GENERAL INFORMATION
TRANSDUCERS SHOULD NOT BE DISCONNECTED FROM THE
WARNING
PRESSURE SOURCE WHILST CONTAINING GASES WHICH ARE
DANGEROUS AND/OR UNDER PRESSURE.
DO NOT APPLY PRESSURES BEYOND THE OVER RANGE MAXIMUM FOR THE TRANSDUCER OR DAMAGE WILL OCCUR.
REFER TO SECTION 3 - CONNECTION TO PRESSURE SOURCES
BEFORE OPERATING THE INSTRUMENT.
TO AVOID ELECTRIC SHOCK DO NOT ALLOW ANY TRANSDUCER
OR CABLE TO COME INTO CONTACT WITH LIVE ELECTRICAL
POWER CONDUCTORS WITH VOLTAGES IN EXCESS OF 30V AC
RMS OR 60V DC.
CAUTION
Pressure Transducers
This precision instrument has been designed for use with the
extensive range of Comark pressure transducers. The use of
other transducers may impair the performance and accuracy of the
instrument. Full details of Comark transducers may be obtained
from Comark Customer Support department or your local distributor.
Repeated sharp exing can break transducer connection cables.
To prolong lead life, avoid sharp bends or kinks in the leads, especially near the connectors.
This instrument is manufactured in accordance with the company’s ISO 9001 Quality Approved System.
CERTIFICATE No. FM29700
This instrument complies with the Electromagnetic Compatibility Directive EN61326-1:2006.
Declarations of Conformity available. Contact Comark Customer Support or your local Distributor.
In line with its policy of continuous development, Comark Limited reserves the
right to alter the instrument specication without prior notice. Further information
is available from Comark Limited or your distributor.
3
Page 5
CALIBRATION, CERTIFICATION AND SERVICE
Certication
Comark can provide certicates of calibration for its whole product
range, to suit ISO 9000 and other quality assurance procedures, food
hygiene regulations, HACCPs and environmental regulations. Comark
certicates are produced by independent quality controlled processes
which compare product performance against agreed National Standards. For peace of mind and best practice Comark recommend annual
certication/recalibration.
Two levels of certication are available for infra-red temperature and
non temperature instruments, excluding humidity:- UKAS certicates via
an external accredited laboratory and NPL traceable certicates from
the Comark calibration laboratory.
Three levels of certication are available for contact temperature and
humidity products and these are detailed here:
a) UKAS Temperature Certication
The Comark UKAS (United Kingdom Accreditation Service)
accredited temperature calibration laboratory is one of the nest in the UK. Comark UKAS certicates can offer the lowest
uncertainty of 0.01°C and provide independent proof of correct
calibration using equipment and procedures audited by UKAS
inspectors. The equipment used is fully traceable to the National Physical Laboratory.
b) UKAS Humidity Certication
In addition to the Comark temperature laboratory, the humidity
laboratory continues the tradition of high accuracy certication
and a wide range (25% to 90%RH) with uncertainties of 2.8% of
reading. This range also includes Dew point measurements.
c) NPL Traceable Certication
Comark NPL certicates are traceable to the National Physical
Laboratory and can offer uncertainty as low as 0.3°C.
4
Page 6
Conformance
Certicates of conformance can be supplied for new, serviced and
recalibrated instruments. These conrm that instruments are within their
original manufactured specication.
Service/Repairs
Regular servicing and any required repairs, under warranty or after, are
available from the Comark Service Department.
For more information on all Comark certication, calibration and service
facilities please call Comark Customer Support or contact your local
distributor.
5
Page 7
1 INTRODUCTION
1.1 Transducer Input
The C9550 external pressure meter accepts all Comark
pressure transducers model 3392, 3393, 3397 or 3398.
Connection is made by means of the 8-pin socket and
the 1.5 metre interface cable. An inbuilt serial interface
allows the C9550 to communicate with the transducer
and provide correct display of the pressure value in any
one of the pressure units selected. Suitable pressure
transducers are:
Model Pressure range, bar Diaphragm Type
3392/PG Series 0.1 to 4000 Internal Gauge
3393/PA Series 0.25 to 16 Internal Absolute
3397/PGF Series 0.1 to 600 Front Flush Gauge
3398/PAF Series 0.25 to 16 Front FlushAbsolute
Table 1. Transducer Model References
1.2 Transducer Operator Manual
The separate Transducer Operator Manual provides
details of the installation, troubleshooting and calibration of transducers.
1.3 Transducer Range Data Sheet
Pressure ranges extend from 0.1 to 4000 bar. The
separate Transducer Range Data sheet, Ref: C089,
lists all pressure ranges available in each of Gauge /
Absolute, Internal Diaphragm / Front Flush Diaphragm,
0.5% / 0.25% accuracy types.
1.4 Ranges, scales and resolutions
A tabular listing of the scales and resolutions of Comark
stocked ranges is provided in section 5.3. The complete listing for all 24 ranges is available separately as
Engineering Note 18, and is available from the Customer Service Department.
6
Page 8
2 KEY FUNCTIONS
2.1 ON/OFF Key
Press once to switch on the instrument. Repeat to
switch off.
Holding down the key when switching on will display the
maximum pressure range for the attached transducer in
Bar. When the key is released the previously set pressure units will be employed.
Holding down the key when switching off will display all
segments as a test.
2.2 ZERO Key
Press to zero the current readings on the display to nullify any offsets caused by large temperature variations,
or to provide a tare function. The instrument discards
any such offset when switched off, and will apply no offset when switched back on. The zero key can be used
whenever offsets occur. Please note that pressing the
zero key has no effect whilst the instrument is in hold or
record modes.
Note that if the zero function is used to set to zero
(tare) a standing pressure reading, the -02- overpressure warning will continue to occur at 110% of full scale
transducer pressure, although the numerical value displayed will not then be 110% of full scale. This ensures
that the display warning for overpressure will apply to
the safety of the transducer.
2.3 SCALE Key
Press to display the pressure measurement in each of
the different scales available. The selected scale unit
is retained in memory even after switch off, so that the
selected unit can be displayed when the instrument is
switched on again.
See Section 5 for a fuller explanation of the C9550
pressure scaling system.
7
Page 9
2.4 HOLD Key
Press to hold the displayed reading and press again to
resume measurement. The scale key will operate while
the instrument is in hold mode so that the held reading
can be displayed in different scales.
2.5 FILTER Key
Press to lter out transient readings. In lter mode the
instrument will average the readings taken over the
last 4 seconds (16 readings) to provide a more stable
reading on the display. Press the key again to resume
normal measurement.
2.6 RECORD Key
Press once to put the instrument into record mode.
This will be indicated by the word RECORD on the display. Maximum and minimum pressure readings can be
stored and displayed while the instrument is in record
mode. When the instrument is in record mode, press
key once to display the current maximum reading and
press again to display the current minimum reading.
Pressing the key a third time will return the display to
normal measurement.
Record mode can be cancelled and the max/min readings cleared by pressing and holding down the key until
the word RECORD disappears from the display.
The scale key will operate while the instrument is in
record mode so that the max/min readings can be
displayed in different scales.
2.7 AUTO Switch Off
The instrument can be programmed to auto switch off by
pressing and holding down the HOLD key while using the
ON/OFF key to switch on the instrument. The word AUTO
appears on the display to indicate auto switch off mode and
the instrument will switch itself off approximately 3 minutes
after the last use of the keypad. Auto switch off mode can
be cancelled by pressing and holding down the HOLD key
while switching on manually with the ON/OFF key.
8
Page 10
3 PRESSURE SOURCE CONNECTION
Transducers for use with the C9550 instruments are tted with
G½A male connectors The following checks must be made
before connecting them to a pressure source:-
•See warnings on page 2.
•Refer to the separate Transducer Operator
Manual.
•Please ensure that you check your own safety
standards before connecting to a pressure source.
•Check the security of all ttings.
•Do not undo pressure connections until the
pressure is released rst.
•Check the pressure rating on the transducer label.
•Check that the pressure source does not exceed
the ratings for the transducer as set out in Table 2.
•Check that the gases to be measured will not
damage or corrode the stainless steels in the
transducer.
TransducerMaximum
Pressure RangesOverpressure
0.1 to 16 bar350% of Full Scale25 to 600 bar200% of Full Scale1000 bar150% of Full Scale
1600 to 4000 bar120% of Full Scale
Table 2. Overpressure limits for ranges
9
Page 11
4 OTHER FUNCTIONS
4.1 Battery Low
Low battery is indicated by the symbol BAT on the
display. Replace the battery as soon as possible after
this symbol appears.
4.2 Over 9999 Reading Indication
If the numeric value for the display exceeds ±9999, with
or without a decimal point, the display will blink and
suppress the top (5th) digit, whilst continuing with correct values of the lower 4 digits.
4.3 Absolute or Gauge Measurement
C9550 Series instruments can be used with either
gauge or absolute pressure transducers; no adjustment
is required between use of either type. Refer to the
transducer label to determine the measurement type.
As the transducer only has a single port a direct differential measurement is not possible; in this case the
zero function may be used between measurement from
two transducers.
4.4 Error Messages
The following error messages may appear on the display to indicate incorrect operation or malfunction:
-00- Sensor not connected
-01- Overow of A/D converter
-02- Overpressure (110% of Full Scale pressure)
-03- Mode not feasible
SCL Invalid Scale Units
10
•Error -00- indicates no sensor is connected.
•Error -01- occurs if connected transducer
signal is too high (for example because of a
transducer defect or a pressure overload).
•Error -02- indicates an attempt to measure
a pressure (either positive or negative) that
exceeds the range of the transducer by
more than 10%. If this occurs the pressure
must be reduced to avoid damage to the
sensor in the transducer.
Page 12
•Error -03- indicates either
a) an attempt to attach an incompatible sensor type.
b) an interrupted data transmission between sensor and instrument.
•Error SCL - indicates the scale units are
not available on this transducer.
In the case of b) either reconnect the transducer or
switch OFF then ON again. In the case of c) press the
SCL key for the next supported scale units.
4.5 Battery Replacement
Unscrew the two screws at the base of the instrument
and slide off the battery cover. Replace the battery with
a 9V PP3 type battery. Replace the battery cover taking care not to over tighten the two screws.
11
Page 13
5. UNITS OF MEASUREMENT
5.1 The C9550 Pressure Scaling System
Units other than the SI-units bar will be calculated by
the C9550 using internationally agreed conversion factors. Once powered up the instrument will display values in the same scale as last selected. One or multiple
depressions of the SCALE key will change the display
to read other units of measurement.
For a given transducer, only those pressure scales with
full scale numeric values less than 11,000 are available for selection; thus the lower pressure transducers
have progressively more scales, to a maximum of 12
for the 0.1 bar transducer. The full display sequence of
pressure scales is Kgcm-2, mBar, Bar, inHg, Torr, inH2O,
kPa, Pa, mmHg, PSI, mmH2O, mH2O. For a complete
listing of all scales and resolutions available for each
pressure transducer, please consult the Comark Customer Service Department.
Because pressure scaling is a function of the transducer connected, the instrument automatically detects
the transducer type at switch-on, or if a transducer is
connected when the instrument is already switched
on. This is known as ‘hot swapping’ when the full scale
bar pressure is displayed for a period of 1.5 seconds to
conrm proper detection of the new transducer.
The error message -03- is displayed either if the selected scale is not available with the attached transducer or
if the valid transducer type was not detected. To clear
the message reconnect the transducer and/or press the
scale key.
12
Page 14
5.2 Calibration Scale
A ±500.0 mV scale for checking instrument calibration may
be reached by holding down the SCL key when switching ON.
This scale is cancelled by switching the instrument OFF. The
units displayed for this scale are simply shown as ‘m’. Details
of a suitable calibration lead may be obtained from the Comark
Customer Service Department.
A full 24 range, 12 scale tabular listing is available from
Comark as Engineering Note 18.
Page 16
6. Optional Equipment
Transducer interface cable, ADP22.
This is available from Comark Limited.
See Section 7 for the wiring of the interface cable.
Pressure transducers
These are not supplied with the instrument, and can
be ordered from the wide range available from Comark
Limited. For details of all ranges, 0.1 to 4000 bar refer
to the Transducer Range Data Sheet, Ref: C089 or contact Comark Limited. Each internal diaphragm trans-
ducer is supplied complete with one G1/2-W gasket.
Transducer interface cable, ADP22.
This is available from Comark Limited.
See Section 7 for the wiring of the interface cable.
Pressure transducers
These are not supplied with the instrument, and can
be ordered from the wide range available from Comark
Limited. For details of all ranges, 0.1 to 4000 bar refer
to the Transducer Range Data Sheet, Ref: C089 or contact Comark Limited. Each internal diaphragm transducer is supplied complete with one G1/2-W gasket.
Adaptor G½A female to to G¼A male, ADP23
This is supplied complete with one G1/4-W gasket.
Comark do not supply any other adaptors. If you require
adaptors other than ADP23, these should be sourced
locally.
Transducer gaskets
These are required in order to obtain a seal. Copper
gaskets may be used up to 3 times, O-ring gaskets may
be used repeatedly, and comprise:
G1/2-W Copper gasket for G½A
G1/4-W Copper gasket for G¼A
G1/2-OW O-ring gasket for G½A.
15
Page 17
7. WIRING OF INTERFACE CABLE
The standard interface cable has a length of 1.5 metres which
is adequate for normal use. For greater lengths two cables may
be joined end to end as the plug on one can be connected to
the socket on the next. Alternatively cables up to 5 metres may
be made by the user.
Wiring of interface cable
Plug Socket Wire Signal
Pin Pin Colour Function
1 1 Black Data I/O
2 2 Brown Serial Clock
3 3 Red 4 4 Orange Chip Select
5 5 Yellow + supply
6 6 Green + signal
7 7 Blue - supply
8 8 Purple - signal
16
Page 18
8. INSTRUMENT SPECIFICATION
CALIBRATED ACCURACY ±0.06% of reading± 1 digit
at +23°C
TEMPERATURE <0.01% of reading per °C
COEFFICIENT change from 23°C
OF READING
RESPONSE TIME 4 readings per second
OPERATING RANGE -20 to +50°C
ENVIRONMENTAL RATING IP67
POWER SOURCE 9V PP3 6F22 - MN1604 UCAR9V
BATTERY LIFE 140 hours (using MN1604)
LOW BAT CHECK Displays BAT just before
end point of battery
DISCONNECTED TRANSDUCER - 0 0 -
INDICATION
OVER PRESSURE - 0 2 -
INDICATION
OVER 9999 Lowest 4 digits blink
READING INDICATION