COSA Xentaur 83 User Manual

H
H
ygrocontrol
HYGROMESS TYPE 83 TRANSMITTER
GmbH
MANUAL
Feuchte- und Temperatur-Meßtechnik Moselstraße 2 B, D-63452 Hanau,
e-mail:
info@hygrocontrol.de
Version 06 / 10
06181-92790
Tel.:
Internet:
Fax.:
www.hygrocontrol.de
HYGROCONTROL Manual for Type 83 Page - 1 -
TABLE OF CONTENTS:
1. Fundamental Features of Series 83 2
2. Preparations before operating 2
2.1. Unpacking and Checking 2
2.2. Set-up for Operation 2
2.3. Instructions for Dispatch 3
3. Technical Data 3
4. HART 4
5. Operation 4
5.1 Mode "MEASUREMENT" 5
5.2 Mode "CALIBRATION" 5
5.2.1 General Notes concerning Calibration 5
5.2.2 How often Calibrations should be done 5
5.2.3 Calibration Chamber 6
5.2.4 Mounting the Calibration Device 6
5.2.5 Operation via BUS Commands 7
5.2.6 List of possible Commands 8
5.2.7 Output Examples 8
5.2.8 List of used Abbreviations 9
6. Practical Instructions and Limits 10
6.1 Direct Contact with Liquids 10
6.2 Filters 10
6.2.1 Resistance against high Air Velocities 10
6.2.2 Resistance against Dust and Aerosols 10
6.3 Resistance against Chemical Attack 11
6.4 Protection of Transmitter Electronics 11
6.5 Influence of Temperature on Humidity 11
6.6 Humidity Measurements over Ice 11
6.7 Measurements under Pressure and Vacuum 12
HYGROCONTROL Manual for Type 83 Page - 2 -
1. Fundamental Features of Series 83 Transmitters
HYGROMESS Humidity and Temperature Meters detect relative Humidity and Temperature. For this purpose the probes are equipped with a capacity type humidity sensor of very small mass. The temperature detection is done by a Pt-1000 RTD , which is placed very close to the humidity sensor to avoid differences in temperature between the two sensors. Incoming sensor signals are temperature compensated and transferred into linearised analogue or digital outputs by digital electronics. We use a microprocessor for the control of all internal functions.
The electronics for transmitting the temperature and humidity sensor signals are built into a IP-65 housing of 80 x 160 x 55 mm. Two different operational modes may be chosen by the user which are:
"MEASURING" and "CALIBRATION"
"MEASURING" is the normal mode for picking up humidities and temperatures, compensating, linearising and transferring the values to the outputs.
"CALIBRATION" mode is used for recalibration of sensor characteristics by special prepared LiCl solutions.
2. Preparations before Operating
2.1 Unpacking and Checking
Every instrument has been examined before dispatch in respect to its electrical and mechanical functions. After receipt of instrument, please check its operation. In case of any malfunction or damage, return the instrument with a detailed description of faults.
2.2 Set-up for Operation
The instrument has to be connected to a power supply of 9 to 36 VDC. The correct voltage is marked on the type label you find on one side of the housing. Do not use voltages other than the marked one! Severe damage would be the result of applying the wrong voltage. The cross section of your conductors should meet the dimensions of
0.5 up to 1.5 mm2 (AWG 16 to AWG 20). You will have to connect the 2 wires to the terminals marked (+) and (-), the shield should be connected at the cable gland. Opening of the housing is done by turning the four screws (quick-release types) with a screwdriver to the left.
HYGROCONTROL Manual for Type 83 Page - 3 -
Please take care that cable diameter corresponds with the dimension of the gland and so will guarantee a dense screwing. During
operation the housing of the instrument should be closed.
2.3 Instructions for Dispatch
If there any problems you are not able to correct, we ask you to contact our technical department or to return the instrument with a short description of the fault. When returning please do not forget to pack the instrument safely in a cardboard box and fill up holes with soft packing material to avoid damage.
3. Technical Data
3.1 Humidity
Humidity Range 0....100 % RH.
Resolution 0,1 % RH. Humidity Sensor SE - 02 Temperature Error of Humidity ± 0,01 % RH./ °C Repeatability < 0,5 % RH. Hysteresis after 4 hour Cycle 10 % - 95 % - 10 % r. F. < 0,5 % RH. Linearity of Electronic / Humidity < 1 % RH.
3.2 Temperature
Range -50.....+150 °C
Resolution 0,01 °C Temperature Sensor RTD: Pt-1000 System Accuracy for Temperature over the total Range ± 0,35 °C Linearity of Electronic / Temperature ± 0,15 °C
3.4 Maximum environmental conditions during operation
Transmitter Electronics -10...+ 50 °C Probe Tube with Sensors -50...+150 °C Max. Humidity 95%RH
3 .5 Connections
Screw Terminals
Marked with + and –
HYGROCONTROL Manual for Type 83 Page - 4 -
4. HART
Following universal commands are available:
0 Read unique identifier 1 Read primary variable 3 Read current and four variables 12 Read message 13 Read tag, description, date 14 Read primary variable sensor information 15 Read output information 16 Read finally assembly number 17 Write message 18 Write tag, description, date 19 Write finally assembly number 34 Write damping value 35 Write range values 44 Write primary variable units 49 Write primary variable sensor serial number 59 Write numbers of response preambles 109 Burst mode control
Sending following user messages will not cause their storage but result in starting the functions below:
%Hxxx starts a humidity calibration with xxx * 0,1 %RH %Txxxx will do a temperature adjustment to xxxx * 0,01 °C %Y positive confirmation of humidity calibration %N negative confirmation of humidity calibration %F end of calibration and show of original users message
Command 3 results in the output of measurement values as follows:
primary variable Humidity in %RH shown unit = % secondary variable Temperature in °C tertiary variable Humidity digits shown unit = none quaternary variable Reference digits shown unit = none
5. Operation
HYGROCONTROL Humidity and Temperature Meters detect relative Humidity and Temperature. For this purpose the probes are equipped with a capacity type humidity sensor of very small mass. The temperature detection is done by a Pt-1000 type RTD , which is placed very close to the humidity sensor to avoid differences in temperature between the two sensors. Incoming sensor signals are temperature compensated and transferred into linearised analogue or digital outputs by digital
Loading...
+ 9 hidden pages