CHINO LE5000 Specifications

LE5000 SERIES
250MM CHART HYBRID RECORDER
MODEL LE5100
LE5000 series are 250mm hybrid recorders
with multi-range input. Innovative design high
performance recorder provides high accuracy, ±0.05%; high speed scanning, 0.1second for
36 points and high speed recording, 1 line in 3
seconds. Simple operational keys and PC setting functions drastically improved usability
of recording system.
FEA TURES
y High speed scanning at 36 points/sec and
high-speed recording
Rapid changes of process data such as lab test results can be scanned simultaneously at 36 points/sec and recorded at about 1 line/ 3 sec. Data for each channel is displayed in 10 different colors which is user selectable.
y High accuracy of 0.05%
The accuracy is ±0.05% and the resolution is 1µV or
0.1ºC
y Various industrial values can be measured at the
same time with selectable ranges
With 35 thermocouple ranges and 8 DC voltage ranges, a total of 43 input ranges are provided which enables universal input and optional mixed input: current inputs are also possible,
y Superior ease of operation
Operation keys are functionally designed for ease of use.
y Engineering port is provided (USB)
A personal computer can be used as an engineering tool and parameter setting and data collecting is available.
y Anti-noise countermeasures
High effective anti-noise countermeasures are taken; suppressive induced noise by 130 dB or more in the common mode while 50dB or more is achieved in the series mode. Effective countermeasures are taken against impulse noise.
y Communication interfaces are available (Option)
RS422A, RS485 and Ethernet can be provided to meet various customers’ needs.
y Recording and calculation of data communication
input (Option)
Data input by communications from a host can be recorded as analog and digital values at the same time with measuring data. Mathematical process of the data communications input from a host can be processed in parallel.
1
MODELS
LE51-N
Input points
1 : 12 points
2 : 24 points
3 : 36 points
Alarm output points (Option)
0: None
1: 12 points
2: 24 points
3: 36 points
Communication interface/ contact output (Option)
N: None (Standard)
1: RS422A/ RS485
Ethernet +1a contact output (Mechanical relay)
External drive / Chart speed change ( Option)
N: None (Standard)
1: Provided
PSE-352A
NAMES AND FUNCTIONS OF EACH PART
Display
Operation keys
z DISPLAY
Three types of displays are available according to user’s demand.
Chart speed and time clock are always displayed on an upper part of
screen and an operational instruction of a setting key is displayed on a
lower part of screen.
Engineering port
zDisplay of 1 channel
1 channel of consecutive or sequential display is available.
zSimultaneous display of 12 channels
12 channels of consecutive or sequential display are available.
z Simultaneous display of 36 channels
36 channels of consecutive display is available. 24 channels display
is also available for 24 points input. (In the case of 24 channels, the
part of CH 25 to 36 is blank)
2
z Operation key
The operation keys are functionally laid out.
Names of keys Functions
Enter key
Escape key
Menu key
Up/ Down and
Left/Right key
Function 1 key
Function 2 key
Recording key
Data print key
Feed key
Shift key
Numeric key
Alphabetic key
Symbol key
Used to set each function.
Each time this key is pressed, it returns to previous page.
Used to display settings for each function.
Used to move a cursor up/ down and left/ right, and also to chose setting items and value.
Used to set and change setting for each function. Data is indicated in a lower part of screen.
Used to set and change setting for each function. Data is indicated in a lower part of screen.
Each time this key is pressed, recording is switched ON or OFF. Used with Enter key.
When this key is pressed, data is simultaneously printed. Used with Enter key.
While this key is pressed, chart paper is fed with a speed of 750mm/min.
Used to switch number key, alphabetic key and other symbol keys.
Used to input numeric value. ( used together with Shift key)
Used to input alphabet. (used together with Shift key)
Used to input symbols. (used together with shift key)
z Engineering port
Engineering port allows parameter setting, setting confirmation and measuring data transmission in connection with PC.
Engineering port
3
RECORDING FORMAT
z Digital recording
Format 1
In the left margin of the chart, the tag number and measuring data are digitally recorded at a specified interval.
z Digital recording
Format 2
The tag number, measuring data and unit are digitally recorded 6 channels/ line at a specified interval superimposed on the analog recording.
z Digital recording
Format 3
The tag number and measuring data are digitally recorded 10 channels/line at a specified interval superimposed on the analog recording.
4
z Data print
Format 1
When Data print key is pressed, analog recording is interrupted and the latest data is printed digitally 6 channels/ line.
z Data print
Format 2
When Data print key is pressed, analog recording is interrupted and the latest data is printed digitally 10 channels/ line.
z Logging recording
Format 1
The tag number, data and unit are recorded digitally at a specified interval 6 channels/ line. Analog recording is not performed.
z Logging recording
Format 2
The tag number, data and unit are recorded digitally at a specified interval 10 channels/ line. Analog recording is not performed.
5
INPUT SIGNALS
Measuring points : 12, 24 and 36 points Input : Multi-channel data range
Range setting: Input type and range are set with front keys Scale setting: The minimum and maximum values and unit are
Indication accuracy:Refer to items of measuring ranges, accuracy
Temperature drift: 0.1% FS/ 10°C Measuring period: 0.1 sec/ channel Reference junction compensation accuracy:
Input resolution: Approx. 1/40000 (Standard range conversion) Burnout: Select with/ without burnout for each input Allowable signal source resistance:
Input resistance: Thermocouple input, DC voltage input --- approx.1M Maximum input applied voltage: ±20V DC Input correction: Zero/span correction and shift correction for each
Maximum common mode voltage:
Common mode rejection ratio: 130dB Series mode rejection ratio:
Terminal board: Detachable type, removable for wire connection
DC voltage --- ±10mV, ±20mV, ±40mV, ±80mV, ±1.25V, ±2.5 V, ±5 V, ±10V DC current --- Shunt resistor (100, 250) needs to be mounted externally Thermocouple --- B, R, S, K, E, J, T, N, PtRh40-PtRh20, NiMo-Ni,WRe5 – WRe26, W-WRe26, Platinel II, U, L Resistance thermometer --- Pt 100, JPt 100
set for each point with front keys Setting range -30000 to 30000 Decimal points Optional setting
rating and display resolutions
K, E, J, T, N, Platinel II --- ±0.5°C or less (0°C or more when measuring)
R, S, WRe5-WRe26, NiMo-Ni, U, L --- ±1.0°C or
less (Only when the ambient temperature is 23°C±5°C)
Thermocouple inputs, DC voltage input (10mV) --- 500 or less (without burnout) DC Voltage input (except 10mV) --- 100or less Resistance thermometer inputs --- 10or less/ line Three lines are common, Pt100, JPt100
channel
30V AC (support LVD) *250V AC at evaluation
50dB (Only when the peak value of noise is below standard range.)
RECORDING SPECIFICATIONS
Recording system : Raster scan system, 10-color wire dot printing Recording and recording color :
Chart paper: Fan-fold type, Overall width 318 mm, total length 20m; Effective
Chart speed: 1 to 1500mm/H (in 1mm/H steps) Skip function: Analog recording, digital recording and digital
Recording compensation: Independent setting of zero spans are available.
Analog recording --- color can be specified for each channel as required. 10 colors (red, purple-red, orange, brown, green, yellow-green, blue-green, purple, purple-blue, black) Digital recording and logging recording - Black Message printing --- Black List printing --- Black
recording width 250mm (analog recording)
display can be set independently from recording slip.
DISPLAY SPECIFICATIONS
Digital display: Color LCD panel RGB (640 x 240 dot)
Setting display: Common to digital display Display contents: Digital display
Status display: RECORD ON ( lights during recording) LED
Display size W149.8 x H57.4 mm
Channel display ( One-point/ multiple points continuous/sequential indication change) Display measuring value of each channel ( One-point/ multiple points continuous/sequential indication change) Clock display (Hour/Minute/Second/Tag/Unit) Chart speed display
KEY LOCK ( lights during key lock) ALARM ( lights during alarm activated) LED CHART END ( lights just before record ending) FAIL ( lights during unit abnormal time) * Sharing LED and setting display
ALARM SPECIFICATIONS
Alarm display : Occurrence CH No., data is displayed in red when
Alarm types: High limit, low limit Alarm setting method: Individual setting for each point four levels/ channels Alarm output: See option specification ( Option)
alarm occurs
SETTING AND OPERATIONAL SPECIFICATIONS
Key types, operation:
Func1 --- Switching each function
Func2 --- Switching each function
Enter --- Setting a change of parameter for each mode
Menu --- Specifying each setting function
Esc --- Used to escape in the middle of setting
--- Used to switch channels when specifying the
parameter on cursor
--- Used to switch channels when specifying the
parameter on cursor
--- Used to move cursor to the right
--- Used to move cursor to the left
Rec --- Analog recording, digital recording, printing,
switching chart ON/OFF
DataP --- Digital recording of latest data
Feed --- Fast-forwarding chart paper
Shift --- Specifying key
. _ = --- Setting characters of “. _ =”
@ + - --- Setting characters of “@ + -”
0 * / --- Setting parameter value 0 and character of “* / ”
1ABC --- Setting parameter value 1 and character of “ABC ”
2DEF --- Setting parameter value 2 and character of “DEF”
3GHI --- Setting parameter value 3 and character of “GHI”
4JKL --- Setting parameter value 4 and character of “JKL”
5MNO --- Setting parameter value 5 and character of “MNO”
6PQR --- Setting parameter value 6 and character of “PQR”
7STU --- Setting parameter value 7 and character of “STU”
8VWX --- Setting parameter value 8 and character of “VWX”
9YZ --- Setting parameter value 9 and character of “YZ”
Recording operation:
Setting contents:
Engineering port (USB) :
RECORD ON/OFF --- recording operation ON/OFF* DATA PRINT --- printing measuring data* FEED --- Fast-forwarding chart paper * Two actions are taken to operate
Parameter setting --- Clock time, chart speed, digital recording at set time range, scale, unit, tag, alarm, ºC, pass word (for option communication and recording format, message printing, calculation)
Setting of a whole parameter is available using engineering software (PASS) from PC
6
GENERAL SPECIFICATIONS
Rated power voltage: 100 to 240V AC (universal power supply) 50/60Hz Maximum power consumption:
Reference operating condition:
Power voltage: 90 to 264V
Power frequency: 50/60Hz ±2%
Warm-up time: 1 hour or longer Normal operating condition:
Power voltage: 90 to 264V Power frequency: 50/60Hz ±2%
Attitude: Forward/ Backward/ Left/ Right within 3°
Transportation condition:
Vibration: 10 to 60 Hz, 4.9m/ S Impact: 392m/S Storage condition: Ambient temperature
-20 to 60°C, 5 to 90%RH
Working condition: Working temperature range 0 to 40°C
Power failure protection:
Insulation resistance: Between primary terminals and protective
Dielectric strength: Between primary terminals and protective
Case assembly material:
Color: Door(frame) --- White (Equivalent to DIC546 1/2), Front panel --- Transparent, Back case --- White (Equivalent to DIC546 1/2) Mounting: Panel mounting Weight: About 15kg (Full option) Dimensions, panel cut: W400 x H260 x D300 mm (Dimensions)
Terminal screws: Measuring input, alarm terminals --- M3.5
Chart paper illumination : White LED
STANDARDS
CE marking: Conformity pending
100V A
Ambient temperature/ humidity range: 21 to 25°C, 45 to 65%RH
Attitude: Forward/ Backward/ Left/ Right within 3°
Ambient temperature/humidity range 0 to 40°C, 20 to 80% RH
At the packed condition on shipment from our factory Ambient temperature/ humidity range:
-20 to 60°C, 5 to 90%RH (No dew condensation)
2
(Approx. 40G or less)
(No dew condensation)
Working humidity range 20 to 80%RH
Programmed parameters stored into EEPROM memory Clock circuit sustained for 5 years or longer by a lithium battery (at the operation of 8 hours or longer per day)
conductor terminals --- 20M or more at 500V DC Between secondary terminals and protective conductor terminals --- 20M or more at 500V DC Between primary terminals and secondary terminals --- 20M or more at 500V DC
conductor terminals --- 1 minute at 1500V AC Between secondary terminals and protective conductor terminals --- 1 minute at 500V AC Between primary terminals and secondary terminals --- 1 minute at 1500V AC Note 1: Primary terminals: power terminal, alarm output terminal, output relay terminal Secondary terminals: measuring input terminal, communication terminal, external drive terminal Note 2: When testing insulation resistance and dielectric strength, please short-circuit every terminals of primary and secondary terminals before the test. Test without short-circuiting terminals can damage instruments.
Door (frame) --- ABS resin, Front panel --- Soda glass, Back case --- Normal steel
388 x 248mm (Panel cut)
Power, protective conductor terminal, external drive terminal, communication terminal --- M4
2
(0.5G or less)
OPTION SPECIFICATIONS
Options Contents
External drive
Alarm output
Communication interface
Chart end output
FAIL output
Receiving resistance for current input
zCommunication interface specification
Ethernet
RS-422A RS-485
Specification
Function
Specification
Function
Specification
USB
Function
Chart 3-speed, chart stop, data printing, list printing, message printing 5 types, operation recording Mechanical relay ---- 12, 24, 36 points output, max contact capacity of 100 to 240V AC, 3A resistance load RS422A or RS485 + Ethernet + 1a contact output (1a contact output is contact output of mecha relay
CHART END relay output when chart paper ended (communication interface is required)
FAIL relay output when abnormality ( communication interface is required) 250( for 20mA) or 100(for 50mA) are externally mounted to measure current
With communication interface
Ethernet10BASE-T/ 100BASE-T, automated recognition, TCP, IP, HTTP, exclusive protocol Data display, parameter setting, with browser Data display, parameter setting on exclusive application RS422A, RS485, Communication protocol: MODBUS Communication specification: 9600 bps to 19200 bps 7E1 to 8N2 Data display and parameter setting using exclusive application Inside of front door, USB1.1, Full speed 12Mbps, Bulk transfer, Control transfer
Parameter setting for exclusive application
Without communication interface
7
MEASURING RANGE, ACCURCY RATING, AND DISPLAY RESOLUTION
Input type Measuring range
DC voltage
T/C
W-Wre26
Wre5-Wre26
PtRh40-PtRh20
NiMo-Ni -50 to 1310 ºC
Platinel
RTD
JPt100
Note 1: Ambient temperature/ humidit y range: 23℃±2 Note 2: For thermocouple input, the accurac y of reference junction compensation is not included with the accuracy ratings. Note 3: Accuracy rating is the percentag e of measuring range K,E,J,T,R,S,B,N : IEC584,JIS C 1602-199 5 W-Wre26,Wre5-WRs26,PtRh40-PtRh20,N iMo-Ni, Platinel: ASTM Vol.14.03 U(Cu-CuNi),L(Fe-CuNi) : DIN43710 Pt100 : IEC751,JIS C 1604-1997 JPt100 : JIS C 1604-1981, JIS C 1606-1986
zExceptions of accuracy ratings
Note : Refer to T/C input accuracy is calcu lated based on standard range.
Input types Measuring range Accuracy ratings
K,E,J,T,L -200 to 0ºC
R,S 0 to 400ºC
B
U -200 to 0ºC
W-WRe26 0 to 300ºC
PtRh40-PtRh20
NiMo-Ni -50 to 100ºC ±0.2%+1digit
-10.0 to 10.0mV ±10mV
-20.0 to 20.0mV
-40.0 to 40.0mV
-80.0 to 80.0mV
-1.25 to 1.25V
-2.5 to 2.5V
-5.0 to 5.0V
-10.0 to 10.0V
-200 to 500ºC ±20mV
K
-200 to 900ºC
-200 to 1370ºC
-200 to 250ºC ±20mV
E
-200 to 500ºC
-200 to 900ºC
-200 to 350ºC ±20mV
J
-200 to 700ºC
-200 to 1200ºC
T -200 to 400ºC ±20mV ±0.05%+0.7ºC
R 0 to 1760ºC
S 0 to 1760ºC
B 0 to 1820ºC
N
U
L
Pt100
0 to 600ºC
0 to 1000ºC
0 to 1300ºC ±80mV
0 to 2315ºC
0 to 2315ºC
0 to 1888ºC ±20mV
0 to 500ºC ±20mV ±0.1%+0.1ºC
0 to 950ºC
0 to 1395ºC
-200 to 350ºC ±20mV
-200 to 600ºC
-200 to 350ºC
-200 to 700ºC
-200 to 900ºC
-50 to 50ºC
-100 to 130ºC
-200 to 250ºC
-200 to 550ºC
-50 to 50ºC
-100 to 130ºC
-200 to 250ºC
-200 to 550ºC
0 to 400ºC None
400 to 800ºC ±0.15%+1digit
0 to 300ºC ±1.5%+1digit
300 to 800ºC ±0.8%+1digit
Standard
range
±20mV
±40mV
±80mV
±1.25V
±2.5V
±5V
±10V
±40mV
±80mV
±40mV
±80mV
±40mV
±80mV
±20mV ±20mV ±20mV ±20mV
±40mV
±80mV
±80mV
±80mV
±40mV
±80mV
±40mV
±20mV
±40mV
±80mV
50
100 200
300
50
100
200
300
Accuracy
rating
±0.05%+1digit
±0.05%+0.5ºC
±0.05%+1ºC
±0.05%+0.7ºC
±0.05%+1ºC
±0.05%+0.7ºC
±0.05%+1ºC
±0.05%+1ºC
±0.1%+0.1ºC
±0.1%+1ºC
±0.1%+1ºC
±0.05%+1ºC
±0.05%+0.3ºC 0.1ºC
Display resolution
±0.2%+1digit
±0.3%+1digit
V
10µV
100µV
1mV
0.1ºC
Specifications subject to change without notice. Printed in Japan (I) 2006. 8 Recycled Paper
TERMINAL BOARD
DIMENSIONS
zPanel cut-out and mounting minimum clearance
Unit: mm
32-8, KUMANO-CHO, ITABASHI-KU, TOKYO 173-8632
PHONE: +81-3-3956-2171
FAX: +81-3-3956-0915
E-mail: inter@chino.co.jp
Website: http://www.chino.co.jp