64- or 32 Pin Connectors (19"-Versions) C1 + C2
Connections / Dimensions of type "72 - A" C3
Connections / Dimensions of type "72 - B" Analog C4
Connections of type "72 - B" Digital C5
Connections of special Versions C6
Accu Version of 19"-Transmitters C7
M1 - M9
C1 - C7
Series 72 Manual page 1
1. Fundamental Features of Series 72
Series 72 Humidity and Temperature Meters detect relative Humidity and
Temperature. For this purpose the probes - either fixed mounted or
interchangeable ones - are equipped with a capacity type humidity sensor of
very small mass. The temperature detection is done by a RTD of type Pt-1000,
which is placed very close to the humidity sensor to avoid differences in
temperature between the two sensors. To assure rapid response to temperature
changes, a thinfilm Pt-1000 on a substrate of ceramic was chosen.
Incoming sensor signals are temperature compensated and transfered into
linearized analog or digital outputs by digital electronics. We use a
microprocessor for the control of all internal functions. Programing and
configurating of the desired ranges and functions is easy due to the dialog
orientated software. Only four buttons are required. In addition, the
microprocessor provides absolute humidity values by calculating them from
relative humidity and temperature, if atmospheric pressure is assumed, or
pressure is known and typed in by pushbuttons. As a standard, the dual line
display is able to show the temperature in °C or °F and the humidity as:
a) rel. Humidity: in %
b) abs. Humidity: in g/m
c) Watercontent: in g/kg
d) Dewpoint: in °C or °F
e) Enthalpy: in Joule
The software of series 72 instruments is divided into 4 modules which can be
called by the operator. These are named:
a) "RUN" = normal measuring mode
b) "LIST" = display of all actual parameters
c) "CALIB" = calibration of a probe
d) "CONFIG" = configuration of display and outputs
By using SMD-techniques, it was possible to build in all the electronics for
transmitting the temperature and humidity sensor signals into a IP-65 housing of
80 x 82 x 55 mm. Only special options, like RS-232 / V24 output, need a larger
housing (80 x 160 x 55 mm).
2. Preparations before Operating
3
2.1. Unpacking and Checking
Every instrument has been examined before dispatch in respect to its electrical
and mechanical functions. After receipt of instrument, please check its
operation. In case of any malfunction or damage, return the instrument with a
detailed description of faults.
Series 72 Manual page 2
2.2. Setup for Operation
The instrument is to be connected to main powers either by a connector or by
using the terminals inside the housing. The correct voltage is marked on the
type label you find on one side of the housing. Do not use voltages other than
the marked one! Severe damage would be the result of applying the wrong
voltage.
If a type 72B. instrument with screw terminals is used, the cross section of your
conductors should meet the dimensions of 0.5 up to 1.5 mm
20) You will find these terminals on the ground PC-board after opening the
housing by turning the four screws (quick-release types) with a screwdriver to
the left.
Please take care that cable diameter corresponds with PG-dimension and so
will guarantee a dense screwing. During operation the housing of the instrument
should be closed.
2.3. Instructions for Dispatch
If there any problems you are not able to correct, we ask you to contact our
technical department or to return the instrument with a short description of the
fault. When returning please do not forget:
a) to add the description of faults
b) to pack the instrument safely in a cardboard box and fill up
holes with soft packing material to avoid damage.
2
(AWG 16 to AWG
3. Dialog orientated Software
3.1. Program Structure
For simplier operation, the software of type 72 transmitters is devided into 4
main parts which are:
a) "RUN" = normal operation for measuring Humidity and Temp.
b) "LIST" = listing all parameters used for operation
c) "CALIB" = calibration of a sensor probe
d) "CONFIG" = setup or change of parameters as:
ranges of humidity and temperature,
lower and upper limits of analog outputs,
dimensions of humidity and temperature values
Each of this main programs is devided into submenus to create a better
overview of its functions and to simplify the operation.
3.2. Pushbutton Operation:
The four pushbuttons are located on the display board and can only be seen
after removing the cover of the transmitter housing. Opening of the cover is
Series 72 Manual page 3
done by turning the four fast mounting screws 90 degrees to the left with a
screwdriver. The red coloured buttons are marked on the display - board with:
1) "E" = enter
2) "M" = mode
3) "
" = encrease
4) "
" = next decimal
The two arrows are the only pushbuttons which have an "enhanced speed"
function, which operates after pressing them for a time period of more than 1
second. In the "LIST" submenus the "
to the previous display point.
4. Measurement Mode "RUN"
"-button leads to the next, the ""-button
4.1. Conditions after Power Connection
After connecting the Hygrocontrol 72 to power, the transmitter is in its normal
measurement mode, called "RUN". First all the segments of the dual line
display are tested for a short period, then a variety of tests occur, which give
information about the probe connected to the microprocessor. If any test fails,
an error message is displayed. See error messages (Chapter 10.).
4.2. Values in the Display
After passing the tests, the display will show in its upper line the actual value of
Humidity with a dimension. This dimension may be - depending on
configuration:
a) % RF (German for rel. Humidity)
b) % RH
c) g/kg (Watercontent in 1kg dry air)
d) g/m
e) °C
f ) °F
The lower line of display shows the Temperature of the probe head with
dimension according to configuration:
3
(Amount of water in 1m3)
Td
(Dewpoint in Fahrenheit)
Td
(Dewpoint in Centigrade)
g) °C (Centigrade)
h) °F (Fahrenheit)
5. Software Module "LIST"
5.1. General Informations about the "LIST" Menu
Using the "LIST" module, users of series 72 transmitters are able to display but not to change - all the actual parameters the instrument has in its memory.
Series 72 Manual page 4
This is a feature which enables the user, supplier and manufacturer - even after
a malfunction - to analyse what has happened with the transmitter.
For better understanding, the "LIST" menu is divided into three submenus which
are named according to their content:
a) "MINMAX" = Memory of the Min and Max values
b) "CONDAT" = Memory of all the configuration datas
c) "CALDAT" = Memory of all the calibration datas
Besides these three submenus of "LIST", there are two ways of jumping out of
the "LIST" menu. These are:
d) "RETURN" = jumping back to the begin of "List"
e) "RUN" = jumping back to measurement mode
5.2. Entering "LIST"
As shown in the survey diagram of all menus (page M1), the entrance into a
menu or submenu allways is done by pressing the "E" (enter - button). The
display now shows "LIST" and the next hitting of "E" brings "MINMAX" into the
display. As in all other submenus you now may enter the "MINMAX" submenu
by pressing "E", or go to other menus by pressing "M". If you once stepped into
a submenu, you may jump back by pressing "E" again.
5.3. Data in "LIST" - "MINMAX"
"MINMAX" holds the MIN and MAX values since the last start of the transmitter
or the last erasing of this memory. The data is shown in the following order:
1) Minimum of Temperature = "T - min", for example "19.5°C"
2) Maximum of Temperature = "T - max", for example "91.1°C"
3) Minimun of Humidity = "H - min", for example "10.1 %"
4) Maximum of Humidity = "H - max", for example "95.1 %"
After these values, the display shows the last three readings of temperature and
humidity. These are kept in memory to enable the transmitter, in case of a
sensor break, to decide which values will be given to the outputs if the option
"keep steady outputs if sensor break occurs" was chosen.
5) last Temperature value = "T - 1", for example "20.3°C"
6) second last Temp.value = "T - 2", for example "20.4°C"
7) third last Temp. value = "T - 3", for example "20.5°C"
8) last Humidity value = "H - 1", for example "79.3 %"
9) second last Hum. value = "H - 2", for example "79.2 %"
10) third last Hum. value = "H - 3", for example "79.1 %"
Quitting "MINMAX" by pressing the "E" button, leads to the display of "Cl Y N". It
is now up to you to decide whether the memory of Min Max Datas should be
erased or not. Pressing "
cases the program will jump back to "MINMAX", which is the start of the "LIST"
submenus.
" will erase, "E" will keep the Min Max Values. In both
Series 72 Manual page 5
5.4. Data in "LIST" - "CONDAT"
Submenu "CONDAT" displays all data, which describe the actual configuration
of the transmitter. After chosing "CONDAT" and pressing "E", you will find the
data as follows:
1) lower limit of Humidity range = "H out “
2) upper limit of Humidity range = "H out "
3) reaction at sensor break Hum. = "H err" with arrow
4) lower limit of Temp. range = "T out "
5) upper limit of Temp. range = "T out "
6) reaction at sensor break Temp. = "T err " with arrow
7) offset analog output Temp. low = "T SET "
8) offset analog output Temp.high = "T SET "
9) offset analog output Hum. low = "H SET "
10) offset analog output Hum. high = "H SET "
11) Display code = "C CODE" f.e. "0"
12) Baud rate of digital output = f.e. "4800" "RS232"
13) address of digital output = f.e. "4" "ADRESS"
14) software version number = "VS" f.e. "T.21"
15) date of software version = f.e. "29.05.91"
16) serial number of transmitter = f.e. "AT1234"
17) serial number of probe = f.e. "AS4321"
or or
or or
With the "
previous point. With "E" you may leave the submenu (at any display!) and jump
back to "CONDAT".
5.5. Data in "LIST" - "CALDAT"
After entering "CALDAT", the display will first show the number of calibration
points - this is a number inbetween "00" and "11" with an additional "NUM" in
the display. Pressing of "
which are marked with the letters a) to k):
a) to k) = calibration datas from 0% to 100%.
After that, the socalled "straight line characteristic" is displayed as for example:
"-23.4" "1.34455".
The next two data points are the initial calibrations at 0% and 80% done at the
factory.
2) and 3) = first and second "initial calibration points".
These two values cannot be erased - not even with the submenu "DELET" of
the "CALIB" program. This is to avoid, that after deleting all calibration data the
transmitter has no straight line information to calculate humidity values from an
actual sensor reading. In addition, the manufacturer may get information about
the sensor changes by comparison of actual and initial calibration data.
" button you step to the following point, with the "" button to the
" leads to the calibration values at these points,
After that, the display shows two numbers relating to the current sensor- and
reference- oscillations. A letter in the last decimal marks:
Series 72 Manual page 6
F) = sensor oscillator
R) = reference oscillator
Finally the display shows the actual pressure factor, marked by "PRESS". This
factor is used to calculate humidity values at atmospheric pressures, if
measurement occurs at other pressure conditions.
The calibration data display may be left by pressing "E", the program then
jumps back to "CALDAT".
6.Calibration
6.1. General Notes about Calibration Standards
There is a lack of international standards for rel. Humidity calibrations,
therefore, different countries may have different methods of calibrations. Since
the users of humidity meters need a simple method to check the accuracy of
their instruments, we are supplying accessories for calibration or testing. As a
standard method we have chosen unsaturated salt solutions to create rel.
humidities of known amount - but saturated salt solutions may also be used.
Series 72 instruments are to be calibrated with rel. humidity - since their
sensors respond to relative, not to absolute, humidity values.
6.2. How often Calibrations should be done
Before dispatch, all probes have been calibrated at 0, 20, 35, 50, 65, 80 and
95%. Changes of sensor characteristic are hard to predict, as they depend on
mechanical and chemical attack on the probe. If stresses are high (such as
temperature, vibration, dust and aggressive media like acids, solvents and
highly reactive gases), we recommend a test within 3 to 12 months. If deviations
of more than 2% are found, a recalibration makes sense.
In some cases, when the humidity range the probe has to detect is a narrow
one - then a partial calibration may be helpful. If you want the best accuracy
over the total range (from 0 to 100%), then as many points as possible should
be used for calibration.
6.3. Salt Solutions for Calibration
The solutions we supply for calibration create rel. humidities from 0 to 95%. We
are using the following:
The marked rel. humidities relate to a temperature of 22°C and do have a
temperature dependency. The accuracy of the solutions at 22°C is +/- 0.5%,
they are not toxic and are not dangerous to the environment.
Series 72 Manual page 7
To assure the easiest handling and highest accuracy, the solutions are
enclosed in glass ampoules, which should be opened only before usage.
Opening is simplified by the weakened middle part of the ampoules. Once
opened the solutions may catch humidity from or give humidity to their
environment - so the time of usage is restricted and they should be used for one
calibration only.
6.4. Calibration Chamber
The calibration chambers we are supplying under No. 1810 (or 4716 for 12 mm
diameter probes) are hermetically sealed chambers which consist of three
parts:
1) main part with thread for coupling with probe tip
2) removeable bottom to put the solution in
3) O-ring to give tight connection between parts 1) and 2).
The chamber is made from Aluminium to achieve
a) temperature equilibrium
b) simpliest cleaning procedure.
Mechanical dimensions of the chamber are optimized to form a closed body
without edges. By this form a quick and stable atmosphere of the desired rel.
humidity over the solution is guaranteed.
6.5. Usage of Saturated Salt Solutions
There exist a number of different salts where the rel. humidity over the surface
of their saturated solutions is known and reported by organisations like NBS or
DIN. Lately the temperature dependencies and accuracies of this solutions were
verified - so series 72 transmitters are able to be tested and calibrated with
these solutions too. As preparation and handling of these saturated solutions
require knowledge and care, we do not provide these directions in this manual.
If you need any help in this matter, please contact our technical staff.
6.6. Temperature Equilibrium
Due to the temperature dependency of rel. humidity over salt solutions, you
should assure a constant temperature of 22°C with deviations of not more than
1°C. This can be done by placing the probe in an area free from draughts, direct
sunlight, radiators and any other factor that might cause temperature
fluctuations. Place the instrument on an insulating base such as polystyrene
and assure that the instrument, calibration chamber and solutions are
conditioned to the same temperature before starting the calibration.
6.7. Mounting the Calibration Device
First remove the filter from the tip of the probe. The chamber may now be
screwed to the thread of the probe. The bottom of the chamber is removeable
and it is here that the textile pad and the solution are to be placed. To avoid
improper humidity values, always clean the bottom of the chamber thoroughly,
use new pads, and new solutions.
Series 72 Manual page 8
Ensure that the calibration chamber faces downwards, to avoid direct contact of
the solution with the sensor. Before inserting the bottom into the chamber,
unscrew the chamber slightly, to help escaping air, when you close the
chamber. Tighten the chamber again after closing!
7. Program "CALIB"
7.1. Password Protection
To protect series 72 instruments against unwanted changes of their calibration,
this part of the program is protected by a password. You should give this
password only to trained and authorized personnel. The password has to be
calculated by a simple mathematical operation from the numeric last 4 digits of
the serial number of the transmitter. This is demonstrated by the following
example:
Let the serial number of the transmitter be: T23456
The first two digits of the password are formed by the sum of the digits number
3 and 4 of the serial number, the last two digits of the password are formed by
the sum of the last two digits of the serial number. In the above given example
of serial number we get:
3 + 4 = 07 for the first two digits,
5 + 6 = 11 for the last two digits.
The password in our example is then: 0711
7.2. Facilities of "CALIB"
The program "CALIB" holds all the facilities to:
a) recalibrate one or more points with menu "CHANGE"
b) add more calibrations with "CHANGE"
c) erase all humidity calibrations with "DELETE"
d) recalibrate temperature with "TEMPER"
e) enter a pressure factor, if measurement is done at a
constant, but not atmospheric pressure, with "PRESS"
f) return to measurement mode with "RUN"
g) return to beginning of "CALIB" with "RETURN"
To avoid any "uncertain" or "dubious” calibrations, the program uses two
principles, which are:
1) If one of the standard fixpoints is recalibrated, the former
value in memory is replaced.
2) If a nonstandard point is calibrated (for example a
saturated LiCl-solution), a standard fixpoint closer than
5% to the new calibration point will be replaced.
Two examples should illustrate this rules:
Series 72 Manual page 9
1) A calibration with saturated LiCl-solution of 11.3%
will replace the former 10%-value.
2) A calibration with saturated NaCl-solution of 73.6%
will not replace the two former values at 65% and 80%.
7.3. Entering "CALIB"
From normal measuring mode you enter the main program area by pressing
"E". The display now shows "LIST". Pressing "M" now switches over to "CALIB"
and the first line of the display shows "0000" with a flashing first digit. It is
necessary now to enter the password by using "
flashing and "" to go to the next decimal. If the displayed number matches the
desired password - one digit always flashes - then pressing "E" will bring you
into the program. The display shows "CHANGE" as the first submenu of
"CALIB". By pressing "M" the other submenus of "CALIB" may be chosen.
7.4. Calibration Procedure ("CHANGE")
If the calibration chamber is fixed to the probe and filled with the solution you
want to calibrate with, the calibration is initiated by "E".
The first line of display will show the measured rel. humidity, the second line will
propose a standard fixpoint which is closest to the actual measurement. If this
proposal does not correspond to the solution, you may change it by hitting "M"
as often as necessary to give the display you want.
If a humidity value you want to calibrate, is not in the proposal list, you may
choose "OTHER" and acknowledge with "E". Now the humidity value is shown
with a flashing first decimal, and adjustment may be done, to achieve the
display you need, by using "" and "" buttons. "" increases the flashing
decimal, "" steps over to the next one. If the desired value is in the display,
press "E". Now a safety check is seen in the form of "N ok Y". Pressing "M" will
jump back to "OTHER", pressing "E" will step into the calibration procedure, as
explained next.
" to change the digit which is
Now pressing "E" will start the calibration. The display shows in the first line
(flashing) the rel. humidity measured, and in the second line (alternating) the
chosen point and a number corresponding to the actual oscillator frequency of
the sensor. This display allows you to see, when stability of humidity in the
chamber is achieved.
When stable conditions are reached, the flashing display ends and you will see
the symbols "Y" and "N". If you want to accept this calibration, you hit "E" now, if
you want to reject it, you hit "M". In both cases the program jumps back to
"CHANGE". If "E" was "E" hit the new calibration is kept in memory. If "M" was
hit the former calibration point is still in memory and no change occured. The
time to reach sufficient equilibrium varies with the solutions and the
environmental conditions and may be 10 minutes or 2 hours.
To calibrate more points you must change the solution in the calibration
chamber. After spilling the old solution, the chamber should be cleaned with
water, and dried before filling with another solution. After closing the chamber
you may start with the next calibration.
Loading...
+ 28 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.