The Intelli-Tank 4 light tank level is designed to display a liquid’s volume to an
eighth of a tank level accuracy through 180-degree viewable ultra-bright LEDs.
The unit set as a Master uses a 0 – 5 PSI pressure transducer to obtain tank
level information and then relays that information along the data line to units set
as Remotes. Multiple Remote units can be linked to the Master tank level unit.
Modes of Operation:
MasterWhen the unit is calibrated with a proper pressure signal it
automatically becomes a Master Unit and will send tank level
information along the 1-wire data transfer line to all other Remote
units on the line.
RemoteUnits are initially shipped as Remote units. A Remote unit only
requires power, ground and the 1-wire data transfer line. The
Remote unit mimics the Master unit’s display by reading the
appropriate information on the 1-wire data transfer line.
Self TestWhen the self test password is entered the unit will perform a self
test and display the results on the display. A Master unit will
command all Remote units to perform the self test also.
Note: All units are shipped as Remote units. A valid calibration will set the unit
as a Master. A Master can be reset as a Remote by entering a password.
Intelli-Tank Manual (106759) 021502– 2 –Class 1
When the unit is first powered up the LEDs will cycle on individually starting with
the EMPTY (bottom) LED and then the LEDs will show current status.
• A Master unit properly connected to a functioning transducer will display
current tank level information.
• A Master unit not connected to a pressure transducer will alternately flash the
bottom two LEDs.
• A Remote unit connected to a Master unit (through the 1-wire data line) will
mimic the Master unit’s LED condition and flash rate.
• A Remote unit not connected to a Master unit will alternately flash the upper
two LEDs and the lower two LEDs. This indicates a “no data” condition.
Level indications:
Intelli-Tank Manual (106759) 021502– 3 –Class 1
Miscellaneous indications:
ConditionLED 4LED 3LED 2LED 1
Invalid calibrationAlternaFLASHAlternaFLASH
EEPROM errorAlternaFLASH AlternaFLASH
Signal voltage above
4.8V
Signal voltage below
.4V
Remote “NO DATA”AlternaFLASHAlternaFLASH
User Error (wave off)ONON
4 quick cyclesONON
The tank level has two flash rates: FAST (1.6Hz) and SLOW (.8Hz). AlternaFLASH flashes two lights
alternately at the SLOW flash rate.
“Wave off” pattern : two center LEDs, two outer LEDs flashing quickly for 4 cycles.
AlternaFLASH AlternaFLASH
AlternaFLASH AlternaFLASH
Intelli-Tank Manual (106759) 021502– 4 –Class 1
Intelli-Tank displayThe display requires a cutout as shown. The unit is water tight and may
be mounted in any location on the operator’s panel.
Outer bezel dimensions.
Intelli-Tank Manual (106759) 021502– 5 –Class 1
Before mounting the display and adhering the label insure that the unit is situated correctly (TOP
is UP). Refer to the drawing for orientation.
Pressure Transducer
The transducer has a ¼” NPT mount and must be mounted vertically as depicted to insure an
accurate reading.
Intelli-Tank Manual (106759) 021502– 6 –Class 1
Power and Ground
It is imperative that a system utilizing Master and Remote tank level units connected by the 1-wire
data line have a common ground.
Pin 1Vehicle power
Pin 2Ground
Dim Function
The LEDs on the tank level unit can be dimmed to a user selectable dim setting by applying
vehicle voltage to the Dim display Input.
Pin 3Dim display input (VIGN)
Transducer Connection
Pin 6Sensor power (+5)
Pin 7Sensor signal
Pin 8Sensor ground
1-Wire Data Line
Pin 4Master to Remote data transfer line
Intelli-Tank Manual (106759) 021502– 7 –Class 1
Typical installation layout.
Intelli-Tank Manual (106759) 021502– 8 –Class 1
The Intelli-Tank display can be calibrated three different ways: 2-point (level
calibration), 5-point and 9-point (volume calibration).
To enter calibration mode use a magnet and activate the magnetic switches in
the order of the appropriate password.
When a magnetic switch is activated the display will indicate which switch was activated
(Left = upper two LEDs, Right = lower two LEDs) for approximately half a second and
then the display will go blank. There is a limit of 2 seconds between switch activation’s
before the the unit reverts to normal operation.
Entering an invalid password will initiate a “wave off” pattern on the display. (Two center
LEDs, two outer LEDs flashing quickly for 4 cycles.)
Calibrate the unit by entering the desired point calibration –
2 point RLLR LLRL
5 point RLLR LRLR
9 point RLLR RLLR
Intelli-Tank Manual (106759) 021502– 9 –Class 1
The unit will respond by flashing the two center LEDs the same number of times
as the desired calibration. The unit will then indicate the level to be calibrated by
lighting the proper LEDs for that level starting with empty (see page 3, tank levelindications). When the tank is filled to the proper level for calibration (dictated by
the LEDs) activate the RIGHT switch to store that point. The unit will flash the
top LED and then set the LEDs for the next calibration point. Continue filling the
tank to the proper level dictated by the LEDs and storing that point with the
RIGHT switch until finished (after the FULL point has been calibrated). The unit
will respond to a completed calibration procedure by lighting the two center LEDs
and then reverting to normal operation by displaying the proper tank level, which
at this point should indicate FULL (all LEDs on) because the tank level is full.
To get out of calibration mode at any time without saving, activate the LEFT
switch.
Calibration data is saved in non-volatile EEPROM so the unit does not need
power to retain calibration data.
Calibration automatically makes the unit a master if the calibration is valid. An
invalid calibration is determined if any point is not at a higher level than the
previous, or if the transducer voltage falls outside of the valid minimum (.4V) or
maximum (4.8V) range. An invalid calibration is acknowledged by giving the
“wave off” and if this was a master unit previously will alterna-flash the outer two
LEDs, while a Remote will revert to Remote operation.
Calibration in a nutshell.
1. Enter the password for the desired calibration (wait for center LEDs to
flash back the calibration number).
2. Ensure tank level is at proper level for calibration point as indicated by the
LEDs.
3. Activate the RIGHT switch to store point (Top LED flashes).
4. Continue steps 2 and 3 until finished (Full has been calibrated).
Intelli-Tank Manual (106759) 021502– 10 –Class 1
Dimming the Display
The display can be dimmed by applying VIGN to pin 3 (Dim Display input). To
select the dim level of the display use the magnetic switches to enter the
password
RLLR LLLR
all of the LEDs will come on. Hold the magnet against the RIGHT switch and the
display will either brighten or dim. Release the magnet and again hold it against
the RIGHT switch and the display’s brightness will move in the opposite
direction. When the dim level is at the desired point activate the LEFT switch.
Each unit has its own calibrated dim level and dim input.
Self Testing the Unit
The Tank Level can check its hardware for proper operation by entering the
password
RLLR LLRR
the unit will then cycle each LED on individually starting with the bottom LED and
then all LEDs will come on and begin flashing between full bright and the
calibrated dim level for 5 seconds. The display will then show the condition of
the self test for 5 seconds.
A PASS condition is indicated when only the top LED (LED 4) is on.
A FAIL condition exists if LED 4 is off and any other LED is on.
LED 3 ONMemory failure.
LED 2 ONData line error.
LED 1 ONTransducer signal line out of tolerance or shorted to
+5V.
LED 1 FlashingTransducer signal shorted to ground (too low).
If the self test password is used on a Master unit all Remote units will also
perform their self test.
Intelli-Tank Manual (106759) 021502– 11 –Class 1
RLLR LLRL2 point calibration
RLLR LRLR5 point calibration
RLLR RLLR9 point calibration
RLLR LLRRSelf test
RLLR LLLLConfigure unit as Remote unit
RLLR LLLRConfigure dim level
Software revision check
Hold a magnet on the LEFT magnetic switch while powering the unit. The LEDs
will display the software revision. (Use the chart below to decipher).
Example. (LED 4 – OFF, LED 3 – ON, LED 2 – OFF, LED 1 – ON) = Ver 1.1
Intelli-Tank Manual (106759) 021502– 12 –Class 1
ConditionEvaluate
Bottom two LEDs
alternate flashing.
Check transducer wiring. Ensure +5V at pin A, ground at pin B and
at least .4V at pin C (Signal).
Unit fails self test,
LED 1 flashing.
Top two LEDs
alternate flashing.
Check transducer wiring. Ensure +5V at pin A, ground at pin B and
no more than 4.8V at pin C (Signal).
Unit fails self test,
LED 1 on.
Middle two LEDs
Perform self test. If it fails with LED 3 on replace unit.
alternate flashing.
Outer two LEDs
alternate flashing.
Try to recalibrate. If condition remains, check if transducer signal
voltage (pin C) changes as tank level increases. If it doesn’t, replace
transducer. If it does, verify depth of tank. It may be impossible to
calibrate a tank with a depth of less than 6 inches.
Bottom two and
Upper two LEDs
alternate flashing.
The unit is configured as a Remote. If it required to be a Master
then recalibrate. If the unit is required to be a Remote check Data
line (Pin 4) continuity and insure line is not grounded.
No LEDS on.Check power (Pin 1) and ground (Pin 2) connection.
Master Tank level
display does not
change when actual
tank level is
changing.
Remote Tank level
display does not
follow Master
display.
Unit fails self test,
LED 2 on.
No passwords are
accepted.
Check transducer wiring. Ensure transducer signal voltage (Pin C)
is varying. If it does, check for same signal changes at Pin 6 of tank
level connector (if it is not the same repair wiring). If signal is good
at both locations try re-calibrating.
Perform self test. If self test is good, check pin 4 (data line) for
continuity and insure it is not shorted to ground or power. Insure
data line is not routed near noisy power or RF sources.
Check pin 4 (data line) for continuity and insure it is not shorted to
ground or power.
If the display issues the “wave off” after entering a password, insure
the unit is installed upright. During power-up the display should
cycle on each LED individually starting with the bottom LED.
Check that the left and right magnetic switches are recognized by
activating each switch and verifying that the associated LEDs
illuminate.
Intelli-Tank Manual (106759) 021502– 13 –Class 1
ConditionEvaluate
The bottom two
LEDs are on and
occasionally they go
out and the top two
flash and then return
to the bottom two
LEDs on (or viceversa). (REMOTE).
The points
calibrated seemed
to have changed.
Unit will not dim
display.
Check for large noise spikes on the data line.
Insure that the unit’s ground potential is the same as the Master’s.
Insure that the data line is not chaffed and making contact with
other electrical wires.
Self test the unit to check for any malfunctions.
Check the pressure transducer for problems.
Recalibrate the unit and take a voltage reading from the transducer
(pin 7 on the unit’s connector) at each calibration point. When the
calibration points again look wrong check the voltages at those
points and determine if they are the same as the voltage reading
taken during calibration.
Insure the Dim input voltage on Pin 3 is at least 9V. Recalibrate dim
setting (RLLR LLLR). If unit does not dim LEDs while in dim
calibrate mode, replace unit.
Intelli-Tank Manual (106759) 021502– 14 –Class 1
Intelli-Tank Manual (106759) 021502– 15 –Class 1
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.