Dakota PVX User Manual

Page 1
OPERATION MANUAL
DAKOTA ULTRASONICS
PPVVXX
Precision A / B Scan Thickness Gauge
P/N P-157-0002 Rev 1. 10, January 2008
Page 2
Page 3
CHAPTER ONE INTRODUCTION.......................................................................1
CHAPTER TWO QUICK STARTUP GUIDE........................................................2
CHAPTER THREE KEYBOA RD, MENU, & CONNECTOR REFERENCE.......12
CHAPTER FOUR PRINCIPALS OF ULTRASONIC ME ASUREMENT .............22
CHAPTER FIVE SELECTING THE MEASUREMENT MODE..........................27
CHAPTER SIX MAKING MEASURE MENTS.....................................................30
CHAPTER SEVEN USING THE A-SCAN & B-SCAN DISPLAYS.....................42
CHAPTER EIGHT THRU PAINT MEASUREMENT TECHNIQUE....................68
CHAPTER NINE ADDITIO NAL FEATURES OF THE PVX...............................71
CHAPTER TEN DATA STORAGE – SETUP, EDIT, & VIEW GRIDS...............81
CHAPTER ELEVEN SETUPS – CREATE, STORE, EDIT, & RECALL............99
CHAPTER TWELVE USING THE UTILITY SOFTWARE................................106
APPENDIX A - VELOCITY TABLE ...................................................................107
APPENDIX B - SETUP LIBRARY.....................................................................109
Page 4
Page 5
CHAPTER ONE INTRODUCTION
The Dakota Ultrasonics model PVX is a visual precision A / B scan ultrasonic thickness gauge. Based on the same operating principles as SONAR, the PVX is capable of measuring the thickness of various materials with accuracy as high as ±
0.0001 inches, or ± 0.001 millimeters. The principle advantage of ultrasonic measurement over traditional methods is that ultrasonic measurements can be
performed with access to only one side of the material being measured. Dakota Ultrasonics maintains a customer support resource in order to assist users
with questions or difficulties not covered in this manual. Customer support may be reached at any of the following:
Dakota Ultrasonics Corporation, 1500 Green Hills Road, #107 Scotts Valley, CA 95066 USA
Telephone: (831) 431- 9722
Facsimile: (831) 431-9723
www.dakotaultrasonics.com
1.1 Disclaimer
The PVX is a full-featured scope that allows a great deal of control over the functionality and electronic performance of the instrument. However, as this provides the user greater control and versatility for a variety of applications, it also requires that the user be comfortable with operation and waveform interpretation of the instrument. It is strongly recommended that the contents of this manual be read in its entirety. It is also recommended that a sufficient amount of time be allocated to working with the instrument in a test environment prior to use in the field. Responsibility for proper use of the instrument and recognition of this phenomenon rest solely with the user of the instrument.
1
Page 6
CHAPTER TWO QUICK STARTUP GUIDE
Turn the PVX on and off using the swit ch located on the bottom right corner of the keypad. When PVX is initially turned on, a flash logo and blinking lights will be displayed prior to entering into the main measurement screen. Note: This section is primarily written as a basic startup guide only.
2.1 Selecting The Transducer Type
The first step in using the PVX is to select the transducer type stored in a setup. There are 14 factory setups containing different transducer types that can be loaded, modified and stored in a new setup location. By selecting the transducer type from a predefined setup list, the PVX can recall specific properties about the transducer, setup the display area and predefined scope settings. Note: Once the transducer has been selected, the PVX will store and recall th is setup, containing the transducer type, every time the PVX is powered on/off. This setup will only change if the user physically selects another setup from the list. Therefore, if you have previously gone through this section and selected the appropriate setup, proceed to the next section. Use the following steps to select your setup (transducer type):
Selecting a Setup for use with a specific Transducer Type
1) Press the MULTI MODE key once to activate the load setup list containing the transducer types and settings.
2) Use the UP and DOWN arrow keys to scroll through the setup list until the
2
target setup is highlighted.
Page 7
PVX Precision Visual Thickness Gauge
3) Press the ENTER key to display the confirmation screen.
4) Press the OK key to select the setup and return to the measure screen, or
ESC to cancel the selecting a setup.
2.2 Calibration
The next step is to calibrate the PVX to the material and transducer being used. If a sound velocity is not known, the PVX can be calibrated to a known thickness sample. This demo will briefly explain each of possible calibration techniques. Read through the various calibration techniques and select one of the following:
Known Velocity
If the material velocity is known, the user may wish to simply enter the velocity number into the PVX, rather than have the PVX calculate the velocity value using a know thickness. The steps for entering the velocity are outlined below:
Using a Known Material Velocity
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the CAL menu is highlighted and displaying the submenu items.
3
Page 8
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until VELOCITY is highlighted.
3) Press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps 4 & 5 until the velocity number is correctly displayed.
7) Press the OK key to set the velocity and return to the menu screen, or ESC to cancel entering the velocity.
8) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
Known Thickness
Sometimes the sound velocity of a material is not known. In this case a sample with a known thickness can be used to determine the sound velocity. It would be very handy to carry a set of mechanical calipers to use in conjunction with the PVX for calibration in the field:
Using a Known Thickness
4
Page 9
PVX Precision Visual Thickness Gauge
1) Physically measure an exact sample of the material or a location directly on the material to be measured using a set of calipers or a digital micrometer.
2) Apply a drop of couplant on the transducer and place the transducer in steady contact with the sample or actual test material. Be sure that the reading is stable and the repeatability indicator, in the top left corner of the display, is fully lit and stable. Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the CAL menu is highlighted and displaying the submenu items.
3) Use the UP and DOWN arrow keys to scroll through the sub menu items until ONE POINT is highlighted.
4) Press the ENTER key to display the Digits Edit Box.
5) Press the UP and DOWN arrow keys to scroll the highlighted value.
6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.
5
Page 10
8) Press the OK key to calculate the velocity and return to the menu screen, or ESC to cancel entering the velocity.
9) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
Note: CHECK YOUR CALIBRATION! Place the transducer back on the calibration point. The thickness reading should now match the known thickness. If the thickness is not correct, repeat the steps above.
6
Page 11
PVX Precision Visual Thickness Gauge
Basic Material Type
If the material velocity is unknown, and a sample thickness cannot be taken from the material, the user may opt to choose a basic material type from a list with approximate velocity numbers. It’s important to note that these velocities will not always be an exact representation of the material being tested. Use these values only if a close approximation is acceptable. Follow the steps below to select a basic material type:
Selecting a Basic Material Type
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the CAL menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until MATERIAL is highlighted.
3) Press the ENTER key to display the list of material types.
4) Press the UP and DOWN arrow keys to scroll through the material list until the appropriate material is highlighted.
5) Press the ENTER key to display the confirmation screen.
6) Press the OK key to select the material and return to the menu screen, or ESC to cancel the material selection.
7) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
7
Page 12
2.3 Measure
The PVX is now ready to measure. There are four different measurement view options, each with a specific purpose. The steps below outline how to toggle between the different view mode options:
Selecting the Measurement View Option
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the DISP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until VIEW is highlighted.
3) Use the LEFT and RIGHT arrow keys to scroll the view options.
4) Once the view is displayed, press the MEAS key to return to measurement
mode.
RF: Is useful to see exactly what the signal looks like directly around the detect point. This view is the preferred view for precision measurements, as it allows the user to view smaller ranges (zoomed in) around the detection point (displayed as a vertical dashed line). This can be done manually by adjusting the delay and width settings. It also provides the user with the ability to view and select either the positive or negative polarity option.
RECTIFIED (RECT): Displays the entire range being scanned. This screen is useful to “get the big picture” when viewing wide ranges (zoomed out). The point which is
triggering the digital thickness reading (called the detect) is displayed as a vertical dashed line.
8
Page 13
PVX Precision Visual Thickness Gauge
BSCAN: The Time Based B-Scan provides the user with a cross sectional view of the mat erial being tested. This mode is useful when there is concern regarding the profile of the blind surface.
DIGITS: Displays the digital thickness value using a larger font size. This view is useful when the PVX is being used as a basic precision thickness gauge following
the initial setup using the scope. Once the view has been selected according to the application requirements, the
delay and width of the screen will potentially need to be adjusted, if the view has been set to RF, RECT, or BSCAN. Use the following steps to adjust these settings
directly from the measurement screen as follows:
Adjusting the Delay & Width
1) Press the MEAS key once to activate the measure menu items. Press the
MEAS key multiple times to move right and the ESC key multiple times to
move left, until the either the DELAY or WIDTH cell is highlighted.
2) Use the UP, DOWN, LEFT, or RIGHT arrow keys to scroll the DELAY and
WIDTH values.
3) Repeat steps 1 & 2 until the range is correctly being displayed.
Alternatively, the DELAY and WIDTH values can be changed using the Digit Edit Box as follows:
9
Page 14
1) Press the MEAS key once to activate measure menu items. Press the
MEAS key multiple times to move right and the ESC key multiple times to move left, until the either the DELAY or WIDTH cell is highlighted.
2) Press the ENTER key to display the digits edit box.
3) Press the UP and DOWN arrow keys to scroll the highlighted value.
4) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
5) Repeat steps 3 & 4 until the DELAY or WIDTH value is correctly displayed.
6) Press the OK key to set the DELAY and WIDTH value and return to the measure screen, or ESC to cancel entering the DELAY or WIDTH value.
7) Finally, press the MEAS key to return to the measurement screen and begin
taking readings.
Note: The DELAY & WIDTH can also be adjusted from the menu tab items. However, using the hot menu keys is the easiest method.
RF RECTIFIED
10
Page 15
PVX Precision Visual Thickness Gauge
B-SCAN DIGITS
In the upper left corner of each of the mode photos above, is the repeatability indicator. The repeatability indicator is represented by six vertical bars and represents how repeatable the measurements are. In regular measurement mode, the PVX makes 4 measurements a second. In scan mode, the PVX makes 32 measurements a second. When the PVX is idle, only the left vertical bar and the underline will be displayed. However, when the PVX is making a measurement, five or six of the bars should be displayed on the repeatability indicator. If fewer than five bars are showing, the PVX is having difficulty achieving a stable measurement and the thickness value displayed is potentially be unstable. Another thing to note is the vertical broken line in the RF and RECTIFIED photos above. This line represents the
actual point of detection. The PVX uses a zero crossing or flank detection method, and presents the actual point of measurement on the display.
11
Page 16
CHAPTER THREE KEYBOARD, MENU, & CONNECTOR REFERENCE
3.1 Menu Key (Operation & Sub Menus )
The Menu key activates the primary menu structure containing 9 menu tab groups. These tab groups then contain sub menu items, or functions. The sub menu items have been organized in tab groups according to how closely they are related to the individual tab group names. Let’s first get familiar with how to move around in these tabs before continuing on to the sub menu functions. This procedure is outlined below:
Activating and Getting Around in the Menu Items
12
Page 17
PVX Precision Visual Thickness Gauge
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left until the desired tab group is highlighted and displaying the submenu items.
Now that your familiar with activating and moving amongst the tab groups, let’s have a look at how to move around in the sub menu items as follows:
Getting Around in the Sub Menu Items
1) Use the UP and DOWN arrow keys to scroll through the sub menu items until the desired function is highlighted.
2) Depending on which function is highlighted, use the LEFT, RIGHT, and Enter keys to scroll the options or activate the Digit Edit and List Box
options.
The sections to follow will provide the user with an explanation of the sub menu functions:
3.2 Probe – Menu
Zero: The PVX is zeroed in much the same way that a mechanical micrometer is zeroed. If the PVX is not zeroed correctly, all of the measurements made using the
PVX may be in error by some fixed value. Refer to the section on page 33, for an explanation of this important procedure.
Note: This function is only used when using a contact style transducer with a pulse­echo contact setup selected. All the other transducer types use multiple echo
modes, and the zero function is not needed. Pulse: The PVX has an adjustable pulse width for both high penetration and resolution applications. The pulse width refers to the duration of time the pulser is
13
Page 18
on. The options are Spike, Thin, and Wide. Refer to page 71 for a further explanation.
3.3 CAL – Menu
Units: Toggle between English or Metric units and low and hi resolutions. The
readout will change from inches to millimeters (IN, IN HR, MM, MM HR). The high resolution options are indicated with HR following the units, and offer an additional decimal place of overall resolution (i.e. IN = .001” and IN HR = .0001”).
Velocity: Function to calibrate the PVX by setting the velocity to a known material velocity. Refer to page 35 for further info.
One Point: Performs a single point calibration. This option allows the user to automatically calculate the velocity by entering a known sample thickness. Refer to page 37 for further info.
Two Point: Performs a two-point calibration. This option allows the user to automatically calculate the velocity by entering a second known sample thickness. Refer to page 38 for further info.
Material: Select the material velocity from a chart of basic material types, when a known sample thickness, or material velocity cannot be obtained. Refer to page 41 for further info.
3.4 DISP (display) – Menu
View: Selectable RF wave, RECT (rectified) wave, BSCAN (cross section), and
DIGITS (large digits) views. Refer to page 43 for further info. Backlight: Selectable OFF, ON, AUTO, or INVERT backlight option. The INVERT
option is actually a screen inversion that toggles background and foreground colors.
Contrast: Adjustable display contrast for variable light conditions. Delay: Provides the user the ability to change where the left side of the display
window starts according to thickness, in inches or millimeters. Refer to page 53 for further info.
Width: Provides the user the ability to change the overall size of the viewable measurement area. It functions a lot like a zoom on a camera. Refer to page 50 for further info.
14
Page 19
PVX Precision Visual Thickness Gauge
3.5 TUNE – Menu
Gain: Increases or decreases the overall amplitude of the signal. Much like turning
the volume up or down on a stereo receiver. Refer to page 55 for further info. AGC: The PVX is equipped with an automatic gain control when operating in
interface-echo and echo-echo modes only. This is much like turning the volume up or down on a stereo receiver. However, the PVX will automatically control how much the volume is turned up or down. Alternatively, the AGC can be manually controlled using the same procedures as GAIN described above. Refer to page 55 for further info.
Threshold: Enables the user to set the sensitivity level of the PVX. The amplitude of the signal must reach and exceed the threshold level before a measurement is detected. Refer to page 58 for further info.
Polarity: The PVX operates on a zero crossing detection principle. This feature toggles which stroke of the cycle the crossing detection uses, either positive or
negative. Refer to page 76 for further info. Rect Waveform: This option provides the user an outlined or filled view option when the display setting is in RECT (rectified) wave mode only. Refer to page 78 for
further info.
3.6 GATE – Menu
Measure Mode: Used to select the measurement mode for different application
requirements. The modes are P-E (pulse-echo), P-E GT(pulse-echo w/gate), and E­E(echo-echo). Note: I-E or Interface echo mode is also a multiple echo mode that
uses two gates like E-E mode(echo-echo). Refer to page 23 for further info. Gate 1: Gates allow the user to view areas or sections of the waveform and ignore
others. Gate 1 can be used in both pulse-echo and echo-echo measurement modes. Refer to page 61 for further info.
Gate 2: Gates allow the user to view areas or sections of the waveform and ignore others. Gate 2 is only used in echo-echo (interface-echo) measurement modes. Refer to page 65 for further info.
15
Page 20
3.7 SETUP – Menu
Open: Displays a list of factory and user defined setups currently stored in memory. These setups can be recalled and used at any time. Refer to page 99 for further info. Save: Provides the user with the ability to save a custom setup that has been
modified or created by the user. Refer to page 101 for further info. Default Setup: Loads a basic default setup. Use only as a last resort when the
setups in the PVX have been corrupted and a computer is not accessible. Refer to page 104 for further info.
Language: Provides the user with three different language options: English, German, and Spanish.
3.8 DATA – Menu
New: Allows the user the ability to create a new alpha numeric grid with custom
parameters, rows, and columns depending on their application reporting requirements. Refer to page 81 for further info.
Edit: Gives the user the ability to change parameters of grid that have been previously saved. Note: Pre-defined coordinates cannot be changed once they have been created. Refer to page 94 for further info.
Open: This function provides the user with the ability to recall grids that currently exist in memory from a list of grids. Refer to page 96 for further info.
Delete One Grid: This function provides the user with the ability to delete one individual grid from a list of multiple grids previously saved in memory. Refer to page 92 for further info.
Delete All Grids: This function provides the user with the ability to delete all grids currently stored in memory. Refer to page 92 for further info.
3.9 UTIL (utilities) – Menu
16
Page 21
PVX Precision Visual Thickness Gauge
Auto Find: Automatically locates the detection point if the measurement is out of the viewable display area. Refer to page 72 for further info. Scan Mode: This function enables a hi speed scan mode that increases the overall sample rate from the standard 4 measurements a second to 32 measurements a
second. Refer to page 73 for further info.
Alarm Status: Toggles alarm mode on or off. Refer to page 74 for further info. Alarm LO Limit: Gives the user the ability to set the LO limit parameter. If the
measurement falls below this value, a red light will illuminate and sound the internal beeper. Refer to page 75 for further info.
Alarm HI Limit: Gives the user the ability to set the HI limit parameter. If the measurement exceeds this value, a red light will illuminate and sound the internal beeper. Refer to page 76 for further info.
3.10 XFER (transfer) – Menu
Backup Setups: Enables the user the ability to backup the setups currently stored in the PVX to a PC via RS232 port. Refer the help section of the DakView 3 or PVXview software for a complete elect ronic manual.
Restore Setups: Enables the user the ability to restore the setups currently saved on a PC to an PVX via RS232 port. Refer the help section of the DakView 3 or PVXview software for a complete electronic manual.
Backup Grid: Enables the user the ability to backup grids currently stored in the
PVX to a PC via RS232 port. Refer the help section of the DakView 3 or PVXview software for a complete electronic manual.
Restore Grid: Enables the user the ability to restore grids currently saved on a PC to an PVX via RS232 port. Refer the help section of the DakView 3 or PVXview software for a complete electronic manual.
About: Provides the user with Dakota Ultrasonics contact information and the PVX software version. Refer the help section of the DakView 3 or PVXview software for a
complete electronic manual.
3.11 CLR (clear) Key
The primary functions of the CLR key, is to clear a measurement from a grid cell location or set obstruct, and backspace in an Alpha Edit Box. If a user has already
17
Page 22
saved a measurement and waveform to a cell location, use this key to clear the measurement at any time.
3.12 MEAS (measurement mode) Key
The MEAS key puts the PVX into it’s primary mode of operation. In this mode, the user has a complete view of the LCD, as well as control of the Hot Menu Functions. These hot functions provide the user with the ability to make crucial adjustments to the display, waveform settings, and grid storage locations without having to search through the menu and sub menu items. It’s important to first get familiar with moving around in the hot functions as follows:
Getting Around in the Hot Menus
1) Press the MEAS key at any time to return to the primarily measurement
mode.
2) Press the MEAS key multiple times to tab right through the menu fields, and
the ESC key multiple times to tab left through the menu fields, until the desired hot function is highlighted.
Now that you’re familiar with activating and moving amongst the hot function fields, let’s have a look at how to adjust or change the values of these fields:
Adjusting the Values of the Hot Menus
1) Use the UP, DOWN, LEFT, and RIGHT arrow keys to increase and decrease the values of the hot function fields.
2) Repeat step 1 until the desired value has been achieved.
18
Page 23
PVX Precision Visual Thickness Gauge
Alternatively, the delay, width, gain, and threshold hot function fields can be changed using the Digit Edit Box as follows:
1) Press the ENTER key to display the Digits Edit Box.
2) Press the UP and DOWN arrow keys to scroll the highlighted value.
3) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
4) Repeat steps 2 & 3 until the delay, width, gain, or threshold number is correctly displayed.
5) Press the OK key to set and return to the measure screen, or ESC to cancel entering the delay, width, gain, or threshold value.
3.13 OK Key
The primary function of the OK key is confirmation of a change or selection. Also toggles display views (Rect, RF, Digits, B-Scan) while in a highlighted grid location.
3.14 ESC Key
The ESC key is used in the MENU, MEAS, and EDIT functions as a back or escape function. Also toggles display views (RECT, RF, DIGITS, or B-SCAN) while in a
highlighted grid location. Note: If a B-SCAN was originally saved in grid location, only DIGITS view can be toggled in conjunction with the B-SCAN.
3.15 Arrow Keys
The Arrow Keys are used to navigate through the menus, increase/decrease values, and toggle specific function keys.
19
Page 24
3.16 ENTER key
The ENTER key is used in the overall menu selection process, to activate list and edit boxes, and save measurements to grid locations.
3.17 MULTI MODE Key
The MULTI MODE key activates the Load Setup List to select a factory or previously saved user setup for a variety of different transducer types and scope settings. This is a quick and efficient way to change modes and transducer types without having to manually adjust PVX. The factory setups serve as the perfect starting point for immediate use or further fine-tuning.
3.18 ON/OFF Key
The ON/OFF key simply powers the unit either ON or OFF. Note: Unit will automatically power off when idle for 5 minutes. All current settings are saved prior to powering off.
3.19 Top & Bottom End Caps
The top & bottom end panels are where all connections are made to the PVX. The diagram above shows the layout and description of the conn ectors:
Transducer Connectors
Refer to Diagram: The transducer connector and battery cover/probe zero disk are located on the PVX’s top end cap. The transducer connector is of type Lemo “00”.
20
Page 25
PVX Precision Visual Thickness Gauge
Probe Zero Disk & Battery Cover Refer to Diagram: The Battery cover is the large round disk shown in the diagram.
Note: This same disk is also used as a probe zero disk. Simply remove the cover
when replacing the batteries (3 AA cells). When performing a probe zero function, simply place the transducer on the disk making firm contact.
Important: Be sure to follow the polarity labels located on the back label of the PVX. Note: Rechargeable batteries can be used, however they must be recharged outside
of the unit in a stand alone battery charger.
RS-232 Connector Refer to Diagram: The RS-232 connector, located on the bottom end cap, is a 2 pin
female Lemo connector. It is designed to connect directly from the PVX to a standard AT serial port on a PC. The cable supplied with the PVX is a Lemo to 9 pin serial cable. Note: This connector is also used to upgrade the PVX with the latest version of firmware.
21
Page 26
CHAPTER FOUR PRINCIPALS OF ULTRASONIC MEASUREMENT
4.1 Time versus thickness relationship
Ultrasonic thickness measurements depend on measuring the length of time it takes for sound to travel through the material being tested. The ratio of the thickness versus the time is known as the sound velocity. In order to make accurate measurements, a sound velocity must be determined and entered into the instrument.
The accuracy of a thickness measurement therefore depends on having a consistent sound velocity. Some materials are not as consistent as others and accuracy will be marginal. For example, some cast materials are very granular and porous and as a result have inconsistent sound velocities.
While there are many different ultrasonic techniques to measure thickness, which will be discussed below, all of them rely on using the sound velocity to convert from time to thickness.
4.2 Suitability of materials
Ultrasonic thickness measurements rely on passing a sound wave through the material being measured. Not all materials are good at transmitting sound. Ultrasonic thickness measurement is practical in a wide variety of materials including metals, plastics, and glass. Materials that are difficult include some cast materials, concrete, wood, fiberglass, and some rubber.
4.3 Range of measurement and accuracy
The overall measurement capabilities, based on the wide variety of materials, is determined by the consistency of the material being measured
The range of thickness that can be measured ultrasonically depends on the material as well as the technique being used and the type of transducer. Thickness measurements can be made from a minimum of 0.010 inch to 9.999” in steel. However, the maximum attainable thickness is much less for more attenuative materials (materials that absorb sound).
Accuracy, is determined by how consistent the sound velocity is through the sound path being measured and, is a function of the overall thickness of th e material. For example, the velocity in steel is typically within 0.5% while the velocity in cast iron can vary by 4%.
4.4 Couplant
All ultrasonic applications require some medium to couple the sound from the transducer to the test piece. Typically a high viscosity liquid is used as the medium. The sound frequencies used in ultrasonic thickness measurement do not travel
22
Page 27
PVX Precision Visual Thickness Gauge
through air efficiently. By using a liquid couplant between the transducer and test piece the amount of ultrasound entering the test piece is much greater.
4.5 Temperature
Temperature has an effect on sound velocity. The higher the temperature, the slower sound travels in a material. High temperatures can also damage transducers and present a problem for various liquid couplants.
Since the sou nd velocity varies with temperature it is important to calibrate at the same temperature as the material being measured.
Normal temperature range
Most standard transducers will operate from 0°F to 180°F.
High temperature measurements
Special transducers and couplants are available for temperatures above 180°F up to 650°F with intermittent contact. It is necessary to cool the transducer, by submerging the transducer in water between readings, when measuring high temperatures.
Modes and temperature errors
In addition to errors caused by velocity changing with temperature, some modes (measurement techniques) are affected more than others. For example, dual element mode has larger errors due to changes in temperature of the delay line. However, multi-echo tec hniques help to minimize these errors. The PVX operates in multi-echo modes for a variety of transducer types.
4.6 Measurement Modes
In this section we will discuss the different measurements modes the PVX is capable of operating in, the transducers required, and the reasons for using specific modes:
Pulse-Echo Mode (flaws & pits)
Pulse-echo mode measures from the initial pulse (sometimes referred to as an artificial zero) to the first echo (reflection). In this mode, the transducer is placed on a reference disk, located on top of the PVX, and a key is pressed to establish a zero point for the particular transducer.
In pulse-echo mode errors result from surface coatings and temperature variations. Since pulse-echo only requires one reflection, it is the most sensitive mode for
measuring weak reflections (flaws) typically found when measuring heavily corroded metals.
Echo-Echo (Interface-Echo) Modes
The echo-echo modes measure between two reflections. This technique is commonly used to eliminate errors from surface coatings in echo-echo mode, as well as make measurements in multiple layered materials. The disadvantage is that two
23
Page 28
echoes are needed which requires a much stronger echo (reflection). In interface­echo mode, it moves the interface further out in time eliminating any noise that’s close to the initial pulse.
Important: The primary difference between interface-echo and echo-echo mode, is interface-echo mode measures from the end of an interface (delay line), to the first
return reflection of the test material. However, in echo-mode, the measurement is made between two return reflections (echoes). Although both modes use two gates and virtually look the same on the scope, they are measuring in different ways and are used to accomplish different application tasks. The diagrams below will offer an additional explanation of the differences in these modes as follows:
Delay Line Single Element Transducer in Interface-Echo mode
Diagram 1
Refer to diagram 1. This diagram illustrates the sound path being measured in
interface-echo mode. (A) is the transducer. Point (B) is the piezoelectric crystal that emits the sound wave. (C) is the delay line, most commonly made of acrylic or graphite. The delay tip will act as the interface and the measurement will be made starting from the difference of the reflection off the end of the delay tip (E). The sound will also travel through the delay tip (C) and test material (F), and reflect back to the piezoelectric crystal off the back wall of the test material. The measurement will be made from the reflection off the end of the delay tip and back wall of the test material.
24
Page 29
PVX Precision Visual Thickness Gauge
Diagram 2
Refer to diagram 2. Referring to (Idle), you’ll notice that when the PVX is in
interface-echo mode a waveform from the end of the delay line will always be displayed and active on the screen when idle (A). Since the gates are also used in interface-echo mode, they will be displayed (B). When measuring, two detects will be displayed between two waveforms at (C) & (F). The measurement starts at (C), the end of the delay line, and measures to the back wall of the test material at (F). Point (D) represents the gate 1 start position, and (E) the gate 2 hold-off. Notice that the start position of the gate 2 hold-off (E) has been adjusted to the right of the delay line waveform. This is done to avoid potential detection on the ring down of the delay line interface, and allow a detection to occur on the back wall of the test material at (F).
Diagram 2
25
Page 30
Refer to diagram 3. The above diagram illustrates a typical view of echo-echo mode using a delay line style transducer. The end of the interface or delay line is displayed at (B). However, unlike interface echo mode the measurement is occurring between two return reflections (A) and (B). Recall that in interface echo mode, the measurements occurred between the interface (B) and the first return reflection (A). This is the primary difference between the two modes. An important thing to note is that the ring down cycles from the transducer are much less than in interface echo mode. This allows the PVX to see much closer to the first return reflection making it possible to measure very thin materials. This mode is also used for thru-paint measurements. By measuring between two echoes, the thickness of paint & coatings will be eliminated completely. Therefore, any paint or coating on the surface of the test material will not be added to actual thickness of the test material.
4.7 Glossary of terms
Repetition Rate & Display Update
The repetition rate is how often the PVX generates a burst of ultrasound. The PVX has a fixed repetition rate of 1 kHz, or 1000 times per second. The PVX
updates the display 4 times per second in regular measurement mode and 32 times a second in hi speed scan mode. Therefore, the scan mode is most suitable for fast scans of a test piece at a lower resolution.
Variable Pulser Width
Transducers have a center frequency where they are most efficient at transmitting sound (i.e. they are resonant). The PVX has a variable width pulser with three setting options (spike, thin, wide). These settings determine the period of time the pulser turned is on. High frequency applications, where resolution is a requirement, may require the spike or thin settings to achieve optimal results. Low frequency applications, where more penetration is a needed, may require the thin or wide settings, offering increased penetration. The variety of transducer frequencies and diameters used in conjunction with the pulser settings, enable users to fine tune the PVX for their application needs.
Gates
Gates control the time measurement process. They enable the user to select measurement regions and threshold levels (sensitivity). The PVX is equipped with two gates. In pulse-echo mode a single gate can be used to eliminate noise. In echo-echo (interface-echo) mode both gates are used. In this mode the second gate starts automatically after a signal is detected during the gates region, and after a user adjustable delay, sometimes called a hold-off.
26
Page 31
CHAPTER FIVE SELECTING THE MEASUREMENT MODE
5.1 The setup library
The PVX contains 16 factory setups and 48 user definable preset locations to store custom setups for easy recall. These setups can be optimized for the user’s specific application needs and can also be stored on a PC and transferred bi-directionally using Dakota’s PC interface software included with the instrument.
The setups supplied with the instrument are optimized for specific transducer types and serve as a starting point for some of the typical applications commonly used with this type of instrument. These setups can be recalled, modified, and overwritten to one of 48 setup locations. The PC software includes a default setup file that can be uploaded to the gauge at any time to restore factory settings. Note: The first 16 setups cannot be overwritten. Once again, these factory settings are excellent starting points for custom setups.
5.2 Which mode & transducer do I use for my application?
High penetration plastics and castings
The most common mode for these types of applications is pulse-echo. Select a low frequency transducer. Depending on the thickness of the material, use a 2.25MHz or lower frequency transducer. Larger diameters offer greater penetration power because of the crystal size, for difficult to measure materials.
Thin Materials & Tough Access Areas
Use any of the delay-line transducers for materials under 1 inch. Tough access areas with minimal clearances, are perfect applications for our pencil style transducers. However, the pencil transducer have a very small thickness range and are only suitable for thin materials less than .300”.
Thru Paint & Coatings
Often times users will be faced with applications where the material will be coated with paint or some other type of epoxy material. Since the velocity of the coating is approximately 2.5 times slower than that of steel, pulse-echo mode will induce error if the coating or paint is not completely removed. By using echo-echo mode, the user is able to successfully measure through both, the coating and steel, and completely eliminate the thickness of the paint or coating. Therefore, the steel can be measured without having to remove the coating prior to measuring.
Both the delay-line and contact transducers will work fine for thru coating measurements. When using contact style transducers, it’s best to consider a high damped model for thinner materials to minimize the potential ring down noise of the transducer.
27
Page 32
Plastics
The PVX works extremely well on thin plastics using a delay line transducer and special graphite tip. Because the velocity of the standard acrylic tip has a similar velocity to most plastics, the PVX is unable to measure the material because it simply can't see it. By using the graphite tip in place of th e acrylic tip, the velocity of the graphite tip allows an impedance mismatch to occur between materials, hence the plastic can successfully be measured.
When measuring thicker plastics, a lower frequency contact transducer should be used – typically 1 - 2. 25MHz. Larger crystal diameters offer greater penetration capabilities. Therefore, for thicker and hard to penetrate plastics, use larger diameter transducers if possible.
High temperature
High temperature applications require a special delay line transducer using a high temperature delay tip with a very low temperature coefficient. In other words, the delay tip is not absorptive at high temperatures.
Noisy Material
Materials such as titanium, stainless steel, and aluminum may have inherent surface noise issues that appear close to the initial pulse, or near the surface. This can be avoided by using a delay style transducer in either echo-echo or interface-echo modes. However, when using a contact style transducer in pulse-echo mode, it’s best to consider a higher frequency transducer. The PVX is also equipped with a single gate in pulse-echo mode. Enabling a single gate is also an effective tool for
blocking unwanted surface noise. See the section Gate 1 on page 61 for more information.
Restricted access
In order to measure materials with lots of curvature or restricted access, the use of a delay line transducer with a cone delay tip or a pencil style transducer, can provide additional access.
28
Page 33
PVX Precision Visual Thickness Gauge
E DELAY LINE 3/8
E DELAY LINE 1/2
5.3 Factory Setup Chart
Num Name Comment 1 Gn/AGC Velocity
1 E­2 I-E DELAY LINE 3/8 3 0.2330 in/uSec 3 E-E DELAY LINE 1/2 3 0.2330 in/uSec 4 I­5 E-E GRAPHITE STEEL 3 0.0940 in/uSec 6 I-E GRAPHITE STEEL 3 0.0940 in/uSec 7 I-E GRAPHITE PLASTIC 3 0.0920 in/uSec 8 E-E PENCIL PTR-100E 3 0.2330 in/uSec 9 I-E PENCIL PTR-100E 3 0.2330 in/uSec 10 E-E PENCIL PTR-300E 6 0.2330 in/uSec 11 I-E PENCIL PTR-300E 5 0.2330 in/uSec 12 E-E CONTACT STEEL 6 0.2330 in/uSec 13 P-E CONTACT STEEL 33 0.2330 in/uSec 14 P-E CONTACT PLASTIC 38 0.0920 in/uSec 15 16
3 0.2330 in/uSec
3 0.2510 in/uSec
29
Page 34
CHAPTER SIX MAKING MEASUREMENTS
The steps involved in making measurements are detailed in this section. The following sections outline how to setup and prepare your PVX for field use.
In pulse-echo modes, the probe zero must be measured on the reference disk (battery disk) attached to the top of the instrument. This compensates for variation in the transducer. In all modes the sound velocity must be determined. The sound velocity is used to convert the transit time to a physical length. The sound velocity can be selected from a material chart in the manual, selected from a material list in the PVX, or for greater precision, determined from a sample of the test material that has been mechanically measured. To enter the velocity from a table, look up the material on the chart in the appendix of this manual and refer to the section below on Calibration to a Known Velocity. T o determine the velocity of a single sample, refer to the Material Calibration section on page 37.
When measuring curved materials, it is more accurate to calibrate from two test points, one at the minimum limit of the target thickness and one at the maximum limit. In this case the reference disk mounted to the PVX is not used. This is called two­point calibration and is described on page 38.
6.1 Selecting The Setup Based on Transducer Type
The fir st step in using the PVX is to select setup according to the type of transducer, and or delay tip length from the list of factory setups. There are delay tip lengths of 3/8” and ½”. There are also two different pencil style transducer options: sphirical focused (PTR -100E) and flat (PTR-300E). The PTR-100E is the most common and popular option available. This option offers both a .060” and .125” diameter contact area tip. By selecting the setup associated with a specific transducer type, all the scope settings will also be loaded and setup. This will give the user an excellent
place to start, requiring only minor adjustments of the delay and width, if any. Note: Once the setup has been selected, the PVX will store and recall this setup every time the PVX is powered on/off. If changes are made to the setup, they must be saved in a setup location in or to recall them at a later time. The setup will only change if the user physically selects another setup from the list. Therefore, if you have previously gone through this section and selected a setup, proceed to the next section:
Selecting a Setup for use with a specific Transducer Type
30
Page 35
PVX Precision Visual Thickness Gauge
1) Press the MULTI MODE key once to activate the load setup list containing the transducer types and settings.
2) Use th e UP and DOWN arrow keys to scroll through the setup list until the target setup is highlighted.
3) Press the ENTER key to display the confirmation screen.
4) Press the OK key to select the setup and return to the measure screen, or ESC to cancel the selecting a setup.
31
Page 36
1) Alternately, the setups can be selected from the menu items. However,
this is not as efficient. Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the SETUP menu is highlighted and displaying
the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until OPEN is highlighted.
3) Press the ENTER key to display the list of stored setups.
4) Press the UP and DOWN arrow key s to scroll through the list of setups until the appropriate setup is highlighted.
5) Press the ENTER key to display the confirmation screen.
6) Press the OK key to select the setup and return to the menu screen, or ESC to cancel the selection process.
7) Press the MEAS key to return to the measurement screen.
Finally, in an effort to be sure that the user is aware of this option, the measurement mode can be manually selected from the menu items. Important Note: The only
time this option should be considered, is if the user plans to manually setup and configure the PVX from scratch – the beginning. If the user decides to do a setup from scratch, all other settings must be tuned from scratch as well. This is not an effort to discourage the user from learning or choosing to setup the PVX from the beginning, however it’s certainly the most time consuming option. The setups already contain the majority of the appropriate settings for the specific transducer, MODE, gates, gain, threshold, etc. Therefore, selecting a factory setup is a much easier option to start. Nevertheless, the measurement mode can be selected from the menu items as follows:
32
Page 37
PVX Precision Visual Thickness Gauge
Selecting the Measure Mode Using The Tabbed Submenu Items
Note: The E-E mode should also be selected for setting up the PVX in I-E
mode.
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the GATE menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until MEASURE MODE is highlighted.
3) Use the LEFT and RIGHT arrow keys to scroll the mode options. Once the
mode is displayed, press the MEAS key to return to the measurement mode screen.
6.2 Probe zero
A probe zero should only be done if a contact style transducer is being used in pulse­echo mode. As noted in chapter 3, the probe zero function is a very important and necessary function that must done prior to calibration. It should also be done on a regular basis. If the PVX is not zeroed correctly, all of the measurements taken may be in error by some fixed value. In order to perform a probe zero, the PVX must be in pulse-echo mode. If echo-echo mode was previously selected in the last section, skip this section and continue on to the next. Note: If you intend to switch between measurement modes simultaneously, setup up the PVX in pulse-echo mode, perform the probe zero as necessary, and save the setup to one of the empty setup locations. Once this is complet ed, the user can switch between modes toggling between two setups. The following steps outline the procedure for performing a probe zero:
33
Page 38
Performing a Probe Zero
1) Clean the surface of the Probe Zero Disk, located on top of the PVX
(battery cover).
2) Apply a drop of couplant to the end of the transducer, or probe disk, and hold the transducer in steady contact with the reference disk.
3) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the PROBE menu is highlighted and displaying the submenu items.
4) Use the UP and DOWN arrow keys to scroll through the sub menu items until ZERO TRANSDUCER is highlighted.
5) Press the ENTER key to display the confirmation screen.
6) Press the OK key to zero the transducer and return to the menu screen, or ESC to cancel the zeroing process.
7) Remove the transducer from the probe zero disk.
8) Press the MEAS key to return to the measurement screen.
34
Page 39
PVX Precision Visual Thickness Gauge
6.3 Material Calibration
In order for the PVX to make accurate measurements, it must be set to the correct sound velocity of the material being measured. Different types of materials have different inherent sound velocities. For example, the velocity of sound through steel is about 0.233 inches per microsecond, versus that of aluminum, which is about
0.248 inches per microsecond. If the gauge is not set to the correct sound velocity,
all of the measurements the gauge makes will be erroneous by some fixed percentage.
The One Point calibration is the simplest and most commonly used calibration method - optimizing linearity over large ranges. The Two Point calibration allows for greater accuracy over small ranges by calculating the probe zero and velocity. The PVX provides four simple methods for setting the sound-velocity outlined below:
Calibration to a known velocity
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the CAL menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until VELOCITY is highlighted.
3) Press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
35
Page 40
6) Repeat steps 4 & 5 until the velocity number is correctly displayed.
7) Press the OK key to set the velocity and return to the menu screen, or ESC
to cancel entering the velocity.
8) Finally, press the MEAS key to return to the measurement screen and begin
taking readings.
36
Page 41
PVX Precision Visual Thickness Gauge
One Point Calibration
Note: Be sure that a probe zero has been done prior to performing this
calibration procedure if a contact transducer in pulse-echo mode is being used.
1) Physically measure a location on the sample to be measured using a set of calipers or a digital micrometer. Note: It’s always best to calibrate on thickest location on the sample, rather than on the thinnest location, as the percentage error is much less with respect to the PVX’s resolution of +/ - .0001” (0.001mm).
2) Apply a drop of couplant on the transducer. Place the transducer in steady contact on the known thickness location. Be sure that the reading is stable and the repeatability indicator, located in the top left corner of the display, is fully lit and stable. Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the CAL menu is highlighted and displaying the submenu items.
3) Use the UP and DOWN arrow keys to scroll through the sub menu items until ONE POINT is highlighted.
37
Page 42
4) Press the ENTER key to display the Digits Edit Box.
5) Press the UP and DOWN arrow keys to scroll the highlighted value.
6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.
8) Press the OK key to calculate the velocity and return to the menu screen, or ESC to cancel entering the velocity.
9) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
Note: CHECK YOUR CALIBRATION! Place the transducer back on the calibration point. The thickness reading should now match the known thickness. If the thickness is not correct, repeat the steps above.
Two Point Calibration
Note: Be sure that a probe zero has been done prior to performing this
calibration procedure, if a contact transducer in pulse-echo mode is being used.
38
Page 43
PVX Precision Visual Thickness Gauge
Note: If the user opted to do a setup from scratch for interface-echo (I-E) mode, a two point calibration must be performed to set the probe zero.
1) Using a set of calipers, physically measure two known thickness locations on the sample of material. Note: The two locations should represent the
desired thickness range being measured. If the range is from .050” - .200”, then the two known locations should represent this range, or be slightly outside or inside of this range. A typical example would be to find two locations at .063” and the other at .247”. These thickness points will satisfy the required thickness range noted above.
2) Apply a drop of couplant on the transducer. Place the transducer in steady contact on thickest known location first. Be sure that the reading is stable and the repeatability indicator, located in the top left corner of the display, is fully lit and stable. Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the CAL menu is highlighted and displaying the submenu items.
3) Use the UP and DOWN arrow keys to scroll through the sub menu items until ONE POINT is highlighted.
4) Press the ENTER key to display the Digits Edit Box.
5) Press the UP and DOWN arrow keys to scroll the highlighted value.
6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.
8) Press the OK key to calculate the velocity for the ONE POINT calibration and return to the CAL menu screen, or press ESC to cancel entering the
velocity.
39
Page 44
9) Apply a drop of couplant on the transducer and place the transducer in steady contact with the sample or actual test material on the thinnest, or second known thickness location. Be sure that the reading is stable and the repeatability indicator, located in the top left corner of the display, is fully lit and stable.
10) Use the UP and DOWN arrow keys to highlight the TWO POINT calibration submenu item.
11) Press the ENTER key to display the Digits Edit Box.
12) Press the UP and DOWN arrow keys to scroll the highlighted value.
13) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
14) Repeat steps 12 & 13 until the known thickness value is correctly displayed.
15) Press the OK key to calculate the velocity for the TWO POINT calibration and return to the CAL menu screen, or press ESC to cancel entering the velocity.
16) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
Note: CHECK YOUR CALIBRATION! Place the transducer back on both
calibration points. The thickness readings should now match the known thickness locations. If they are not correct, repeat the steps above.
40
Page 45
PVX Precision Visual Thickness Gauge
Built in Material Selection
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the CAL menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until MATERIAL is highlighted.
3) Press the ENTER key to display the list of Material Types.
4) Press the UP and DOWN arrow keys to scroll through the material list until the appropriate material is highlighted.
5) Press the ENTER key to display the confirmation screen.
6) Press the OK key to select the mat erial and return to the menu screen, or press ESC to cancel the material selection.
7) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
41
Page 46
CHAPTER SEVEN USING THE A-SCAN & B-SCAN DISPLAYS
A key feature of the PVX is the waveform display. The waveform is a graphical representation of the sound reflections returning to the transducer. Consider standing at the base of a canyon and screaming “Hello There”. After a brief delay, you will here multiple echoes, “Hello There’s” repeated back to you. The waveform display shows the amplitude of the signal received on the vertical (Y) axis and time (shown in units of thickness) on the horizontal (X) axis.
The waveform display is very useful for viewing and adjusting the location of the gates. The gates are typically used to eliminate potential surface noise by adjusting the starting point in single echo modes, for multi echo measurement modes, as well as to adjust the threshold (sensitivity) in either mode. The waveform display is also very useful for locating internal flaws in materials, and is almost mandatory for use with pencil style transducers.
The B-Scan display is also very useful when scanning surfaces and viewing the cross section of the test material. It provides a convenient way of profiling the blind surfaces during a scan. The B-Scan display is also equipped with a scan bar representing the overall thickness. The scan bar gives the user a visual indication when a flaw or defect is passed over during the scan process. The scan bar will deflect off of the defect and return back to the overall thickness. Visually, this is much easier to notice than watching for changes in the digital value displayed. The scan bar has also been included in the large digits display mode for the same purpose.
Note: The following chapter outlines some of the fine adjustment features of the
PVX. The PVX has four different display options (RF A-Scan, Rectified A-Scan, B- Scan, and Large Digits). We’ll take a better look at these options in this chapter. Note: In order to recall and use the new adjustments made to the PVX at a later
time, the user must save the modified settings in one of the setup locations prior to powering off the unit. Refer page 99 for more information on setups.
7.1 Selecting a Display Option
The following procedure outlines how to select or toggle display options:
Changing Display Options
42
Page 47
PVX Precision Visual Thickness Gauge
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the DISP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until VIEW is highlighted.
3) Use the LEFT and RIGHT arrow keys to scroll th e view options. Once the appropriate view is displayed, press the MEAS key to return to the measurement screen.
43
Page 48
7.2 Display Views
RF View
RF
The RF mode shows the waveform in a similar fashion to an oscilloscope. It shows both the positive and the negative peaks. The peak (either positive or negative) selected for measurement is shown in upper portion of the display. This is the preferred mode for thin precision measuring modes, using delay line and pencil style transducers.
It’s important to note tha t a measurement must fall inside the displays viewable range in order to see the waveform. However, even if the waveform is outside of the viewable range of the display, a measurement can be taken and viewed on the digital readout. If the wave is outside of the display range, you can manually change the
range by adjusting the Delay and Width values, or you can use the Auto Find feature located in the UTIL menu of the tabbed menu items. Refer to page 72 for additional information on the Auto Find feature.
The following is a list of the viewable features on the display:
A) Stability of Reading Indicator – Indicates the stability of the return echo on a
scale of 1 to 6 – the solid bars displayed in the figure above indicate a repeatable signal. If the PVX is displaying a reading from memory, the repeatability indicator will be replaced by the text “MEM”.
B) Battery life Indicator – Fully charged batteries will appear filled in solid. Note:
The diagram shows the batteries at approximately 50%.
44
Page 49
PVX Precision Visual Thickness Gauge
C) Thickness Reading – Digital readout of thickness (inches or millimeters). D) Detect Indicator – The broken vertical line displays the zero crossing (flank)
detection point on the waveform where the measurement has been obtained. Notice that the digit al thickness readout is the same as the location of the detect indicator according to the measurements labels F, also in the diagram.
E) Echo Signal – Graphical full waveform representation of the echo signal
displayed as an amplitude (vertical or “Y” axis) plotted against time (horizontal or “X” axis), this time is then converted to a physical measurement and displayed according.
F) Measurement Labels – The measurement labels are calculated and displayed
based on where the left side of the display has been set (Delay), and the overall viewable area (Width) of the display. Note: The gray hash marks break the
display up into 5 segments, or quadrants. The measurement labels correspond to
the measurement at each hash mark. G) Units Label - Displays the current units of measurement (inches or millimeters). H) Hot Menus - Each of the fields located under the waveform display are called the
Hot Menu Fields. These fields allow quick control of some of the fine
adjustments needed to control the display settings, measurement modes, and grid
memory control. All of these fields can be adjusted without having to activate the
tabbed menu items and searching through a variety of menus to make
adjustments.
45
Page 50
Rectified
Rectified View
The rectified A-Scan mode shows half of the waveform. Either the positive or negative peaks are displayed, depending on the polarity selected. This is the preferred display view for flaw detection applications.
It’s important to note that a measurement must fall inside the viewable display range in order to see the waveform. However, even if the waveform is outside of the viewable area, a measurement can be taken and viewed on the digital readout. If the waveform is outside of the display range, you can manually change the range by adjusting the Delay and Width values, or you can use the Auto Find feature located in the UTIL menu in the tabbed menu items. Refer to page 72 for additional information on the Auto Find feature.
The following is a list of the viewable features on the display:
A) Stability of Reading Indicator – Indicates the stability of the return echo on a
scale of 1 to 6 – the solid bars displayed in the figure above indicate a repeatable signal. If the PVX is displaying a reading from memory, the repeatability indicator
will be replaced by the text “MEM”.
B) Battery life indicator – Fully charged batteries will appear filled in solid. Note:
The diagram shows the batteries at approximately 50%.
C) Thickness reading – Digital readout of thickness (inches or millim eters). D) Detect Indicator – The broken vertical line displays the zero crossing (flank)
detection point on the waveform where the measurement has been obtained. Notice that the digital thickness readout is the same as the location of the detect indicator according to the measurements labels F, also in the diagram.
46
Page 51
PVX Precision Visual Thickness Gauge
E) Echo Signal – Graphical full waveform representation of the return echo
displayed as an amplitude (vertical or “Y” axis) plotted against time (horizontal or
“X” axis), this time is then converted to a physical measurement. F) Measurement Labels – The measurement labels are calculated and displayed
according to where the left side of the display has been set (Delay), and the
overall viewable area (Width) of the display. Note: The gray hash marks break
the display up into 5 segments, or quadrants. The measurement labels
correspond to the measurement at each hash mark. G) Units Label - Displays the current units of measurement (inches or millimeters).
H) Hot Menus - Each of the fields located under the waveform display are called the
Hot Menu Fields. These fields allow quick control of some of the fine
adjustments needed to control the display settings, measurement modes, and grid
memory control. All of these fields can be adjusted without having to activat e the
tabbed menu items and searching through a variety of menus to make
adjustments.
B-Scan View
B-Scan
The B-Scan displays a time based cross section view of test material. This view is commonly used to display the contour of the blind, or underside, surface of the material being tested. It is very similar to a fish finder. If a flaw is located during a scan, the B-Scan will draw the flaw on the screen. The solid black rectangle in the diagram at location E represents the cross section, or side v iew of the material. You will notice that the overall thickness of the material is .500”, and the display range is
0.00” to 1.00” respectively. The B-Scan view draws at a rate of 15 seconds per
47
Page 52
screen from right to left. Also notice at location J, the contour thickness changes of the bottom surface of the material.
It’s important to note that the measurement range on the display be set wide enough, so that the maximum thickness of the material can be viewed on the display. Using the diagram above, if th e material thickness was actually 1.75”, the underside of the material would not be viewable according to the current range at 0.00” – 1.00”. All the user would see is a black screen from 0.00” – 1.00” with no view of the bottom contour at 1.75”.
The following is a list of the viewable features on the display:
A) Stability of Reading Indicator – Indicates the stability of the echo signal on a
scale of 1 to 6 – the solid bars displayed in the figure above indicate a repeatable signal. If the PVX is displaying a reading from memory, the repeatability indicator will be replaced by the text “MEM”.
B) Battery life indicator – Fully charged batteries will appear filled in solid. Note:
The diagram shows the batteries at approximately 50%. C) Thickness reading – Digital readout of thickness (inches or millimeters). D) B-Scan Display Area – This is the area representing where the B-Scan will be
drawn. Notice the range of the area in the diagram at 0.00” – 1.0” respectively. E) Displayed B-Scan – Graphical presentation of the cross section, or side view of
the test material. The top, or accessible side of the material represented as 0.00”,
and the bottom, or blind surface at .500”. You will notice the contour in the B-
Scan, indicating a very irregular bottom surface. The B-Scan is displayed from
right to left at a rate of 15 screens per second. F) Measurement Labels – The measurement labels are calculated and displayed
based on where the left side of the display has been set (Delay), and the overall
viewable area (Width) of the display. Note: The gray hash marks break the
display up into 5 segments, or quadrants. The measurement labels correspond to
the measurement at each hash mark.
G) Units Label - Display’s the current unit of measurement (inches or millimeters). H) Hot Menus - Each of the fields located under the B-Scan display are called the
Hot Menu Fields. These fields allow quick control of some of the fine
adjustments needed to control the display settings, measurement modes, and grid
memory control. All of these fields can be adjusted without having to activate the
tabbed menu items and searching through a variety of menus to make
adjustments. I) Scan Bar – The scan bar corresponds directly with thickness represented by the
horizontal labels on the on the display. As you can see from the diagram above,
the filled horizontal bar is displayed at .500”. This display can be very useful
when scanning materials in either B-Scan or Digits display views. It is much
easier to watch and notice deflections for flaws and pits using the scan bar, rather
than trying to notice changes in digital values. If the user is scanning a material,
48
Page 53
PVX Precision Visual Thickness Gauge
for instance, and passes over a flaw in the process, the scan bar will quickly deflect and alarm the user to go back and search for the defect.
J) Contour (blind surface) – A view of the B-Scan displaying the blind side surface
contour of the material.
DIGITS VIEW
DIGITS
The Digits view is a basic digital thickness gauge look and feel. The larger digits make it much easier for the operator to monitor th e thickness readings. The Scan Bar has also been added to the Digits view to provide the user with yet another visual tool for easily monitoring changes in thickness readings due to internal flaws or defects.
The following is a list of the viewable features on the display:
A) Stability of Reading Indicator – Indicates the stability of the return echo on a
scale of 1 to 6 – the solid bars displayed in the figure above indicate a repeatable signal. If the PVX is displaying a reading from memory, the repeatability indicator
will be replaced by the text “MEM”.
B) Battery life indicator – Fully charged batteries will appear filled in solid. Note:
The diagram shows the batteries at approximately 50%.
C) Thickness reading – Digital readout of thickness (inches or millimeters). D) Large Digits – Large readout of thickness (inches or millimeters). E) Units Label – Display’s the current unit of measurement (inches or millimeters). F) Measurement Labels – The measurement labels are calculated and displayed
based on where the left side of the display has been set (Delay), and the overall viewable area (Width) of the display. Note: The gray hash marks break the
49
Page 54
display up into 5 segments, or quadrants. The measurement labels correspond to
the measurement at each hash mark. G) Scan Bar – The scan bar corresponds directly with thickness represented by the
horizontal labels on the on the display. As you can see from the diagram above,
the filled horizontal bar is displayed at .500”. This display can be very useful
when scanning materials in either B-Scan or Digits display views. It is much
easier to watch and notice deflections from flaws and general changes in
thickness using the scan bar, rather than trying to notice changes in the digital
value. If the user is scanning a material, for instance, and passes over a flaw
during the process, the scan bar will quickly deflect and alarm the user to go back
and search for the defect.
H) Hot Menus - Each of the fields located under the display are called the Hot Menu
Fields. These fields allow quick control of some of the fine adjustments needed
to control the display settings, measurement modes, and grid memory control.
All of these fields can be adjusted without having to activate the tabbed menu
items and searching through a variety of menus to make adjustments.
7.3 Adjusting the screen
The width refers to the overall viewable range, with respect to depth, being displayed on the screen. The starting depth is displayed at the lower left side of the screen, and the ending depth is shown at the lower right hand side of the screen. The difference between these values is called the Width. The procedures to adjust the
Width are outlined below:
Adjusting the Width using the Hot Menus
50
Page 55
PVX Precision Visual Thickness Gauge
1) Press the MEAS key once to activate measure menu items. Press the MEAS key multiple times to move right and the ESC key multiple times to move left, until the WIDTH cell is highlighted.
2) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the highlighted value.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps 4 & 5 until the WIDTH value is correctly displayed.
7) Press the OK key to return to the measurement screen, or ESC to cancel
entering the WIDTH.
The user can also access and adjust the width from the tabbed menus. However, this method is a more tedious process than making the adjustments using the Hot Menus. The procedure using the tabbed menus is outlined below:
51
Page 56
Adjusting the Width using the Tabbed Menus
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the DISP menu is highlighted and displ aying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until WIDTH is highlighted.
3) Press the LEFT and RIGHT arrow keys to scroll the value. When the correct width is being displayed, proceed to step 8.
4) Alternatively, press the ENTER key to display the Digits Edit Box.
5) Press the UP and DOWN arrow keys to scroll the highlighted value.
6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
7) Repeat steps 5 & 6 until the Width value is correctly displayed.
8) Press the OK key to set the Width and return to the menu screen, or ESC to
52
cancel entering the Width.
Page 57
PVX Precision Visual Thickness Gauge
9) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
Starting delay
The starting DELAY, or starting depth, is the value displayed on the bottom lower left side of the display in both RF and RECT (Rectified) views. The DELAY will also be represented in both the BSCAN and DIGITS (Scan Bar) views, as the starting thickness depth. The starting depth is the minimum thickness value that can be viewed on the display. In DIGITS view, the delay will only affect the starting depth of the Scan Bar.
Note: Once the range is set, it will remain the same for all the views respectively.
The procedures to adjust the Delay are outlined below:
Adjusting the Delay using the Hot Menus
1) Press the MEAS key once to activate measure menu items. Press the MEAS key multiple times to move right and the ESC key multiple times to
move left, until the DELAY cell is highlighted.
2) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the highlighted value.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
53
Page 58
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit location s.
6) Repeat steps 4 & 5 until the DELAY value is correctly displayed.
7) Press the OK key to return to the measurement screen, or ESC to cancel
entering the DELAY.
The user can also access and adjust the delay from the tabbed menus. However, this method is a more tedious process than making the adjustments using the Hot Menus. The procedure using the tabbed menus is outlined below:
Adjusting the Delay using the Tabbed Menus
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the DISP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until DELAY is highlighted.
3) Press the LEFT and RIGHT arrow keys to scroll the value. When the correct Delay is being displayed, proceed to step 8.
54
Page 59
PVX Precision Visual Thickness Gauge
4) Alternatively, press the ENTER key to display the Digits Edit Box.
5) Press the UP and DOWN arrow keys to scroll the highlighted value.
6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
7) Repeat steps 5 & 6 until the Delay number is correctly displayed.
8) Press the OK key to set the Delay and return to the menu screen, or ESC to cancel entering the Delay.
9) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
7.4 Gain
The gain, or amplification of the return echoes, can be adjusted over a wide range. The setting of the gain is crucial in order to obtain valid readings during the measurement process. Too much gain may result in erroneous measurements, by detecting on noise rather than the actual material back wall itself. Too little gain may result in detection on an undesirable section of the waveform. It may also result in lack of detection on internal flaws or other specific applications using pencil style transducers. The gain can easily be compared to the volume control of a home stereo system. If you turn it up too much, you can’t hear the music clearly. If it’s turned down too much, you can’t hear it at all.
The procedures to adjust the Gain are outlined below:
Adjusting the Gain using the Hot Menus
55
Page 60
1) Press the MEAS key once to activate measure menu items. Press the MEAS key multiple times to move right, and the ESC key multiple times to move left until the GAIN cell is highlighted.
2) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the highlighted value.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps 4 & 5 until the GAIN value is correctly displayed.
7) Press the OK key to return to the measurement screen, or ESC to cancel entering the GAIN.
The user can also access and adjust the gain from th e tabbed menus. However, this method is more tedious than making the adjustments using the Hot Menus. The
procedure using the tabbed menus is outlined below:
56
Adjusting the Gain using the Tabbed Menus
Page 61
PVX Precision Visual Thickness Gauge
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the TUNE menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until GAIN is highlighted.
3) Press the LEFT and RIGHT arrow keys to scroll the value. When the correct Gain is being displayed, proceed to step 8.
4) Alternatively, press the ENTER key to display the Digits Edit Box.
5) Press the UP and DOWN arrow keys to scroll the highlighted value.
6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
7) Repeat steps 5 & 6 until the Gain value is correctly displayed.
8) Press the OK key to set the Gain and return to the menu screen, or ESC to cancel entering the Gain.
9) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
57
Page 62
7.5 Threshold
The Threshold is the level (sensitivity) of the signal amplitude required to trigger the thickness reading. This level can be used in conjunction with the gain. Example:
suppose the user can visually see the waveform they need to detect on, but the PVX is not detecting because the Gain is too low, or the Threshold to high. The Threshold level can be decreased (lower sensitivity) in order to detect signals with lower amplitudes. This will allow the amplitude from the waveform to reach and exceed the Threshold level, and enable the PVX to detect on the proper return echo. This can also be accomplished by increasing the Gain, allowing the signal to exceed the current Threshold level. Lowering the Threshold (increases sensitivity), and raising the Threshold decreases the sensitivity respectively.
The procedures to adjust the Threshold are outlined below:
58
Page 63
PVX Precision Visual Thickness Gauge
Adjusting the Threshold using the Hot Menus
1) Press the MEAS key once to activate measure menu items. Press the MEAS key multiple times to move right and the ESC key multiple times to
move left, until the THR (threshold) cell is highlighted.
2) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the highlighted value.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps 4 & 5 until the THR value is correctly displayed.
7) Press the OK key to return to the measurement screen, or ESC to cancel entering the THR.
The user can also access and adjust the Threshold from the tabbed menus. However, this method is more tedious than making the adjustments using the Hot Menus. The procedure using the tabbed menus is outlined below:
59
Page 64
Adjusting the Threshold using the Tabbed Menus
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab
left, until the TUNE menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until THRESHOLD is highlighted.
3) Press the LEFT and RIGHT arrow keys to scroll the value. When the correct threshold is being displayed, proceed to step 8.
4) Alternatively, press the ENTER key to display the Digits Edit Box.
5) Press the UP and DOWN arrow keys to scroll the highlighted value.
6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
7) Repeat steps 5 & 6 until the Threshold number is correctly displayed.
8) Press the OK key to set the Threshold and return to the menu screen, or ESC to cancel entering the Threshold.
9) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
60
Page 65
PVX Precision Visual Thickness Gauge
7.6 Gates
The PVX is equipped with two gates that can be adjusted by the user to measure a specific region or area on and between waveforms. In pulse - echo, only a single gate can be used. The single gate has a threshold and a start point. I n echo – echo (interface-echo) mode, two gates are used. The same threshold level is associated with both gates. The first gate has a starting point and threshold, and the second gate has a start time that is determined by adding an adjustable HOLD OFF, or delay, to the return echo measured by the first gate. These features are described in more detail below:
Note: In order to adjust the gates they must be activated. The gates will only be active if the measurement mode is set to P-E GT (pulse – echo w/gate1), or E-E
(echo – echo/interface-echo w/ gates 1&2). If the gates are not active, they will be grayed out in the tabbed menus. Note: The setups contain the proper modes according specific transducer types. The explanations that follow assume that the proper measurement mode has been selected by selecting a specific setup. Refer to page 27 for additional information on selecting a setup for the transducer being used, which contains the correct measurement mode setting.
Gate 1
GATE 1 is used in both pulse-echo and echo-echo (interface-echo) modes. It has a
threshold and a start time. The purpose of GATE 1 is to prevent the PVX from measuring unwanted surface noise in pulse-echo mode. This is accomplished by
changing the start time of GATE 1. It may be advantageous to activate a gate when measuring aluminum, titanium, and stainless steel using lower frequency
transducers. However, if a high frequency transducer is being used, the gate may not be necessary. Aluminum sometimes has an inherent surface noise problem that can cause erroneous readings from the noise just following the initial pulse. By blocking the surface noise with GATE 1 this error is eliminated. GATE 1 is also used
in echo - echo and interface-echo mode to force the PVX to measure between the first and second back wall echoes or from the end of an interface (delay line), by moving the start to the first back wall reflection of the test material.
The procedures to adjust Gate 1 are outlined below:
61
Page 66
SURFACE NOISE PROBLEM
NO GATE W/GATE
The diagrams above illustrate a typical surface noise condition. Refer to the NO GATE diagram: (A) refers to the noise in front of the actual back wall signal (C).
Notice that without a gate activated, the PVX is detecting on the noise (A) as shown at point (B). However, the true measurement should be taken at point (C). Given the threshold and gain levels the PVX is currently set to in the NO GATE diagram, the amplitude from the noise is sufficient enough to cause the PVX to detect, or measure
the noise rather than the true back wall thickness (C). Therefore, the PVX is making an incorrect reading at point (B).
Now refer to the W/GATE diagram. The horizontal line at the top of (D), is GATE1. The start point of GATE1 has been adjusted to just beyond the noise (A) so that the PVX ignores the noise and detects the true back wall (C). Note: the PVX will only detect on signals that are located inside the dimensions of GATE1 (B). Therefore, the PVX cannot see (A) at all, with respect to the starting point of (B). Also notice, the THR (threshold) level is the height of the distance (D) from the baseline. Zero threshold is indicated by the bottom of the range (D), and THR: 6 (threshold) is indicated at the top of the range at (D). Therefore, the vertical height of GATE1 is the THR (threshold) level. The threshold level can be increased to decrease sensitivity, or decreased to increase sensitivity.
If the threshold level was increased in the NO GATE diagram, so that the level was higher than the amplitude of the noise (A), the PVX would have detected on the true back wall (C). Alternatively, if the gain level was decreased, the signal amplitude of
62
Page 67
PVX Precision Visual Thickness Gauge
the noise (A) would have decreased below the threshold level, and the PVX would have also detected the true back wall (C). This example brings all the fine
adjustments into consideration, and demonstrates the versatility of having fully functional scope rather than a basic digital thickness gauge.
Adjusting Gate1 using the Hot Menus
1) Press the MEAS key once to activate measure menu items. Press the MEAS key multiple times to move right and the ESC key multiple times to
move left until the GATE 1 cell is highlighted.
2) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the highlighted value.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps 4 & 5 until the GATE 1 value is correctly displayed.
7) Press the OK key to return to the measurement screen, or ESC to cancel entering the GATE 1 value.
63
Page 68
The user can also access and adjust the gate from the tabbed menus. However, this method is more tedious than making the adjustments using the Hot Menus. The
procedure using the tabbed menus is outlined below:
Adjusting Gate1 using the Tabbed menus
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab
left until the GATE menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until GATE is highlighted.
3) Press the LEFT and RIGHT arrow keys to scroll the value. When the correct gate is being displayed, proceed to step 8.
4) Alternatively, press the ENTER key to display the Digits Edit Box.
5) Press the UP and DOWN arrow keys to scroll the highlighted value.
6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
7) Repeat steps 5 & 6 until the gain number is correctly displayed.
8) Press the OK key to set the Gate 1 value and return to the menu screen, or ESC to cancel entering the Gate 1 value.
64
Page 69
PVX Precision Visual Thickness Gauge
9) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
Gate 2
The GATE 2 HOLDOFF is only used in the echo-echo (interface-echo) mode. Note: The PVX must be in E-E (echo-echo) mode to be activated. GATE 2 is sometimes called a Hold-Off or Delay. The function of the GATE HOLDOFF is to prevent the second gate from triggering on noise associated with ring down of the first echo. The GATE 2 HOLDOFF is visible as the distance from the end of GATE 1 at the first detect, to the start of GATE 2 as display by the horizontal bars on the display. A further explanation suggests the HOLDOFF is a time delay that simply extends the end of GATE 1 further out in time to prevent GATE 2 from starting. This is commonly used to avoid detecting interface noise or transducer ring down. The diagrams below provide a visual example of the gates, followed by the procedures to adjust the gates for different applications:
Adjusting Gate2 using the Hot Menus
1) Press the MEAS key once to activate measure menu items. Press the MEAS key multipl e times to move right and the ESC key multiple times to
move left, until the GATE2 cell is highlighted.
65
Page 70
2) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the highlighted value.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps 4 & 5 until the GATE2 value is correctly displayed.
7) Press the OK key to return to the measurement screen, or ESC to cancel entering the GATE 2 value.
Alternatively, Gate 2 can be selected from the menus and adjusted as follows:
Adjusting the Gate 2 from the Menu Items
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab
left until the GATE menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until GATE 2 is highlighted.
66
Page 71
PVX Precision Visual Thickness Gauge
3) Press the LEFT and RIGHT arrow keys to scroll the value. When the correct gate value is being displayed, proceed to step 8.
4) Alternatively, press the ENTER key to display the Digits Edit Box.
5) Press the UP and DOWN arrow keys to scroll the highlighted value.
6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
7) Repeat steps 5 & 6 until the GATE 2 value is correctly displayed.
8) Press the OK key to set the gate and return to the menu screen, or ESC to cancel entering the gate value.
9) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
67
Page 72
CHAPTER EIGHT THRU PAINT MEASUREMENT TECHNIQUE
8.1 Introduction to Thru Paint Measurement
The principle behind thru paint measurement is by measuring the time between two backwall echoes returning from the test material. Since both of these backwall echoes travel the same path through the paint or coating, the thickness of the coating is subtracted out of the measurement so that only the actual material thickness can be measured. This feature saves the user a great deal of time scraping and removing the coating from various test materials during the inspection process. The primary purpose of thru paint measurement is to determine actual or overall material thickness by eliminating the coating thickness.
ECHO – ECHO MODE (THRU PAINT ) – TWO GATES
Refer to the diagram above. This is a typical example of an echo-echo mode setup commonly used for a through paint or coatings application. Once again, the definition of an echo-echo measurement, is a measurement taken between to return echoes, rather than from the initial pulse to a single echo. Therefore, the measurement is taken between the first and second echo reflections. Point (C) represents the detection on the back wall of the first reflection. Notice that GATE1 has been has been set up so that the PVX can only see the first back wall reflection (C), and cannot see anything before the first reflection. Remember that the PVX can only see those signals inside the dimensions of the gate.
GATE2 (hold-off) has been set at point (B), so that the transducer ring down noise following the first detection (C) can’t interfere with the second back wall reflection (D). The distance from point (C), which is the end of the first gate (A), to the start of GATE2 (B) , is the hold-off time or delay. This type of setup allows the user to eliminate the paint or coating, and measure only metal the thickness of the metal.
8.2 Setting Up Thru Paint Mode
68
Page 73
PVX Precision Visual Thickness Gauge
The most straight forward way of setting up the PVX for thru paint measurements is to select a factory setup with a corresponding transducer type from the list of setups included with the PVX. These setups will serve as a good starting point for setting customized applications. However, fine adjustments may be necessary in order to be suitable for your specific applications. These configurations are general setups only, but have been setup and optimized for the majority of common measurement. Once a setup has been selected and the appropriate calibration procedure completed, the user can make any fine adjustments to the setup and begin taking measurements.
When configuring the PVX for specific thru paint applications, all of the scope parameters will potentially be needed. The delay, width, gain, threshold, gate1, and gate2 features will be subject to change. For this reason, they have all been added to Hot Menu fields located directly beneath the A-Scan display for quick and easy access. Note: Once the values of the fields have been changed or modified, these changes must be saved to a setup location prior to powering the PVX off. Failure to
do so will result in losing your changes.
THRU-PAINT (INCORRECT) THRU-PAINT (CORRECT)
Refer to the incorrect diagram above. Point (A) represents the detection on the first back wall echo. The true, second back wall, reflection should be detecting at point (D). However, the hold-off (C) is setup incorrectly and the PVX is detecting the ring down noise of the transducer, on the first reflection, rather than the true second back
wall reflection shown at (D). Before we look at the diagram with the correct configuration, let’s consider all of our
options on how to fix the problem beforehand. Our delay and width will simply change the view options of the screen – not needed in this example. Will a gain or threshold adjustment fix the problem? Unfortunately, not. Why? Notice the amplitude of the cycle just to the left of (B). If we tried to increase the threshold level above the height of the cycle, we would lose our detection for both echoes resulting
in no reading at all. If we decrease our gain, reducing the signal amplitudes, we would also lose our detection of both back wall echoes.
69
Page 74
Notice that the amplitude of the cycle just to the left of (C). The amplitude of the cycle, at (C), is less than the amplitude of the cycle at (B), in terms of overall height. If we decreased our gain, we would lose the first cycle (C) long before losing the second cycle (B). Therefore, adjusting the gain or threshold levels will not help us in this example. Our final option is adjusting the GATE2 hold -off, shown at point (C) in the incorrect diagram, to point (B) in the correct diagram. If we adjust the GATE2 hold-off further to the right of the first reflection, the ring down noise from the first back wall echo no longer interferes with the true detection (D), shown in the correct diagram.
The adjustment considerations in the example above will typically be used for all thru paint applications respectively. These examples are true of all pot ential echo-echo setups with the PVX. In some applications the hold-off may be sufficient, while a gain or threshold adjustment will solve the problem. A similar process of elimination should be considered for all thru paint and general echo-echo applications.
70
Page 75
CHAPTER NINE ADDITIONAL FEATURES OF THE PVX
9.1 Pulse Width
The PVX has an adjustable pulse width option, located as a submenu item in the PROBE tab group. Pulse width, refers to the duration of time the pulser is left on. This time results in increased energy sent into the test material. There are three width options (SPIKE, THIN, and WIDE). The SPIKE setting may be desirable for high resolution and general applications to decrease the overall noise. This can be considered the normal or standard setting. When additional energy is needed, more penetration, the THIN and WIDE options may be necessary. The procedure to change the pulse width is outlined below:
Selecting the Pulse Width
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the PROBE menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until PULSE is highlighted.
3) Use the LEFT and RIGHT arrow keys to scroll the Pulse options.
4) Once the Pulse is displayed, press the MEAS key to return to measurement
screen.
71
Page 76
9.2 Auto Find
The PVX is equipped with a very handy feature for locating the detection point of a back wall echo that is not currently in the viewable region of the display. Suppose the PVX is currently setup with a delay of 0.0” and a width of 0.500”, and the user is measuring material thickness of .625”. The PVX will make the correct measurement and display it digitally. However, the waveform will not show up at all in both A-Scan
views, and incorrectly in the B-Scan view without adjusting the range of the Delay and Width accordingly.
The AUTO FIND feature is a convenient way to let the PVX find the detection point and bring the waveform signal into view automatically. The user can then make
small adjustments to the range once the signal is displayed on the screen. The procedure to use the AUTO FIND feature is outlined below:
Using Auto Find
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the UTIL menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until AUTO FIND is highlighted.
3) Place the transducer on the test material and obtain a reading.
4) Press the ENTER key to automatically adjust the display settings and bring
the waveform into view.
5) Once the signal is displayed, press the MEAS key to return to measurement
mode.
72
Page 77
PVX Precision Visual Thickness Gauge
9.3 High Speed Scan
The High Speed Scan feature of the PVX increases the overall sample rate from the standard 4 measurements a second to 32 measurements a second. This feature enables a user to make scanned passes over an arbitrary length of the test material, while still maintaining a reasonable representation of thickness over the scanned area or region. The display will continue to be updated during the scan process, and when the transducer is removed, the PVX will display the minimum reading found. The procedure to use the SCAN MODE feature is outlined below:
Using the Scan Mode
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until the UTIL menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until SCAN MODE is highlighted.
3) Use the LEFT and RIGHT arrow keys to toggle the SCAN MODE on/off.
4) Press the MEAS key to return to the measurement screen.
9.4 Alarm Mode
The Alarm Mode feature of the PVX provides the user with a method of setting tolerances, or limits, for a particular application requirement. This feature may be used for incoming material inspections to verify the material received is within the manufacturer specifications. There are two limits, or alarm values, that can be setup
in the PVX – ALARM LO LIMIT and ALARM HI LIMIT. However, the user may choose to activate and utilize only one of the limit values, depending on their specific
application requirements. The procedures to use the ALARM MODE feature are outlined below:
73
Page 78
UTIL MENU – ALARM MODE
Toggle Alarm Status (on/off)
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the UTIL menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until ALARM STATUS is highlighted.
3) Use the LEFT and RIGHT arrow keys to toggle the ALARM STATUS
on/off/audible.
74
Page 79
PVX Precision Visual Thickness Gauge
4) Continue on to the next section “Setting the Alarm Lo Limit”.
Setting the Alarm LO Limit
1) Assuming the ALARM STATUS is ON, use the UP and DOWN arrow keys to scroll through the sub menu items until ALARM LO LIMIT is highlighted.
2) Press the LEFT and RIGHT arrow keys to scroll the value. When the correct alarm value is being displayed, proceed to step 7.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps 4 & 5 until the ALARM LO value is correctly displayed.
7) If only one limit will be used, press the MEAS key to return to the
measurement screen and begin taking readings. Otherwise, continue on to set the ALARM HI LIMIT.
75
Page 80
Setting the Alarm HI Limit
1) Assuming the ALARM STATUS is ON,use the UP and DOWN arrow keys to scroll through the sub menu items until ALARM HI LIMIT is highlighted.
2) Press the LEFT and RIGHT arrow keys to scroll the value. When the correct alarm value is being displayed, proceed to step 8.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps 4 & 5 until the ALARM HI value is correctly displayed.
7) Press the OK key to set the ALARM HI value and return to the menu
screen, or ESC to cancel entering the ALARM HI value.
8) Press the MEAS key to return to the measurement screen and begin taking
readings.
9.5 Polarity
The PVX is equipped with an option to select the polarity, or phase +/ -, for the purpose of detection. The phase is a very valuable feature to have when the signal returning from the test material is marginal, a low frequency transducer is being used, and the user has a very weak positive or negative cycle, while trying to measure very thick materials. There is a possibility that signal will become so weak that it falls below the threshold, and peak jumps to the next cycle altogether. This peak jump represents a shift of the detection further out in time, resulting incorrect measurements. This feature is also critical for thin measurements where accuracy is required. Once again, if the phase is not properly selected, a peak jump can occur making the test material appear to be thicker than it actually is. Refer to the diagram below:
76
Page 81
PVX Precision Visual Thickness Gauge
Notice the +/- positions in the diagram. The positive phase is everything above the horizontal center line, and the negative everything below the center line. The PVX uses a zero crossing (flank) method for detection. Therefore, the detect line (A)
is represented by the broken vertical dotted line, and is currently detecting on the negative portion of the waveform (B) at the zero crossing (C). The signal amplitude
(B) is shown as the negative vertical height at (B) in the diagram. Refer to the height of the second negative peak at (G). It’s clear that the amplitude of (G) is much greater than that of (B). Therefore, if the user was measuring thick attenuative material, and the amplitude of (B) decreased substantially, the PVX would lose the first cycle (B) and peak jump to the second cycle (F). The detect (A), would move to (F), resulting in incorrect measurements.
If the user were to select the positive phase in the diagram above, the detect would measure at (H). (A) would move to (H). Notice the height of (D) with respect to the
height of (E) in the diagram. If the user was measuring thick attenuative material using this phase, the signal (E) will certainly be lost long before (D). Therefore, the
positive phase is a much better choice in the diagram above.
Toggle Polarity (+/-)
Note: Before toggling the Polarity, the PVX should be set to the RF display
view option. The RF view will give the user the best opportunity to correctly
77
Page 82
view the positive and negative cycles of the waveform. Please refer to page 43 for information on selecting the Display Views.
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the TUNE menu is highlighted and displaying the submenu items.
2) Use th e UP and DOWN arrow keys to scroll through the sub menu items
until POLARITY is highlighted.
3) Use the LEFT and RIGHT arrow keys to toggle the POLARITY negative or
positive.
4) Press the MEAS key to return to the measurement screen and begin taking
readings.
Important: If a contact transducer is being used in pulse-echo mode, be sure to do a Probe Zero after change the polarity.
9.6 Rectified Waveform (Filled/unfilled)
The PVX has yet another feature primarily included for user personal viewing preference. When the PVX is in the Rectified Display View, the waveform can be
completely filled in black, or simply outlined. The procedures to toggle this feature, are outlined below:
Toggle Rect Wave (filled / Unfilled)
78
Page 83
PVX Precision Visual Thickness Gauge
Note: Before toggling the Rect Wave, the PVX should be set to the RECT display view option. Please refer to page 43 for information on selecting the
Display Views.
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the TUNE menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until RECT WAVEFORM is highlighted.
3) Use the LEFT and RIGHT arrow keys to toggle the RECT WAVEFORM
filled or unfilled.
4) Press the MEAS key to return to the measurement screen and begin taking
readings.
79
Page 84
9.7 Language Selection
The PVX is equipped with a multi language selection feature. This is an added feature for to accommodate everyone. The procedures are outlined below:
Selecting the Language using the Tabbed Menus
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab
left, until the SETUP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until LANGUAGE is highlighted.
3) Press the LEFT and RIGHT arrow keys to scroll the available language options.
4) Press the OK key to set the Language and return to the menu screen, or ESC to cancel selecting the Language.
5) Finally, press the MEAS key to return to the measurement screen and begin taking readings.
80
Page 85
10.1 Introduction to Grid (spreadsh eet) format
Data is stored in the PVX in a spreadsheet like format called a GRID. A GRID is simply a table of readings. The location in a grid is specified by giving a row and
column coordinate. The rows are numbered from 1 to 99 and the columns are labeled from A to Z, then AA to ZZ giving a maximum of 52 possible columns.
The largest grid that can be held in the PVX is limited by the total memory of the PVX. Multiple grids can be created and stored until the PVX’s memory is full. If the user attempts to store a grid in the PVX when the memory is full, the PVX will respond with an error message indicating that the memory is unable to store the new grid.
The PVX can store a total of 12,000 readings with a corresponding waveform stored with every reading. The waveform stored with the reading will depend on what view the PVX was in at the time it was saved. For example, If the view is RECT, then a rectified waveform will be saved with the reading.
10.2 Creating a new Grid
Important Note: This entire section is a step by step guide to successfully create a
grid. The instructions must be used in the sequential order specified, as follows:
Creating a Grid Name
81
Page 86
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the DATA menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until NEW is highlighted.
3) Press the ENTER key to display the new Grid Edit Box.
4) Use the UP and DOWN arrow keys to scroll through the new Grid List Items until NAME is highlighted.
5) Press the ENTER key to activate the Alpha Edit Box.
6) Use the UP, DOWN, LEFT, & RIGHT arrow keys to highlight the appropriate alpha characters.
7) Press the ENTER key to select a character and advance to the next field of the Grid Name.
8) Use the CLR key to backspace if necessary.
9) Repeat steps 6 - 8 until the Grid Name is completed.
82
Page 87
PVX Precision Visual Thickness Gauge
10) Press the OK key to save the Grid Name and return to the Grid List Items menu, or ESC to cancel entering the Grid Name.
Creating a Grid Note
1) Use the UP and DOWN arrow keys to scroll through the new Grid List Items until NOTE is highlighted.
2) Press the ENTER key to activate the Alpha Edit Box.
83
Page 88
3) Use the UP, DOWN, LEFT, & RIGHT arrow keys to highlight the appropriate alpha characters.
4) Press the ENTER key to select a character and advance to the next field of the Grid Note.
5) Use the CLR key to backspace if necessary.
6) Repeat steps 3 through 5 until the Grid Note is complet ed.
7) Press the OK key to save the Grid Note and return to the Grid List Items menu, or ESC to cancel entering the Grid Note.
Setting the Coordinates of the Grid
Note: A grid is defined by using coordinates to define the Top Left and the Bottom Right corners of the grid. Alpha coordinates are horizontal across the
top, and numeric coordinates are vertical down the side. Therefore, to define the top left corner of the grid, there will be an (X,Y) coordinate. Where X is an alpha column location across the top and Y is a numeric row location down the side. Use the same logic when choosing the lower right corner.
Setting the Top Left Coordinate
84
Page 89
PVX Precision Visual Thickness Gauge
1) Use the UP and DOWN arrow keys to scroll through the new Grid List Items until TOP LEFT is highlighted.
2) Press th e ENTER key to activate the Coordinate Edit Box.
3) Use the LEFT, & RIGHT arrow keys to scroll the Columns, and the UP, DOWN arrow keys to scroll the Rows.
4) Press the OK key to select the coordinate and return to the Grid List Items screen, or ESC to cancel the selection and return to the Grid List Items menu.
Setting the Lower Right Coordinate
85
Page 90
1) Use the UP and DOWN arrow keys to scroll through the new Grid List Items until LOWER RIGHT is highlighted.
2) Press the ENTER key to activate the Coordinate Edit Box.
3) Use the LEFT, & RIGHT arrow keys to scroll the Columns, and the UP, & DOWN arrow keys to scroll the Rows.
4) Press the OK key to select the coordinate and return to the Grid List Items
86
screen, or ESC to cancel the selection and return to the Grid List Items menu.
Page 91
PVX Precision Visual Thickness Gauge
Note: If there is not enough memory available to create the grid, an error message box “NOT ENOUGH MEMORY“ will be displayed. Press the OK or ESC key to return to the Grid List Items menu. It may be necessary to free some memory in the PVX at this time. Refer to page 92 for more information on Deleting a Grid.
87
Page 92
Selecting the Auto Increment Direction
The Auto Increment feature gives the user the ability to specify which direction to advance the cursor after storing a reading.
5) Use the UP and DOWN arrow keys to scroll through the new Grid List Items until INCR. DIR is highlighted.
6) Use the LEFT & RIGHT arrow keys to toggle the Increment direction NONE, NORTH, EAST, SOUTH, or WEST.
7) When the correct Increment direction is displayed, continue on to the next section “Saving the Grid”.
88
Page 93
PVX Precision Visual Thickness Gauge
Saving the Grid
Once all the parameters are set, the user has the option of saving or canceling the new grid.
1) Use the UP and DOWN arrow keys to scroll through the new Grid List Items until CREATE NEW GRID? is highlighted.
2) Press the ENTER key to accept the grid parameters, and activate the confirmation screen.
3) Press the OK key to save the New Grid, or the ESC key to cancel the New Grid setup and return to the DATA menu.
89
Page 94
4) Press the MEAS key to return to the measurement screen and begin storing readings.
10.3 Storing a reading
Now that a grid has been created, it’s time to make some measurements and store the readings. The following procedures outline this process:
Storing a Reading
1) Press the MEAS key once to activate measure menu items. Press the MEAS key multiple times to move right and the ESC key multiple times to
move left until the LOC cell is highlighted.
2) Press the ENTER key to display the Grid View Box.
3) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the target cell cursor to the desired storage location.
90
Page 95
PVX Precision Visual Thickness Gauge
5) Press the ENTER key to save the current reading in the highlighted cell location.
6) Abort the Grid View Box by pressing the MEAS key at any time.
Note: Every time a reading is saved to memory, a picture of the display is automatically saved with the reading.
10.4 Viewing stored readings and waveforms
It is sometimes necessary to go back and view the stored readings and waveforms using the PVX without a PC. The following procedures outline this process:
Viewing Stored Readings & Waveforms
1) Press the MEAS key once to activate measure menu items. Press the MEAS key multiple times to move right and the ESC key multiple times to
move left until the LOC cell is highlighted.
2) Press the ENTER key to display the Grid View Box.
3) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the stored readings and corresponding display view. Notice as the cursor is moved to
91
Page 96
a different cell, the display will be updated with the display view saved with the reading.
Note: The user can also toggle the display view between RF, RECT, DIGiTS if previously saved in one of the listed modes, or B-SCAN and DIGITS if saved in B-Scan Mode, by pressing the ESC and OK keys while a location is highlighted.
4) The user may opt to clear a specific reading and save a new one at any time. Press the CLR key in the appropriate cell location to clear the reading, take a new measurement, and press the OK key to save the new reading.
5) Abort the Grid View Box by pressin g the MEAS key at any time.
10.5 Deleting a grid
Deleting One or All Grids from Memory
Deleting One Grid
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the DATA menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
92
until DELETE ONE GRID is highlighted.
Page 97
PVX Precision Visual Thickness Gauge
3) Press the ENTER key to display the Grid List Box.
4) Use the UP and DOWN arrow keys to scroll through the stored Grids until the target Grid to delete is highlighted.
5) Press the OK key to delete the Grid.
6) Press the MEAS key to return to the measurement screen.
Deleting All Grids
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the DATA menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until DELETE ALL GRIDS is highlighted.
93
Page 98
3) Press the ENTER key to activate the confirmation screen.
4) Press the OK key to delete All Grids from memory, or the ESC key to abort.
5) Press the MEAS key to return to the measurement screen.
10.6 Editing a grid
Once a grid has been created and saved to memory, the user can edit the Comments or Increment Direction at a later time. The following procedures outline this process:
Editing a Grid
94
Page 99
PVX Precision Visual Thickness Gauge
1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the DATA menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items until EDIT is highlighted.
3) Press the ENTER key to display the Edit List Box.
4) Use the UP and DOWN arrow keys to scroll through the Edit options until
NOTE or INCR. DIR is highlighted.
Note: If editing the INCR. DIR, simply use the LEFT or RIGHT arrow keys to
scroll NONE, NORTH, EAST, SOUTH, or WEST. Proceed to step 10.
5) Press the ENTER key to activate the Alpha Edit box – Only used when editing the NOTE.
6) Use the UP, DOWN, LEFT, & RIGHT arrow keys to highlight the appropriate alpha characters.
7) Press the ENTER key to select a character and advance to the next field of the Comments.
8) Use the CLR key to backspace if necessary.
95
Page 100
9) Repeat steps 6 - 8 until the Comments are completed.
10) Press the UP or DOWN arrow key to highlight SAVE CHANGES, and the OK key to activate the confirmation screen.
11) Press the OK to save the changes or the ESC key to cancel editing the grid parameters.
12) Press the MEAS key to return to the measurement screen.
10.7 Changing the active grid (Open)
96
Loading...