C9040 Operator Instructions
1. INTRODUCTION
The C9040 is a handheld, thermocouple simulator. It has
been designed specically for the simple checking of thermocouple thermometers by simulating one of ve selectable
temperatures to give a reading on the test instrument. The
C9040 is available in three versions.
C9040K type K sub-miniature plug.
C9040TM type T sub-miniature plug.
C9040TL type T lumberg plug.
The C9040 is highly stable and is supplied with a UKAS
certicate of calibration.
2. PREPARATION FOR USE
A BATTERY TEST must be performed before use to ensure
that enough power is available for temperature simulations.
Switch on the C9040 and set the rotary control to BATTERY
TEST.
The LED on the front of the instrument marked TEST should
glow brightly. If the LED is dim or does not glow at all, then
the battery should be replaced.
Note: Do not leave the instrument in BATTERY TEST mode
as this drains the battery quickly.
3. CONNECTING THE C9040 TO THE
THERMOMETER TO BE TESTED.
Two calibration checks are possible with the C9040:
a) A quick calibration check to an uncertainty +/- 0.5°C
b) A longer test giving readings to an uncertainty +/-
0.2°C
The two types of test are described as follows:
a) Quick Test to +/- 0.5°C
This is the most likely test and will allow the
instrument to be tested for basic functionality to an
uncertainty of +/- 0.5°C. The C9040 and the
instrument on test must have been in the same
ambient conditions for at least 15 minutes. Connect
the C9040 to the thermometer to be tested at least 2
minutes prior to checking the calibration. Switch on
the C9040 using the Blue button on the side of the
instrument once. This will allow 3 minutes of
continuous operation, before auto-switch off. Switch
on the thermometer to be tested. Select the
appropriate output on the C9040 using the rotary
control. Check the reading on the thermometer to be
tested. The reading should be within +/- 0.5°C +/- the
instrument under test specication, of the selected
temperature. No reference to the UKAS certicate
is required.
b) Thorough Test to an uncertainty of +/- 0.2°C
This test is performed exactly as the above except for the
following changes:
The two instruments must be left in the same environment for
at least 2 Hrs and during this time the C9040 should remain
connected. For best results the test should be conducted in a
stable temperature environment.
NOTE (Thorough Test): When assessing the reading on the
thermometer under test, an allowance must be made for the
accuracy of the C9040. Refer to your UKAS certicate for
details.
NOTE (All Tests): Just prior to auto-switch off readings may
become erratic. To continue correct operation simply press
the on/off button once more to provide another 3 minutes of
continuous operation.
4. UKAS CERTIFICATION
The C9040 is supplied complete with a UKAS certicate
of calibration produced in the Comark UKAS accredited
calibration laboratory. To ensure continued correct
operation and optimum performance it is recommended
that the C9040 is returned at least annually for a calibration
check and issue of new UKAS certicate.
5. SPECIFICATION
Temperature Range
Type K -20°C, 0°C, +50°C, +100°C, +500°C
Type T -18°C, 0°C, +5°C, +70°C, +100°C
Each output is calibrated to be accurate to ±0.2°C at +23°C. Refer to the
UKAS certicate of calibration for calibrated results.
Cold Junction
Specication 0.05°C/°C away from +23°C
Ambient Operating
Conditions
Temperature +18°C to +28°C
Humidity 10-90%RH Non-condensing
Storage 0 to +40°C
Battery PP3,6LF22
Battery life 500 hrs normal use.
Note: Extensive use of the BATTERY TEST function will seriously reduce
battery life.
EMC Emission — EN 50081-1
No emissions above EN 55 022
Class B limits
Immunity — EN 50082-1
Performance to Criterion B
Dimensions L 117mm x W 70mm x D 35mm
Weight 120g excluding battery