Copyright 2008 Dakota Ultrasonics. All rights reserved.
No part of this publication may be reproduced, translated into another
language, stored in a retrieval system, or transmitted in any form or by any
means; electronic, mechanical, photocopying, recording, or otherwise,
without the prior written consent of Dakota Ultrasonics.
Every precaution has been taken in the preparation of this publication.
Dakota Ultrasonics assumes no responsibility for errors or omissions.
Neither is any liability assumed for damages resulting from the use of
information contained herein.
Any brand or product names mentioned herein are used for identification
purposes only, and are trademarks or registered trademarks of their
respective holders.
1500 Green Hills Road, #107 Scotts Valley, CA 95066 USA
Tel (831) 431-9722 Fax (831) 431-9723
http://www.dakotaultrasonics.com
MMX-6 DL Ultrasonic Thickness Gauge
CONTENTS
INTRODU CTION
OPERATION
THE KEYPAD
THE DISPLAY
THE TRANSDUCER
MAKING MEASUREMENTS
CONDITION AND PREPARATION OF SURFACES
PROBE ZERO
CALIBRATION
UNITS,BACKLIGHT,SCAN MODE,ALARM MODE,DUAL-
MULTI MODE,BEEPER,RS232PORT &DATA LOGGER
TRANSDUCER SELECTION
APPENDIX A:PRODUCT SPECIFICATIONS
1
3
3
7
10
11
13
14
15
22
34
38
APPENDIX B:APPLICATION NOTES
APPENDIX C:SOUND VELOCITIES OF COMMON MATERIALS
WARRANTY INFORMATION
39
43
45
Dakota Ultrasonics
DISCLAIMER – Very Important!
Inherent in ultrasonic thickness measurement is the possibility that the
instrument will use the second rather than the first echo from the back
surface of the material being measured while in standard pulse-echo mode.
This may result in a thickness reading that is TWICE what it should be.
In addition, measurements through very thick paint or coatings while
using echo-echo mode, may result in the paint or coating being measured
rather than the actual material intended. The Responsibility for proper use
of the instrument and recognition of these types of phenomenon’s rest
solely with the user of the instrument.
MMX-6 DL Ultrasonic Thickness Gauge
1
INTRODUCTION
The Dakota Ultrasonics model MMX-6 DL is a precision Ultrasonic
Micrometer. Based on the same operating principles as SONAR, the
MMX- 6 DL is capable of measuring the thickness of various materials with
accuracy as high as ± 0.001 inches, or ± 0.01 millimeters. The principle
advantage of ultrasonic measurement over traditional methods is that
ultrasonic measurements can be performed with access to only one side of
the material being measured.
This manual is presented in three sections. The first section covers
operation of the MMX- 6 DL, and explains the keypad controls and display.
The second section provides guidelines in selecting a transducer for a
specific application. The last section provides application notes and a table
of sound vel ocity values for various materials.
Dakota Ultrasonics maintains a customer support resource in order to
assist users with questions or difficulties not covered in this manual.
Customer support may be reached at any of the following:
Dakota Ultrasonics, 1500 Green Hills Road, #107
Scotts Valley, CA 95066 USA
The MMX-6 DL interacts with the operator through the membrane
keypad and the LCD display. The functions of the various keys on the
keypad are detailed below, followed by an explanation of the display and its
various symbols.
The Keypad
This key is used to turn the MMX-6 DL on and off. When the tool is
turned ON, it will first perform a brief displ ay test by illuminating all of the
segments in the display. After one second, the tool will display the internal
software version number and the current file location and status. After
displaying the version number, the display will show "0.000" (or "0.00" if
using metric units), indicating the tool is ready for use.
The MMX-6 DL is turned OFF by pressing the ON/OFF key. The tool
has a special memory that retains all of its settings even when the power is
off. The tool also features an auto-powerdown mode designed to conserve
battery life. If the tool is idle for 5 minutes, it will turn itself off.
Dakota Ultrasonics
4
The PRB- 0 key is used to "zero" the MMX- 6 DL in much the same way
that a mechanical micrometer is zeroed. If the tool is not zeroed correctly,
all of the measurements that the tool makes may be in error by some fixed
value. Refer to page 14 for an explanation of this important procedure.
The CAL key is used to enter and exit the MMX- 6 DL's calibration
mode. This mode is used to adjust the sound-velocity value that the MMX-6 DL will use when calculating thickness. The tool will either calculate the
sound-velocity from a sample of the material being measured, or allow a
known velocity value to be entered directly. Ref er to page 15 for an
explanation of the two CAL functions available.
The MODE key is used to toggle through the various features and
settings of the MMX-6 DL (gate, alarm mode, beeper, back light, units,
scan mode, and differential mode ). The MODE key is used in conjunction
with the arrow and send keys to enable/disable the features and settings.
MMX-6 DL Ultrasonic Thickness Gauge
5
The UP arrow key has three functions. When the MMX- 6 DL is in
calibration mode, this key is used to increase numeric values on the
display. An auto-repeat function is built in, so that when the key is held
down, numeric values will increment at an increasing rate. When MODE is
activated, the UP arrow key scrolls through the various features and
settings of the MMX-6 DL. When the data logging feature has been
activated by pressing the MEM key, the UP arrow is used to scroll through
the various files, storage locations, and functions of the data logger. Refer
to page 28 for further information regarding the use of the UP arrow key
and the data logger.
The DOWN arrow key has three functions. When the MMX- 6 DL is in
the CAL mode, this key is used to decrease numeric values on the display.
An auto-repeat function is built in, so that when the key is held down,
numeric values will decrement at an increasing rate. When MODE is
activated, the DOWN arrow scrolls through the various features and
settings of the MMX -6 DL. When the data logging feature has been
activated by pressing the MEM key , the UP arrow is used to scroll through
the various files, storage locations, and functions of the data logger. Refer
to page 28 for further information regarding the use of the DOWN arrow
key and the data logger.
Dakota Ultrasonics
6
The MEM key enables/disables the data logging feature of the MMX-6
DL. This key is used in conjunction with the UP/DOWN arrows, SEND,
AND CLR keys (hi-lighted in green). The combination of these keys
control the data logging features of the MMX-6 DL. Refer to the section on
data logging page 28.
The CLR key is specifically used with the data logging feature of the MMX- 6 DL. This key clears the contents of an entire file, or individual
storage locations. The CLR key is also used to send an obstruct (ObSt) to
an individual storage location. The ObSt symbol would indicate that a the
user was unable to take a reading at a particular location. Refer to the
section on data logging page 28.
The SEND key is used for sending data to internal storage locations,
and external peripheral devices ( serial printer / computer ). The SEND key
is also used to select data logging functions in the MMX- 6 DL page 28.
MMX-6 DL Ultrasonic Thickness Gauge
7
The Display
The numeric portion of the display consists of 4 complete digits
preceded by a leading "1", and is used to display numeric values, as well
as occasional simple words, to indicate the status of various settings.
When the MMX- 6 DL is displaying thickness measurements, the display
will hold the last value measured, until a new measurement is made.
Additionally, when the battery voltage is low, the entire display will begin to
flash. When this occurs, the batteries should be replaced.
These eight vertical bars form the Stability Indicator. When the MMX- 6 DL is idle, only the left -most bar and the underline will be on. While the
gauge is taking a measurement, six or seven of the bars should be on. If
fewer than five bars are on, the MMX-6 DL is having difficulty achieving a
stable measurement, and the thickness value displayed will most likely be
erroneous.
Dakota Ultrasonics
8
When the IN symbol is on, the MMX- 6 DL is displaying a thickness
value in inches. The maximum thickness that can be displayed is 19.999
inches.
When the MM symbol is on, the MMX- 6 DL is displaying a thickness
value in millimeters. If the displayed thickness exceeds 199.99 millimeters,
the decimal point will shift automatically to the right, allowing values up to
1999.9 millimeters to be displayed.
When the IN symbol is on, in conjunction with the /µs symbol, the MMX-6 DL is displaying a sound-velocity value in inches-per-microsecond.
When the M symbol is on, in conjunction with the /s symbol, the MMX-6 DL is displaying a sound-velocity value in meters- per-second.
MMX-6 DL Ultrasonic Thickness Gauge
9
When the + symbol is on and blinking, this indicates that the MMX- 6 DL
is currently operating in echo- echo ( Thru-paint/coating ) mode.
Dakota Ultrasonics
10
The Transducer
The transducer is the "business end" of the MMX- 6 DL. It transmits and
receives ultrasonic sound waves that the MMX-6 DL uses to calculate the
thickness of the material being measured. The transducer connects to the
MMX- 6 DL via the attached cable, and two coaxial connectors. When
using transducers manufactured by Dakota Ultrasonics, the orientation of
the dual coaxial connectors is not critical: either plug may be fitted to either
socket in the MMX- 6 DL.
The transducer must be used correctly in order for the MMX-6 DL to
produce accurate, reliable measurements. Below is a short description of
the transducer, followed by instructions for its use.
This is a bottom view of a typical transducer. The two semicircles of the
wearface are visible, as is the barrier separating them. One of the
semicircles is responsible for conducting ultrasonic sound into the material
being measured, and the other semicircle is responsible for conducting the
echoed sound back into the transducer. When the transducer is placed
against the material being measured, it is the area directly beneath the
center of the wearface that is being measured.
MMX-6 DL Ultrasonic Thickness Gauge
11
This is a top view of a typical transducer. Press against the top with the
thumb or index finger to hold the transducer in place. Moderate pressure is
sufficient, as it is only necessary to keep the transducer stationary, and the
wearface seated flat against the surface of the material being measured.
Making Measurements
In order for the transducer to do its job, there must be no air gaps
between the wear-face and the surface of the material being measured.
This is accomplished with the use of a "coupling" fluid, commonly called
"couplant". This fluid serves to "couple", or transfer, the ultrasonic sound
waves from the transducer, into the material, and back again. Before
attempting to make a measu rement, a small amount of couplant should be
applied to the surface of the material being measured. Typically, a single
droplet of couplant is sufficient.
After applying couplant, press the transducer (wearface down) firmly
against the area to be measured. The Stability Indicator should have six or
seven bars darkened, and a number should appear in the display. If the
MMX- 6 DL has been properly "zeroed" (see page 14) and set to the correct
sound velocity (see page 15), the number in the display will indicate the
actual thickness of the material directly beneath the transducer.
Dakota Ultrasonics
12
If the Stability Indicator has fewer than five bars darkened, or the
numbers on the display seem erratic, first check to make sure that there is
an adequate film of couplant beneath the transducer, and that the
transducer is seated flat against the material. If the condition persists, it
may be necessary to select a different transducer (size or frequency) for
the material being measured. See page 30 for information on transducer
selection.
While the transducer is in contact with the material that is being
measured, the MMX-6 DL will perform four measurements every second,
updating its display as it does so. When the transducer is removed from
the surface, the display will hold the last measurement made.
IMPORTANT
Occasionally, a small film of couplant will be drawn out between the
transducer and the surface as the transducer is removed. When this
happens, the MMX- 6 DL may perform a measurement through this
couplant film, resulting in a measurement that is larger or smaller than it
should be. This phenomenon is obvious when one thickness value is
observed while the transducer is in place, and another val ue is observed
after the transducer is removed.
In addition, measurements through very thick paint or coatings may
result in the paint or coating being measured rather than the actual material
intended. The responsibility for proper use of the instrument, and
recognition of these types of phenomenon’s, rest solely with the user of the
instrument.
Loading...
+ 35 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.