COMPAT 199255 Operating Instructions

Operating Instructions
199255 COMPAT DualFlo
Enteral Delivery Pump
®
1
Table of Contents
For Model Number
199255 COMPAT DualFlo
Enteral Delivery Pump U.S. Patent No. 5,634,907
Section Page
®
Controls 4
Preparation of COMPAT DualFlo
®
Delivery Set 8
Operating Instructions 9
Quick Reference:
Flow Rate, Dose Limit, Formula Delivered 12
Hydration with Automatic Flush 13
Formula Strength, Programmable Flush, Water Delivered 14
Alarms/Indicators 15
Troubleshooting 17
Specifications 18
Self Test 21
Factory Service 21
Safety and Warnings 21
NOT FOR PARENTERAL USE
INTENDED EXCLUSIVELY FOR ENTERAL TUBE FEEDING AND PATIENT HYDRATION.
For More Information
Contact your local Nestlé HealthCare Nutrition Sales Representative or call Customer Service from anywhere in the U.S. and Canada at 1-877-338-4238.
Limited Warranty 22
Cleaning 22
Pump is not intended for neonatal use. Caution: Federal law restricts this device to sale by or on the order of a physician.
Introduction
2
The COMPAT DualFlo administration of all enteral feeding solutions. Key features of this device include:
Patient hydration with automatic flush
Patient hydration needs are facilitated by administering up to 500 mL of water over a selected time interval (in hours). Time interval is adjustable in 1 hour increments from 2 to 12 hours. In addi­tion, the final 25 mL of water is automati­cally flushed through the line just before the conclusion of the time interval.
Formula dilution
Provide formula dilution when necessary by selecting Formula Strength. Formula can be delivered at full strength (FS), or diluted to (all water delivery, no formula).
3
/4,2/3,1/2,1/3,1/4 strength, or H20
Programmable Flush after Dose is Delivered
Deliver up to 500 mL of water in a programmable water delivery to hydrate the patient or flush the line after the for­mula dose limit is delivered. Water is delivered at a fixed rate of 375 mL/hr.
®
Enteral Delivery Pump is an easy to use, versatile device for the
Delivery Set Security Doors
Reduces the likelihood of accidental dislodgment of sets and enables immediate detection of a free flow condition.
Automatic priming
Automatic priming of formula and water through the administration set allows nurse time to perform other bedside duties and may help reduce wasted sets due to improper manual priming.
Versatility
• Formula rate adjustable in 1 mL/hr increments from 1–295.
• Dose Limit adjustable in 1 mL/hr increments from 1–2000.
• Formula Strength options include FS (Full Strength), 3/4,2/3,1/2,1/3,1/4 strength, or H20 (water only).
• Water volume adjustable in 1 mL/hr increments from 0–500.
• Time Interval adjustable in 1 hour increments from 2 to 12 hours, with programmable flush available as dose is delivered.
• Ability to select basic full strength formula delivery when water delivery options are not desired.
3
Introduction (cont.)
Lifetime memory retention
Infusion Rate, Dose Limit, Formula Strength, Water Volume Set, Time Interval, Formula and Water Volume Delivered, and Accumulated Volume are automatically recalled when the pump is turned back on.
Audio and visual alarms when pump is turned on:
• Feed Error for detecting empty container or occluded line
1
/2 minute Hold Alarm if user
•2 forgets to start the pump
• Dose Delivered
• Low Battery
• No Set
Designed for ease of use
• Large LED display window, alternates displaying delivery rate and volume delivered when pump is running.
• Dose Limit for Intermittent Feeding with choice of 3 dose complete alarm modes.
• Ability to quickly view formula and water settings without interruption of feeding.
• Adapter Cap Holder.
• Smooth surface of pump makes cleaning easy.
• Lightweight and portable, the pump is 7pounds and can clamp to any IV pole.
• Adjustable Alarm Volume Control knob.
• 8-hour battery life.
• Self-diagnostic mode allows for com­plete circuitry check when pump is turned on.
Controls
Keypad Controls
1. ON, turns power on, memory recalls previous settings.
2. OFF, turns power off, battery
charges when plugged in to an electrical outlet.
3. SELECT, selects a delivery function
to display and/or change.
4. CLEAR, clears displayed value of
selected function to 0. (If RATE is selected, pressing and holding CLEAR for 4 seconds will reset all parameters to default settings.)
5. increases flow rate, dose limit,
formula strength, water volume set, or time interval when pressed. Change is more rapid when pressed and held.
6. decreases flow rate, dose limit,
formula strength, water volume set, or time interval when pressed. Change is more rapid when pressed and held.
7. PRIME, automatically primes both
formula and water through the administration set.
8. START/HOLD, starts and/or holds
pump operation. HOLD is used to temporarily stop feeding, or silence an alarm condition. Parameters can only be changed when pump is in HOLD mode. An audio and visual alarm is activated if the pump has been in HOLD mode for 2 1/2 min­utes. Press START/HOLD to resume normal operation.
9
10 11
1
3
2
4
12
7
5
8
6
Function Displays
9. RUNNING, STOPPED, indicates status of the pump when power is on.
10. AC, indicates the unit is plugged into an electrical outlet.
11. LOW BAT, indicates when the battery charge is below performance level.
12. Display panel, indicates selected parameter.
4
5
Controls (cont.)
Front of the pump
13. Flow Sensors and Drip Chamber Guide
14. Tubing Guide
15. Drip Chamber Brackets
16. Rotary Peristaltic Pump Mechanism
17. Door Latch
18. Delivery Set Security Door
Back of the Pump
19. Set Adapter Cap Holder
20. Power Cord
21. Alarm Volume Control
22. Pole Clamp
23. Directions
17
13
14
15
16
18
22
21
19
20
23
Controls (cont.)
6
29
30
31
32
25
27
24
26
28
7
Controls (cont.)
Formula Side Displays
24. RATE, displays the formula infusion rate. Select from 1–295 mL/hr in 1 mL increments.
25. VOL. DELIVERED, displays full strength or diluted formula volume delivered.
26. DOSE LIMIT, displays dose limit. Select from 1–2000 mL in 1 mL increments. Dose is completed when Volume Delivered equals the Dose Limit set. Refer to Dose Delivered under Alarms/Indicators. If nOnE is displayed, this feature is deactivated.
27. ACCUM. VOLUME, displays full strength or diluted formula volume delivered over multiple feedings.
28. STRENGTH, (Formula Dilution) displays formula strength to be delivered. Select from FS (Full Strength), or H
3
/4,2/3,1/2,1/3,1/4 strength,
0 (water only).
2
Water Side Displays
29. VOLUME SET, displays the volume of water to be delivered over the selected Time Interval or at the end of the dose limit. Select from 0–500 mL in 1 mL increments.
30. TIME INTERVAL, displays the time interval (2 to 12 hours in 1 hour increments) over which the selected water volume is delivered. When doSE is displayed, indicates Programmable Flush is activated. When nOnE is displayed, all water side settings are deactivated.
31. VOL. DELIVERED, displays amount of selected water volume that has been delivered. At the con­clusion of the Time Interval selected, the Volume Delivered turns to 0 and a new accumulation begins.
32. ACCUM. VOLUME, displays water volume delivered over multiple Time Intervals.
Preparation of COMPAT DualFlo®Delivery Set
8
DIAGRAM 1
Open pump doors by pulling door latch, insert pump set drip chamber into drip chamber bracket.
DIAGRAM 3
Insert pump adapter into bridge and place tubing into tubing guide. Close and latch delivery set security door.
DIAGRAM 2
Assemble silicone tubing around rotor.
DIAGRAM 4
Completely assembled pump set. Check or change Pump Settings to prepare enteral delivery.
9
Detailed Operating Instructions
Attach pump to IV pole using built-in pole clamp. Plug power cord into a 120 VAC, 60 Hz. three wire grounded electrical outlet if available.
Prepare COMPAT DualFlo
®
Delivery Set
1. Close both roller clamps on delivery set prior to setup. Prepare and hang DualFlo delivery set from an IV pole.
2. Open delivery set security doors by pulling the latch on each door.
3. Insert water drip chamber into water side drip chamber bracket, and formula drip chamber into formula side drip chamber bracket.
4. Carefully stretch water set silicone tubing around rotor. Insert pump adapter into bridge. Place tubing into tubing guide, close and latch delivery set security door.
5. Repeat step 4 with formula side delivery set.
Check pump settings
1. Press ON- alarm sounds briefly and displays .8.8.8.8. The pump has memory and displays previous settings, and returns to flashing Flow Rate on the display panel.
Note: Changes can only be made to parame­ters at the START of a new feeding protocol to avoid improper water delivery. To facili­tate this safety feature, parameters can only be changed when amount displayed in Vol. Delivered is 0.
If desired, clear Vol. Delivered or reset all parameters to default settings by pressing and holding CLEAR for 4 seconds when Rate is displayed.
2. Check Formula Volume Delivered. Press SELECT until Vol.
Delivered is displayed. If desired, reset to 0 by pressing CLEAR. This also resets Water Volume Delivered to 0. (Accumulated Volume is not affected.)
3. Check Dose Limit. Press SELECT until Dose Limit is displayed. To change, press (increase) or (decrease). For continuous feeding where no dose limit is desired, press until nOnE is dis­played.
4. Check Formula Accumulated Volume. Press SELECT until Accum. Volume is displayed. If desired, reset to 0 by pressing CLEAR.
Detailed Operating Instructions (cont.)
10
Formula Strength (dilution)
5. To dilute formula, press SELECT until strength is displayed. Press or to set desired formula strength. Options available are FS (Full Strength-no formu­la dilution),
or H
In steps 6–8, you have the option to select Hydration with Automatic Flush or Programmable Flush after dose is delivered.
3
/4,2/3,1/2,1/3,1/4 strength,
0 (water only).
2
Hydration with automatic 25 mL flush
Administers water with formula over a selected time interval during continu­ous feeding, concluding with an automat­ic 25 mL water flush.
6. To set water volume, press SELECT until Volume Set is displayed. Press or to set the desired volume of water to be delivered. Adjustable in 1 mL increments from 0–500 mL.
7. To select time interval, in hours, press SELECT until Time Interval is displayed. Change by pressing or . Adjustable in 1 hour increments from 2 to 12 hours. MAXIMUM WATER DELIV­ERY IS 75 ML PER HOUR UP TO A MAXI­MUM OF 500 ML. (See page 13 for maxi­mum water volume setting at specific time intervals.) The final 25 mL of
water is automatically flushed through the line before the con­clusion of the selected time inter­val.
Programmable Flush after dose is delivered
Selected water volume delivered after the formula dose limit is delivered. (For this feature, water is delivered at a fixed rate of 375 mL/hr.)
8. Follow steps 6 and 7. At Volume Set, enter the desired volume of water to be deliv­ered. When Time Interval is selected, press
until doSE is displayed. The
amount in Volume Set is delivered after formula dose limit is reached.
To deactivate Hydration and Programmable Flush, at Time Interval,
press until nOne is displayed.
9. Check Water Volume Delivered. Press SELECT until Vol. Delivered is dis­played. This will be reset to 0 when Formula Vol. Delivered is cleared (see step
2).
10 . Check Water Accumulated Volume.
Press SELECT until Accum. Volume is displayed. If desired, reset to 0 by pressing CLEAR.
11. Check Infusion Rate. Press SELECT until Rate is displayed. To change, press or .
11
Detailed Operating Instructions (cont.)
Auto Priming
Automatic priming results in a small amount of fluid being discharged beyond the delivery set adapter.
12. Open both roller clamps on the delivery set. Remove the luer adapter cap from the delivery set, and place the cap in the luer adapter cap holder available on the pump.
13. Press PRIME for automatic priming of the delivery set. The Pump displays a series of P’s during the prime cycle and the rotor begins to turn. Formula is primed first, then water (if necessary) through the deliv­ery line. This fluid is not counted in Vol. Delivered. The prime cycle com­pletes in approximately 3 minutes. An audi­ble alarm sounds when priming is com­plete, the pump goes to Hold mode, and the Flow Rate flashes.
Begin Feeding
1. Connect delivery set adapter to the patients enteral feeding tube.
CAUTION: Verify enteral tube placement before beginning feeding.
2. Press START/HOLD to begin enteral feeding. The Pump primarily displays the flow rate. It briefly displays the formula volume delivered every 7 seconds.
3. Press START/HOLD and/or OFF to stop the pump at any time (Flow Rate, Dose Limit, Formula and Water Vol. Delivered, Formula and Water Accum. Volume, Volume Set, and Time Interval are retained in pump memory).
Pressing SELECT allows the user to scroll through and view Formula and Water side settings at any time.
Quick Reference
(How to set Flow Rate, Dose Limit,
or CLEAR Formula Volume Delivered)
Note: Changes can only be made to parameters at the START of a new feed­ing protocol to avoid improper water delivery. To facilitate this safety feature, parameters can only be changed when amount displayed in Vol. Delivered is 0.
If desired, clear Volume Delivered or reset all parameters to default settings by pressing and holding CLEAR for 4 seconds when Rate is displayed.
12
To set the Flow Rate
1. Press SELECT until Rate is displayed.
2. Enter a new flow rate by pressing (increase) or (decrease). Change is more rapid when the key is pressed and held.
3. Adjustable in 1 mL/hr increments from 1–295.
To set the Dose Limit
1. Press SELECT until Dose Limit is displayed.
2. Enter a new dose limit by pressing or
. Change is more rapid when the key is
pressed and held.
3. Adjustable in 1 mL increments from 1–2000.
4. For continuous feeding when no dose limit is desired, press until nOnE is dis­played.
To clear Volume Delivered and/or Accumulated Volume
If Formula Strength is activated to deliver diluted formula, Formula Vol. Delivered and Accum. Volume Delivered display the TOTAL DILUTED FEEDING (formula AND water) delivered.
1. Press SELECT until Vol. Delivered is displayed.
2. Press CLEAR.
3. Follow steps 1 and 2, using SELECT until Accum. Volume is displayed, and clear if desired.
Any time Formula Vol. Delivered is cleared, Water Vol. Delivered is automatically cleared. Likewise, if Water Vol. Delivered is cleared, Formula Vol. Delivered is automatically cleared.
Accumulated Volume displays full strength or diluted formula volume delivered over multiple dose limit feedings.
13
Quick Reference
(How to set Hydration with Automatic Flush)
Hydration with automatic 25 mL flush
Administers water volume with formula over a selected time interval during con­tinuous feeding. The Time Interval con­cludes with an automatic 25 mL water flush.
Select Water Volume to be delivered
Hydration with automatic flush is only recommended when Full Strength for­mula is being delivered. (If used with diluted formula, will result in further dilu­tion of the formula.)
1. Press SELECT until Formula Strength is displayed, and press , if necessary, to display FS (Full Strength) formula deliv­ery.
2. Press SELECT until Volume Set is displayed.
3. Press or until desired water volume is displayed. Adjustments can be made in 1 mL increments from 0–500 mL. Change is more rapid when the key is pressed and held.
Select Time Interval (in hours) over which water volume will be delivered
4. Press SELECT until Time Interval is displayed.
5. Press or . Time intervals are adjustable in 1 hour increments from 2 to 12 hours.
Maximum water delivery is 75 mL per hour
In the last 5 minutes of the selected time interval, formula delivery stops and the final 25 mL of water is auto­matically flushed through the line. Water Vol. Delivered resets to 0 after the time interval is completed and the entire water volume has been delivered.
Maximum Hydration Settings
Maximum
Time Water
Interval (hr) Volume (mL)
2 150 3 225 4 300 5 375 6 450 7 500 8 500
9 500 10 500 11 500 12 500
DOSE 500
Quick Reference
(How to set Formula Strength, Programmable
Flush, or CLEAR Water Volume Delivered)
14
Formula Strength (Dilution)
1. Press SELECT until Formula Strength is displayed.
2. Enter the formula strength required by pressing or .
3. Options are FS (Full Strength-No formula dilution), 3/4,2/3,1/2,1/3,1/4 strength, or H20 (no formula delivery, only water is delivered).
To clear Volume Delivered and/or Accumulated Volume
Water Volume Delivered is used to moni­tor water delivered when Hydration with Automatic Flush, or Programmable Flush options are activated.
1. Press SELECT until Vol. Delivered is displayed.
2. Press CLEAR.
3. Follow steps 1 and 2, using the SELECT key to move to Accum. Volume, and clear if desired.
Any time water volume delivered is reset to 0, formula volume delivered is auto­matically reset to 0. Likewise, when for­mula volume delivered is reset to 0, water volume delivered is automatically reset to
0.
Programmable flush after dose is delivered
A set volume of water for hydration or flushing that is automatically adminis­tered after the dose limit is delivered at a rate of 375 mL/hr.
1. Press SELECT until Volume Set is displayed.
2. Press or . The Water Volume is adjustable in 1 mL increments from 0–500. Change is more rapid when the key is pressed and held.
3. Press SELECT until Time Interval is displayed.
4. Press until doSE is displayed.
To deactivate Hydration and Programmable Flush (water side settings)
1. At Time Interval, press until nOnE is displayed.
2. Clear Formula Vol. Delivered and Accum. Volume. Set Rate and Dose Limit. Verify formula strength.
Water Accumulated Volume is used to monitor the total water volume delivered over multiple time intervals.
15
Alarms/Indicators
When turned on, the COMPAT DualFlo visible alarms under several conditions where desired performance cannot be maintained. Pressing START/HOLD after the alarm condition is identified silences the alarm and returns the pump to hold mode. Perform the necessary procedures to correct the alarm condition. Press START/HOLD again to resume normal operation.
no SEt (No set)
An audible alarm sounds, and no SEt is displayed. Indicates one of the following conditions:
A COMPAT DualFlo set is not being used.
One or both of the delivery set security doors are not closed and latched securely after the START/HOLDkey has been pressed. Immediately determine if the delivery set has become dislodged from the pump.
Caution: If formula is free flowing, take cor­rective action immediately. Stop the formu­la flow using the roller clamps on the deliv­ery set. Monitor patient for signs and symp­toms of formula/water over-delivery.
®
Enteral Feeding Pump is designed to provide audible and
LOW BAT illuminated (Low Battery)
The LOW BAT indicator on the display panel comes on 15 minutes before the pump ceases operation due to a low battery condition. After 15 minutes, the indicator flashes, an audible alarm sounds, and the pump stops operation.
Whenever LOW BAT is indicated, plug the pump into an electrical outlet, and normal feeding may continue. The battery
recharges anytime the pump is connected to an AC outlet regardless of whether the pump is turned on or off.
To avoid degrading battery life, it is impor­tant not to run down the battery.
FEEd Err (Feed error)
The pump stops, an audible alarm sounds, and FEEd Err is displayed.
Possible Condition
1. Water or Formula feeding container
is empty
2. An occlusion has caused delivery to stop
3. Drip chamber is overfilled
4. Drip chamber walls are heavily coated
with formula
5. Drip chamber bracket sensors need
cleaning
6. Drip chamber is improperly placed
Corrective Action
1. Refill container or start new delivery set. Follow instructions to prime the delivery set, and resume feeding
2. Identify point of occlusion and correct
3. Insert new administration set (use automatic priming to avoid “flooding” of drip chamber)
4. Manipulate chamber to clear drops in sensor path
5. Clean sensor with cotton swab and isopropyl alcohol
6. Reposition drip chamber
Alarms/Indicators (cont.)
16
doSE dEL (Dose delivered)
The pump stops, doSE dEL is displayed, and an audible alarm sounds when the pump has delivered an amount of formula equal to the Dose Limit.
Three alarm modes are available when the dose is complete:
Mode 1–Audible alarm sounds once, then stops
Mode 2–Audible alarm sounds once, stops, and then alarms at 5 minute intervals
Mode 3–Audible alarm sounds continuously
To select the desired mode, turn pump off. Then:
1. Turn the pump on by pressing ON.
2..8.8.8.8 is immediately displayed. While dis-
played, press and hold the and keys simultaneously for 4 seconds.
3. 1, 2, or 3 is displayed, which reflects the
corresponding alarm mode (Mode 1, Mode 2, or Mode 3).
4. Select the desired alarm mode by
pressing (increase) or (decrease).
5. Press SELECT to save the new alarm
mode. The pump audibly alarms once, ver­ifying the new alarm mode is saved. (If SELECT is not pressed, the new alarm mode is not saved.)
6. Press OFF to turn the pump off.
7. When the pump is turned back on, it is set
to operate with the new alarm mode.
HoLd Err (Hold error)
Pump is left on hold and START/HOLD is not pressed within 2 START/HOLD to silence the alarm, press again to start feeding if desired.
1
/2 minutes. Press
Running/Stopped illuminated
Indicates the status of the pump.
AC illuminated
Indicates the pump is operating on electrical power. Plug power cord into an AC electrical outlet whenever possible. No audible alarm sounds.
17
COMPAT®Enteral Feeding Pumps are reliable electromechanical devices. As with any mechanical device, problems may occur. The following tips will help you correct the minor in-service problems that may arise.
Troubleshooting
TROUBLESHOOTING GUIDE
SITUATION
Low battery
• alarm activated
(
LOW BAT
POSSIBLE CAUSE
• Battery charge is below performance level
)
SOLUTION
• Plug cord into AC outlet
illuminated
(
FEEd Err
alarm activated
)
• Formula or water container is empty
• Occlusion is restricting flow
• Replace or refill formula or water container
• Identify point of occlusion and free up
- Pump Set
• kinked tubing
• blocked tubing
• closed roller clamp
- Feeding tube
- Feeding container
• Drip chamber not properly
• Assemble properly
placed in pump
• Manipulate chamber to clear drops or condensation in sensor path
• Clean with cotton swab damp­ened with isopropyl alcohol
• Press START/HOLD button
• Clear Volume Delivered and start a new dose
• Examine tubing for proper placement, close door
• Load an appropriate set
• Assemble tubing properly around pump rotor. Close and
(
HoLd Err
) Pump
is left on hold
•(doSE dEL) Dose Complete
no SEt) alarm
•( activated
• Drip chamber walls coated with formula drops or condensation in sensor path
• Flow sensors housing need cleaning
• Pump is on hold for more than 2
1
/2 minutes
• Pump has delivered preset volume
• Door not closed and latched
• Wrong set is used
• Set is not properly in place
latch door.
Specifications
18
Delivery Sets
The DualFlo Enteral Delivery Pump offers a choice of delivery sets, including:
• 199407 COMPAT DualFlo®Piercing Spike and 1000 mL Vinyl Water Bag with preattached pump set
• 199447 COMPAT DualFlo 1000 mL Formula Vinyl Bag and 1000 mL Vinyl Water Bag with preat­tached pump set
®
• 199607 Compat® SpikeRight™ Spike Connector and 1000 mL Vinyl Water Bag with Preattached Pump Set.
19
Specifications (cont.)
Type of Device . . . . . . . . . . . . . . . . .
Volumetric enteral feeding pump
Mechanism . . . . . . . . . . . . . . . . . . . Rotary Peristaltic
Infusion Rate . . . . . . . . . . . . . . . . . 1–295 mL/hr in 1 mL increments
Dose Limit . . . . . . . . . . . . . . . . . . . . 0–2000 mL in 1 mL increments
Formula Strength . . . . . . . . . . . . . . FS (Full Strength),
1
/4, H20
3
/4, 2/3, 1/2, 1/3,
Water Volume Set . . . . . . . . . . . . . . 0–500 mL in 1 mL increments
Time Interval . . . . . . . . . . . . . . . . . 2 to 12 hours in 1 hour increments,
dose, none
Water Delivery . . . . . . . . . . . . . . . . 375 mL/hr fixed rate when
Programmable Flush after Dose Delivered is activated 75 mL per hour maximum water delivery when Hydration with Automatic Flush is activated
Accuracy . . . . . . . . . . . . . . . . . . . . . +/- 10% of selected flow rate
Occlusion Pressure . . . . . . . . . . . . Does not exceed 20 p.s.i.
Battery . . . . . . . . . . . . . . . . . . . . . . . Sealed lead acid Rechargeable
Automatically recharges when con­nected to an AC outlet Recharge time is 12 hours from complete discharge to full recharge Operating time is 8 hours at 100 mL/hour (fully charged battery) (Charge unit fully every 6 months)
Dimension . . . . . . . . . . . . . . . . . . . . 6.4” High x 9.75” Wide x 5.0” Deep
Weight . . . . . . . . . . . . . . . . . . . . . . . 7.0 pounds
Case Material . . . . . . . . . . . . . . . . . Fire and impact resistant plastic
Power Requirement . . . . . . . . . . . 120V , 187 mA, 60 Hz
Operating Temperature . . . . . . . . . 10
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 75% R.H. non-condensing
Storage Temperature . . . . . . . . . . . 0
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 95% R.H. non-condensing
o
–40oC (50o– 104oF)
o
–50oC (32o– 122oF)
Type of Protection Against Electrical Shock
. . . . . . . . . . . . . . . . . . . . . .
Class 1, Internally-Powered Equipment
Degree of Protection Against Electrical Shock
. . . . . . . . . . . . . . . . . . . .
Type BF
Degree of Protection Against Ingress of Fluids
. . . . . . . . . . . . . . . . . . . .
Ordinary Equipment – IPX0
Symbols on Pump and Accessorries
The following symbols may be found on the pump:
Refer to Accompanying Documents
Alternating Current
Type BF Protection (degree of protection against electrical shock—there is no conductive connection to the patient)
Medical Electrical Equipment
1) Classified with respect to electrical shock, fire and mechanical hazards in accordance with UL60601-1
20
2) Classified with respect to electrical shock, fire, mechanical and other specified hazards in accordance with CAN/CSA C22.2 No. 601.1
21
Self Test
The major functions of the pump may be tested by the following steps:
1. Turn the pump on by pressing ON.
2..8.8.8.8 is displayed immediately. While displayed, press and simulta­neously hold the and keys for 4 seconds. The display changes to either a 1, 2, or 3 (the dose delivered alarm mode).
3. Press and simultaneously hold the and keys again for 4 seconds.
4. After 4 seconds, an audible alarm will sound, ‘teSt’ is displayed momentarily, and then all eights (.8.8.8.8).
Factory Service
All service work should be performedby authorized personnel only. If the pump does not function properly:
• Contact your supplier of the COMPAT DualFlo Pump to arrange for Pump return and replacement.
• If your pump was supplied directly from Nestlé HealthCare Nutrition, contact your Nestlé HealthCare Nutrition Sales Representative or call Customer Service for instructions from anywhere in the U.S. and
®
Enteral Delivery
5. The pump rotor should begin to turn. To test the visual alarms, press either SELECT, CLEAR, , , PRIME, or START/HOLD. When any of these keys are pressed or the sensor is momentarily blocked, all visual alarm indicators should light, an audible alarm sounds, and the rotor will halt momentarily.
6. Any failure indicates a service check must be performed. See the Factory Service section for instructions on where to send the pump.
7. Turn the pump off by pressing OFF.
Canada at 1-877-338-4238 from 8:00 a.m. to 4:30 p.m. CST.
• If it becomes necessary to return a pump for factory service, a replace­ment pump will be sent to you. Carefully pack the defective pump in the carton the replacement pump arrived in and ship according to instructions given by the Nestlé HealthCare Representative.
• Unauthorized repairs void warranty coverage (see section on limited war­ranty).
Safety and Warnings
NOT FOR PARENTERAL USE
INTENDED EXCLUSIVELY FOR ENTERAL TUBE FEEDING AND PATIENT HYDRATION.
Do not use this device near flammable anesthetics, possible explosion hazard.
This device is designed to minimize the effects of uncontrolled electromagnetic
interference and other types of interfer­ence from external sources. Avoid use of other equipment that may cause erratic operation or degradation in the perfor­mance of this device.
If the integrity of the external protective conductor in the installation is in doubt, equipment shall be operated from it’s internal electrical power source.
Limited Warranty
22
Nestlé HealthCare Nutrition warrants the COMPAT DualFlo
®
Enteral Feeding Pump against defects in material and workmanship under normal use and ser­vice for a period of one (1) year from date of delivery. This warranty is valid only to the original purchaser and does not extend to any product or part thereof, which has been subjected to accident, abuse, alteration, misuse, or which has not been operated and maintained in accordance with prescribed instructions. This warranty shall not apply if the prod­uct has been repaired by anyone other than an authorized service representative
Cleaning
Routine cleaning of the COMPAT
®
DualFlo
pump should be an ongoing process in order to ensure maximum performance.
Caution
Always disconnect the pump from an electrical outlet before cleaning to avoid electric shock hazard.
Clean the pump housing and rotor regularly using a cloth or sponge with warm soapy water. (Prolonged exposure to alcohol, household detergents or strong cleansers can result in damage to pump housing.)
Clean drip sensors with a cotton swab and isopropyl alcohol.
Do not autoclave. Do not immerse pump in cleaning solutions.
of Nestlé HealthCare Nutrition.
Nestlé HealthCare Nutrition reserves the right to repair or replace (at its option) any pump which fails to meet the forego­ing warranty.
This warranty is in lieu of all other warranties expressed or implied includ­ing, without limitation, warranties of mer­chantability and of fitness for a particular purpose.
Disinfecting
If it is necessary to disinfect the entire pump, the following is recommended:
General Disinfectants
®
Cidex
(Advanced
Sterilization Products) pHisohex Hibiclens Isopropyl alcohol
AIDS and Hepatitis Disinfectants
10% concentration of 5.25% sodium
Tuberculosis Disinfectants
70% concentration of isopropyl alcohol
After application of these disinfecting agents, allow to air dry. These recom­mendations should not be substituted for official procedures that may differ among institutions. To determine the proper pro­cedure in a particular institution, contact the Infection Control Department.
®
(Winthrop-Breon Lab.)
®
(Stuart Pharmaceuticals)
hypochlorite (household bleach)
CMPT-10267-0308