Cornelius FRUITISTA 3 User Manual

FRUITISTA VIPER 3 FLAVOR
Service Manual
Release Date: September 9, 2010
Publication Number: 621360041TBSER
Revision Date: N/A
Revision: A
Visit the IMI Cornelius web site at www.cornelius.com
for all your Literature needs.
This Product is warranted only as provided in Cornelius’ Commercial Warrant applicable to this Product and is subject to all of the restrictions and limitations contained in the Commercial Warranty.
Cornelius will not be responsible for any repair, replacement or other service required by or loss or damage resulting from any of the following occurrences, including but not limited to, (1) other than normal and proper use and normal service conditions with respect to the Product, (2) improper voltage, (3) inadequate wiring, (4) abuse, (5) accident, (6) alteration, (7) misuse, (8) neglect, (9) unauthorized repair or the failure to utilize suitably qualified and trained persons to perform service and/or repair of the Product, (10) improper cleaning, (11) failure to follow installation, operating, cleaning or maintenance instructions, (12) use of “non-authorized” parts (i.e., parts that are not 100% compatible with the Product) which use voids the entire warranty, (13) Product parts in contact with water or the product dispensed which are adversely impacted by changes in liquid scale or chemical composition.
Contact Information:
To inquire about current revisions of this and other documentation or for assistance with any Cornelius product contact:
www.cornelius.com
800-238-3600
Trademarks and Copyrights:
This document contains proprietary information and it may not be reproduced in any way without permission from Cor­nelius.
Printed in U.S.A.
Copyright © 2010, All rights reserved, IMI Cornelius Inc.
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Manual Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Safety Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Read and Follow ALL Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Safety Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Recognition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Different Types of Alerts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Safety Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Qualified Service Personnel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Safety Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Shipping And Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
CO2 (Carbon Dioxide) Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Mounting in or on a Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Dispensed Product Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Overrun, as Applied to Carbonated Beverages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Overrun Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Overrun is a Variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Specific Product Ingredients Affect Overrun . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
BRIX Affects Overrun. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Low Dispensing Volume Affects Overrun. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Carbonation Level in Liquid Product Affects Overrun . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Freezing Affects Overrun . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
CO2 System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Water System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Syrup System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Control Panel Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Control Panel Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Control Panel Menu Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
System Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Unit Data Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Error Status Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Error Log Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Clock Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Setting the Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Setting Daylight Savings Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Events Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Setting Events . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Setting the Sleep and Wakeup Times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Setting Defrost Lockout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Viscosity Setup Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
CO2 Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Option Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Setting the Temperature Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Setting the Date Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Setting the Time Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Setting the POS Lighting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Setting the Type of Syrup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Barrel Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Maintenance Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Barrel Maintenance Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Purging a Barrel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Filling a Barrel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Run the Barrel Motor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Rinsing a Barrel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Manual Diagnostic Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Totals Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
BRIX Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Motor Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Calibrating a Motor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
System Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Pressure Sensors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Maintaining Product Quality. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Dispensed Product Throughput . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Programmed Defrost Scheduling. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Sleep Mode Recommendations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Viscosity Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
CO2 Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Starting the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Preventative Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Monthly Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Cleaning Air Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Semi-Annual Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Inspecting and Replacing Scraper Blades . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Cleaning the Syrup Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Servicing Motorman Dispensing Valves. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Sanitizing the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Emptying a Barrel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Flushing the System of Syrup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Sanitizing the Barrel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Flushing the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Cleaning the Water Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Component Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Barrel Motor Seal Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Removing the Existing Seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Installing a New Seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Motor Seal Leak Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Barrel Motor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Motor Run Capacitor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Water Pump Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
CO2 Pulse Valve Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Syrup and Water Shutoff Solenoid Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Condenser Fan Motor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Compressor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Hot Gas Solenoid Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Liquid Line Solenoid Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Syrup Pressure Switch Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Water Pressure Switch Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Display Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Control Board Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Software Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Motor Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Troubleshooting the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Troubleshooting Product Not Cold. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Viper Installation Manual

MANUAL OVERVIEW

INTRODUCTION

The organization of this manual allows the user to scan quickly to the subject of interest along the left side of a page and to read the detail about the subject or procedure on the right side of the page. The manual provides the detail needed for newcomers to the industry while allowing experienced technicians to skip over the details and move quickly through the material.
This manual is designed as a guide to the technician in maintaining and servicing the Viper system. The Viper system is simple in design and has built-in features and diagnostic controls to help the service technician quickly and accurately service the machine.
The unit consists of multiple freeze barrels that each contain an internal beater driven by a rear mounted electric motor, a refrigeration system, timer-controlled, automatic hot gas defrost system and interconnecting tubing and controls required to dispense the product.
Some of the system features and functions are listed below:
Simple User Interface LCD Display
Real Time Clock
Incoming Line Voltage Sensing
Pressure Sensing of Incoming Water, Syrup and CO
System Error Handling
Error Log
Viscosity Control
2
© 2010, IMI Cornelius Inc. - 1 - Publication Number: 621360041TBSER
Viper Installation Manual

SAFETY INSTRUCTIONS

READ AND FOLLOW ALL SAFETY INSTRUCTIONS

Safety Overview Read and follow ALL SAFETY INSTRUCTIONS in this manual and any

warning/caution labels on the unit (decals, labels or laminated cards).
Read and understand ALL applicable OSHA (Occupational Safety and
Health Administration) safety regulations before operating this unit.

Recognition

Recognize Safety Alerts
!

Different Types of Alerts

SAFETY TIPS

This is the safety alert symbol. When you see it in this manual or on the unit, be
!
Indicates an immediate hazardous situation which if not avoided WILL result in serious injury, death or equipment damage.
!
Indicates a potentially hazardous situation which, if not avoided, COULD result in serious injury, death, or equipment damage.
!
Indicates a potentially hazardous situation which, if not avoided, MAY result in minor or moderate injury or equipment damage.
alert to the potential of personal injury or damage to the unit.
DANGER:
WARNING:
CAUTION:
Carefully read and follow all safety messages in this manual and safety
signs on the unit.
Keep safety signs in good condition and replace missing or damaged
items.
Learn how to operate the unit and how to use the controls properly.
• Do not let anyone operate the unit without proper training. This appli-
ance is not intended for use by very young children or infirm persons
Publication Number: 621360041TBSER - 2 - © 2010, IMI Cornelius Inc.
Viper Installation Manual
without supervision. Young children should be supervised to ensure that they do not play with the appliance.
Keep your unit in proper working condition and do not allow unauthor-
ized modifications to the unit.

QUALIFIED SERVICE PERSONNEL

!
WARNING:
Only trained and certified electrical, plumbing and refrigeration technicians should service this unit. ALL WIRING AND PLUMBING MUST CONFORM TO
NATIONAL AND LOCAL CODES. FAILURE TO COMPLY COULD RESULT IN SERIOUS INJURY, DEATH OR EQUIPMENT DAMAGE.

SAFETY PRECAUTIONS

This unit has been specifically designed to provide protection against personal injury. To ensure continued protection observe the following:
Disconnect power to the unit before servicing following all lock out/tag out procedures established by the user. Verify all of the power is off to the unit before any work is performed.
Failure to disconnect the power could result in serious injury, death or equipment damage.
Always be sure to keep area around the unit clean and free of clutter. Failure to keep this area clean may result in injury or equipment damage.

SHIPPING AND STORAGE

Before shipping, storing, or relocating the unit, the unit must be sanitized and all sanitizing solution must be drained from the system. A freezing ambient environment will cause residual sanitizing solution or water remaining inside the unit to freeze resulting in damage to internal components.
!
WARNING:
!
CAUTION:
!
CAUTION:

CO2 (CARBON DIOXIDE) WARNING

© 2010, IMI Cornelius Inc. - 3 - Publication Number: 621360041TBSER
Viper Installation Manual
!
DANGER:
CO
2 displaces oxygen. Strict attention MUST be observed in the prevention of
CO
2 gas leaks in the entire CO2 and soft drink system. If a CO2 gas leak is
suspected, particularly in a small area, IMMEDIATELY ventilate the contaminated area before attempting to repair the leak. Personnel exposed to high concentrations of CO of consciousness and DEATH.

MOUNTING IN OR ON A COUNTER

!
WARNING:
When installing the unit in or on a counter top, the counter must be able to support a weight in excess of 450 lbs. to insure adequate support for the unit. FAILURE
TO COMPLY COULD RESULT IN SERIOUS INJURY, DEATH OR EQUIPMENT DAMAGE.
Note: Many units incorporate the use of additional equipment such as icemakers. When any addition equipment is used you must check with the equipment manufacturer to determine the additional weight the counter will need to support to ensure a safe installation.
2 gas experience tremors which are followed rapidly by loss

DISPENSED PRODUCT CONDITIONS

Overrun, as Applied to Carbonated Beverages

Overrun Definition
Overrun is defined as product expansion that takes place in the frozen carbonated drink. It is caused primarily by CO2 gas breakout and secondarily by freezing.
Overrun is a Variable
The percentage or degree of overrun depends on a number of factors. The specific syrup, BRIX, low dispensing volume, carbonation level in the liquid product and freezing of the product. These items all affect overrun. After these factors hav been considered, desired viscosity (product consistency) adjustment may be made on the unit. The viscosity adjustment adjusts product texture from very wet to light.
Specific Product Ingredi­ents Affect Overrun
Each syrup has its own specific formulation of makeup. Fruit flavors contain citric acids that colas do not. Colas also differ in ingredients from one brand to another.
Publication Number: 621360041TBSER - 4 - © 2010, IMI Cornelius Inc.
BRIX Affects Overrun
Viper Installation Manual
Each product formulation has its own peculiarities regarding the way the product absorbs carbonation and the way it releases carbonation.
Sugar in carbonated drinks is like anti-freeze in water. The higher the BRIX, the greater the resistance of the product to freezing. Conversly, in products with lower BRIX, freezing takes place at higher termperatures than for high-BRIX products. Thus, BRIX affects overrun because the amount of sugar in a drink has a direct bearing on the product’s freezing characteristics.
Low Dispensing Volume Affects Overrun
Carbonation Level in Liq­uid Product Affects Over­run
Freezing Affects Overrun
When a unit sits idle fo a period of time with no drinks being dispensed, CO2 gas in the system takes a “set”. When the first few drinks are drawn off after an idle period, CO
2 gas has less tendancy to break out as the drink is dispensed. The
result is that these first drinks hav less overrun than drinks dispensed during peak­use periods.
The higher the specific carbonation level in a given product, the greater the potential for carbonation breakout in frozen carbonated form of that drink. For example, drinks with 3.0 volume of carbonation have more gass breakout in frozen carbonated form and more overrun than drinks that contain 2.0 volumes of CO
2 gas.
Freezing causes approximately a 5-7 percent expansion in dispensed frozen carbonated drinks. The degree of freezing is limited because the finished drink is intended to be sipped through a straw. This is not possible if the product is too “solid”.
© 2010, IMI Cornelius Inc. - 5 - Publication Number: 621360041TBSER
Viper Service Manual

SYSTEM OVERVIEW

INTRODUCTION

The Viper unit consists of the following systems and hardware:
Multiple freeze barrels, each containing an internal scraper bar driven by an AC motor.
A refrigeration system and an intelligent, hot gas defrost system.
The components are enclosed in a powder-coated steel frame to prevent corro­sion. It is covered with ventilated cladding panels and a lighted merchandiser. The cladding is easily removable to facilitate installation, service and mainte­nance.
Each barrel has a transparent faceplate, with an integral relief valve and a removable, self-closing dispensing valve mounted on the front. A removable drip tray, with cup rest is located directly below the dispensing valves.
A programmable control system with a control panel that controls operational and diagnostic functions and settings is located behind the merchandiser.
Publication Number: 621360041TBSER - 6 - © 2010, IMI Cornelius Inc.
Viper Service Manual

THEORY OF OPERATION

The refrigeration system schematic is shown in Figure 1. It shows the basic con­figuration for the Viper refrigeration system.
The wiring diagram of the 3-Barrel Viper unit is shown in Figure 2. This diagram shows the details of the electrical connections in the unit.
Figure 1. Fruitista Viper System 3-Barrel Refrigeration Schematic
© 2010, IMI Cornelius Inc. - 7 - Publication Number: 621360041TBSER
Viper Service Manual
Figure 2. Fruitista Viper 3-Barrel Wiring Diagram
Publication Number: 621360041TBSER - 8 - © 2010, IMI Cornelius Inc.

CO2 System

Viper Service Manual
An overall schematic of the three delivery systems contained in the unit are shown in Figure 5. The CO and the syrup system is at the bottom of the diagram. The CO
2 system is on top, the water system is in the middle
2 system interacts
with both the water and syrup systems. It provides pressure and carbonation for the syrup/water product mix.
A CO2 tank or bulk CO2 supply delivers carbon dioxide gas (CO2) to an adjust­able secondary CO Figure 3 and Figure 4. The CO pump, the expansion tank regulator and the secondary CO
CO
2 enters the expansion tank regulator and is reduced to approximately 30
2 regulator assembly that is attached to the tank as shown in
2 system also supplies CO2 to the water boost
2 tank regulators.
psig to feed holding pressure on the expansion tank. This provides a force to work against the barrel pressure when the product freezes and expands.
CO
2 also enters the secondary regulators. These regulators are used to adjust
barrel overrun/expansion for various products. The pressure settings for various types of syrup are shown in Table 1. Overrun CO
2 pressure is applied to the CO2
control solenoids through preset orifices and on to the in-line check valves.
From the in-line check valve, the CO
2 flows into the product line.
Figure 3. Typical Bulk Cylinder CO2 Connection
© 2010, IMI Cornelius Inc. - 9 - Publication Number: 621360041TBSER
Viper Service Manual
Table 1 provides guidelines for machine settings based on general syrup type. Several factors, including syrup formulation, level of citric acids, etc, will impact settings. These settings are to provide initial adjustments to achieve product overruns in the 80-120% range.
Figure 4. Typical Tank CO2 Connection
Table 1
Settings
Nominal Settings FCB 3 50% 7 PSIG
NOTE: For citric syrups, adjust the CO
the above to compensate for the lower CO
Syrup
Type Set
Viscosity CO
2 pressures down by 2-4 PSIG from
2 Setting Expansion
2 adsorption.
Publication Number: 621360041TBSER - 10 - © 2010, IMI Cornelius Inc.
Viper Service Manual
Figure 5. 3-Barrel System Schematic
© 2010, IMI Cornelius Inc. - 11 - Publication Number: 621360041TBSER
Viper Service Manual

Water System

Incoming water flows to a water pressure regulator that is preset to 30 psig. It flows through the water booster pump, to the regulator and through a sold-out switch.
Once through the boost pump, the water supply is split to each barrel and is fed to the water flow controls. From the flow rate control, the water passes through a single ball check valve and mixes with the syrup in a wye fitting for injection into the freeze barrel.

Syrup System

Syrup enters the unit through a sold-out switch. It enters a similar flow control to the water system, passes through a single-ball check valve and into the wye fit­ting where it meets the water for injection into the freeze barrel.

CONTROL PANEL

Introduction

The Viper unit uses a microprocessor based control system that monitors and controls all of the major systems and components of the machine. Temperatures and pressures are monitored, along with pumps, valves and the refrigeration system. They are managed by the control system to provide a consistently high quality product with optimal efficiency.
The control system is set up by the service provider to perform the tasks neces­sary to keep the Viper unit operating correctly. In addition to controlling the unit, the control system keeps track of the diagnostic information used when adjust­ing and/or repairing the machine.
The control system needs to be accessed in the following situations:
• Installing the Viper
• Modifying Operating Characteristics
• Checking Performance
• Servicing/Repairing the Machine
• Checking for Error Messages
The control system is accessed using the control panel located behind the lighted merchandiser. The control panel contains an LCD display and buttons shown in Figure 6.
There are 2 levels of access to the control panel: The first level can be accessed by the operator for normal operation and the second level is used by qualified service technicians for installation and service functions. The service functions can be secured (locked out) so that an operator does not have access to them.
Publication Number: 621360041TBSER - 12 - © 2010, IMI Cornelius Inc.
Viper Service Manual
Control Panel Display
The control panel has a structured organization of menus. The outline of this structure is shown in Figure 7. The Maintenance sub-menu is not visible when the security feature is on.
The first menu that is displayed after the unit is powered up and stabilized is the BARREL STATUS or HOME menu, shown in Figure 6. This menu is displayed when the unit is running in normal operation.
The control panel display has two main areas. The first area is the menu display area. This area presents information about the status and settings of the machine. It also displays menus of actions that are taken to modify the function­ing of the machine.
Control Panel Buttons
Menu
BARREL STATUS
#1 FREEZE 87 #2 FREEZE 68
Display
Area
12:51P MAR 04
MENU ON OFF DFRST SPIN
Menu Buttons
Figure 6. Control System Display
The second area are the buttons and arrows located across the bottom and on the right side of the control panel. There are up to five buttons and four direc­tional arrows that may be used on a screen to activate and control various the functions of the system. Each button that is active for a given menu has a label directly above it. The label describes what that button controls. For example, from the Barrel Status menu, pressing button 1 (labeled MENU) displays the Select menu with more options — pressing button 3 (labeled OFF) turns off the highlighted barrel.
On each menu below the Select menu, there is a button (labeled BACK) to save any settings and return to the previous menu. There is another button (GO) to advance to the highlighted menu.
© 2010, IMI Cornelius Inc. - 13 - Publication Number: 621360041TBSER
Viper Service Manual
Initialization
Main
Sub-Menus
Screens
Unit Data
Barrel Data
Barrel
Status
(Home Screen)
Setup
Sub-Menus
Clock Setup
Event Setup
Option Setup
Maintenance
Sub-Menus
Barrel Maint.
Manual Diag
Totals
Error Status
Error Log
Unit Data
Software Ver.
Figure 7. Software Structure
Viscosity Setup
CO2 Setup
Barrel Maint.
Brix Setup
Motor Setup
System Setup
Publication Number: 621360041TBSER - 14 - © 2010, IMI Cornelius Inc.
Viper Service Manual

CONTROL PANEL MENU DESCRIPTIONS

The following section describes the information displayed on each control panel menu and the interactions and settings that are controlled by that menu.

System Menus

The system menu structure allows the user to control the unit through the control panel. Operational settings, optional parameters, troubleshooting and error information are controlled through this menu system.
The Barrel Status menu, shown in Figure 8, is the home screen for the unit. During normal operation, this screen is displayed on the control panel.
BARREL STATUS
12:51P MAR 04
MENU ON OFF DFRST SPIN
#1 FREEZE 87 #2 FREEZE 68
Figure 8. Main Screen
Table 2
Button Description
MENU Displays the Main menu
ON Turns on the highlighted barrel
OFF Turns off the highlighted barrel
DFRST Initiates a manual defrost on the highlighted barrel. Manual defrost is lim-
SPIN Turns off refrigeration, if on and mixes the product in the barrel.
ited to a two minute cycle. If the barrel is off, this selection initiates a full defrost cycle.

Main Menu

To access the menus for the system from the Barrel Status menu, press the MENU button. This displays the Select menu, shown in Figure 9. The Setup and Maintenance menus are accessed through the buttons shown at the bottom of the display.
NOTE: If security is turned on, the MAINT button is not displayed.
© 2010, IMI Cornelius Inc. - 15 - Publication Number: 621360041TBSER
Viper Service Manual
SELECT
12:51P MAR 04
BACK MAIN SETUP MAINT GO
UNIT DATA ERROR STATUS ERROR LOG
Figure 9. Select Screen
When the Select menu is displayed, the Unit Data is highlighted. To access the Unit Data menu, press the GO button. Refer to Table 3, the Select menu, for button functions.
Table 3
Button Description
BACK Returns the display to the Barrel Status menu
MAIN Highlighted to indicate current menu
SETUP Opens the Setup menu
MAINT Displays and opens Maintenance menu, if security is off or accessed
GO Opens the highlighted selection on the current menu
Unit Data Menu
The Unit Data screen, shown in Figure 10, provides information about the individual barrels in the system. When DATA is highlighted and the GO button is pressed on the Select menu, the display shows the first barrel in the system and some of its parameters. To view the next barrel, press the Down arrow on the right side of the display. Continue pressing the Down arrow to view the data on all barrels in the system. Refer to Table 4 for button functions.
UNIT DATA
12:51P MAR 04
BACK BRL UNIT VER
#1 FREEZE OUTLET TEMP 33
VISCOSITY: 9 REFRIGERATION: ON AV OZ/HR/3 HR: 80 DEFROST TIMER: 2:50:23
o
F
Figure 10. Unit Data Screen
Table 4
Button Description
BACK Returns the display to the Select menu
BRL Displays the data for a single barrel, including Outlet Temperature, Vis-
cosity, if Refrigeration is On or Off, Average Ounces of product dispensed per hour for the last three hours and Defrost Lock Time until next defrost
Publication Number: 621360041TBSER - 16 - © 2010, IMI Cornelius Inc.
Error Status Menu
Viper Service Manual
Table 4
Button Description
UNIT Displays the real time input voltage and frequency being supplied to the
unit and the low, high or both refrigeration pressures for the unit, depend­ing on the settings of the compressor sensors in the Unit Data screen, as shown in Figure 31.
VER Displays the version number of the UI, I/O, Motor and Status boards
installed in the unit
The Error Status menu, shown in Figure 11, displays any current errors con­tained in the Error Messages List.
There are three types of errors that can be generated by the system. These error types and a description of their meaning are listed in Table 5.
Table 5
Error Type Description
System A System error displays the same information across all barrels in the unit.
Barrel A Barrel error is limited to a specific barrel
Limp A Limp error is an error that doesn’t prevent the unit from operating, but limits
functionality.
The possible messages from the control system and a description of the errors are shown in Table 6. The button functions are described in Table 7.
ERROR STATUS
12:51P MAR 04
BACK
#1 MOTOR PWR COMP HIGH
Figure 11. Error Status Screen
Table 6
Displayed
Message
HOT GAS Off System Occurs when the barrel outlet temperature exceeds
HOT BARREL Off System Occurs when the barrel outlet temperature exceeds
MTR CTL FAIL System The motor controller reported a fault, triacs are
TCO OPEN System TCO (Temperature Cut Off) has occurred on the
Action Type Description
o
F for more than one minute and the compres-
115 sor is on.
o
115
F for more than one minute and the compres-
sor is off.
shorted and there is a loss of motor on/off function­ality.
refrigeration oulet of a barrel or the 24VAC power has failed.
© 2010, IMI Cornelius Inc. - 17 - Publication Number: 621360041TBSER
Viper Service Manual
Table 6
Displayed
Message
COMM ERR IO System A communications error occurred between the IO
COMM ERR MTR
BARREL REFRG
MTR CAL ERR Barrel Occurs if a calibration initiated from the Motor
CMP HI-PRESS System This error is generated if the low side pressure is
BRL TEMP OUT Barrel Indicates thermistor reading is at min. or max. value
MOTOR STALL Barrel Motor stalled. A two minute defrost and two retries
METER FAIL Barrel Metering chip is nonresponsive.
BRL RECOVER Barrel Barrel viscosity is too high. Motor cannot rotate.
COM ERR STAT Limp Communications error, the system continues to run.
HIGH VOLTAGE System The unit is experiencing voltage greater than
LOW VOLTAGE System The unit is experiencing voltage less than 200VAC.
NO MOTOR Barrel The motor is reporting no current draw when acti-
FILL ERROR Barrel Barrel has not filled properly within 2 min.
H
2O OUT System Water pressure is too low for proper operation. This
2 OUT System CO2 pressure is below required operating pressure.
CO
SYRUP OUT Barrel Syrup pressure is below required operating pres-
CLOCK System The real time system clock is not functioning.
MTR CTL EEPROM
Action Type Description
board and the UI board.
System UI board cannot communitcate with the motor
Barrel Compressor has been on for more than 45 minutes
System Motor control EEPROM is bad or missing.
board.
and the viscosity of the barrel is not satisfied with no draws.
Setup menu fails. This indicates that the power draw of the motor is outside the allowable calibra­tion range.
greater than 55 psi for greater than 30 sec. Only checked when barrels are in freeze mode and com­pressor is running more than two minutes.
for 30 sec. without changing.
are allowed, then the message is displayed,
Automatic two minute defrost to thaw barrel.
260VAC. System is enabled when line voltage drops below 260VAC for two minutes.
System is enabled when line voltage rises above 200VAC for two minutes.
vated. Motor is unplugged or thermal cutoff has tripped.
error occurs when the H water pressure for more than one second. Error is cleared 10 sec. after restoral.
This error occurs when the CO or low CO Error is cleared 10 sec. after restoral.
sure. This error occurs when the syrup sensor reports no or low syrup pressure for more than one second. Error is cleared 10 sec. after restoral.
2 pressure for more than one second.
2O sensor reports no or low
2 sensor reports no
Table 7
Button Description
BACK Returns the display to the Select menu
RESET Allows the user to reset the highlighted error (This button only appears on
Publication Number: 621360041TBSER - 18 - © 2010, IMI Cornelius Inc.
certain manually resettable errors.)
Error Log Menu
Viper Service Manual
The Error Log screen displays information about any errors generated by the unit, as shown in Figure 12. If no errors have been detected, the right side of the display is blank. To view more than one error, press the Up or Down arrows on the right side of the display to scroll through the error log. Button functions are described in Table 8.

Setup Menu

ERROR LOG
12:51P MAR 04
BACK
ERROR NUMBER 2 #3 COMM FAIL
OCCUR: 2:44 PM 3/04/08 CLEAR: 2:52 PM 3/04/08
Figure 12. Error Log Screen
Table 8
Button Description
BACK Returns the display to the Select menu
The Setup menu provides the means to change various settings to local preferences, such as time, date and temperature formats. It also allows the user to adjust the unit for the type of syrup being served in each barrel.
To access the Setup menu from the Barrel Status menu, press the MENU button. This displays the Select menu, shown in Figure 9. Press the SETUP button to display the Setup menu, shown in Figure 13. To set up the unit options, use the Up and Down arrows on the right side of the control panel to highlight the choices on the display. When the desired option is highlighted, press the GO button to access that menu.
SELECT
12:51P MAR 04
BACK MAIN SETUP MAINT GO
BACK MAIN SETUP MAINT GO
CLOCK SETUP EVENTS SETUP VISC SETUP CO2 SETUP OPTION SETUP BARREL MAINT
Figure 13. Setup Screen
© 2010, IMI Cornelius Inc. - 19 - Publication Number: 621360041TBSER
Viper Service Manual
Button Description
BACK Returns the display to the Select menu
MAIN Opens the Main menu
SETUP Highlighted to indicate that the Setup menu is displayed
MAINT Displays button and opens Maintenance menu, if security is off or
GO Opens the highlighted selection on the current menu
Clock Setup Menu
Select the CLOCK SETUP menu from the Select menu, shown in Figure 13. This displays the Clock Setup menu, shown in Figure 14. Button functions are described in Table 10.
Table 9
accessed
Setting the Clock
CLOCK SETUP
12:51P MAR 04
BACK CLOCK DST - +
TIME: 11:00 AM DATE: 01/01/00
Figure 14. Clock Setup Screen
Table 10
Button Description
BACK Returns the display to the Select menu
CLOCK Highlighted to indicate that the Clock menu is displayed
DST Opens the Daylight Savings Time menu
- Decrements the highlighted field
+ Increments the highlighted field
Select the CLOCK SETUP menu from the Select menu, shown in Figure 13. This displays the Clock Setup menu, shown in Figure 14. To set the time, perform the procedure in Table 11.
Table 11
Step Action Procedure
Set clock time Use up and down arrows on the right side of the con-
1
Select hour field Use the left and right arrows to select hour field
2
Publication Number: 621360041TBSER - 20 - © 2010, IMI Cornelius Inc.
trol panel to highlight TIME display on screen.
Viper Service Manual
Table 11
Step Action Procedure
Set correct hour Use the + or - buttons at the bottom of the display to
3
Select minute field Use the left and right arrows to select minute field.
4
Set correct minute Use the + or - buttons at the bottom of the display to
5
Select AM/PM field If the 12 hour clock option is selected, use the left and
6
Set AM/PM Use the + button at the bottom of the display to set the
7
set the proper hour.
set the proper minute.
right arrows to select the AM/PM field.
AM/PM setting.
To set the date, perform the procedure in Table 12 and refer to Figure 14.
Table 12
Step Action Procedure
Set date Use up and down arrows on the right side of the con-
1
Select month field Use the left and right arrows to select month field
2
Set correct month Use the + or - buttons at the bottom of the display to
3
Select day field Use the left and right arrows to select day field.
4
Set correct day Use the + or - buttons at the bottom of the display to
5
Select year field Use the left and right arrows to select year field.
6
Set correct year Use the + or - buttons at the bottom of the display to
7
trol panel to highlight DATE display on screen.
set the correct month.
set the correct day.
set the correct year.
Setting Daylight Sav­ings Time
Once the date and time are set properly, daylight savings time settings can be done. The Daylight Savings Time menu is shown in Figure 15. If daylight savings time is off, press the + button to activate it. The button functions are described in Table 13. To set daylight savings time, perform the procedure in Table 14.
NOTE: Default daylight savings settings are for the U.S.
CLOCK SETUP
12:51P MAR 04
BACK CLOCK DST - +
Figure 15. Daylight Savings Setup Screen
DST: ON SPRING MONTH: MAR
SPRING WEEK: 1 FALL MONTH: OCT FALL WEEK: L
Table 13
Button Description
BACK Returns the display to the Select menu
© 2010, IMI Cornelius Inc. - 21 - Publication Number: 621360041TBSER
Viper Service Manual
Button Description
CLOCK Opens the Clock menu
DST Highlighted to indicate DST menu
- Decrements the highlighted field by one increment
+ Increments the highlighted field by one increment
Step Action Procedure
1
2
3
4
5
6
7
8
9
10
11
Table 13
Table 14
Set daylight savings time Press DST button at the bottom of display to open day-
light savings time display, shown in Figure 15.
Select DST Use the up and down arrows to select DST.
Set DST on Use the + button to turn on daylight savings time.
Select SPRING MONTH Use up and down arrows to select SPRING MONTH.
Set SPRING MONTH Use the + or - buttons at the bottom of the display to set
the correct month.
Select SPRING WEEK Use up and down arrows to select SPRING WEEK.
Set SPRING WEEK Use the + or - buttons at the bottom of the display to set
the correct week. The choices are 1, 2, 3 or L.
Select FALL MONTH Use the up and down arrows to select FALL MONTH.
Set FALL MONTH Use the + or - buttons at the bottom of the display to
set the correct month.
Select FALL WEEK Use up and down arrows to select FALL WEEK.
Set FALL WEEK Use the + or - buttons at the bottom of the display to set
the correct week. The choices are 1, 2, 3 or L.
Events Setup Menu
When the daylight savings settings are complete, press the BACK button to save the settings and return to the Select menu, shown in Figure 13.
The Events Setup menu, shown in Figure 16, allows the user to set sleep periods for the unit and to lock out the defrost cycle during peak busy times. Sleep periods and defrost lockouts may be programmed for any day of the week or for all days of the week, depending on business requirements. Button functions are described in
EVENT SETUP
12:51P MAR 04
BACK CLEAR - +
Table 15.
DAY SUN SLEEP 10:00 PM
WAKEUP 08:00 AM DEFROST LOCK 1: 05:15 PM DEFROST LOCK 2: 09:30 AM DEFROST LOCK 3: ------------
Figure 16. Event Setup Screen
Publication Number: 621360041TBSER - 22 - © 2010, IMI Cornelius Inc.
Setting Events
Viper Service Manual
Table 15
Button Description
BACK Returns the display to the Select menu and saves current settings
CLEAR Clears the highlighted field on the menu
- Decrements the highlighted field
+ Increments the highlighted field
When the sleep and wakeup settings and the defrost lockout settings are complete, press the BACK button to save the settings and return to the Select menu, shown in Figure 13.
Events setup allows the user to set sleep periods for the unit and to lock out the defrost cycle during peak busy times. Sleep periods and defrost lockouts may be programmed for individual days of the week or for all days of the week, depending on location requirements.
Setting the Sleep and Wakeup Times
Sleep and wakeup times are set on the Event Setup menu shown in Figure 16. To set the sleep and wakeup times, perform the procedure in Table 16.
When the unit goes into a sleep period, the barrel automatically defrosts for four minutes. Also, if a barrel is off when the sleep cycle starts, the barrel continues to be in the off state when the sleep cycle ends. If a barrel is running when the sleep cycle starts, the barrel starts back up when the sleep cycle ends.
Wakeup should be scheduled for approximately one hour before product is required.
NOTE: Setting the sleep time after midnight requires you to set it dur­ing the next day. (i.e. Saturday night at 1am must be set as 1am on Sunday, etc.) If this type of setting is required, you cannot set another sleep cycle during that following day, i.e. Sunday.
Table 16
Step Action Procedure
1
2
3
4
5
6
7
8
Set sleep and wakeup times
Select DAY The DAY field is highlighted.
Set DAY Use the + or - buttons at the bottom of the display
Select SLEEP Use the up and down arrows to highlight SLEEP.
Set hour field Use the left and right arrows to select the hour field
Set minute field Use the left and right arrows to select the minute
Select AM/PM field If the 12 hour clock option is selected, use the left and
Select DAY for Wakeup Repeat Steps 2 and 3 as needed.
Open the Events Setup menu, shown in Figure 16 by pressing the GO button.
to set the desired day or all days.
and press the + or - buttons to change the hour.
field and press the + or - buttons to change the minutes (15 min. increments)
right arrows to select the AM/PM field. Then press the + or - buttons to change between AM and PM.
© 2010, IMI Cornelius Inc. - 23 - Publication Number: 621360041TBSER
Viper Service Manual
Step Action Procedure
9
When the sleep and wakeup settings are complete, press the BACK button to save the settings and return to the Select menu, shown in Figure 13.
Setting Defrost Lockout
From the Barrel Status menu, shown in Figure 8, press the MENU button and then press the SETUP button to display the Setup menu. Use the up and down arrows on the right of the control to highlight the Events Setup menu, then press GO to enter the menu (Figure 16).
The lockout can also be set day by day or for all days with up to three lockout periods of four hours each per day. These lockout periods may be overlapped to provide from four to 12 hours of defrost lockout, if desired. When the periods are overlapped, it is recommended that the defrost lockouts for sequential lockout periods be overlapped by 15 minutes. The defrost lockout affects all barrels in the unit. To set the defrost lockouts, perform the procedure in Table 17.
Table 16
Select WAKEUP Use the up and down arrows to highlight
WAKEUP and repeat Steps 5 through 7 to set the WAKEUP times.
Table 17
Step Action Procedure
1
2
3
4
5
6
7
8
9
10
11
12
Set defrost lockout Open Events Setup menu, shown in Figure 16
Select DAY The DAY field is highlighted.
Set DAY Use the + and - buttons at the bottom of the display
to set the desired day or all days.
Select DEFROST LOCK 1Use the up and down arrows to highlight
DEFROST LOCK 1.
Set hour field Use left and right arrows to select hour field
Use the + and - buttons at the bottom of the display to set the desired hour.
Set minute field Use left and right arrows to select minute field
Use the + and - buttons at the bottom of the display to set the desired minute (in 15 min. increments).
Select AM/PM field If the 12 hour clock option is selected, use the left
and right arrows to select the AM/PM field.
Use the + button at the bottom of the display to set the AM/PM field.
Select DEFROST LOCK 2Repeat Steps 2 through 10 for the DEFROST
LOCK 2 time, if desired.
Select DEFROST LOCK 3Repeat Steps 2 through 10 for the DEFROST
LOCK 3 time, if desired.
When the defrost lockout settings are complete, press the BACK button to save the settings and return to the Select menu, shown in Figure 13.
Viscosity Setup Menu
The viscosity maintained in the freeze barrels depends on the type of product being served. Some products are served best at a higher viscosity, while others require a lower viscosity for best quality. The Viscosity menu, shown in Figure
Publication Number: 621360041TBSER - 24 - © 2010, IMI Cornelius Inc.
Viper Service Manual
17, allows the user to adjust the viscosity in each barrel to the optimum setting for each type of syrup.
VISC SETUP
12:51P MAR 04
BACK ALL - +
#1 VISC RANGE 6 48-110 #2 VISC RANGE 6 48-110
Figure 17. Viscosity Setup Screen
Refer to the Installation manual for the recommended settings based on syrup type.
Table 18
Step Action Procedure
Set viscosity range From the Setup screen (Figure 13), open the Viscos-
1
Select barrel Use the Up and Down arrows to highlight the desired
2
Select range field Use Left and Right arrows to select range field
3
Set range Use the + or - buttons at the bottom of the display to
4.
Select barrel Repeat Steps 2 through 4 for all barrels.
5
ity Setup screen, shown in Figure 17.
barrel.
set the desired range.
To set all barrels in the system to the same viscosity setting, perform Steps 1 through 4 in Table 18 and then press the ALL button at the bottom of the display while highlighting the viscosity setting you desire for all the barrels. When the viscosity settings are complete, press the BACK button to save the settings and return to the Select menu, shown in Figure 13.
CO2 Setup Menu
The CO2 Setup menu, shown in Figure 18, allows the user to adjust the percentage of time that the CO for the unit when using different syrups. The valves use a one second interval dura­tion time. Setting a valve to 45 activates the valve for 450 msec. out of each second.
The CO
2 pulse valves may be set between 0 and 100%, in 5% increments. To
change the settings of the CO
© 2010, IMI Cornelius Inc. - 25 - Publication Number: 621360041TBSER
2 pulse valves are open. This provides an adjustment of the overrun
2 pulse valves, perform the procedure in Table 19.
CO2 SETUP
12:51P MAR 04
BACK ALL - +
#1 CO2 PULSE %: 45 #2 CO2 PULSE % 45
#3 CO2 PULSE % 45
Figure 18. CO2 Pulse Valve Setup Screen
Viper Service Manual
Step Action Procedure
1
2
3
4
5
To set all barrels in the system to the same duration setting, perform Steps 1 through 4 in Table 19 and then press the ALL button at the bottom of the display while highlighting the duration setting you desire for all the barrels. When the duration settings are complete, press the BACK button to save the settings and return to the Select menu, shown in Figure 13.
Option Setup Menu
Table 19
Set CO2 Pulse valve dura­tion
Select barrel Use the Up and Down arrows to highlight the desired
Select duration field Use Left and Right arrows to select duration field
Set duration Use the + or - buttons at the bottom of the display to
Select barrel Repeat Steps 2 through 4 for all barrels.
From the Setup screen (Figure 13), select the CO2 Setup screen, shown in Figure 18.
barrel.
set the desired duration.
The Option Setup menu allows the user to set the various options available in the system. These options are listed in Table 20. The Option Setup menu is shown in Figure 19.
The functions of the display buttons change, depending on the highlighted selection on the Option Setup screen. Refer to Table 20 for a listing of the button functions for the various field selections.
o
OPTION SETUP
12:51P MAR 04
BACK oF oC
TEMP FORMAT DATE FORMAT USA TIME FORMAT 12 HR POS LIGHTING ALWAYS #1 SYRUP TYPE: FCB #2 SYRUP TYPE: FCB
F
Figure 19. Option Setup Screen
Table 20
Option Button 2 Button 3 Button 4 Button 5
Tem p Fo r ma t
Date Format USA EURO
Time Format 12 HR 24 HR
POS Lighting OFF ALWAYS SLEEP
#1 SYRUP TYPE FCB FCB-L FUB FUB-L
#X SYRUP TYPE FCB FCB-L FUB FUB-L
°F °C
When all the options are set to the desired settings for the unit, press the BACK button to store these settings and return to the Select menu, shown in Figure 13.
Publication Number: 621360041TBSER - 26 - © 2010, IMI Cornelius Inc.
Viper Service Manual
Setting the Temperature Format
Setting the Date Format
Setting the Time Format
The temperature format displayed by the unit may be set to either Centigrade or Farenheit. Press the °F button to display readings in Farenheit and press the °C button to display readings in Centigrade.
The date format can be displayed in either United States or European format. To display U.S. date format, press the USA button. This displays the date in mm/dd/ yy format. Press the EURO button to display the date in dd/mm/yy format.
Time format can be displayed in either 12 or 24 hour format. To display the clock settings in 12 hour format (1:08 P), press the 12 HR button. To display settings in 24 hour format (23:05), press the 24 HR button.
Setting the POS Light­ing
Setting the Type of Syrup
Barrel Maintenance
POS Lighting is controlled by the POS LIGHTING field on the Option Setup menu (Figure 19). To turn off the merchandiser lighting, press the OFF button while the POS LIGHTING field is highlighted. To turn on the merchandiser lighting permanently, press the ALWAYS button. To turn the merchandiser lighting on and off with the Sleep settings, press the SLEEP button.
Syrup type for each barrel may be selected by highlighting the desired barrel and pressing the appropriate button, FCB, FCB-L, FUB or FUB-L. FCB is for Frozen Carbonated Beverage, FCB-L is for Frozen Carbonated Beverages ­Light (diet), FUB is for Frozen Uncarbonated Beverages and FUB-L is for Fro­zen Uncarbonated Beverages - Light (diet). Each of these settings provides the proper viscosity and temperature settings for the type of syrup being used.
Refer to the next section “Barrel Maintenance Menu” to perform barrel mainte­nance on the system.

Maintenance Menu

The Maintenance menu, shown in Figure 20, is only available to service person­nel or other authorized users, if security is enabled.
© 2010, IMI Cornelius Inc. - 27 - Publication Number: 621360041TBSER
Viper Service Manual
The Maintenance menu is used for barrel maintenance, diagnostics, BRIX setup and system information. It allows the technician access to the unit during peri­odic maintenance or troubleshooting problems. It sometimes overrides the nor­mal inputs from the system in order to perform this task.
Barrel Maintenance Menu
SELECT
12:51P MAR 04
BACK MAIN SETUP MAINT GO
BARREL MAINT MANUAL DIAG TOTALS BRIX SETUP MOTOR SETUP SYSTEM
Figure 20. Select Screen
The Barrel Maintenance menu, shown in Figure 21, allows the service techni­cian to clean, rinse and sanitize the system on a barrel by barrel basis. The bar­rel must be OFF in order to perform the barrel maintenance functions. As a warning to the technician, the “Do Not Drink” status indicators are illuminated on the barrels when entering the Barrel Maintenance menu.
The procedure for purging a barrel is shown in Table 21. The procedure for filling a bar­rel is shown in Table 22. The procedure for running the scraper blade only on a barrel is shown in Table 23 and the procedure for waking up a barrel is shown in Table 25.
Purging a Barrel
BARREL MAINT
12:51P MAR 04
BACK PURGE FILL SPIN
#1 PURGE #2 OFF
Figure 21. Barrel Maintenance Screen
Table 21
Step Action Procedure
1 Purge a barrel From the Select menu (Figure 20), open the Barrel
2 Select the barrel Use the Up and Down arrows to highlight the desired
3 Select Purge function Press the PURGE button at the bottom of the display
Maintenance menu, shown in Figure 21.
barrel
to select purge for the highlighted barrel
Publication Number: 621360041TBSER - 28 - © 2010, IMI Cornelius Inc.
Filling a Barrel
Viper Service Manual
Table 21
Step Action Procedure
4 Empty the barrel Place a container under the barrel dispensing valve,
5 Complete the procedure When the barrel is completely empty, close the dis-
open the dispensing valve and dispense all the product from the barrel. As the product level lowers in the bar­rel, partially close the valve to avoid spurting.
pensing valve and press the STOP button at the bot­tom of the display to turn off the purge. Press the BACK button to return to the Select menu.
NOTE: When the PURGE button is pressed, it changes to a STOP but­ton.
Table 22
Step Action Procedure
Fill a barrel From the Select menu (Figure 20), open the Barrel
1
Select the barrel Use the Up and Down arrows to highlight the desired barrel
2
Select Fill function Press the FILL button at the bottom of the display to fill
3
Finish filling the barrel When the fill completes, intermittently open and close
4
Complete the procedure Press the STOP button at the bottom of the display to
5
Maintenance menu, shown in Figure 21.
the highlighted barrel
the relief valve to bleed CO2 from the barrel.
stop the fill. Press the BACK button to return to the Select menu.
Run the Barrel Motor
Rinsing a Barrel
NOTE: When FILL button is pressed, it changes to a STOP button.
Table 23
Step Action Procedure
Run a barrel From the Setup menu (Figure 20), open the Barrel
1
Select the barrel Use the Up and Down arrows to highlight the desired
2
Select Run function Press the SPIN button at the bottom of the display to
3
Stop the procedure Press the STOP button at the bottom of the display to
4
Maintenance menu, shown in Figure 21.
barrel
start the motor in the highlighted barrel
stop the motor. Press the BACK button to return to the Select menu.
NOTE: When RUN button is pressed, it changes to a STOP button.
When rinsing a barrel, the Barrel Maintenance menu cannot be used. The pro­cedure must be done manually, as described in Table 24.
© 2010, IMI Cornelius Inc. - 29 - Publication Number: 621360041TBSER
Viper Service Manual
Step Action Procedure
1
2
3
4
5
6
7
Manual Diagnostic Menu
The Manual Diagnostic menu, shown in Figure 22, is used for troubleshooting and testing the system on a component basis. It allows the service technician to turn individual components of the system on and off for troubleshooting pur­poses. Pressing the BACK button returns the display to the Main menu and turns off all the active loads.
This screen should only be used by qualified technicians to troubleshoot the system.
Table 24
Empty a barrel If the barrel has liquid in it, use the Purge procedure in
Table 21 to empty the barrel.
Rinse a barrel Go to the Barrel Status menu (Figure 6).
Select the barrel Use the Up and Down arrows to highlight the desired
barrel
Turn off the barrel Press the OFF button at the bottom of the display to turn
off the barrel.
Override Water Solenoid Press the manual bypass on the water solenoid and fill
the barrel with plain water. Use the relief valve on the faceplate of the barrel to relieve pressure and com­pletely fill the barrel.
Run the barrel motor Use the procedure in Table 23 to rinse the barrel.
Complete the procedure Repeat the Purge procedure in Table 21 to empty the
barrel.
MANUAL DIAG
12:51P MAR 04
BACK ON OFF
POS LIGHTING OFF COMPRESSOR #1 OFF
HOT GAS SOL #1 ON PRODUCT SOL #1 OFF CO2 VALVE #1 OFF LLS VALVE #1 OFF
Figure 22. Manual Diagnostic Screen
The items shown in Table 25 are for the first barrel in the unit only. The list includes additional components with their related barrel number, i.e. Hot Gas Sol #2, etc.
CAUTION: The compressor can suffer damage if it is allowed to run without either a hot gas defrost or liquid line solenoid open. The software is designed to turn on the hot gas sole­noid for a barrel to prevent damage from occurring.
Table 25
Display Description
POS LIGHTING This selection turns the Point-Of-Sale lighting on and off
COMPRESSOR #1 This selection turns the compressor on and off
HOT GAS SOL #1 This selection opens and closes the hot gas solenoid for barrel 1
Publication Number: 621360041TBSER - 30 - © 2010, IMI Cornelius Inc.
Totals Menu
Viper Service Manual
Table 25
Display Description
PRODUCT SOL #1 This selection opens and closes the product solenoids and valves
CO2 VALVE #1 This selection opens and closes the CO
LLS VALVE #1 This selection opens and closes the Liquid Line solenoid and
BRL MOTOR #1 This selection turns the barrel 1 motor on and off
for barrel 1
2 solenoic and valve for
barrel 1
valve for barrel 1
To move between one set of barrel parameters to the next, press the Right or Left arrow keys. This moves the highlight to the first entry of the second barrel, etc.
The Totals menu (Figure 23) shows the cumulative run time of the system, a component of the system or an error. Some individual items may be reset, where available, by highlighting the item and pressing the RESET button. The RESET button is only displayed for items that may be reset, such as hours since the motor seal was changed or the number of compressor cycles. Pressing the BACK button returns to the Choose menu and saves the changes. The descrip­tion of each item in the Totals list is shown in Table 26.
To move between one set of barrel parameters and the next, press the Right or Left arrow keys. This highlights the first entry of the second barrel, etc.
TOTALS
12:51P MAR 04
BACK
POWER ON 00079:34 SLEEP 00034:21
SYSTEM ERROR 00003:45 COMP RUN #1 00006:53 COMP CYCLES #1 0000012 BARREL REFG #1 00006:54
Figure 23. Totals Screen
Table 26
Display Value Description
POWER ON HHHHH:MM Shows how long the machine has been powered
SLEEP HHHHH:MM Shows how long the machine has been in sleep
SYSTEM ERROR HHHHH:MM Shows how long the machine has had a system
COMP RUN #1 HHHHH:MM Shows how long the compressor has been run-
COMP CYCLES #1 ######## Shows the number of compressor cycles since
BARREL REFG #1 HHHHH:MM Shows how long the barrel has been in refrigera-
MOTOR ON #1 HHHHH:MM Shows how long the motor has been running on a
MOTOR SEAL #1 HHHHH:MM Shows how long it has been since the motor seal
up
mode
error
ning since the last reset
the last reset
tion mode
specific barrel
was inserted on a specific barrel
© 2010, IMI Cornelius Inc. - 31 - Publication Number: 621360041TBSER
Viper Service Manual
DEFROST #1 HHHHH:MM Shows how long the barrel has been in defrost
FILL CYCLES #1 ######## Shows the number of times the barrel has filled
SYRUP RUN #1 HH:MM:SS Shows how long the syrup solenoid has been
SYRUP OUT #1 HHHHH:MM Shows how long the syrup has been out
ERROR #1 HHHHH:MM Shows how long the barrel has had an error
BRIX Setup Menu
BRIX is important to the quality of the final product. The BRIX menu facilitates the extraction of a sample of product from the unit for BRIX measurement. There is an automatic three second dispense of product that produces a constant volume so that a BRIX comparison can be made between samples. To perform a BRIX test, perform the procedure in Table 27.
Table 26
Display Value Description
mode
with product
open
Table 27
Step Action Procedure
Set status of barrels Make sure the status of all barrels is OFF on the con-
10
11
12
13
14
1
Remove drip tray/access
2
panel
Set product supply valve Turn product supply valve to the 180 degree (BRIX)
3
Access BRIX menu Press MENU. Then press the SETUP button.
4
Set barrel for BRIX From the Setup screen (Figure 13), open the BRIX
5
6
Select barrel Use the Up and Down arrows to highlight the desired
7
Perform BRIX procedure Locate the appropriate barrel sample tube and hold a
8
9
Measure BRIX Place adequate amount of the product on a refracto-
Adjust BRIX If BRIX is not within the proper range, adjust the
Clear hose When BRIX measurement is complete, press the
Set product supply valve Turn product supply valve back to the (BARREL)
trol display.
Remove the drip tray, loosen the two screws holding the access panel and remove it. (Figure 25)
position for the barrel you are testing. (See Figure 26)
Setup screen, shown in Figure 24
Use the Up and Down arrows to highlight BRIX SETUP. Press the GO button at the bottom of the dis­play.
barrel for BRIXing.
cup under it.
Open the valve at the end of the sample tube. Press the BRIX button, and wait 3-5 seconds. The product pump pumps product for approximately 3 seconds. After the sample is dispensed Press BRIX twice more to dispense product two more times. Discard all three of these samples.
Press the BRIX button a fourth time. Collect this sample in a cup and close the valve at end of the sample tube.
meter and read the BRIX value. A target BRIX read­ing of 13.0 (+/- 1.0) is normally desired for sugar­based syrups. Lower values for some diet syrups can be specified. Check with the syrup manufacturer if you are not sure.
syrup/water mixture by opening or closing the syrup valve, shown in Figure 27, and then retest.
valve on the bottom of the appropriate water solenoid to flush out the hose before closing the valve at the end of the hose.
position for the barrel you tested. (See Figure 26)
Publication Number: 621360041TBSER - 32 - © 2010, IMI Cornelius Inc.
Viper Service Manual
Table 27
Step Action Procedure
Replace access panel Replace the hoses and the access panel. Secure the
15
panel with the two screws removed in Step 2. This completes the procedure.
BRIX SETUP
12:51P MAR 04
BACK BRIX STOP
Splash Panel
Retaining Screws
Figure 25. Splash Panel Mounting Screws
#1 OFF #2 OFF
Figure 24. BRIX Setup Screen
Flow
Product
Supply
Val ve
Barrel 3
Product
Supply
Val ve
Barrel 1
Product
Supply
Val ve
Barrel 2
Flow
Controls
Barrel 1
Flow
Controls
Barrel 2
Controls
Barrel 3
Valves shown in BARREL position w/ Splash Panel & Solenoid Covers Removed
Figure 26. Product Control Valves
© 2010, IMI Cornelius Inc. - 33 - Publication Number: 621360041TBSER
Viper Service Manual
Motor Setup Menu
Manual
Syrup Control
Val ve
Syrup Flow
Control
Figure 27. Flow Controls with Panel Removed
Manual
Water Control
Val ve
Water Flow
Control
(Do Not Adjust)
Motor setup is initially done at the factory and does not need to be changed unless a barrel motor or barrel components are replaced, adjusted or removed and reinstalled.
This menu allows you to select different motor types for each barrel in the unit and run the calibration procedure on that motor. This procedure correlates the actual viscosity of the barrel and its contents with the electrical characteristics of the motor. These characteristics are stored by the control system and are accessed at the time of calibration for each particular motor.
Because of the differences between motor characteristics from different motor suppliers, the Motor menu, shown in Figure 28, is used to adjust the unit settings to match the motor type being installed. It provides the correct table of electrical characteristics from imbedded settings located in non-volatile memory. Perform the procedure described in Table 30 to change the motor type setting. Insure that the motor type matches the settings listed in Table 28 based on model.
NOTE: If the Motor Type shown is “DEFAULT”, it indicates that the EEPROM is not installed or is faulty.
Table 28
Model Motor Type
All 60 Hz 1
All 50 Hz 2
Publication Number: 621360041TBSER - 34 - © 2010, IMI Cornelius Inc.
Viper Service Manual
MOTOR SETUP
12:51P MAR 04
BACK CAL - +
#1 MOTOR TYPE 1 #2 MOTOR TYPE 2
Figure 28. Motor Setup Screen
Table 29
Button Description
BACK Returns the display to the Select menu
CAL Verifies the motor settings of a new motor.
- Decrements the highlighted field
+ Increments the highlighted field
Table 30
Step Action Procedure
Set barrel for Motor Type From the Maintenance menu (Figure 20), open the
1
Select the barrel Use the Up and Down arrows to highlight the desired
2
Select the proper motor
3
type
Complete the procedure Press the BACK button at the bottom of the display to
4
Motor Setup menu, shown in Figure 28.
barrel
Press the + or - buttons at the bottom of the display to select the proper motor type
save the settings and return to the Setup menu.
Calibrating a Motor
Calibration of motors is typically required when a motor or any associated barrel components (i.e., scraper blades, seal, faceplate, etc.) is removed or replaced. Calibration establishes a new baseline for the motor assembly. This allows the system to determine proper viscosity settings for the new motor. When a new motor is installed in the unit or any of the barrel components are changed or adjusted, the motor should be calibrated by performing the procedure described in Table 31.
!
CAUTION:
Calibration must be performed when the product in the barrel is com­pletely liquid. There must not be any ice on the scraper blade.
Table 31
Step Action Procedure
Re-assemble barrel It is VERY IMPROTANT that calibration be performed
1
Defrost the barrel Select DFRST on the Barrel Status menu (Figure 8). It is
2
© 2010, IMI Cornelius Inc. - 35 - Publication Number: 621360041TBSER
when the product in the barrel is in a 100% liquid state. There can be no ice in the barrel or on the beater bar.
very important that calibration be performed on a fully defrosted barrel (outlet temperature must be 41 must be purged and refilled with new product). Selecting DFRST while the barrel is off initiates a full defrost cycle.
°F or barrel
Viper Service Manual
Step Action Procedure
3
4
5
6
7
System Menu
The System menu is used during production to set up the unit. It also allows the technician to view and change the number of barrels, the number of compres­sors and the number of pressure sensors present in the sytem, as shown in Fig­ure 29. The description of these items is shown in Table 32.
NOTE: Changing these values to a number that doesn’t match the physical number of items in the system generates errors or turns off some of the equipment present in the system.
When the individual fields are highlighted, the display buttons change. Refer to Table 33 for a listing of the button functions for the various field selections.
Table 31
Select Motor Setup Use the Up and Down arrows to highlight MOTOR
SETUP on the Maintenance menu.
Select the barrel Use the Up and Down arrows to highlight the desired
barrel
Start calibration Press the CAL button at the bottom of the Motor Setup
menu to start the calibration process.
The calibration cycle continues for five (5) minutes to allow the gearbox to stabalize at temperature and ensure correct callibration. At that time the motor stops and it is calibrated.
Complete the procedure Press the BACK button at the bottom of the display to
return to the Maintenance menu.
SYSTEM
12:51P MAR 04
BACK - +
# OF BARRELS 2 # OF COMPRESSOR 1
SECURITY OFF CMP #1 SENSORS BOTH CMP #X SENSORS LOW
Figure 29. System Screen
Table 32
Display Description
# OF BARRELS This selection shows the number of barrels in the system and
# OF COMPRESSORS This selection shows the number of compressors in the sytem.
SECURITY This selection allows the user to turn security on or off.
CMP #X SENSORS This selection allows the user to choose which pressure sensors
allows the user to change the quantity.
Currently this number is fixed at one and cannot be changed.
are read for the compressor.
Table 33
Option Button 2 Button 3 Button 4 Button 5
# of Barrels - +
# of Compressors - +
Security OFF ON
Publication Number: 621360041TBSER - 36 - © 2010, IMI Cornelius Inc.
Security
Viper Service Manual
Table 33
Option Button 2 Button 3 Button 4 Button 5
Compressor Sensors NONE LOW HIGH BOTH
System security is located on the System menu. It allows a supervisor or service technician to keep unauthorized personnel from accessing the Maintenance menu. This feature is activated on the System menu. When security is turned on, users can only access the Main and Setup menus.
Pressing the extreme Left and Right buttons (Buttons 1 and 5) on the bottom of the display simultaneously and holding them for approximately five (5) seconds unlocks the Security menu. If security is left ON on the System menu, when the system times out or when the user goes back to the Main menu, security is re­activated and the Maintenance menu is not accessable. To disable security, access the System menu, highlight Security and press the OFF button.
Pressure Sensors
Pressure sensors allow the technician to view the high and low-side pressures in the refrigeration system without having to connect external gauges to the sys­tem. To view this information, go to the Unit Data screen shown in Figure 10 and press the UNIT button. The System screen, shown in Figure 30, is displayed.
The number and type of sensors in the unit are typically set at the factory and should not be changed. Data is only available on the sensors installed. If a sen­sor that is not installed in the unit is activated, incorrect data is displayed. It is possible to install sensors in the field, but it is not recommended because sys­tem refrigerant charge may be lost.
The low side pressure sensor is used to monitor compressor reversal. If it is not installed on the unit, the unit may not operate properly.
SYSTEM
12:51P MAR 04
BACK NONE LOW HIGH BOTH
# OF BARRELS 2 # OF COMPRESSOR 1
SECURITY OFF CMP #1 SENSORS BOTH
Figure 30. System Screen
When the pressure sensors selection is highlighted, pressing the NONE button removes the pressure readings from the Unit Data screen, shown in Figure 31. If LOW is selected, only the low pressure reading is displayed. If HIGH is selected, only the high pressure reading is displayed.
NOTE: If a pressure sensor is open, the pressure reading is 386 PSIG.
© 2010, IMI Cornelius Inc. - 37 - Publication Number: 621360041TBSER
Viper Service Manual
UNIT DATA
12:51P MAR 04
BACK BRL UNIT VER
LINE VOLTAGE: 229 LINE FREQUENCY: 60
LOW PSIG #1: 60 HIGH PSIG #1: 174
Figure 31. Unit Data Screen
Publication Number: 621360041TBSER - 38 - © 2010, IMI Cornelius Inc.
Viper Service Manual

OPERATION

MAINTAINING PRODUCT QUALITY

It has been determined that the following factors can affect the rate at which product quality diminishes (as indicated by a change in product appearance).
1. Dispensed Product Throughput
2. Programmed Defrost Scheduling
3. Viscosity Setting
Cornelius recommends the following instructions be read and followed relative to operating and establishing settings for the FCB equipment. Cornelius equipment service manuals contain instructions on how to program settings within the control system. Operators who have not been trained on servicing Cornelius FCB equip­ment should not attempt to modify equipment settings but should contact an accred­ited service provider.
Cornelius makes the following recommendations to help assure maximum product quality.

Dispensed Product Throughput

FCB equipment is designed to provide a high throughput of frozen carbonated prod­uct to meet peak draw demands. Where low product throughput is experienced, there is the potential for product quality to diminish. The information shown in Table 34 outlines the minimum throughput per barrel that must be dispensed on a 24 hour basis.
Volume of dispensed product per barrel per 24 hours required to maintain product quality.
NOTE: Cornelius recommends that, in conditions where the FCB machine is
operational and the minimum throughput (as described in Table 34 is not met on a per barrel basis, product should be dispensed and dis­carded to increase throughput and help assure that product quality is maintained.
NOTE: Data in Table 34 assumes equipment has been correctly installed,
commissioned and calibrated as per directions contained in all techni­cal literature published by Cornelius and the recommendations con­tained in this document have been followed.
Cornelius recommends that, in conditions where the FCB machine is operational and the minimum throughput is not met on a per barrel basis, product should be dis­pensed and discarded to increase throughput and help assure that product quality is maintained.
Table 34
Viper Viscosity <
48 oz. 60 oz
4Viscosity > 4
© 2010, IMI Cornelius Inc. - 39 - Publication Number: 621360041TBSER
Viper Service Manual

Programmed Defrost Scheduling

The control system in the Viper system includes a function to automatically defrost product in the barrel at programmed intervals based on the viscosity of the product. Programmed defrosts are scheduled frequently to ensure that product quality within the barrel is maintained. Failure to defrost regularly during periods of low throughput allow increased ice crystal size, with a possible decrease in product quality. Regular throughput of dispensed product replenishes the barrel frequently with liquid and reduces the requirements for programmed defrosts. The unit automatically senses the lack of throughput and maintains the product quality by defrosting the barrels more frequently when throughput is slow.
Sleep Mode Recom­mendations
Cornelius recommends programming a sleep period for the machine during any amount of time over 3 hours that the unit is not used. This increases the life of the machine and reduces energy consumption.
A wakeup time must be programmed after the sleep period to return the unit to nor­mal operation. Cornelius recommends programming the wakeup time for approxi­mately 20 minutes before the product is needed. This is the time recommended if the ambient temperature is at 75°F. The times will vary depending on the ambient temperature (a higher ambient temperature requires more time for the product to cool).

Viscosity Setting

CO2 Setting

The control system includes a function to select the desired product viscosity. This function is referred to as “Viscosity Setting”. There are selectable viscosity ranges from 1-9. The higher the number selected, the more viscous the frozen product in the barrel becomes. This increased viscosity is achieved by freezing the product in the barrel to a lower temperature thereby increasing ice crystal size/growth. As the ice crystal size increases, however, there is potential for product quality to diminish.
Cornelius recommends that the viscosity settings be set at the lowest possible set­ting to achieve the desired drink quality. In most typical installations using a sugar­based syrup, acceptable drink quality can be achieved by programmed viscosity set­tings in the range of 3-5 for Viper.
Diet FCB syrups freeze much more readily than sugar based syrups, so the viscosity set­ting should be selected at the minimum value available (which is 1 or 2 for the Viper).
The control system includes a function to adjust the desired product overrun. This function is referred to as the “CO in 5% increments.
Typically, 40 to 45% settings on the pulse valves creates 100% overrun.
2 Setting”. This setting is adjustable from 0 to 100%
Publication Number: 621360041TBSER - 40 - © 2010, IMI Cornelius Inc.
Viper Service Manual

STARTING THE UNIT

The following table, Table 32, describes the basic procedure for starting the Viper. Refer to the Operator’s Manual (P/N 621360041TBOPR) for details on operation of the Viper unit.
Table 32
Step Action Procedure
1. Open the merchandiser Open the merchandiser and expose the control panel.
2. Turn on the barrels Turn all barrels on by pressing the buttons labeled ON to start the normal refrigeration process.
3. Close the merchandiser. Close the merchandiser. In approximately 20 minutes, product is ready to serve.
© 2010, IMI Cornelius Inc. - 41 - Publication Number: 621360041TBSER
Viper Service Manual

PREVENTATIVE MAINTENANCE

SUMMARY

There are no daily maintenance procedures required on the Viper unit other than normal cleanup of spills or overspray and emptying the drip tray. Normal equip­ment maintenance intervals are listed in Table 33. It is recommended that a pre­ventative maintenance procedure be performed every six (6) months. This procedure should include all of the maintenance items described in Table 33 and the following sections.
Maintenance Procedure Frequency of Maintenance
Table 33
Preventative Maintenance Summary
Clean Air Filter Monthly or more often, as necessary (See
Check BRIX Every six months or when changing syrup
Seal Change Every six months
Clean Condenser Coil Every six months or as necessary
Sanitize Unit Every six months or when changing syrup
Check Water Filter Every six months or if water pressure in the
Change or Rotate Scraper Blades Every six months
Check for Leaks Every six months
Clean BIB Connectors & check opera­tion of syrup sold out switches
Check Clock Setting Every six months or when changing to or
Change caged o-ring on models w/ Motorman Valves.
!
CAUTION:
Table 34)
types (See “BRIX Setup Menu” on page 32.)
types
system is low.
Every six months or when changing syrup types.
from daylight savings time
Every six months or when changing syrup types.
Only trained and certified electrical, plumbing and refrigeration technicians should service this unit. ALL WIRING AND PLUMBING MUST CONFORM TO NATIONAL AND LOCAL CODES.
Publication Number: 621360041TBSER - 42 - © 2010, IMI Cornelius Inc.
Viper Service Manual

MONTHLY MAINTENANCE

A general inspection of the machine for leaks and cleanliness should be done on a weekly basis. Any abnormal conditions should be reported and corrected when noticed.

Cleaning Air Filter

The air filter should be cleaned at least once a month (more often in harsh envi­ronments). Perform the procedure in Table 34 to clean the air cleaner.
Step Action
Table 34
1. Open the merchandiser
2. Remove the filter, shown in Figure 32, by grasping the two tabs and sliding it straight out the front of the unit.
3. Wash the filter with clean water. Shake out the excess water.
4. Reinstall the air filter, mesh side down.
5. Close the merchandiser.
Ta b Ta b
Figure 32. Air Filter Location
© 2010, IMI Cornelius Inc. - 43 - Publication Number: 621360041TBSER
Viper Service Manual

SEMI-ANNUAL MAINTENANCE

There are several procedures that should be completed on a semi-annual basis. There procedures help to insure proper operation of the system on a long term basis. Refer to Table 33 for a listing of the semi-annual maintenance items.

Inspecting and Replacing Scraper Blades

Scraper blades should be inspected for wear during the semi-annual mainte­nance check. To inspect the blades, perform the procedure in Table 35
Step Action
1. See “Purging a Barrel” on page 28.
2. When purging is complete, remove power from the unit.
Table 35
3. Remove the four (4) 3/4 in. nuts from the faceplate and slowly remove the faceplate from the unit.
4. Slowly pull the blade assembly out of the barrel.
5. Rinse and inspect the blade assembly for signs of wear.
6. If there is significant wear on the blades, Reverse or replace them, depending on the wear pattern.
7. Replace the blade assembly and bushings (PN 2392) onto the motor shaft in the barrel.
8. Re-install the faceplate and replace the four (4) mounting nuts. Hand tighten the nuts untill the faceplate makes contact with the gasket, then use a wrech to tighten the nuts an additional 1/4 turn. Be careful not to over tighten the nuts or cracking of the faceplate may result.
9. Refill the barrel with product. See “Filling a Barrel” on page 29.
10. When barrel fill is complete, perform a motor calibration. See “Calibrating a Motor” on page 35.

Cleaning the Syrup Connections

Syrup connections should be cleaned at least every six months, or when syrup types are changed. Perform the procedure in Table 36 to clean the BIB connec­tors.
Table 36
Step Action
1. Open the merchandiser and select OFF for the barrel to be cleaned.
2. Remove the quick disconnect from the BIB container.
3. Fill a suitable pail or bucket with soap solution.
Publication Number: 621360041TBSER - 44 - © 2010, IMI Cornelius Inc.
Viper Service Manual
Step Action
4. Submerge the disconnect in the soap solution and then clean it using a nylon bristle brush. (Do not use a wire brush). Rinse with clean water and dry.

Servicing Motorman Dispensing Valves

NOTE: Dispensing valves with caged o-rings should be serviced (lubri-
cated) every six months and o-rings should be replaced once a year.
Refer to Figure 34 and perform the procedure in Table 37 to lubricate or change the caged o-rings in each dispensing valve on the unit.
Table 36
Figure 33. Self-Closing Dispensing Valve
Table 37
Step Action
1. Defrost freeze cylinders, shut unit down, disconnect electrical power from Unit.
2. Perform the procedure in Table 38 to empty the barrels.
3. Remove the hex nuts and flat washers securing the faceplate to the freeze barrel, then remove the faceplate from the barrel.
4. Carefully remove the large o-ring from the faceplate.
5. Unscrew the relief valve from the faceplate.
6. Disassemble the dispensing valve (see Figure 33).
7. Remove the two screws and hold-down plates securing the spring housing to the dispensing valve body, then remove the housing.
8. Remove the torsion spring from the dispensing valve.
© 2010, IMI Cornelius Inc. - 45 - Publication Number: 621360041TBSER
Viper Service Manual
Step Action
9. Remove the knob, spring, sleeve, and lever from the dispensing valve.
10. Remove spring fitting from dispensing valve.
11. Press the valve with the caged O-ring, down and out of the dispensing valve body.
12. Carefully remove the caged O-ring from the valve.
13. Wash all the parts in warm water. Remove all traces of syrup and lubricant, especially from the faceplate, o-rings, and dispensing valve. If parts are excessively coated, wipe clean with a paper towel to remove excess syrup and lubricant, especially from caged o-ring and dispensing valve. Use a brush (provided with the unit) to clean the faceplate relief valve passages and the inside of the knob.
14. Submerge all the parts in a 4-percent solution of approved sanitizing agent for the amount of time recommended by the sanitizer manufacturer.
15. Remove the parts from the sanitizing solution and place them on clean paper towels.
16. Assemble the dispensing valve.
17. Lubricate the caged o-ring. Carefully install the caged o-ring onto the valve from the straight end (opposite tapered end). Lubricate the grooves that the o­ring rides to fill in all void areas around the o-ring.
Table 37

Sanitizing the System

Emptying a Barrel
18. Carefully install the valve with the caged o-ring in the dispensing valve body.
19. Install the spring fitting, knob and lever parts, torsion spring and spring hous­ing assembly by reversing the removal procedure. Do not tighten down the hold-down plates securing the spring housing at this time.
20. After re-installing the faceplate, turn the dispensing valve spring housing to the left (counterclockwise) to put tension on the tension spring. Apply just enough tension so that the dispense valve shaft returns to the closed position after release. Do not overtighten. Overtightening results in a high activation force on the valve.
NOTE: Use Dow-Corning DC-111 (P/N 321471000) light grade silicone
lubricant to lubricate the o-rings.
The syrup systems should be sanitized every 180 days by a qualified service technician following the sanitizer manufacturer’s recommendations or when changing syrup types.
The sanitizing process consists of emptying the barrel, washing the lines and barrel, cleaning the BIB connectors, rinsing and refilling the system.
To empty the barrel, perform the procedure in Table 38.
Table 38
Step Action
1. From the Barrel Status menu, press the DFRST button.
2. When the barrel is defrosted, go to the Maintenance menu. If the security fea­ture is active, access the Maintenance menu by by pressing and holding the far left and right buttons simultaneously for approximately five seconds.
Publication Number: 621360041TBSER - 46 - © 2010, IMI Cornelius Inc.
Viper Service Manual
Table 38
Step Action
3. Remove the splash panel. See “BRIX Setup Menu” on page 32.
4. Turn the Product/BRIX valve 90 degrees clockwise to shutoff product to the barrel.
5. Place a large waste container under the dispense valve and drain as much product as possible from the barrel.
6. When the pressure in the barrel drops, from the Barrel Maintenance menu, press the PURGE button to repressurize the barrel with CO lowers in the barrel, partially close the valve to avoid spurting.
7. Disconnect the BIB from the unit.
2. As product level
NOTE: The unit should be sanitized every 180 days by a qualified service technician following the sanitizer manufacturer’s recommendations.
Flushing the System of Syrup
After emptying the barrel, the barrel should be flushed of product before pro­ceeding with the sanitizing procedure. Perform the procedure in Table 39.
Table 39
Step Action
1. Fill a clean 5-gallon pail with plain water.
2. Connect a sanitizing fitting (p/n cc 28688) to the BIB connector. Put the con­nector in the bucket of water.
3. Make sure the Product/BRIX valve is in the BRIX position.
4. Hold a waste container under the brix tube to collect syrup from the syrup line and open the manual syrup flow valve (see Figure 27) to start filling the syrup line with plain water. Continue to hold the manual syrup flow valve open until clean water starts coming out of the BRIX tube.
5. Release the manual syrup flow valve and turn the Product/BRIX valve to the Product position.
6. Open the manual water flow valve (see Figure 27) to start filling the barrel with water. At the same time, open the faceplate relief valve until water comes out.
7. When the barrel is full, press the SPIN button on the Barrel Status menu, while highlighting the appropriate barrel. This starts the scraper blade. Allow blade to operate for fifteen seconds.
8. Turn barrel OFF by pressing the OFF button.
9. Place a waste container under the barrel dispensing valve. Open the dispens­ing valve and dispense all wash water from the barrel. When the pressure in the barrel drops, from the Barrel Maintenance menu, press the PURGE button to re-pressurize the barrel with CO rel, partially close the valve to avoid spurting.
2. As the wash water level lowers in the bar-
10. Perform rear barrel seal replacement. See “Barrel Motor Seal Replacement” on page 51.
11. Perform inspection and replacement of scraper blades. See “Inspecting and Replacing Scraper Blades” on page 44.
12. Perform a leak test on the barrel. See “Motor Seal Leak Test” on page 53.
Sanitizing the Barrel
Sanitize the syrup system and barrel by performing the procedure shown in Table 40.
© 2010, IMI Cornelius Inc. - 47 - Publication Number: 621360041TBSER
Viper Service Manual
Step Action
1. Use a clean 5-gallon pall filled with a sanitizing solution and water at a temper­ature of 90º F to 110º F (32º C to 43º C). The sanitizing solution must have 200-ppm available chlorine.
2. Connect a sanitizing fitting (p/n cc 28688) to the BIB connector. Put the con­nector in the bucket of sanitizing solution.
3. Make sure the Product/BRIX valve in the BRIX position.
4. Hold waste container under the BRIX tube to collect flush water from the syrup line and open the manual syrup flow valve to start filling the syrup line with sanitizing solution. Continue to hold the syrup flow valve open until sanitizing solution starts coming out of the BRIX tube.
5. Turn the Product/BRIX valve in the Product position.
6. Manually override (open) the syrup flow valve to fill the barrel with sanitizing solution.
7. Fill the barrel with sanitizing solution by opening the faceplate relief valve until sanitizing solution comes out of the relief port.
8. Hold a 16 oz. cup under the dispense valve. Hold the dispense valve fully open until the cup is full.
9. Use the brush provided with the unit to clean the relief port and clean the out­let of the dispense valve with sanitizing solution.
10. From the Barrel Status menu, press the SPIN button while highlighting the
appropriate barrel. This starts the scraper blade. Allow blade to operate for minimum of 10, but no more than 15 minutes. Turn the scraper blade off by pressing the OFF button.
11. Place a large container under the dispense valve and drain as much sanitizing solution as possible from the barrel.
12. When the pressure in the barrel drops, from the Barrel Maintenance menu, press the PURGE button to re-pressurize the barrel with CO solution level lowers in the barrel, partially close the valve to avoid spurting.
Table 40
2. As sanitizing
Flushing the System Flush the wash water from the system by performing the procedure in Table 41.
!
CAUTION:
Flush the system thoroughly, residual sanitizing solution left in the system may create a health hazard.
Table 41
Step Action
1. Turn the Product/BRIX valve to the Product position.
2. Manually open the manual water flow valve to start filling the barrel with wash water. At the same time, open the faceplate relief valve until water comes out.
3. From the Barrel Status menu, press the SPIN button while highlighting the appropriate barrel. This starts the scraper blade. Allow the blade to operate for fifteen seconds, then turn barrel OFF by pressing the OFF button.
4. To drain the water from the system, turn the barrel OFF.
5. Place a container under the barrel dispensing valve. Open the dispensing valve and dispense all rinse water from the barrel. When the pressure in the barrel drops, from the Barrel Maintenance menu, press the PURGE button to re-pressurize the barrel with CO partially close the valve to avoid spurting.
6. Remove the sanitizing fitting (p/n cc 28688) from the BIB connector and con­nect a BIB containing syrup to the syrup line.
2. As the wash water level lowers in the barrel,
Publication Number: 621360041TBSER - 48 - © 2010, IMI Cornelius Inc.
Viper Service Manual

Cleaning the Water Filter

The water filter screen in the rear of the unit should be removed and cleaned every six months, during the preventative maintenance procedures. To inspect and clean the filter screen, perform the procedure in Table 42. Refer to Figure 34 for water filter location.
Table 41
Step Action
7. Rotate the Product/BRIX valve to the BRIX position and open the valve at the end of the tube.
8. Hold a waste container under the BRIX tube to collect the sanitizing solution from the syrup line and open the manual syrup flow valve to start filling the syrup line with syrup. Continue to hold open the syrup flow valve until syrup starts coming out of the BRIX tube.
9. Perform a BRIX setup. See “BRIX Setup Menu” on page 32.
10. Fill the barrel with product as described in Table 22.
11. Perform motor calibration. See “Calibrating a Motor” on page 35.
Water
Filter
Figure 34. Water Filter Location
Table 42
Step Procedure
1.
Disconnect power from the unit.
2. Turn off and disconnect the water supply to the unit.
3. Remove the rear cover of the unit.
4. Disassemble the water filter. (Use two wrenches, if necessary)
5. Remove the filter and rinse it under running water to remove any debris.
6. Reassembe the filter.
7. Connect and turn on the water supply to the unit.
8. Check the filter for any leaks.
9. Replace the rear cover.
10. Reconnect power to the unit.
© 2010, IMI Cornelius Inc. - 49 - Publication Number: 621360041TBSER
Viper Service Manual

COMPONENT REPLACEMENT

The following are procedures for servicing the major components of the Viper system.

BARREL MOTOR SEAL REPLACEMENT

The barrel motor seal is typically replaced during the semi-annual preventative maintenance procedure. The seal location is shown in Figure 40.

Removing the Existing Seal

To remove the existing barrel seal, perform the procedure in Table 42.
Step Action
1. Purge the barrel. See “Purging a Barrel” on page 28.
2 When purging is complete, disconnect power from the unit.
3 Open the relief valve on the front of the barrel to ensure that pressure is
4 Remove the four (4) 3/4 in. nuts from the faceplate and slowly remove the
5 Carefully pull the blade assembly out of the barrel.
6 Remove any remaining product from the barrel.
7 Rinse and inspect the blade assembly and barrel for signs of wear. (Replace if
8 Remove the seal and spacer from the rear of barrel.
9 When old seal and spacer are removed, use a clean soft cloth to clean
Motor
Seal
Figure 40. Front View Inside of Barrel w/ Faceplate off
Motor Shaft
Table 42
released. CAUTION: Barrel pressure MUST be relieved before removing the faceplate!
faceplate from the unit.
necessary)
between the motor shaft and the barrel to remove any remaining product.
Publication Number: 621360041TBSER - 50 - © 2010, IMI Cornelius Inc.
Viper Service Manual

Installing a New Seal

Perform the procedure in Table 43, Figure 41 and Figure 42 to replace the barrel seal with a new seal. When installing a new seal, NEVER use oil or silicon based lubricants. This can cause the new seal to rotate and LEAK!
Table 43
Step Action
1 Remove the old seal and the spacer (see Table 42).
2 Remove the new seal from its packaging.
3 Clean the spacer.
4 Slide the spacer (with slots toward the motor housing) over the motor shaft.
5 Lubricate the inside of the seal using Dow Corning 111 Lubricant and Sealant.
(Note, Do not lubricate the static side(outer portion) of the seal. This could cause the seal to rotate in the bore and leak.)
6 Carefully slide the seal over the motor shaft and press it firmly until is fully
seated against the spacer and the seal cannot be pushed any further.
7 Run your finger around the perimeter of the seal to insure it is flush against the
back of the barrel.
8 Reinstall the blade assembly. Rotate the assembly while inserting it to insure
that it is fully seated on the motor shaft. (Figure 30)
9 Lubricate the faceplate o-ring.
10 Re-install the faceplate and replace the four (4) mounting nuts. Hand tighten
the nuts until the faceplate makes contact with the gasket, then use a wrench to tighten the nuts an additional 1/4 turn. Be careful not to overtighten the nuts or cracking of the faceplate may result.
11 Clean the drip tray and tube using warm water.
12 Perform the seal leak test in Table 44.
Spacer
(Note Orientation)
Motor Shaft
Figure 41. View of Motor Shaft with Spacer on Shaft
© 2010, IMI Cornelius Inc. - 51 - Publication Number: 621360041TBSER
Viper Service Manual

Motor Seal Leak Test

After replacing the motor seal, it is advisable to test the seal under pressure before refilling the barrel with product. Perform the procedure in Table 44 to ver­ify that the seal is installed properly and not leaking. When the seal passes the leak test, perform the motor calibration routine on page 35.
Seal
Spacer
Figure 42. Side View of Seal and Spacer on Motor Shaft
Step Action
1 Turn the 3-way Product Supply valve behind the splash panel (see Figure 26)
to the Off (horizontal) position.
2 From the Maintenance menu, press the FILL button for the barrel. This pres-
surizes the barrel with CO
3 Observe the water and syrup fill solenoids.
4 If there are NO seal leaks, these solenoids cycle off within a minute and
remain off.
5 If the solenoids cycle off and remain off, the barrel seal is not leaking and the
barrel is ready to be filled with product.
6 If the solenoids do not cycle off, or cycle off and on, the seal is leaking and the
seal installation process described in Table 42 and Table 43 must be rea­peated.
7 If there are no leaks, turn the 3-way Product Supply valve to the BRIX posi-
tion.

BARREL MOTOR REPLACEMENT

The barrel motor does not require any special alignment when being replaced. It mounts on four (4) bolts. These bolts are screwed into threaded bosses in the foam pack barrel assembly, as shown in Figure 43. To replace a barrel motor, perform the procedure in Table 45.
Table 44
2.
Publication Number: 621360041TBSER - 52 - © 2010, IMI Cornelius Inc.
Viper Service Manual
Barrel Motor
Mounting
Bolts
Foam
Pack
Figure 43. Side View of Motor Showing Two of the Four Mounting Bolts
Table 45
Step Action
1 Perform the procedure in Steps 1 through 8 of Table 42.
2 Disconnect the motor from the circuit by unplugging the connector.
3 Remove the motor by removing the four (4) bolts holding the motor to the bar-
rel assembly.
4 Install the replacement motor and connect the connector.
5 Replace the motor seal, scraper blade assembly onto the motor shaft in the
barrel, as described in Table 42.
6 Reconnect power to the unit.
7 Refill the barrel with product. See “Filling a Barrel” on page 29.
8 When barrel fill is complete, perform a motor calibration. See “Calibrating a
Motor” on page 35.

MOTOR RUN CAPACITOR REPLACEMENT

The motor run capacitors, shown in Figure 44, are mounted at the rear of the unit between the barrel motors. To replace a run capacitor, perform the proce­dure in Table 46.
Figure 44. View of Motor Run Capacitors
Step Action
1 Disconnect power from the unit.
Motor Run Capacitors
Table 46
© 2010, IMI Cornelius Inc. - 53 - Publication Number: 621360041TBSER
Viper Service Manual
Step Action
2 Remove the rear cover.
3 Disconnect the spade clips from the defective capacitor.
4 Remove the mounting screw from the base of the capacitor.
5 Replace the capacitor.
6 Reverse the procedure to install the new capacitor.
7 Reconnect power to the unit.
8 Refill the barrel with product. See “Filling a Barrel” on page 29.
9 When barrel fill is complete, perform a motor calibration. See “Calibrating a

WATER PUMP REPLACEMENT

The water pump is located just behind the water filter in the center of the unit. It is mounted on a bracket facing the rear of the unit. To replace the pump, perform the procedure in Table 47.
Table 46
Motor” on page 35.
Water Pump
Figure 45. Rear View of Water Pump
Table 47
Step Action
1 Disconnect power from the unit.
2 Remove the rear cover and the side panels.
3 Disconnect the CO
4 Remove and drain the water supply from the unit.
5 Disconnect the quick disconnect input and output water lines from the bottom
of the pump.
6 Disconnect quick disconnect CO
7 Remove the four (4) screws and locknuts holding the pump to the mounting
bracket.
8 Replace the pump.
9 Reverse the procedure to install the new pump.
2 supply from the unit.
2 input from the bottom of the pump.
Publication Number: 621360041TBSER - 54 - © 2010, IMI Cornelius Inc.
Viper Service Manual

CO2 PULSE VALVE REPLACEMENT

The CO2 valves, shown in Figure 46, are located near the center of the unit and are mounted on a bracket to the bottom of the chassis. They control the injection of CO
2 with the syrup/water combination to provide the final mixture for the prod-
uct. There are one of these solenoids for each barrel in the system. When working in this area, care should be taken not to damage or misalign the CO orfice. To replace one of these valves, perform the procedure in Table 48.
2 metering
Mounting Tie-Wrap
Figure 46. Side View of CO2 Pulse Valves
Pulse Valve
(1
OF 3)
Pulse Valves
Figure 47. Front View of CO2 Pulse Valves
Table 48
Step Action
1 Disconnect power from the unit.
2 Disconnect the CO
3 Remove the rear cover and side panels.
4 Carefully remove the tubing from the rear of the valve.
5 Remove the mounting tie-wrap from the rear of the valve.
6 Replace the valve.
7 Reverse the procedure to reinstall the valve into the unit.
© 2010, IMI Cornelius Inc. - 55 - Publication Number: 621360041TBSER
2 supply from the unit.
Viper Service Manual
Step Action
Table 48
8 Test the unit for CO
9 Replace the rear and side panels.
10 Reconnect power to the unit.
2 leaks.

SYRUP AND WATER SHUTOFF SOLENOID REPLACEMENT

The syrup and water shutoff valves and solenoids are mounted on a single bracket, as shown in Figure 48. Perform the procedure in Table 49 to replace the valves.
Figure 48. View of Shutoff Solenoids w/ Access Panel Removed
Syrup
Shutoff
Valves &
Solenoids
Table 49
Step Procedure
1 Disconnect power from the unit.
2 Disconnect and bleed the water and syrup lines to remove pressure from the
system.
3 Remove the rear cover and side panels.
4 Remove the water and syrup lines from the input to the valves.
5 Remove the Wye tube from the output (top) of the valves by loosening the
nuts between the Wye and the check valves.
Check
Wye
Retaining
Nuts
6 Remove the two (2) locknuts from the base of the assembly and remove it
from the unit.
7 Remove the check valve from the valve being replaced.
Val ves
Publication Number: 621360041TBSER - 56 - © 2010, IMI Cornelius Inc.
Viper Service Manual
Step Procedure
8 Install the check valve on the new valve and solenoid.
9 Remove the valve and solenoid from the mounting bracket and replace it.
10 Reverse the procedure to reinstall the bracket into the unit.
11 Reconnect power to the unit and test the system.

CONDENSER FAN MOTOR REPLACEMENT

The fan motor is located on the top of the unit, as shown in Figure 49. If the motor needs replacement, perform the procedure in Table 50 to remove and reinstall a motor.
Table 49
Locking
Nuts
Figure 49. Condenser Fan Mounting Nuts
Locking
Table 50
Step Action
1 Disconnect power from the unit.
2 Disconnect the fan by unplugging the connector.
3 Remove the four locking nuts securing the fan to the top of the condenser.
4 Lift the fan and shroud off the unit.
5 Remove the four screws and washers securing the fan to the shroud.
6 Replace the fan.
7 Reverse the procedure to install the new fan.
8 Reconnect power to the unit.
Nuts

COMPRESSOR REPLACEMENT

Caution should be used when working with refrigerants. To replace a compres­sor, perform the procedure in Table 51.
NOTE: Be sure to reclaim refrigerant when replacing compressor.
© 2010, IMI Cornelius Inc. - 57 - Publication Number: 621360041TBSER
Viper Service Manual
Step Action
1 Disconnect power to the unit.
2 Remove the rear cover and side panels.
3 Label and disconnect all wires from the compressor.
4 Remove the four (4) mounting nuts from the base of the compressor.
5 Remove all insulation from around fittings.
6 Unsweat the suction line from the compressor.
7 Cut the discharge line on top of the compressor (Make sure line can be
8 Remove the compressor and unsweat the discharge line.
9 Reverse the procedure to install the new compressor.
10 Replace the dryer.
11 Evacuate the system and recharge it according to the Cornelius nameplate on
12 Replace the side panels and rear cover.
13 Reconnect power to the unit and test the unit.
Table 51
spliced back together.)
the unit.

HOT GAS SOLENOID REPLACEMENT

The hot gas valves and solenoids, shown in Figure 50, are located near the cen­ter of the unit on the right and left sides. Whenever a hot gas valve and solenoid is replaced, the dryer should be replaced at the same time.
Hot Gas Solenoid
Step Action
1 Disconnect power from the unit.
2 Remove the side panels.
3 Reclaim refrigerant from the system.
4 Carefully remove the insulation and remove the hot gas valve and solenoid.
5 Replace the hot gas valve and solenoid.
6 Replace the insulation around the valve and solenoid.
7 Replace the filter-dryer.
Figure 50. Hot Gas Solenoid (1 per barrel)
Table 52
Publication Number: 621360041TBSER - 58 - © 2010, IMI Cornelius Inc.
Viper Service Manual
Step Action
8 Evacuate the system.
9 Restore the proper refrigerant charge to the system and verify proper system
operation.

LIQUID LINE SOLENOID REPLACEMENT

The liquid line valves and solenoids, shown in Figure 51, are located near the center of the unit on the right and left sides. Whenever a liquid line valve and solenoid is replaced, the dryer should be replaced at the same time.
Table 52
Liquid Line
Solenoid
Figure 51. Liquid Line Solenoid (1 per barrel)
Table 53
Step Action
1 Disconnect power from the unit.
2 Remove the side panels.
3 Reclaim refrigerant from the system.
4 Carefully unwrap the foam insulation from around the valve and solenoid.
5 Remove the liquid line valve and solenoid.
6 Replace the liquid line valve and solenoid and insulation.
7 Replace the filter-dryer.
8 Evacuate the system.
9 Restore the proper refrigerant charge to the system and verify proper system
operation.

SYRUP PRESSURE SWITCH REPLACEMENT

The syrup pressure switches (Sold Out Indicators) are located along the left and right sides of the unit near the front. There is one switch for each barrel in the system. Figure 52 shows the syrup pressure switch on the left side of the unit. To replace the syrup pressure switch, perform the procedure in Table 54.
© 2010, IMI Cornelius Inc. - 59 - Publication Number: 621360041TBSER
Viper Service Manual
Step Action
1 Disconnect power to the unit
2 Remove the appropriate side panel.
3 Turn off the syrup supply to the unit.
4 Drain the syrup from the system.
5 Disconnect the wires on the pressure switch.
6 Remove the switch, using two wrenches.
7 Replace the switch with a new unit.
8 Reverse the procedure to install the new switch.
9 Restore the syrup supply and check for leaks.
10 Check switch operation.
Syrup
Pressure
Switch
Figure 52. View of Syrup Pressure Switch and Valve
Table 54

WATER PRESSURE SWITCH REPLACEMENT

The water pressure switch, shown in Figure 53, is located in back of the water filter, when viewed from the rear of the unit. Access to the switch is obtained by removing the left side panel of the unit, when viewed from the rear. To replace the water pressure switch, perform the procedure in Table 55.
Water
Pressure
Switch
Figure 53. View of Water Pressure Switch
Publication Number: 621360041TBSER - 60 - © 2010, IMI Cornelius Inc.
Viper Service Manual
Step Action
1 Disconnect power from the unit.
2 Remove the side panel from the unit.
3 Turn off the water supply to the unit.
4 Drain the water from the system.
5 Disconnect the wires from the switch.
6 Remove the switch, using two wrenches.
7 Replace the switch with a new unit.
8 Reverse the procedure to install the new switch.
9 Restore the water supply and check for leaks.
10 Check to ensure that the new switch is working properly.

DISPLAY BOARD REPLACEMENT

Table 55
When replacing or testing electronic components, be sure to wear a static strap that is connected to a chassis ground. This protects the electronic components from any static charge while working on the unit.
The display board is located on the back of the control panel display. It is mounted to the panel with four (4) phillips head screws, shown in Figure 54.
Mounting
Screws
Figure 54. View of Display Board
Mounting
Screws
Table 56
Step Action
1 Disconnect power from the unit.
2 Connect a static strap to your wrist and any convenient chassis ground on the
unit.
3 Unplug the two (2) connectors from the display board.
4 Remove the four (4) screws from the corners of the board.
5 Replace the board with a new board.
6 Reverse the procedure to install the new board.
© 2010, IMI Cornelius Inc. - 61 - Publication Number: 621360041TBSER
Viper Service Manual

CONTROL BOARD REPLACEMENT

When replacing or testing electronic components, be sure to wear a static strap that is connected to a chassis ground. This protects the electronic components from any static charge while working on the unit.
The control board is located in the E-Box. It is mounted to the panel on four (4) plastic mounting studs, shown in Figure 55.
Step Action
1 Disconnect power from the unit.
2 Connect a static strap to your wrist and any convenient chassis ground on the
unit.
3 Unplug the eight (8) connectors from the control board.
4 Squeeze each of the mounting studs at the corners of the board while care-
fully lifting the board.
5 Replace the board with a new board.
6 Reverse the procedure to install the new board.
Table 57

Software Compatibility

Mounting
Studs
Figure 55. View of Control Board in E-Box
Fruitista sofware is not compatable with standard Viper sofware versions. When ordering replacement/service parts for any unit check the software version of the UI and IO boards and be sure to order the correct board.
Publication Number: 621360041TBSER - 62 - © 2010, IMI Cornelius Inc.
Viper Service Manual

MOTOR BOARD REPLACEMENT

When replacing or testing electronic components, be sure to wear a static strap that is connected to a chassis ground. This protects the electronic components from any static charge while working on the unit. The motor board is located in the E-Box, to the right of the control board. It is mounted to the panel on three (3) plastic mounting studs and a threaded stand­off, shown in Figure 55.
Step Action
1 Disconnect power from the unit.
2 Connect a static strap to your wrist and any convenient chassis ground on the
unit.
3 Unplug the eight (8) connectors from the control board.
4 Squeeze each of the mounting studs at the corners of the board while care-
fully lifting the board.
5 Replace the board with a new board.
6 Reverse the procedure to install the new board.
Table 58
Mounting
Studs
Figure 56. View of Motor Board in E-Box
© 2010, IMI Cornelius Inc. - 63 - Publication Number: 621360041TBSER
Viper Service Manual

TROUBLESHOOTING

TROUBLESHOOTING THE SYSTEM

Problem Probable Cause Remedy
Unit will not run. A. Unit not plugged in.
B. Circuit breaker. C. No power at L1 or L2 on contac-
tor.
D. Low voltage at T1 or T2.
“Sleep” display on bar­rel status menu.
Barrel status OFF. A. Not activated.
Low/High voltage ERROR.
Frozen Barrel A. Low Brix
A. Sleep time set. B. Clock incorrectly set. C. No wake time set.
B. Error has shut off motors.
C. Unit in Diagnostics.
D. Defective touch switch.
A. Line Voltage out of SPEC (spec is
205-255 VAC for 60 Hz and 196­265 VAC for 50 Hz).
B. Error range 180/260.
B. Broken beater bar C. Broken drive coupler. D. No defrost. E. Are defrosts programmed all
days?
A. Plug in unit. B. Reset/replace circuit breaker C. Check voltage at outlet.
Check wiring to contactor.
D. Replace the contactor.
A. Check sleep programming. B. Check sleep programming. C. Check sleep programming.
A. Turn to ON, DFRST or SPIN. B. Correct error & turn to ON,
DFRST or SPIN.
C. Exit diagnostics & turn to ON,
DFRST or SPIN.
D. Replace.
A. 1 - Check line voltage.
2 - Check T1/T2 with Compressor running.
B. Correct line voltage.
A. Correct cause and reset Brix. B. Replace beater bar. C. Replace the drive coupler. D. Program defrosts in unit. E. Reprogram to all days.
Publication Number: 621360041TBSER - 64 - © 2010, IMI Cornelius Inc.
Viper Service Manual

Troubleshooting Product Not Cold

Problem Probable Cause Remedy
Compressor not Run­ning
Compressor Running but not Cooling
Restricted Air Flow A. Dirty filter.
Fan Motor A. Bad connection
A. Barrel not in ON mode. B. No voltage to compressor.
C. Bad start components. D. Compressor’s thermal overload
protector “open”.
E. Open or shorted compressor wind-
ings.
F. Bad control board.
A. Low refrigerant.
B. Restricted condenser/filter. C. Condenser fan motor/blade defec-
tive. D. Hot gas valve leaking or open. E. Defective compressor.
B. Dirty condenser. C. Damaged fins. D. Not enough “clearance” around
unit.
B. Bad motor C. Cracked or bent fan blade
A. Turn barrel to ON. B. Check power at contactor L1, L2 -
T1, T2. C. Check components and wiring. D. Check resistance of compressor
windings and check incoming line
voltage. E. Check resistance of compressor
windings. F. Troubleshoot, replace if necessary.
A. Repair leak and weigh in new
charge. B. Clean or repair. C. Repair or replace.
D. Replace. E. Repair or replace.
A. Clean filter. B. Clean condenser. C. Repair/replace if necessary. D. Ensure proper spacing around
unit.
A. Check/connect B. Replace motor C. Replace fan blade
Hot Gas Valve A. Miswired.
B. Defective coil. C. Defective product delivery board.
No/Low Refrigerant A. Leak. A. Repair and weigh in new charge.
Sensors A. Bad connection.
B. Bad sensor. C. Sensor out of position. D. Defective product delivery board.
A. Correct wiring. B. Replace coil. C. Replace.
A. Correct wiring. B. Replace sensor. C. Reposition sensor and clip. D. Replace.
© 2010, IMI Cornelius Inc. - 65 - Publication Number: 621360041TBSER
Viper Service Manual
Publication Number: 621360041TBSER - 66 - © 2010, IMI Cornelius Inc.
IMI Cornelius Inc.
www.cornelius.com
Loading...