Title: Hyfrecators Date: 10-4-2019
By: Birtcher and Conmed
File = hyfrecator-guide.doc
Rev 1.0
1. General
Service Guideline for models 733, Hyfrecator Plus, and Hyfrecator
2000.
2. Reference Documents
Operator Manual, Hyfrecator 2000, P/N 7-900-OM-ENG (1998,
Available at frankshospitalworkshop.com)
Service Manual Hyfrecator Plus (available at medwrench.com)
Electrical Apparatus Subcourse MD0363, U.S. Army
3. Tools / Fixtures / Accessories
Electrosurgical Analyzer: ex: Dempsey 443 or better.
Optional Footswitch.
Hand Control Pencil/Cable
Safety Tester
DISCLAIMER: THIS PROCEDURE PROVIDED "AS IS" AND
WITH POSSIBLE FAULTS. USER MUST VERIFY BEFORE USE.
NEITHER PROVIDER NOR WEBSITE ASSUMES ANY
RESPONSIBILITY FOR ITS USE.
4. Performance Verifications
4.1. Inspection
Check unit for any damage; i.e. connectors, knobs, the power
cord and the footswitch with its cable and strain reliefs.
4.2. Preliminary
With no accessories plugged in, turn unit on.
With no handpiece connected, verify that the up and down arrow
power-set buttons (Model 733 has knob and no display) change
the display count. For the Hyfrecator Plus and 2000, the
maximum display is 30.
Connect the handpiece and verify that its up/down buttons
function as above (skip if 733).
With the blade of the handpiece free, verify that the handpiece
power activation button activates the unit, noting the alert tone.
AC Line Leakage acceptable if less than 50 microamps.
4.3. RF Output Check
Connect analyzer (wattmeter / 500 ohm load) to bipolar
terminals. According to the user manual, output power should
track display by 1 watt +/- 10% of reading.
Model 733
Hyfrecator Plus
Hyfrecator 2000
1
Alternate Method: when the movable lead is in the “HI” socket, and the analyzer connected
between chassis and the pencil tip, power should range from 1 – 30 watts.
Optional Check: Connect a 25watt light bulb across bipolar terminals. Light bulb should light
at ¾ brightness.
Note: For low power, range is 0 to 15 with increments of 0.1 up to 5 watts (note the decimal point). Beyond
5, increment is one.
4.4. Neutral/Patient Plate/ Return Electrode
Note: The Hyfrecator 2000 is the only version of this device to include a jack for a patient plate
connection. All units internally connect the return lead of the RF output transformer to chassis ground.
“Return electrode” is
chassis itself
4.5. Calibration
See Service manual.
Output vs Dial, Model 2000 Peak Voltages, Model 2000
2