. The transducer may be "scrubbed" across a surface, and any brief interruptions in the signal will be ignored. When the transducer loses contact with the surface for more than a second, the VX will display the fastest sound velocity it found.
When the VX is not in calibration mode, press the UP arrow key to turn Scan Mode on and off. A brief message will appear in the display confirming the operation. While scanning, the display will show a moving series of dashes instead of a sound value.
Page 21
VX Ultrasonic Velocity Gauge
17
TRANSDUCER SELECTION
The VX is inherently capable of performing measurements on a wide range of materials, from various metals to glass and plastics. Different types of material, however, will require the use of different transducers. Choosing the correct transducer for a job is critical to being able to easily perform accurate and reliable measurements. The following paragraphs highlight the important properties of transducers, which should be considered when selecting a transducer for a specific job.
Generally speaking, the best transducer for a job is one that sends sufficient ultrasonic energy into the material being measured such that a strong, stable echo is received by the VX. Several factors affect the strength of ultrasound as it travels. These are outlined below:
Initial Signal Strength
The stronger a signal is to begin with, the stronger its return echo will be. Initial signal strength is largely a factor of the size of the ultrasound emitter in the transducer. A large emitting area will send more energy into the mater ial being measured than a small emitting area. Thus, a so-called "1/2-inch" transducer will emit a stronger signal than a "1/4-inch" transducer.
Absorption and Scattering As ultrasound travels through any material, it is partly absorbed. If
the material through which it travels has any grain structure, the sound waves will also experience scattering. Both of these effects reduce the
Page 22
Dakota Ultrasonics
18
strength of the waves, and thus, the VX's ability to detect the returning echo.
Higher frequency ultrasound is absorbed and scattered more than
ultrasound of a lower frequency. While it may seem that using a lower frequency transducer might be better in every instance, low frequencies are less directional than high frequencies. Thus, a higher frequency transducer would be a better choice for detecting the exact location of small pits or flaws in the material being measured.
Geometry of the Transducer The physical constraints of the measuring environment sometimes
determine a transducer's suitability for a given job. So me transducers may simply be too large to be used in tightly confined areas. Also, the surface area available for contacting with the transducer may be limited, requiring the use of a transducer with a small wearface. Measuring on a curved surface, such as an engine cylinder wall, may require the use of a transducer with a matching curved wearface.
Temperature of the Material When it is necessary to measure on surfaces that are exceedingly
hot, high temperature transducers must be used. These transducers are built using special materials and techniques that allow them to withstand high temperatures without damage. Additionally, care must be taken when performing a "Probe-Zero" or "Calibration to Known Thickness" with a high temperature transducer. Se e Appendix B for more information on measuring materials with a high temperature transducer.
Page 23
VX Ultrasonic Velocity Gauge
19
Selection of the proper transducer is often a matter of tradeoffs between various characteristics. It may be necessary to experiment with a variety of transducers in order to find one that works well for a given job. Dakota Ultrasonics can provide assistance in choosing a transducer, and offers a broad selection of transducers for evaluation in specialized applications.
Page 24
Dakota Ultrasonics
20
Physical
Weight: 10 ounces Size: 2.5W x 4.75H x 1.25D inches (63.5W x 120.7H x 31.8D mm).
Operating Temperature: -20 to 120 °F (-20 to 50 °C) Case: Extruded aluminum body / nickel plated aluminum end caps.
APPENDIX A
Product Specifications
Keypad
Sealed membrane, resistant to water and petroleum products.
Power Source
Two “AA” size, 1.5 volt alkaline or 1.2 volt NiCad cells. 200 hours typical operating time on alkaline, 120 hours on NiCad.
Display
Liquid-Crystal-Display, 4.5 digits, 0.500-inch high numerals. LED backlight.
Measuring
Range: 0.025 to 19.999 inches (0.63 to 500 millimeters) Resolution: 0.001 inch (0.01 millimeter)
Accuracy: ±0.001 inch (0.01 millimeter), depends on material
and conditions
Sound Velocity Range: .0492 to .3930 IN/ µs (1250 to 10000 M/s)
Page 25
VX Ultrasonic Velocity Gauge
21
blank page
Page 26
Dakota Ultrasonics
22
APPENDIX B
Application Notes
Measuring hot surfaces
The velocity of sound through a substance is dependant upon its temperature. As materials heat up, the velocity of sound through them decreases. In most applications with surface temperatures less than about
200°F (100 °C), no special procedures must be observed. At temperatures above this point, the change in sound velocity of the material being measured starts to have a noticeable effect upon ultrasonic measurement.
At such elevated temperatures, it is recommended that the user perform a calibration procedure (refer to page 11) on a sample piece of known thickness, which is at or near the temperature of the material to be
measured. This will allow the VX to correctly calculate the velocity of sound through the hot material.
When performing measurements on hot surfaces, it may also be necessary to use a specially constructed high -temperature transducer. These transducers are built using materials that can withstand high temperatures. Even so, it is recommended that the probe be left in contact with the surface for as short a time as needed to acquire a stable measurement. While the transducer is in contact with a hot surface, it will begin to heat up itself, and through thermal expansion and other effects may begin to adversely affect the accuracy of measurements.
Page 27
VX Ultrasonic Velocity Gauge
23
Measuring laminated materials
Laminated materials are unique in that their density (and therefore sound-velocity) may vary considerably from one piece to another. Some laminated materials may even exhibit noticeable changes in sound-velocity across a single surface. The only way to reliably measure such materials is by performing a calibration procedure on a sample piece of known thickness. Ideally, this sample material should be a pa rt of the same piece being measured, or at least from the same lamination batch. By calibrating to each test piece individually, the effects of variation of sound-velocity will be minimized.
An additional important consideration when measuring laminates, is that any included air gaps or pockets will cause an early reflection of the ultrasound beam. This effect will be noticed as a sudden increase in sound speed in an otherwise regular surface. While this may impede accurate measurement of the material, it does provide the user with positive indication of air gaps in the laminate.
Page 28
Dakota Ultrasonics
24
blank page
Page 29
VX Ultrasonic Velocity Gauge
25
in/us
m/s
Aluminum
0.250
6350
Brass
0.173
4394
Cast Iron
0.180
(apprx)
4572
Copper
0.184
4674
Epoxy resin
0.100
(apprx)
2540
Glass, crown
0.223
5664
Gold
0.128
3251
Iron
0.232
5893
Magnesium
0.228
5791
Nylon
0.102
(apprx)
2591
Platinum
0.156
3962
Polystyrene
0.092
2337
PVC
0.094
2388
Rubber, vulcanized
0.091
2311
Silver
0.142
3607
Steel, stainless
0.223
5664
Teflon
0.056
1422
Titanium
0.240
6096
Zinc
0.166
4216
APPENDIX C
Sound Velocities of some Common Materials
Material sound velocity
Bismuth 0.086 2184 Cadmium 0.109 2769 Constantan 0.206 5232
German silver 0.187 4750 Glass,flint 0.168 4267 Ice 0.157 3988 Lead 0.085 2159 Mercury 0.057 1448
Nickel 0.222 5639 Paraffin 0.087 2210 Plexiglass 0.106 2692 Porcelain 0.230 (apprx) 5842 Quartz glass 0.222 5639
Steel, common 0.233 5918 Stellite 0.275 (apprx) 6985 Tin 0.131 3327 Tungsten 0.210 5334 Water 0.058 1473
Page 30
Dakota Ultrasonics
26
blank page
Page 31
VX Ultrasonic Velocity Gauge
27
WARRANTY INFORMATION
Warranty Statement
Dakota Ultrasonics warrants the VX against defects in materials and workmanship for a period of five years from receipt by the end user. Additionally, Dakota Ultrasonics warrants transducers and accessories against such defects for a period of 90 days from receipt by the end user. If Dakota Ultrasonics receives notice of such defects during the warranty period, Dakota Ultrasonics will either, at its option, repair or replace products that prove to be defective.
Should Dakota Ultrasonics be unable to repair or replace the product within a reasonable amount of time, the customer's alternative exclusive remedy shall be refund of the purchase price upon return of the product.
Exclusions
The above warranty shall not apply to defects resulting from: improper or inadequate maintenance by the customer; unauthorized modification or misuse; or operation outside the environmental specifications for the product.
Dakota Ultrasonics makes no other warranty, either express or implied, with respect to this product. Dakota Ultrasonics specifically disclaims any implied warranties of merchantability or fitness for a particular purpose. Some states or provinces do not allow limitations on the duration of an implied warranty, so the above limitation or exclusion may not apply to you. However, any implied warranty of merchantability or fitness is limited to the five-year duration of this written warranty.
This warranty gives you specific legal rights, and you may also have other rights, which may vary from state to state or province to province.
Obtaining Service During Warranty Period
If your hardware should fail during the warranty period, contact Dakota Ultrasonics and arrange for servicing of the product. Retain proof of purchase in order to obtain warranty service.
For products that require servicing, Dakota Ultrasonics may use one of the following methods:
- Repair the product
- Replace the product with a re-manufactured unit
- Replace the product with a product of equal or greater performance
- Refund the purchase price.
After the Warranty Period
If your hardware should fail after the warranty period, contact Dakota Ultrasonics for details of the services available, and to arrange for non-warranty service.
Page 32
Dakota Ultrasonics
28
blank page
Page 33
MATERIAL SAFETY DATA SHEET
N/A = not applicable or not available (To comply with 29 CFR 1910.1200)
SECTION 1 – PRODUCT IDENTIFICATION
Product Name: SOUNDSAFE Generic Name: Ultrasonic Couplant Manufacturer: Sonotech, Inc. 774 Marine Dr., Bellingham, WA 98225 (360) 671-9121
SECTION 2 – HAZARDOUS INGREDIENTS
This material does not contain any ingredients having
known health hazards in concentrations greater than 1%.
This material does not contain any known or suspected
carcinogens.
SECTION 4 – FIRE AND EXPLOSION
HAZARD DATA
Flash Point : none Upper Exposure Limit: none Lower Exposure Limit: none Special Fire Fighting Procedures : N/A Extinguishing media: N/A Unusual Fire and Explosion Hazards : none
NFPA Hazardous Materials
Identification System (est)
Health……………………0 Flammability…………….0 Reactivity………………..0
SECTION 3 – PHYSICAL DATA
(nominal)
Boiling Point: >220 °F pH: 7.35 – 7.9 Freezing Point: <20°F Acoustic Imp.: 1.726x10 Vapor Pressure: N/A Vapor Density: N/A Evaporation Rate: N/A Specific Gravity: >1.02 Solubility in Water: complete Appearance and Odor: water white, opaque gel; bland odor
6
SECTION 5 – REACTIVITY DATA
Stability: Stable Conditions to Avoid: none Incompatibility (Materials to Avoid): none known Hazardous Polymerization: will not occur Hazardous Decomposition or Byproducts: none known
SECTION 6 – HEALTH HAZARD AND FIRST AID DATA
Routes of Entry : Skin: not likely Ingestion: not normally Eyes: not normally Inhalation: no Effects of Overexposure: Acute: May cause temporary
1
eye irritation
Chronic: none expected
SECTION 7 – STORAGE AND HANDLING
INFORMATION
Precautions to be taken in handling and storage: Store between 20°F and 120 °F. Spills are slippery and should
be cleaned up immediately. Steps to be taken in case material is released or spilled: Pick up excess for disposal. Clean with water. Waste disposal method: Dispose of in accordance with federal, state, and local regulations.
1
SOUNDSAFE contains only food grade and cosmetic grade ingredients.
SONOTECH, INC.
Toll Free: 1-800-458-4254
774 Marine Dr., Bellingham, WA 98225
Telephone: (360) 671-9121
First Aid Procedures : Skin: Remove with water if desired. Eyes: Flush with water for 15 minutes. Ingestion: For large quantities, induce vomiting and
call a physician.
Inhalation: N/A
SECTION 8 – CONTROL MEASURES
Respiratory Protection: not required Ventilation: not required Protective Gloves : on individuals demonstrating
sensitivity to SOUNDSAFE Eye Protection: as required by working conditions Other Protective Equipment : not required
Fax: (360) 671-9024
Page 34
Loading...