ConMed Hall MicroPower High Speed Drill, Hall MicroPower 6020-026, Hall MicroPower 6021-026 Instruction Manual

Hall
®
MicroPower™ High Speed Drill
Instruction Manual
(6020-026 / 6021-026)
Proprietary Information
This manual contains information deemed proprietary to Linvatec Corporation. The information contained herein, including all of the designs and related materials, is the sole property of ConMed Linvatec and/or its licensors. ConMed Linvatec and/or its licensors reserve all patent, copyright and other proprietary rights to this document, including all design, manufacturing methodology and reproduction.
This document, and any related materials, is confidential and is protected by copyright laws and shall not be duplicated, transmitted, transcribed, stored in a retrieval system, or translated into any human or computer language in any form or by any means, electronic, mechanical, magnetic, manual or otherwise, or disclosed to third parties, in whole or in part, without the prior express written consent of ConMed Linvatec.
ConMed Linvatec r eserv es the right to r e vise this publication and to make changes fr om time to time in the contents hereof without obligation to notify any person of such revision or changes, unless otherwise required by law.
Linvatec, Hall, Advantage, PowerPro, E9000, Surgairtome Two, Micro 100, Ultrapower, MicroPower, and SmartGuard are trademarks or registered trademarks of Linvatec Corporation.
©
ConMed Linvatec Corporation 2006. All Rights Reserved. Printed in USA
Record the Model and Serial Numbers of the handpiece(s), and date received. Retain for future reference.
Handpiece Model No. Serial No Date Handpiece Model No. Serial No Date Handpiece Model No. Serial No Date Handpiece Model No. Serial No Date Handpiece Model No. Serial No Date Handpiece Model No. Serial No Date Handpiece Model No. Serial No Date Handpiece Model No. Serial No Date
Table of Contents
1.0 INTRODUCTION
1.1 Intended Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2 Warnings and Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2.2 Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Symbol Definitions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.4 MicroPower High Speed Drill . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.5 Irrigation Tubing Set (E9418) for use with the MicroPower High Speed Drill . . . . . .5
2.0 HANDPIECE INSTALLATION and OPERATION
2.1 Handpiece Cord Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2.2 Hall Bur Guard Assembly Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2.1 Assembly Instructions for the Medium Bur Guard (1375-012), Long Bur Guard (1375-011), Cloward Extra-Long Bur Guard (1375-023),
Tissue Retractor Guard (1387-022). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.2.2 Assembly Instructions for the Laminectomy Bur Guard (1375-020) . . . . . . 10
2.2.3 Assembly Instructions for the 20° (1375-032) and 20° Extra Long
(1375-033) Angle Attachments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
2.2.4 Assembly Instructions for the 70° Contra-Angle with Surgical Head (1375-034), 70° Contra-Angle with Dental Head (1375-035),
90° Angle (1375-036) Attachments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
2.3 Handpiece Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
2.3.1 Tubing Set Attachment (E9418) for the MicroPower High Speed Drill. . . . 13
2.3.2 Preoperative Functional Test for the MicroPower High Speed Drill . . . . . .14
2.3.3 High Speed Drill Attachments and Accessories . . . . . . . . . . . . . . . . . . . . . .15
2.3.4 Removing the Detachable Handpiece Activation Lever . . . . . . . . . . . . . . . .16
2.4 Footswitch Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.4.1 Operation of the 2-Pedal Footswitch (5020-053) . . . . . . . . . . . . . . . . . . . . .17
2.4.2 Operation of the 3-Pedal Footswitch (C9863). . . . . . . . . . . . . . . . . . . . . . . .17
2.5 Handpiece Cord Button Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Page
i
Table of Contents
3.0 MAINTENANCE
3.1 Cleaning and Sterilizing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1.1 Cleaning Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1.2 Handpiece Cleaning Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1.3 Attachment Lubricating Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.1.4 Sterilization Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.1.4.1 Sterilization Warnings, Precautions and Notes. . . . . . . . . . . . . . . 23
3.2 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.3 Bur Guard Testing Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.3.1 Bur and Neuro Guard Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.3.2 Medium Bur Guard (1375-012) Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.0 TECHNICAL SPECIFICATIONS and ACCESSORIES
4.1 System Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.2 Handpiece Cord (MC5057) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.3 Handpieces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.4 Handpieces, Attachments and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Page
ii
1.0
I
Drill and accessories. The safe and effecti v e use of this equipment requires the understanding of and compliance with all warnings, cautionary notices, and instructions marked on the product, and included in this manual.
INTRODUCTION
t is recommended that personnel study this manual before attempting to operate, clean or sterilize the MicroPower™ High Speed
1.1 Intended Use
he MicroPower High Speed Drill, which operates in conjunction with assorted bur
T
guards, burs and drill bits, is designed to perform bone cutting, drilling, and soft tissue resection.
This handpiece can be operated using the
PowerPro
(E9000, E9000G, E9000J) or Advantage (D3000, D3000I, D3000J) Controllers.
The MicroPower High Speed Drill is intended for small bone procedures such as: Arthroscopic, Neurosurgical, Orthopedic (hand, wrist, foot and ankle) and Plastic / Reconstructive.
®
(PRO2000/PRO2000I), the E9000
®
®
1.2 Warnings and Cautions
This equipment is designed for use by medical professionals completely familiar with the required techniques and instructions for use of the equipment. Read and follow all warning
and cautionary notices and instructions marked on the product and included in this manual.
Service intervals, as listed in “Table 2:
Maintenance Schedule” on page 20 , are
required to keep the MicroPower High Speed Drill and attachments at their optimum operating performance.
1.2.1 Warnings
1. Eye protection is always necessary when operating equipment. Eye injury may result.
2. Continually check all handpieces and attachments for overheating. If overheating is sensed, immediately discontinue use and return equipment for service. Overheating of the bur, bit or blade may cause damage to the bur, bit or blade and may cause thermal necrosis.
3. Do not attach, insert or remove attachments or accessories while the handpiece is oper­ating. Serious injury may occur. Place the
handpiece safety mechanism to the safe position prior to installation or removal of items .
4. Do not immerse handpieces in fluids.
1
2
5. Prior to each use, perform the following:
Inspect all equipment for proper operation.
Ensure all attachments and accessories are correctly and completely attached to the handpiece.
Perform the required Performance Tests for each associated handpiece prior to each use.
6. Do not use burs for plunge cutting. Injury or damage may occur.
1.2.2 Cautions
1. Handpieces are factory sealed. Do not disassemble or lubricate, as this may void the warranty. There are no user­serviceable parts inside.
2. Use only associated ConMed Linvatec attachments and accessories (i.e. drill bits) as defined within the descriptions of each attachment.
3. Handle all equipment carefully. If any equipment is dropped or damaged in any way, return it immediately for service.
4. Never operate the MicroPower High Speed Drill without a bur and appropriate bur guard in place and the collet locked . Damage will occur.
7. Do not excessively bend or kink the hand­piece cord. Always inspect cords for signs of excessive wear or damage. If wear or
damage is found, discontinue use and replace immediately .
8. Always inspect for bent, dull or damaged burs, blades or drill bits before each use. Do not attempt to straighten or sharpen. Do not use if damaged. After use, dispose of properly.
9. Always use a bur of the proper length for the specific guard.
10. Bur and neuro guards should be checked frequently. Overheating can occur if bur guard bearings are worn or not kept clean. Follow the recommended maintenance schedule on page 20.
11. Perform the following steps to verify guards are in good operating condition.
NOTE: Bur and neuro guards are to be returned to the factory or a ConMed Linvatec authorized service facility for rou­tine maintenance every six (6) months.
(a) Remove the guard from the handpiece
and insert a bur into the nose of the guard.
(b) While holding the bur, spin the guard.
The guard should spin freely around the bur shaft without resistance.
5. Do not stall handpieces, damage can occur.
6. Do not handle the handpieces by the cord. Do not pull on the cord to remove it from the handpiece or controller. Use the connec­tor portion of the cord for disconnection.
3
(c) Attach the guard and bur to the
handpiece.
1.3 Symbol Definitions
(d) Operate the handpiece for
approximately 30 seconds, then stop the handpiece. Once the bur has stopped moving, carefully feel the end of the guard where the guard encases the shaft of the bur for overheating. If overheating is noticed, return guard for service.
12. After each use, thoroughly clean the con­troller, handpieces, attachments and acces­sories (See “3.1 Cleaning and Sterilizing”
on page 21 ).
13. Do not use in the presence of flammable anesthetics, gases, disinfecting agents, cleaning solutions, or any material suscepti­ble to ignition due to electrical sparking.
Attention, consult accompany­ing documents
Type B equipment
Single Use Only
Eye Protection Required
No user service recommended. Refer servicing to qualified
ConMed
sonnel. Indicates product component
should not be sterilized.
Indicates product component should not be immersed in any type of fluid.
Indicates handpiece should not be immersed in any fluid.
Linvatec service per-
Rx ONLY
Indicates product should not be oiled or lubricated.
Not to be used for plunge cut­ting
Caution: Federal Law restricts this device to sale by or on the order of a physician.
WEEE (Waste Electronics and Electrical Equipment) Symbol. Regarding Euro­pean Union end-of-life of product.
1.4 MicroPower High Speed Drill
Bur Lock Collar — Rotate to lock or
unlock a bur.
Activation Lever — Depress to operate
the handpiece. The activation lever can be removed for leverless operation on the 6020-026 handpiece (Reference “2.3.4
Removing the Detachable Handpiece Activation Lever” on page 16 ).
Safe/Run Slide — Place in the “SAFE”
position prior to inserting or removing attachments and/or accessories (blades, bits, burs) and when the handpiece is not in use. Place in the “RUN” position to activ ate the handpiece.
High Speed Drill
(6020-026 / 6021-026)
Handpiece Cord Connector — The hand-
piece cord connects here to provide power to the handpiece. Uses the MC5057 Hand­piece Cord.
NOTE: Do not allow saline to enter into the handpiece cord connector and cord. Damage may occur to the handpiece.
4
E9418 Irrigation Tubing Set
1.5 Irrigation Tubing Set (E9418) for
use with the MicroPower High Speed Drill
The E9418 Tubing Set is used to provide irrigation for blades, bits and burs while using the MicroPower High Speed Drill.
Spike Line — Connects to the fluid supply
for fluid distribution to the surgical site.
Irrigator Tip — Accepts various irrigation
tips (See “4.4 Handpieces, Attachments
and Accessories” on page 30 ) for the
MicroPower High Speed Drill for fluid distribution to the surgical site.
Cassette - Attaches to the peristaltic pump
of the controller to aid in fluid delivery.
Roller Clamp — Used to adjust the fluid
flow to the surgical site.
Handpiece Cord Clips — Used to attach
the tubing to the handpiece cord.
5
2.0
HANDPIECE INSTALLATION and
OPERATION
2.1 Handpiece Cord Installation
1. To attach the handpiece cord: (a) Ensure pins on cord and handpiece are
completely dry prior to connecting.
(b) Insert the handpiece cord into the cord
receptacle of the handpiece. Push together until fully seated.
NOTE: Do not force the cord into the hand­piece cord connector. This may bend the pins and damage the handpiece.
3. To remove the handpiece cord: (a) Press the latch and remove the cord
from the handpiece.
2. Connect the handpiece cord to the appropri­ate console.
6
2.2 Hall Bur Guard Assembly Instructions
WARNINGS:
1. Do not operate the drill without the appropriate bur guard. Always use a bur of the proper length. The tip of the bur guard should cover the safe line on the bur, if applicable. Without the stabilization that the proper guard provides, the bur can break and be propelled with great force.
2. To reduce the risk of injury, prior to surgery, spin the bur guard on a bur. If the bur guard spins freely, the bearing is still good. Otherwise, the bur guard must be sent for repair immediately. DO NOT USE.
3. When not in use and prior to connecting or removing attachments and accessories, always place the handpiece in its safe, or off position, if applicable.
Table 1: Bur Guard and Attachment Guide
REF Description Instructions
1375-012
1375-011 Long Bur Guard - Use only 5092 series long burs See page 8 1375-023
1375-020 Laminectomy Bur Guard - Use only 5092-103 Long Carbide Bur See page 10 1387-022
1375-032 20° Angle Attachment - Use only 5092 series long burs See page 11 1375-033 20° XL Angle Attachment - Use only 5093 series extra-long burs See page 11
1375-034
1375-035
Medium Bur Guard - Use only 5091 series medium burs except
5091-080, 5091-081, 5091-082 and 5091-083
Cloward Extra-Long Bur Guard - Use only 5093 series extra-long
burs
Tissue Retractor Guard - Use only with long burs (5092 series)
which have a diameter of 4.0mm or less. Recommended bur: 5092-136
70° Contra-Angle Attachment with Dental Head
Use friction grip burs only - size 1/16” or 1.58mm in diameter
70° Contra-Angle Attachment with Surgical Head
Use only 5089 extra-short burs and 5090 series short burs
See page 8
See page 8
See page 8
See page 12
See page 12
1375-036
90° Angle Attachment
Use only 5089 extra-short burs and 5090 series short burs
See page 12
7
2.2.1 Assembly Instructions for the:
• Medium Bur Guard (1375-012)
• Long Bur Guard (1375-011)
• Cloward Extra-Long Bur Guard (1375-023)
• Tissue Retractor Guard (1387-022)
8
Photo below shows the Tissue Retractor Guard.
NOTE: When using the Tissue Retractor Guard, use only 5092 series long burs which have a diameter of 4.0 mm or less. The recommended bur is the 5092-136 Long Oval Cutting Bur.
1. Place the handpiece in the “SAFE” position.
2. To install a bur guard: (a) Slide the appropriate bur guard over the
end of the drill. Ensure it is seated completely. Do not operate the drill without the bur guard completely seated to the nose of the drill.
3. Prior to installing a bur, but only while using the 1375-012 Medium Bur Guard, it
is recommended that a SmartGuard
®
Protector Sleeve (1375-112) be attached. Reference the associated insert that comes packaged with the SmartGuard for addi­tional information.
NOTES:
SmartGuard Protector Sleeves are designed to be attached to the Medium Bur Guard during surgical applications. They are intended to be used as a heat­sensing and insulating device that changes color from PURPLE (SAFE to use bur guard) to PINK (REPLACE bur guard) to indicate over-heating of the attached bur guard.
Overheating might occur if bearings are worn or are not kept clean. If overheat­ing occurs while a SmartGuard Protector Sleeve is attached to the Medium Bur Guard, the SmartGuard Protector Sleeve will change color from PURPLE to PINK. Immediately discontinue use of the bur guard to prevent possible injury to the patient and/or operator. Replace the bur guard and return original bur guard for service.