Formula Strength, Programmable
Flush, Water Delivered14
Alarms/Indicators15
Troubleshooting17
Specifications18
Symbols on Pump and Accessories20
Self Test21
Factory Service21
Safety and Warnings21
NOT FOR PARENTERAL USE
INTENDED EXCLUSIVELY FOR
ENTERAL TUBE FEEDING AND
PATIENT HYDRATION.
For More Information
Contact your local Nestlé HealthCare
Nutrition Sales Representative or call
Customer Service from anywhere in the
U.S. and Canada at 1-877-338-4238.
Limited Warranty22
Cleaning22
Pump is not intended for neonatal use.
Caution: Federal law restricts this device to
sale by or on the order of a physician.
Introduction
2
The COMPAT DualFlo
administration of all enteral feeding solutions. Key features of this device include:
Patient hydration with
automatic flush
Patient hydration needs are facilitated by
administering up to 500 mL of water
over a selected time interval (in hours).
Time interval is adjustable in 1 hour
increments from 2 to 12 hours. In addition, the final 25 mL of water is automatically flushed through the line just before
the conclusion of the time interval.
Formula dilution
Provide formula dilution when necessary
by selecting Formula Strength. Formula
can be delivered at full strength (FS), or
diluted to
(all water delivery, no formula).
3
/4,2/3,1/2,1/3,1/4 strength, or H20
Programmable Flush after Dose is
Delivered
Deliver up to 500 mL of water in a
programmable water delivery to hydrate
the patient or flush the line after the formula dose limit is delivered. Water is
delivered at a fixed rate of 375 mL/hr.
®
Enteral Delivery Pump is an easy to use, versatile device for the
Delivery Set Security Doors
Reduces the likelihood of accidental
dislodgment of sets and enables
immediate detection of a free flow
condition.
Automatic priming
Automatic priming of formula and water
through the administration set allows
nurse time to perform other bedside
duties and may help reduce wasted sets
due to improper manual priming.
Versatility
• Formula rate adjustable in 1 mL/hr
increments from 1–295.
• Dose Limit adjustable in 1 mL/hr
increments from 1–2000.
• Formula Strength options include FS
(Full Strength), 3/4,2/3,1/2,1/3,1/4
strength, or H20 (water only).
• Water volume adjustable in 1 mL/hr
increments from 0–500.
• Time Interval adjustable in 1 hour
increments from 2 to 12 hours, with
programmable flush available as
dose is delivered.
• Ability to select basic full strength
formula delivery when water delivery
options are not desired.
3
Introduction (cont.)
Lifetime memory retention
Infusion Rate, Dose Limit, Formula
Strength, Water Volume Set, Time
Interval, Formula and Water Volume
Delivered, and Accumulated Volume are
automatically recalled when the pump is
turned back on.
Audio and visual alarms when
pump is turned on:
• Feed Error for detecting empty
container or occluded line
1
/2 minute Hold Alarm if user
•2
forgets to start the pump
• Dose Delivered
• Low Battery
• No Set
Designed for ease of use
• Large LED display window, alternates
displaying delivery rate and volume
delivered when pump is running.
• Dose Limit for Intermittent
Feeding with choice of 3 dose
complete alarm modes.
• Ability to quickly view formula and
water settings without interruption of
feeding.
• Adapter Cap Holder.
• Smooth surface of pump makes
cleaning easy.
• Lightweight and portable, the
pump is 7pounds and can clamp
to any IV pole.
• Adjustable Alarm Volume
Control knob.
• 8-hour battery life.
• Self-diagnostic mode allows for complete circuitry check when pump is
turned on.
Controls
Keypad Controls
1. ON, turns power on, memory recalls
previous settings.
2. OFF, turns power off, battery
charges when plugged in to
an electrical outlet.
3. SELECT, selects a delivery function
to display and/or change.
4. CLEAR, clears displayed value of
selected function to 0. (If RATE is
selected, pressing and holding
CLEAR for 4 seconds will reset all
parameters to default settings.)
5.increases flow rate, dose limit,
formula strength, water volume set,
or time interval when pressed.
Change is more rapid when pressed
and held.
6.decreases flow rate, dose limit,
formula strength, water volume
set, or time interval when pressed.
Change is more rapid when
pressed and held.
7. PRIME, automatically primes both
formula and water through the
administration set.
8. START/HOLD, starts and/or holds
pump operation. HOLD is used to
temporarily stop feeding, or silence
an alarm condition. Parameters can
only be changed when pump is in
HOLD mode. An audio and visual
alarm is activated if the pump has
been in HOLD mode for 2 1/2 minutes. Press START/HOLD to
resume normal operation.
9
1011
1
3
2
4
12
7
5
8
6
Function Displays
9. RUNNING, STOPPED, indicates
status of the pump when power is on.
10. AC, indicates the unit is plugged into
an electrical outlet.
11. LOW BAT, indicates when
the battery charge is below
performance level.
12. Display panel, indicates selected
parameter.
4
5
Controls (cont.)
Front of the pump
13. Flow Sensors and Drip
Chamber Guide
14. Tubing Guide
15. Drip Chamber Brackets
16. Rotary Peristaltic Pump
Mechanism
17. Door Latch
18. Delivery Set Security Door
Back of the Pump
19. Set Adapter Cap Holder
20. Power Cord
21. Alarm Volume Control
22. Pole Clamp
23. Directions
17
13
14
15
16
18
22
21
19
20
23
Controls (cont.)
6
29
30
31
32
25
27
24
26
28
7
Controls (cont.)
Formula Side Displays
24. RATE, displays the formula infusion
rate. Select from 1–295 mL/hr in 1
mL increments.
25. VOL. DELIVERED, displays full
strength or diluted formula volume
delivered.
26. DOSE LIMIT, displays dose limit.
Select from 1–2000 mL in 1 mL
increments. Dose is completed when
Volume Delivered equals the Dose
Limit set. Refer to Dose Delivered
under Alarms/Indicators. If nOnE is
displayed, this feature is deactivated.
27. ACCUM. VOLUME, displays full
strength or diluted formula volume
delivered over multiple feedings.
28. STRENGTH, (Formula Dilution)
displays formula strength
to be delivered. Select from FS (Full
Strength),
or H
3
/4,2/3,1/2,1/3,1/4 strength,
0 (water only).
2
Water Side Displays
29. VOLUME SET, displays the volume
of water to be delivered over the
selected Time Interval or at the end
of the dose limit. Select from 0–500
mL in 1 mL increments.
30. TIME INTERVAL, displays the time
interval (2 to 12 hours in 1 hour
increments) over which the selected
water volume is delivered. When
doSE is displayed, indicates
Programmable Flush is activated.
When nOnE is displayed, all water
side settings are deactivated.
31. VOL. DELIVERED, displays
amount of selected water volume
that has been delivered. At the conclusion of the Time Interval selected,
the Volume Delivered turns to 0 and
a new accumulation begins.
32. ACCUM. VOLUME, displays
water volume delivered over multiple
Time Intervals.
Preparation of COMPAT DualFlo®Delivery Set
8
DIAGRAM 1
Open pump doors by pulling door latch,
insert pump set drip chamber into drip
chamber bracket.
DIAGRAM 3
Insert pump adapter into bridge and
place tubing into tubing guide. Close and
latch delivery set security door.
DIAGRAM 2
Assemble silicone tubing around rotor.
DIAGRAM 4
Completely assembled pump set. Check
or change Pump Settings to prepare
enteral delivery.
9
Detailed Operating Instructions
Attach pump to IV pole using built-in pole
clamp. Plug power cord into a 120 VAC,
60 Hz. three wire grounded electrical
outlet if available.
Prepare COMPAT DualFlo
®
Delivery Set
1. Close both roller clamps on delivery set
prior to setup. Prepare and hang DualFlo
delivery set from an IV pole.
2. Open delivery set security doors by pulling
the latch on each door.
3. Insert water drip chamber into water side
drip chamber bracket, and formula drip
chamber into formula side drip chamber
bracket.
4. Carefully stretch water set silicone
tubing around rotor. Insert pump adapter
into bridge. Place tubing into tubing
guide, close and latch delivery set security
door.
5. Repeat step 4 with formula side
delivery set.
Check pump settings
1. Press ON- alarm sounds briefly and
displays .8.8.8.8. The pump has memory
and displays previous settings, and returns
to flashing Flow Rate on the display panel.
Note: Changes can only be made to parameters at the START of a new feeding protocol
to avoid improper water delivery. To facilitate this safety feature, parameters can
only be changed when amount displayed in
Vol. Delivered is 0.
If desired, clear Vol. Delivered or reset all
parameters to default settings by pressing
and holding CLEAR for 4 seconds when Rate
is displayed.
2. Check Formula Volume
Delivered. Press SELECT until Vol.
Delivered is displayed. If desired, reset to
0 by pressing CLEAR. This also resets
Water Volume Delivered to 0.
(Accumulated Volume is not affected.)
3. Check Dose Limit. Press SELECT
until Dose Limit is displayed. To change,
press (increase) or (decrease). For
continuous feeding where no dose limit is
desired, press until nOnE is displayed.
4. Check Formula AccumulatedVolume. Press SELECT until Accum.
Volume is displayed. If desired, reset to 0
by pressing CLEAR.
Detailed Operating Instructions (cont.)
10
Formula Strength (dilution)
5. To dilute formula, press SELECT
until strength is displayed. Press or
to set desired formula strength. Options
available are FS (Full Strength-no formula dilution),
or H
In steps 6–8, you have the option to
select Hydration with Automatic
Flush or Programmable Flush after
dose is delivered.
3
/4,2/3,1/2,1/3,1/4 strength,
0 (water only).
2
Hydration with automatic
25 mL flush
Administers water with formula over
a selected time interval during continuous feeding, concluding with an automatic 25 mL water flush.
6. To set water volume, press
SELECT until Volume Set is displayed.
Press or to set the desired volume of
water to be delivered. Adjustable in 1 mL
increments from 0–500 mL.
7. To select time interval, in hours,
press SELECT until Time Interval is
displayed. Change by pressing or .
Adjustable in 1 hour increments from
2 to 12 hours. MAXIMUM WATER DELIVERY IS 75 ML PER HOUR UP TO A MAXIMUM OF 500 ML. (See page 13 for maximum water volume setting at specific
time intervals.) The final 25 mL of
water is automatically flushed
through the line before the conclusion of the selected time interval.
Programmable Flush after dose
is delivered
Selected water volume delivered after the
formula dose limit is delivered. (For this
feature, water is delivered at a fixed rate
of 375 mL/hr.)
8. Follow steps 6 and 7. At Volume Set, enter
the desired volume of water to be delivered. When Time Interval is selected, press
until doSE is displayed. The
amount in Volume Set is
delivered after formula dose
limit is reached.
To deactivate Hydration and
Programmable Flush, at Time Interval,
press until nOne is displayed.
9. Check Water Volume Delivered.
Press SELECT until Vol. Delivered is displayed. This will be reset to 0 when
Formula Vol. Delivered is cleared (see step
2).
10 . Check Water Accumulated Volume.
Press SELECT until Accum. Volume is
displayed. If desired, reset to 0 by pressing
CLEAR.
11. Check Infusion Rate. Press
SELECT until Rate is displayed. To
change, press or .
11
Detailed Operating Instructions (cont.)
Auto Priming
Automatic priming results in a small
amount of fluid being discharged
beyond the delivery set adapter.
12. Open both roller clamps on the delivery
set. Remove the luer adapter cap from the
delivery set, and place the cap in the luer
adapter cap holder available on the pump.
13. Press PRIME for automatic priming of
the delivery set. The Pump displays a series
of P’s during the prime cycle and the rotor
begins to turn. Formula is primed first,
then water (if necessary) through the delivery line. This fluid is not counted inVol. Delivered. The prime cycle completes in approximately 3 minutes. An audible alarm sounds when priming is complete, the pump goes to Hold mode, and the
Flow Rate flashes.
Begin Feeding
1. Connect delivery set adapter to the
patients enteral feeding tube.
CAUTION: Verify enteral tube
placement before beginning feeding.
2. Press START/HOLD to begin enteral
feeding. The Pump primarily displays the
flow rate. It briefly displays the formula
volume delivered every 7 seconds.
3. Press START/HOLD and/or OFF to
stop the pump at any time (Flow Rate,
Dose Limit, Formula and Water Vol.
Delivered, Formula and Water Accum.
Volume, Volume Set, and Time Interval are
retained in pump memory).
Pressing SELECT allows the user to
scroll through and view Formula and
Water side settings at any time.
Quick Reference
(How to set Flow Rate, Dose Limit,
or CLEAR Formula Volume Delivered)
Note: Changes can only be made to parameters at the START of a new feeding protocol to avoid improper water delivery. To facilitate this safety feature,
parameters can only be changed when amount displayed in Vol. Delivered is 0.
If desired, clear Volume Delivered or reset all parameters to default settings
by pressing and holding CLEAR for 4 seconds when Rate is displayed.
12
To set the Flow Rate
1. Press SELECT until Rate is displayed.
2. Enter a new flow rate by pressing
(increase) or (decrease). Change is
more rapid when the key is pressed and
held.
3. Adjustable in 1 mL/hr increments from
1–295.
To set the Dose Limit
1. Press SELECT until Dose Limit is
displayed.
2. Enter a new dose limit by pressing or
. Change is more rapid when the key is
pressed and held.
3. Adjustable in 1 mL increments from
1–2000.
4. For continuous feeding when no dose
limit is desired, press until nOnE is displayed.
To clear Volume Delivered and/or
Accumulated Volume
If Formula Strength is activated to deliver
diluted formula, Formula Vol. Delivered
and Accum. Volume Delivered display the
TOTAL DILUTED FEEDING (formula
AND water) delivered.
1. Press SELECT until Vol. Delivered is
displayed.
2. Press CLEAR.
3. Follow steps 1 and 2, using SELECT
until Accum. Volume is displayed, and
clear if desired.
Any time Formula Vol. Delivered is cleared,
Water Vol. Delivered is automatically cleared.
Likewise, if Water Vol. Delivered is cleared,
Formula Vol. Delivered is automatically
cleared.
Accumulated Volume displays full strength or
diluted formula volume delivered over multiple
dose limit feedings.
13
Quick Reference
(How to set Hydration with Automatic Flush)
Hydration with automatic
25 mL flush
Administers water volume with formula
over a selected time interval during continuous feeding. The Time Interval concludes with an automatic 25 mL
water flush.
Select Water Volume to be delivered
Hydration with automatic flush is only
recommended when Full Strength formula is being delivered. (If used with
diluted formula, will result in further dilution of the formula.)
1. Press SELECT until Formula Strength
is displayed, and press , if necessary, to
display FS (Full Strength) formula delivery.
2. Press SELECT until Volume Set is
displayed.
3. Press or until desired water
volume is displayed. Adjustments can be
made in 1 mL increments from 0–500 mL.
Change is more rapid when the key is
pressed and held.
Select Time Interval (in hours) over
which water volume will be delivered
4. Press SELECT until Time Interval is
displayed.
5. Press or . Time intervals are
adjustable in 1 hour increments from
2 to 12 hours.
Maximum water delivery is 75 mL
per hour
In the last 5 minutes of the selected
time interval, formula delivery stops
and the final 25 mL of water is automatically flushed through the line.
Water Vol. Delivered resets to 0 after
the time interval is completed and
the entire water volume has been
delivered.
Maximum Hydration Settings
Maximum
TimeWater
Interval (hr)Volume (mL)
2150
3225
4300
5375
6450
7500
8500
9500
10500
11500
12500
DOSE500
Quick Reference
(How to set Formula Strength, Programmable
Flush, or CLEAR Water Volume Delivered)
14
Formula Strength (Dilution)
1. Press SELECT until Formula Strength
is displayed.
2. Enter the formula strength required by
pressing or .
3. Options are FS (Full Strength-No
formula dilution), 3/4,2/3,1/2,1/3,1/4
strength, or H20 (no formula delivery,
only water is delivered).
To clear Volume Delivered and/or
Accumulated Volume
Water Volume Delivered is used to monitor water delivered when Hydration with
Automatic Flush, or Programmable Flush
options are activated.
1. Press SELECT until Vol. Delivered is
displayed.
2. Press CLEAR.
3. Follow steps 1 and 2, using the SELECT
key to move to Accum. Volume, and clear
if desired.
Any time water volume delivered is reset
to 0, formula volume delivered is automatically reset to 0. Likewise, when formula volume delivered is reset to 0, water
volume delivered is automatically reset to
0.
Programmable flush after dose is
delivered
A set volume of water for hydration or
flushing that is automatically administered after the dose limit is delivered at a
rate of 375 mL/hr.
1. Press SELECT until Volume Set is
displayed.
2. Press or . The Water Volume is
adjustable in 1 mL increments from
0–500. Change is more rapid when the
key is pressed and held.
3. Press SELECT until Time Interval is
displayed.
4. Press until doSE is displayed.
To deactivate Hydration and
Programmable Flush (water side
settings)
1. At Time Interval, press until nOnE
is displayed.
2. Clear Formula Vol. Delivered and Accum.
Volume. Set Rate and Dose Limit. Verify
formula strength.
Water Accumulated Volume is used to
monitor the total water volume delivered
over multiple time intervals.
15
Alarms/Indicators
When turned on, the COMPAT DualFlo
visible alarms under several conditions where desired performance cannot be maintained.
Pressing START/HOLD after the alarm condition is identified silences the alarm and returns the
pump to hold mode. Perform the necessary procedures to correct the alarm condition. Press
START/HOLD again to resume normal operation.
no SEt (No set)
An audible alarm sounds, and no SEt is
displayed. Indicates one of the following
conditions:
•A COMPAT DualFlo set is not being used.
•One or both of the delivery set security
doors are not closed and latched securely
after the START/HOLDkey has been
pressed. Immediately determine if the
delivery set has become dislodged from
the pump.
Caution: If formula is free flowing, take corrective action immediately. Stop the formula flow using the roller clamps on the delivery set. Monitor patient for signs and symptoms of formula/water over-delivery.
®
Enteral Feeding Pump is designed to provide audible and
LOW BAT illuminated (Low Battery)
The LOW BAT indicator on the display
panel comes on 15 minutes before
the pump ceases operation due to a low
battery condition. After 15 minutes, the
indicator flashes, an audible alarm
sounds, and the pump stops operation.
Whenever LOW BAT is indicated, plug
the pump into an electrical outlet, and
normal feeding may continue. The battery
recharges anytime the pump is connected to
an AC outlet regardless of whether the
pump is turned on or off.
To avoid degrading battery life, it is important not to run down the battery.
FEEd Err (Feed error)
The pump stops, an audible alarm sounds, and FEEd Err is displayed.
Possible Condition
1. Water or Formula feeding container
is empty
2. An occlusion has caused delivery to stop
3. Drip chamber is overfilled
4. Drip chamber walls are heavily coated
with formula
5. Drip chamber bracket sensors need
cleaning
6. Drip chamber is improperly placed
Corrective Action
1. Refill container or start new delivery
set. Follow instructions to prime the
delivery set, and resume feeding
2. Identify point of occlusion and correct
3. Insert new administration set (use
automatic priming to avoid “flooding” of
drip chamber)
4. Manipulate chamber to clear drops in
sensor path
5. Clean sensor with cotton swab and
isopropyl alcohol
6. Reposition drip chamber
Alarms/Indicators (cont.)
16
doSE dEL (Dose delivered)
The pump stops, doSE dEL is displayed,
and an audible alarm sounds when the
pump has delivered an amount of formula
equal to the Dose Limit.
Three alarm modes are available when
the dose is complete:
•Mode 1–Audible alarm sounds once, then
stops
•Mode 2–Audible alarm sounds once,
stops, and then alarms at 5
minute intervals
•Mode 3–Audible alarm sounds
continuously
To select the desired mode, turn pump
off. Then:
1. Turn the pump on by pressing ON.
2..8.8.8.8 is immediately displayed. While dis-
played, press and hold the and keys
simultaneously for 4 seconds.
3. 1, 2, or 3 is displayed, which reflects the
corresponding alarm mode
(Mode 1, Mode 2, or Mode 3).
4. Select the desired alarm mode by
pressing (increase) or
(decrease).
5. Press SELECT to save the new alarm
mode. The pump audibly alarms once, verifying the new alarm mode is saved. (If
SELECT is not pressed, the new alarm
mode is not saved.)
6. Press OFF to turn the pump off.
7. When the pump is turned back on, it is set
to operate with the new alarm mode.
HoLd Err (Hold error)
Pump is left on hold and START/HOLD
is not pressed within 2
START/HOLD to silence the alarm,
press again to start feeding if desired.
1
/2 minutes. Press
Running/Stopped illuminated
Indicates the status of the pump.
AC illuminated
Indicates the pump is operating on
electrical power. Plug power cord into an
AC electrical outlet whenever possible.
No audible alarm sounds.
17
COMPAT®Enteral Feeding Pumps are reliable electromechanical devices. As with any
mechanical device, problems may occur. The following tips will help you correct the
minor in-service problems that may arise.
Troubleshooting
TROUBLESHOOTING GUIDE
SITUATION
Low battery
•
alarm activated
•
(
LOW BAT
POSSIBLE CAUSE
• Battery charge is below
performance level
)
SOLUTION
• Plug cord into AC outlet
illuminated
•
(
FEEd Err
alarm activated
)
• Formula or water container
is empty
• Occlusion is restricting flow
• Replace or refill formula or
water container
• Identify point of occlusion
and free up
- Pump Set
• kinked tubing
• blocked tubing
• closed roller clamp
- Feeding tube
- Feeding container
• Drip chamber not properly
• Assemble properly
placed in pump
• Manipulate chamber to clear
drops or condensation in
sensor path
• Clean with cotton swab dampened with isopropyl alcohol
• Press START/HOLD button
• Clear Volume Delivered and
start a new dose
• Examine tubing for proper
placement, close door
• Load an appropriate set
• Assemble tubing properly
around pump rotor. Close and
•
(
HoLd Err
) Pump
is left on hold
•(doSE dEL)
Dose Complete
no SEt) alarm
•(
activated
• Drip chamber walls coated
with formula drops or
condensation in sensor path
• Flow sensors housing need
cleaning
• Pump is on hold for more
than 2
1
/2 minutes
• Pump has delivered
preset volume
• Door not closed and latched
• Wrong set is used
• Set is not properly in place
latch door.
Specifications
18
Delivery Sets
The DualFlo Enteral Delivery Pump offers a choice of delivery sets, including:
• 199407 COMPAT DualFlo®Piercing
Spike and 1000 mL Vinyl Water Bag
with preattached
pump set
• 199447 COMPAT DualFlo
1000 mL Formula Vinyl Bag and
1000 mL Vinyl Water Bag with preattached pump set
®
• 199607 Compat® SpikeRight™
Spike Connector and 1000 mL Vinyl
Water Bag with Preattached Pump
Set.
Automatically recharges when connected to an AC outlet
Recharge time is 12 hours from
complete discharge to full recharge
Operating time is 8 hours at 100
mL/hour (fully charged battery)
(Charge unit fully every 6 months)
Dimension . . . . . . . . . . . . . . . . . . . .6.4” High x 9.75” Wide x 5.0” Deep
Type BF Protection (degree of protection against
electrical shock—there is no conductive connection
to the patient)
Medical Electrical Equipment
1) Classified with respect to electrical shock, fire
and mechanical hazards in accordance with
UL60601-1
20
2) Classified with respect to electrical shock, fire,
mechanical and other specified hazards in
accordance with CAN/CSA C22.2 No. 601.1
21
Self Test
The major functions of the pump may be
tested by the following steps:
1. Turn the pump on by pressing ON.
2..8.8.8.8 is displayed immediately.
While displayed, press and simultaneously hold the and keys for
4 seconds. The display changes to
either a 1, 2, or 3 (the dose delivered
alarm mode).
3. Press and simultaneously hold the
and keys again for 4 seconds.
4. After 4 seconds, an audible alarm will
sound, ‘teSt’ is displayed momentarily,
and then all eights (.8.8.8.8).
Factory Service
All service work should be performedby
authorized personnel only. If the pump
does not function properly:
• Contact your supplier of the
COMPAT DualFlo
Pump to arrange for Pump return
and replacement.
• If your pump was supplied directly
from Nestlé HealthCare Nutrition,
contact your Nestlé HealthCare
Nutrition Sales Representative or call
Customer Service for instructions
from anywhere in the U.S. and
®
Enteral Delivery
5. The pump rotor should begin to
turn. To test the visual alarms, press
either SELECT, CLEAR, , ,
PRIME, or START/HOLD. When
any of these keys are pressed or the
sensor is momentarily blocked, all
visual alarm indicators should light,
an audible alarm sounds, and the
rotor will halt momentarily.
6. Any failure indicates a service check
must be performed. See the Factory
Service section for instructions on
where to send the pump.
7. Turn the pump off by pressing OFF.
Canada at 1-877-338-4238 from
8:00 a.m. to 4:30 p.m. CST.
• If it becomes necessary to return a
pump for factory service, a replacement pump will be sent to you.
Carefully pack the defective pump in
the carton the replacement pump
arrived in and ship according to
instructions given by the Nestlé
HealthCare Representative.
• Unauthorized repairs void warranty
coverage (see section on limited warranty).
Safety and Warnings
NOT FOR PARENTERAL USE
INTENDED EXCLUSIVELY FOR
ENTERAL TUBE FEEDING AND
PATIENT HYDRATION.
Do not use this device near flammable
anesthetics, possible explosion hazard.
This device is designed to minimize the
effects of uncontrolled electromagnetic
interference and other types of interference from external sources. Avoid use of
other equipment that may cause erratic
operation or degradation in the performance of this device.
If the integrity of the external protective
conductor in the installation is in doubt,
equipment shall be operated from it’s
internal electrical power source.
Limited Warranty
22
Nestlé HealthCare Nutrition warrants the
COMPAT DualFlo
®
Enteral Feeding
Pump against defects in material and
workmanship under normal use and service for a period of one (1) year from
date of delivery. This warranty is valid
only to the original purchaser and does
not extend to any product or part thereof,
which has been subjected to accident,
abuse, alteration, misuse, or which has
not been operated and maintained in
accordance with prescribed instructions.
This warranty shall not apply if the product has been repaired by anyone other
than an authorized service representative
Cleaning
Routine cleaning of the COMPAT
®
DualFlo
pump should be an ongoing
process in order to ensure maximum
performance.
Caution
Always disconnect the pump from an
electrical outlet before cleaning to
avoid electric shock hazard.
Clean the pump housing and rotor
regularly using a cloth or sponge
with warm soapy water. (Prolonged
exposure to alcohol, household
detergents or strong cleansers can
result in damage to pump housing.)
Clean drip sensors with a cotton swab
and isopropyl alcohol.
Do not autoclave. Do not immerse
pump in cleaning solutions.
of Nestlé HealthCare Nutrition.
Nestlé HealthCare Nutrition reserves the
right to repair or replace (at its option)
any pump which fails to meet the foregoing warranty.
This warranty is in lieu of all other
warranties expressed or implied including, without limitation, warranties of merchantability and of fitness for a particular
purpose.
Disinfecting
If it is necessary to disinfect the entire
pump, the following is recommended:
After application of these disinfecting
agents, allow to air dry. These recommendations should not be substituted for
official procedures that may differ among
institutions. To determine the proper procedure in a particular institution, contact
the Infection Control Department.
®
(Winthrop-Breon Lab.)
®
(Stuart Pharmaceuticals)
hypochlorite (household bleach)
CMPT-10267-0308
SN1032