The information in this document is subject to change without notice.
Chroma ATE INC. makes no warranty of any kind with regard to this manual, including, but not
limited to, the implied warranties of merchantability and fitness for a particular purpose.
Chroma ATE INC. shall not be held liable for errors contained herein or direct, indirect, special,
incidental or consequential damages in connection with the furnishing, performance, or use of
this material.
CHROMA ATE INC.
66 Huaya 1st Road, Guishan, Taoyuan 33383, Taiwan
Copyright Notices. Copyright 2014 - 2015 Chroma ATE INC., all rights reserved. Reproduction,
adaptation, or translation of this document without prior written permission is prohibited,
except as allowed under the copyright laws.
ii
Warranty
All of Chroma’s instruments are warranted against defects in material and workmanship for a
period of one year from date of shipment. Chroma agrees to repair or replace any assembly
or component found to be defective, under normal use during this period. Chroma’s obligation
under this warranty is limited solely to repairing any such instrument, which in Chroma’s sole
opinion proves to be defective within the scope of the warranty when returned to the factory or
to an authorized service center. Purchaser is responsible for the shipping and cost of the
service item to Chroma factory or service center. Shipment should not be made without prior
authorization by Chroma.
This warranty does not apply to any products repaired or altered by persons not authorized by
Chroma, or not in accordance with instructions furnished by Chroma. If the instrument is
defective as a result of misuse, improper repair, or abnormal conditions or operations, repairs
will be billed at cost.
Chroma assumes no responsibility for its product being used in a hazardous or dangerous
manner either alone or in conjunction with other equipment. High voltage used in some
instruments may be dangerous if misused. Special disclaimers apply to these instruments.
Chroma assumes no liability for secondary charges or consequential damages and in any
event, Chroma’s liability for breach of warranty under any contract or otherwise, shall not
exceed the purchase price of the specific instrument shipped and against which a claim is
made.
Any recommendations made by Chroma regarding the use of its products are based upon
tests believed to be reliable; Chroma makes no warranty of the results to be obtained. This
warranty is in lieu of all other warranties, expressed or implied, and no representative or
person is authorized to represent or assume for Chroma any liability in connection with the
sale of our products other than set forth herein.
“O” indicates that the level of the specified chemical substance is less than the threshold level
specified in the standards of SJ/T-11363-2006 and EU 2005/618/EC.
“” indicates that the level of the specified chemical substance exceeds the threshold level
specified in the standards of SJ/T-11363-2006 and EU 2005/618/EC.
The recycling label shown on the product indicates the Hazardous Substances contained in
the product as the table listed below.
: See <Table 1>.
: See <Table 2>.
<Table 1>
Disposal
Do not dispose of electrical appliances as unsorted municipal waste; use separate collection
facilities. Contact your local government for information regarding the collection systems
available. If electrical appliances are disposed of in landfills or dumps, hazardous substances
can leak into the groundwater and get into the food chain, damaging your health and
well-being. When replacing old appliances with a new one, the retailer is legally obligated to
take back your old appliances for disposal free of charge.
iv
<Table 2>
Part Name
Hazardous Substances
Lead
Mercury
Cadmium
Hexavalent
Chromium
Polybrominated
Biphenyls
Polybromodiphenyl
Ethers
Pb
Hg
Cd
Cr6+
PBB
PBDE
PCBA
O O O O O
CHASSIS
O O O O O
ACCESSORY
O O O O O
PACKAGE
O O O O O
O
“O” indicates that the level of the specified chemical substance is less than the threshold level
specified in the standards of SJ/T-11363-2006 and EU 2005/618/EC.
“” indicates that the level of the specified chemical substance exceeds the threshold level
specified in the standards of SJ/T-11363-2006 and EU 2005/618/EC.
1. Chroma is not fully transitioned to lead-free solder assembly at this moment; however,
most of the components used are RoHS compliant.
2. The environment-friendly usage period of the product is assumed under the operating
environment specified in each product’s specification.
Disposal
Do not dispose of electrical appliances as unsorted municipal waste; use separate collection
facilities. Contact your local government for information regarding the collection systems
available. If electrical appliances are disposed of in landfills or dumps, hazardous substances
can leak into the groundwater and get into the food chain, damaging your health and
well-being. When replacing old appliances with a new one, the retailer is legally obligated to
take back your old appliances for disposal free of charge.
v
vi
Safety Summary
BEFORE APPLYING POWER
Verify that the power is set to match the rated input of this power supply.
PROTECTIVE GROUNDING
Make sure to connect the protective grounding to prevent an electric
shock before turning on the power.
NECESSITY OF PROTECTIVE GROUNDING
Never cut off the internal or external protective grounding wire, or
disconnect the wiring of protective grounding terminal. Doing so will
cause a potential shock hazard that may bring injury to a person.
FUSES
Only fuses with the required rated current, voltage, and specified type
(normal blow, time delay, etc.) should be used. Do not use repaired
fuses or short-circuited fuse holders. To do so could cause a shock or
fire hazard.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or
fumes. The instrument should be used in an environment of good
ventilation.
DO NOT REMOVE THE COVER OF THE INSTRUMENT
Operating personnel must not remove the cover of the instrument.
Component replacement and internal adjustment can be done only by
qualified service personnel.
The following general safety precautions must be observed during all phases of operation,
service, and repair of this instrument. Failure to comply with these precautions or specific
WARNINGS given elsewhere in this manual will violate safety standards of design,
manufacture, and intended use of the instrument. Chroma assumes no liability for the
customer’s failure to comply with these requirements.
vii
Safety Symbols
DANGER – High voltage.
Explanation: To avoid injury, death of personnel, or damage to
the instrument, the operator must refer to the explanation in the
instruction manual.
High temperature: This symbol indicates the temperature is
hazardous to human beings. Do not touch it to avoid any
personal injury.
Protective grounding terminal: This symbol indicates that the
terminal must be connected to ground before operation of the
equipment to protect against electrical shock in case of a fault.
Functional grounding: To identify an earth (ground) terminal in
cases where the protective ground is not explicitly stated. This
symbol indicates the power connector does not provide
grounding.
Frame or chassis: To identify a frame or chassis terminal.
Alternating Current (AC)
Direct Current (DC) / Alternating Current (AC)
Direct Current (DC)
Push-on/Push-off power switch
The WARNING sign highlights an essential operating or
maintenance procedure, practice, condition, statement, etc.,
which if not strictly observed, could result in injury to, or death of,
personnel or long term health hazards.
The CAUTION sign highlights an essential operating or
maintenance procedure, practice, condition, statement, etc.,
which if not strictly observed, could result in damage to, or
destruction of, equipment.
The Notice sign highlights an essential operating or maintenance
procedure, condition, or statement.
WARNING
CAUTION
viii
Revision History
Date
Version
Revised Sections
Oct. 2014
1.0
Complete this manual.
Mar. 2015
1.1
Add the functions for 11050-5M
The following lists the additions, deletions and modifications in this manual at each revision.
7.2.4.5 PARAmeter:SWEEP Command Set .............................................................. 7-40
7.2.4.6 DUAL:FREQuency Command Set ................................................................ 7-46
xii
Preface
11050-5M
11050
60Hz 5MHz
1kHz 10MHz
1. Preface
1.1 Overview
The 1105X Series is high frequency passive series instrument used for testing and analyzing
components. Take the 11050 of measurement frequency up to 10MHz as a basis, ”-5M”
after the model name indicates the maximum measurement frequency of the meter is 5MHz
and the rest series products may be deduced by analogy. The device was designed to solve
the problems of low labor efficiency and low product quality that have occurred since the
electronics industry began to flourish. It is able to increase the work efficiency and enhance
the product quality.
The test parameters equipped with this device containing L, C, R, Z, Y, DCR, etc., which are
perfect functions for the production line and quality control.
Via the instrument is capable of providing fast, highly accurate, convenient and reliable tests.
It has the functions of selection controls for test frequency and test voltage, LIMIT TEST, BIN
TEST, LOAD, LIST SWEEP, STORE/RECALL and RS232/LAN/GPIB. The handler interface
is able to communicate with automation machine.
The instrument provides DC bias (I0.02A) output for measuring coil inductance under bias
current source. It also can coordinate with Chroma’s DC bias source devices.
The multi-function test device, ergonomic keyboard design, guided panel operation,
extra-large LCD, locked key and password protection makes the 11050 instrument very easy
to operate. Its protection enables the test results to be shown on the display clearly.
If the instrument requires testing lead extension, use 4 connection points for connecting test
correctly. It needs to consider high frequency response of test wire as measuring under high
frequency.
GPIB interface (option), Handler interface, Bias & Bias Current control interface.
1-1
HF LCR Meter 1105X Series User’s Manual
RS-232
USB-Host
Memory
LAN & USB-H
GPIB
Handler
Bias
Bias
Current
11050-5M/
11050
Optional
Optional
Item
Q’ty
Description
Power Cord
1
A 1 meter bend power cord
Slow Blow Fuse 2.5A
2
For rated voltage AC 100 240V use
User’s Manual CD
1
CD for user’s manuals in English and Traditional
Chinese
D-SUB 15Pin link cable - 50CM
1
The cable is for linking the monitor and the
instrument.
Adapter
1
Power plug 3P to 2P
M4 x 10 pan head screw
5
Securing display unit and measurement unit.
Item
Q’ty
Description
Quick Start Guide
1
One Traditional Chinese version
A133509
1
GPIB card
A133510
1
LAN & USB-H card
D-SUB 15Pin link cable-2M
1
For hanging monitor connection use
RS232 link cable
1
RS-232 link cable
Cable CEN-36M/CEN-36M
150CM
1
Cable for connecting Handler interface
Test Cable-BNC+ Alligator Clip
1
BIAS cable
A110242
1
ESR test fixture for supporting 100kHz
A110234
1
High frequency test cable for supporting 1MHz
A133004
1
Four-terminal SMD test kit for supporting 1MHz.
A110211
1
Four-terminal DIP test kit for supporting 10MHz
A110501
1
Four-terminal SMD test kit for supporting 10MHz.
Note
Only name item is required when ordering any the accessories.
Table 1-1 Interface requirement
1.3 Checking before Use
Before shipment, this instrument was inspected and found to be free of mechanical and
electrical defects. As soon as the instrument is unpacked, inspect for any damage that may
have occurred in transit. Save all packing materials in case the instrument has to be
returned. If damage is found, please file claim with carrier immediately. Do not return the
instrument to Chroma without prior approval.
Upon receipt of the instrument, please check on the following items:
(1) If there are any damages or scratches on the product surface.
(2) The standard and optional accessories came with the instrument as listed in Table 1-2
and Table 1-3. .
Table 1-2 Standard Accessories
Table 1-3 Optional Accessories
1-2
2. Specifications
Primary Parameter
L C R
Z,Y
DCR
Secondary Parameter
D,Q,R,θ,Q+R
D,Q,R,θ
Q,X
θ
- - -
Range
Resolution
Unit
60.00 ~ 99.99
0.01
Hz
100.0 ~ 999.9
0.1
Hz
1.000 ~ 9.999
0.001
kHz
10.00 ~ 99.99
0.01
kHz
100.0 ~ 99.9
0.1
kHz
1.000 ~ 9.999
0.001
MHz
10.00
---
MHz
Test Voltage
Model
Frequency
Range
Resolution
Set Accuracy
10mVrms ~
999mVrms
11050-5M
60Hz~5MHz
1mV
± (10 + 1×fs) % ±1mV
11050
1kHz~10MHz
2.1 Measurement Function
Measurement Parameter:
Z : Impedance
L : Inductance
C : Capacitance
R : Resistance
Y : Admittance
X : Reactance
θ: Phase
Q : Quality Factor
D : Loss Factor
DCR : DC Resistance
Parameter Combination:
Specifications
Equivalent Circuit: Parallel and Series
Range: Auto and Manual
Trigger Mode: Internal, Manual and External
Measurement Terminals: Four-terminal measurement
Measurement Speed: Fast, Medium and Slow
Measurement accuracy is specified when all of the following conditions are satisfied:
(1) Warm-up time: 30 minutes
(2) Two sides of tester should keep ventilation at least 10cm.
(3) Single unit ambient measurement, open and short corrections have been performed.
(4) Single unit + long lead ambient measurement, open and short as well as load offsets have
been performed.
2.5 Accuracy
Within 1 year after factory calibrated
Temperature: 23ºC 5ºC
Relative Humidity: 75% maximum
2-2
Specifications
Note
Ze is the error corresponding to impedance which is represented by absolute
error, ex. Z error = 1%, Ze = 0.01.
Warm up: 30 minutes minimum
Test condition for accuracy specification: Tes voltage is 1Vrms and measurement speed
with slow.
1. Z/ Accuracy
The basic accuracy of 11050 is listed in Table 2-1.
The basic accuracy of 11050-5M is listed in Table 2-2.
Measurement with fast rate, accuracy is multiplied by 2.
Measurement with medium rate, accuracy is multiplied by 1.5.
Measurement with slow rate, accuracy is multiplied by 1.
Test voltage 1Vrms, accuracy is multiplied by 1.
Test voltage < 1Vrms, accuracy is multiplied by (1/test voltage).
2. L, C, R Accuracy
For Q 10 (loss factor 0.1), correspond to accuracy of Z, where
ZL=2fL
ZC=1/ (2fC)
Accuracy specification table of 11050: See Table 2-1 Z- Accuracy of 11050.
Accuracy specification table of 11050-5M: See Table 2-2 Z- Accuracy of 11050-5M.
If Q <10 (loss factor > 0.1), L accuracy is multiplied by (1+1/Q).
(1) Z stands for actual measurement impedance.
(2) Ze is the error corresponding to impedance which is represented by absolute
error, ex. Z error = 1%, Ze = 0.01.
10uH
100uH
1mH
10mH
100mH
1uH
10nH
100nH
L
1pF
10pF
100pF
1nF
10nF
100nF
1uF
10uF
100uF
1mF
10mF
100mF
100fF
10fF
1H
10H
100H
1kH
10kH
100kH
C
100mΩ
1MΩ
10MΩ
1kΩ
10kΩ
100kΩ
1Ω
10Ω
100Ω
|Z|
4. ESR Accuracy
ESR = Z×Ze (Q 10)
5. DCR Accuracy
100 kΩ < RDC ≤ 10 MΩ: 0.1%×(
10Ω < RDC ≤ 100 kΩ: 0.1%
1Ω < RDC ≤ 10Ω :( 0.2%+2mΩ)
RDC ≤ 1Ω :( 0.5%+0.2mΩ)
R
DC
100 k
Specifications
)
Figure 2-1 Impedance Planar Graph
2.6 Correction
Open Correction:
Remove measurement errors due to open stray impedances of test fixture.
Short Correction:
Remove measurement errors due to short residual impedances of test fixture.
Load Correction:
Place and load the known reference load, the system use the load value as standard.
2-5
HF LCR Meter 1105X Series User’s Manual
Open/Short/Load Corrections are recommended to use under the conditions below:
Connect DC bias measurement passive component additionally.
Place component test machine
Test fixture connected externally (user-defined included)
The condition contains long wire and fixture
Due to there are stray and residual impedance effects in fixtures connected externally, it is
suggested to perform Open/Short/Load Correction before measuring DUT. For operation
method, see 5.3.1.2 LCR MODE- OPEN /SHORT CORRECTION.
2.7 Interface
RS-232 interface : Standard RS-232 interface
USB-1 interface : Standard FLASH MEMORY interface
Handler interface : Output signal of test result and I/O interface of
external trigger
Bias interface : DC Bias BNC interface
DC Bias Current control interface : DC bias current control interface
USB-2 interface (option) :Bar Code Reader interface
LAN interface (option) :10/100 type network interface
GPIB interface (option) : Standard GPIB interface
2.8 Others
Display: Color 640x480 LCD Panel
Power: 100 ~ 240V± 10%, 47~63Hz, power consumption is 60VA, maximum.
Environment: Operating 0C ~ 40C, 10% ~ 90% relative humidity
Storage -10C ~ 70C, 10% ~ 90% relative humidity
Dimension: 11050 entire device installation 428mm (W) x 230mm (H) x 290mm (D)
Weight: 11050 entire device installation 8.0Kg approximately
4. Humidity: Maximum 80%RH at 31°C decreasing to 50%RH at 40°C
5. Pollution Degree: 2
3.1 Ambient Environment
(1) Do not use the meter in a dusty or vibrating
location. Do not expose it to sunlight or corrosive
gas. Be sure that the ambient temperature is 0 ~
40C and that the relative humidity is below 90%.
(2) Two sides of the meter is equipped with a cooling
fan to keep the internal temperature down, so
adequate ventilation should be ensured. The
meter should be located at least 10cm from any
object or wall behind it. Do not block the left and
right ventilation holes to keep the meter in good
precision.
(3) The meter has been carefully designed to reduce
the noise from the AC power source. However, it
should be used in a noise-free or as low as
possible environment. If noise is inevitable,
please install a power filter.
(4) The meter should be stored within the temperature
range -10ºC ~ +70ºC. In case it is not to be used
for a long time, please store it in the original or
similar package, and keep it from direct sunlight
and humidity to ensure a good condition for later
use.
Installation
3.2 Power Line Connection
Before plugging in the power cord, make sure the power switch is off. Please use the power
supply frequency of 50Hz or 60Hz.
3.3 Fuse
There is one fuse installed in the power inlet on the rear panel, please be aware of the
following when replacing it:
3-1
HF LCR Meter 1105X Series User’s Manual
1. First turn off the power and unplug the power cord before changing the fuse.
2. The specification of fuse: AT2.5A 250V
For safety and noise reduction, it is necessary to use a 3-pin power cord to connect the power
inlet on the rear panel for AC line input.
3.4 Power Regulation
As this instrument belongs precision electronical test device, so the accuracy is possible to be
influenced lower by input power unstable after testing. There is ±10% changeable power
even in the laboratory, so we suggest that use the regulator in power and test devices to be
the only one way to avoid power voltage affecting measured data.
3.5 Entire Device Installation & Allocation
The 11050-5M/11050 entire device installation includes Display Unit and Measurement Unit
(abbreviation: Meas. Unit, use 11050 illustration below as an example).
(1) Entire device installation diagram
Display unit and meas. unit are only fixed by monitor supporting frame when delivery.
User can remove pan head screw on monitor supoorting frame to separate display unit
and meas. unit by actual demand. Disassemble diagrams are as the following.
3-2
Installation
(2) Wiring inspection on rear of entire device
Display unit and meas. unit are only fixed by monitor supporting frame when delivery.
Users need to use standard accessory D-SUB 15Pin connecting wire 50CM to connect
REMOTE CONTROL terminals on the rear of display unit and meas. unit.
3-3
HF LCR Meter 1105X Series User’s Manual
Other wires installation are as the following: From left to right sequentially - AC power cord,
RS-232 cable (option), D-SUB 15Pin connecting wire 50CM, Cable CEN-36M/ CEN-36M
150CM (option) and test cable BNC + alligator clip (option).
(3) Allocate display unit at proximal to the user
Purchase D-SUB 15Pin -2M connecting wire additionally to move display unit out that its
allocation and meas. unit are as the figure below.
(4) Hanging Requirements
The separate display unit can hang on the location where monitor supporting frame locked
and it also can lock wall rack as the diagram below. The wall rack is locked in screw hole
with anti-baked coating.
3-4
Installation
The selection of wall rack is by users’ requirement. Screw hole locations on rear panel of
display unit as below shows, dimension unit is mm.
3.6 Power-on Procedure
After powering on the tester, the power indicator on 11050 MEASUREMENT UNIT front panel
will be lit and the inner fan at two sides will be operating. The meter performs a series of
self-test, 11050 display unit will be shown as below.
3-5
HF LCR Meter 1105X Series User’s Manual
If “COMMUNICATION ERROR, PLEASE REBOOT” message pops up, the screen shows as
below. Please power off the meter, confirm whether display unit and measurement unit
connections are normal as well as reboot it.
3-6
Installation
1. Troubleshooting:
(1) When FAN LOCK protection is occurred, please power off the
meter and then reboot to observe if it is caused by error action.
(2) If FAN LOCK is occurred again, please contact the distributor or
Chroma’s service department for service.
2. After powering on the meter and start operating, please keep
ventilation at left and right sides of meter at least 10cm.
CAUTION
3.7 FAN LOCK Protection
For keeping internal heat dissipation and getting the stable measurement, FAN LOCK
protection will be occurred if there is fan failure or foreign material invasion. The meter can’t
be operated and outputted at the same time, operation screen shows the protection message
as below.
3.8 Contrast Adjustment of LCD Display
To adjust the contrast of LCD display that can be used by the system configuration in system
setup, but notice that the adjustment should depend on personal acceptable brightness.
3-7
HF LCR Meter 1105X Series User’s Manual
Press this key to increase brightness.
Press this key to decrease brightness.
Software/hardware version of display unit
Maximum frequency of the meter
Maximum DCA of the meter
Software/hardware version of meas. unit
Display unit BOOT
Model number/name
Company name
3.9 Query Version
Query version can via ABOUT in system setup, the screen as below shows. The information
in the screen below users can get model number/name, program version and maximum
measurement frequency.
3.10 Maintenance & Cleaning
Remove all connected wires and cables on the instrument before cleaning. Use a brush to
clean the dust on it. If the cover with dirt, use cleaning naphtha to wipe it. Don’t use
corrosive liquid to avoid the cover damaged. For internal cleaning, use a low-pressure air
gun to vacuum the dust inside or send it back to the distributors or agents of Chroma for
cleaning.
3.11 Notices for Using Fixture & Test Cable
When the test cable or fixture connects DUT for proceeding measurement, please keep the
test fixture away from the wire with high current, high voltage or high frequency signal to avoid
the interference on the measurement.
3-8
Descriptions of Display Unit & Meas. Unit
Press this key to allow entry to LCR MODE.
Press this key to allow entry to LCRZ SCAN
MODE.
Press this key allows entry to BIAS SCAN MODE.
Press this key to allow entry to PARAMETER
SWEEP MODE.
Press this key to allow entry to DUAL FREQUENCY
MODE.
4. Descriptions of Dispaly Unit & Meas.
Unit
4.1 Display Unit
4.1.1 Front Panel
(1) LCD Display
The display of this test device is a color 640 X 480 LCD Panel with backlight function.
(2) Main Index
Pressing this key allows entry to the main index screen. In this screen you may select
the test function you wish to use.
(3) Meas. Display
Upon pressing this key, the instrument is in main index and setup screens switching.
4-1
HF LCR Meter 1105X Series User’s Manual
Main Index
LCRZ MODE
PARAMETER SWEEP MODE
[Meas Display]
[Meas Display]
[Meas Display]
[Meas Display]
LCRZ SETUP
PARAMETER SWEEP SETUP
(4) Numbers and Symbol Keys ( [0] …… [9] [ . ] [ - ] )
These numbers and symbol keys are for each setting condition which needs to input
value and symbol.
(5) Enter Key
Its function is to check the value setting or condition selection and parameters setting.
(6) Reset Key
Upon pressing the Reset key, the measurement function stops when performing scan &
sweep tests.
(7) Clear Key (CLR)
It clears the inputted values before defining it (ENTER key unpressed). The value is at
the cursor located place on the display.
(8) Start Key
This is the key to trigger the start of measurement.
4-2
Descriptions of Display Unit & Meas. Unit
(9) , , ,
There are , , , four arrow keys. These keys set and control the cursor
movement, which can be useful when inputting various parameters.
(10) Lock Key
It is a panel key lock function. When pressing this key, the upper right side of the display
will show lock pattern indicating the key locked. If users desire to cancel this condition
just press this key again.
(11) System Setup
Press this key the tester will be under system setup screen. The options can be selected
including relate information of main unit, primary system parameter, password setting and
memory management.
(12) Selection key
It includes seven keys. The major function shows different conditions of each function or
other options which needs to be selected by the user’s requirement.
4.1.2 Bottom View of Display Unit
(1) Buzzer beeping location
The location of buzzer sounds.
(2) Earphone Connector
The connector is for connecting earphone.
(3) REMOTE CONTROL
Use D-SUB 15Pin connection cable to connect to measurement unit.
(4) Ground terminal
This terminal connects the cover of this instrument directly, then connect to the ground to
avoid exterior obstruction that makes instrument connection uncompleted and influences
the safety.
4-3
HF LCR Meter 1105X Series User’s Manual
(1) Model number of LAN & USB-H interface card is A133510 (option).
Its main function is network communication and for connecting Bar
Code Reader interface.
(2) Model number of GPIB interface card is A133509 (option).
4.1.3 Left Side View of Display Unit
(1) Interface Slot
It is a slot for plugging in LAN & USB-H or GPIB interface. Unscrew two screws on light
gray cover marked 1 of above figure and plug in optional interfaces as well as lock the
screws back.
(2) USB Interface
Standard FLASH MEMORY interface
(3) RS-232 Interface
Standard RS-232 interface
4.1.4 Rear Panel of Display Unit
(1) REMOTE CONTROL
Use D-SUB 15Pin connection cable to connect to measurement unit.
(2) Bracket Mounting Hole
The screw holes are for locking bracket of MEASUREMENT UNIT and DISPLAY UNIT.
4-4
Descriptions of Display Unit & Meas. Unit
That “high” terminals for (+) and “low” terminals for () polarity as marked
on the front panel while polarized component is under test.
1. It is prohibited to connect the output over 50V PEAK MAX.
Discharge first before measuring the charged component to avoid
damaging the instrument. Test the charged component with
A110242 fixture (option) is recommended and the withstand voltage
is up to 100V.
2. The cover and axis of LCUR test terminal can’t be connected over
1V. Make sure the ground integrity to avoid it burning down when
Bias Current system test is required.
CAUTION
WARNING
There are four screw holes at top side and four at bottom side. The bracket locked to
screw holes at top or bottom side is adjusted by users’ visual angle.
4.2 Measurement Unit
4.2.1 Front Panel
(1) UNKNOWN
Four individual BNC sockets, connect an external test device or wire for DUT
measurement.
H
: Current drive terminal, high potential terminal.
CUR
H
: Potential detection terminal, high potential terminal.
It indicates the meter under measuring while this LED is blinking.
(3) POWER LED
It indicates power on while this LED lit continuously.
(4) POWER Switch
This pressed indicates power on and unpressed indicates power off.
4-5
HF LCR Meter 1105X Series User’s Manual
1. Power on sequence: 11050 first and then 1320 when 11050 adds
DC bias current 1320.
2. Use 2014 delivered version of DC Bias Current Source 1320.
Please contact Chroma’s customer service if there is problem about
1320 version.
CAUTION
4.2.2 Rear Panel of Measurement Unit
(1) EXT. BIAS
DC bias input for connecting externally, 35V maximum input voltage.
(2) HANDLER
It is a I/O interface for outputting test result signal and triggering externally. For detail
descriptions, please see chapter 6 HANDLER Interface.
(3) REMOTE CONTROL
Use D-SUB 15Pin connection cable to connect to display unit.
(4) EXT -BIAS CURRENT
It connects and controls external DC bias current 1320.
(5) Ground terminal
This terminal connects the cover of this instrument directly, then connect it to the ground
to avoid exterior obstruction that makes instrument connection uncompleted and
influences the safety.
(6) AC Line Socket
It meets the International Electromechanical Commission 320 standard for 3-wire socket.
Please use the power cord of Belden SPH-386 or similar.
4.2.3 Side View of Measurement Unit
(1) Fan Inlet/Outlet
Air inlet/outlet of fan at two sides
4-6
Descriptions of Display Unit & Meas. Unit
The meter should be located at least 10cm from any object or wall behind
it. Do not block the left and right ventilation holes to keep the meter in
good precision.
(2) Rear Foot Pad
It’s fixed foot pad.
(3) Front Foot Pad
It’s a adjustable foot pad that provides appropriate angle to users.
4-7
Setting for Operation
Press this key to enter system relate setup screen.
Press this key to enter measurement selection screen.
Press this key to enter test screen.
Press this key to clear count.
Press this key to select PAGE or EDIT
interface.
Press this key to select DISPLAY MODE.
Press this key to enable BIN MODE.
Press this key to enable LOAD CORRECTION.
Press this key to enable LIMIT MODE.
5. Setting for Operation
5.1 Basic Operation
The tester mainly provides the right key of Display Unit front panel to users’ control interface.
Its operation is easy and convenience. Next to introduce several basic operation modes as
below.
1. Main Menu Key Setting:
[MEAS DISPLAY], [MAIN INDEX] and [SYSTEM SETUP] are function keys for three main
screens shown as below. [MEAS DISPLAY] can return the previous selected test screen
or quick enter basic setting screen at the present mode. [MAIN INDEX] is a selection
screen of measurement mode and [SYSTEM SETUP] is the system relate setup menu of
the meter.
Test screens will be varied by test function selection, LCR parameter test screen is as the
following.
MAIN INDEX menu as below shows.
5-1
HF LCR Meter 1105X Series User’s Manual
System Setup menu as below shows.
2. Basic Operation:
(1) Function Selection:
Selecting the screen option needs to use selection key, direction control keys
and to move the highlight to the function desire to select and then press ENTER
for entry. System Setup menu as below shows.
(2) Key in Setting:
To set the setting or parameter in function option needs to use selection key,
direction control key, ENTER, and numeral keys. The following screen is
STORE/RECALL file name setting, the right of the screen is function option by using
selection key to select the function to be used. When moving the highlight to the
set option, the bottom of screen has instruction such as “Key in file name”.
5-2
Setting for Operation
Press this key to exit the screen (back
to previous page).
Press this key to store the file.
Press this key to copy the file.
Press this key to search the file.
Press this key to recall the file.
Press this key to paste the file.
Press this key to delete the file.
Press this key to cancel input.
Press this key to enter option setting.
Press this key to move the option downward.
Press this key to move the option upward.
Meanwhile, File Name needs to perform editing, input English alphabet or numeral.
After pressing ENTER, select English alphabet at the right of screen or numeral key
for editing as below shows.
(3) Shortcut Key Operation:
Take SETUP menu of LCR Mode as an example, the function or table desires to set
with shortcut key function. The bottom of screen shows the option that can be set.
The front of option has a numeral indicating its settings, key in the option numeral
key to be set thus complete the setting. Press numeral key [1] as the screen below
shows, the test parameter will be immediate set to C parameter.
Press ENTER on the function or table to be set, the right side of screen shows all
shortcut keys.
5-3
HF LCR Meter 1105X Series User’s Manual
PARAMETER1
PARAMETER2
0:L
0:D / 1:Q / 2:R / 4:X / 5:θ / 6:Q+R
1:C
0:D / 1:Q / 2:R / 5:θ
2:R
1:Q / 4:X
3:Z
5:θ
4:Y
5:θ
5:DCR
Press this key to return to previous
option.
Press this key for increment – fine tune
Press this key for discrement – fine
tune
Press this key for discrement – coarse
tune
Press this key for increment – coarse
tune
The various parameters will map to different shortcut keys as the Table 5-1 shows.
The unused shortcut keys will be disabled and not displayed, other settings can be
done in the same manner.
If the option setting is a value, the bottom of screen shows the range of value. The
following screen shows test voltage settings in SETUP page of LCR Mode.
Use the value option at the right screen to select test voltage to be set by scrolling
up and down.
Key in the test voltage to be set and press mV selection key if mV required then the
setting is completed.
5-4
Press this key to return to previous
option.
Press this key to select unit.
Press this key to select BIN SORTING
SETTING.
Press this key to select CORRECTION
function.
Press this key to select PAGE or EDIT
interface.
Press this key to select SETUP function .
Press this key to select STORE/RECALL
function.
Press this key to select LIMIT SETTING.
Press this key to clear count.
Press this key to select PAGE or EDIT
interface.
Press this key to select DISPLAY MODE.
Press this key for BIN MODE ON.
Press this key for LOAD CORRECTION
ON.
Press this key for LIMIT MODE ON.
(5) EDIT and PAGE Interfaces Operation:
There are PAGE and EDIT interfaces at the right side of screen, use selection key to
select the interface as the below shows LCR Mode test screen. Select any
function option under PAGE interface thus it enters the selected function setting
screen.
Setting for Operation
The function set can quick enable/disable and switch along with clear display
screen mode if EDIT interface selected.
3. Description for System Setup:
The upper right corner of each operation screen has five icons for showing operation
status currently of the meter as below red frame shows. The hightlighted area indicates
it is in use or is with the function currently whereas this function isn’t equipped if it is
unhightlighted. The detail descriptions are as the following.
5-5
HF LCR Meter 1105X Series User’s Manual
1
2 3 4
5
Zoom in icon
(1) GPIB function icon
It indicates the meter is with GPIB connection function and equipped with GPIB
interface. (option).
(2) RS232 function icon
RS232 function icon. RS232 interface is the standard equipment of the meter, the
icon will be highlighted after powered-on.
(3) LAN function icon
It indicates the meter is with LAN connection function and equipped with LAN &
USB-H interface. (option).
(4) REMOTE function icon
When the external connection device controls the meter, it is already controlled by
external connection device. If there is error command, this icon will turn into red
shown as .
(5) LOCK function icon
Press LOCK key while the tester is under setting or measuring. The icon will be
displayed as highlighted.
5.2 Setup Description
5.2.1 System Configuration
Power on the meter, wait the display until for entry operation screen. Press [SYSTEM
SETUP] to enter SYSTEM SETUP screen as below shows.
5-6
Press this key to exit from this screen.
Entering System Configuration screen as below shows.
Setting for Operation
The parameters of SYSTEM CONFIGURATION are given as the following.
AUTO REPORT:
It auto generates measurement result and auto outputs data result under measuring via
RS232 interface on Display Unit. The control range is “ON/OFF” and default value when
delivery is with “OFF”.
HANDLER JUDGMENT:
Output Limit Test result of measurement data Judgment to external fixture or device via
Handler interface on Measurement Unit . Please see chapter 6 HANDLER Interface.
Users select judgment signal level with “HIGH” or “LOW” as output. The default setting is
“HIGH”.
BRIGHTNESS:
To adjust the contrast of the display. The control range is “0~10”, the value 0 indicates
darkest and the value 10 indicates brightest. The default setting is “5”.
BUZZER:
It sets the warning beep volume to SMALL, LARGE, and OFF. The default setting is
LARGE.
SOUND MODE:
It selects warning beep of judgment mode. The function needs to coordinate with Buzzer
fuction for activation, there are two selections of “PASS” and “FAIL” can be set. The meter
sounds beep if the setting is PASS and DUT measurement result meets setting condition.
Otherwise, the meter also sounds beep when the setting is FAIL and DUT measurement
5-7
HF LCR Meter 1105X Series User’s Manual
result can’t meet setting condition. The default setting is PASS.
ALARM MODE:
It sets the beep of buzzer. The function needs to coordinate with Buzzer fuction and SOUND
MODE setting for activation, its selections are PULSE and LEVEL. The default setting is
PULSE.
SYSTEM TIME YEAR:
The system running year currently. The factory default is actual year.
SYSTEM TIME DATE:
The system running date currently. The factory default is actual month/date.
SYSTEM TIME TIME:
The system running time currently. The factory default is actual time.
USER PASSWORD:
It is for keying in user exclusive password when the meter powered-on or modifying setting
parameter. Its control range is two selections of ON and OFF. The operation is to press
[LOCK] to lock all control interfaces when the setting is “ON”. Press [LOCK] again if it needs
to be unlocked and key in exclusive password. For password setting, please see section
5.2.2. The default setting is OFF.
LANGUAGE:
It selects the interface displayed language. The setting range includes two selections of
English and S.CHINESE. The default setting is English.
5.2.2 Password Setup
Select PASSWORD SETUP to enter password setting under SYSTEM SETUP, meanwhile
input password window will be popped up. Key in PASSWORD for entering the setup. The
meter initial setting exclusive SUPERVISOR PASSWORD is 201211050 as the screen below.
After entering PASSWORD SETUP screen, there is SUPERVISOR PASSWORD for users
keying in exclusive password, 12 digits at most. The password in the following is default
setting that we recommend users to change it at first time use.
5-8
Setting for Operation
If PASSWORD is missing, the instrument needs to be return to Chroma
for reset.
CAUTION
Press this key to exit.
Press this key to enter DEFAULT LOAD.
Press this key to enter DEFAULT RAM.
Press this key to exit.
Press this key to enter DEFAULT
OPEN/SHORT.
When USER PASSWORD function in SYSTEM CONFIGURATION is activated, press [LOCK]
to lock all control interfaces to avoid users’ misaction. Press [LOCK] again if it needs to be
unlocked and key in exclusive password.
5.2.3 Memory Management
Select MEMORY MANAGEMENT under SYSTEM SETUP status. Key in SUPERVISOR
PASSWORD to enter memory management function as below shows.
DEFAULT OPEN/ SHORT:
It sets open and short values initialized. Clear open/short offset value of fixture and restore
to default setting of the meter.
“DEFAULT OPEN/ SHORT, ARE YOU SURE ?” window pops up as operating DEFAULT
OPEN/ SHORT. Click “YES” to delete it, otherwise click “NO”.
5-9
HF LCR Meter 1105X Series User’s Manual
YES
NO
Press this key to enter DEFAULT LOAD.
Press this key to enter DEFAULT RAM.
Press this key to exit.
Press this key to enter DEFAULT
OPEN/SHORT.
YES
NO
“Waiting 3 seconds” screen pops up when “YES” clicked, the clearance action is completed
when ending second count down.
DEFAULT LOAD:
Load offset data initialization, clear load offset value.
“DEFAULT LOAD, ARE YOU SURE ?” window pops up as operating DEFAULT LOAD. Click
“YES” to delete it, otherwise click “NO”.
“Waiting 3 seconds” screen pops up when “YES” clicked, the clearance action is completed
when ending second count down.
5-10
Setting for Operation
Press this key to enter DEFAULT LOAD.
Press this key to enter DEFAULT RAM.
Press this key to exit.
Press this key to enter DEFAULT
OPEN/SHORT.
YES
NO
Press this key to enter DEFAULT LOAD.
Press this key to enter DEFAULT RAM.
Press this key to exit.
Press this key to enter DEFAULT
OPEN/SHORT.
DEFAULT RAM:
Memory initialization, all operation settings will be returned to default. SUPERVISOR
PASSWORD and USER PASSWORD still remains original setting, the test file saved in the
meter won’t be deleted.
“DEFAULT RAM, ARE YOU SURE ?” window pops up as operating DEFAULT RAM. Click
“YES” to delete it, otherwise click “NO”.
“Waiting 10 seconds” screen pops up when “YES” clicked, the clearance action is completed
when ending second count down.
5-11
HF LCR Meter 1105X Series User’s Manual
Press this key to enter RS232 setiing.
Press this key to enter LAN setting.
Press this key to enter GPIB setting.
Press this key to exit.
5.2.4 Interface Setup
Select INTERFACE SETUP under SYSTEM SETUP status, the right of screen shows three
interfaces of RS232, LAN (option) and GPIB (option) as the screen below.
5.2.4.1 RS232 Setting
Select RS232 setting to enter INTERFACE SETUP-RS232 as below screen shows.
Parameter and setting are as the following.
BAUD RATE:
Baud rate setting, there are 9600, 19200, 38400, 57600and 115200 for selection with factory
default to 115200 when shipped.
WORD LENGTH:
The WORD LENGTH is fixed to 8 which is the same as ASCII code.
PRRITY CHECK:
There are NONE, ODD and EVEN for selection with factory default to NONE when shipped.
STOP BIT:
Stop bit is fixed to 1.
FLOW CONTROL:
There are NONE and H/W(hardware) for selection with factory default to NONE when
shipped.
5-12
The baud rate setting should be same as the RS-232C interface card, or
there may have data transmission problem.
Pin
Name
Description
Ground
5
GND
Grounding wire
Data
2
RxD
Receiving data
3
TxD
Transmitting data
7
RTs
Transmitting request
8
CTs
Transmitting clear
When the instrument is connected via RS232 and PC, it is recommended
both use the same system power voltage. This is for avoiding burn down
the instrument or component inside PC terminal interface.
CAUTION
LINE
PC
2232
3323
5555
7787
8878
RTs
CTs
GND
RxD
TXD
RxD
TXD
GND
9 Pin(M)
RTs
CTs
9 Pin(F)
9 Pin(F)
9 Pin(M)
Instrument
CAUTION
Press this key to CONFIRM.
Press this key to exit.
Connection of RS-232C signal line:
Setting for Operation
Table of RS-232C signal line and pin:
5.2.4.2 LAN Setting
Select LAN setting to enter INTERFACE SETUP-LAN as screen below shows.
5-13
HF LCR Meter 1105X Series User’s Manual
Note
PORT is fixed used in 16881.
Press this key to exit.
TCP/IP MODE:
Transmission Control Protocol and Internet Protocol setting methods. The meter includes
two modes of Automatic and Manual. If Automatic mode is selected, a DHCP (Dynamic Host
Configuration Protocol) server needs to be set up on network. The meter will get IP
ADDRESS, SUBNET MASK and GATEWAY settings from DHCP server automatically. After
getting the settings, please press CONFIRM for confirmation. If Manual mode is selected,
set IP ADDRESS and SUBNET MASK by users. After the setting is completed, please press
CONFIRM for confirmation. The default setting is Manual.
IP ADDRESS:
Internet communication protocol address.
SUBNET MASK:
It is for subnet mask.
GATEWAY:
Communication gateway
5.2.4.3 GPIB Setting
GPIB ADDRESS:
It selects GPIB address. The factory default setting is 17 and the range is 1~30.
5.2.5 SELF TEST RESULT
Select SELT TEST RESULT under SYSTEM SETUP as the screen below shows.
5-14
Setting for Operation
The calibration data should be shown as PASS. If it is FAIL, please
contact the distributor for the instrument calibration.
CAUTION
Press this key to exit.
Enter SELT TEST RESULT, the screen shows the meter’s status currently and if calibration
data is normal operation as below.
LCR CALIBRATION DATA: Parameter calibration data of passive component
DCR CALIBRATION DATA: DC resistance calibration data
LCR OPEN DATA: Parameter open data of passive component
LCR SHORT DATA: Parameter short data of passive component
DCR SHORT DATA: DC resistance short data
DCA CALIBRATION DATA: DC current calibration data
LOAD DATA: Load offset data
LCR CALIBRATION OPEN: Open calibration of passive component
LCR CALIBRATION SHORT: Short calibration of passive component
SOURCE LEVEL CALIBRATION DATA: Level calibration of measurement signal source
LINK 1320: The connection status when it connects Bias Current Source 1320.
5.3 Operation of Main Index
Press [MAIN INDEX] to enter main index menu, where four modes of LCR MODE, LCRZ
SCAN MODE, BIAS SCAN MODE and PARAMETER SWEEP MODE for selection as the
screen below shows.
5-15
HF LCR Meter 1105X Series User’s Manual
Press this key to enter LCR MODE.
Press this key to enter LCRZ SCAN MODE.
Press this key to enter BIAS SCAN MODE.
Press this key to enter PARAMETER SWEEP MOD.
Press this key to enter DUAL FREQUENCY MODE.
LCR MODE:
It sets measurement parameter and environment for quick measuring DUT parameter and
providing judgment results of test value comparison or binning.
LCRZ SCAN MODE:
Users set multiuple sets (9 sets at most) of various test frequencies, voltage sizes and
parameters for DUT performing compare test by DUT frequency and voltage characteristic.
It shows test value of DUT by the table and provides compare judgment result of test value.
BIAS SCAN MODE:
This meter is able to connect Chroma’s DC Bias Current Source 1320 for performing
inductance DC bias current scanning test. Users set multiuple sets (27 sets at most) of
various test frequencies, current sizes and parameters for performing compare test. It
shows test value of DUT by the table and provides compare judgment result of test value.
PARAMETER SWEEP MODE:
Users sets test frequency, test voltage or test current in a certain range for DUT performing
multiple points (401 points at most) sweeping measurement that provides detail graph
analysis to users’ reference.
It is general common used measurement display screen for setting basic test parameters.
The related operation descriptions are described as the following.
The upper left side of LCR DISPLAY screen is measurement parameter and environment
setting, the lower left side is measurement result display area as well as the right side is
open/short offset or value operation statistical function as below shows.
5-16
Setting for Operation
Note
Resistance of measurement range will be limited by varied test frequency, range
descriptions are as follows.
1. When FREQUENCY50kHz, there are 12 ranges for selection - “0.3”, “1”,
3. When 200kHz < FREQUENCY < 1MHz, there are 8 ranges for selection –
“1”, “3”, “10”, “30”, “100”, “300”, “1k” and “3k”.
4. When FREQUENCY 1MHz, there are 4 ranges for selection - “10”, “30”,
“100” and “300”.
Press this key to select PAGE or EDIT
interface.
Press this key to select DISPLAY MODE.
Press this key to CLEAR COUNT.
Press this key to enable LOAD function.
Press this key to enable LIMIT function.
Press this key to enable BIN function.
Descriptions for measurement parameter and environment setting are as the following.
PARAMETER1:
It is primary parameter that includes six measurement parameters of “L”, “C”, “R”, “Z”, “Y” and
“DCR” for selection. The default setting is “R”.
PARAMETER2:
It is secondary parameter that follows primary parameter setting coordinating with different
secondary measurement parameters. The default setting is “X”.
RANGE MODE:
It is for switching range mode that includes “AUTO” and “HOLD”. The default setting is
“AUTO”.
RANGE:
It is for selecting resistance of measurement range. The RANGE MODE will be changed
from “AUTO” to “HOLD” when adjusting resistance of measurement range. The default
setting is “300”.
CKT(S/P):
It is for selecting equivalent circuit SERIAL or PARALLEL mode. The default setting is
SERIAL.
SPEED:
It sets measurement speed that includes FAST, MEDIUM and SLOW for selection. The
default setting is FAST.
5-17
HF LCR Meter 1105X Series User’s Manual
1. Power-on sequence is 11050 first and 1320 followed when adding
DC Bias Current Source 1320 to 11050.
2. Use 2014 delivered version of DC Bias Current Source 1320.
Please contact Chroma’s customer service if there is problem about
1320 version.
CAUTION
Press this key to select PAGE or EDIT
interface.
Press this key to enter SETUP.
Press this key to process CORRECTION
function.
Press this key to enable LIMIT setting.
Press this key to enable BIN sorting
setting.
Press this key to enable store/recall
function.
FREQUENCY:
It sets test frequency for setting test frequency range of 60Hz~5MHz /1KHz~10MHz. For the
limitation of usage range, please see section 2.2 Test Signal. The default setting is “1M” Hz.
LEVEL:
It sets test voltage for selecting test voltage range of 10mV~1V. For the limitation of usage
range, please see section 2.2 Test Signal. The default setting is “1.000” V.
TRIGGER:
It selects trigger mode that includes “INT”, “MAN”, “EXT” and “BUS”. The defaule setting is
“INT”.
BIAS V ENABLE:
It enables bias voltage. When the DC bias input connected externally is enabled, for detail
description see EXT. BIAS BNC connector in section 4.2.2 Rear Panel of Measurement Unit.
Its operation includes ON and OFF, the default setting is “OFF”.
BIAS I:
It sets bias current. The range of the meter is 0mA~20mA when DC Bias Current Source
1320 doesn’t be connected. Select the number in LINK 1320 option first and then set bias
current value in BIAS I option when connecting DC Bias Current Source 1320. The default
setting is “0.000” A.
BIAS I ENABLE:
It enables bias current. When it is enabled under the condition of bias current input need
which includes the operations of “ON” and “OFF”. The default setting is “OFF”.
Press [MEAS DISPLAY] on LCD DISPLAY screen or select PAGE interface at the right side of
screen for entering SETUP function as below shows.
For entry measurement environment setting as the screen below shows.
5-18
Setting for Operation
Press this key to exit.
Descriptions for SETUP page are as the following.
The page above divides into upper and lower areas. The upper area contains 12 test
parameters of PARAMETER1, PARAMETER2, RANGE MODE, RANGE, CKT (S/P), SPEED,
FREQUENCY, LEVEL, TRIGGER, BIAS V ENABLE, BIAS I and BIAS I ENABLE. These
parameters are the same as LCR DISPLAY thus don’t take more description here. The
lower area belongs advanced setting and the descriptions are as the following.
CONST Rsou:
It selects output resistance mode including three modes of “OFF”, “100” and “25” for
selection. The default setting is “25”.
DCR MODE:
It selects DC resistance mode. The meter measures output test signal sources of DCR
parameter that includes “SMALL” (small DC signal source≒20mA) and “NORMAL” (normal
DC signal source ≒100mA) two selections. Select the appropriate measurement signal
source by DUT’s resistance. The default setting is “NORMAL”.
TRIGGER DELAY:
When it connects with automatic machine, external trigger EXT for coordinating with its action.
Meanwhile, the trigger action of meter needs to be delayed thus users can adjust TRIGGER
DELAY time. Its setting range is “09999” ms and default setting is “0” ms.
MEAS. DELAY:
It is for PARAMETER1 as additional function of DCR measurement. The inductance is with
transient effect thus the instrument confirm the stability of DUT signal first in testing
inductance. If DUT signal is stable thus it enters measurement at once, it enters
measurement after MEAS. DELAY TIME if the signal is unstable. Control range of MEAS.
DELAY TIME is “099999”ms. The default setting is “0” ms.
AVERAGE:
It sets average time of measurement. Select the average setting as “1”, the program will
take average in each time of measurement. If the setting is “2”, the program will take
average in two times of measurements. Following test can be deduced accordingly. The
setting range is “1~10”, the default setting is “1”.
LINE FREQ.:
It selects input AC power mode. The correct power frequency selection helps stability and
correctness of test value. Three selections of “50Hz”, “60Hz” and “N/A” for range. “N/A” indicates
not consider power frequency but test speed will be faster. The default setting is “60Hz”.
5-19
HF LCR Meter 1105X Series User’s Manual
Note
1. A set of 1320 provides 20A.
2. Power-on sequence is 11050 first and 1320 next when adding DC Bias
Current Source 1320 to 11050.
3. Use 2014 delivered version of DC Bias Current Source 1320. Please
contact Chroma’s customer service if there is problem about 1320 version.
Press this key to select PAGE or EDIT interface.
Press this key to enter SETUP.
Press this key to process CORRECTION
functrion.
Press this key to enable LIMIT setting.
Press this key to enable BIN sorting setting.
Press this key to enable Store/Recall function.
Press this key to select PAGE or EDIT interface.
Press this key to enter SETUP.
Press this key to process CORRECTION function.
Press this key to enable LIMIT setting.
Press this key to enable BIN sorting setting.
Press this key to enable Store/Recall function.
Vm/Im DISPLAY:
It displays voltage and current value. VM and IM at the bottom of LCR DISPLAY show
voltage and current value on the DUT if “ON” is selected. The default setting is “OFF”.
LCR DISPLAY Page shows as below after the function is enabled.
Z/θ DISPLAY:
It displays impedance and phase angle. Z and θ at the bottom of LCR DISPLAY will be
shown as below if “ON” is selected. The default setting is “OFF”.
LCR DISPLAY Page shows as below after the function is enabled.
LINK 1320:
It sets the number for connecting 1320. BIAS I setting current follows linked 1320 quantity to
provide bias current for setting if 1320 number is set. The maximum setting is 15 sets and
the default setting is “0”.
BIAS I READY:
It sets current stable time that is for coordinating DELAY.T setting after outputting bias current
source 1320. The 11050 needs to be set 10s for measurement when 1320 DELAY TIME
5-20
Setting for Operation
The screen shows ”ALC unable to regulate” message indicating ALC
function can’t modify test voltage as below shows.
CAUTION
Press this key to select PAGE or
EDIT interface.
Press this key to enter SETUP.
Press this key to process
CORRECTION function.
Press this key to enable LIMIT
setting.
Press this key to enable BIN
sorting setting.
Press this key to enable Store/
Recall function
setting is 10s. BIAS I READY control range is “0.1100”s and the default setting is “0” s.
D.P. FIX1:
The displayed digit of primary parameter after decimal point. It follows users’ requirement to
set measurement of primary parameter and to display digit length with 1~5 digit can be
selected. The default setting is “4”.
D.P. FIX2:
The displayed digit of secondary parameter after decimal point. It follows users’ requirement
to set measurement of secondary parameter and to display digit length with 1~5 digit can be
selected. The default setting is “4”.
ALC:
It auto compensates voltage value of DUT in measuring. ALC (Auto Level Control) function
is for setting voltage level and getting stable test voltage for DUT that won’t be varied by
DUT’s impedance level. The default setting is OFF.
Except for basic measurement screen in LCR MODE and setting method of measurement
parameter, other two displayed modes are also included. Select DISPLAY MODE in EDIT
interface at right menu of LCR DISPLAY. “NORMAL”, “JUDGMENT” and “SPECIAL”are in
DISPLAY MODE for switching which will be described as follows.
JUDGMENT Mode:
This screen is mainly applied on LIMIT TEST or BIN SORTING TEST to count pass, fail and
DUT total number.
LIMIT TEST screen:
5-21
HF LCR Meter 1105X Series User’s Manual
Note
If LIMIT TEST is selected thus BIN SORTING TEST will be disabled. On the
contrary, LIMIT TEST will be disabled if BIN SORTING TEST is selected.
Press this key to select PAGE or EDIT
interface.
Press this key to select DISPLAY MODE.
Press this key to CLEAR COUNT.
Press this key to enable LOAD function.
Press this key to enable LIMIT mode.
Press this key to enable BIN mode.
Press this key to select PAGE or EDIT
interface.
Press this key to select DISPLAY MODE.
Press this key to CLEAR COUNT.
Press this key to enable LOAD function.
Press this key to enable LIMIT mode.
Press this key to enable BIN mode.
Press this key to select PAGE or EDIT
interface.
Press this key to select DISPLAY MODE.
Press this key to CLEAR COUNT.
Press this key to enable LOAD function.
Press this key to enable LIMIT mode.
Press this key to enable BIN mode.
BIN SORTING TEST screen:
SPECIAL Mode:
This screen simplify a lot of settings that only remains displayed measurement for users
observing measurement result clearly.
“Operation Sample”
Take measurement power inductance as an example, measurement and setting procedures
are as the following.
(1) Click [MAIN INDEX] on DISPLAY UNITto enter LCR MODE.
5-22
Step
Function
Parameter
Unit
Remark
1
PARAMETER1
Ls
2
PARAMETER2
Q or Rs
3
RANGE MODE
Auto
4
CKT(S/P)
SERIAL
5
SPEED
SLOW
6
FREQUENCY
1.000M
Hz
7
LEVEL
1.000
V
Press this key to enter LCR MODE.
(2) Enter LCR MODE initialation screen
Setting for Operation
(3) Measure L/(Q or Rs) of DUT to enter LCR DISPLAY screen. Follow DUT – inductance
characteristics separately to set primary/secondary parameter, range mode, equivalent
circuit (series/parallel) mode, measurement speed, test frequency, test voltage, etc.
Please refer the Table 5-2 DUT – Inductance L/(Q or Rs) Setting to set the figure below by
sequential.
Table 5-2 DUT – Inductance L/(Q or Rs) Setting
5-23
HF LCR Meter 1105X Series User’s Manual
Note
The initial setting of the meter is Internal Trigger (INT.) after powered on entering
LCR mode.
1
Press this key to select L.
Step 1: Operate the direction control keys of , , or on DISPLAY UNIT to
highlight the column of PARAMETER1.
Press [ENTER] on DISPLAY UNIT and select L parameter at the right side of wondow.
Step 2: Press the right key on DISPLAY UNIT to highlight the column of
PARAMETER2 and [ENTER] to select Q or Rs parameter (Q parameter is selected in the
example).
5-24
Setting for Operation
Press this key to select Q or Rs.
Press this key to select Serial.
Press this key to select Slow.
Step 3: Press the direction control keys of , , or on DISPLAY UNIT to
highlight the column of CKT(S/P) and [ENTER] to select equivalent circuit to Serial or
Parallel. Serial is selected in the example.
Step 4: Press the right key on DISPLAY UNIT to highlight the column of SPEED and
[ENTER] to select SLOW.
Step 5: Press the direction control keys of , , or on DISPLAY UNIT to
highlight the column of FREQUENCY. Up/down scroll bar will be shown on the right of
window. The highlighted value will be varied by users’ scrolling and changed it to
1.000MHz.
5-25
HF LCR Meter 1105X Series User’s Manual
Press this key for decrement in fine tuning
Press this key for decrement in coarse
tuning
Press this key for increment in fine tuning
Press this key for increment in coarse tuning
MHz
1
Press this key for decrement in fine tuning
Press this key for decrement in coarse
tuning
Press this key for increment in fine tuning
Press this key for increment in coarse tuning
Or use numerical key [1] on DISPLAY UNIT to select MHz at the right of window.
Step 6: Press the right key on DISPLAY UNIT to highlight the column of LEVEL.
Up/down scroll bar will be shown on the right of window. The highlighted value will be
varied by users’ scrolling and changed it to 1.000V.
Or use numerical keys on DISPLAY UNIT to input the setting value and press [ENTER].
5-26
Setting for Operation
Step 7: Connect the DUT to UNKNOWN four ends BNC terminal of meter for measuring
its value.
5.3.1.2 LCR MODE- OPEN /SHORT CORRECTION
This section is aimed to introduce open/short correction for measuring ACR. The tester only
performs machine inner calibration when delivery. It also executes open/short correction for
external connected test fixtures that avoids the fixture affecting DUT and causing additional
measurement error. For DCR correction, please see the next section. The diagram as
below shows.
5-27
HF LCR Meter 1105X Series User’s Manual
CORRECTION
Press this key to enter LOAD correction.
Press this key to exit.
Open/short correction operation: Select CORRECTION at the right Page of LCR Dispaly to
enter CORRECTION-OPEN/SHORT function as the screen below shows.
Descriptions for CORRECTION-OPEN/SHORT screen are as the following.
1st LCR FREQUENCY:
It sets 1st LCR correction frequency. Users perform 1st OPEN and SHORT single frequency
correction for the required two sets of specific frequency. The setting frequency ranges of
11050-5M/11050 are 60Hz~5MHz and 1kHz~10MHz. The default settings are “100kHz” and
“1MHz”.
2nd LCR FREQUENCY:
It sets 2nd LCR correction frequency. Users perform 2nd OPEN and SHORT single
frequency correction for the required two sets of specific frequency. The setting frequency
ranges of 11050-5M/11050 are 60Hz~5MHz and 1kHz~10MHz. The default settings are
“100kHz” and “1MHz”.
LCR OPEN:
It sets LCR for performing open correction on fixture. It can select SINGLE OPEN or
MULTIPLE OPEN, SINGLE OPEN needs to coordinate with LCR FREQUENCY set frequency
to do correction. Change “OFF” in LCR OPEN column to “ON” after SINGLE OPEN or
MULTIPLE OPEN executed. The default setting is “OFF”.
Take a test fixture to execute MULTIPLE OPEN correction as an example, connect the fixture
to UNKNOWN four ends of BNC connector. The fixture is short-circuited with HCUR and
5-28
Setting for Operation
Press this key to enable open correction.
Press this key to execute multiple open
correction.
Press this key to return to the previous selection.
Press this key to disable open correction.
Press this key to execute 2nd single frequency
open correction.
Press this key to execute 1st single frequency
open correction.
Press this key to execute multiple open
correction.
Press this key to return to the previous
selection.
Press this key to execute 2nd single
frequency open correction.
Press this key to execute 1st single
frequency open correction.
Press this key to enable open correction.
Press this key to disable open correction.
HPOT, LCUR and LPOT as the diagram in the following.
Use the direction control key (,) of Display Unit to switch it to LCR OPEN column.
Press MULTIPLE OPEN to perform open correction and the screen shows “MEASURING…”.
The range for executing correction frequency is 1kHz 10MHz, the status screen is as the
following.
When the screen shows “OPEN CORRECTION COMPLETED.” that indicates open
correction completed. The execution status screen shows as below:
5-29
HF LCR Meter 1105X Series User’s Manual
Press this key to execute multiple open
correction.
Press this key to return to the previous
selection.
Press this key to execute 2nd single frequency
open correction.
Press this key to execute 1st single frequency
open correction.
Press this key to enable open correction.
Press this key to disable open correction.
Press this key to disable open correction.
Press this key to enable open correction.
Press this key to execute multiple open
correction.
Press this key to return to the previous selection.
Press this key to execute 2nd single frequency
open correction.
Press this key to execute 1st single frequency
open correction.
Press this key to execute multiple open
correction.
Press this key to return to the previous
selection.
Press this key to execute 2nd single
frequency open correction.
Press this key to execute 1st single frequency
open correction.
Press this key to disable open correction.
Press this key to enable open correction.
Change LCR OPEN column from “OFF” to “ON”.
The screen shows “OPEN CORRECTION UNDER LIMIT!” as the screen in the following
when open correction is with error. It represents open test action encounterd problem,
please check whether the test cable is broken or test clip bad contact. Execute open test
again after inspecting.
LCR SHORT:
It sets LCR for performing short correction on fixture. It can select SINGLE SHORT or
MULTIPLE SHORT, SINGLE SHORT needs to coordinate with LCR FREQUENCY set
frequency to do correction. Change “OFF” in LCR SHORT column to “ON” after SINGLE
SHORT or MULTIPLE SHORT executed. The default setting is “OFF”.
5-30
Setting for Operation
Press this key to execute multiple short
correction.
Press this key to return to the previous
selection.
Press this key to execute 2nd single frequency
short correction.
Press this key to execute 1st single frequency
short correction.
Press this key to enable short correction.
Press this key to disable short correction.
Press this key to execute multiple short
correction.
Press this key to return to the previous
selection.
Press this key to execute 2nd single frequency
short correction.
Press this key to execute 1st single frequency
short correction.
Press this key to enable short correction.
Press this key to disable short correction.
Take a test fixture to execute 1MHz SINGLE SHORT correction as an example, connect the
fixture to UNKNOWN BNC connector. HCUR, HPOT, LCUR andLPOT will be short-circuited as
the diagram in the following.
Use the direction control key (, ) of Display Unit to switch it to LCR SHORT column.
Press SINGLE SHORT to perform 1MHz short correction and the screen shows
“MEASURING…”. The processing status screen is as the following.
When the screen shows “SHORT CORRECTION COMPLETED.” that indicates short
5-31
HF LCR Meter 1105X Series User’s Manual
1. The measured frequency will affect the selection of test fixture. The stray
capacitance effect of DUT can’t be ignored under the circumstance of
fixture far away the main unit and higher frequency of measurement.
2. It is recommended to perform MULTIPLE OPEN and MULTIPLE SHORT if
test item includes parameter test with multiple frequency.
Press this key to disable short correction.
Press this key to enable short correction.
Press this key to execute multiple short
correction.
Press this key to return to the previous
selection.
Press this key to execute 2nd single short
correction.
Press this key to execute 1st single short
correction.
Press this key to execute multiple short
correction.
Press this key to return to the previous
selection.
Press this key to disable short correction.
Press this key to enable short correction.
Press this key to execute 2nd single short
correction.
Press this key to execute 1st single short
correction.
Press this key to execute multiple short
correction.
Press this key to return to the previous
selection.
Press this key to execute 2nd single short
correction.
Press this key to execute 1st single short
correction.
Press this key to disable short correction.
Press this key to enable short correction.
correction completed. The execution status screen shows as below:
Change LCR SHORT column from “OFF” to “ON”.
The short correction screen shows “SHORT CORRECTION OVER LIMIT!” (as the screen
below) or “SHORT CORRECTION OVER FAIL”. It represents short test action encounterd
problem, please check whether the test cable is broken or test clip bad contact. Execute
short test again after inspecting.
5-32
Setting for Operation
3. Users need to perform open/short correction again after powered on each
time or before replacing/testing measurement fixture.
4. Users can press [RESET] on Display Unit when open/short to be aborted
by requirement. The screen shows “MEASUREMENT ABORT!” as the
figure in the following.
Note
Here, we only introduce environmental operation of LCR MODE whereas other
modes also use the same interface and operation. The sequential section won’t
describe it repeatedly.
Press this key to enter LOAD
Press this key to execute
multiple open correction.
Press this key to exit.
Press this key to enter LCR MODE
Press this key to execute 1st
single open correction.
Press this key to execute 2nd
single open correction.
The usage of LCRZ SCAN MODE, BIAS SCAN MODE and PARAMETER SWEEP MODE
can be done in the same manner with OPEN/SHORT correction of LCR MODE.
“Operation Example”
Before the meter starts measuring (DUT unplaced) in single unit(fixture 0m included)
application, it is need to execute open/short correction to get the optimum measurement
accuracy. Operation for open/short correction is as the following.
(1) Place the test fixture on UNKNOWN four ends BNC terminal of the meter after
powered-on.
(2) Press [MAIN INDEX] on DISPLAY UNIT, it enters LCR MODE.
(3) Enter initial screen in LCR MODE.
5-33
HF LCR Meter 1105X Series User’s Manual
Press this key to select PAGE.
Press this key to execute CORRECTION
function.
2
(4) Change the menu to PAGE interface at the right window and press CORRECTION to
enter CORRECTION-OPEN/SHORT screen.
(5) Open correction: Use the direction control key – down key () to highlight LCR
OPEN column after entered CORRECTION-OPEN/SHORT screen. The right side
of window will show MULTIPLE OPEN. Make sure whether HCUR is short-circuited
with HPOT and LCUR with LPOT.
Next to press MULTIPLE OPEN to start open correction function, the screen shows
“MEASURING…” at this time. The ranges of 11050-5M/11050 correction frequency
are 60Hz 5MHz /1kHz 10MHz and the screen is as the following.
5-34
Setting for Operation
Press this key to execute multiple open
correction.
Press this key to execute multiple short
correction.
For a few minutes later, the screen shows “OPEN CORRECTION COMPLETED.”
thus open correction is completed.
(6) Short correction: Use the direction control key – down key () to hightlight LCR
SHORT column. The right side of window will show MULTIPLE SHORT. Make
sure whether HCUR is short-circuited with HPOT and LCUR with LPOT.
Next to press MULTIPLE SHORT to start short correction function, the screen shows
“MEASURING…” at this time. The ranges of 11050-5M/11050 correction frequency
are 60Hz 5MHz /1kHz 10MHz and the screen is as the following.
5-35
HF LCR Meter 1105X Series User’s Manual
Press this key to execute multiple
short correction.
Press this key to return to the previous
selection.
Press this key to enable correction.
5
The screen shows “SHORT CORRECTION COMPLETED.” thus short correction is
completed.
(7) Use the direction control key – up/down key (,) to highlight LCR OPEN column
after open/short correction is done. It sets the right menu of window to ON and
enables open correction function. In the same way, it sets LCR SHORT to ON and
enables short correction function as well as presses RETURN to exit from the setting.
(8) Press EXIT or [MEAS DISPLAY] to exit from CORRECTION-OPEN/SHORT screen.
5-36
Setting for Operation
Press this key to exit.
Press this key to enter LOAD correction.
Press this key to execute DCR short
correction.
Press this key to exit.
(9) It returns LCR DISPLAY to confirm whether measurement quantity level of fixture
short is small.
(10) Place the DUT and start measuring after executed the above described open/short
correction.
5.3.1.3 LCR MODE - DCR SHORT CORRECTION
This section DCR Short Correction continues the concept of previous section Open/Short
Correction that is aimed at DCR test such as DCR measurement of inductance.
DCR short correction operation: Select CORRECTION at the right Page of LCR Dispaly to
enter CORRECTION-OPEN/SHORT screen.
5-37
HF LCR Meter 1105X Series User’s Manual
1. Users need to do DCR short correction at each time powered-on or
replacing measurement fixture if test item includes DCR test item.
2. Users can press [RESET] on Display Unit when DCR short correction to
be aborted by requirement. The screen shows “MEASUREMENT
ABORT!” as the figure in the following.
Press this key to enter LOAD correction.
Press this key to execute DCR short correction.
Press this key to exit.
Press this key to enter LOAD correction.
Press this key to execute DCR short correction.
Press this key to exit.
DCR SHORT:
It sets LCR for performing DCR correction on fixture. Change “OFF” in DCR SHORT column
to “ON” after DCR Short Correction executed. The default setting is “OFF”.
It selects DCR SHORT in selection key to do correction. The screen shows “DCR SHORT
CORRECTION COMPLETED.” that indicates DCR short correction completed.
Next to change “OFF” in DCR SHORT column to “ON” thus it enables DCR short correction
function.
The screen shows “DCR SHORT CORRECTION FAIL!” as the screen in the following when
DCR short correction is with error. It represents DCR short test action encounterd problem,
please check whether the test cable is broken or test clip bad contact. Execute short test
again after inspecting.
5-38
Setting for Operation
Press this key to disable DCR
short correction.
Press this key to enable DCR
Press this key to execute DCR
short correction.
Press this key to return to the
previous selection.
Press this key to enter OPEN/SHORT
correction .
Press this key to execute single open
correction.
Press this key to execute single open correction.
Press this key to exit.
Press this key to execute load correction.
The usage of LCRZ SCAN MODE, BIAS SCAN MODE and PARAMETER SWEEP MODE
can be done in the same manner with DCR correction of LCR MODE.
5.3.1.4 LCR MODE - OPEN /SHORT/LOAD CORRECTION
LOAD Correction is mainly applied on known reference load (GOLDEN SAMPLE). It is
loaded into the machine as standard value which coordinates with open/short correction to
reduce further stray and residual impedance of external connected fixture/long wire effect on
the accuracy.
The operation method is to select CORRECTION in Page interface of LCR Dispaly and press
CHANGE TO LOAD at the right of CORRECTION-OPEN/SHORT window thus entered
CORRECTION-LOAD screen as the following.
The left side of CORRECTION-LOAD window includes four areas from up to down as the
following.
(1) Parameter Setting Table
5-39
HF LCR Meter 1105X Series User’s Manual
There are ten different test items can be set as the figure above shows. Input
reference load by users and load it into parameter setting table based on known
parameters as standard value.
Parameter functions in the table are introduced as the following.
ON: It enables the load of this column. The default setting is “OFF”.
P1: The primary measurement parameters include five parameters - “L”, “C”, “R”, “Z”
and “Y” for selection. The default setting is “R”.
P2: According to the primary measurement parameter setting, it coordinates with
different secondary measurement parameters. The default setting is “X”.
FREQ.: The frequency ranges of 11050-5M/11050 can set including 60Hz~5MHz and
1kHz~10MHz. The default setting is “1.000M” Hz.
CKT(S/P): It selects DUT equivalent circuit mode. SERIAL or PARALLEL two
modes are for selection. The default setting is SERIAL.
RANGE: It sets measurement range impedance value. The default setting is
“100”.
REF. P1: Correction value of primary measurement parameter
REF. P2: Correction value of secondary measurement parameter
(2) Display Area of Correction Parameter Result
It displays load correction status of primary/secondary parameter until it meets the
value in parameter setting table.
(3) Display Area ofOpen/Short Correction Result
The result after open/short correction executed will be shown in this area.
OPEN P1: OPEN primary measurement parameter correction value, the displayed
value 1/, unit: S.
OPEN P2: OPEN secondary measurement parameter correction value, the
displayed value 1/, unit: S.
SHORT P1: SHORT primary measurement parameter correction value, unit: .
SHORT P2:SHORT secondary measurement parameter correction value, unit: .
(4) Status Display Area
5-40
Setting for Operation
Note
Here, we only introduce environmental operation of LCR MODE whereas other
modes also use the same interface and operation. The sequential section won’t
describe it repeatedly.
Press this key to enter LCR MODE.
Press this key to select PAGE.
The displayed statuses are as the following:
MEASURING…
LOAD OPEN CORRECTION COMPLETED.
LOAD OPEN CORRECTION UNDER LIMIT!
LOAD SHORT CORRECTION COMPLETED.
LOAD SHORT CORRECTION OVER LIMIT!
LOAD CORRECTION COMPLETED.
MEASUREMENT ABORT!
“Operation Example”
Before the meter starts measuring (DUT unplaced) in long wire application, it is need to
execute open/short/load correction (Load ON) to get the optimum measurement accuracy.
The operation method is as the following.
(1) Place the test long wire and its clamp/fixture(DUT not included) on UNKNOWN four
ends BNC terminal after powered-on.
(2) Press [MAIN INDEX] on DISPLAY UNIT, it enters LCR MODE.
(3) Enter initial screen in LCR MODE.
5-41
HF LCR Meter 1105X Series User’s Manual
Press this key to execute CORRECTION
function.
2
Press this key to LOAD correction.
(4) Change the menu to PAGE interface at the right window and press CORRECTION
to enter CORRECTION-OPEN/SHORT screen.
(5) Select CHANGE TO LOAD at the right of window after entered
CORRECTION-OPEN/SHORT screen, thus it enters CORRECTION-LOAD function
screen.
(6) Enter initial screen in CORRECTION-LOAD.
(7) It selects a standard inductance (or GOLDEN SAMPLE of known test value) as the
table below shows. There are three test items corresponding parameters with
different frequencies in the following table. Input the value to the table based on
the parameter such as P1/P2 (measurment parameter), FREQ. (test frequency),
5-42
CKT(S/P) (serial/parallel mode) and RANGE along with primary/secondary
P1
P2
FREQ.
CKT(S/P)
RANGE
REF.P1
REF.P2
1
Ls
Rs
1M
SERIAL
10Ω
296n
3.4
2
Ls
Rs
3M
SERIAL
10Ω
275n
4.5 3 Ls
Rs
10M
SERIAL
30Ω
246n
5.3
Note
The selection of RANGE needs to close to impedance value of DUT, it is for
getting the optimum accuracy. It follows the known parameter to convert DUT
impedance value and selects the required range by Figure 5-1 Range & Impedance (Z) Reference Table.
1030100300
CONST Rsou=25Ω
FREQUENCY=1M~10MHz
Unit:Ω
100Ω
Range
300Ω
Range
30Ω
Range
10Ω
Range
Figure 5-1 Range & Impedance (Z) Reference Table
The impedance(Z) is related to resistance & reactance. Take inductance as an
example, its impedance is related to resistance & reactance(inductive reactance).
Formula of resistance(Rs) and reactance (XL) is as the following for serial model.
The set value is converted into inductive reactance and total impedance. Please
bring the values into the formula below to get the required total impedance.
XL= 2πfL
Z=(Rs2+XL2)
Item
FREQUENCY(Hz)
L(H)
Rs()
XL ()
Z()
1
1M
296n
3.4
1.86
3.88
2
3M
275n
4.5
5.18
6.86
3
10M
246n
5.3
15.46
16.34
measurement parameter known value of reference load (GOLDEN SAMPLE).
Table 5-3 Standard Inductance (or GOLDEN SAMPLE of Known Test Value)
After inputting all values of three test items to the corresponding table, press RETURN
key at the right of screen that all parameters input are completed. The input
completed status is as the figure below.
Setting for Operation
5-43
HF LCR Meter 1105X Series User’s Manual
Table 5-4 Impedance Counting Example of Open/Short/Load Correction Test Item
Get the required total impedance to correspond Figure 5-1 Range & Impedance
(Z) Reference Table to find the corresponding range, for example total impedance
of item 1 is 3.88 which is below of 10 so the range should be set to 10. In
the same way, the range of item 2 also select 10. Total impedance of item 3 is
16.34 which is below of 30 and its measureable range is 10~100 so the range
should be set to 30.
1. For long wire and its clamp/fixture (DUT not included) of open/short
correction, please see the connection in 5.3.1.2 LCR MODE- OPEN /SHORT CORRECTION.
2. If the same long wire and its clamp/fixture (DUT not included) are
used. It is recommended to execute open corrections of all test
items first and then all short corrections. For example, execute open
correction of three test items first and adjust fixture for executing short
correction as well as do load correction for each test item when
2
Press this key to return to the previous
(8) Use direction control key (, , , ) on DISPLAY UNIT to highlight any
setting of item 1. The right option of window shows parameter setting
corresponded table during the cursor moving on the table. Press RETURN to
switch it to the correction setting can be operated.
(9) Before executing load correction, please execute single open correction and single
short correction for long wire and its clamp/fixture (DUT not included) in sequence.
The correction value will be recorded into the screen after open/short correction
ended.
5-44
Setting for Operation
processing single open/short correction.
Note
The measurement of meter at the beginning has some differences with target
known test value. The last offset value displayed on the screen will close to
GOLDEN SAMPLE setting of known test value during executing load offset
procedure. The operating point of DUT is related to offset time, offset results of
various inductance level and operating frequency has some differences. The
second offset can be performed for obtaining better load offste result if the last
offset result has some differences with GOLDEN SAMPLE setting.
Note
Make sure if offset result approximates to the setting when the offset ends.
Press this key to execute single open
correction.
Press this key to execute single short
correction.
1
3
Press this key to execute load correction.
Test Item1
(10) Place standard inductance (or GOLDEN SAMPLE of known test value) into long
wire and its clamp/fixture and next to execute load correction. The screen shows
“LOAD CORRECTION COMPLETED.” that indicates load correction completed.
(11) Two test items follows step (8)~(10) for execution and its’ offset result as below
shows.
Load offset result of test item 2 and 3.
5-45
HF LCR Meter 1105X Series User’s Manual
Test Item2
Test Item3
3
Press this key to exit.
Press this key to open load offset.
(12) Load offset function is enabled after all open/short/load offsets are completed.
Use the direction key of (, , , ) on DISPLAY UNIT to move the cursor to
ON column of test item 1. The right side shows OFF/ON for selection that ON
colum will be changed from OFF to ON after ON selected. The other two test items
also are managed by the same method. Press RETURN key at right screen after it
is completed.
(13) Open/short/load offset execution of three test items are done to remove standard
inductance (or GOLDEN SAMPLE of known test value). Press EXIT or [MEAS
DISPLAY] key to return to LCR MODE screen.
(14) After arranging the previous setting of open/short/load offset, please refer Table 5-5
Open/short/load offset test item 1 setting and correspond the numerical sequence in
it.
5-46
Setting for Operation
Step
Function
Parameter
Setting
Unit
Remark
1
PARAMETER1
Ls
2
PARAMETER2
Rs
3
RANGE MODE
HOLD
4
RANGE
10Ω
5
CKT(S/P)
SERIAL
6
SPEED
SLOW
7
FREQUENCY
1.000M
Hz
8
LEVEL
1.000
V
1 4 7
Press this key to select PAGE or EDIT
interface.
Press this key to enable LOAD function.
Table 5-5 Open/short/load offset test item 1 setting
(15) Next to switch it to EDIT interface in order to set LOAD CORRECT setting to ON.
(16) Place the DUT to start measuring by open/short/load offset test item 1 after the
above execution is completed. To check if it meet to calculation result via Table 5-4 Impedance Counting Example of Open/Short/Load Correction Test Item.
5-47
HF LCR Meter 1105X Series User’s Manual
1. Parameter setting of LCR DISPLAY need to notice that primary/
secondary parameter, test frequency and range settings should be
the same as load offset setting. It can express load offset function,
otherwise the load offset function will be lost.
2. LOAD CORRECT in EDIT interface of LCR DISPLAY screen needs
to be enabled simultaneous with load enable in load offset screen
that load offset function can be worked.
3. Execute open/short offset first before operating load offset.
CAUTION
Test Item1
Test Item3
Test Item2
(17) Test item 2 and 3 use the same method for operation which its measured result as
below shows.
5.3.1.5 LCR MODE – LIMIT TEST Function
LIMIT TEST needs to use with DISPLAY MODE in EDIT interface of LCR DISPLAY that are
applied in JUDGMENT mode. It can calculate total counts of good product/no good product
and DUT.
LIMIT TEST setting screen need to via LIMIT SETTING selection in PAGE interface of LCR
DISPLAY to enter LIMIT Test setting screen as the following.
5-48
Setting for Operation
Note
PARA.1 represents primary parameter and PARA.2 represents secondary
parameter.
Press this key to exit.
Press this key to select upper/lower display
mode.
Press this key to store or recall file.
Press this key to select PAGE or EDIT
interface.
Press this key to select DISPLAY MODE.
Press this key to CLEAR COUNT.
Press this key to enable LOAD function.
Press this key to enable LIMIT function.
Press this key to enable BIN function.
The upper screen reserves parameter setting of LCR DISPLAY and it can’t be used here.
The lower screen can enter STANDARD, HIGH LIMIT and LOW LIMIT for measurement
primary/secondary parameter.
High/low limit value follows LIMIT MODE at lower right of screen to select ABS (absolute
value) or % (percentage) mode. Column HIGH LIMIT and LOW LIMIT enter a value if ABS
mode is selected. Column HIGH LIMIT and LOW LIMIT enter percentage compared nwith
parameter standard value if % mode is selected.
Press EXIT or [MEAS. DISPLAY] key to return to LCR DISPLAY screen after the setting is
completed. It is switched to EDIT interface to select DISPLAY MODE as JUDGMENT mode
and enabled LIMIT MODE by selection key for starting the test. The lower screen displays
LIMIT test, Pass/Fail and Count Result as below.
It selects CLEAR COUNT via selection key and “CLEAR ALL COUNT, ARE YOU SURE?”
message popped up if compare result to be calculated again. Select YES if all counts to be
cleared. Select NO if all counts not to be cleared.
5-49
HF LCR Meter 1105X Series User’s Manual
Press this key to clear count value.
Press this key not to clear count value.
Press this key to select PAGE or EDIT interface.
Press this key to select DISPLAY MODE.
Press this key to CLEAR COUNT.
Press this key to enable LOAD function.
Press this key to enable LIMIT function.
Press this key to enable BIN function.
Press this key to select PAGE or EDIT interface.
Press this key to select DISPLAY MODE.
Press this key to CLEAR COUNT.
Press this key to enable LOAD function.
Press this key to enable LIMIT function.
Press this key to enable BIN function.
LIMIT MODE of LCR DISPLAY includes four modes of “OFF”, “ABS”, “Δ” and “Δ%”. Except
for “OFF” mode belongs to compare test function not to be enabled, other three modes are
described as the following.
ABS mode: It displays the measurement value of primary/secondary parameter at present
and calculates Pass, Fail, DUT total count, etc.
Δ mode: It displays the difference of primary/secondary parameter and nominal standard
value along with calculates Pass, Fail, DUT total count, etc.
Calculation method:
Δ primary parameter = Primary parameter of actual DUT – Nominal standard value of primary
parameter
Δ secondary parameter = Secondary parameter of actual DUT – Nominal standard value of
secondary parameter
5-50
Setting for Operation
100%
parameterprimary of valuestandard Nonimal
parameter primary of valuestandard Nominal - valueDUT actual ofparameter Primary
=parameter%ΔPrimary
%100
parametersecondary of valuestandard Nominal
parametersecondary of valuestandard Nominal - valueDUT actual ofparameter Secondary
=parameter% ΔSecondary
Press this key to select PAGE or EDIT interface.
Press this key to select DISPLAY MODE.
Press this key to CLEAR COUNT.
Press this key to enable LOAD function.
Press this key to enable LIMIT function.
Press this key to enable BIN function.
Press this key to exit.
Press this key to select limit display mode.
Press this key to store or recall file.
Press this key to select INPUT MODE.
Δ% mode: It displays the difference percentage of primary/secondary parameter and nominal
standard value along with calculates Pass, Fail, DUT total count, etc.
Calculation method:
5.3.1.6 LCR MODE - BIN SORTING TEST Function
BIN SORTING TEST needs to use with JUDGMENT mode in EDIT interface of LCR DISPLAY.
Follow the parameter level set to sort DUT and judge if it is Pass or Fail based on DUT’s
primary/secondary parameter as well as display DUT total count. For setting and operation
are as follows.
The upper screen remains parameter setting of LCR DISPLAY and the setting here can’t be
used. The lower screen includes STANDARD (nominal standard value), HIGH LIMIT and
LOW LIMIT of primary/secondary parameter bin sorting. AUX indicates secondary
parameter setting and BIN1~BIN8 indicates various bin sorting setting of primary parameter.
The lower right screen includes INPUT MODE, LIMIT MODE and STORE/RECALL functions
for selection. The functions are described as follows.
5-51
HF LCR Meter 1105X Series User’s Manual
Press this key to select PAGE or EDIT interface.
Press this key to select DISPLAY MODE.
Press this key to CLEAR COUNT.
Press this key to enable LOAD function.
Press this key to enable LIMIT function.
Press this key to enable BIN function.
Press this key to clear Count.
Press this key to clear Count.
INPUT MODE:
It sets input mode selection. It only can set single column if SINGLE setting is selected. It
can edit the entire line for increasing operation speed set by users if ALL setting is selected.
The default setting is “SINGLE”.
LIMIT MODE:
It sets high/low limit mode selection. There are two modes of “ABS” and “%” for selection
which applies on HIGH LIMIT and LOW LIMIT settings. The default setting is “ABS”.
STORE/RECALL:
It stores or recalls the data. For more detail, please see 5.3.1.7 LCR MODE STORE/RECALL Operation.
Press EXIT or [MEAS. DISPLAY] key to return to LCR DISPLAY screen after the setting is
completed. It selects DISPLAY MODE to JUDGMENT mode in EDIT interface and enables
BIN MODE at lower right screen for starting the test. The lower screen displays Bin Sorting
and Calculation Result as the following.
It selects CLEAR COUNT via EDIT interface and “CLEAR ALL COUNT, ARE YOU SURE?”
message popped up if compare result to be calculated again. Select YES if all counts to be
cleared. Select NO if all counts not to be cleared.
BIN MODE of LCR DISPLAY includes four modes of “OFF”, “ABS”, “Δ” and “Δ%”. Except for “OFF” mode belongs to compare test function not to be enabled, other three modes are
described as the following.
ABS mode: It displays the measurement value of primary/secondary parameter at present
5-52
Setting for Operation
Press this key to select PAGE or EDIT interface.
Press this key to select DISPLAY MODE.
Press this key to CLEAR COUNT.
Press this key to enable LOAD function.
Press this key to enable LIMIT function.
Press this key to enable BIN function.
Press this key to select PAGE or EDIT interface.
Press this key to select DISPLAY MODE.
Press this key to CLEAR COUNT.
Press this key to enable LOAD function.
Press this key to enable LIMIT function.
Press this key to enable BIN function.
Press this key to select PAGE or EDIT interface.
Press this key to select DISPLAY MODE.
Press this key to CLEAR COUNT.
Press this key to enable LOAD function.
Press this key to enable LIMIT function.
Press this key to enable BIN function.
and calculate Pass, Fail, DUT total count, etc.
Δ mode: It displays the difference of primary/secondary parameter and nominal standard
value along with calculates Pass, Fail, DUT total count, etc.
Calculation method:
Δ primary parameter = Primary parameter of actual DUT – Nominal standard value of primary
parameter
Δ secondary parameter = Secondary parameter of actual DUT – Nominal standard value of
secondary parameter
Δ% mode: It displays the difference percentage of primary/secondary parameter and nominal
standard value along with calculates Pass, Fail, DUT total count, etc.
5-53
HF LCR Meter 1105X Series User’s Manual
100%
parameterprimary of valuestandard Nonimal
parameter primary of valuestandard Nominal - valueDUT actual ofparameter Primary
=parameter%ΔPrimary
%100
parametersecondary of valuestandard Nominal
parametersecondary of valuestandard Nominal - valueDUT actual ofparameter Secondary
=parameter% ΔSecondary
Note
Select selection key EXIT first for exiting if USB can’t be read. It enters STORE/
RECALL after a few seconds interval that USB can be seen as the figure below
shows.
Press this key to exit.
Press this key to store the file.
Press this key to copy the file.
Press this key to search the file.
Press this key to recall the file.
Press this key to paste the file.
Press this key to delete the file.
Calculation method:
5.3.1.7 LCR MODE - STORE/RECALL Operation
It selects STORE / RECALL selection in PAGE interface of LCR DISPLAY thus enters data
access function screen as the screen below.
There is a location for inputting FILE NAME at the upper of STORE/RECALL screen that can
input the file name to be stored or searched.
There are SEARCH, RECALL, STORE, COPY, PASTE and DELETE functions at the right of
STORE / RECALL screen. The memory includes INTERNAL and USB.
SEARCH: It selects this function for inputting complete file name or keyword to search file.
RECALL: It recalls the file.
STORE: It stores the file. The internal memory can store 10 files and external memory is depended by memory capacity.
COPY: It copies the file.
PASTE: It pastes the file.
DELETE: It deletes the file.
There is also STORE / RECALL function for selection in LIMIT SETTING and BIN SORTING
SETTING.
5-54
Press this key to exit.
Press this key to store the file .
Press this key to copy the file.
Press this key to search the file.
Press this key to recall the file.
Press this key to paste the file.
Press this key to delete the file.
Press this key to enter LIMIT SETTING function.
Press this key to enter CORRECTION function.
Press this key to select PAGE or EDIT interface.
Press this key to enter SETUP function.
Press this key to enter STORE/RECALL function.
5.3.2 LCRZ SCAN MODE Operation
Setting for Operation
5.3.2.1 LCRZ SCAN DISPLAY Description
It enters LCRZ SCAN DISPLAY and the table in the middle of screen displays test frequency
users’ set along with the measurement result voltage corresponded. The lower of screen
displays test result of comparison function and the right of screen displays parameter setting
of test function as the screen below.
Select PAGE interface at right screen and related descriptions as the following.
SETUP:
The basic setup includes parameter setting tio be measured such as measurement parameter
(L, C, R… ), AVERAGE and SPEED…
LIMIT SETTING:
It sets test frequency, test voltage, nominal value of DUT and high/low limit of nominal value. .
CORRECTION:
It corrects open/short circuit. See 5.3.1.2 for operation description.
STORE/RECALL:
It archieves the data. See 5.3.1.7 for operation description.
5-55
HF LCR Meter 1105X Series User’s Manual
Press this key to clear count.
Press this key to select TEST MODE.
Press this key to select PAGE or EDIT interface.
Press this key to enable or disable LOAD
CORRECTION function.
Press this key to select trigger mode.
Press this key to select measurement speed.
Press this key not to clear all counts of DUT.
Press this key to clear all counts.
It selects EDIT interface at right screen as the following.
The functions in EDIT interface except for “CLEAR COUNT” function, other functions of
“LOAD CORR”, “TEST MODE”, “TRIGGER” and “SPEED” will be synchornized with SETUP
under PAGE interface. Users can fast operate it under EDIT interface. Here
supplementary explanation for “CLEAR COUNT” function.
CLEAR COUNT:
It clears count. It selects CLEAR COUNT and “CLEAR ALL COUNT, ARE YOU SURE?”
message popped up if compare result to be calculated again. Select YES if all counts to be
cleared. Select NO if all counts not to be cleared.
5.3.2.2 LCRZ SCAN SETUP
It enters LCRZ SCAN SETUP screen if PAGE interface at right screen is selected and SETUP
function key is pressed as the screen below shows.
5-56
Setting for Operation
Press this key to exit.
LCRZ SCAN SETUP descriptions are as follows.
PARAMETER1:
It selects primary parameter which includes L, C, R, Z and Y for selection. The default
setting is “L”.
PARAMETER2:
It selects secondary parameter which follows the setting of primary parameter to coordinate
with different secondary parameter. The default setting is “Q”.
CKT (S/P):
It selects equivalent circuit (series/parallel) mode which includes “SERIAL” and “PARALLEL”
modes. The default setting is “SERIAL”.
CONST Rsou:
It selects output impedance mode which includes three modes of “OFF”, “100” and “25” for
selection. The default setting is “25”.
AVERAGE:
It sets average time for measurement. By choosing the average setting such as “1”, the
program will take average in each time of measurement. If it is set to “2”, the program will
take average in two times of measurement. Following test can be deduced accordingly and
the setting range is ‘1~10”. The default setting is “1”.
LINE FREQ.:
It selects input AC power mode. To select the correct line frequency will be helpful to the
stability and accuracy of test value. It includes three selections of “50Hz”, “60Hz” and “N/A”; “N/A” indicates not to consider power frequency but test speed increased. The default
setting is “60Hz”.
TRIG. DELAY:
When it is used with auto machinery, external trigger EXT for coordinating with the action of
auto machinery needs to delay the meter trigger. Thus users can adjust TRIGGER DELAY
time, its setting range is “09999”ms. The default setting is “0”ms.
TEST MODE:
It executes test mode which includes “SEQU” or “STEP” measurement for selection. The
default setting is “SEQU”.
SPEED:
5-57
HF LCR Meter 1105X Series User’s Manual
Press this key to exit.
Press this key to select INPUT MODE function.
Press this key to clear DUT setting to empty
status.
Press this key to select LIMIT MODE.
Press this key to select RANGE MODE.
It measures speed mode which includes “FAST”, “MEDIUM” and “SLOW” for selection. The
default setting is “FAST”.
TRIGGER:
It selects trigger mode which includes “INT”, “MAN”, “EXT” and “BUS” for selection. The
default setting is “INT”.
LOAD CORR.:
It is the function of load correction which includes “ON” or “OFF”. The default setting is
“OFF”.
MEAS. DELAY:
It sets the time delayed for executing next test item as the figure below shows. MEAS.
DELAY control range is “0~99999”ms. The default setting is “0” ms.
5.3.2.3 LCRZ SCAN LIMIT SETTING
It enters LCRZ SCAN LIMIT SETTING screen if PAGE interface at right screen is selected
and LIMIT SETTING function key is pressed as the screen below shows.
5-58
Setting for Operation
The LCRZ SCAN DISPLAY can’t display measurement result and
respresented by blank when COMP. is set to to OFF.
CAUTION
Press this key to return to the previous item.
Press this key for adding - fine tuning
Press this key for reducing – fine tuning
Press this key for reducing – coarse tuning
Press this key for adding - coarse tuning
“P1 LOW”, “P2 STD”, “P2 HIGH” and “P2 LOW” parameter settings can’t be displayed on the
same screen due to the restriction of column width. Control the cursor direction control key
of (, ) on DISPLAY UNIT unitl parameter columns of “P1 LOW”, “P2 STD”, “P2 HIGH”
and “P2 LOW” are seen.
The screen is divided into central table and right column that will be explained separately as
the following.
The central table displays ten setting items of “COMP. ”, “Freq.”, “LEVEL”, “RANGE”, “P1
STD”, “P1 HIGH”, “P1 LOW”, “P2 STD”, “P2 HIGH”, “P2 LOW”, etc. The left screen labelled
column 1~9 for users additional setting 9 different test conditions and comparisons.
Descriptions for the detail parameter settings are as the following.
COMP.:
It sets compare mode. Users can select single parameter or two parameters for comparison,
operation modes are: “OFF”, “PARA.1”, “PARA.2” and “P1+P2”. The default setting is “OFF”.
Freq.:
It can set measurement frequency range from 1 kHz to 10MHz. The default setting is
“1.000M”.
LEVEL:
It can select Test voltage range of 10mV~1V. The default setting is “1V”.
RANGE:
It selects impedance value of measurement range. Use selection key to switch RANGE
MODE to “HOLD” mode for selecting measurement range its impedance adjustment is need.
The measurement range can’t be adjusted if RANGE MODE is in “AUTO” mode. The default
setting is “100”.
P1 STD:
The standard value of primary parameter includes five measurement parameters of L, C, R, Z
and Y for selection. This column is for inputting primary standard value of DUT. The default
setting is empty.
P1 HIGH:
This column is for setting the high limit of “P1 STD” standard value. The default setting is
5-59
HF LCR Meter 1105X Series User’s Manual
Note
“CLEAR ALL” function can’t clear the measured setting parameters of “COMP. ”,
“Freq.” and “LEVEL”.
Press this key not to clear all settings of DUT.
Press this key to clear all settings of DUT.
empty.
P1 LOW:
This column is for setting the low limit of “P1 STD” standard value. The default setting is
empty.
P2 STD:
The standard value of secondary parameter follows “P1 STD” parameter to select
corresponded “P2 STD” parameter (PARAMETER2 description in 5.3.2.2 LCRZ SCAN SETUP). The column is for inputting secondary standard value of DUT. The default setting
is empty.
P2 HIGH:
This column is for setting the high limit of “P2 STD” standard value. The default setting is
empty.
P2 LOW:
This column is for setting the low limit of “P2 STD” standard value. The default setting is
empty.
The right screen includes function items of “INPUT MODE”, “CLEAR ALL” and “LIMIT MODE”
for selection. Its’ descriptions are as the following.
INPUT MODE:
It sets input mode selection. It only can set single column if SINGLE setting is selected. It
can edit the entire line for increasing operation speed set by users if ALL setting is selected.
The default setting is “SINGLE”.
CLEAR ALL:
It clears all “P1 STD”, “P1 HIGH”, “P1 LOW”, “P2 STD”, “P2 HIGH” and “P2 LOW” parameter
settings of DUT. “CLEAR ALL COUNT, ARE YOU SURE?” message will be popped up if “CLEAR ALL” function used. Select YES if all settings to be cleared. Select NO if all
settings not to be cleared.
It clears the parameters of “P1 STD”, “P1 HIGH”, “P1 LOW”, “P2 STD”, “P2 HIGH” and “P2 LOW” when executing “CLEAR ALL” function.
LIMIT MODE:
5-60
Setting for Operation
Press this key to enter LCRZ SCAN MODE.
Press this key to select PAGE.
It selects two high/low limit modes of “ABS” and “%” for selection that applies on the settings
of “P1 HIGH”, “P1 LOW”, “P2 HIGH” and “P2 LOW”. The default setting is “ABS”.
“Operation Example”
Here, use a inductance to do measurement for setting multiple sets (9 sets at most) of
different test frequency, voltage level and parameter on DUT processing compare test in long
wire application. Before the meter starts measuring (DUT unplaced), it is need to execute
open/short/load correction (Load ON) to get the optimum measurement accuracy. The
operation method is as the following.
(1) Place the test long wire and its clamp/fixture(DUT not included) on UNKNOWN four ends
BNC terminal after powered-on.
(2) Press [MAIN INDEX] on DISPLAY UNIT, it enters LCRZ SCAN MODE.
(3) It enters initial screen in LCRZ SCAN MODE.
(4) It selects a standard inductance (or GOLDEN SAMPLE of known test value), its
parameter value can refer Table 5-3 Standard Inductance (or GOLDEN SAMPLE of Known Test Value). Change the menu to PAGE interface at the right window and press
CORRECTION to start processing load correction.
5-61
HF LCR Meter 1105X Series User’s Manual
Note
(1) For detail operation, refer the setting in 5.3.1.4 LCR MODE - OPEN
/SHORT/LOAD CORRECTION.
(2) This example continue to use the operation example in 5.3.1.4 LCR MODE -
OPEN /SHORT/LOAD CORRECTION.
The parameter used as Table 5-3 Standard Inductance (or GOLDEN SAMPLE of
Known Test Value) shows.
P1
P2
FREQ.
CKT(S/P)
RANGE
REF.P1
REF.P2
1
Ls
Rs
1M
SERIAL
10Ω
296n
3.4 2 Ls
Rs
3M
SERIAL
10Ω
275n
4.5
3
Ls
Rs
10M
SERIAL
30Ω
246n
5.3
FREQ
LEVEL
RANGE
Ls
Ls Error
Rs
Rs Error
1
1.000M
1.000
10Ω
296n
±1%
3.4
±1%
2
3.000M
1.000
10Ω
275n
±1%
4.5
±1%
3
10.00M
1.000
30Ω
246n
±1%
5.3
±1%
Press this key to perform CORRECTION
function.
Press this key to perform CORRECTION
function.
1
2
(5) Open/short/load correction are done to remove standard inductance (or GOLDEN
SAMPLE of known test value). Press EXIT or [MEAS DISPLAY] key to return to LCRZ
SCAN DISPLAY screen. Inductance parameter specifications of DUT are as the
following.
Table 5-6 DUT – parameter specification of inductance
(6) It enters LCRZ SCAN SETUP screen if PAGE interface is selected and SETUP function
key is pressed as the screen below shows.