COSA Xentaur SF6 Purity Pro User Manual

Optical Gas Monitor
Model RK60000
COSA SF6 Purity Pro
Instruction Manual
84G Horseblock Rd
Yaphank, NY 11980
(631) 345-3434 Fax: (631) 924-7337
7125 North Loop East
Houston, TX 77028
(713) 947-9591 Fax.: (713) 947-7549
email: sales@cosaxentaur.com
www.cosaxentaur.com
1
2
Introduction
*Note: Advice on Usage
Thank you for purchasing the Model RK60000. This instrument is a portable gas monitor to measure “Helium in N2,” “CH4 in air,” and other gas concentration and calorie (valid only to the measurement mentioned in the attached measuring gas specifications) intermittently.
This instruction manual is a guidebook on how to operate the Model RK60000. Both beginner users and experienced users must read this manual and understand the contents thoroughly prior to operation.
This instrument is to be used solely for the purposes mentioned in this manual and the attached measuring gas specifications.
For your safety, the following marks have been included in this manual.
Danger
The liability or exposure to harm or injury.
Warning
If the instrument is not operated accordingly, it may serve as a serious hazard or health risk.
Caution
If the instrument is not operated accordingly, it may malfunction or serve as a serious hazard.
As the kinds of measurable gas and range vary depending on the type of RK60000, refer to the attached measuring gas specification. For standard models, refer to the attached “each TYPE measuring gas specification.”
3
Index
1. PARTS AND PART FUNCTIONS
2. MEASURING MODE (POWER KEY)
3. SETTING MODE (ENTER + POWER)
4. MAINTENANCE
5. DISPOSING OF THE INSTRUMENT
6. TROUBLE SHOOTING
7. WARNINGS ON USAGE
8. DEFINITION OF TERMS
9. SPECIFICATIONS
10. MEASURING PRINCIPLE
11. WARRANTY
2-1. Basic Display for Measuring Mode and Explanation………………………….6 2-2. Procedures from Power ON to Measurement……………………………………7 2-3. Saving Data………………………………………………………………………………9 2-4. Initial Display (Self-Diagnostic Display)………………………………………….9 2-5. ERROR Message Display……………………………………………………………10 2-6. AIR CAL. CAUTION…………………………………………………………………….11 2-7.Affection on Measurement from Pressure………………………………………12
3-1. Changing the Measuring Gas (SELECT GAS)…………………………………..14 3-2. Adjusting the Time (SET DATE / TIME)…………………………………………..15 3-3. Confirming the Saved Data (VIEW SAVED DATA)……………………………..15 3-4. Clearing the Saved Data (CLEAR SAVED DATA)…………………….…………15 3-5. Start the Measurement (START MEAS.)…………………………………………15
4-1.Replacing Batteries……………………………………………..……………………16 4-2. Confirming Sensitivities…………………………………………………………….16 4-3. Daily Check………………………………………………………………………..…..16 4-4. Frequency / Standard for Replacing Parts…………………………………….17
9-1. Specifications………………………………………………………….………………21 9-2. Standard Accessories……………………………………………………………….21
……………………………………………………………………………………..14
…………………………………………………………..…………………18
…………………………………………………………………………..19
………………………………………………………………………..…20
…………………………………………………………………………………..21
………………………………………………………………………………………..23
…………………………………………………………………..4
………………………………………………………………6
……………………………………………………………14
…………………………………………………….………17
……………………………………………………………………….22
4
1. PARTS AND PART FUNCTIONS
(1)
POWER key:
enter the “SETTING MODE.”
(2)
LCD Display: (3)
LED for “SAMPLING”: The LED is on while the internal pump is working.
(4)
PUMP key (CANCEL) :
Contents selected by the keys are cancelled by this key.
(5)
SAVE key ( key) :
(This key is used to enter numbers or to move the cursor.)
(6)
RANGE key ( key):
(This key is used to enter numbers or to move the cursor.)
(7)
AIR CAL. key (ENTER key): (8)
Buzzer
place.]
(9)
There is a data logger interface under the cover.
TOP
Press this key to turn the power on until the buzzer beeps. Keep pressing the key for approximately three seconds to turn the power off. When pressing the POWER key and (7) ENTER key, you can
Concentration is displayed. The time, battery remainder, and measures of gas and base gas are displayed as well.
When this key is pressed, the internal pump starts to work. When the key is pressed again, the pump stops. (Pump is automatically turned OFF after three minutes of operation to conserve
energy.)
Information about the Time/Date, Measuring Gas, base Gas, and Displayed Concentration is saved by this key. The maximum number of data points is 100.
Position of the decimal point on the concentration display can be moved by this key.
When calibration is performed with standard gas, this key is used. Contents selected by the keys are fixed by this key.
Beeps when the power is turned on or off, or when special control or incorrect operation takes
5
***Explanations in the ( ) are for use during SETTING MODE.
(10) GAS IN
Inlet for measuring gas. A specific absorption tube must be connected to the
(11) REF. OUT
Outlet connected to the reference gas chamber. The exhaust from the outlet
(12) GAS OUT
Outlet for measuring gas sucked from the GAS IN.
(13) DC6V
6VDC Interface for the specific AC adaptor.
(14) REC.
A jack for 0-1VDC output. (cable : option)
(15) Battery Cover
Cover for the batter box.
(16) Screw
Fixing screw for the battery cover. It can be opened or closed by a screwdriver
inlet, or the reading may go wrong.
Tubing Structure
must be emitted into the fresh air at atmospheric pressure.
or coin.
6
2. MEASURING MODE (POWER KEY)
Display Contents
Status
Gas Name and Base Gas
Standard measuring mode.
CAUTION CHECK AIR CAL.
Zero drifting might be occurring. Let the monitor sample fresh air and perform AIR CAL.
CAUTION ABNORMAL TEMP.
Internal thermistor detects an abnormal temperature.
CAUTION LOW BATT.
Battery voltage is low. Replace the batteries.
CAUTION LOW CONTRASUT
The interferometer sensor is getting contaminated. Before measuring, the sensor should be replaced or repaired.
CAUTION
The brightness is getting low. Before measuring, the sensor
possible.
2-1. Basic Display for Measuring Mode and Explanati on
LOW BRIGHTNESS
should be replaced or repaired.
Caution
When the message “CAUTION ~ ” appears on the LCD, make an appropriate adjustment as soon as
7
2-2. Procedures from Power ON to Measurement
selected properly. Confirm if [Measuring Gas] and [base gas] are the ones selected.
3) Press the PUMP Key and let it sample-draw enough
the buzzer beeps. “AIR CAL value” will then be displayed.
1) The power is turned on by pressing the POWER key until the buzzer beeps. The initial display (self-diagnostic display) will show up as well. A display that confirms the setting will then appear. After two seconds, it will proceed to the basic display of measuring mode.
Self-Diagnostic Display Confirmation Display
Caution
This instrument does not indicate a correct reading if the measuring gas and base gas are not
2) Two of the messages, [Meas Gas/Base Gas] and [Air CAL], are displayed as seen below. (Refer to WARNIN AIR CAL.)
reference gas (fresh air) into the instrument for about two minutes (the color of the display reverses while the pump is operating.)
4) After the indication is stable, press the PUMP key for the pump to stop operating. Press the AIR CAL key until
5) After the AIR CAL is
8
completed, press the PUMP key and let the sample gas draw into the
After the indication on the reverse display becomes stable, the pump will stop the operation and read out the indication as a measuring result at this pump operation stop. At this pump operation, stop under the gas flow ceasing condition. This is the measuring result.
measurement cannot be achieved.
*Note
reading. Read the indication after the gas flow stops.
*Note
= ______
instrument. The color of the display will reverse while the pump is operating. *Note that the indication on display under pump operation is not measuring result.
6)
Caution
Press the AIR CAL key after confirming that there is enough fresh air in the chamber, or else the correct
If the [AIR] is selected as the base gas, the indication will show a zero when the AIR CAL key is pressed. However, if the [N2] or other gases except [AIR] are selected as the base gas, the indication will not show a zero, bur rather other values determined based on the combination of the measuring gas and the base gas. This is caused by sampling the gas which is not on target (in this case, the air.) Note that the instrument is not broken. Once it samples the target gas, it shows a correct reading. Depending on the combination of the measuring gas and the base gas, the AIR CAL value may be a negative value. For further information, refer to the measuring gas specification.
Caution
RK60000 performs best when both the Gas Chamber and Reference Chamber are at the atmospheric pressure (101.3 kPa.) While the measuring gas is being sampled, the pressure in the Gas Chamber will become slightly different from the atmospheric pressure, and the analyzer will not show the correct
RK60000 measures the gas best when both the Gas Chamber and Reference Chamber are at
101.3 kPa. If precise measurement is required, perform the pressure correction by the following method:
Pressure Correction
Pressure at measurement [kPa]
101.3 kPa______ X (Reading – Air Cal. Value) + Air Cal Value
9
2-3. Saving Data
2-4. Initial Display (Self-Diagnostic Display)
Once the SAVE key is pressed, the buzzer will beep and the display shown right will appear. The time, data, and the measurement result are saved chronologically from No. 001 to No. 100. The saved data can be seen at “VIEW SAVED DATA” in the SETTING MODE. Also, the saved data can be deleted with “CLEAR SAVED DATA” in the SETTING MODE.
After starting with the MEASURING MODE by pressing the POWER key, the self-diagnostic function will automatically begin to work. If the result is normal, the display shown to the right will appear.
If any problems occur, it will be indicated as shown in the display at right. The display exemplifies a contamination of the interferometer. If the problem becomes serious, the display will not switch to the basic display , but instead the ERROR message display.
10
2-5. ERROR Message Display
ERROR message and the meanings are as follows:
Contents of Display
The Meanings and Required Solutions
ERROR: LOW BATT.
Battery voltage is too low to perform the measurement. Replace the batteries.
ERROR: CONTRAST
Contamination of the optic interferometer sensor is too severe to perform the measurement. Replace the sensor or have it repaired.
ERROR : BRIGHTNESS
The brightness is too low to perform the measurement. Replace the sensor or have it repaired.
ERROR: SETTING
The setup data saved in the setting mode has an error. Confirm the setup data in the setting mode and set the data again.
ERROR: SYSTEM
A system error has occurred. Repair work is required.
ERROR: PUMP
There is a possibility that the pump is not working. Pump replacement is required.
If the problem proceeds to affect the measurement, an ERROR message (as shown to the right) will
appear.
11
2-6. AIR CAL CAUTION
1. Press the PUMP key and let it sample fresh air.
key
minutes.) Otherwise, the correct measurement cannot be acquired.
sure to perform the AIR CAL procedures correctly.
If the temperature inside of the sensor has changed over ±10ºC from the last time the AIR CAL was performed, [CAUTION/CHECK AIR CAL] will be displayed in order to prevent zero drifting. [Measuring Gas/Base Gas] and [AIR CAL] are displayed consecutively.
If the [CAUTION/CHECK AIR CAL] is displayed, perform the AIR CAL based on the procedure below:
AIR CAL Procedure
2. When the indication becomes stable, stop the pump with the PUMP key and press the AIR CAL
Caution
Before pressing the AIR CAL key, allow the instrument to sample enough fresh air (ideally for two
Caution
When the measurement starts, confirm if the AIR CAL value is displayed after the fresh air is sampled enough. If the AIR CAL value is not indicated, or if [AIR CAL CAUTION] is not displayed, be