Comark 3392-PG Installation Manual

Page 1
C9550 EXTERNAL
PRESSURE METER
OPERATING MANUAL
1
Page 2
CONTENTS
CALIBRATION, CERTIFICATION
1.1 Transducer Input . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2 Transducer Operator Manual .................6
1.3 Transducer Range Data Sheet ...............6
1.4 Ranges, scales and resolutions ...............6
2.1 ON/OFF Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2 ZERO Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.3 SCALE Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.4 HOLD Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.5 FILTER Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.6 RECORD Key . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.7 AUTO Switch Off . . . . . . . . . . . . . . . . . . . . . . . . . 8
4 OTHER FUNCTIONS .................... 10
4.1 Battery Low . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.2 Over 9999 Reading Indication ...............10
4.3 Absolute or Gauge Measurement . . . . . . . . . . . . 10
4.4 Error Messages ..........................10
4.5 Battery Replacement ......................11
5 UNITS OF MEASUREMENT .............. 12
5.1 The C9550 Pressure Scaling System . . . . . . . . . 12
5.2 Calibration Scale . . . . . . . . . . . . . . . . . . . . . . . . . 13
Table 3. Scale Conversion Factors ...............13
5.3 Table of Full Scale Pressures
and Resolutions . . . . . . . . . . . . . . . . . . . . . . . . . 14
1
Page 3
6 OPTIONAL EQUIPMENT . . . . . . . . . . . . . . . . 15
7. WIRING OF INTERFACE CABLE .......... 16
8. INSTRUMENT SPECIFICATION ........... 17
Declaration of Conformity . . . . . . . . . . . . . . . . . . 18
FIGURE 1 View of the measurement system . . . . . . . . 19
2
Page 4
GENERAL INFORMATION
TRANSDUCERS SHOULD NOT BE DISCONNECTED FROM THE
WARNING
PRESSURE SOURCE WHILST CONTAINING GASES WHICH ARE DANGEROUS AND/OR UNDER PRESSURE.
DO NOT APPLY PRESSURES BEYOND THE OVER RANGE MAXI­MUM FOR THE TRANSDUCER OR DAMAGE WILL OCCUR.
REFER TO SECTION 3 - CONNECTION TO PRESSURE SOURCES BEFORE OPERATING THE INSTRUMENT.
TO AVOID ELECTRIC SHOCK DO NOT ALLOW ANY TRANSDUCER OR CABLE TO COME INTO CONTACT WITH LIVE ELECTRICAL POWER CONDUCTORS WITH VOLTAGES IN EXCESS OF 30V AC RMS OR 60V DC.
CAUTION Pressure Transducers This precision instrument has been designed for use with the extensive range of Comark pressure transducers. The use of other transducers may impair the performance and accuracy of the instrument. Full details of Comark transducers may be obtained from Comark Customer Support department or your local distribu­tor.
Repeated sharp exing can break transducer connection cables.
To prolong lead life, avoid sharp bends or kinks in the leads, espe­cially near the connectors.
This instrument is manufactured in accordance with the company’s ISO 9001 Quality Approved System.
CERTIFICATE No. FM29700
This instrument complies with the Electromagnetic Compatibility Directive EN61326-1:2006.
Declarations of Conformity available. Contact Comark Customer Support or your local Distributor.
In line with its policy of continuous development, Comark Limited reserves the
right to alter the instrument specication without prior notice. Further information
is available from Comark Limited or your distributor.
3
Page 5
CALIBRATION, CERTIFICATION AND SERVICE
Certication
Comark can provide certicates of calibration for its whole product
range, to suit ISO 9000 and other quality assurance procedures, food hygiene regulations, HACCPs and environmental regulations. Comark
certicates are produced by independent quality controlled processes
which compare product performance against agreed National Stand­ards. For peace of mind and best practice Comark recommend annual
certication/recalibration.
Two levels of certication are available for infra-red temperature and non temperature instruments, excluding humidity:- UKAS certicates via an external accredited laboratory and NPL traceable certicates from
the Comark calibration laboratory.
Three levels of certication are available for contact temperature and
humidity products and these are detailed here:
a) UKAS Temperature Certication
The Comark UKAS (United Kingdom Accreditation Service)
accredited temperature calibration laboratory is one of the n­est in the UK. Comark UKAS certicates can offer the lowest
uncertainty of 0.01°C and provide independent proof of correct calibration using equipment and procedures audited by UKAS inspectors. The equipment used is fully traceable to the Na­tional Physical Laboratory.
b) UKAS Humidity Certication
In addition to the Comark temperature laboratory, the humidity
laboratory continues the tradition of high accuracy certication
and a wide range (25% to 90%RH) with uncertainties of 2.8% of reading. This range also includes Dew point measurements.
c) NPL Traceable Certication
Comark NPL certicates are traceable to the National Physical
Laboratory and can offer uncertainty as low as 0.3°C.
4
Page 6
Conformance
Certicates of conformance can be supplied for new, serviced and recalibrated instruments. These conrm that instruments are within their original manufactured specication.
Service/Repairs
Regular servicing and any required repairs, under warranty or after, are available from the Comark Service Department.
For more information on all Comark certication, calibration and service
facilities please call Comark Customer Support or contact your local distributor.
5
Page 7
1 INTRODUCTION
1.1 Transducer Input
The C9550 external pressure meter accepts all Comark pressure transducers model 3392, 3393, 3397 or 3398. Connection is made by means of the 8-pin socket and the 1.5 metre interface cable. An inbuilt serial interface allows the C9550 to communicate with the transducer and provide correct display of the pressure value in any one of the pressure units selected. Suitable pressure transducers are:
Model Pressure range, bar Diaphragm Type
3392/PG Series 0.1 to 4000 Internal Gauge 3393/PA Series 0.25 to 16 Internal Absolute 3397/PGF Series 0.1 to 600 Front Flush Gauge 3398/PAF Series 0.25 to 16 Front Flush Absolute
Table 1. Transducer Model References
1.2 Transducer Operator Manual
The separate Transducer Operator Manual provides details of the installation, troubleshooting and calibra­tion of transducers.
1.3 Transducer Range Data Sheet
Pressure ranges extend from 0.1 to 4000 bar. The separate Transducer Range Data sheet, Ref: C089, lists all pressure ranges available in each of Gauge / Absolute, Internal Diaphragm / Front Flush Diaphragm,
0.5% / 0.25% accuracy types.
1.4 Ranges, scales and resolutions
A tabular listing of the scales and resolutions of Comark stocked ranges is provided in section 5.3. The com­plete listing for all 24 ranges is available separately as Engineering Note 18, and is available from the Cus­tomer Service Department.
6
Page 8
2 KEY FUNCTIONS
2.1 ON/OFF Key
Press once to switch on the instrument. Repeat to switch off.
Holding down the key when switching on will display the maximum pressure range for the attached transducer in Bar. When the key is released the previously set pres­sure units will be employed.
Holding down the key when switching off will display all
segments as a test.
2.2 ZERO Key
Press to zero the current readings on the display to nul­lify any offsets caused by large temperature variations, or to provide a tare function. The instrument discards any such offset when switched off, and will apply no off­set when switched back on. The zero key can be used whenever offsets occur. Please note that pressing the zero key has no effect whilst the instrument is in hold or record modes.
Note that if the zero function is used to set to zero (tare) a standing pressure reading, the -02- overpres­sure warning will continue to occur at 110% of full scale transducer pressure, although the numerical value dis­played will not then be 110% of full scale. This ensures that the display warning for overpressure will apply to the safety of the transducer.
2.3 SCALE Key
Press to display the pressure measurement in each of the different scales available. The selected scale unit is retained in memory even after switch off, so that the selected unit can be displayed when the instrument is switched on again.
See Section 5 for a fuller explanation of the C9550 pressure scaling system.
7
Page 9
2.4 HOLD Key
Press to hold the displayed reading and press again to resume measurement. The scale key will operate while the instrument is in hold mode so that the held reading can be displayed in different scales.
2.5 FILTER Key
Press to lter out transient readings. In lter mode the
instrument will average the readings taken over the last 4 seconds (16 readings) to provide a more stable reading on the display. Press the key again to resume normal measurement.
2.6 RECORD Key
Press once to put the instrument into record mode. This will be indicated by the word RECORD on the dis­play. Maximum and minimum pressure readings can be stored and displayed while the instrument is in record mode. When the instrument is in record mode, press key once to display the current maximum reading and press again to display the current minimum reading. Pressing the key a third time will return the display to normal measurement.
Record mode can be cancelled and the max/min read­ings cleared by pressing and holding down the key until the word RECORD disappears from the display.
The scale key will operate while the instrument is in record mode so that the max/min readings can be displayed in different scales.
2.7 AUTO Switch Off
The instrument can be programmed to auto switch off by pressing and holding down the HOLD key while using the ON/OFF key to switch on the instrument. The word AUTO appears on the display to indicate auto switch off mode and the instrument will switch itself off approximately 3 minutes after the last use of the keypad. Auto switch off mode can be cancelled by pressing and holding down the HOLD key while switching on manually with the ON/OFF key.
8
Page 10
3 PRESSURE SOURCE CONNECTION
Transducers for use with the C9550 instruments are tted with
G½A male connectors The following checks must be made before connecting them to a pressure source:-
•See warnings on page 2.
•Refer to the separate Transducer Operator
Manual.
•Please ensure that you check your own safety
standards before connecting to a pressure source.
•Check the security of all ttings.
•Do not undo pressure connections until the
pressure is released rst.
•Check the pressure rating on the transducer label.
•Check that the pressure source does not exceed
the ratings for the transducer as set out in Table 2.
•Check that the gases to be measured will not
damage or corrode the stainless steels in the transducer.
Transducer Maximum Pressure Ranges Overpressure
0.1 to 16 bar 350% of Full Scale 25 to 600 bar 200% of Full Scale 1000 bar 150% of Full Scale
1600 to 4000 bar 120% of Full Scale
Table 2. Overpressure limits for ranges
9
Page 11
4 OTHER FUNCTIONS
4.1 Battery Low
Low battery is indicated by the symbol BAT on the display. Replace the battery as soon as possible after this symbol appears.
4.2 Over 9999 Reading Indication
If the numeric value for the display exceeds ±9999, with or without a decimal point, the display will blink and suppress the top (5th) digit, whilst continuing with cor­rect values of the lower 4 digits.
4.3 Absolute or Gauge Measurement
C9550 Series instruments can be used with either gauge or absolute pressure transducers; no adjustment is required between use of either type. Refer to the transducer label to determine the measurement type. As the transducer only has a single port a direct dif­ferential measurement is not possible; in this case the zero function may be used between measurement from two transducers.
4.4 Error Messages
The following error messages may appear on the dis­play to indicate incorrect operation or malfunction:
-00- Sensor not connected
-01- Overow of A/D converter
-02- Overpressure (110% of Full Scale pressure)
-03- Mode not feasible SCL Invalid Scale Units
10
• Error -00- indicates no sensor is connected.
• Error -01- occurs if connected transducer
signal is too high (for example because of a transducer defect or a pressure overload).
• Error -02- indicates an attempt to measure
a pressure (either positive or negative) that exceeds the range of the transducer by more than 10%. If this occurs the pressure must be reduced to avoid damage to the sensor in the transducer.
Page 12
• Error -03- indicates either
a) an attempt to attach an incompatible sensor type. b) an interrupted data transmission between sensor and instrument.
• Error SCL - indicates the scale units are
not available on this transducer.
In the case of b) either reconnect the transducer or switch OFF then ON again. In the case of c) press the SCL key for the next supported scale units.
4.5 Battery Replacement
Unscrew the two screws at the base of the instrument and slide off the battery cover. Replace the battery with a 9V PP3 type battery. Replace the battery cover tak­ing care not to over tighten the two screws.
11
Page 13
5. UNITS OF MEASUREMENT
5.1 The C9550 Pressure Scaling System
Units other than the SI-units bar will be calculated by the C9550 using internationally agreed conversion fac­tors. Once powered up the instrument will display val­ues in the same scale as last selected. One or multiple depressions of the SCALE key will change the display to read other units of measurement.
For a given transducer, only those pressure scales with full scale numeric values less than 11,000 are avail­able for selection; thus the lower pressure transducers have progressively more scales, to a maximum of 12 for the 0.1 bar transducer. The full display sequence of pressure scales is Kgcm-2, mBar, Bar, inHg, Torr, inH2O, kPa, Pa, mmHg, PSI, mmH2O, mH2O. For a complete listing of all scales and resolutions available for each pressure transducer, please consult the Comark Cus­tomer Service Department.
Because pressure scaling is a function of the trans­ducer connected, the instrument automatically detects the transducer type at switch-on, or if a transducer is connected when the instrument is already switched on. This is known as ‘hot swapping’ when the full scale bar pressure is displayed for a period of 1.5 seconds to
conrm proper detection of the new transducer.
The error message -03- is displayed either if the select­ed scale is not available with the attached transducer or if the valid transducer type was not detected. To clear the message reconnect the transducer and/or press the scale key.
12
Page 14
5.2 Calibration Scale
A ±500.0 mV scale for checking instrument calibration may be reached by holding down the SCL key when switching ON. This scale is cancelled by switching the instrument OFF. The units displayed for this scale are simply shown as ‘m’. Details of a suitable calibration lead may be obtained from the Comark Customer Service Department.
1 bar = Unit Conversion Factor
mbar 1000
inHg 29.530 Torr 750.06 inH2O 401.46 kPa 100 Pa 100000 mmHg 750.06 psi 14.5038 mmH2O 10197.2 mH2O 10.1972 kgcm-2 1.01972
Table 3. Scale Conversion Factors from 1 bar.
13
Page 15
5.3 Table of Full Scale Pressures and Resolutions
Scale FS,bar 2.500 6.000 9.999 9.99 400.0
mBar Range 2500 6000 10000 NS NS
Resolution 1 1 1
Bar Range 2.500 6.000 10.000 100.00 400.0
Resolution 0.001 0.001 0.001 0.01 0.01
inHg Range 73.82 177.2 295.3 2953 NS
Resolution 0.01 0.1 0.1 1
Torr Range 1875 4500 7501 NS NS
Resolution 1 1 1
inH2O Range 1003.7 2409 4015 NS NS
Resolution 0.1 1 1
kPa Range 250.0 600.0 1000.0 10000 NS
Resolution 0.1 0.1 0.1 1
Pa Range NS NS NS NS NS
Resolution
mmHg Range 1875 4500 7501 NS NS
Resolution 1 1 1
PSI Range 36.26 87.02 145.0 145.0 5802
Resolution 0.01 0.01 0.1 1 1
mmH2O Range NS NS NS NS NS
Resolution
mH2O Range 25.49 61.18 101.97 1019.7 4079
Resolution 0.01 0.01 0.01 0.1 1
-2
Kgcm
Resolution 0.001 0.001 0.001 0.01 0.1
Range 2.549 6.118 10.197 101.97 407.9
Table 4. Full scale pressures and resolutions
14
Notes about table 4:
• The bold FS, bar value listed appears on
display at power-on when the ON button is held.
• Key: NS = No Scale.
• Underline means this digit is absent, and
the remaining digits blink.
A full 24 range, 12 scale tabular listing is available from Comark as Engineering Note 18.
Page 16
6. Optional Equipment
Transducer interface cable, ADP22.
This is available from Comark Limited. See Section 7 for the wiring of the interface cable.
Pressure transducers
These are not supplied with the instrument, and can be ordered from the wide range available from Comark Limited. For details of all ranges, 0.1 to 4000 bar refer to the Transducer Range Data Sheet, Ref: C089 or con­tact Comark Limited. Each internal diaphragm trans-
ducer is supplied complete with one G1/2-W gasket.
Transducer interface cable, ADP22.
This is available from Comark Limited. See Section 7 for the wiring of the interface cable.
Pressure transducers
These are not supplied with the instrument, and can be ordered from the wide range available from Comark Limited. For details of all ranges, 0.1 to 4000 bar refer to the Transducer Range Data Sheet, Ref: C089 or con­tact Comark Limited. Each internal diaphragm trans­ducer is supplied complete with one G1/2-W gasket.
Adaptor G½A female to to G¼A male, ADP23
This is supplied complete with one G1/4-W gasket. Comark do not supply any other adaptors. If you require adaptors other than ADP23, these should be sourced locally.
Transducer gaskets
These are required in order to obtain a seal. Copper gaskets may be used up to 3 times, O-ring gaskets may be used repeatedly, and comprise: G1/2-W Copper gasket for G½A G1/4-W Copper gasket for G¼A G1/2-OW O-ring gasket for G½A.
15
Page 17
7. WIRING OF INTERFACE CABLE
The standard interface cable has a length of 1.5 metres which is adequate for normal use. For greater lengths two cables may be joined end to end as the plug on one can be connected to the socket on the next. Alternatively cables up to 5 metres may be made by the user.
Wiring of interface cable
Plug Socket Wire Signal Pin Pin Colour Function
1 1 Black Data I/O 2 2 Brown Serial Clock 3 3 Red ­4 4 Orange Chip Select 5 5 Yellow + supply 6 6 Green + signal 7 7 Blue - supply 8 8 Purple - signal
16
Page 18
8. INSTRUMENT SPECIFICATION
CALIBRATED ACCURACY ±0.06% of reading± 1 digit at +23°C
TEMPERATURE <0.01% of reading per °C COEFFICIENT change from 23°C OF READING
RESPONSE TIME 4 readings per second
OPERATING RANGE -20 to +50°C
ENVIRONMENTAL RATING IP67
POWER SOURCE 9V PP3 6F22 - MN1604 UCAR9V
BATTERY LIFE 140 hours (using MN1604)
LOW BAT CHECK Displays BAT just before
end point of battery
DISCONNECTED TRANSDUCER - 0 0 - INDICATION
OVER PRESSURE - 0 2 - INDICATION
OVER 9999 Lowest 4 digits blink READING INDICATION
CONNECTIONS 8 way Lumberg socket
KFV81 Mating plug, Lumberg SV81
OVERALL DIMENSIONS Height 173mm, Width 70/83mm, Depth 32mm
WEIGHT 255 gm including battery
17
Page 19
Declaration of Conformity
We:
Comark Limited Bury Mead Road Hitchin Herts SG5 1RT
Declare that the product:
Comark C9550
Is in conformity with the requirements of the follow­ing documents (Directives):
EMC Directive 2004/108/EEC
Standards: EN61326-1:2006
Signed on behalf of Comark Limited
at:
David French Quality Manager
18
Bury Mead Road Hitchin Herts SG5 1RT
Page 20
Fig. 1 View of the complete measurement system.
19
Page 21
20
Page 22
Comark Limited 52 Hurricane Way Norwich, Norfolk, NR6 6JB England Tel: 01603 (+44 1603) 256647 Fax: 01603 (+44 1603) 256744 Email: service@comarkltd.com
Website: www.comarkltd.com
Comark Instruments PO Box 9090, Everett, WA 98206, USA Tel: (503) 643 5204 Fax: (503) 644 5859 Email: sales@comarkUSA.com
Website: www.comarkUSA.com
© Comark Limited Stock No. 16986 Issue 8 11/09