Dakota PX-7 User Manual

Page 1
OPERATION MANUAL
-
129
-
0002
Rev
1.60
January
200
8
DAKOTA ULTRASONICS MODEL PX-7 PRECISION THICKNESS GAUGE
Page 2
Copyright 2008 Dakota Ultrasonics. All rights reserved.
No part of this publication may be reproduced, translated into another language, stored in a retrieval system, or transmitted in any form or by any means; electronic, mechanical, photocopying, recording, or otherwise, without the prior written consent of Dakota Ultrasonics.
Every precaution has been taken in the preparation of this publication. Dakota Ultrasonics assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from the use of information contained herein.
Any brand or product names mentioned herein are used for identification purposes only, and are trademarks or registered trademarks of their respective holders.
1500 Green Hills Road, #107 Scotts Valley, CA 95066 USA
Tel (831) 431-9722 Fax (831) 431-9723
www.dakotaultrasonics.com
Page 3
CONTENTS
PX-7 Ultrasonic Thickness Gauge
INTRODU CTION OPERATION
THE KEYPAD THE DISPLAY THE TRANSDUCER MAKING MEASUREMENTS CONDITION AND PREPARATION OF SURFACES SELECTING MEASUREMENT MODE & THE LOCK MODE CALIBRATION
SCAN MODE, ALARM MODE, DIFF MODE, RS232 PORT TRANSDUCER SELECTION APPENDIX A: PRODUCT SPECIFICATIONS APPENDIX B: APPLICATION NOTES
1 3 3 7
9 10 12 13 14 17 22 25 27
APPENDIX C: SOUND VELOCITIES OF COMMON MATERIALS WARRANTY INFORMATION
29 31
DISCLAIMER
Inherent in ultrasonic thickness measurement is the possibility that the instrument will use the second rather than the first echo from the back surface of the material being measured. This may result in a thickness reading that is TWICE what it should be. Responsibility for proper use of the instrument and recognition of this phenomenon rests solely with the user of the instrument.
Page 4
Dakota Ultrasonics
blank page
Page 5
PX-7 Ultrasonic Thickness Gauge
1
INTRODUCTION
The Dakota Ultrasonics model PX-7 is a precision Ultrasonic Micrometer. Based on the same operating principles as SONAR, the PX-7 is capable of measuring the thickness of various materials with accuracy as
high as ± 0.0001 inches, or ± 0.001 millimeters. The principle advantage of ultrasonic measurement over traditional methods is that ultrasonic measurements can be performed with access to only one side of the material being measured.
This manual is presented in three sections. The first section covers operation of the PX- 7, and explains the keypad controls and display. The second section provides guidelines in selecting a transducer for a specific application. The last section provides application notes and a table of sound velocity values for various materials.
Dakota Ultrasonics maintains a customer support resource in order to assist users with questions or difficulties not covered in this manual. Customer support may be reached at any of the following:
Dakota Ultrasonics, 1500 Green Hills Road, #107 Scotts Valley, CA 95066 USA
Telephone: (831) 431- 9722
Facsimile: (831) 431-9723
www.dakotaultrasonics.com
Page 6
Dakota Ultrasonics
2
blank page
Page 7
PX-7 Ultrasonic Thickness Gauge
3
OPERATION
The PX-7 interacts with the operator through the membrane keypad and the LCD display. The functions of the various keys on the keypad are detailed below, followed by an explanation of the display and its various symbols.
The Keypad
This key is used to turn the PX-7 on and off. When the tool is turned ON, it will first perform a brief display test by illuminating all of the segments in the display. After one second, the tool will display the internal software version number. After displaying the version number, the display will show "0.0000" (or "0.000" if using metric units), indicating the tool is ready for use.
The PX-7 is turned OFF by pressing the ON/OFF key. The tool has a special memory that retains all of its settings even when the power is off. The tool also features an auto-powerdown mode designed to conserve battery life. If the tool is idle for 5 minutes, it will turn itself off.
Page 8
Dakota Ultrasonics
4
The CAL key is used to enter and exit the PX-7's calibration mode. This mode is used to adjust the sound-velocity value that the PX-7 will use when calculating thickness. The tool will either calculate the sound-velocity from a sample of the material being measured, or allow a known velocity value to be entered directly. Refer to page 14 for an explanation of the two CAL functions available.
The IN/MM key is used to switch back and forth between English and metric units. This key may be used at any time, whether the tool is
displaying a thickness (IN or MM) or a velocity value (IN/µs or M/s).
The UP arrow key has two functions. When the PX-7 is in calibration mode, this key is used to increase numeric values on the display. An auto­repeat function is built in, so that when the key is held down, numeric
values will increment at an increasing rate. When the PX-7 is in measurement mode, this key is used to select between three different
modes of measurement ( echo-echo, interface-echo, or auto ). Refer to page 14 for an explanation of the calibration procedure and page 13 to review the various measurement modes.
Page 9
PX-7 Ultrasonic Thickness Gauge
5
The SCAN key switches the SCAN measurement mode on and off. Refer to page 17 for an explanation of the SCAN measurement mode.
The DOWN arrow key has three functions. When the PX-7 is in the CAL mode, this key is used to decrease numeric values on the display. An auto-repeat function is built in, so that when the key is held down, numeric
values will decrement at an increasing rate. When the PX- 7 is not in calibration mode, the DOWN arrow key switches the display backlight between three available settings. OFF will be displayed when the backlight is switched off. AUTO will be displayed when the backlight is set to automatic mode, and ON will be displayed when the ba cklight is set to stay
on. In the AUTO setting, the backlight will illuminate while the PX-7 is making a measurement and turn off after several seconds to conserve battery life. Finally, if the DOWN arrow key is pressed and held down while
powering the PX-7 on, a LOC mode will be toggled on/off.
Page 10
Dakota Ultrasonics
6
The ALRM key has two functions. By holding down the ALRM key when powering up the PX- 7, the audible beeper will be turned on or off
accordingly. After the unit has been turned on, pressing the ALRM key will toggle the alarm mode to the on/off positions and allow the user to enter a nominal thickness value. Refer to page 18 for details on how to use the alarm feature.
The DIFF key is used to enable or disable the Differential mode and allow the user to enter a nominal value as a comparison against actual thickness measurements. Refer to page 19 for a complete explanation of this feature.
The SEND key is used to send the cu rrently displayed thickness measurement to an external storage device via the PX-7's RS232 port.
Refer to page 20 for an explanation of how to use this feature.
Page 11
PX-7 Ultrasonic Thickness Gauge
7
The Display
The numeric portion of the displ ay consists of 4 complete digits preceded by a leading "1", and is used to display numeric values, as well as occasional simple words, to indicate the status of various settings. When the PX-7 is displaying thickness measurements, the display will hold the last value measured, until a new measurement is made. Additionally, when the battery voltage is low, the entire display will begin to flash. When this occurs, the batteries should be replaced.
These eight vertical bars form the Stability Indicator. When the PX-7 is idle, only the left -most bar and the underline will be on. While the gauge is
taking a measurement, six or seven of the bars should be on. If fewer than five bars are on, the PX-7 is having difficulty achieving a stable
measurement, and the thickness value displayed will most likely be erroneous.
Page 12
Dakota Ultrasonics
8
When the IN symbol is on, the PX-7 is displaying a thickness value in inches. The maximum thickness that can be displayed is 1.0000 inches.
When the MM symbol is on, the PX- 7 is displaying a thickness value in millimeters. If the displayed thickness exceeds 20.00 millimeters, the decimal point will shift automatically to the right, allowing values up to 25.40 millimeters to be displayed.
When the IN symbol is on, in conjunction with the /µs symbol, the PX-7 is displaying a sound-velocity value in inches- per-microsecond.
When the M symbol is on, in conjunction with the /s symbol, the PX- 7 is displaying a sound-velocity value in meters-per-second.
Page 13
PX-7 Ultrasonic Thickness Gauge
9
The Transducer
The transducer is the "busi ness end" of the PX-7. It transmits and receives ultrasonic sound waves that the PX-7 uses to calculate the thickness of the material being measured. The transducer connects to the PX-7 via the attached cable, and one coaxial connector.
The transducer must be used correctly in order for the PX-7 to produce accurate, reliable measurements. Below is a short description of the transducer, followed by instructions for its use.
The diagram above is a side view of a typical single element delay line transducer. This delay line is fastened to the transducer with a retainer ring. A drop of couplant is applied between the delay line and transducer body. The transducer body contains one crystal element that is responsible for conducting ultrasonic sound into and back from the material being measured. When the transducer is placed against the material being measured, it is the area directly beneath the delay line that is being measured.
Page 14
Dakota Ultrasonics
10
When measuring, press against the top of the transducer with the thumb or index finger to hold the transducer in place. Moderate pressure is sufficient, as it is only necessary to keep the transducer stationary, and the delay line seated flat against the surface of the material being measured.
Making Measurements
In order for the transducer to do its job, there must be no air gaps between the wear-face and the surface of the material being measured. This is accomplished with the use of a "coupling" fluid, commonly called "couplant". This fluid serves to "couple", or transfer , the ultrasonic sound waves from the transducer, into the material, and back again. Before attempting to make a measurement, a small amount of couplant should be applied to the surface of the material being measured. Typically, a single droplet of coupl ant is sufficient.
After applying couplant, press the transducer (wearface down) firmly against the area to be measured. The Stability Indicator should have six or seven bars darkened, and a number should appear in the display. If the PX-7 has been set to the correct sound velocity (see page 14), the number
in the display will indicate the actual thickness of the material directly beneath the transducer.
Page 15
PX-7 Ultrasonic Thickness Gauge
11
If the Stability Indicator has fewer than five bars darkened, or the numbers on the display seem erratic, first check to make sure that there is an adequate film of couplant beneath the transducer, and that the transducer is seated flat against the material. If the condition persists, it may be necessary to select a different transducer (size or frequency) for the material being measured. See page 22 for information on transducer selection.
While the transducer is in contact with the material that is being measured, the PX-7 will perform four measurements every second, updating its display as it does so. When the transducer is removed from the surface, the display will hold the last measurement made.
IMPORTANT
Occasionally, a small film of couplant will be drawn out between the transducer and the surface as the transducer is removed. When this happens, the PX-7 may perform a measurement through this couplant film, resulting in a measurement that is larger or smaller than it should be. This phenomenon is obvious when one thickness value is observed while the transducer is in place, and another value is observed after the transducer is removed.
Page 16
Dakota Ultrasonics
12
Condition and Preparation of Surfaces
In any ultrasonic measurement scenario, the shape and roughness of the test surface are of paramount importance. Rough, uneven surfaces may limit the penetration of ultrasound through the material, and result in unstable, and therefore unreliable, measurements. The surface being measured should be clean, and free of any small particu late matter, rust, or scale. The presence of such obstructions will prevent the transducer from seating properly against the surface. Often, a wire brush or scraper will be helpful in cleaning surfaces. In more extreme cases, rotary sanders or grinding wheels may be used, though care must be taken to prevent surface gouging, which will inhibit proper transducer coupling.
Extremely rough surfaces, such as the pebble-like finish of some cast iron, will prove most difficult to measure. These kinds of surfa ces act on the sound beam like frosted glass on light:, the beam becomes diffused and scattered in all directions.
In addition to posing obstacles to measurement, rough surfaces contribute to excessive wear of the transducer, particularly in situations where the transducer is "scrubbed" along the surface.
Page 17
PX-7 Ultrasonic Thickness Gauge
13
Selecting Measurement Mode
The PX-7 is equipped with four measurement mode options (echo­echo, interface- echo, auto, or plas). In echo-echo mode, the gauge has the ability to read thin metals dow n to .006 inches (.15 millimeters). The echo­echo mode also allows the user to measure the thickness of metals that have been previously coated or painted on the surface. This enables the user to determine the thickness of the metal without having to remove the paint. In interface-echo mode, the gauge has the ability to read plastics and thicker materials. The PX-7's auto mode will automatically switch between echo-echo and interface-echo according to the different materials being measured. Finally, the plas mod can be used to measure thin plastics using a special graphite tip. The following section outlines how to switch between measurement modes:
Measurement Modes
1) Make sure the PX-7 is on.
2) Press the UP arrow key to toggle between measurement m odes. The gauge will display GatE followed by I-E, E-E, Auto, or PLAS depending on which mode the PX-7 was last used in.
3) Repeat step 2 until the desired mode has been displayed.
Note: Once the mode has been selected, the user can use the built-in
LOC feature to prevent additional users from accidentally changing
the mode. When the PX-7 is in LOC mode, all features (with the exception of the calibration and units), are locked and cannot be
Page 18
Dakota Ultrasonics
14
altered by the user. Press and hold the DOWN arrow key while
powering up the PX-7 to toggle the LOC feature on/off. Repeat procedure to toggle on/off status.
Calibration
In order for the PX- 7 to make accurate measurements, it must be set to the correct sound-velocity for the material being measured. Different types of material have different inherent sound-velocities. For example, the velocity of sound through steel is about 0.233 inches-per-microsecond, versus that of aluminum, which is about 0.248 inches-per-microsecond. If the tool is not set to the correct sound-velocity, all of the measurements the tool makes will be erroneous by some fixed percentage. The one point calibration is the simplest and most commonly used calibration procedure -
optimizing linearity over large ranges. The PX-7 provides two simple methods for setting the sound-velocity, described in the following pages.
Page 19
PX-7 Ultrasonic Thickness Gauge
15
Calibration to a known thickness
NOTE: This procedure requires a sample piece of the specific
material to be measured, the exact thickness of which is known, e.g. from having been measured by some other means.
1) Make sure the PX-7 is on.
2) Apply couplant to the sample piece.
3) Press the transducer against the sample piece, making sure that the transducer sits flat against the surface of the sample. The display sh ould show some (probably incorrect) thickness value, and the Stability Indicator should have nearly all its bars on.
4) Having achieved a stable reading, remove the transducer. If the displayed thickness changes from the value shown while the transducer was coupled, repeat step 3.
5) Press the CAL key. The IN (or MM ) symbol should begin flashing.
6) Use the UP and DOWN arrow keys to adjust the displayed thickness up or down, until it matches the thickness of the sample piece.
7) Press the CAL key again. The IN/µs (or M/s) symbols should begin flashing. The PX-7 is now displaying the sound velocity
value it has calculated based on the thickness value that was entered in step 6.
8) Press the CAL key once more to exit the calibration mode. The PX-7 is now ready to perform measurements.
Page 20
Dakota Ultrasonics
16
Calibration to a known velocity
NOTE: This procedure requires that the operator know the sound-
velocity of the material to be measured. A table of common materials and their sound-velocities can be found in Appendix C.
1) Make sure the PX-7 is on.
2) Press the CAL key to enter calibration mode. If the IN (or MM) symbol is flashing, press the CAL key again, so that the IN/µs (or M/s) symbols are flashing.
3) Use the UP and DOWN arrow keys to adjust the displayed velocity up or down, until it matches the sound-velocity of the material to be measured.
4) Press the CAL key once more to exit the calibration mode. The PX-7 is now ready to perform measurements.
NOTE: At any time during the calibration procedure (IN, MM, IN/ µs, or
M/s flashing in the display), pressing the SCAN key will restore the tool to the factory default sound- velocity for steel (0.233 IN/µs).
To achieve the most accurate measurements possible, it is generally
advisable to always calibrate the PX-7 to a sample piece of known thickness. Material composition (and thus, its sound-velocity) sometimes varies from lot to lot and from manufacturer to manufacturer. Calibration to a sample of known thickness will ensure that the tool is set as closely as possible to the sound velocity of the material to be measured.
Page 21
PX-7 Ultrasonic Thickness Gauge
17
PX-7 Modes
Scan Mode
While the PX-7 excels at making single point measurements, it is
sometimes desirable to examine a larger region, searching for the thinnest point. The PX-7 includes a feature, called Scan Mode, which allows it to do
just that.
In normal operation, the PX-7 performs and displays four measurements every second, which is quite adequate for single measurements. In Scan Mode, however, the tool performs eight measurements every second, but does not display them. While the transducer is in contact with the material being measured, the PX-7 is keeping track of the lowest measurement it finds. The transducer may be "scrubbed" across a surface, and any brief interruptions in the signal will be ignored. When the transducer loses contact with the surface for more than
a second, the PX-7 will display the smallest measurement it found.
When the PX-7 is not in calibration mode, press the SCAN key to turn Scan Mode on and off. A brief message will appear in the display
confirming the operation. While scanning, the display will show a moving series of dashes instead of a thickness value. When the transducer is removed from the material being scanned, the PX-7 will (after a brief pause) display the smallest measurement it found.
Page 22
Dakota Ultrasonics
18
Alarm Mode
The Alarm Mode feature of the PX-7 allows the user to set an audible and visual Hi / Lo parameter when taking measurements. If the measurement falls below or above the Hi / Lo limits, set by the user, a red light will be illuminated on the front panel of the gauge and the beeper sounded. This improves the speed and efficiency of the inspection process by eliminating constant viewing of the actual reading displayed. The following pages outline how to enable and set up this feature:
Using the Beeper
1) While the unit is off, press and hold down the ALRM key.
2) Press ON/OFF key to power up the unit.
3) Release the ALRM key - BEEP ON or BEEP OFF will be displayed enabling or disabling the beeper.
4) Repeat step s 1 through 3 to toggle between BEEP ON or BEEP OFF.
Page 23
PX-7 Ultrasonic Thickness Gauge
19
Alarm Mode
1) Press ON/OFF key to power up the PX-7.
2) Press the ALRM key. ALAr OFF, or ALAr followed by a thickness value and flashing IN (or MM ) symbol will be displayed - indicating the alarm mode is enabled.
3) Repeat step 2 to toggle between ALAr OFF or ALAr value and flashing IN / MM symbol (On).
4) Assuming ALAr value flashing IN is displayed, use the UP and DOWN arrow keys to scroll to the desired Lo alarm value.
5) Press the SEND key to select the Lo value entered. The Hi value will now be displayed. Repeat step 4 select the Hi alarm value.
6) The PX-7 is now ready to perform measurements using the Alarm feature.
Differential Mode
In the Quality Control environment, it is sometimes necessary to know the difference between a nominal (target) thickness value and an actual thickness value. This feature is also included in the PX-7. With the Differential Mode enabled, the PX-7 will display the positive or negative difference from an entered nominal value. The following steps outline the procedure for setting up this feature:
Page 24
Dakota Ultrasonics
20
Differential Mode
1) Press ON/OFF key to power up the PX-7.
2) Press the DIFF key. DIFF OFF or DIFF followed by a thickness value and flashing IN (or MM ) will be displayed.
3) Repeat step 2 to toggle between DIFF OFF or DIFF value and flashing IN.
4) Assuming DIFF value flashing IN is displayed, use the UP and DOWN arrow keys to scroll to the desired nominal thickness value.
5) Press the SEND key once again to select the nominal value entered.
6) The PX-7 is now ready to perform measurements using the Differential feature.
RS232 Serial Port
The PX-7 is equipped with an RS232 serial port. Using the accessory cable (part# N-306- 0010), the PX-7 has the ability to connect to a computer, or ex ternal storage device. The following section outlines the procedure for connecting the PX- 7 to a computer, and how to collect data using any standard communications program:
Page 25
PX-7 Ultrasonic Thickness Gauge
21
Connecting To a Computer
1) Connect the accessory cable (part# N-306- 0010) to the 2 pin jack located on the bottom of the PX-7, and the 9 pin connector to a
serial port on the computer.
2) Start the communications software that will be used to collect the measurements (i.e. Windows 95/98 - HyperTerminal).
3) Setup the communications software using the following parameters: Data Bits - 8, Parity - None, Stop Bits - 1, Baud Rate 1200.
4) Set the communications software COMM port to the port number that the PX-7 is connected.
5) After taking a measurement, press the SEND key to send the measurement to the computer. The measurement will be displayed on the computer screen.
NOTE: Communications software packages generally have the ability
to capture the screen data to a common text file. This text file, containing the measurements, can then be imported into any common spreadsheet program (i.e. Excel, Quattro Pro, Lotus123) for further reporting requirements.
Page 26
Dakota Ultrasonics
22
TRANSDUCER SELECTION
The PX-7 is inherently capable of performing measurements on a wide range of materials, from various metals to glass and plastics. Different types of material, however, will require the use of different transducers. Choosing the correct transducer for a job is critical to being able to easily perform accurate and reliable measurements. The following paragraphs highlight the important properties of transducers, which should be considered when selecting a transducer for a specific job.
Generally speaking, the best transducer for a job is one that sends sufficient ultrasonic energy into the material being measured such that a strong, stable echo is received by the PX-7. Several factors affect the
strength of ultrasound as it travels. These are outlined below:
Initial Signal Strength
The stronger a signal is to begin with, the stronger its return echo will be. Initial signal strength is largely a factor of the size of the ultrasound emitter in the transducer. A large emitting area will send more energy into the material being measured than a small emitting area. Thus, a so-called "1/4-inch" transducer will emit a stronger signal than a "1/8-inch" transducer.
Absorption and Scattering As ultrasound travels through any material, it is partly absorbed. If
the material through which the sound travels has any grain structure, the sound waves will experience scattering. Both of these effects reduce the strength of the waves, and thus, the PX-7's ability to detect the returning echo.
Page 27
PX-7 Ultrasonic Thickness Gauge
23
Higher frequency ultrasound is absorbed and scattered more than
ultrasound of a lower frequency. While it may seem that using a lower frequency transducer might be better in every instance, low frequencies are less directional than high frequencies.
Geometry of the Transducer The physical constraints of the measuring environment sometimes
determine a transducer's suitability for a given job. Some transducers may simply be too large to be used in tightly confined areas. Also, the surface area available for contacting with the transducer may be limited, requiring the use of a transducer with a small cone tipped delay line. Measuring on a curved surface, may require the use of a transducer with a matching curved wearface.
Temperature of the Material When it is necessary to measure on surfaces that are exceedingly
hot, special delay lines may be neccessary. Additionally, care must be taken when performing a "Calibration to Known Thickness" with a high temperature application. See Appendix B for more information on measuring materials with a high temperatures.
Selection of the proper transducer is often a matter of tradeoffs between various characteristics. It may be necessary to experiment with a variety of transducers in order to find one that works well for a given job. Dakota Ultrasonics can provide assistance in choosing a transducer, and offers a broad selection of transducers for evaluation in specialized applications.
Page 28
Dakota Ultrasonics
24
blank page
Page 29
25
Physical
Weight: 10 ounces(with batteries). Size: 2.5W x 4.5 H x 1.24 D inches (63.5W x 114.3 H x 31.5 D mm). Operating Temperature: -20 to 120 °F ( -30 to 50 °C) Case: Extruded aluminum body / nickel plated aluminum end caps.
PX-7 Ultrasonic Thickness Gauge
APPENDIX A
Product Specifications
Keypad
Sealed membrane, resistant to water and petroleum products.
Power Source
Two “AA” size, 1.5 volt alkaline or 1.2 volt NiCad cells. 150 hours typical op erating time on alkaline, 100 hours on NiCad.
Display
Liquid-Crystal-Display, 4.5 digits, 0.500 inch high numerals. LED backlight.
Measuring
Range: 0.0060 to 1.0000 inches (0.15 to 25.40 millimeters) -Steel Resolution: 0.0001 inch (0.001 millimeter) Accuracy: ±0.0001 inch (0.001 millimeter), depends on material
and conditions
Sound Velocity Range: 0.0492 to 0.3937 in/µs (1250 to 10000m/s)
Page 30
Dakota Ultrasonics
26
blank page
Page 31
PX-7 Ultrasonic Thickness Gauge
27
APPENDIX B
Application Notes
Measuring tubing
When measuring a piece of tubing for wall thickness, it may pr ove beneficial to have multiple delay lines with different radiuses for different tubing diameters. The delay lines can be easily radiused by placing a piece of emery cloth around the tubing and moving the transducer back and forth until a radius has formed on the tip of the delay line.
Measuring hot surfaces
The velocity of sound through a substance is dependant upon its
temperature. As materials heat up, the velocity of sound through them decreases. In most applications with surface temperatures less than about
200°F (100 °C), no special procedures must be observed. At temperatures above this point, the change in sound velocity of the material being measured starts to have a noticeable effect upon ultrasonic measurement.
At such elevated temperatures, it is recommended that the user perform
a calibration procedure (refer to page 14) on a sample piece of known thickness, which is at or near the temperature of the material to be measured. This will allow the PX-7 to correctly calculate the velocity of sound through the hot material.
When performing measurements on hot surfaces, it may also be
necessary to use a specially constructed high -temperature delay line. It is recommended that the probe be left in contact with the surface for as short
Page 32
Dakota Ultrasonics
28
a time as needed to acquire a stable measurement. While the transducer is in contact with a hot surface, it will begin to heat up, and through thermal expansion and other effects, may begin to adversely affect the accuracy of measurements.
Measuring laminated materials
Laminated materials are unique in that their density (and therefore
sound-velocity) may vary considerably from one piece to another. Some laminated materials may even exhibit noticeable changes in sound-velocity across a single surface. The only way to reliably measure such materials is by performing a calibration procedure on a sample piece of known thickness. Ideally, this sample material should be a part of the same piece being measured, or at least from the same lamination batch. By calibrating to each test piece individually, the effects of variation of sound-velocity will be minimized.
An additional important consideration when measuring laminates, is that
any included air gaps or pockets will cause an early reflection of the ultrasound beam. This effect will be noticed as a sudden decrease in thickness in an otherwise regular surface. While this may impede accurate measurement of total material thickness, it does provide the user with positive indication of air gaps in the laminate.
Page 33
PX-7 Ultrasonic Thickness Gauge
29
in/us
m/s
Aluminum
0.250
6350
Brass
0.173
4394
Cast Iron
0.180
(apprx)
4572
Copper
0.184
4674
Epoxy resin
0.100
(apprx)
2540
Glass, crown
0.223
5664
Gold
0.128
3251
Iron
0.232
5893
Magnesium
0.228
5791
Nylon
0.102
(apprx)
2591
Platinum
0.156
3962
Polystyrene
0.092
2337
PVC
0.094
2388
Rubber, vulcanized
0.091
2311
Silver
0.142
3607
Steel, stainless
0.223
5664
Teflon
0.056
1422
Titanium
0.240
6096
Zinc
0.166
4216
APPENDIX C
Sound Velocities of some Common Materials
Material sound velocity
Bismuth 0.086 2184 Cadmium 0.109 2769 Constantan 0.206 5232
German silver 0.187 4750 Glass,flint 0.168 4267 Ice 0.157 3988 Lead 0.085 2159 Mercury 0.057 1448
Nickel 0.222 5639 Paraffin 0.087 2210 Plexiglass 0.106 2692 Porcelain 0.230 (apprx) 5842 Quartz glass 0.222 5639
Steel, common 0.233 5918 Stellite 0.275 (apprx) 6985 Tin 0.131 3327 Tungsten 0.210 5334 Water 0.058 1473
Page 34
Dakota Ultrasonics
30
blank page
Page 35
PX-7 Ultrasonic Thickness Gauge
31
WARRANTY INFORMATION
Warranty Statement
Dakota Ultrasonics warrants the PX-7 against defects in materials and workmanship
for a period of five years from receipt by the end user. Additionally, Dakota Ultrasonics warrants transducers and accessories against such defects for a period of 90 days from receipt by the end user. If Dakota Ultrasonics receives notice of such defects during the warranty period, Dakota Ultrasonics will either, at its option, repair or replace products that prove to be defective.
Should Dakota Ultrasonics be unable to repair or replace the product within a
reasonable amount of time, the customer's alternative exclu sive remedy shall be refund of the purchase price upon return of the product.
Exclusions
The above warranty shall not apply to defects resulting from: improper or
inadequate maintenance by the customer; unauthorized modification or misuse; or operation outside the environmental specifications for the product.
Dakota Ultrasonics makes no other warranty, either express or implied, with respect
to this product. Dakota Ultrasonics specifically disclaims any implied warranties of merchantability or fitn ess for a particular purpose. Some states or provinces do not allow limitations on the duration of an implied warranty, so the above limitation or exclusion may not apply to you. However, any implied warranty of merchantability or fitness is limited to the five-year duration of this written warranty.
This warranty gives you specific legal rights, and you may also have other rights
which may vary from state to state or province to province.
Obtaining Service During Warranty Period
If your hardware should fail during the warranty period, contact Dakota Ultrasonics
and arrange for servicing of the product. Retain proof of purchase in order to obtain warranty service.
For products that require servicing, Dakota Ultrasonics may use one of the following
methods:
- Repair the product
- Replace the product with a re-manufactured unit
- Replace the product with a product of equal or greater performance
- Refund the purchase price.
After the Warranty Period
If your hardware should fail after the warranty period, contact Dakota Ultrasonics for
details of the services available, and to arrange for non-warranty service.
Page 36
MATERIAL SAFETY DATA SHEET
N/A = not applicable or not available (To comply with 29 CFR 1910.1200)
SECTION 1 – PRODUCT IDENTIFICATION
Product Name: SOUNDSAFE Generic Name: Ultrasonic Couplant Manufacturer: Sonotech, Inc. 774 Marine Dr., Bellingham, WA 98225 (360) 671-9121
SECTION 2 – HAZARDOUS INGREDIENTS
This material does not contain any ingredients having
known health hazards in concentrations greater than 1%.
This material does not contain any known or suspected
carcinogens.
SECTION 4 – FIRE AND EXPLOSION
HAZARD DATA
Flash Point : none Upper Exposure Limit: none Lower Exposure Limit: none Special Fire Fighting Procedures : N/A Extinguishing media: N/A Unusual Fire and Explosion Hazards : none
NFPA Hazardous Materials
Identification System (est)
Health……………………0 Flammability…………….0 Reactivity………………..0
SECTION 3 – PHYSICAL DATA
(nominal)
Boiling Point: >220 °F pH: 7.35 – 7.9 Freezing Point: <20°F Acoustic Imp.: 1.726x10 Vapor Pressure: N/A Vapor Density: N/A Evaporation Rate: N/A Specific Gravity: >1.02 Solubility in Water: complete Appearance and Odor: water white, opaque gel; bland odor
6
SECTION 5 – REACTIVITY DATA
Stability: Stable Conditions to Avoid: none Incompatibility (Materials to Avoid): none known Hazardous Polymerization: will not occur Hazardous Decomposition or Byproducts: none known
SECTION 6 – HEALTH HAZARD AND FIRST AID DATA
Routes of Entry : Skin: not likely Ingestion: not normally Eyes: not normally Inhalation: no Effects of Overexposure: Acute: May cause temporary
1
eye irritation
Chronic: none expected
SECTION 7 – STORAGE AND HANDLING
INFORMATION
Precautions to be taken in handling and storage: Store between 20°F and 120 °F. Spills are slippery and should
be cleaned up immediately. Steps to be taken in case material is released or spilled: Pick up excess for disposal. Clean with water. Waste disposal method: Dispose of in accordance with federal, state, and local regulations.
1
SOUNDSAFE contains only food grade and cosmetic grade ingredients.
SONOTECH, INC.
Toll Free: 1-800-458-4254
774 Marine Dr., Bellingham, WA 98225
Telephone: (360) 671-9121
First Aid Procedures : Skin: Remove with water if desired. Eyes: Flush with water for 15 minutes. Ingestion: For large quantities, induce vomiting and
call a physician.
Inhalation: N/A
SECTION 8 – CONTROL MEASURES
Respiratory Protection: not required Ventilation: not required Protective Gloves : on individuals demonstrating
sensitivity to SOUNDSAFE Eye Protection: as required by working conditions Other Protective Equipment : not required
Fax: (360) 671-9024
Loading...