3.1.1 Cut Major Modes ..............................................................................................................................3-1
3.1.2 COAG Major Modes .........................................................................................................................
3.1.3 Bipolar Major Modes ........................................................................................................................
System 5000™ functions and essential circuit information are provided in this section. This section begins
with a description of the key parameters for each mode. This is followed by an overview of how the system
functions and some key operational information for the modules within the system.
3.1 Mode Descriptions
The key functional parameters for each mode are presented here. Nominal mode specifications are provided
in section 1.2.11.
3.1.1 Cut Major Modes
Major
mode
CUTPURE391 KHzNoneNone
Minor
Mode
BLEND 1391 KHz16 pulses
BLEND 2391 KHz11 pulses
BLEND 3391 KHz10 pulses
Activation of Pulse Cut will make the selected cut
mode, Pure Cut, Blend 1, Blend 2, or Blend 3
active for 70 milliseconds every 600 milliseconds.
NOTE: The low duty cycle of Pulsed Cut mode
makes the average power very low – about 12%–
when compared with the power displayed on the
RF frequencyModulation: Number of
Pulses, Time on/off
40µs/10µs
28µs/23µs
26µs/24µs
front panel. The period is also long causing most
ESU analyzers to provide erratic or erroneous
readings. Correct power can be verified by mea
suring the peak to peak current and comparing the
value with the current measured in the non-pulsed
mode.
Modulation: Frequency &
period
20 KHz
50µs
20 KHz
50µs
20 KHz
50µs
3.1.2 COAG Major Modes
Major
mode
COAGPINPOINT391 KHz4 pulses
Minor ModeRF frequencyModulation: Number
of Pulses, Time on/off
10µs/40µs
STANDARD562 KHzSingle pulse39 KHz
SPRAY562 KHzSingle pulse19 KHz
-
Modulation:
Frequency & period
20 KHz
50µs
Activation of Pulse Coag will make the selected
coag mode, either Standard or Spray, active for
2.5 milliseconds every 5 milliseconds. Displayed
power setting will represent the average power
being delivered which is approximately half the
power delivered during the pulses.
Standard and Spray Coag modes are fundamental
ly different from the Cut modes in that the reso
nant circuit of the RF Amplifier and Transformer
combination is excited by the energy of a single
pulse, causing the resonant circuit to ring until
the energy is dissipated. Circuitry in the amplifier
3-1
-
provides further damping to dissipate the energy more quickly to minimize RF leakage effects. Spray Coag
provides the maximum open circuit voltage for which the system is rated.
3.1.3 Bipolar Major Modes
Major
mode
BIPOLARMACRO391 KHzNoneNone
Minor
Mode
MICRO391 KHzNoneNone
RF fre
quency
Modulation: Number of
Pulses, Time on/off
Modulation: Frequency &
period
3.1.4 Advanced Specialty Modes
Specialty ModeEffect
GeneralNormal open surgery mode – Parameters noted above.
FluidsTemporarily increases power upon activation for faster initiation. Duration and power increase
vary with mode and power setting
LapLimits maximum peak voltage for safer laparoscopic surgery. This action does affect the load
curves when in high impedance tissue, or using normally high voltage modes.
3.2 System Overview
Mains power is converted to electrosurgical output
power through the High Voltage Power Supply
(HVPS), the RF Amplifier, and the Transformer
and Output sections of the system.
Mains power is converted to high voltage direct
current power in the HVPS to supply the RF
Amplifier. This universal input power factor cor
rected, single output, switch mode power supply
is adjustable under software control with 10-bit
resolution. The HVPS output and power factor
correction sections of the HVPS can be enabled
or disabled under software control. The HVPS
uses a current mode two-switch forward converter
topology with short circuit protection and over
voltage limiting.
Pulses generated in the RF Controller are ampli
fied to electrosurgical power and voltage levels in
the RF Amplifier and Transformer portions of the
power train. The RF Amplifier and Transformer
form a resonant switched mode amplifier with
multiple outputs that are selected on a mode-bymode basis using relays on the primary and sec
ondary side of the transformers. One transformer
is used for monopolar outputs, while the other
transformer is for the bipolar output.
Electrosurgical power flows from the RF
Amplifier and Transformer sections to the Output
section where the power is switched to the specific
electrosurgical outputs. The Output section also
has circuitry to detect activations from accessories
-
-
-
and the circuitry to perform the Automatic Return
Monitor (A.R.M.™) function to ensure the integ
rity of the dispersive electrode connection.
The power section also includes a number of out
put voltage and current sensors that are used by
the RF Controller for control of power delivery
and by the Monitor to detect errant output condi
tions.
The RF Controller is a Digital Signal Processor
(DSP) that generates an RF Amplifier drive sig
nal based upon measured parameters compared
with settings-based parameters. The pulse train
sequence is a settings-based parameter that is
dependent on the selected mode. Target power,
current limit, voltage limit, and impedance thresh
olds are all settings-based parameters derived
from a load curve that is specific to the front
panel power setting. The RF Controller samples
electrosurgical output voltage and output cur
rent from sensors over 450,000 times per second
and uses these sampled values to calculate output
power and sensed impedance. The output power,
output current, output voltage, and sensed imped
ance are compared with corresponding settingsbased parameters of target power, current limit,
voltage limit, and impedance threshold; respec
tively; and the RF Controller adjusts the width of
individual pulses within each mode-based pulse
train sequence in a closed-loop fashion to control
corresponding output power. The RF Controller
also adjusts the HVPS output more slowly, allow
ing adjustment of the RF Amplifier drive pulses
-
-
-
-
-
-
-
-
-
3-2
Voltage
sampled
450,000
times per
second
Current
sampled
450,000
times per
second
Power
setting –
watts
displayed
on the front
panel
Calculate
measured
resistance
Calculate
measured
power
Compare
Desired
Power
Measured
Power
Adjust
output
waveform
(power)
Patient
0
20
40
60
80
100
120
140
160
180
200
0500 1000 1500 2000
90
35
180
Figure 3.1 RF Controller Block Diagram
to optimize the electrosurgical output waveform.
Finally, the RF Controller minimizes RF leakage
currents using the CONMED Leakage Abatement
System (CLAS™), which imposes a duty cycle on
the electrosurgical output when sensed impedance
and output voltage exceed settings-based imped
ance thresholds and voltage limits in the Coag
modes.
The RF Monitor is also a DSP, but it is used to
monitor the system for a variety of conditions that
could lead to safety problems, including:
• The Monitor has independent sensors for out
put voltage and current, which it uses to cal
-
culate power for comparison with the power
that the RF Controller senses and for com
-
parison with the generator power setting.
• To ensure that the correct outputs are acti
vated, the Monitor also independently
senses current at each of the outputs, look
ing for current flow that would indicate
electrosurgical power at outputs other than
the selected output.
• The Monitor senses the voltage at the output
of the HVPS to ensure that it is reasonable
for the power setting.
• The Monitor senses the audio output to ensure
that a tone occurs whenever electrosurgical
outputs are active.
• The RF Amplifier drive signal is sensed by
the Monitor to detect improper frequencies
or improper pulse sequences for the selected
mode.
• The Monitor independently compares the
activation signal with that seen by the System
Controller to ensure that the activation signal
is consistent.
The Monitor has the capability to independently
disable the electrosurgical output if a problem is
detected.
The System Controller provides the primary con
trol interface to the user and other outside sys
tems, including the serial interface, the activation
relay, tone generation, and displays.
Finally, the Display accepts all user input and provides all user feedback. The Display is controlled
by the System Controller through a serial interface
and illuminates the LED display elements in a
time division multiplexed fashion; the illuminated
LED display elements are actually on less than half
the time. The Display also provides for user input
through the buttons on the control panel, includ
ing switch de-bouncing and conditioning.
Figure 3.2 illustrates the key elements of the sys
tem in block diagram form.
3.2.1 High Voltage Power Supply (HVPS)
The HVPS is comprised of a Power Factor
Control (PFC) section and a Forward Converter
(FC) section. The PFC converts Mains power
to approximately 400 volts using techniques
that ensure the mains current into the supply is
sinusoidal and in phase with the mains voltage.
By doing so, RMS current and harmonic distor
tion are reduced. The Forward Converter then
-
-
-
-
-
3-3
Real Time
Clock (5K)
Indicators:
Power, Mode,
& bipolar
current
Displays
Keyboard
Modes / Power
Activation
Request
System
Controller
RF Controller
RF Monitor
RF Amp
RF Output
Board
HV Power
Supply
Bip
Mon
RFHVSup
MRF H1 SN
MRF H2 SN
MRF FT SN
MRF BP ISN
MRF BP VSN
MRF MP ISN
MRF MP VSN
M HVDC
/HV EN
Tone
Mon
WF EN
WFORM
DAMP
CNTL
RLY DRV
RF BP IS
RF BP VS
RF MP IS
RF MP VS
HV SET
PFC EN
BIP TONE
ACT TONE
AL TONE
RS232
ACT RLY
Serial Interface
Connector
Activation Relay
Connector
Host
Bus
SPI
Patient
RF INH
Mon
RF
Transformer
Board
VARM
Figure 3.2 System Block Diagram
converts the PFC output to an adjustable DC
voltage for use by the RF amplifier.
The System Controller can enable or disable the
PFC section of the HVPS. The PFC is normally
enabled during operation to ensure a resistive load
is presented to the Mains.
The Forward Converter is a switch-mode power
converter that adjusts its operating frequency
between 25KHz and 100KHz to ensure proper
resolution for the commanded output volt
age. Isolation between Mains power and the
-
3-4
HVPS output occurs in the Forward Converter.
Forward Converter output voltage is set from the
RF Controller by the /HVSET signal. The RF
Monitor enables the output of the HVPS. The
forward converter includes current limiting on the
output and has provisions to shutdown when the
output of the Low Voltage Supply exceeds limits.
3.2.2 RF Amplifier and Transformer
The RF Amplifier and Transformer portions use
a switch-mode resonant amplifier to convert the
power from the HVPS to the RF energy neces
-
sary for electrosurgery. One may think of the
amplifier as a high-speed switch that pulses cur
rent through a resonant circuit, which is formed
by the monopolar or bipolar transformer together
with capacitors and inductors connected to the
transformer primary and secondary windings.
Two Metal-Oxide-Semiconductor Field Effect
Transistors (MOSFETs) are connected in a paral
lel fashion provide the switching. The pulses to
drive the gates on the MOSFETs in this arrange
ment come from the RF Controller. Adjusting
the width of the drive pulses regulates the output
power in this arrangement; as the pulses become
longer, the output power increases. As noted in
the RF Controller discussion, the RF Controller
compares the output power with the desired
power and adjusts the pulse width to minimize
the difference.
A drive of several pulses at a frequency that closely
matches the resonant frequency of the amplifier
characterize Cut, Blend, and Pinpoint modes, and
the output pulses substantially correspond to the
drive. Spray and Standard Coag modes, how
ever, are characterized by pulses that occur less
frequently where the amplifier is allowed to “ring”
at its resonant frequency. A damping capability is
provided to enhance the surgical effect by damp
ing the ringing effect for each drive pulse.
Because the transformer windings and the reso
nant frequencies between the cut and coag modes
are different, a method of selecting resonating
components is implemented using relays. These
relays switch in the relevant components for the
selected mode based upon commands from the
System Controller.
A Balun transformer is provided for the
monopolar modes to reduce differences between
the source and return currents, thus reducing RF
leakage.
Finally, the RF Amplifier and Transformer provide
capabilities for sensing RF output current and
voltage. These are transformer-isolated representations of the current flowing in the leads and the
voltage across the output, which are supplied to
the RF Controller and the Monitor processors.
3.2.3 Electrosurgical Outputs
Relays are provided to isolate electrosurgical out
puts and select which outputs are active. The
System Controller selects the appropriate output
relays based upon activation command inputs.
-
-
-
-
-
-
-
The Monitor utilizes sensors implemented on each
electrosurgical output to determine whether cur
rent is flowing only to the correct outputs. In the
event that current flows in an output that is not
selected, the Monitor can independently disable
RF.
The System 5000™ output panel connectors are
illuminated to aid visibility in low lighting situ
ations. This illumination is provided by a single
LED on the display board that is distributed to
the receptacles through a fiber-optic bundle.
3.2.4 Activation Command Sensing
Each of the Hand Controlled Accessory recep
tacles incorporate inputs that are used to sense
an activation command from the user. Each
monopolar hand controlled accessory receptacle
has an input for cut and an input for coag. The
bipolar receptacle incorporates a single activation
input. Each of these five inputs is isolated from
the other electrosurgical outputs and from other
low-level circuitry in the system. All are powered
by a multiple output isolated power supply. The
footswitch activation inputs on the back panel are
configured in a similar way and share one of the
isolated power supply outputs.
3.2.5 Automatic Return Monitor (A.R.M.™)
The patient return connector interfaces to single
and dual dispersive electrodes using a two-pin
connector. A.R.M.™ circuitry uses an actively
driven impedance measurement circuit, which
allows the System Controller to detect the type
of dispersive electrode connected and verify its
integrity.
3.2.6 Low Voltage Power Sources
The low voltage power supply is a medical-grade
universal input offline triple output switching
power supply. The power supply is active anytime
Mains power is connected to the unit with the
Mains power switch turned on.
3.2.7 System Controllers and Monitor
Three processors are used for system interface &
control, RF control, and system monitor func
tions. The ESU control section consists of dual
channel architecture with two independent chan
nels where one is used exclusively for RF output
control and the other is used for safety monitor
ing. All three of these processors are located on
-
-
-
-
-
-
3-5
the Control board, along with circuitry to isolate
them from RF noise.
• System Controller (System Microcontroller):
A dedicated microcontroller that handles the
entire user interface, Serial Interface, real
time clock functions, and enables/disables the
power factor control section of the HVPS
using the PFC_EN signal. The System
Controller can also disable the signal used to
drive the RF Amplifier and can terminate RF
drive at any time without interaction from
either the RF Controller or the Monitor. The
System Controller is comprised of an standard
architecture microprocessor together with
Field Programmable Gate Array (FPGA),
which provides interface logic to a variety of
signals, a 3.68MHz oscillator, independent
voltage regulators, a processor supervisory
reset circuit, and other interface logic.
• RF Controller: A DSP that is dedicated to
the output and control of RF power using
the DAMPCNTRL and RF_DRV outputs.
To reduce the effects on the microproces
sor circuits on the Controller board from
RF noise at the output, DAMPCNTRL and
RF_DRV are both differential mode signals
running between the RF Controller and the
RF Amplifier. The RF Controller is capable
of disabling RF output power and put
ting the system into a safe state without any
interaction from the Monitor or the System
Controller. The RF Controller indepen
dently monitors the RF output voltage and
current for control purposes through several
scaled inputs. It sets the output voltage of the
HVPS using the HV_SET signal dependent
on the output Mode and power selected. The
RF Controller controls the fan based upon
temperature measurements supplied from the
RF Amplifier through the System Controller.
The RF Controller is comprised of a DSP,
together with circuitry necessary for convert
ing the signals used for control purposes
between analog and digital form, independent
voltage regulators, and other interface logic.
• RF Monitor: A DSP that is dedicated to safety
monitoring activities. The Monitor is capable
of disabling RF output power and putting the
system into a safe state without any interac
tion from the RF Controller or the System
Controller. To ensure that the Monitor can
correctly perform its function, the Monitor is
-
-
-
-
resistively isolated from the other two proces
sors and has independent voltage regulation.
The RF Monitor independently monitors a
variety of inputs to detect safety problems
and has control of disable signals for both the
HVPS and RF Amplifier drive. The Monitor
is comprised of the same DSP as the RF
Controller, together with circuitry necessary
for converting the signals monitored between
analog and digital form, an FPGA to provide
interface logic, independent voltage regula
tors, isolation resistors and other interface
logic.
3.2.8 Low Voltage Power Monitoring
The low voltage power supply is monitored in
hardware and resets the processors if it is out of
range. The microprocessor supervisory device on
the Controller board monitors +5V and +3.3V
and will reset the system should the levels drop
approximately 0.3V. The Controller assembly
has the circuit that will reset the system should
the 3.3V supply exceed 3.6V. The High Voltage
power supply has a circuit that will inhibit HVDC
should the +5V supply exceed 5.7V.
3.2.9 Operator Control Panel
Keyboard: The main operator input device for
choosing operating modes and settings is the
membrane keyboard panel. Tactile-feedback
mechanical switches allow the operator to set
modes and adjust power settings.
Display Panel: Consists of 7-segment displays,
discrete dual colored LED’s, and light bars that
will display all controls and settings. LED display
elements are illuminated in a time division multi
plexed fashion; the illuminated LED display ele
ments are actually on less than half the time.
Bipolar Current Meter: The System 5000™
has a bargraph display that provides an indication of measured bipolar impedance. A special
-
tone works in conjunction with this bargraph to
indicate when the measured bipolar impedance
exceeds a particular limit.
3.2.10 Activation Tones
Tone is generated for all activation requests, fault
detection and changes made on the Control Panel.
The System Controller generates the tone signal
(ACT_TONE, AL_TONE, & BP_TONE), which
is amplified by a driver on the Backpanel PCB
Assembly. The activation tone and bipolar tone
-
-
-
-
3-6
are individually adjustable, but alarm tones are
not adjustable and are set to generate tone greater
than 65 dB.
Circuitry on the Backpanel PCB permits the
Monitor to verify the oscillation from voltage
measured across the speaker, which provides
confirmation that the speaker is indeed generat
ing audible tones during activation. RF output
is inhibited should the speaker drive current be
absent or too low.
3.2.11 Activation Relay Connector
There is an Accessory Relay Connector, which
provides a relay closure (SPST switch) that may
be used for activating external accessories such as
smoke evacuation units.
-
3.3 Optional System Configurations
An eight-position configuration dipswitch (S2),
located on the Controller PCB Assembly (A4)
allows a qualified service technician to change
some of the factory default settings. With the
exception of the DACview switch, the configura
tion dipswitch settings are only detected when
power is initialized, so any changes to the switch
positions must be made with the main power
off. Each switch is OFF in the Down position
and ON in the UP position. The system detects
changes in the DACview switch while power is
on, so it is treated differently. Relevant informa
tion for the configuration dipswitches appears in
Section 4.9.
-
-
3-7
This page has been intentionally left blank.
3-8
Maintenance
Section 4.0
TM
This section contains information useful in the
maintenance and repair of the System 5000™.
WARNING: High voltages are present at the
connections and within the System 5000™.
Maintenance personnel should take precautions
to protect themselves. Read the safety sum
mary in Section 1.1.4 before working on the
ESU.
-
4.1 General Maintenance Information
Although the System 5000™ has been designed
and manufactured to high industry standards, it is
recommended that periodic inspection and perfor
mance testing be performed to ensure continual
safe and effective operation.
Ease of maintenance was a primary consideration
in the design of the System 5000™. Maintenance
features of this unit include microprocessor aided
troubleshooting aids and push button calibration,
built in fault detection, circuit protection, and easy
access to circuitry while the unit is operational.
These features, coupled with the warranty, local
support, loaner equipment, factory support, toll
free phone service to the factory and available fac
tory training ensure the user of a minimal mainte
nance effort with extensive support available.
-
-
-
4.3.1 Top Cover Removal and Replacement
Top Removal:
1) Remove the two screws located on rear of
unit as shown.
2) Pull back and up to remove top.
Cover Screws
Cover
Screws
Cover
Screws
4.2 Maintenance Personnel
Only qualified biomedical engineers should perform service on the System 5000™. Refer all
servicing to a qualified biomedical engineer. If
necessary, your CONMED sales representative will
be happy to assist you in getting your equipment
serviced.
4.3 Assembly Breakdown/Parts Access
CAUTION: This device contains components that can be damaged by static electricity.
Proper handling by grounding of personnel
during servicing is mandatory.
Following are instructions for unit disassembly
and reassembly instructions.
Side Clips
4-1
Top Replacement:
1) Place top approximately ¾” from front bezel
on top of unit.
2) Press forward, aligning lip of front bezel with
groove in top and side clips with tabs on cast
ing.
3) Re-install screws.
3) Unlatch display ribbon cable, dispersive electrode connector, ReadiPlug™ cable connector
and two ground connectors.
4) In most situations, it is not necessary to
-
remove the four power switch connectors.
The bezel can be rotated off to the right side
for output board removal. To fully remove the
bezel, these connectors must be disconnected.
Slide top forward.
4.3.2 Bezel Removal and Replacement
Bezel Removal
:
1) Remove Top.
2) Remove two flat-head screws on side of bezel
and two pan-head screws on bottom of bezel.
AC Power Switch
White
Blue
Black
Brown
Bezel Replacement:
1) Connect power switch connectors as shown,
if required.
2) Connect dispersive electrode connector and
ReadiPlug
™ cable connector prior to sliding
bezel into place.
3) Slide bezel into unit. As shown in figure, the
output board insulator is positioned between
the sheet metal base and the Output Board.
4-2
Insulating Sheet
4) Reconnect cables and replace and tighten
screws.
Spring contacts are exposed and
can be deformed, causing erratic
operation. Handle with care.
4.3.3 Processor Board Removal and
Replacement
Processor Board Removal
:
1) Remove Top.
2) Loosen the two screws holding the board to
slots in the brackets.
3) Unlatch the ribbon cable going to the display.
4) Pull board up and out of unit.
Processor Board Replacement:
1) Align board into the two slots of the brackets
attached to the heatsinks. Align with con
nector on Output Board and press firmly to
engage it fully. Tighten the two screws.
4.3.4 Transformer Board Removal and
Replacement
3) Unlatch the ribbon cable and power cable on
the top of the board.
4) Pull board towards inside of unit so screws
line up with keyholes, then up and out of
unit. The power cable to RF Amp must be
unplugged to completely remove transformer
board.
Transformer Board Replacement
:
1) Reverse board removal operation.
NOTE: When servicing unit, board can be sup
ported in heatsink as shown. This will provide
access to the lower boards while the unit is func
tional.
-
-
Transformer Board Removal
:
1) Remove Top.
2) Loosen the two screws mounting the board
to the keyhole slots on standoffs and two
screws mounting the board to the heatsink.
Align slot
in board
with rib on
heatsink.
Route cable through slot.
4-3
4.3.5 Output Board Removal and
Replacement
4.3.6 RF Amp Board Removal and
Replacement
Output Board Removal
:
1) Remove Top, Bezel and Processor Board.
Note: It is not necessary to remove power switch
connections from the bezel.
2) Remove the seven cables along the rear side
of the board and the three screws shown.
Spring contacts are exposed and
can be deformed, causing erratic
operation. Handle with care.
Output Board Replacement:
1) Prior to replacing board, assure that the insu
lator sheet is positioned properly as shown.
RF Amp Board Removal
:
1) Remove Top, Bezel and Transformer Board.
2) Remove the four screws that attach the
heatsink to the sheet metal chassis. One for
the handle, one on the back panel, and two
below the heatsink. Remove the two screws
that attach the board to the sheet metal chassis as shown. Note: These screws are in
holes, not slots.
3) Unlatch three cables – two cables from the
RF Output Board and a cable from the RF
Power Supply Board.
4) Loosen the smaller hex standoff.
5) Slide RF Amp Board with heatsink off the
base as shown.
-
Note: The insulator
sheet provides an
important dielectric
barrier. For safe
operation, position
over center standoff as
shown.
2) Replace board on standoffs.
3) Route cables as they were prior to removal
and connect them back to their proper con
nectors.
4) Replace and tighten the three screws.
4-4
RF Amp Board Replacement:
1) Slide board with heatsink back into its previ
ous position on the sheet metal chassis. The
heatsink has two pins that align into holes on
the sheet metal chassis.
2) Tighten hex standoff onto board. It provides
an electrical ground to Transformer Board.
-
3) Reinstall and tighten the six screws and latch
the three cables.
4.3.7 Low Voltage Power Supply Module
Removal and Replacement
NOTE: This module is not user serviceable at
the component level. If faulty, the entire cir
cuit board must be replaced. Replacements are
available from CONMED Customer Service.
Do not discard the module cover, mount
ing plate and hardware; the replacement part
includes only the circuit board.
Low Voltage Power Supply Module Removal:
1) Remove Top Cover.
2) Loosen four screws located in slots and
unlatch the two cables.
3) Slide Low Voltage Power Supply Module
inward off the slots, then upward to remove.
Circuit Board
4) Remove the cover by removing four screws.
Then remove the four standoffs to separate
the Low Voltage Power Supply from its
mounting plate.
NOTE: When servicing unit, board can be held in
heatsink as shown. This will provide access to the
lower boards while the unit is functioning.
Low Voltage Power Supply Module Replacement:
1) Replace Low Voltage Power Supply on
mounting plate, fasten standoffs, replace cover
and tighten screws.
2) Place Low Voltage Power Supply Module
into unit on standoffs. Tighten the loose
screws and latch the two connectors.
4-5
4.3.8 High Voltage Power Supply Removal
and Replacement
High Voltage Power Supply Board Removal
:
1) Remove Top Cover. Position Transformer
Board in its servicing position to gain access
to screws. If desired, remove the Low
Voltage Power Supply.
2) Unlatch the Mains Power Cable, RF Amp
Cable, and Output Board Ribbon Cable.
3) Loosen the two screws holding the edge of
the board to the chassis.
4) Remove the screw to the handle, the screw to
the back panel, the two screws on the bottom
of the heatsink and slide out the heatsink with
the board attached.
Note: Observe the position of the insulating sheet
under the High Voltage Power Supply. If the
insulating sheet is removed, replace it as shown.
It is important to maintain its function as a
dielectric barrier and to protect the ribbon cable
from the leads of the High Voltage Power Supply
Board.
Insulating Sheet
4.3.9 Rear Panel with Board Removal and
Replacement
High Voltage Power Supply Board Replacement:
1) Slide board with heatsink back into its previ
ous position on the sheet metal chassis. The
heatsink has two pins that align into holes on
the sheet metal chassis.
2) Reinstall and tighten the six screws and latch
the three cables.
Rear Panel with Back Panel Board Removal
:
1) Remove Top Cover.
2) Unlatch the Ribbon Cable.
3) Remove the four screws on the bottom of the
Rear Panel, and the three screws shown on
the Rear Panel.
-
4) Slide the Rear Panel with Back Panel Board
back toward the handle and then down to
remove.
4-6
Rear Panel with Back Panel Board Replacement:
1) Slide Rear Panel with Back Panel Board back
into place on the unit.
2) Reinstall and tighten the seven screws and
latch the ribbon cable.
4.3.10 Back Panel Board Removal and
Replacement
Back Panel Board Removal
1) Remove Top Cover and Rear Panel.
2) Unlatch the Activation Relay and Fan connectors.
3) Remove the four screws for the foot switch
connectors, the two nuts on the volume
potentiometers and the two screws to the rear
panel sheet metal standoffs. This will free the
Back Panel Board from the Rear Panel.
Back Panel Board Replacement:
1) Install Back Panel Board to Rear Panel
2) Reinstall and tighten the six screws, two nuts.
Latch the RSA and fan cables.
:
7) Remove the Display Controller Board by separating it from the display board and pulling
it off the studs. There are two 40-pin connec
tors between these boards that may require
separation by prying with a blunt object.
8) Remove the five hex standoffs and pull the
Display board off the studs.
Display Boards Replacement
Caution: When reinstalling Boards and display
shield, take care to route fiber optic cable as
shown to avoid crimping it between the bezel
and display shield.
:
-
4.3.11 Display Boards Removal and
Replacement
Display Boards Removal
1) Remove the Top Cover and Bezel.
2) Remove 8 nuts and two ground cables on the
back of the display shield.
3) Disconnect the ribbon cable and remove the
sheet metal display shield.
4) Slide the round spacers off their studs.
5) Disconnect the flex circuit connector.
Caution: The flex circuit has a short service
loop and is fragile; handle with care.
6) Pull to disconnect the fiber optic cable from
LED. Caution: The fiber optic cable is frag
ile; handle with care.
:
1) Reinstall boards in the reverse order described
above. Latch the flex circuit connector, press
fiber optic cable onto the LED and press the
Display Controller Board firmly into place on
the two 40 pin connectors.
-
4-7
4.3.12 Power Transistor Replacement
Caution: This device contains components that
can be damaged by static electricity. Proper
handling by grounding of personnel during
servicing is mandatory.
All RF Power Supply and RF Amp components
mounted to the heatsink may be replaced.
only components supplied by CONMED.
Follow these instructions for replacement:
1) No thermal compound is necessary, but the
mating surfaces of the transistor, insulator
pad and surface of casting should be clean.
Always replace the insulator pad associated
with the transistor. Always fasten or clamp
the part to the heat sink surface prior to sol
dering it to the board. This will assure good
thermal contact is maintained.
2) In order to maintain alignment with the heat
sink surface, the leads of these parts have been
bent to the proper shape. They should be
purchased from CONMED with bent leads.
3) When installing the RF Amp transistors or
diodes, be sure to orient the Bellville washer
as shown with the convex surface next to the
head of the screw. Tighten screws to 5-7 inch
pounds.
Use
-
4) When installing the RF Power Supply transistors or diodes, replace components as shown
and ensure the insulating tube is installed over
the clip. Locate the part on the clip so that
the bend of the clip is approximately centered
on the body of the part as shown. Tighten
screw to 8-10 in-lbs. When tightening screw,
hold the clip to prevent it from rotating.
Clamp the part to the heatsink surface prior
to soldering to the board.
4.4 Cleaning
The interior of the unit may be vacuumed or
blown out as required. The exterior of the unit
may be cleaned by wiping it with a cloth that
has been dampened (not dripping) with a mild
detergent such as Windex® or Formula 409®.
Windex® is a registered trademark of the S.C.
Johnson Company. Formula 409® is a registered
trademark of the Clorox Company.
4.5 Periodic Inspection
The System 5000™ should be visually inspected
at least every six months. This inspection should
include checks for the following:
1) Damage to the power cord and plug.
2) The proper mating and absence of damage to
the accessory connectors.
3) Any obvious external or internal damage to
the unit.
4) An accumulation of lint or debris within the
unit or heatsink.
5) Control Panel cuts, punctures, or dents.
4-8
4.6 Periodic Performance Testing
4.6.3 Output Power
The System 5000™ should be tested for correct
performance at least once every year. Every unit
is supplied with a serialized Production Test Data
Sheet that tabulates the results of the factory tests
that were performed on the unit. This data is
supplied so that it may be used as a reference for
subsequent tests. Recommended periodic perfor
-
mance tests are listed in the following sections.
4.6.1 Chassis Ground Integrity
Connect a standard ohmmeter between the
earth ground prong on the power plug and the
Equipotential Ground Connection. Compensate
for lead resistance. Confirm less than 0.2 ohms
resistance is measured.
4.6.2 Displays, Alarms and Commands
Perform the Preliminary Functional Test procedure
described in section 2.3.1 of this manual to verify
proper operation of displays, alarms and com
-
mands.
1) Equipment Requirements:
a) Monopolar Footswitch
b) Bipolar Footswitch
c) Commercial ESU Tester (e.g. Dynatech
454A or equivalent) with 50 and 300
ohm loads for bipolar modes and a 500
ohm load for monopolar modes.
Note: Micro Bipolar is particularly sensitive to the
load resistance. A 50 ohm load should be used for
checking power to obtain the best results.
2) Use test leads to connect the ESU tester to
the unit’s return electrode output and the
footswitch controlled active output. Set the
Load resistance per mode as indicated in
Tables 4.1 and 4.2.
3) Perform the monopolar power tests indicated
in Tables 4.1 and 4.2. The acceptance range
is given in both Watts and Amps to accom
modate available test equipment. It is not nec
essary to test for both power and current.
Table 4.1 Monopolar Cut Mode RF Output Power Accuracy
ModeLoad (ohms) Power SettingWatts (min) Watts (max)Amps (min)Amps (max)
Pure50010713.00.1180.161
Standard
Blend 150010713.00.1180.161
Blend 250010713.00.1180.161
Blend 350010713.00.1180.161
500201723.00.1840.214
5005045550.3000.332
500100901100.4240.469
5002001802200.6000.663
5003002703300.7350.812
500201723.00.1840.214
5005045550.3000.332
500100901100.4240.469
5002001802200.6000.663
500201723.00.1840.214
5005045550.3000.332
500100901100.4240.469
5002001802200.6000.663
500201723.00.1840.214
5005045550.3000.332
500100901100.4240.469
5002001802200.6000.663
-
-
4-9
Table 4.2 Monopolar Coag Mode RF Output Power Accuracy
ModeLoad (ohms) Power SettingWatts (min) Watts (max)Amps (min)Amps (max)
Spray50010713.00.1180.161
500201723.00.1840.214
5005045550.3000.332
5008072880.3790.420
Standard
50010713.00.1180.161
500201723.00.1840.214
5005045550.3000.332
500100901100.4240.469
5001201081320.4650.514
ModeLoad (ohms) Power SettingWatts (min)Watts (max)Amps (min)Amps (max)
Pinpoint50010713.00.1180.161
500201723.00.1840.214
5005045550.3000.332
500100901100.4240.469
5001201081320.4650.514
Standard
pulse
Spray pulse
50010713.00.1180.161
5002017230.1840.214
5006054660.3290.363
50010713.00.1180.161
5002017230.1840.214
5004036440.2680.297
4) Disconnect the ESU tester from the unit.
5) Use test leads to connect the ESU tester to
the Bipolar Accessory outputs.
6) Perform the bipolar power tests indicated in
Table 4.3. This table only provides the mini
mum number of points to be tested.
Table 4.3 Bipolar Mode RF Output Power Accuracy
ModeLoad (ohms) Power SettingWatts (min) Watts (max)Amps (min)Amps (max)
Macro
Bipolar
Micro
Bipolar
4.6.4 RF Leakage Measurement
RF Leakage can present a hazard in the operating
room because electrosurgical currents can flow
to the patient and operating room staff through
unintended paths, which can cause injury. RF
30010713.00.1530.208
300201723.00.2380.277
3005045550.3870.428
3009081990.5200.574
5010713.00.3740.510
50252228.00.6630.748
505045550.9491.049
leakage occurs because the total energy in the
output voltage waveform is provided with a con
ductive path through stray parasitic capacitance
distributed within the generator and along the
length of the leads. Table 4.4 presents the allowed
RF leakage currents to ground.
-
-
4-10
Loading...
+ 51 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.