Formula Strength, Programmable
Flush, Water Delivered14
Alarms/Indicators15
Troubleshooting17
Specifications18
Symbols on Pump and Accessories20
Self Test21
Factory Service21
Safety and Warnings21
NOT FOR PARENTERAL USE
INTENDED EXCLUSIVELY FOR
ENTERAL TUBE FEEDING AND
PATIENT HYDRATION.
For More Information
Contact your local Nestlé HealthCare
Nutrition Sales Representative or call
Customer Service from anywhere in the
U.S. and Canada at 1-877-338-4238.
Limited Warranty22
Cleaning22
Pump is not intended for neonatal use.
Caution: Federal law restricts this device to
sale by or on the order of a physician.
Introduction
2
The COMPAT DualFlo
administration of all enteral feeding solutions. Key features of this device include:
Patient hydration with
automatic flush
Patient hydration needs are facilitated by
administering up to 500 mL of water
over a selected time interval (in hours).
Time interval is adjustable in 1 hour
increments from 2 to 12 hours. In addition, the final 25 mL of water is automatically flushed through the line just before
the conclusion of the time interval.
Formula dilution
Provide formula dilution when necessary
by selecting Formula Strength. Formula
can be delivered at full strength (FS), or
diluted to
(all water delivery, no formula).
3
/4,2/3,1/2,1/3,1/4 strength, or H20
Programmable Flush after Dose is
Delivered
Deliver up to 500 mL of water in a
programmable water delivery to hydrate
the patient or flush the line after the formula dose limit is delivered. Water is
delivered at a fixed rate of 375 mL/hr.
®
Enteral Delivery Pump is an easy to use, versatile device for the
Delivery Set Security Doors
Reduces the likelihood of accidental
dislodgment of sets and enables
immediate detection of a free flow
condition.
Automatic priming
Automatic priming of formula and water
through the administration set allows
nurse time to perform other bedside
duties and may help reduce wasted sets
due to improper manual priming.
Versatility
• Formula rate adjustable in 1 mL/hr
increments from 1–295.
• Dose Limit adjustable in 1 mL/hr
increments from 1–2000.
• Formula Strength options include FS
(Full Strength), 3/4,2/3,1/2,1/3,1/4
strength, or H20 (water only).
• Water volume adjustable in 1 mL/hr
increments from 0–500.
• Time Interval adjustable in 1 hour
increments from 2 to 12 hours, with
programmable flush available as
dose is delivered.
• Ability to select basic full strength
formula delivery when water delivery
options are not desired.
3
Introduction (cont.)
Lifetime memory retention
Infusion Rate, Dose Limit, Formula
Strength, Water Volume Set, Time
Interval, Formula and Water Volume
Delivered, and Accumulated Volume are
automatically recalled when the pump is
turned back on.
Audio and visual alarms when
pump is turned on:
• Feed Error for detecting empty
container or occluded line
1
/2 minute Hold Alarm if user
•2
forgets to start the pump
• Dose Delivered
• Low Battery
• No Set
Designed for ease of use
• Large LED display window, alternates
displaying delivery rate and volume
delivered when pump is running.
• Dose Limit for Intermittent
Feeding with choice of 3 dose
complete alarm modes.
• Ability to quickly view formula and
water settings without interruption of
feeding.
• Adapter Cap Holder.
• Smooth surface of pump makes
cleaning easy.
• Lightweight and portable, the
pump is 7pounds and can clamp
to any IV pole.
• Adjustable Alarm Volume
Control knob.
• 8-hour battery life.
• Self-diagnostic mode allows for complete circuitry check when pump is
turned on.
Controls
Keypad Controls
1. ON, turns power on, memory recalls
previous settings.
2. OFF, turns power off, battery
charges when plugged in to
an electrical outlet.
3. SELECT, selects a delivery function
to display and/or change.
4. CLEAR, clears displayed value of
selected function to 0. (If RATE is
selected, pressing and holding
CLEAR for 4 seconds will reset all
parameters to default settings.)
5.increases flow rate, dose limit,
formula strength, water volume set,
or time interval when pressed.
Change is more rapid when pressed
and held.
6.decreases flow rate, dose limit,
formula strength, water volume
set, or time interval when pressed.
Change is more rapid when
pressed and held.
7. PRIME, automatically primes both
formula and water through the
administration set.
8. START/HOLD, starts and/or holds
pump operation. HOLD is used to
temporarily stop feeding, or silence
an alarm condition. Parameters can
only be changed when pump is in
HOLD mode. An audio and visual
alarm is activated if the pump has
been in HOLD mode for 2 1/2 minutes. Press START/HOLD to
resume normal operation.
9
1011
1
3
2
4
12
7
5
8
6
Function Displays
9. RUNNING, STOPPED, indicates
status of the pump when power is on.
10. AC, indicates the unit is plugged into
an electrical outlet.
11. LOW BAT, indicates when
the battery charge is below
performance level.
12. Display panel, indicates selected
parameter.
4
5
Controls (cont.)
Front of the pump
13. Flow Sensors and Drip
Chamber Guide
14. Tubing Guide
15. Drip Chamber Brackets
16. Rotary Peristaltic Pump
Mechanism
17. Door Latch
18. Delivery Set Security Door
Back of the Pump
19. Set Adapter Cap Holder
20. Power Cord
21. Alarm Volume Control
22. Pole Clamp
23. Directions
17
13
14
15
16
18
22
21
19
20
23
Controls (cont.)
6
29
30
31
32
25
27
24
26
28
7
Controls (cont.)
Formula Side Displays
24. RATE, displays the formula infusion
rate. Select from 1–295 mL/hr in 1
mL increments.
25. VOL. DELIVERED, displays full
strength or diluted formula volume
delivered.
26. DOSE LIMIT, displays dose limit.
Select from 1–2000 mL in 1 mL
increments. Dose is completed when
Volume Delivered equals the Dose
Limit set. Refer to Dose Delivered
under Alarms/Indicators. If nOnE is
displayed, this feature is deactivated.
27. ACCUM. VOLUME, displays full
strength or diluted formula volume
delivered over multiple feedings.
28. STRENGTH, (Formula Dilution)
displays formula strength
to be delivered. Select from FS (Full
Strength),
or H
3
/4,2/3,1/2,1/3,1/4 strength,
0 (water only).
2
Water Side Displays
29. VOLUME SET, displays the volume
of water to be delivered over the
selected Time Interval or at the end
of the dose limit. Select from 0–500
mL in 1 mL increments.
30. TIME INTERVAL, displays the time
interval (2 to 12 hours in 1 hour
increments) over which the selected
water volume is delivered. When
doSE is displayed, indicates
Programmable Flush is activated.
When nOnE is displayed, all water
side settings are deactivated.
31. VOL. DELIVERED, displays
amount of selected water volume
that has been delivered. At the conclusion of the Time Interval selected,
the Volume Delivered turns to 0 and
a new accumulation begins.
32. ACCUM. VOLUME, displays
water volume delivered over multiple
Time Intervals.
Loading...
+ 16 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.