COVIDIEN 3000 AD-ADA Electronic Thermometer User Manual

Operating Manual
TM
Filac
3000 AD/ADA Electronic Thermometer
Table of Contents
Operating Manual
Filac
3000 AD/ADA Electronic Thermometer
en
Table of Contents
Section Page
I. About Body Temperatures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
II. General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
III. Safety and Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
IV. Icon Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
V. Instructions For Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Probe Covers — Applying & Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Changing Isolation Chambers and Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Oral & Axillary Temperature Taking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Rectal Temperature Taking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Direct Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Quick Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Cold Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Recall Last Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Pulse Timer Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Bio-Tech Mode (Option Configuration Menu) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
VI. Cleaning and Sterilization. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
VII. Battery Replacement / Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
VIII. Mounting Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
IX. Anti-Theft Features & Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
X. Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
XI. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
XII. Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
XIII. Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
XIV. Electromagnetic Conformity Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
This product contains software solely owned by Covidien. Covidien grants the user a non-exclusive, limited license to use the software according to the operating instructions. A copy of the license can be obtained from Covidien.
Filac
3000 AD/ADA Electronic Thermometer
I. About Body Temperature
It is a common myth that 98.6°F (37°C) is the “normal body temperature.” The truth is that 98.6°F (37°C) is an average body temperature. Normal body temperature is actually a range that varies with age, gender, and measurement site. Body temperatures also fluctuate through out the day, typically cooler temperatures in the morning, warmer in the afternoon, and cooling down again in the evening. Other factors that can influence body temperature are: the patient’s recent level of activity, metabolism rate, or medications. Normal body temperatures also tend to decrease with age.
Please see the chart below for normal temperature ranges by patient age and site. Readings from different body sites, even when taken at the same time, should not be directly compared; body temperatures will vary by site.
Temperature Site
Ear
Oral
Core
Rectal
Axillary
References:
Chamberlain, J, Terndrup, T., et al. (1995) Determination of Normal Ear Temperature with an Infrared Emission Detection Thermometer. Annals of Emergency Medicine, Volume 25, Issue 1, Pages 15-20
Braim. S., Preston, P., and Smith, R. (1988) Getting a better read on thermometry. RN. 1998 Mar; 61(3):57-60
Munk, P, Woods, S, Leduc, D., et al Canadian Paediatric Society Statement; Temperature Measurement in Paediatrics; Paediatric Child Health Vol 5 No 5: July/ August, 2000
Brunner, L. and Suddarth, Dl, et al. (1982) The Lippincott Manual of Nursing Practice, Third Edition; J. B. Lippincott Company, Philadelphia, PA; 1982; p.1145
Houdas, Y. and Ring, E. F. J. Human Body Temperature, Its Measurement and Regulation; Plenum Press, NY, 1982; p.81-87.
Normal Body Temperatures by Patients Age
0-2 Years
97.5°- 100.4°F
36.3°- 38.0°C
-
-
97.5° - 100.0°F
36.3°- 37.7°C
97.9° - 100.4°F
36.6°- 38.0°C
94.5° - 99.1°F
34.7°- 37.2°C
3-10 Years
97.0°- 100.0°F
36.1°- 37.7°C
95.9° - 99.5°F
35.3°- 37.5°C
97.5° - 100.0°F
36.3°- 37.7°C
97.9°- 100.4°F
36.5°- 38.0°C
96.6° - 98.0°F
35.8°- 36.6°C
11-65 Years
96.6° - 99.7°F
35.8°- 37.6°C
97.6° - 99.6°F
36.7°- 37.8°C
98.2° - 100.2°F
36.7°- 37.8°C
98.6° - 100.6°F
37.0°- 38.1°C
95.3° - 98.4°F
35.1°- 36.8°C
> 65 Years
96.4° - 99.5°F
35.7°- 37.5°C
96.4° - 98.5°F
35.7°- 36.9°C
96.6° - 98.8°F
35.8°- 37.1°C
97.1° - 99.2°
36.1°- 37.3°C
96.0° - 97.4°F
35.5°- 36.3°C
II. General Information
1. The Filac 3000 AD/ADA Electronic Thermometer is a fast, highly accurate, and easy to use clinical instrument for measuring patient temperatures by Oral, Axillary or Rectal means.
2. The electromagnetic compatibility of this device has been verified by test according to the EN60601-1-2:2001 requirements.
3. This device requires no user maintenance other than periodic cleaning and replacement of expired batteries.
Features
• Fasttemperaturemeasurements
• Measurementrange:30°Cto43°C(86°Fto109°F)
• PushbuttonalternatesCelsius/Fahrenheitscale
• Audibleindicationofcompletion(“beep”sound)
• Interchangeable,color-codedisolationchamberswithmatchingprobes.
Helps keep probe covers and probes together for aiding in infection control (Blue -oral/axillary; Red -rectal)
• Easilywipedclean
• Usesfourstandarddisposable“AA”batteries
• “Auto-On”whenprobeiswithdrawnfromtheprobewell
• “Auto-Off”conservesbatterylife
• LowBatteryandDeadBatteryindicators
• LasttemperatureRecall
• 15,30,45and60secondpulseTimer
• Easilyaccessibleprobecoverbox
• EasytoreadLCDdisplaywithbacklight
• ColdMode–lowerpre-heattemperatureoptionprovidesquicker,accuratereadingsforpost-ORpatients
• Theftpreventionlockoutandanti-theftaccessories
Filac
3000 AD/ADA Electronic Thermometer
Table of Contents
1
III. Safety and Warnings
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
1. Read this booklet thoroughly before using the Filac 3000 AD/ADA Electronic Thermometer.
2. Do not use this device near flammable anesthetics. Not suitable for use in the presence of flammable anesthetic mixture with air, oxygen or nitrous oxide, or in an oxygen rich environment.
3. Do not use this thermometer without first installing a new Filac 3000 AD/ADA Electronic Thermometer probe cover.
4. Use only Filac 3000 AD/ADA Electronic Thermometer probe covers with this device. Use of any other probe cover will result in erroneous temperature readings.
5. The device and probe covers are Non-sterile. Do not use on abraded tissue.
6. To limit cross contamination, use Blue devices for Oral and Axillary temperature taking only.
7. Use
8. Thoroughly dry all electrical contacts on both probe and thermometer after washing, or device may fail to function properly.
9. For re-calibration, service or integrity checks, refer to a qualified Biomedical Technician or return to manufacturer.
10.
11. Disposal of used probe covers must be performed in accordance with current medical practices or local regulations regarding
12. Do not use lithium batteries with this device.
13. Do not mix alkaline and rechargeable batteries.
14. Removal of the batteries is recommended if the unit is not going to be used for an extended period of time.
15. Dispose of batteries in a manner consistent with local environmental and institutional policy for Alkaline batterydisposal.
16. Dispose of old battery-powered electronic equipment in a manner consistent with institutional policy for expired
17. Cleaning frequency and practices must be consistent with institutional policy for cleaning of non-sterile devices.
18. Device to be used by trained personnel.
red
devices only for
Warning:
may affect calibration and voids warranty.
disposal of infectious, biological medical waste.
equipmentdisposal.
No modification of this equipment is allowed. Do not open unit. No user-serviceable parts inside. Opening of device
rectal
temperatures.
Note: Even though this device has been designed to minimize the effects of electromagnetic interference, it does generate radio frequency
energy. If not used in accordance with the instructions, the device could cause interference in other equipment operating within its vicinity. If the device is causing interference, the following actions may be taken in an attempt to correct the interference:
• Re-orientorre-locatethereceivingdevice.
• Increasetheseparationbetweenthedevices.
• Consultacustomerservicerepresentative.
Filac
3000 AD/ADA Electronic Thermometer
Table of Contents
2
IV. Icon Identification
ORAL Temperature Mode
Works only with BLUE probe and isolation chamber
AXILLARY Temperature Mode
Works only with BLUE probe and isolation chamber
RECTAL Temperature Mode
Works only with RED probe and isolation chamber
Site Button
Use to change measurementlocation
°C/°F Button
Use to change temperaturescale
Pulse Timer Button
Use to start and stop pulsetimer
Low Battery
Batteryvoltageislow–replace
batteries as soon as possible
Dead Battery
Unit will not operate until new batteries are installed
Progress Bar
Temperature measurement status indication
Return to Base
Return unit to Anti-theft base to clear anti-theft lock outfeature
Out of Range - High
Temperature measurement is above 109°F (43°C)
Out of Range - Low
Temperature measurement is below 86°F (30°C)
10%
-13°F
-25°C
Follow instructions for use -
This manual must be read prior to use of the thermometer (Symbol appears blue ondevice)
95%
Humidity limitations
131°F 55°C
Temperature limitations
Keep away from sunlight
Keep dry
EN 50419
Dispose of as Electrical and Electronic waste
Pulse Timer
15 or 30 second timer
Quick Mode
Fast predictive measurement
mode–displaystemperature
measurement in 3.5 seconds
Cold Mode
Post-operative mode. Pre-heat temperature of 91°F (33°C)
Direct Mode
Non-predictive temperature measurement mode
Install/Remove Probe Cover
Flashing probe cover reminder to install probe cover prior to use and remove cover after use
Authorized representative in the European Community
Type BF Equipment
Degree of protection against
Non-Sterile
Latex-free
Single use
By prescription only
electricalshock–no conductive connection to the patient
Non-ionizing electromagnetic radiation
Medical Electrical Equipment
CE mark
0123
DEHP-free
Manufacturer Date of Manufacture
Manufacturer Serial Number
Filac
3000 AD/ADA Electronic Thermometer
Table of Contents
3
Loading...