COMPAT
®
FEEDING PUMP
Operating Instructions
199235
Operating Instructions:
1. Plug the power cord into a 115 VAC., 60 HZ. three wire grounded outlet.
2. Press ON; alarm will sound briefly. Remember, the pump has lifetime memory and will
automatically recall the previous parameter settings. (Diagram 1)
3. Check Infusion Rate. If change is desired, press INC or DEC to change rate setting.
(Diagram 1)
4. Check Dose Limit by pressing DOSE LIMIT SET. If change is desired, press INC or DEC
within 3 seconds to change limit setting (Diagram 1). Zero Dose Limit implies no dose
limit monitoring by pump.
5. Press TOTAL RESET to clear Total Volume, if desired.
6. Prepare and hang COMPAT® Delivery Set from an IV pole. Close roller clamp.
7. Insert drip chamber into drip chamber brackets. (Diagram 2)
8. Open roller clamp. Fill tubing with formula until all air is expelled from tubing. Close
clamp.
9. Stretch silicone tubing around rotor. Insert pump adaptor into bridge. (Diagram 3)
10. Place tubing into tubing guide. (Diagram 4)
11. Attach pump set to patient’s enteral feeding tube.
12. Open roller clamp.
13. Press RUN/HOLD to start pump.
14. OFF will turn off the pump at any time. (Infusion Rate, Dose Limit, Volume Delivered, and
Accumulated Volume are retained.)